K Number
K231148Device Name
ScleroSafe 150 mm, ScleroSafe 350 mm
Manufacturer
VVT Medical Ltd.
Date Cleared
2023-06-20
(60 days)
Product Code
KRA, FMF
Regulation Number
870.1210Why did this record match?
Applicant Name (Manufacturer) :
VVT Medical Ltd.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
ScleroSafe Peripheral Venous Aspiration & Infusion Kit with Dual Procedure Syringe (ScleroSafe) is intended for the delivery of Asclera®, an FDA-approved sclerosant (Asclera®), in the treatment of varicosities in superficial veins with a diameter of 2 to 3mm.
Device Description
The subject device, the ScleroSafe™ 150 mm, ScleroSafe™ 350 mm (or ScleroSafe Peripheral Venous Aspiration & Infusion Kit with Dual Procedure Syringe) is intended for the delivery of FDA-approved sclerosant Asclera® in the treatment of varicosities in superficial veins with a diameter of 2 to 3mm.
The ScleroSafe device includes a 5 Fr dual lumen catheter ("DLC"), mountable on a 0.018 inch straight-tipped guidewire in an over-the-wire configuration. The DLC is connected to a hub that splits the catheter into two separate tubes. One of the tubes ends with a blue luer lock for the aspiration extension and the other with a transparent luer lock for the injection. Following needle insertion into the guidewire is removed and a Dual Procedure Syringe ("DPS") is connected to the DLC hub via dedicated luers. The larger syringe (10 ml) is connected to blue luer lock connector (Aspiration Extension) of the smaller syringe (5 ml) is connected to a transparent luer lock connector of the catheter (Injection Extension) so fluids are simultaneously injected through one lumen and aspirated through another lumen. ScleroSafe is provided in two sizes: 150mm and 350mm.