K Number
K981748
Manufacturer
Date Cleared
1998-07-22

(65 days)

Product Code
Regulation Number
866.5660
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wielisa ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

Device Description

The Wielisa ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

The wells of the microtiter strips are coated with purified (Human Neutrophil source) proteinase 3, and MPO (Human Neutrophil source) antigen. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating.

The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled (Goat) antibodies to human IgG binds to the antibodies in the wells in this second incubation.

After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated and the results are given as a ratio to the negative control.

AI/ML Overview

Here's an analysis of the acceptance criteria and study detailed in the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity or specificity. However, it presents the results of a comparative study against clinical characterization and a predicate semi-quantitative ELISA. We can infer the performance targets based on the documented predicate device equivalence. For the purpose of this response, I'll interpret the "acceptance criteria" as meeting or exceeding the performance of the predicate devices or demonstrating high clinical validity.

Performance MetricSpecificity/ThresholdAcceptance Criteria (Inferred)Reported Device Performance
Clinical Sensitivity - PR3-ANCAWG samplesHigh (e.g., >80%)92.9% (95% CI: 84.9-100%)
MP samplesModerate (e.g., >50%)51.2% (95% CI: 35.6-66.8%)
Clinical Sensitivity - MPO-ANCAWG samplesLow (e.g., <20%)9.8% (95% CI: 4.9-19.0%)
MP samplesModerate (e.g., >40%)46.5% (95% CI: 31.3-61.7%)
Clinical Specificity - PR3-ANCASLEVery High (e.g., >95%)100% (95% CI: 90.4-100%)
RAVery High (e.g., >95%)100% (95% CI: 92.7-100%)
NS (Blood Donors)Very High (e.g., >95%)100% (95% CI: 97.6-100%)
Clinical Specificity - MPO-ANCASLEHigh (e.g., >80%)82.8% (95% CI: 68.7-96.8%)
RAVery High (e.g., >95%)100% (95% CI: 92.6-100%)
NS (Blood Donors)Very High (e.g., >95%)100% (95% CI: 97.6-100%)
Relative Sensitivity - PR3-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)High (e.g., >95%)98.3% (95% CI: 95.0-100%)
Relative Sensitivity - MPO-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)High (e.g., >95%)95.8% (95% CI: 87.7-100%)
Relative Specificity - PR3-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)Very High (e.g., >98%)100% (95% CI: 98.0-100%)
Relative Specificity - MPO-ANCA (vs. Semi-quantitative ELISA)(Implied equivalence)Very High (e.g., >98%)100% (95% CI: 98.4-100%)
Batch to Batch Variation (CV%)PR3-ANCALow (e.g., <20%)3-11%
MPO-ANCALow (e.g., <25%)7-24%
Inter-assay Precision (CV%)PR3-ANCALow (e.g., <15%)10-11%
MPO-ANCALow (e.g., <25%)5-21%
Intra-assay Precision (CV%)PR3-ANCAVery Low (e.g., <10%)5-6%
MPO-ANCAVery Low (e.g., <10%)3-5%

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Sensitivity and Specificity Study (Table 1):
    • Sample Size: A total of 288 sera.
    • Data Provenance: Frozen retrospective sera with clinical characterization. The country of origin is not specified, but the manufacturer is based in Sweden.
  • Relative Sensitivity and Specificity Study (Table 2):
    • Sample Size: A total of 216 frozen retrospective sera.
    • Data Provenance: Frozen retrospective sera. The country of origin is not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number of experts or their qualifications. For the "Clinical sensitivity and specificity" study (Table 1), the ground truth is based on "clinical characterisation." For the "Relative sensitivity and specificity" study (Table 2), the ground truth is established by a "semi-quantitative ELISA," which implies a laboratory-based gold standard. There is no mention of expert review in either case for establishing ground truth, beyond the initial clinical diagnoses.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method involving human experts for the test set results. The ground truths are either based on clinical diagnoses or the results of a predicate semi-quantitative ELISA.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This device is an ELISA test kit, not an AI-powered diagnostic tool that assists human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the device performance presented is for the standalone "algorithm" (the ELISA test kit itself). The results for sensitivity, specificity, and precision metrics are reported for the kit's performance independent of human-in-the-loop interpretation beyond standard laboratory procedures for operating an ELISA test.

7. The Type of Ground Truth Used

  • Clinical Sensitivity and Specificity Study (Table 1): Clinical diagnoses/characterization (e.g., diagnosed with Wegener's granulomatosis (WG), microscopic polyangiitis (MP), Systemic Lupus Erythematosus (SLE), Rheumatoid Arthritis (RA), or healthy blood donors (NS)).
  • Relative Sensitivity and Specificity Study (Table 2): Results from a "semi-quantitative ELISA" (the predicate devices, Wielisa PR-3 ANCA ELISA Kit and Wielisa MPO ANCA ELISA Kit).
  • Precision Studies (Table 3, 4, 5): The ground truth for these studies is the inherent signal/value of the control samples used to assess reproducibility.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI. As this is an ELISA test kit, not an AI software, the concept of a training set for an algorithm is not applicable in the conventional sense. The studies described are performance validation studies.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, the concept of a "training set" for an AI algorithm is not applicable here. The establishment of ground truth for the validation studies is described in point 7.

{0}------------------------------------------------

JUL 22 1998

Summary of Safety and Effectiveness Information ANCA Screening ELISA Test Kit

I. Wieslab AB Ideon Research Park S-223 70 Lund Sweden Contact person: Dr. Jorgen Wieslander Telephone: 46-46-182840 Date of preparation: May 1, 1998

Description of Device: The Wielisa ANCA Screening Test Kit is an enzyme-linked II. immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

The wells of the microtiter strips are coated with purified (Human Neutrophil source) proteinase 3, and MPO (Human Neutrophil source) antigen. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating.

The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled (Goat) antibodies to human IgG binds to the antibodies in the wells in this second incubation.

After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated and the results are given as a ratio to the negative control.

III. Predicate Device

The ANCA Screening test is substantially equivalent to the Wielisa PR-3 ANCA ELISA Kit and the Wielisa MPO ANCA ELISA Kit. Equivalence is demonstrated by the following comparative results:

{1}------------------------------------------------

Control andDisease groupsTotalNegative <3PR3 MPOEquivocal 3-4PR3 MPOPositive >4PR3 MPO
----------------------------------------------------------------------------------------------------------------
Table 1. Clinical sensitivity and specificity. A total of 288 frozen retrospective sera with clinical
characterisation were assayed. The following table summarises the results
Control andDisease groupsTotalNegative +3Equivocal +?Positive +?
PR3MPOPR3MPOPR3MPO
Blood donors: (NS)1311311270400
WG:4233701394
MP:432023202120
SLE:313124020*5
RA:4141400100

WG = Wegener's granulomatosis, MP = microscopic polyangiitis RA = rheumatoid arthritis

SLE = systemic lupus erythematosus GBM = glomerular basement membrane

*All samples were positive in the semi-quantitative MPO-ELISA.

Clinical sensitivity (Equivocal samples not included in the calculations)

PR3-ANCA:WG = 39/42 = 92.9 %95% CI = 84.9-100%
MP = 21/41 = 51.2 %95% CI = 35.6-66.8%
MPO-ANCA:WG = 4/41 = 9.8 %95% CI = 4.9-19.0%
MP = 20/43 = 46.5 %95% CI = 31.3-61.7%

Clinical specificity (Equivocal samples not included in the calculations)

PR3-ANCA: SLE = 31/31 = 100 %95% CI = 90.4-100%
RA = 41/41 = 100 %95% CI = 92.7-100%
NS = 131/131 = 100 %95% CI = 97.6-100%
MPO-ANCA: SLE = 24/29 = 82.8 %95% CI = 68.7-96.8%
RA = 40/40 = 100 %95% CI = 92.6-100%
NS = 127/127 = 100 %95% CI = 97.6-100%

{2}------------------------------------------------

Table 2. Relative sensitivity and specificity of the Wielisa ANCA screen kit compared to an semi-quantitative ELISA. A total of 216 frozen retrospective sera were assayed. The following table summarises the results.

Wielisa ANCA screen
Semi-quantitative ELISANegative <3Equivocal 3-4Positive >4
PR3MPOPR3MPO
MPO-ANCAPositive0100023
PR3-ANCAPositive1000590
Negative1521822400
Equivocal140111
Total154187256024

Sera falling in the equivocal range were are not including in the calculations:

Relative sensitivity PR3-ANCA = 59/60 = 98.3 %95% CI = 95.0-100%
Relative sensitivity MPO-ANCA = 23/24 = 95.8 %95% CI = 87.7-100%
Relative specificity PR3-ANCA = 152/152 = 100 %95% CI = 98.0-100%
Relative specificity MPO-ANCA = 182/182 = 100 %95% CI = 98.4-100%

.

{3}------------------------------------------------

SamplePR3MeanOD ratioSDCV %SampleMPOMeanOD ratioSDCV %
234.51.03316.33.924
519.82.211627.82.17
833.81.549232.913

Table 3. Batch to batch variation was determined by testing three different samples. Results were obtained for 4 different batches.

Table 4. Inter-assay precision was determined by testing one sample. Results were obtained for six different runs.

SampleODratioSDCV %SampleMPOMeanOD ratioSDCV %
PK27.32.911PK17.33.821
K512.11.210K616.50.845

Table 5. Intra-assay precision was determined by testing one sample in 22 wells.

SampleMeanODSDCV %SampleMeanODSDCV %
PK1.30.076PK1.90.063
K51.30.065K61.460.075

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged around the circumference of the circle.

JUL 2 2 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Weislab AB c/o William L. Boteler, Jr. IMMUNO PROBE, INC. 1306 Bailes Lane, Suite F Frederick, MD 21701

K981748 Re: Trade Name: Wielisa ANCA Screening Test Kit Regulatory Class: II Product Code: MOB Dated: May 3, 1998 Received: May 18, 1998

Dear Mr. Boteler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such Existing major regulations affecting your device additional controls. can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its

internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

K981748 510(k) Number: Not known

Device Name: Wielisa ANCA Screening Test Kit

Indications For Use: The Wielisa ANCA Screening Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the qualitative detection of antibodies to Proteinase-3 (PR-3) and Myeloperoxidase (MPO) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of systemic vasculitis, especially Wegener's granulomatosis (WG) and : microcopic polyangiitis (MP). The assay is intended for use in patients with signs and symptoms consistent with systemic vasculitis. It is not intended for screening a healthy population. A positive result should always be confirmed by a semi-quantitative assay.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) _____________________________________________________________________________________________________________________________________________________________________________

Concurrence of CDRH, Office of Device Evaluation (ODE)

11 Prescription Use (Per 21 CFR 801.109)

OR

Over-The Counter Use (Optional Format 1-2-96)

Ji Ma
(Division Sign-Off)
Divotlinical Laboratory Devices K98,748
510(k) Number

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).