Search Results
Found 2 results
510(k) Data Aggregation
(163 days)
THEROX, INC.
The infusion catheter is intended to assist in the controlled infusion of diagnostic or therapeutic solutions into the peripheral or coronary vasculature. The target vessel effective diameter is ≥ 2.0 mm. The catheter is placed by a physician using a guide catheter and over a guidewire that extends beyond the final position of the infusion catheter.
The TherOx® Infusion Catheter is a sterile, single-use, 4.6 French (F) intravascular catheter with a 127 cm usable length. The catheter is comprised of three sections; 1) a distal atraumatic soft tip 2) a semi-rigid but flexible shaft; and 3) a proximal luer hub for connection of solution-delivery devices. A radiopaque band at the distal tip allows fluoroscopic visualization of the catheter's location in the vessel.
TherOx, Inc. sought 510(k) clearance for their TherOx® Infusion Catheter, indicating substantial equivalence to the Boston Scientific/Target Therapeutics Tracker® - 38 Infusion Catheter. The study performed was a non-clinical test data summary comparing the TherOx® Infusion Catheter to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance:
Characteristic | Acceptance Criteria (based on predicate device and standards) | TherOx® Infusion Catheter Performance |
---|---|---|
Physical/Material | ||
Outer Diameter | Comparable to predicate to allow intended use | 4.6 F (0.060 in) overall (smaller than predicate's 5.3F/5F) |
Inner Diameter | Sufficient for recommended guidewire | 0.046 in overall, 0.037 in min at marker band (slightly smaller than predicate at marker band) |
Usable Length | Comparable to predicate for physician manipulation | 127 cm (longer than predicate's 115cm/117cm) |
Materials | Biocompatible and suitable for intravascular use | High density polyethylene (HDPE) shaft, rigid HDPE luer, LDPE plasticized tip (minor differences from predicate, but compliant) |
Radiopacity | Visible under fluoroscopy | Radiopaque band at distal tip allows visualization |
Sterile | Sterile | Sterilized with ethylene oxide |
Single-Use | Single-use only | Single-use only |
Shelf Life | Established shelf life | 3-year shelf life established |
Functional/Performance | ||
Strength | Meets standards (ISO 10555 series) for intravascular catheters | Test results met acceptance criteria |
Flexibility | Meets standards (ISO 10555 series) for intravascular catheters | Test results met acceptance criteria |
Pressure | Meets standards (ISO 10555 series) for intravascular catheters | Test results met acceptance criteria |
Flow | Meets standards (ISO 10555 series) for intravascular catheters | Test results met acceptance criteria |
Biocompatibility | Meets ISO 10993-1 | Satisfies ISO 10993-1 requirements |
Sterilization | Validated per ISO 11135 | Validated per ISO 11135 |
Packaging | Designed and tested per ISO 11607 | Designed and tested per ISO 11607 |
Joint Strength | Within specified requirements | Test results met acceptance criteria |
Dimensional Conformance | Within specified requirements | Test results met acceptance criteria |
2. Sample size used for the test set and the data provenance:
- Sample Size: The document does not explicitly state the exact sample size (number of catheters) used for each individual test. It mentions "production samples that were sterilized twice in a validated sterilization cycle" and "samples that were sterilized twice and then thermally aged to simulate an expected shelf life." This implies a sufficient number of devices were tested to validate the performance and shelf life.
- Data Provenance: The data is from non-clinical bench testing conducted by TherOx, Inc. explicitly for this 510(k) submission. It is prospective in the sense that the tests were performed to demonstrate compliance of the TherOx® Infusion Catheter. The country of origin of the data is implied to be where TherOx, Inc. is located (Irvine, CA, USA) or where their testing facilities are.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This was a non-clinical bench study, not a clinical study involving human expert interpretation. Therefore, there were no experts used to establish a "ground truth" in the sense of clinical diagnosis or interpretation. The "ground truth" or acceptance criteria were derived from international standards (ISO 10555 series, ISO 10993-1, ISO 11135, ISO 11607) and internal device specifications, as well as comparison to the predicate device. The engineers and scientists conducting the tests would have been qualified in their respective fields of materials science, biomechanics, and medical device testing.
4. Adjudication method for the test set:
- Not applicable. As a non-clinical bench study, there was no adjudication method involving multiple human readers or experts to determine a consensus "ground truth." Test results were compared directly to the predetermined acceptance criteria outlined in the relevant standards and internal specifications.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not performed. This submission is for a physical medical device (an infusion catheter) and evaluates its mechanical and material properties, not an AI or imaging diagnostic tool that would involve human reader interpretation.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm or software. No standalone algorithm performance was evaluated.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the non-clinical tests was established by predetermined engineering specifications and compliance with recognized international and national standards for intravascular catheters and medical device testing. These standards dictate acceptable ranges for physical dimensions, material properties, strength, flexibility, pressure resistance, flow rates, biocompatibility, sterilization efficacy, and packaging integrity. The predicate device's characteristics also served as a reference point.
8. The sample size for the training set:
- Not applicable. This is a non-clinical bench study for a physical device, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable, as there was no training set for a machine learning model.
Ask a specific question about this device
(178 days)
THEROX, INC.
TherOx Infusion Guidewires are intended for use in facilitating the placement of interventional devices in the cardiovascular and neurovascular system. TherOx Infusion Guidewires are designed for small vessel use.
TherOx Infusion Guidewires are also intended to be used for the controlled selective infusion of physician-specified fluid agents before, during or after interventional procedures in the cardiovascular or neurovascular system.
TherOx Infusion Guidewires are not intended for delivery of contrast agents. Fluid agents utilized with TherOx Infusion Guidewires should be fully prepared and used according to their manufacturer's instructions for use.
TherOx Infusion Guidewires are intended for use in facilitating the placement of interventional devices in the cardiovascular and neurovascular system. TherOx Infusion Guidewires are designed for small vessel use.
TherOx Infusion Guidewires are also intended to be used for the controlled selective infusion of physician-specified fluid agents before, during or after interventional procedures in the cardiovascular or neurovascular system.
TherOx Infusion Guidewires are not intended for delivery of contrast agents. Filuid agents utilized with TherOx Infusion Guidewires should be fully prepared and used according to their manufacturer's instructions for use.
TherOx Infusion Guidewires are not intended for neonatal or pediatric use.
The TherOx Infusion Guidewire has an outside diameter of .014 in. and is available 180 cm and 300 cm in length, and is intended to pass freely thru standard catheters that have a thru lumen of 016 in. or larger.
The 300 cm TherOx Exchange Guidewire differs only in its overall length from the 180 cm TherOx Infusion Guidewire. The distal 180 cm of the TherOx Exchange Guidewire is identical to the 180 cm TherOx Infusion Guidewire. Only the proximal hypotube has been extended by 120 cm to achieve the 300 cm length. This extended proximal portion always remains outside the body and facilitates the replacement of one interventional device for another.
A proximal fluid connection is secured by connecting the proximal open end to a simple Tuohy-Borst connection. The infusion guidewire may be torqued by rotation of the torquing knob to direct the orientation of the distal tip.
Here's a breakdown of the acceptance criteria and the study information for the TherOx Infusion Guidewires, based on the provided text:
Acceptance Criteria and Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it describes various tests performed and concludes that the device meets its functional specifications and is safe for its intended use, or performed comparably to predicate devices.
However, we can infer the acceptance criteria from the "Summary of Studies" section and present the reported performance against these implicit criteria:
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
In Vitro Tests: | |
Functionality and Durability | Met or exceeded functional specification, safe for intended use. |
Tensile Strength | Not explicitly detailed, but implied to meet functional specifications. |
Guidewire Torque Response | Not explicitly detailed, but implied to meet functional specifications. |
Tip Deflection and Strength | Not explicitly detailed, but implied to meet functional specifications. |
Joint Integrity and Strength | Not explicitly detailed, but implied to meet functional specifications. |
Handling Characteristics | Not explicitly detailed, but implied to meet functional specifications. |
Fluid Infusion Characteristics | Not explicitly detailed, but implied to meet functional specifications. |
Product Failure Modes | Not explicitly detailed, but implied to meet functional specifications. |
Dimensional Conformance (Length, O.D., joint sizes, etc.) | All dimensions were within the design specifications. |
Radiopacity | Demonstrated adequate radiopacity when compared with commercially available guidewires and vascular catheters. |
Biocompatibility (for short-term vascular use, |
Ask a specific question about this device
Page 1 of 1