K Number
K210160

Validate with FDA (Live)

Device Name
AlertWatch:AC
Manufacturer
Date Cleared
2021-09-10

(232 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AlertWatch:AC is intended for use by physicians for secondary monitoring of ICU patients. AlertWatch:AC is also intended for use by physicians providing supplemental remote support to bedside care teams in the management and care of ICU patients. AlertWatch:AC is not intended for use in monitoring pediatric or neonatal patients. AlertWatch:AC is a software system that combines data from the electronic medical record, networked physiologic monitors, and ancillary systems, and displays them on a dashboard view of the unit and patient. The clinical decision support is generated to aid in understanding the patient's current condition and changes over time. Once alerted by AlertWatch:AC, the physician must refer to the primary monitor, device or data source before making a clinical decision.

Device Description

AlertWatch:AC is a secondary monitoring system used by physicians to monitor adult patients in an ICU environment. The purpose of the device is to synthesize a wide range of patient data and inform physicians of potential problems. Once alerted, the physician is instructed to refer to the primary monitoring device or EMR before making a clinical decision. The software design includes a default set of rules and alerts that can be configured by the hospital during the installation process. AlertWatch:AC is intended to supplement, not replace, a hospital's primary EMR. The device retrieves data from the electronic medical record (EMR) system and networked physiologic monitors, integrates this data, and performs a series of calculations to assess potential clinical issues. The information is conveyed both via organ colors and messages in the alert panel. Any alert can also be configured to send pages to physicians assigned to the patient.

AI/ML Overview

The provided text describes the 510(k) clearance for AlertWatch:AC, a secondary monitoring system for ICU patients. However, it does not contain information about acceptance criteria or a specific study that proves the device meets such criteria in terms of the accuracy or performance of its clinical decision support algorithms.

The document focuses on regulatory compliance, outlining the device's intended use, technological comparison to a predicate device, and various verification and validation activities (software V&V, human factors study, default limits review, and wireless co-existence testing).

Therefore, I cannot provide the requested information regarding a table of acceptance criteria and reported device performance using figures like sensitivity, specificity, or AUC, nor can I detail the sample size, ground truth establishment, or expert qualifications for such a study, because this information is not present in the provided text.

Based on the document, here's what can be inferred or explicitly stated about the device's validation:

  1. A table of acceptance criteria and the reported device performance: Not available in the provided text. The document refers to "software verification and validation testing" and "performance testing" but does not provide specific quantitative acceptance criteria or results for the clinical decision support functionality (e.g., accuracy of alerts). It states that "the results of performance testing demonstrate that the subject device performs in accordance with specifications and meets user needs and intended uses," but no specifics are given.

  2. Sample sizes used for the test set and the data provenance:

    • Software Verification and Validation Testing: Performed with "both constructed data and data from the EMR." No specific sample size for the "data from the EMR" portion is provided.
    • Human Factors Study:
      • Summative usability study: 18 users.
      • Summative usability study on verbal alarm signals: 15 users.
    • Data Provenance: Not explicitly stated beyond "data from the EMR." No geographical origin (e.g., country) or retrospective/prospective nature is specified.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Expert Committee for Default Limits and Thresholds: "Acute care physicians at the University of Michigan Health System." The exact number of physicians is not given, but it implies multiple experts. Their specific qualifications (e.g., years of experience, board certifications) are not detailed beyond "acute care physicians."
  4. Adjudication method for the test set: Not explicitly mentioned for any testing related to the clinical decision support's accuracy. For the "Expert Committee" review of default limits, it states clinicians "reviewed the limits, provided feedback, and reviewed the final results," implying a consensus-based approach without detailing a specific adjudication method like 2+1 or 3+1.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, the document explicitly states "Clinical Data: Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." This indicates that no MRMC study comparing human readers with and without AI assistance was performed or presented.

  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The document mentions "Software verification and validation testing... ensure that the product works as designed, and was tested with both constructed data and data from the EMR." While this implies algorithm-only testing as part of V&V, it doesn't present it as a separate performance study with metrics suitable for standalone performance (e.g., sensitivity, specificity for specific conditions detected by the algorithm). The device is positioned as clinical decision support where the physician "must refer to the primary monitor, device or data source before making a clinical decision," suggesting it's not intended for standalone diagnostic use.

  7. The type of ground truth used:

    • For the "Default Limits and Thresholds": Ground truth was established by "Review of References" (published studies) and "Expert Committee" (consensus/feedback from acute care physicians). This suggests a form of expert consensus and literature-based validation for the rule-based alerts.
    • For "Software Verification and Validation Testing" using EMR data: The method for establishing ground truth for this EMR data is not described.
  8. The sample size for the training set: Not applicable. The AlertWatch:AC is described as a "software system that combines data... and performs a series of calculations to assess potential clinical issues." It uses "a default set of rules and alerts" and "established patient risk and acuity calculations (SOFA and SIRS)." This indicates a rule-based or calculational system rather than a machine learning model that would typically have a "training set." Therefore, no training set size is mentioned.

  9. How the ground truth for the training set was established: Not applicable, as it's not a machine learning model with a distinct training set. The "default limits and thresholds" and "established patient risk and acuity calculations" are based on literature review and expert consensus rather than labelled training data for an AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 10, 2021

AlertWatch, Inc. % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 1555 King Street. Suite 300 Alexandria, Virginia 22314

Re: K210160

Trade/Device Name: AlertWatch:AC Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MSX Dated: August 11, 2021 Received: August 12, 2021

Dear Donna-Bea Tillman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Jennifer Shih External Heart Rhythm and Rate Team Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210160

Device Name AlertWatch:AC

Indications for Use (Describe)

AlertWatch:AC is intended for use by physicians for secondary monitoring of ICU patients. AlertWatch:AC is also intended for use by physicians providing supplemental remote support to bedside care teams in the management and care of ICU patients. AlertWatch:AC is not intended for use in monitoring pediatric or neonatal patients. AlertWatch:AC is a software system that combines data from the electronic medical record, networked physiologic monitors, and ancillary systems, and displays them on a dashboard view of the unit and patient. The clinical decision support is generated to aid in understanding the patient's current condition and changes over time. Once alerted by AlertWatch:AC, the physician must refer to the primary monitor, device or data source before making a clinical decision.

Type of Use (Select one or both , as applicable)
----------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for AlertWatch:AC is provided below.

SUBMITTER 1.

Applicant:AlertWatch, Inc.330 East Liberty St.Fourth FloorAnn Arbor, MI 48104
Contact:Justin AdamsAlertWatch, Inc.CEOPhone: 734-998-8344Justin.adams@alertwatch.com
Submission Correspondent:Donna-Bea Tillman, Ph. DSenior ConsultantBiologics Consulting Group, Inc.1555 King Street, Suite 300Alexandria, VA 22314(410) 531-6542dtillman@biologicsconsulting.com
Date Prepared:September 10, 2021

2. DEVICE

Device Trade Name:AlertWatch:AC
Device Common Name:System, Network and Communication, PhysiologicalMonitors
Classification Name21 CFR 870.2300 - Cardiac monitor (includingcardiotachometer and rate alarm)
Regulatory Class:II
Product Code:MSX

PREDICATE DEVICE 3.

Predicate Device: K171322 - Philips eCareManager 4.1 Reference Device: K153335 - AlertWatch:OR

{4}------------------------------------------------

DEVICE DESCRIPTION 4.

AlertWatch:AC is a secondary monitoring system used by physicians to monitor adult patients in an ICU environment. The purpose of the device is to synthesize a wide range of patient data and inform physicians of potential problems. Once alerted, the physician is instructed to refer to the primary monitoring device or EMR before making a clinical decision.

The software design includes a default set of rules and alerts that can be configured by the hospital during the installation process. AlertWatch:AC is intended to supplement, not replace, a hospital's primary EMR.

The device retrieves data from the electronic medical record (EMR) system and networked physiologic monitors, integrates this data, and performs a series of calculations to assess potential clinical issues. The information is conveyed both via organ colors and messages in the alert panel. Any alert can also be configured to send pages to physicians assigned to the patient.

5. INTENDED USE/INDICATIONS FOR USE

AlertWatch:AC is intended for use by physicians for secondary monitoring of ICU patients. AlertWatch:AC is also intended for use by physicians providing supplemental remote support to bedside care teams in the management and care of ICU patients. AlertWatch:AC is not intended for use in monitoring pediatric or neonatal patients.

AlertWatch:AC is a software system that combines data from the electronic medical record, networked physiologic monitors, and ancillary systems, and displays them on a dashboard view of the unit and patient. The clinical decision support is generated to aid in understanding the patient's current condition and changes over time. Once alertWatch:AC, the physician must refer to the primary monitor, device, or data source before making a clinical decision.

6. SUBSTANTIAL EQUIVALENCE

Comparison of Indications

Both the subject AlertWatch:AC and the predicate eCareManager device are used by physicians in a hospital environment to monitor patients. Both devices are only intended for secondary monitoring and users must refer to the primary monitor or primary clinician before making clinical decisions. Both systems aggregate and summarize information to help users make informed decisions. Therefore, the differences in indications for use do not constitute a new intended use, and eCareManager can be used as a primary predicate for the subject AlertWatch:AC.

Technological Comparisons

The table below compares the key technological feature of the subject devices to the predicate device (Philips eCareManager 4.1, K171322).

{5}------------------------------------------------

Predicate DeviceProposed Device
Device NameeCareManager 4.1AlertWatch:AC v1.00
510(k) numberK171322K210160
ClassificationMSXIdentical
Regulation870.2300Identical
Intended Use environmentIn-Hospital PatientsICU
Intended UsersTrained Medical StaffPhysicians
Intended for primarymonitoring?No - only for secondary monitoring.User must refer to primary monitoror bedside clinician to confirm alldecisions.Identical
Data serverElectronic medical record, patientmonitoring systems and ancillarysystems connected through networksIdentical
Supported EMR systemsEpicAll current versions of Epic (2015+)and Centricity 7.63.299The reference device(AlertWatch:OR) also supportscurrent versions of Epic (2015+) andCentricity 7.63.299
Supported physiologicmonitorsPhilipsGE Solar 9500,GE Carespan B850,Capsule Medical DeviceIntegration System
Organs MonitoredHeart, Lungs, Kidneys, Brain, LiverIdentical
Supported EMR DataElementsDemographic, Diagnosis, Labs, RiskFactors & Comorbidities, Lines,Tubes and Drains, Medications,Infusions, Notes, Images.Demographic, Diagnosis, Labs, RiskFactors & Comorbidities, Lines,Tubes and Drains, Medications,Infusions, Notes.
Patient Risk Scores &Acuity CalculationsProprietary Philips Acuity Score,APACHESOFA and SIRS, established patientrisk and acuity calculations.
Hardware platformssupportedPCPC, iPad, iPhoneThe hardware platforms ofAlertWatch:OR, the referencedevice, also includes the use of iOSmobile devices.
Predicate DeviceProposed Device
Generates clinicaladvisorieseCareManager uses the above data togenerate clinical decision supportnotifications that aid inunderstanding the patient's currentcondition and changes over time.Identical
Re-display of vital sign datafrom primary monitorsYesIdentical
Intended to replace primarymonitorsNoIdentical
PagingUnknownYes. For physicians to page eachother and for AlertWatch:AC totransmit alerts to the bedsideproviders using the hospital-established paging system.
ViewsSingle PatientSingle Patient
Organ SpecificOrgan Specific
CensusCensus
CameraCamera hardware and software for inroom video.None. A camera is not a requiredelement for the device to perform asintended.

Table 1: Technological Comparison

{6}------------------------------------------------

There are slight differences in technological characteristics that reflect the different clinical needs of the user populations, but the general approach of retrieving data from the EMR system, integrating this data, and performing a series of calculations to assess potential patient clinical issues is the same. The differences in technological characteristics do not raise different questions of safety and effectiveness.

7. PERFORMANCE DATA

Biocompatibility Testing

There are no direct or indirect patient-contacting components of the subject device. Therefore, patient contact information is not needed for this device.

Electrical safety and electromagnetic compatibility (EMC)

Not applicable. The subject device is a software-only device. It contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.

However, because AlertWatch:AC includes wireless communication, the system was evaluated in accordance with the FDA guidance document: Radio Frequency Wireless Technology in Medical Devices: Guidance for Industry and Food and Drug Administration Staff (August 14, 2013). Wireless Co-existence testing was performed to establish that the wireless components work effectively in the hospital environment.

{7}------------------------------------------------

Software Verification and Validation Testing

Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification of AlertWatch:AC was conducted to ensure that the product works as designed, and was tested with both constructed data and data from the EMR. Validation was conducted to check the design and performance of the product.

Bench Testing

  • . Human Factors Study: To better understand any potential usability issues that could interfere with the intended use of the product. AlertWatch performed a comprehensive human factors study. AlertWatch recruited 18 users and conducted a summative usability study. The results of the study showed that users with no training were able to successfully perform critical tasks and were able to use the device for its intended purpose- to clarify clinical information and support information access. AlertWatch also recruited 15 users to participate in a summative usability study on the verbal alarm signals. The results of the study showed that users with no training were able to successfully perform critical tasks and were able to use the device for its intended purpose.
  • Default Limits and Thresholds: AlertWatch, Inc. used a two-phased approach to ensure ● that the default limits were clinically valid:
      1. Review of References. AlertWatch sought out definitive published studies that highlighted appropriate limits for certain patient conditions. This includes limits and thresholds for the green/yellow/red organ schema, as well as the alerts and formulae.
      1. Expert Committee. For all limits and thresholds. AlertWatch sought out the opinion and confirmation of acute care physicians at the University of Michigan Health System. Each of these clinicians reviewed the limits, provided feedback, and reviewed the final results.
  • IEC 60601-1-8: A summary of how AlertWatch:AC addresses the primary issues identified in IEC 60601-1-8 is provided in the submission.

Animal Testing

Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.

Clinical Data

Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.

{8}------------------------------------------------

CONCLUSION 8.

The subject AlertWatch:AC and the predicate Philips eCareManager 4.1 (K171322) have the same intended use, namely to provide secondary monitoring of in-hospital patients. The general approach of retrieving data from the EMR system, integrating this data, and performing a series of calculations to assess potential patient clinical issues is the same. The differences in technological characteristics do not raise different questions of safety and effectiveness, and the results of performance testing demonstrate that the subject device performs in accordance with specifications and meets user needs and intended uses. Therefore, AlertWatch:AC has been demonstrated to be substantially equivalent to the predicate Philips eCareManager 4.1.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).