Search Results
Found 4 results
510(k) Data Aggregation
(270 days)
Synaptic Medical Corporation
The Nordica PV Cryo Mapping Catheter is indicated for multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart.
The Nordica PV Cryo Mapping Catheter is a 3F sterile, single use, multi-electrode, diagnostic catheter designed to map cardiac signals during ablation procedures. The catheter is provided with a fixed loop diameter of 15mm or 20mm with either 5 or 4 evenly spaced, radiopaque pairs of electrodes. The proximal end of the catheter contains an electrical connection for the EP electrical cable that integrates with electrophysiology lab recording systems. Once deployed in the vasculature through the central lumen of a catheter, a circular or spiral shape is established such that the electrodes contact the endocardial surface. This allows for recording, pacing, and interrogation of electrical conduction between the left atrium and the pulmonary veins.
The provided text describes the submission for a 510(k) premarket notification for the "Nordica PV Cryo Mapping Catheter" and focuses on demonstrating substantial equivalence to a predicate device.
Crucially, this document focuses on the safety and performance of a medical device (a catheter) through various engineering, sterilization, biocompatibility, and animal studies. It explicitly states: "This section is not applicable as the subject and predicate devices do not contain any software components or utilize software for their intended use." and "Human Clinical Performance Testing: Clinical testing was not required to demonstrate substantial equivalence to the predicate device."
Therefore, the request for acceptance criteria and a study proving a device meets the acceptance criteria in the context of AI/software performance, human readers, ground truth, and training/test sets is not applicable to this document.
The document details the following types of testing for the physical medical device:
1. A table of acceptance criteria and the reported device performance:
The document provides a "Table 1: Comparison of Technological Characteristics with the Predicate Device" which implicitly serves as a comparison of the subject device's features against the predicate's known characteristics, thereby establishing the "acceptance criteria" through equivalence. The "reported device performance" is then demonstrated through various specific tests.
Feature / Acceptance Criteria Category | Predicate Device (Implied Acceptance) | Subject Device (Reported Performance) | Rationale for Substantial Equivalence |
---|---|---|---|
Product Code | DRF | DRF | Same Product Code as Predicate Device. |
Regulation | 21 CFR 870.1220 | 21 CFR 870.1220 | Same Regulation Number as Predicate. |
Indications for Use | "multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The Achieve Advance mapping catheter is designed to obtain electrograms in the atrial regions of the heart." | "multiple electrophysiological mapping of cardiac structures (i.e., recording and stimulation only). The Nordica PV Cryo Mapping Catheter is designed to obtain electrograms in the atrial regions of the heart." | Same Indications for Use as Predicate Device. |
Intended Use | "to map intracardiac structures of the heart." | "to map intracardiac structures of the heart." | Same Intended Use as Predicate Device. |
Sterilization | Ethylene Oxide Sterilized | Ethylene Oxide Sterilized (Validated to ISO 11135:2014+A1:2019, SAL 10^-6 or better) | Same Sterilization Method as Predicate Device (with updated standard compliance). |
Single Use | Yes | Yes | Same Single Use as Predicate Device. |
Biocompatibility | Complies with ISO 10993-1 (as specified in K153139) | Complies with ISO 10993-1 (2018) and FDA Guidance. Tested endpoints: ISO MEM Elution, ASTM Hemolysis, SC5b-9 Complement Activation Assay, ASTM Partial Thromboplastin Time (PTT), In-Vivo Thrombogenicity Evaluation [GLP Animal Study], ISO Guinea Pig Maximization Sensitization Test, ISO Intracutaneous Study in Rabbits, ISO Material Mediated Pyrogen Study (GLP), ISO Acute Systemic Toxicity Study in Mice. All met acceptance criteria. | Same Biocompatibility compliance as Predicate Device - subject device complies to current revision of the standard. |
Performance (General Standard) | Per ISO 10555-1 | Per ISO 10555-1:2013/AMD 1:2017 (for general catheter performance) | Same Performance standards as Predicate Device - subject device complies to current revision of the standard. |
Performance Testing (Specific) | Joint Strength, Corrosion Resistance, Radiopacity, Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Kink Resistance, Stiffness, Life Testing. | Dimensional, Atraumatic Tip & Surface, Joint Strength / Peak Tensile Force, Corrosion Resistance, Radiopacity (verified in GLP animal study), Radial Loop Compliance, Distal Stiffness, Axial Load, Torque Response, Electrical Continuity, Turns to Failure, Device Functionality Kink Resistance Stiffness Life Testing. All met design input acceptance criteria. | Same as predicate device - testing for subject device performed in accordance with latest versions of standards. Requirements established for these tests were validated for their intended use through V&V testing. |
Outer Diameter | 3F (1mm) | 3F (1mm) | Same Outer Diameter size as predicate device. |
Effective Length | 146 cm | 149 cm | Equivalent effective length; demonstrated compatibility, safe and effective use through V&V studies. |
Distal End Shape | Circular Loop | Circular Loop | Same Distal End Shape as predicate device. |
Loop Diameter | 15mm & 20mm (K162892), 25 mm (K153139) | 15mm & 20mm | Same Loop Diameters as predicate device. |
Number of Electrodes | 8 (15mm loop), 10 (25mm loop) | 8 (15mm loop), 10 (20mm loop) | Same number of electrodes as predicate device. (Note: Predicate has 25mm loop option not present in subject device, but the 15mm and 20mm subject options match predicate's corresponding electrode counts). |
Electrode Spacing | Unpaired, Spacing Between Electrodes: 4.1mm (15mm loop), 5.8mm (20mm loop), 5.8mm (25mm loop) | Paired Electrodes, Spacing Between Pairs: 7.3 mm (15mm loop/8 electrodes), 7.3mm (20mm loop/10 electrodes); Spacing Between Electrodes: 2 mm (15mm and 20mm loops) | Similar electrode spacing to predicate device. Differences due to paired electrodes; demonstrated safety and performance through V&V testing. |
Delivery Method | Delivered Through a Delivery Catheter: Yes | Delivered Through a Delivery Catheter: Yes | Same delivery method as predicate device. |
Materials | Catheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel / Polyimide (Proximal). Loop Material: Nitinol insulated with PET (Pebax Covered). Electrode Material: Not Specified. Tip: Not Specified. | Catheter Body Tubing: Nitinol / Pebax (Distal), Stainless Steel (Proximal). Loop Material: Nitinol covered with Pebax. Electrode Material: 90% Platinum / 10% Iridium. Tip: Loctite 4311 Medical Grade Adhesive. | Subject and predicate devices constructed of equivalent materials. Differences qualified through biological safety and V&V testing. |
Sterilization Residuals | (Implicitly meets standards based on predicate's clearance) | Met acceptance criteria below limits specified in EN ISO 10993-7:2008 + A1:2022. | Compliance with updated standard. |
Shelf-Life | (Implicitly demonstrated for predicate) | Accelerated aging per ASTM F1980-21. Package Integrity (label, bubble leak ASTM F2096-11(2019), seal visual, seal strength ASTM F88/F88M-21) and Product V&V performed after aging. | Demonstrated through testing. |
Electrical Safety / EMC | (Implied compliance for predicate) | Complies with IEC 60601-1-2 Ed. 4.1 (2020-09) for Class A Emissions; IEC 60601-1:2005 (and amendments); IEC 60601-1-6:2010 (and amendments) for usability. | Compliance with updated standards. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test set for Biocompatibility: "All tested subject devices met the acceptance criteria for each biological endpoint test". The exact number of devices tested for each biological endpoint (e.g., ISO MEM Elution, Hemolysis, etc.) is not specified, but it implies a sufficient sample size was used for each test to provide robust results.
- Test set for Benchtop Studies: "The results of benchtop testing demonstrated the design outputs of the subject device met the design input acceptance criteria required to demonstrate substantial equivalence with the predicate device." Similarly, exact sample sizes for each benchtop test (e.g., buckle force, catheter deflection, etc.) are not provided.
- Pre-Clinical GLP Animal Study: A "chronic GLP animal study" was performed on a "canine animal model". The exact sample size (number of animals) is not specified.
- Data Provenance: The document does not specify the country of origin for the data or whether the studies were retrospective or prospective, beyond stating it was a "chronic GLP animal study" which implies a prospective, controlled study in a laboratory setting. All testing appears to be internal (Synaptic Medical Corporation) or contracted to ISO/ASTM compliant labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as the studies are engineering, biological, and animal model evaluations, not human-read AI evaluations.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device has no software components or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the various tests, the "ground truth" is defined by the established scientific and engineering standards and methods:
- Biocompatibility: Defined by the acceptance criteria of ISO 10993 series standards (e.g., no cytotoxicity, no hemolysis, no sensitization, no pyrogenicity, etc.).
- Sterilization: Defined by ISO 11135:2014+A1:2019 (achieving SAL of 10^-6 or better) and EN ISO 10993-7:2008 + A1:2022 (residual limits).
- Shelf-Life: Defined by ASTM F1980-21 for accelerated aging and specific packaging integrity tests (ASTM F2096-11(2019), ASTM F88/F88M-21).
- Benchtop Performance: Defined by ISO 10555-1:2013/AMD 1:2017 and internal "design input acceptance criteria" for various physical properties (e.g., buckle force, tensile strength, electrical continuity).
- Pre-Clinical GLP Animal Study: Evaluation of "chronic safety, performance, and usability" in a canine model, presumably against pre-defined physiological endpoints and observational criteria.
8. The sample size for the training set
Not applicable, as this device does not involve AI/machine learning.
9. How the ground truth for the training set was established
Not applicable, as this device does not involve AI/machine learning.
Ask a specific question about this device
(178 days)
Synaptic Medical Corporation
Ask a specific question about this device
(88 days)
Synaptic Medical Corporation
The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit is intended to introduce various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum.
The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit consists of a 1) Steerable Sheath, 2) Dilator and 3) Guidewire, and is indicated for introducing various cardiovascular catheters into the heart, including the left side of the heart through the interatrial septum. When the left side of the heart is accessed, the device is used with a compatible Transseptal Needle to puncture the interatrial septum for transseptal catheterization procedures. The NaviGo 12F Steerable Intracardiac Catheter Introducer Kit is a sterile, single use device, with asymmetrical bi-directional steerable introducer with usable length of 65cm. The Steerable Sheath is fitted with a hemostasis valve to minimize blood loss during catheter introduction and/or exchange. A side port with three-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling, and pressure monitoring. The handle is equipped with a rotating collar to deflect the Steerable Sheath tip clockwise ≥ 180° and counterclockwise ≥ 90°. The steerable introducer features distal vent holes to facilitate aspiration and a radiopaque tip marker to improve fluoroscopic visualization.
The provided text describes the NaviGo 12F Steerable Intracardiac Catheter Introducer Kit and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and the Reported Device Performance
The document describes various performance tests conducted, but it does not explicitly list quantitative acceptance criteria for each test in a table format with corresponding reported performance values. Instead, it generally states that the device "met the acceptance criteria" or "demonstrated robust biocompatibility."
However, based on the descriptions, we can infer some general criteria and outcomes:
Performance Test Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Sterilization | Successful validation per ISO 11135:2014+A1:2019. Bioburden testing per ISO 11737-1. LAL testing per ANSI/AAMI ST72:2019 and FDA Guidance. EO and ECH residuals below limits per EN ISO 10993-7:2008+A1:2022. | Sterilization process developed and validated. Bioburden testing performed. LAL testing performed. EO and ECH residual testing demonstrated results below specified limits. |
Shelf-Life | Package integrity (label, bubble leak, seal visual, seal strength) maintained after accelerated aging (ASTM F1980-21). Product V&V (safety and performance) confirmed after accelerated aging. | Shelf-life testing performed via accelerated aging. Package integrity testing (label, bubble leak, seal visual, seal strength) performed and confirmed. Product V&V performed and confirmed. (Note: 6-month shelf-life initially, with potential for extension). |
Biocompatibility | Compliance with ISO 10993-1:2018 for "Externally Communicating Device, Circulating Blood with Limited Exposure" (e.g., passing results for cytotoxicity, hemolysis, complement activation, thrombogenicity, sensitization, irritation, pyrogenicity, acute systemic toxicity). | All tested subject devices met the acceptance criteria for each biological endpoint test and demonstrated robust biocompatibility in accordance with ISO 10993-1. |
Benchtop Studies | Design outputs met design input acceptance criteria for each specific test (Visual & Dimensional, Buckle Force, Catheter Deflection, Shaft Bending & Kink Resistance, Simulated Use & Compatibility, Freedom from Leakage, Peak Tensile Force & Bond Strength). | Results of benchtop testing demonstrated the design outputs of the subject device met the design input acceptance criteria. |
Pre-Clinical (GLP Animal Study) | Primary Objective: Overall safety demonstrated. Secondary Objective: Performance and usability assessed with "Acceptable" and "Above Average" ratings. Thrombogenicity assessment completed. | Overall safety demonstrated. Performance and usability evaluation met acceptance criteria with "predominant rating of Acceptable and Above Average." Thrombogenicity assessment completed. |
2. Sample Size Used for the Test Set and the Data Provenance
- Sterilization Testing: Not explicitly stated, but bioburden testing, LAL testing, and EO/ECH residual testing were performed on "the subject device."
- Shelf-Life Testing: Not explicitly stated, but accelerated aging studies and various package integrity tests were performed on "the subject device."
- Biocompatibility Testing: Not explicitly stated, but various in-vitro and in-vivo tests were conducted. The "In-Vivo Thrombogenicity Evaluation [GLP Animal Study]" refers to the animal study below. Materials were from the device itself.
- Benchtop Studies: Not explicitly stated, but various devices would have been tested for each category.
- Pre-Clinical (GLP Animal Study):
- Sample Size: Six (6) healthy canines.
- Data Provenance: Prospective animal study conducted in a GLP (Good Laboratory Practice) environment. The location/country of the GLP lab is not specified but it's an animal model, not human data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
- Pre-Clinical (GLP Animal Study):
- Number of Experts: Two (2)
- Qualifications: "Board Certified Electrophysiologists specializing in ablation of pulmonary veins in the management of atrial fibrillation."
4. Adjudication Method for the Test Set
- For the Pre-Clinical (GLP Animal Study), the two electrophysiologists "performed the procedures and rated the overall performance and usability of the subject device." The document does not specify a formal adjudication method (e.g., 2+1, 3+1). It states "predominant rating of Acceptable and Above Average," suggesting a consensus or combined assessment was reached, but the process is not detailed.
- For other tests (benchtop, sterilization, biocompatibility), the ground truth is based on standard test methods and established scientific criteria, not expert adjudication in the same sense.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This device is a physical medical device (catheter introducer kit), not an AI/software device that would typically involve human readers interpreting images with or without AI assistance. The study described is a pre-clinical animal study to assess safety, performance, and usability of the physical device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
No, a standalone algorithm performance study was not done. This is not an AI/software device.
7. The Type of Ground Truth Used
- Sterilization: Regulatory standards (ISO, ANSI/AAMI, EN ISO) for sterilization efficacy, bioburden levels, endotoxin levels, and residual limits.
- Shelf-Life: Regulatory standards (ASTM F1980-21, ASTM F2096-11(2019), ASTM F88/F88M-21) and product V&V testing.
- Biocompatibility: Regulatory standards (ISO 10993-1:2018, ISO 10993-4, -5, -7, -10, -11, -23) and FDA guidance for biological evaluation endpoints.
- Benchtop Studies: Design input acceptance criteria based on engineering specifications and industry standards relevant to catheter function (e.g., buckle force, deflection, kink resistance, leakage, bond strength).
- Pre-Clinical (GLP Animal Study): Direct observation, clinical assessment, and expert rating by Board Certified Electrophysiologists in a canine model for safety, performance, and usability. This includes observations for adverse events, device maneuverability, and overall handling, as well as thrombogenicity assessment.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device. The "training set" concept is typically relevant to AI/machine learning models.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set for an AI/ML model.
Ask a specific question about this device
(413 days)
Synaptic Medical Corporation
The RithmID-SD Steerable Diagnostic Electrophysiology Catheter can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.
The RithmID-SD Steerable Diagnostic Electrophysiology Catheter are biocompatible, flexible, radiopaque electrophysiology catheters that are available in a variety of diameters, lengths, curve shapes, and electrode number and spacing configurations, with a high-torque shaft with an array of platinum iridium alloy electrodes at the distal tip that can be used for recording electrical signals. The catheter is designed to facilitate the electrophysiological mapping of the heart.
The RithmID-SD Steerable Diagnostic Electrophysiology Catheter consist of a handle, a shaft and a steerable diagnostic tip. The catheter is introduced through the sheath and into the femoral vein, from the inferior vena cava into the heart and coronary sinus. The RithmID-SD Steerable Diagnostic Electrophysiology Catheter are available in 6F in various curves, including B, D, Y, and R curves.
The provided text is a 510(k) summary for a medical device called the "RithmID-SD Steerable Diagnostic Electrophysiology Catheter". This document focuses on demonstrating substantial equivalence to predicate devices, primarily through bench testing, biocompatibility testing, and comparisons of technical characteristics.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list acceptance criteria for each test in a separate column. Instead, for most tests, it states "All test samples met the acceptance criteria," indicating that the criteria were satisfied. The performance results are qualitative; for a few tests, comparative statements are made relative to the reference device.
Here's a compilation of the tests and their reported performance:
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Corrosion Resistance | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Particulate Matter Evaluation | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria and is equivalent to the Reference Device. |
Radiopacity Detectability | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Signal Acquisition | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Usability | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. The performance of the Subject Device is better or equivalent to the Reference Device. |
Electrical Safety | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Visual Inspection (Bench Testing) | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Dimensional Verification | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Simulated Use | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Connection Plug Force | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Tip Fatigue Tolerance | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Shaft Fatigue Tolerance | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Flexural Fatigue Tolerance | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Torsional Force | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Electrode Conductor Resistance | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. |
Buckling Force | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. The buckling force of the Subject Device is less than or equal to the Reference Device. |
Peak Tensile Force | Not explicitly stated, implied "met criteria" | All test samples met the acceptance criteria. The peak tensile force of the Subject Device was compared to the Reference Device. |
Torque Strength | Data comparison, not a pass/fail criterion | The average number of 360° rotations when permanent mechanical deformation occurs was recorded for the Subject Device and Reference Device. The torque strength of the Subject device is compared to the Reference Device for reference only. |
Packaging Integrity (Distribution Sim.) | ASTM D4169-16 | All test samples met the acceptance criteria per ASTM D4169-16. |
Packaging Integrity (Visual Insp.) | ASTM F1886/F1886M-16 | All test samples met the acceptance criteria per ASTM F1886/F1886M-16. |
Packaging Integrity (Bubble Leak) | ASTM F2096-11 | All test samples met the acceptance criteria per ASTM F2096-11. |
Packaging Integrity (Tray Seal Strength) | F88/F88M-15 | All test samples met the acceptance criteria per F88/F88M-15. |
Cytotoxicity | Non-cytotoxic (ISO 10993-5) | Non-cytotoxic |
Sensitization | Non-sensitizer (ISO 10993-10) | Non-sensitizer |
Intracutaneous Reactivity | Non-irritant (ISO 10993-10) | Non-irritant |
Acute Systemic Toxicity | Non-Toxic (ISO 10993-11) | Non-Toxic |
Systemic Toxicity (Pyrogenicity) | Non-Pyrogenic (ISO 10993-11) | Non-Pyrogenic |
Hemocompatibility (Hemolysis) | Non-Hemolytic (ISO 10993-4) | Non-Hemolytic |
Hemocompatibility (Thrombogenicity) | Passed (ISO 10993-4) | Passed |
Hemocompatibility (Complement Act.) | Non-Activator (ISO 10993-4) | Non-Activator |
Hemocompatibility (PTT) | Passed (ISO 10993-4) | Passed |
Shelf Life (Accelerated Aging) | Meets design specification at 1 year | All acceptance criteria met; device performs as intended to Design Specification. |
Sterilization (SAL) | SAL of 10^-6 | Achieved a SAL of 10^-6. |
EO and ECH Residuals | Below limits specified in ISO 10993-7 | Below the limits specified in ISO 10993-7. |
Bacterial Endotoxin Levels |
Ask a specific question about this device
Page 1 of 1