Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K981750
    Manufacturer
    Date Cleared
    1998-07-22

    (65 days)

    Product Code
    Regulation Number
    866.5660
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WIESLAB AB

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wielisa Anti-GBM, ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to glomerular basement membrane (GBM), Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of reno-pulmonary syndromes and rapidly progressive glomerulonephritis, especially Goodpasture syndrome (GP), Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with GP, WG, and MP. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

    Device Description

    The Wielisa Anti-GBM, ANCA Screening Test Kit is an II. enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to glomerular basement membrane (GBM), Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of reno-pulmonary syndromes and rapidly progressive glomerulonephritis, especially Goodpasture syndrome (GP). Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with GP, WG, and MP. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

    The wells of the microtiter strips are coated with purified proteinase 3 (Human Neutrophil source). MPO (Human Neutrophil source) and GBM (Bovine source) antigen. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating.

    The wells are then washed to remove unbound antibodies and other components. A coniugate of alkaline phosphatase-labeled (Goat) antibodies to human IgG binds to the antibodies in the wells in this second incubation.

    After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated and the results are given as a ratio to the negative control.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Wielisa Anti-GBM, ANCA Screening Test Kit, based on the provided text:

    Device Description: The Wielisa Anti-GBM, ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to glomerular basement membrane (GBM), Proteinase-3 (PR-3), and Myeloperoxidase (MPO) in human sera. It's used as an aid to diagnose reno-pulmonary syndromes and rapidly progressive glomerulonephritis (Goodpasture syndrome, Wegener's granulomatosis, and microscopic polyangiitis).


    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the clinical sensitivity and specificity results. The device aims to achieve high sensitivity (correctly identifying positive cases) and high specificity (correctly identifying negative cases) for each antibody against various disease states and controls.

    Metric (Antibody)Target/Acceptance Criteria (Implicit)Reported Device Performance
    Clinical Sensitivity
    PR3-ANCA (WG)High92.9% (95% CI: 84.9 – 100%)
    PR3-ANCA (MP)High51.2% (95% CI: 35.6-66.8%)
    MPO-ANCA (WG)High9.8% (95% CI: 4.9 – 19.0%)
    MPO-ANCA (MP)High46.5% (95% CI: 31.3-61.7%)
    Anti-GBM (GP)High100% (95% CI: 92.2 – 100%)
    Clinical Specificity
    PR3-ANCA (SLE)High100% (95% CI: 90.4 – 100%)
    PR3-ANCA (RA)High100% (95% CI: 92.7 - 100%)
    PR3-ANCA (NS)High100% (95% CI: 97.7 - 100%)
    MPO-ANCA (SLE)High82.8% (95% CI: 68.7 - 96.8%)
    MPO-ANCA (RA)High100% (95% CI: 92.6 - 100%)
    MPO-ANCA (NS)High100% (95% CI: 97.6 - 100%)
    Anti-GBM (SLE)High100% (95% CI: 90.4 -100%)
    Anti-GBM (RA)High100% (95% CI: 92.7 - 100%)
    Anti-GBM (NS)High100% (95% CI: 97.6 – 100%)
    Relative Sensitivity (vs. Semi-quantitative ELISA)
    PR3-ANCAHigh98.3% (95% CI: 95.0 - 100%)
    MPO-ANCAHigh95.8% (95% CI: 87.7 - 100%)
    Anti-GBMHigh100% (95% CI: 92.0 - 100%)
    Relative Specificity (vs. Semi-quantitative ELISA)
    PR3-ANCAHigh100% (95% CI: 98.0 - 100%)
    MPO-ANCAHigh100% (95% CI: 98.4 - 100%)
    Anti-GBMHigh100% (95% CI: 97.7 - 100%)
    PrecisionLow CV% (for intra-assay, inter-assay, batch-to-batch)All reported CV% values are below 21%, indicating good precision.

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Sensitivity and Specificity Study:
      • Sample Size: A total of 326 frozen serum samples.
      • Data Provenance: Retrospective sera. The country of origin is not explicitly stated.
        • Blood donors (NS): 131 samples
        • Wegener's granulomatosis (WG): 42 samples
        • Microscopic polyangiitis (MP): 43 samples
        • Systemic lupus erythematosus (SLE): 31 samples
        • Rheumatoid arthritis (RA): 41 samples
        • Goodpasture syndrome (GP): 38 samples
    • Relative Sensitivity and Specificity Study (vs. Predicate Device):
      • Sample Size:
        • 216 frozen retrospective sera for PR-3 and MPO comparison.
        • 169 frozen retrospective sera for anti-GBM comparison.
      • Data Provenance: Retrospective sera. The country of origin is not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The text states that the sera had "clinical characterisation" for the clinical sensitivity and specificity study, and were compared against an "alternative semi-quantitative ELISA" for the relative sensitivity and specificity study. However, the number of experts, their qualifications, or the specific methodology (beyond "clinical characterization") used to establish the ground truth for these samples is not provided in the given text.


    4. Adjudication Method for the Test Set

    The adjudication method for establishing the ground truth is not provided in the given text.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done as this is an in-vitro diagnostic (IVD) device (ELISA kit) that produces quantitative or qualitative results, not interpretations by human readers that would be assisted by AI. Thus, there is no AI assistance or human reader involvement in the primary assay performance.


    6. If a Standalone Study Was Done

    Yes, the studies presented are standalone (algorithm only) performance studies for the ELISA kit. The "device performance" refers to the output of the ELISA kit itself, without human interpretation or intervention in the diagnostic process beyond performing the assay according to instructions. The device generates an optical density (OD) which is then calculated as a ratio, and categorized as negative, equivocal, or positive.


    7. The Type of Ground Truth Used

    • Clinical Sensitivity and Specificity Study: "Clinical characterisation" was used as the ground truth for the disease groups (WG, MP, SLE, RA, GP) and blood donors. This implies diagnosis based on clinical evidence, potentially confirmed by other laboratory tests or pathology (though not specified).
    • Relative Sensitivity and Specificity Study: The ground truth was established by comparison to results from an "alternative semi-quantitative ELISA" for PR-3, MPO, and anti-GBM. This suggests that the predicate device or a similar established method served as the reference standard.

    8. The Sample Size for the Training Set

    The text does not provide any information about a training set. This is typical for ELISA kits, where the assay design, antigen selection, and analytical parameters are developed through R&D, rather than an explicit "machine learning training set." The studies described focus on validation of the final product.


    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned or described, the method for establishing its ground truth is not applicable/provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K981748
    Manufacturer
    Date Cleared
    1998-07-22

    (65 days)

    Product Code
    Regulation Number
    866.5660
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WIESLAB AB

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wielisa ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

    Device Description

    The Wielisa ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

    The wells of the microtiter strips are coated with purified (Human Neutrophil source) proteinase 3, and MPO (Human Neutrophil source) antigen. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating.

    The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled (Goat) antibodies to human IgG binds to the antibodies in the wells in this second incubation.

    After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated and the results are given as a ratio to the negative control.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study detailed in the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity or specificity. However, it presents the results of a comparative study against clinical characterization and a predicate semi-quantitative ELISA. We can infer the performance targets based on the documented predicate device equivalence. For the purpose of this response, I'll interpret the "acceptance criteria" as meeting or exceeding the performance of the predicate devices or demonstrating high clinical validity.

    Performance MetricSpecificity/ThresholdAcceptance Criteria (Inferred)Reported Device Performance
    Clinical Sensitivity - PR3-ANCAWG samplesHigh (e.g., >80%)92.9% (95% CI: 84.9-100%)
    MP samplesModerate (e.g., >50%)51.2% (95% CI: 35.6-66.8%)
    Clinical Sensitivity - MPO-ANCAWG samplesLow (e.g., 40%)46.5% (95% CI: 31.3-61.7%)
    Clinical Specificity - PR3-ANCASLEVery High (e.g., >95%)100% (95% CI: 90.4-100%)
    RAVery High (e.g., >95%)100% (95% CI: 92.7-100%)
    NS (Blood Donors)Very High (e.g., >95%)100% (95% CI: 97.6-100%)
    Clinical Specificity - MPO-ANCASLEHigh (e.g., >80%)82.8% (95% CI: 68.7-96.8%)
    RAVery High (e.g., >95%)100% (95% CI: 92.6-100%)
    NS (Blood Donors)Very High (e.g., >95%)100% (95% CI: 97.6-100%)
    Relative Sensitivity - PR3-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)High (e.g., >95%)98.3% (95% CI: 95.0-100%)
    Relative Sensitivity - MPO-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)High (e.g., >95%)95.8% (95% CI: 87.7-100%)
    Relative Specificity - PR3-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)Very High (e.g., >98%)100% (95% CI: 98.0-100%)
    Relative Specificity - MPO-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)Very High (e.g., >98%)100% (95% CI: 98.4-100%)
    Batch to Batch Variation (CV%)PR3-ANCALow (e.g.,
    Ask a Question

    Ask a specific question about this device

    K Number
    K974166
    Manufacturer
    Date Cleared
    1998-02-17

    (104 days)

    Product Code
    Regulation Number
    866.5660
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WIESLAB AB

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Wielisa MPO ANCA Test Kit. An Enzyme Linked Immunosorbent Assay (ELISA) for the detection and semi-quantitation of IgG antibodies in human serum to MPO (Myeloperoxidase). The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Microscopic polyangiitis.

    Device Description

    The Wielisa MPO ANCA Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of IgG antibodies to myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of microscopic polyangiitis. FOR IN VITRO DIAGNOSTIC USE.

    The wells of the microtiter strips are coated with purified myeloperoxidase. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating.

    The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled antibodies to human IgG binds to the antibodies in the wells in this second incubation.

    After a further washing step. detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated against a calibrator curve and the results are given in arbitrary units.

    AI/ML Overview

    This submission describes the Wielisa MPO ANCA Test Kit, an ELISA for detecting and semi-quantitating IgG antibodies to myeloperoxidase (MPO) in human sera, used as an aid in diagnosing microscopic polyangiitis. While the document doesn't explicitly state "acceptance criteria", it provides performance metrics that serve as the basis for demonstrating substantial equivalence to a predicate device. The studies described assess the device's clinical sensitivity and specificity, relative sensitivity and specificity compared to other methods, and precision (batch-to-batch, inter-assay, intra-assay), and linearity.

    Here's a breakdown of the acceptance criteria (inferred from the studies' findings) and the reported device performance:


    1. Table of Acceptance Criteria and Reported Device Performance
    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Clinical Sensitivity for Microscopic Polyangiitis (MP)Sufficiently high to aid diagnosis (e.g., lower bound of 95% CI > a clinically relevant threshold).47.1% (95% CI: 33.1-61.0%) for MP.
    Clinical Specificity for Healthy NormalsVery high (e.g., lower bound of 95% CI > 95%).100% (95% CI: 97.5-100%) for Healthy Normals.
    Clinical Specificity for SLEVery high (e.g., lower bound of 95% CI > 90%).97.1% (95% CI: 91.3-100%) for SLE.
    Clinical Specificity for RAVery high (e.g., lower bound of 95% CI > 95%).100% (95% CI: 82.7-100%) for RA.
    Relative Sensitivity vs. P-ANCA IFAHigh degree of agreement (e.g., >80% with acceptable CI).84.6% (95% CI: 70.5 - 98.8 %) against P-ANCA IFA.
    Relative Specificity vs. P-ANCA IFAHigh degree of agreement (e.g., >90% with acceptable CI).94.8% (95% CI: 91.7 - 97.9 %) against P-ANCA IFA.
    Relative Accuracy vs. P-ANCA IFAHigh degree of agreement (e.g., >90% with acceptable CI).93.7% (95% CI: 90.5 - 96.8 %) against P-ANCA IFA.
    Relative Sensitivity vs. Alternate ELISAVery high degree of agreement (e.g., >95% with acceptable CI).96.0% (95% CI: 89.2 - 100 %) against an Alternate ELISA.
    Relative Specificity vs. Alternate ELISAVery high degree of agreement (e.g., >95% with acceptable CI).98.8% (95% CI: 96.5 - 100 %) against an Alternate ELISA.
    Relative Accuracy vs. Alternate ELISAVery high degree of agreement (e.g., >95% with acceptable CI).97.7% (95% CI: 95.0 - 100 %) against an Alternate ELISA.
    Batch-to-batch CV %Acceptable variability (e.g., 0.9).All 6 positive sera showed strong linear correlation with r values ranging from 0.893 to 0.996.

    1. Sample sizes used for the test set and data provenance

      • Clinical Sensitivity and Specificity Study (Table 1):
        • Test Set Sample Size: 364 frozen retrospective sera.
        • Data Provenance: Retrospective sera. Country of origin is not specified but given the submitter is from Sweden, it is likely based in Europe or an international panel.
      • Comparison to P-ANCA IFA (Table 2):
        • Test Set Sample Size: 245 sera.
        • Data Provenance: Not explicitly stated as retrospective or prospective, but clinical characterization in other tables suggests a retrospective collection. Country of origin not specified.
      • Comparison to Alternate ELISA (Table 3):
        • Test Set Sample Size: 129 frozen retrospective sera.
        • Data Provenance: Retrospective sera. Country of origin not specified.
    2. Number of experts used to establish the ground truth for the test set and qualifications of those experts

      • This information is not provided in the submission. The classification of patient groups (e.g., Microscopic Polyangiitis, Wegener's Granulomatosis, Systemic Lupus Erythematosus, Rheumatoid Arthritis, Anti-GBM nephritis, Healthy Normals) implies clinical diagnosis by medical professionals, but the number or specific qualifications of these "experts" (e.g., rheumatologists, nephrologists, pathologists) who established the ground truth are not detailed.
    3. Adjudication method for the test set

      • This information is not provided in the submission. The ground truth seems to be based on pre-existing clinical characterization of the serum samples, but how these clinical diagnoses were adjudicated is not described.
    4. Multi-reader multi-case (MRMC) comparative effectiveness study, effect size

      • No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not done. This device is an in vitro diagnostic (IVD) test kit (ELISA), not an AI-powered diagnostic imaging tool or a system designed to assist human readers in interpretation where an MRMC study would be applicable. The performance is assessed on the test's ability to detect an analyte directly.
    5. Standalone (algorithm only without human-in-the-loop performance) study

      • Yes, the studies presented are effectively standalone performance studies for the device. The ELISA kit generates quantitative results (absorbance/units) that are interpreted against a cut-off (e.g., >20 units for positive) directly by the user, without a human "in the loop" for interpretation of the primary signal beyond reading the instrument output and applying the defined interpretation criteria. The device's output is the result.
    6. The type of ground truth used

      • The ground truth for the clinical sensitivity and specificity study (Table 1) was based on clinical characterization of patient groups (e.g., diagnosed cases of microscopic polyangiitis, healthy normals, other autoimmune diseases).
      • For the comparative studies (Tables 2 & 3), the ground truth was the result from a predicate/comparative method:
        • P-ANCA IFA (for Table 2).
        • An alternate commercial MPO ANCA ELISA (for Table 3).
    7. Sample size for the training set

      • This information is not provided in the document. The document describes the performance of the final device, but not its development process including how training sets (if any were used for assay development/optimization, calibration curve establishment) were utilized.
    8. How the ground truth for the training set was established

      • As the sample size and details of a training set are not provided, how its ground truth was established is also not described. For an ELISA kit, "training" typically refers to the optimization of reagents, concentrations, and establishment of cut-off values and calibrator curves, often using well-characterized positive and negative control samples. However, the specifics of this process are not detailed in this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K974167
    Manufacturer
    Date Cleared
    1998-02-17

    (104 days)

    Product Code
    Regulation Number
    866.5660
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WIESLAB AB

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wielisa PR-3 ANCA Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of antibodies to Proteinase-3 (PR-3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis.

    Device Description

    The Wielisa PR-3 ANCA Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of lgG antibodies to Proteinase-3 (PR-3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis. FOR IN VITRO DIAGNOSTIC USE.

    The wells of the microtiter strips are coated with purified proteinase 3. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating,

    The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled antibodies to human IgG binds to the antibodies in the wells in this second incubation.

    After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated against a calibrator curve and the results are given in arbitrary units.

    AI/ML Overview

    The provided text is a summary of safety and effectiveness information for the Wielisa PR-3 ANCA Test Kit, a medical device cleared by the FDA in 1998. It primarily focuses on demonstrating substantial equivalence to a predicate device and presenting performance data from various studies.

    Here's an attempt to extract the requested information, acknowledging that specific "acceptance criteria" (i.e., predefined thresholds for performance that the device must meet) are not explicitly stated as typical in modern regulatory submissions. Instead, the document presents the reported performance and implicitly expects these results to be sufficient to demonstrate substantial equivalence to the predicate.


    Acceptance Criteria and Device Performance for Wielisa PR-3 ANCA Test Kit

    1. Table of Acceptance Criteria and Reported Device Performance

    As explicit, pre-defined acceptance criteria (e.g., "Sensitivity must be >90%") are not stated in the document, the "Reported Device Performance" column reflects what was achieved in the studies, which implicitly served as the basis for FDA clearance. The predicate device's performance is not detailed in this document, so direct comparison against a numerical benchmark from the predicate is not possible from the given text.

    Performance MetricImplicit Acceptance Criteria (based on predicate equivalence)Reported Device Performance (Wielisa PR-3 ANCA Test Kit)Study Reference
    Clinical Sensitivity (WG)Sufficient for aid in diagnosis92.9% (95% CI: 86.7-99.0%)Table 1
    Clinical Sensitivity (MP)Sufficient for aid in diagnosis50.0% (95% CI: 36.1-63.9%)Table 1
    Clinical Specificity (SLE)Sufficient for aid in diagnosis97.4% (95% CI: 92.4-100%)Table 1
    Clinical Specificity (RA)Sufficient for aid in diagnosis100% (95% CI: 82.7-100%)Table 1
    Clinical Specificity (GBM)Sufficient for aid in diagnosis93.6% (95% CI: 87.3-99.8%)Table 1
    Clinical Specificity (Normals)Sufficient for aid in diagnosis100% (95% CI: 97.5-100%)Table 1
    Relative Sensitivity (vs C-ANCA IFA)Comparable to predicate device96.3% (95% CI: 92.1-100.0%)Table 2
    Relative Specificity (vs C-ANCA IFA)Comparable to predicate device91.9% (95% CI: 87.6-96.2%)Table 2
    Relative Accuracy (vs C-ANCA IFA)Comparable to predicate device93.4% (95% CI: 90.2-96.6%)Table 2
    Relative Sensitivity (vs Alternate ELISA)Comparable to predicate device100% (95% CI: 92.0-100%)Table 3
    Relative Specificity (vs Alternate ELISA)Comparable to predicate device97.5% (95% CI: 94.1-100%)Table 3
    Relative Accuracy (vs Alternate ELISA)Comparable to predicate device98.3% (95% CI: 95.9-100%)Table 3
    Batch-to-Batch Mean CVAcceptable for IVD use8.9% - 14.3% (for various samples)Table 4
    Inter-Assay Mean CVAcceptable for IVD useNot clearly reported due to OCR error; likely ≤ 15%Table 5
    Intra-Assay Mean CVAcceptable for IVD useNot clearly reported due to OCR error; likely ≤ 10%Table 6
    Linearity (r value)Demonstrated linear relationship0.811 - 0.945 (for various positive sera)Table 7

    2. Sample Size Used for the Test Set and the Data Provenance

    • Clinical Sensitivity and Specificity (Table 1):
      • Sample Size: 364 frozen retrospective human sera.
      • Provenance: Retrospective, country of origin not specified, but the submission is from a Swedish company (Wieslab AB), implying European or international samples are possible.
    • Relative Sensitivity and Specificity vs. C-ANCA IFA (Table 2):
      • Sample Size: 245 frozen retrospective human sera.
      • Provenance: Retrospective, country of origin not specified.
    • Relative Sensitivity and Specificity vs. Alternate ELISA (Table 3):
      • Sample Size: 120 frozen retrospective human sera.
      • Provenance: Retrospective, country of origin not specified.
    • Batch-to-batch variation (Table 4): 4 samples tested in duplicate across four different batches.
    • Inter-assay precision (Table 5): 2 samples tested in duplicate across six different runs.
    • Intra-assay precision (Table 6): 1 sample tested in 80 wells.
    • Linearity (Table 7): 7 positive sera, serially diluted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    The document states that the sera for clinical sensitivity and specificity (Table 1) had "clinical characterization." However, it does not specify the number or qualifications of experts used to establish the ground truth (diagnosis of Wegener's granulomatosis, microscopic polyangiitis, etc.). It can be inferred that these diagnoses were established through standard clinical practice by medical professionals, but no details are provided regarding the adjudication or consensus process for these diagnoses in the context of this study.

    For the comparative studies in Tables 2 and 3 (C-ANCA IFA and Alternate ELISA), the ground truth was the result of the comparator device. These are not human expert adjudications but rather comparisons against other laboratory tests.

    4. Adjudication Method for the Test Set

    • For the clinical characterization (Table 1), the document only states "clinical characterization." There is no explicit description of an adjudication method (e.g., 2+1, 3+1, none) for the diagnoses of the disease groups.
    • For the comparative studies (Tables 2 and 3), the "ground truth" was established by the predicate or comparator assay (C-ANCA IFA or an alternate commercial ELISA). This is a direct comparison to a reference test, rather than human expert adjudication of images or clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) ELISA kit designed for laboratory use, not for direct human interpretation of medical images or complex clinical cases where MRMC studies are typically performed to assess human reader performance with and without AI assistance. The studies here compare the device's performance against clinical diagnoses or other established laboratory tests.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, this entire submission represents a standalone performance evaluation. The Wielisa PR-3 ANCA Test Kit is a laboratory assay. The reported measurements of sensitivity, specificity, and precision are derived directly from the assay's output (absorbance readings converted to arbitrary units), without inherent human interpretation as part of the primary diagnostic output. Human intervention occurs at the level of performing the test and interpreting the final quantitative result within a clinical context, but the performance metrics presented are of the device itself.

    7. The Type of Ground Truth Used

    • Clinical Sensitivity & Specificity (Table 1): The ground truth was based on clinical characterization of the disease groups (Wegener's granulomatosis, microscopic polyangiitis, systemic lupus erythematosus, rheumatoid arthritis, anti-GBM nephritis, and healthy normals). This implies established clinical diagnoses, likely confirmed by a combination of clinical signs, symptoms, and other diagnostic tests, but not exclusively a single definitive "pathology" in all cases.
    • Relative Sensitivity & Specificity (Tables 2 and 3): The ground truth was established by comparison to other laboratory tests: C-ANCA IFA (Table 2) and an alternate commercial ELISA (Table 3).

    8. The Sample Size for the Training Set

    The document does not provide information on a specific "training set" for the device. As an ELISA kit from 1998, it predates the widespread application of machine learning or AI models that require distinct training sets. The "development" of such a test would involve formulation, optimization of reagents, and establishment of cutoff values, rather than a machine learning training phase on a dedicated dataset. The data presented are performance evaluation data.

    9. How the Ground Truth for the Training Set Was Established

    Since no specific "training set" is mentioned (as the device is not an AI/ML diagnostic model), this question is not applicable in the context of the provided document. The "ground truth" for establishing the kit's parameters (e.g., calibrator values, cutoffs) would have been determined through internal development and validation processes using characterized samples, but these are not explicitly detailed as a "training set" in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K974169
    Manufacturer
    Date Cleared
    1998-02-17

    (104 days)

    Product Code
    Regulation Number
    866.5660
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WIESLAB AB

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wielisa anti-GBM Test Kit is an Enzyme Linked Immunosorbent Assay (ELISA) for the detection and semi-quantitation of IgG antibodies in human serum to GBM (glomerular basement membrane). The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Goodpasture syndrome. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    The Wielisa anti-GBM Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection and semi-quantitation of IgG antibodies to glomerular basement membrane(GBM) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Goodpasture syndrome. FOR IN VITRO DIAGNOSTIC USE. The wells of the microtiter strips are coated with purified GBM antigen. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating. The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled antibodies to human IgG binds to the antibodies in the wells in this second incubation. After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated against a calibrator curve and the results are given in arbitrary units.

    AI/ML Overview

    The Wieslab AB Ideon Research Park's Wielisa anti-GBM Test Kit is an enzyme-linked immunosorbent assay (ELISA) designed for the detection and semi-quantitation of IgG antibodies to glomerular basement membrane (GBM) in human sera. The assay aims to aid in the diagnosis of Goodpasture syndrome. The device applied for 510(k) clearance (K974169) to the FDA. The submission outlines studies demonstrating the performance characteristics of the device.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the provided document. Instead, the performance is demonstrated through comparison with clinical characterization and existing methods. The implied acceptance is that the device demonstrates high sensitivity and specificity in diagnosing anti-GBM nephritis and distinguishing it from other conditions, as well as showing comparable results to a predicate device and an alternative ELISA.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Clinical Sensitivity (anti-GBM nephritis)High sensitivity for detecting anti-GBM nephritis.100% (95% CI: 95.0-100%)
    Clinical Specificity (Blood donors)High specificity for healthy individuals.100% (95% CI: 97.5-100%)
    Clinical Specificity (Systemic vasculitis: WG and MP)High specificity for other autoimmune conditions (Systemic vasculitis).96.9% (95% CI: 90.7-100%)
    Clinical Specificity (SLE)High specificity for other autoimmune conditions (SLE).100% (95% CI: 92.6-100%)
    Clinical Specificity (RA)High specificity for other autoimmune conditions (RA).100% (95% CI: 82.7-100%)
    Relative Sensitivity (vs. GBM IFA)High sensitivity when compared to Immunofluorescence Assay (IFA) as an existing method.100.0% (95% CI: 94.6-100.0%)
    Relative Specificity (vs. GBM IFA)High specificity when compared to Immunofluorescence Assay (IFA) as an existing method.88.9% (95% CI: 67.9-100.0%)
    Relative Accuracy (vs. GBM IFA)High overall accuracy when compared to Immunofluorescence Assay (IFA).98.4% (95% CI: 95.3-100.0%)
    Relative Sensitivity (vs. Alternate ELISA)High sensitivity when compared to another commercial ELISA.92.7% (95% CI: 84.6-100%)
    Relative Specificity (vs. Alternate ELISA)High specificity when compared to another commercial ELISA.100% (95% CI: 95.9-100%)
    Relative Accuracy (vs. Alternate ELISA)High overall accuracy when compared to another commercial ELISA.97.4% (95% CI: 94.3-100%)
    Batch to Batch Variation (CV%)Low coefficient of variation (CV%) to demonstrate consistency across different manufacturing batches.3.1% - 17.7% (for samples with mean values 16-134 units)
    Inter-assay Precision (CV%)Low coefficient of variation (CV%) to demonstrate consistency between different assay runs.3.0% (for 46 units) and 7.6% (for 154 units)
    Intra-assay Precision (CV%)Low coefficient of variation (CV%) to demonstrate consistency within a single assay run.10% (for 77 units)
    Linearity (r value)High correlation (r value close to 1) for serial dilutions.0.823 - 0.994

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Sensitivity and Specificity Study (Table 1):
      • Sample Size: 272 frozen sera.
      • Data Provenance: Retrospective, with clinical characterization. The country of origin is not specified, but the applicant's address is in Sweden.
    • Relative Sensitivity and Specificity vs. GBM IFA (Table 2):
      • Sample Size: 68 frozen retrospective sera.
      • Data Provenance: Retrospective. Country of origin not specified.
    • Relative Sensitivity and Specificity vs. Alternate ELISA (Table 3):
      • Sample Size: 122 frozen retrospective sera.
      • Data Provenance: Retrospective. Country of origin not specified.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not explicitly state the number of experts used or their detailed qualifications (e.g., radiologist with 10 years of experience).

    • For Table 1 (Clinical Sensitivity and Specificity), the "clinical characterization" of the sera implies that a clinical diagnosis was used as ground truth. This would typically involve review by clinicians, potentially specialists in nephrology or rheumatology, based on a combination of clinical symptoms, other laboratory tests, and potentially biopsy results. However, the exact number and qualifications are not provided.
    • For Table 2 (Relative Sensitivity and Specificity vs. GBM IFA), the Immunofluorescence Assay (IFA) results are used as the comparator, which serves as a form of ground truth or an established benchmark. The interpretation of IFA typically requires experienced laboratory personnel or pathologists.
    • For Table 3 (Relative Sensitivity and Specificity vs. Alternate ELISA), another commercial ELISA is used as the comparator, also serving as an established benchmark.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (like 2+1 or 3+1) for resolving discrepancies in diagnostic classifications for the ground truth. It relies on "clinical characterization" or comparison with established methods (IFA, alternate ELISA).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study focuses on the performance of the in-vitro diagnostic device itself, not on the improvement of human readers with or without AI assistance. The device is a laboratory assay, not a tool for interpretation by human readers in the same way an imaging AI might be.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented are essentially standalone performance evaluations of the Wielisa anti-GBM Test Kit. This is an immunoassay that produces quantitative results (optical density, then arbitrary units) and then applies defined cut-offs (e.g., 20 units positive) to classify samples. There is no "human-in-the-loop" interaction in the interpretation of the assay's output as part of the primary device function itself, beyond a clinician interpreting the final numerical result provided by the kit in the context of a patient's overall clinical picture.

    7. The Type of Ground Truth Used

    • Clinical Sensitivity and Specificity (Table 1): Clinical diagnosis based on "clinical characterization" of the sera. This would likely stem from a combination of patient symptoms, medical history, other diagnostic tests, and potentially pathology (e.g., kidney biopsy for anti-GBM nephritis).
    • Relative Sensitivity and Specificity vs. GBM IFA (Table 2): Results from Immunofluorescence Assay (IFA) for GBM antibodies. IFA is considered a standard method for detecting these antibodies.
    • Relative Sensitivity and Specificity vs. Alternate ELISA (Table 3): Results from an alternate commercial ELISA for anti-GBM antibodies. This serves as a comparison against another established in-vitro diagnostic approach.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a separate "training set" or "validation set" in the context of machine learning. For an ELISA kit, development typically involves optimizing reagents and conditions using various samples, but this is not typically referred to as a "training set" in the same way as in AI/ML. The provided sample sizes (272, 68, 122) appear to be for the performance evaluation/testing of the device with established parameters.

    9. How the Ground Truth for the Training Set Was Established

    Since a "training set" in the AI/ML sense is not explicitly described, neither is the method for establishing its ground truth. However, for the development of such an assay, the "ground truth" used during optimization would typically involve well-characterized patient samples with confirmed diagnoses (e.g., biopsy-proven Goodpasture syndrome, healthy controls, or patients with other well-defined autoimmune diseases).

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1