Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K211619
    Date Cleared
    2022-12-29

    (582 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Shenzhen Comen Medical Instruments CO., LTD.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NC10 and NC12 patient monitors are intended to be used for monitoring, displaying, alarming and storing of multiple physiological parameters These parameters include ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), anesthetic gas (AG), and Bispectral index (BIS) for a single patient.

    All parameters can be monitored on single adult, pediatric, and neonatal patients except:

    • · BIS monitoring is intended for adult and pediatric patients only;
    • · C.O. monitoring is restricted to adult patients only;
      · Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
    • · When using COMEM SpO2, the monitor is intended to be used on adult patients only.
    • · NIBP measurement continual mode is not applicable to neonates.

    The monitors are to be used in general healthes by clinical physicians or appropriate medical staff under the direction of physicians.

    The monitors are not intended for helicopter transport, hospital ambulance, or home use.

    The monitors do not measure, display, or trend changes in the ST segment.

    The monitors do not intend for use as apnea monitors.

    The monitors are not intended for use in MRI or CT environments.

    The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.

    Device Description

    The NC10 and NC12 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing multiple physiological parameters. These parameters include ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), anesthetic gas (AG), and Bispectral index (BIS) for a single patient.
    All parameters can be monitored on single adult, pediatric, and neonatal patients except:
    BIS monitoring is intended for adult patients only; C.O. monitoring is restricted to adult patients only; Arrhythmia analysis is intended for use with adult patients only and is not intended and shall not be used on pediatric and neonatal population. When using COMEM SpO2, the monitor is intended to be used on adult patients only. NIBP measurement continual mode is not applicable to neonates. Both models are designed with:
    Same system framework and components
    Same hardware design principle
    Same software platform
    Same parameters measurement subsystems (including parameters modules and accessories)
    The only difference between NC10 and NC12 is the display size.

    AI/ML Overview

    The acceptance criteria and supporting study details for the Multi-Parameter Patient Monitor (NC10 and NC12) are provided below, based on the given FDA 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present specific "acceptance criteria" for each physiological parameter in a tabular format with corresponding "reported device performance." Instead, it compares the subject device's specifications to those of the predicate device, stating that the subject device's performance aligns with or is a subset of the predicate's performance, and that the device meets relevant consensus standards. The "Comparison" column in the provided tables indicates "Same" for most parameters, implying that the subject device's performance is equivalent to the established performance of the predicate device. For the "Comen SpO2" feature, where there's a difference, the document states, "The SpO2 accuracy met ISO 80601-2-61 and was validated by the clinical study," indicating that its performance meets the standard.

    Here's a condensed representation of the key performance specifications for the subject device (NC10 and NC12), which also serve as implied acceptance criteria given the "Same" comparison to the predicate:

    ParameterMethod / Range / Accuracy (NC10 & NC12)
    ECG (Arrhythmia Analysis)Asystole, ventricular fibrillation, R ON T, VT>2, Couplet, PVC, Bigeminy, Trigeminy, Brady, PNC, PNP, Missed Beats, Heart Pause, Irregular Heart Beat, VTAC, Tachy, PVCs Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm.
    RespirationMethod: Trans-thoracic impedance; Range: adult:0-120 rpm, pediatrics:0-150rpm, neonate:0-150rpm; Accuracy: 7 to 150rpm: ±2rpm or ±2%, whichever is greater.
    SpO2 (Masimo)Method: red and infrared light; Range: 1~100%; Accuracy: No motion: 70-100% ±2%(adult/pediatric), 70-100% ±3%(neonate); Motion: 70-100% ±3%.
    SpO2 (Nellcor)Method: red and infrared light; Range: 0~100%; Accuracy: 70-100% ±2%(adult/pediatric), 70-100% ±3%(neonate).
    SpO2 (Comen)Method: red and infrared light; Range: 0~100%; Accuracy: 70-100% ±3% (adult). (Note: Restricted to adult use compared to predicate)
    Pulse Rate (from SpO2, IBP, NIBP)Ranges: 20-350 bpm (variable by source); Accuracy: ±1bpm to ±5bpm or ±1% to ±3% (variable by source and conditions).
    NIBPMethod: Oscillometry; Range: Adult: systolic:40-270 mmHg, diastolic:10-215 mmHg; Pediatrics: systolic:40-200 mmHg, diastolic:10-150 mmHg; Neonate: systolic:40-135 mmHg, diastolic:10-100 mmHg; Error: Max mean error: ±5 mmHg, Max standard deviation: 8 mmHg.
    TemperatureMethod: Thermal resistance; Range: 0-50°C; Accuracy: ±0.1°C.
    CO2 (Masimo)Method: Infrared absorption; Range: 0-190mmHg; AwRR:0-150rpm; Accuracy: 0-114mmHg: ±(2.25mmHg+reading×4%).
    CO2 (Respironics)Method: Infrared absorption; Range: 0-150mmHg; AwRR: 0, 2-150bpm; Accuracy: 0-40mmHg: ±2mmHg, other ranges higher % errors.
    IBPMethod: Direct invasive measurement; Range: -50 to 300 mmHg; Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error).
    Cardiac OutputMethod: Thermodilution; Range: 0.1 to 20 L/min; Accuracy: ±5% or ±0.1 L/min, whichever is greater.
    Anesthetic Gas (AG)Method: Infrared absorption; Ranges for various gases; Accuracy for CO2, N2O, Hal, Enf, Iso, Sev, Des, O2, awRR.
    BISRange and Accuracy: SQI: 0-100%, 1%; EMG: 0100dB, 1%; BIS: 0-100, 1%; SR: 0100%, 1%; ESR: 0~100%, 1%.

    The document implies that the "reported device performance" for the subject device meets or is equivalent to these specified ranges and accuracies through bench testing and clinical studies, confirming compliance with relevant standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify exact numerical sample sizes for each clinical test. It mentions that clinical accuracy of NIBP, SpO2, and respiratory rate were validated for the intended patient population.
      • For SpO2 accuracy, it states the validation was done "using the method outlined in ISO 80601-2-61:2017 and the FDA guidance Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff, March 2013." These standards typically require a certain number of subjects (often healthy volunteers) with induced hypoxemia for desaturation studies to demonstrate accuracy across the specified range. However, the exact number is not provided in this summary.
      • For NIBP accuracy, it states validation was "according to ISO 81060-2 which contains the requirements for clinical accuracy and the protocols for investigating the NIBP determination clinical accuracy." This standard also prescribes specific subject enrollment criteria and measurement methods.
      • For Respiratory Rate (RR) accuracy, it was validated "by clinical testing to compare the measurement of the subject device and that of a clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography." The sample size for this is not detailed.
    • Data Provenance: The document does not explicitly state the country of origin of the data. It also does not explicitly state whether the studies were retrospective or prospective, though clinical validation studies for device clearance are typically prospective. It does say "All clinical accuracy validation studies were conducted in accordance with standard ISO 14155:2020," which governs clinical investigation of medical devices, generally implying prospective collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number or qualifications of experts used for establishing ground truth, as it is a multi-parameter patient monitor.

    • For SpO2, the ground truth would typically be established by a CO-oximeter reading during a controlled desaturation study, as per ISO 80601-2-61. This is a highly objective measurement.
    • For NIBP, ground truth is typically established by direct intra-arterial blood pressure measurements, not by expert consensus.
    • For Respiratory Rate, the ground truth was "clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography." This implies clinically trained personnel, but their specific qualifications or number are not provided.
    • For Arrhythmia Analysis, and other subjective physiological monitoring parameters, the ground truth source is not explicitly mentioned but typically relies on expert interpretation of ECG waveforms or other data.

    4. Adjudication Method for the Test Set

    The document does not detail any adjudication methods (e.g., 2+1, 3+1) for the test set, as most of the parameters are quantitative measurements compared against an objective reference standard rather than subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This document describes a multi-parameter patient monitor, which is a measurement device, not an AI-assisted diagnostic imaging tool. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not applicable to this type of device and was not conducted. The study aims to demonstrate that the device's measurements are accurate and equivalent to predicate devices, not to show an improvement in human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The entire device, including its algorithms for parameter measurement and arrhythmia detection, operates in a "standalone" fashion to generate the values and alarms displayed to the clinician. The performance validated (e.g., accuracy of SpO2, NIBP, RR, arrhythmia detection) is the inherent performance of the device's algorithms and hardware. While a human uses the device and interprets its output, the core measurements are algorithm-driven.

    7. The Type of Ground Truth Used

    • SpO2: CO-oximetry in a controlled desaturation study (objective, gold-standard reference for SpO2 saturation).
    • NIBP: Direct intra-arterial blood pressure measurements (objective, gold-standard).
    • Respiratory Rate: Clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography. (This suggests an expert-derived observation from an objective measurement, or comparison to another well-established measurement device).
    • ECG/Arrhythmia Detection and other parameters: The document implies comparison to established methods and compliance with relevant ISO standards, which would typically involve highly accurate reference measurements and possibly expert review of waveforms for specific event detection.

    8. The Sample Size for the Training Set

    The document does not provide information about a training set or its sample size. This is common for device clearances that focus on performance validation rather than machine learning algorithm development where distinct training and test sets are crucial. The device's algorithms are likely based on established physiological principles and signal processing, rather than deep learning from a massive training dataset.

    9. How the Ground Truth for the Training Set was Established

    As no specific training set is mentioned in the filing summary for this device, information regarding the establishment of its ground truth is not applicable or provided. The device's performance is demonstrated through its adherence to established international standards and clinical testing against reference methods.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Shenzhen Comen Medical Instruments Co.,Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    · The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;

    · NIBP measurement continual mode is not applicable to neonates;

    · Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;

    The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.

    Device Description

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
    • NIBP measurement continual mode is not applicable to neonates;
    • Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
    Those monitors provide patient monitoring capabilities by using corresponding accessories.
    The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
    But the software in the host and modules and components are different.
    The differences between C50 and C80 are ICG, BIS and size of monitor.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.

    FeaturePredicate Device (K170876) PerformanceSubject Device (C50 & C80) PerformanceComparison to Acceptance Criteria (Predicate) & Justification
    Monitor SizePassport 17m: 400x370x193mm; Passport 12m: 297x336x187mmC80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mmThe monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised.
    Integrated display & touch screenColor TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m)Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels)Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability.
    Power supplyTwo rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12mC50 and C80: Powered either by built-in battery or external AC."Same, both powered by battery and AC." Implied acceptance: functional equivalence.
    Battery11.1V, 4500mAh11.1V, 2200mAh/4400mAh"The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained.
    ECG3-lead, 5-lead and 12-lead selectable, heart rate (HR)3-lead, 5-lead and 12-lead selectable, heart rate (HR)"Same." Implied acceptance: functional equivalence.
    ECG (Arrhythmia Analysis)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types)Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types)"C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance.
    RespirationMethod: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified.Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified."Same." Implied acceptance: functional and performance equivalence.
    Pulse oxygen saturation (SpO2)Method: red and infrared light method. Masimo SpO2: Range:1100%. Accuracy: No motion: 70-100% ±2%(adult/pediatric), ±3%(neonate); Motion: 70-100% ±3%. Nellcor SpO2: Range:0100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate). Mindray SpO2: Range:0~100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate).Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion)."Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy.
    Pulse rate (PR)From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%).From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%)."The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61.
    Non-invasive blood pressure (NIBP)Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg.Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg."The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30.
    Temperature (Temp)Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C.Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C."The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56.
    Carbon dioxide (CO2)Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm.Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm."The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55.
    Invasive blood pressure (IBP)Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor)."Same." Implied acceptance: functional and performance equivalence.
    Cardiac output (C.O.)Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater.Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater."Same." Implied acceptance: functional and performance equivalence.
    Anesthetic gas (AG)Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL).Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm)."For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules.
    BIS (C80 only)Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied).Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%."The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules.
    ICG (C80 only)Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified.Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified."The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components.

    Study that Proves the Device Meets Acceptance Criteria

    The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.

    Non-Clinical Performance Data:

    • Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
    • Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
      • ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
      • IEC 60601-1-2 Edition 4: 2014-02
    • Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
    • Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
    • Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).

    Clinical Studies:

    1. NIBP Clinical Study - for adults and children

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 25 patients (19 adults, 6 children).
      • Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    2. NIBP Clinical Study - for neonate and infants

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 20 subjects.
      • Demographics: 10 males, 10 females. 13 subjects
    Ask a Question

    Ask a specific question about this device

    K Number
    K173454
    Date Cleared
    2018-08-16

    (283 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Shenzhen Comen Medical Instruments Co.,Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. It's used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. All parameters can be applied to adult, pediatric or neonate. The Patient Monitor is not intended to be used on the patients with pacemaker.

    The monitor is used for monitoring the clinical patients, so only the doctors and nurses who are qualified through training can use these monitors.

    Device Description

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is intended to be used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Resp), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. Other parameters can be applied to adult, pediatric or neonate.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Shenzhen Comen Medical Instruments Co., Ltd. Multi-parameter Patient Monitor, Model C30. It describes the device's characteristics, intended use, and comparisons to legally marketed predicate devices.

    However, the document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a specific clinical study for the device's performance in isolation. The testing mentioned is primarily related to electrical safety, electromagnetic compatibility, and adherence to performance standards, along with some clinical verification for NIBP and respiration rate accuracy.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of an AI algorithm's performance (such as sensitivity, specificity, or reader improvement) is not directly available or applicable to this type of medical device submission. This document describes a traditional multi-parameter patient monitor, not an AI-powered diagnostic device.

    Based on the provided text, here's what can be extracted and what cannot be provided:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various performance characteristics for each parameter (ECG, Respiration, Temperature, SpO2, Pulse Rate, NIBP, CO2) as part of the technological comparison to predicate devices, including measurement ranges and accuracies. These can be considered the de facto performance criteria. While the document states the device was "tested and found to be in compliance with recognized safety, performance and electromagnetic compatibility standards," it does not provide a specific table of acceptance criteria with corresponding measured performance for each parameter from a single comprehensive study report. Instead, it infers compliance through comparison to predicate device specifications and adherence to standards.

    ParameterAcceptance Criteria (from predicate comparison/standards)Reported Device Performance (Implied from substantial equivalence and compliance statements)
    ECG Heart Rate (HR)Neonate/Pediatric: 15 to 350 bpm; Adult: 15 to 300 bpm; Accuracy: ±1% or ±1 bpm (whichever is larger)The document states the device has "similar" characteristics and "complies" with performance standards, implying it meets these ranges and accuracies, but no direct measured performance values are provided.
    Respiration Rate (Resp)Measurement range: Adult: 0 to 120 rpm; Neonate: 0 to 150 rpm; Accuracy: 7 to 150 rpm: ±1 rpm (for subject device, compared to predicate's ±2 rpm or ±2%); 0 to 6 rpm: Not specified.Clinical study indicated to verify RR accuracy by comparing subject device RR to capnography. Implies accuracy within the stated/expected range, but no specific results are quantified in the summary.
    Temperature (Temp)Measurement range: 0 to 50°C (32 to 122°F); Accuracy: ±0.1°C or ±0.2°F (without probe)No specific measured performance values provided, but implied to be compliant with these specifications.
    SpO2Masimo SpO2: Measurement range: 1 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric, no motion), ±3% (neonate, no motion, and with motion). Nellcor SpO2: Measurement range: 0 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric), ±3% (neonate). 0 to 69%: Not specified.No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Pulse Rate (PR)From Masimo SpO2: Range 25 to 240 bpm; Accuracy: ±3 bpm (no motion), ±5 bpm (with motion). From Nellcor SpO2: Range 20 to 300 bpm (20-250 bpm: ±3 bpm, 251-300 bpm: not specified).No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Non-Invasive Blood Pressure (NIBP)Measurement range (Adult/Pediatric/Neonate systolic/mean/diastolic ranges vary as per table). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. (These are performance requirements based on ISO 81060-2:2013)."According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This statement indicates testing was done according to the standard, implying it met the accuracy requirements specified by that standard, but explicit performance values are not provided.
    Carbon Dioxide (CO2)Respironics: Range 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%. awRR: 0-150 bpm, Accuracy: ±1 rpm. Masimo: Range 0-190 mmHg; Accuracy: 0-114 mmHg: ±1.52 mmHg + 2%. awRR: 0-150 bpm, Accuracy: ±1 rpm.No specific measured performance values provided, but implied to be compliant with these specifications.

    2. Sample size used for the test set and the data provenance:

    • NIBP Clinical Study: The document states "According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This standard specifies requirements for Non-Invasive Sphygmomanometers, including clinical investigation procedures. However, the exact sample size for this specific study and the data provenance (country, retrospective/prospective) are not given in this summary.
    • Respiration Rate Clinical Study: "To verify respiration rate (RR) accuracy, Clinical study was conducted comparing the RR from the subject device with that from capnography." The exact sample size and data provenance are not given.
    • For other parameters, bench testing and compliance with standards are mentioned, but no specific clinical test sets or sample sizes are detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided as this is not an AI diagnostic device and does not inherently involve expert image interpretation or similar processes for ground truth establishment. Clinical studies mentioned (NIBP, RR) would likely use reference measurements from other validated medical devices as a "ground truth" or comparator, but details on expert involvement are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/provided. Adjudication methods are typically used in studies involving subjective assessments, especially common in AI diagnostic imaging where multiple readers reconcile disagreements. This device is a physiological monitor measuring objective parameters.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable/provided. This is a physiological monitoring device, not an AI-powered diagnostic tool intended to assist human readers in interpreting complex medical data or images. Therefore, an MRMC study is beyond the scope of this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable/provided. This device is a multi-parameter patient monitor that measures physiological signals directly rather than an AI algorithm performing a standalone diagnostic function. The device's performance is its direct measurement output, which is then used by healthcare professionals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the NIBP and Respiration Rate clinical studies, the "ground truth" would likely be established through reference measurements from other validated medical devices or established clinical methods. For example, NIBP accuracy usually compares the device's readings to those obtained through an invasive arterial line or another highly accurate, validated NIBP device. For respiration rate, capnography (a precise measurement of CO2 in exhaled breath) was explicitly mentioned as the comparator.

    8. The sample size for the training set:

    The device described is a traditional patient monitor developed through engineering and design principles, not a machine learning or AI model that requires a "training set." Therefore, this concept is not applicable to this product.

    9. How the ground truth for the training set was established:

    As above, the concept of a "training set" and "ground truth for the training set" is not applicable to this traditional medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1