Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K250470
    Date Cleared
    2025-06-23

    (125 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.

    Device Description

    The ear thermometer is a handheld device that displays the temperature of the measured patient by measuring the thermal radiation of the eardrum. Measurement unit: °C or °F. The results can be displayed on LCD. The thermometers are powered by 1.5V×2 (AAA or AA) alkaline batteries, which can be used for people over three months. A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the eardrum, which is converted into temperature measurement with the unit of °C or °F. All the models share the same critical components, intended use, working principle and similar product design, and compose of a sensor, PCB, buttons, LCD display and housing. Functions include temperature measurement, memory reading recall, unit switch, low battery detection and high temperature indicator.

    AI/ML Overview

    The FDA 510(k) clearance letter and accompanying 510(k) Summary for the Ear Thermometers (EAR-E101, EAR-E102, EAR-E103) provide information on acceptance criteria and supporting studies.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Jiangsu Yuyue Medical Equipment & Supply Co., Ltd, Infrared Ear Thermometers YHT101 and YHT200, K203583). Thus, the "acceptance criteria" are largely framed as demonstrating equivalence or adherence to relevant standards. The performance data is presented in comparison to the predicate and overall compliance with standards.

    Acceptance Criteria (Stated or Implied by Comparison)Reported Device Performance (Subject Device)
    Accuracy (Clinical)Measurement Range: 32.0℃~42.9℃ (89.6°F-109.2°F) Accuracy: ±0.2℃/±0.4°F (for 32.0℃ to 42.9℃)
    BiocompatibilityPassed Cytotoxicity, Sensitization, and Irritation tests per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23
    Electrical SafetyCompliant with IEC 60601-1, IEC 60601-1-11, ISO 80601-2-56
    Electromagnetic Compatibility (EMC)Compliant with IEC 60601-1-2
    Operating EnvironmentsTemperature: 15°C~ 40°C Humidity: ≤ 85% RH, non-condensing Atmospheric pressure: 70–106 kPa
    Transport and Storage EnvironmentsAmbient Temperature: -20°C to 55°C Relative Humidity: ≤ 93% RH, non-condensing Atmospheric pressure: 50 kPa to 106 kPa
    Regulatory ComplianceCompliant with 21 CFR 880.2910 (Clinical Electronic Thermometer)
    Indications for Use (Patient Population)Measures human body temperature of people over three months from surface of eardrum. (Narrower than predicate, excluding preterm and newborns)

    2. Sample size used for the test set and the data provenance:

    • Non-clinical Data (Bench Testing): The document does not specify a "sample size" in terms of number of devices for bench testing. It states that "The device has been tested according to the following standards," implying tests were performed on representative samples to ensure compliance.
    • Clinical Data: The summary states, "The clinical testing has been conducted per ISO 80601-2-56 Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement." However, the specific sample size (number of patients/measurements) and data provenance (e.g., country of origin, retrospective/prospective) for this clinical study are NOT provided in the given document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Clinical Data: The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical study. As it's an ear thermometer, the "ground truth" would typically come from a reference temperature measurement method (e.g., rectal thermometry) rather than expert interpretation of an image or signal.

    4. Adjudication method for the test set:

    • Clinical Data: Given that the device measures a quantitative value (temperature), an adjudication method in the context of expert consensus (like multiple readers for an image) is generally not applicable. The comparison would be between the device's reading and the reference method's reading. The document does not describe any specific adjudication method for the clinical test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was NOT done. This type of study (MRMC) is relevant for diagnostic imaging devices where human interpretation plays a significant role, sometimes aided by AI. This document is for an ear thermometer, a direct measurement device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance was done for the ear thermometer. The device itself is designed to provide temperature readings directly, without requiring human interpretation or input to calculate the core measurement. The bench and clinical testing described aim to quantify this standalone performance in terms of accuracy and adherence to standards.

    7. The type of ground truth used:

    • Clinical Data: While not explicitly stated, for a clinical thermometer, the ground truth for performance testing is typically established by comparison against a standardized, highly accurate reference thermometer (e.g., a rectal thermometer or an oral thermometer with known accuracy) or a specialized blackbody calibrator in a controlled environment as specified by standards like ISO 80601-2-56. The document mentions "clinical accuracy test" and "performance test," strongly implying such a comparative methodology.

    8. The sample size for the training set:

    • Not applicable / Not provided. The device is an ear thermometer, which is a sensor-based measurement tool, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. The device's performance is determined by its design, calibration, and adherence to physical principles and engineering specifications, which are validated through bench and clinical testing.

    9. How the ground truth for the training set was established:

    • Not applicable. As noted above, the concept of a "training set" and its associated ground truth establishment is not relevant for this type of medical device submission.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.

    Device Description

    The Wrist Blood Pressure Monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa.

    All the models included in this submission follow the the same intended use, same measurement principle, same blood pressure core algorithm and similar product design. All the models can be used with one cuff size 13.5~19.5 cm (5.3-7.7inches).

    The main differences are appearance, Dimensions and some specifications which will not affect the safety and effectiveness of the device.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated summary pertain to a Wrist Blood Pressure Monitor, which is a medical device for measuring blood pressure and pulse rate. It is not an AI/Software as a Medical Device (SaMD). Therefore, many of the typical acceptance criteria and study designs associated with AI/SaMD (such as multi-reader multi-case studies, ground truth establishment by experts, training set details, or effect sizes of AI assistance) are not applicable to this device.

    The acceptance criteria and study details provided are tailored to the performance of a non-invasive blood pressure measurement system (hardware device), focusing on accuracy, safety, and effectiveness.

    Here's a breakdown of the requested information based on the provided document, addressing the device's specific characteristics as a hardware blood pressure monitor:


    Acceptance Criteria and Device Performance (Wrist Blood Pressure Monitor)

    1. Table of Acceptance Criteria and Reported Device Performance

    As per the 510(k) summary, the device's accuracy is a key performance metric. The acceptance criteria are based on the international standard ISO 81060-2 Third edition 2018-11 [Including AMD1:2020].

    Performance MetricAcceptance Criteria (from ISO 81060-2)Reported Device Performance
    Blood Pressure AccuracyMean error and standard deviation of differences for systolic and diastolic pressure not over the limits specified in ISO 81060-2.All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2.
    Heart Rate Accuracy± 5% of reading± 5% of reading (Same as Predicate, implying met for proposed)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set (Clinical Accuracy Study):
      • Three groups of clinical accuracy research were conducted. Each group included 100 subjects, for a total of 300 subjects across the 13 models.
      • Group 1: 100 subjects (47 Male, 53 Female)
      • Group 2: 100 subjects (54 Male, 46 Female)
      • Group 3: 100 subjects (44 Male, 56 Female)
      • Minimum subjects for each group was 85, as per ISO 81060-2.
    • Data Provenance: The document does not explicitly state the country of origin for the clinical data. However, the manufacturer is "Shenzhen AOJ Medical Technology Co., Ltd." in Shenzhen, Guangdong, China. It is highly probable the data was collected in China.
    • Retrospective or Prospective: The clinical accuracy study, designed to meet ISO 81060-2, is typically conducted prospectively to collect new data for device validation. The wording "clinical accuracy research" and "clinical accuracy test report and data analysis followed the requirements" implies a prospective study.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • This question is not directly applicable in the context of this device because "ground truth" for a blood pressure monitor's accuracy is established against a reference standard, not through expert consensus on interpretations of images or signals (as would be the case for AI/SaMD).
    • For blood pressure monitors, the "ground truth" or reference measurement is typically taken by trained medical professionals using a standardized reference sphygmomanometer (e.g., mercurial or auscultatory method), as per the ISO 81060-2 standard. The document states "The Same Arm Sequential Method was chosen for all studies," which is a standard procedure comparison method against a reference device. The qualifications of the individuals performing these reference measurements would be trained clinicians (e.g., physicians, nurses).

    4. Adjudication Method for the Test Set

    • This question is not applicable for a blood pressure monitor's accuracy testing. Adjudication methods (like 2+1 or 3+1) are used to resolve discrepancies in human expert interpretations, especially in image-based diagnostics.
    • For blood pressure accuracy, deviations are quantified statistically between the device reading and the reference measurement, not through an adjudication process among multiple "readers."

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, this was not done. This type of study is relevant for AI/SaMD devices where AI assists human interpretation and is a key component for assessing the AI's clinical utility. The Wrist Blood Pressure Monitor is a standalone hardware device that provides a measurement; it does not involve human "readers" interpreting data or AI assistance.

    6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    • Yes, in essence. The entire clinical accuracy study described in Section 8 ("Clinical data") evaluates the device's performance (which incorporates its internal oscillometric algorithm) in a standalone manner against a reference standard. The "algorithm" here refers to the embedded software that processes the oscillometric signals to derive blood pressure and pulse rate. The study directly assesses how accurately the device (with its integrated algorithm) measures blood pressure readings compared to the reference.
    • Performance Metrics: The evaluation was based on the "mean error and standard deviation of differences for systolic, diastolic pressure" as per ISO 81060-2.

    7. The Type of Ground Truth Used

    • The ground truth for the clinical accuracy testing was established through concurrent measurements using a standardized reference method (e.g., auscultatory method with a mercurial sphygmomanometer or another validated reference device) on the same arm, sequentially with the test device. This is the standard for blood pressure monitor validation as per ISO 81060-2.
    • It is not "expert consensus" in the sense of subjective medical interpretation, but rather an objective, standardized measurement performed by trained personnel using a calibrated reference instrument.

    8. The Sample Size for the Training Set

    • This concept is not applicable to this type of medical device clearance. The Wrist Blood Pressure Monitor is a hardware device with an embedded algorithm (oscillometric technique) that is based on established physiological principles. It doesn't use machine learning or deep learning in a way that requires a separate "training set" of patient data for an AI model to learn from, as would be the case for AI/SaMD devices. The device's "training" (development and calibration) would involve engineering principles and laboratory testing, rather than a data-driven machine learning process.

    9. How the Ground Truth for the Training Set Was Established

    • As the concept of a "training set" in the context of machine learning is not applicable here (see point 8), the establishment of ground truth for such a set is also not applicable. The device's underlying measurement principle is well-established oscillometric technology. Development and calibration rely on physical models and engineering validation.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232128
    Date Cleared
    2023-10-14

    (89 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared thermometer is intended to measure human body temperature of people over three months old from the eardrum or forehead. It is indicated to be used in homecare and healthcare environments.

    Device Description

    The Infrared thermometer is a handheld device, which can measure human body's temperature either via the eardrum or the forehead for clinical or home use. The results can be displayed on LCD and can also be transmitted to a non-device APP to record and display through Bluetooth.
    There are two models AOJ-20A and AOJ-20Y. Both models share the same construction except differences on product enclosure,PCB layout and button designs.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for an Infrared Thermometer (AOJ-20A; AOJ-20Y). It details the device, its intended use, and the non-clinical testing performed to establish substantial equivalence to predicate devices.

    However, the document does not contain the level of detail typically found in a clinical study report or a more comprehensive validation plan. Specifically, it lacks:

    • Explicit acceptance criteria in a tabular format with corresponding performance results.
    • Specific sample sizes for test sets beyond general statements of testing being conducted.
    • Data provenance (country, retrospective/prospective) for clinical data.
    • Number and qualifications of experts for ground truth.
    • Adjudication methods.
    • Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone algorithm performance details.
    • Specific details on the type of ground truth (e.g., pathology, outcomes data) beyond "performance effectiveness."
    • Training set sample size and ground truth establishment for training.

    The document primarily focuses on non-clinical performance data to support the substantial equivalence claim, rather than a clinical study demonstrating acceptance criteria met in a real-world setting with human subjects. The tests mentioned are:

    • Biocompatibility testing: Cytotoxicity, Skin Sensitization, Skin Irritation per ISO 10993-1.
    • Electrical safety and electromagnetic compatibility (EMC): Compliance with IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2.
    • Bench Testing: Compliance with ISO 80601-2-56 for performance effectiveness.
    • Software Verification and Validation Testing: Conducted as per FDA's "Content of Premarket Submissions for Device Software" guidance, with the device considered "Basic" documentation level.

    Given these limitations in the provided text, I can only extract the general types of tests and the standards they comply with, rather than specific acceptance criteria values and the device's measured performance against them. The information on "study that proves the device meets the acceptance criteria" is limited to the types of non-clinical tests conducted.

    Therefore, direct answers to many of your questions cannot be provided from the given document as it does not contain clinical study data or a detailed validation report with specific performance metrics against acceptance thresholds.

    Based on the provided text, here is what can be inferred and what is explicitly stated:


    1. A table of acceptance criteria and the reported device performance

    The document references compliance with ISO 80601-2-56 for "performance effectiveness" in bench testing. This standard likely specifies accuracy and precision requirements for clinical thermometers. However, the specific numerical acceptance criteria (e.g., ±0.2°C) and the actual measured performance values of the device against these criteria are not provided in this summary.

    Inferred/General:

    Acceptance Criteria (Standard)Reported Device Performance
    Performance EffectivenessComplies with ISO 80601-2-56 (details not provided)
    BiocompatibilityPasses Cytotoxicity, Skin Sensitization, Skin Irritation per ISO 10993-1
    Electrical Safety & EMCComplies with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2
    Software V&VTesting conducted, documentation provided per FDA guidance for "Basic" level device software

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not specified. The document states "Bench testing was conducted on the thermometer device, consisting of all the system." This suggests testing of the physical devices, but no numerical sample count (e.g., number of thermometers tested, number of measurements taken) is given.
    • Data provenance: Not specified. The tests are non-clinical bench and lab tests, not clinical human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable as the testing described is non-clinical/bench testing, not involving human expert interpretation for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a medical device (thermometer), not an AI imaging or diagnostic algorithm used by human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The document describes the performance of the device itself (sensor, electronics, software) in a standalone manner via bench testing, but not an "algorithm only" in the sense of a standalone AI model. The device's function is temperature measurement.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For "performance effectiveness," the ground truth (or reference standard) would typically be calibrated reference thermometers or temperature sources compliant with the ISO standard used for test and measurement, not clinical outcomes or expert consensus.
    • For biocompatibility, the ground truth is the chemical and biological assays.
    • For electrical safety and EMC, the ground truth is the measurement against standard limits using appropriate test equipment.

    8. The sample size for the training set

    • Not applicable, as this is a traditional medical device (thermometer), not an AI/machine learning model requiring a training set for its core function. The "software verification and validation testing" refers to standard software engineering practices, not machine learning model training.

    9. How the ground truth for the training set was established

    • Not applicable for the same reasons as point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222994
    Date Cleared
    2023-07-26

    (301 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arm Blood Pressure Monitor is intended to measure the systolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique at medical facilities or at home.

    Device Description

    AOJ-33series arm blood pressure monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or Kpa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use. All the models included in this submission follow the same software, same measurement principle and same specifications. The main differences are color of the face shell and keys, which will not affect the safety and effectiveness of the device. AOJ-33B include the Bluetooth transmission functionality which can transfer data for Application in the external instruments, and the measuring data, including systolic diastolic pressures and pulse rate can be displayed, stored and reviewed by the Application in the external instruments without any control feature, therefore, no interoperability happened. This function is not available for AOJ-33A.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the Arm Blood Pressure Monitor, models AOJ-33A and AOJ-33B, meets these criteria.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The core performance acceptance criteria for blood pressure monitors are typically defined by international standards like ISO 81060-2. The document states that the device was tested to ISO 81060-2:2018.

    Acceptance Criteria (from ISO 81060-2:2018; implicitly accepted by FDA clearance)Reported Device Performance (from Clinical Study on AOJ-33A)
    Mean error and standard deviation of differences for systolic pressure within specified limits of ISO 81060-2:2018Met: "All data's mean error and standard deviation of differences for systolic... pressure is not over the limits of ISO 81060-2:2018."
    Mean error and standard deviation of differences for diastolic pressure within specified limits of ISO 81060-2:2018Met: "All data's mean error and standard deviation of differences for ...diastolic pressure is not over the limits of ISO 81060-2:2018."
    Blood pressure measurement range: $\pm$ 3 mmHgMet: "$\pm$ 3 mmHg" (This is listed under "Accuracy" in the comparison table, implying it's the expected accuracy).
    Heart rate measurement accuracy: $\pm$ 5% of readingMet: "$\pm$ 5% of reading" (This is listed under "Accuracy" in the comparison table).

    2. Sample size used for the test set and the data provenance

    • Sample Size: 100 adult subjects (49 female, 51 male).
    • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a clinical validation study performed on AOJ-33A, implying it was a prospective study conducted for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document refers to a "Same Arm Sequential Method" clinical validation study in accordance with ISO 81060-2:2018. This standard typically involves multiple trained observers (experts) taking sequential measurements using a reference sphygmomanometer. However, the document does not specify the exact number of experts or their qualifications.

    4. Adjudication method for the test set

    The document states "The Same Arm Sequential Method was chosen." This method, as described in ISO 81060-2, involves comparing the automated device measurements to those taken by trained observers using a reference manual sphygmomanometer on the same arm. While it implies a comparison against expert readings, the document does not explicitly detail an adjudication method (e.g., 2+1, 3+1). The standard outlines procedures for obtaining reference measurements, which inherently involve a form of expert consensus or highly standardized measurement to establish ground truth.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable as the device is an automated blood pressure monitor and not an AI-assisted diagnostic tool for human readers. It directly measures and displays blood pressure and pulse rate.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done. The clinical validation study described is for the device (AOJ-33A) functioning independently to measure blood pressure and pulse rate, without human interpretation or assistance during the measurement process.

    7. The type of ground truth used

    The ground truth for the clinical study was established by comparison with established reference measurements, likely from a manual sphygmomanometer operated by trained observers, as per the ISO 81060-2:2018 standard for non-invasive sphygmomanometers. This effectively represents a validated, expert-derived measurement.

    8. The sample size for the training set

    The document does not specify a separate training set sample size. The clinical data mentioned (100 adult subjects) refers to the testing or validation set for the device's performance against the standard. For medical devices like blood pressure monitors, the "training" (i.e., calibration and algorithm development) often occurs internally during the device design phase, and explicit "training set" data for regulatory submission in the same way as an AI algorithm is typically not detailed.

    9. How the ground truth for the training set was established

    As there is no explicit mention of a separate "training set" in the context of data for regulatory submission (beyond the device's inherent design and calibration), this information is not provided in the document. The general principle for developing such devices involves engineering and clinical testing to ensure accurate readings, which implicitly leverages previous data and established hemodynamic principles to calibrate the oscillometric algorithm.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared thermometer is intended to measure human body temperature of people over three months old from the eardrum or forehead. It is indicated to be used in homecare and healthcare environments.

    Device Description

    The AOJ-20 series Infrared thermometer is a handheld device, which can measure human body's temperature either via the eardrum or the forehead, for clinical or home use. The results can be displayed on LCD. The device operates in adjusted mode, and the reference body site of output temperature is oral. The measurement distance for forehead mode is 0 cm ~ 3 cm. The thermometers are powered by AAA 1.5V×2 alkaline batteries. A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the eardrum or the surface of the skin of forehead, which is converted to the equivalent oral temperature with the unit of °C or ºF. All the models share the similar design and the same critical components. The major differences include: - Mechanical changes and the corresponding hardware adjustment happened to AOJ-20A, AOJ-20B, AOJ-20C, AOJ-20D, AOJ-20E, AOJ-20F, AOJ-20H, AOJ-20M, AOJ-20T, AOJ-20R and AOJ-20Y - In the measurement function, two measurement modes are distinguished between children and adults for AOJ-20B and AOJ-20E.

    AI/ML Overview

    The provided text describes the regulatory clearance of an Infrared Thermometer (AOJ-20 series) and includes detailed performance testing against acceptance criteria. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is an Infrared Thermometer, and its primary performance criterion is accuracy. The relevant standard for accuracy is ISO 80601-2-56.

    Acceptance Criteria (from ISO 80601-2-56)Reported Device Performance
    Accuracy: ±0.2°C (±0.4°F)±0.2°C (±0.4°F)
    Biocompatibility: No cytotoxicityNo potential cytotoxicity
    Biocompatibility: No sensitizationNo sensitization observed
    Biocompatibility: No irritationNegligible irritation
    Electrical Safety & EMC: Compliance with specific IEC standards (60601-1, 60601-1-11, 60601-1-2)Device complies with specified IEC standards
    Software Verification & Validation: Software functions meet requirementsTest results demonstrated software functions met requirements
    Minimum Subject Numbers for Clinical Accuracy Validation (ISO 80601-2-56 Section 201.102)All models met minimum subject number requirements for each age group (refer to table in source for specific counts)

    The text explicitly states: "The system complies with the ISO 80601-2-56 Medical electrical equipment — Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement for performance effectiveness." and "Clinical accuracy validation was carried out on people over three month indicated in the instructions for use. The number of subjects in each age group met the minimum number requirements."

    2. Sample Size Used for the Test Set and Data Provenance

    The "test set" in this context refers to the clinical study subjects for accuracy validation.

    • Sample Size: The sample sizes vary per model, but generally around 135-138 subjects per model. The specific breakdown is provided in the table:
      • AOJ-20A: 138 subjects
      • AOJ-20B: 136 subjects
      • AOJ-20C: 137 subjects
      • AOJ-20D: 138 subjects
      • AOJ-20E: 137 subjects
      • AOJ-20F: 137 subjects
      • AOJ-20H: 135 subjects
      • AOJ-20M: 137 subjects
      • AOJ-20T: 137 subjects
      • AOJ-20R: 137 subjects
      • AOJ-20Y: 137 subjects
    • Data Provenance: The document does not specify the country of origin for the clinical data. It also does not explicitly state whether the study was retrospective or prospective, but clinical accuracy validation per ISO standards is typically done in a prospective manner.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given document. For clinical thermometers, ground truth for body temperature is typically established using a highly accurate reference thermometer (e.g., rectal or oral thermometer depending on the standard's requirements for reference body site), not necessarily by human experts.

    4. Adjudication Method for the Test Set

    This information is not provided in the given document. Adjudication methods like 2+1 or 3+1 are more common in studies involving subjective assessments (e.g., image interpretation by radiologists), which is not directly applicable to a thermometer's clinical accuracy validation where objective measurements against a reference are typically performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or performed. This type of study is more relevant for diagnostic aids where human interpretation is involved. The document describes clinical accuracy validation of the device itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The device is an Infrared Thermometer, which is a standalone measurement device. Its performance is inherently standalone. There is no "algorithm only" component separate from human interaction, as a human user operates the device to take a temperature reading. The clinical and bench testing described assess this standalone performance.

    7. The Type of Ground Truth Used

    The ground truth used for clinical accuracy validation of a thermometer is typically based on a reference body temperature measurement obtained by a highly accurate, calibrated reference thermometer (e.g., a specific type of oral or rectal thermometer). The document states the device "operates in adjusted mode, and the reference body site of output temperature is oral," implying oral temperature is the reference for adjusted readings. The study was conducted "per Section 201.102 of ISO 80601-2-56," which would define the exact method for establishing ground truth.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. Infrared thermometers are hardware devices with embedded algorithms, but they are not typically "trained" on large datasets of temperature readings in the way AI/ML models are. The device relies on physical principles of infrared detection and calibrated conversion algorithms, not a training set in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no mention of a "training set" in the context of this infrared thermometer. The device's internal algorithms and calibrations are developed and verified through engineering principles and bench testing, not machine learning training on a ground truth dataset.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222125
    Date Cleared
    2022-12-01

    (136 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Arm Blood Pressure Monitor is intended to measure the systolic blood pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique in which an inflatable CUFF is wrapped around the upper arm at medical facilities or at home.

    Device Description

    AOJ-30 series Arm blood pressure monitor is designed as a battery driven automatic non-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at upper arm within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg or kPa. The device has the data storage function in order for data reviewing, including the systolic pressure, diastolic pressure, pulse rate and measurement time. The device also has low voltage indication, which will be triggered when the battery is low. The proposed device is intended to be used in medical facilities or at home. And the effectiveness of this sphygmomanometer has not been established in pregnant (including pre-eclamptic) patients. The product is provided non-sterile, and not to be sterilized by the user prior to use. All the models included in this submission follow the similar software, same measurement principle and same specifications. The differences existed between different models included in this submission will not affect the safety and effectiveness of the device. AOJ-30A and AOJ-30B include the Bluetooth transmission functionality which can transfer data for Application in the external instruments, and the measuring data, including systolic diastolic pressures and pulse rate can be displayed, stored and reviewed by the Application in the external instruments without any control feature, therefore, no interoperability happened. This function is not available for other models.

    AI/ML Overview

    The provided text describes the 510(k) submission for the Shenzhen AOJ Medical Technology Co., Ltd. Arm Blood Pressure Monitor (AOJ-30 series). This document focuses on demonstrating the substantial equivalence of the new device to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria in the context of an AI/ML study.

    Therefore, many of the requested details, such as those related to AI/ML performance (e.g., sample size for training set, number of experts for ground truth, MRMC study, standalone performance), are not applicable to this document as it pertains to a traditional non-invasive blood pressure monitor.

    However, I can extract information related to the performance testing and clinical validation for this type of medical device as presented in the document.

    Study Details Proving Device Meets Acceptance Criteria (as per the provided document)

    The document describes the performance testing undertaken to demonstrate substantial equivalence to a predicate device, focusing on accuracy, safety, and effectiveness for a non-invasive blood pressure monitor.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core "acceptance criteria" for a non-invasive oscillometric sphygmomanometer like this are typically defined by international standards for accuracy. The document explicitly references these standards for performance effectiveness.

    CriterionAcceptance Standard (as per referenced ISO 80601-2-30)Reported Device Performance (as per comparison table)
    Blood Pressure Measurement Accuracy$\pm$ 3 mmHg (standard)$\pm$ 3 mmHg
    Heart Rate Measurement Accuracy$\pm$ 5% of reading (standard)$\pm$ 5% of reading
    BiocompatibilityMeets ISO 10993-1, 5, 10 for Cytotoxicity, Skin Sensitization, Skin IrritationTesting conducted per standards, passed
    Electrical SafetyMeets IEC 60601-1, IEC 60601-1-11, ISO 80601-2-30Compliance stated
    Electromagnetic Compatibility (EMC)Meets IEC 60601-1-2Compliance stated
    Software Verification & Validation TestingAligned with FDA Guidance for "Major" Level of ConcernTesting conducted, documentation provided

    Note: The performance values ($\pm$ 3 mmHg for BP, $\pm$ 5% for HR) are commonly accepted accuracy requirements for non-invasive oscillometric blood pressure monitors as per standards like ISO 81060-2. The document states the device has this accuracy, implying it meets these criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Clinical Validation: "100 subjects were involved for each study" (for the upgraded AOJ-30A, AOJ-30C, AOJ-30D, and AOJ-30E models).
    • Data Provenance: Not explicitly stated regarding country of origin, but the submitting company is Shenzhen AOJ Medical Technology Co., Ltd. in China. The study is described as "clinical testing... conducted per IEC 81060-2: 2013." The nature of the study is prospective clinical validation as it describes involving subjects and conducting tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The document implies ground truth (reference blood pressure measurements) would be established by experts as per the standard ISO 81060-2:2018. This standard typically requires a certain number of qualified observers (e.g., trained healthcare professionals) taking reference measurements. However, the exact number of experts and their specific qualifications are not explicitly detailed in this document. It only states: "Observer preparation, Reference determination, Clinical investigation methods and Data analysis are referred to ISO 81060-2: 2018 without any deviation."

    4. Adjudication Method for the Test Set

    • The document refers to ISO 81060-2:2018 for "Reference determination" and "Clinical investigation methods." This standard outlines specific methodologies for obtaining simultaneous reference blood pressure measurements (e.g., using a mercury sphygmomanometer or an automated reference device), which inherently involves a form of "ground truth" or reference standard, but it's not an "adjudication method" in the sense of multiple independent readers evaluating images, as would be the case for an AI/ML study.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML devices where human readers' performance with and without AI assistance is evaluated. This document concerns a standalone blood pressure monitor.

    6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, the core of the clinical validation described is a standalone performance test of the device's measurement accuracy against a reference standard. The device itself is an automated blood pressure monitor without human interpretation requirements beyond reading the displayed values. The accuracy values ($\pm$ 3 mmHg for BP, $\pm$ 5% for HR) are standalone performance metrics.

    7. The Type of Ground Truth Used

    • The ground truth for the clinical validation of blood pressure devices is established through simultaneous, direct measurements by a highly accurate reference method (e.g., auscultatory method with trained observers using a mercury sphygmomanometer, or another validated reference device) conducted according to the specified international standard (ISO 81060-2:2018). This is implied by the statement "Reference determination... referred to ISO 81060-2: 2018 without any deviation." It's essentially "expert consensus" in the sense of adhering to a standardized clinical protocol for reference measurements.

    8. The Sample Size for the Training Set

    • Not applicable. This document describes the validation of a hardware-based blood pressure monitor (oscillometric technique), not an AI/ML device that requires a "training set" in the machine learning sense. The device's algorithm for blood pressure determination is pre-सेट and does not undergo "training" on a dataset in the way an AI model would.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no "training set" for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221170
    Date Cleared
    2022-10-07

    (168 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The thermometer is intended to measure human body temperature of people over one month from surface of forehead. It is a non-sterile, reusable, non-contact and hand-held device. It is indicated to be used in homecare and healtheare environments.

    Device Description

    The Forehead thermometer is a handheld device, which can measure human body's temperature from the forehead for clinical or home use. The results can be displayed on LCD. The measurement is non-contact with a distance of 3-5 cm to measure the temperature. The thermometers are powered by AAA 1.5V×2 alkaline batteries, which can be used for people over one month. A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the surface skin of the forehead, which is converted to a body temperature with the unit of °C or °F. The reference body site of the output temperature is oral. All the models share the similar design and the same critical components, and compose of a sensor, PCB, buttons, LCD display and housing. Functions include temperature measurement, memory reading recall, voice mute/unmute and unit/mode switch, low battery detection and high temperature indicator.

    AI/ML Overview

    The provided text describes the acceptance criteria and a study demonstrating that the device meets these criteria. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Subject Device)Reported Device Performance (Subject Device)
    Measurement Range: 32.0°C ~ 42.9°CEvaluated per ISO 80601-2-56; test results met requirements.
    Accuracy: ±0.2°CEvaluated per ISO 80601-2-56; test results met requirements.
    Memory Data Limit: 40 valuesVerified during design and development; test results met requirements.
    Operating Environments (Temperature): 15°C ~ 40°CPerformance testing conducted under suggested environment; test results met requirements.
    Biocompatibility (In vitro Cytotoxicity): No potential cytotoxicity allowedNo potential cytotoxicity
    Biocompatibility (Skin Sensitization): No sensitization should be observedNo sensitization observed (test sample score 0)
    Biocompatibility (Skin Irritation): No irritation should be observedNegligible (no observed primary irritation, test sample score 0)
    Electrical Safety & EMC: Compliance with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2Complies with specified standards
    Performance Effectiveness: Compliance with ISO 80601-2-56Complies with ISO 80601-2-56
    Software Verification & Validation: Software functions meet requirementsSoftware functions met requirements
    Clinical Accuracy Validation: Meet minimum number requirements for subjects in each age group (A1, A2, B, C)Number of subjects in each age group met minimum requirements.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size for Clinical Accuracy Validation:
      • AOJ-F101: 153 subjects
      • AOJ-F102: 142 subjects
      • AOJ-F103: 132 subjects
    • Data Provenance: The document (K221170) is a 510(k) summary submitted to the FDA by a Chinese company (Shenzhen AOJ Medical Technology Co., Ltd. in Shenzhen, China). The clinical testing was explicitly stated as "Clinical accuracy validation was carried out on people over one month indicated in the instructions for use." While not explicitly stated, clinical studies for FDA submissions typically involve prospective data collection to demonstrate performance. The country of origin for the clinical data is not specified, but usually, it aligns with the manufacturer's location or regions where clinical trials are feasible and recognized for regulatory purposes.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. For a thermometer, "ground truth" for temperature is typically established using a highly accurate reference thermometer (e.g., a rectal thermometer for core body temperature, used in a clinical setting), rather than expert consensus on interpretation.

    4. Adjudication method for the test set:

    This information is not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments where multiple human readers interpret medical images or data. For a thermometer's accuracy, the ground truth is usually a precise measurement from a reference device, making such adjudication unnecessary.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device is a thermometer, a measurement tool, not an AI-assisted diagnostic device that aids human readers in interpreting complex cases. Therefore, an MRMC study comparing human reader performance with and without AI assistance was not conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device itself is a standalone measurement device. The "Clinical data" section describes a clinical accuracy validation study to confirm its performance directly against a clinical reference for body temperature, which is a form of standalone performance evaluation for a thermometer. There is no "algorithm only" in the context of an AI device here, but rather device-only performance for a physical product.

    7. The type of ground truth used:

    For the clinical accuracy validation, the ground truth for body temperature measurements would have been established using a highly accurate reference clinical thermometer (e.g., an oral or rectal thermometer, depending on the standard protocol for comparison) in a controlled clinical setting. The document specifies that clinical testing was conducted "per Section 201.102 of ISO 80601-2-56," which outlines the requirements for clinical investigations of medical electrical equipment used for body temperature measurement. This standard details how to establish accurate reference temperatures.

    8. The sample size for the training set:

    This information is not provided in the document. The document describes premarket testing and validation for a physical medical device, not a machine learning model that requires a training set. The "software verification and validation testing" refers to traditional software engineering V&V, not AI model training.

    9. How the ground truth for the training set was established:

    This information is not applicable as there is no mention of a training set for a machine learning model. The device is a conventional electronic thermometer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213485
    Date Cleared
    2022-05-25

    (208 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AOJ-25 series Digital Thermometer (except AOJ-25B) are designed as reusable battery-operated electronic device, and intended for the measurement of oral, armpit and rectal temperature for people of all ages at home.
    The AOJ-25B is intended to measure temperature rectally only for people of all ages at home as well. The device is reusable.

    Device Description

    The digital thermometer is designed as hand-held device which can measure human body's temperature orally, axillary (under the arm), or rectally. The results can be displayed on LCD.
    The digital thermometers have several functions, such as beep indication, °C and °F unit switchable, low battery detection, memories, three-color backlight, auto automatic power off functions.
    The device is a predictive digital thermometer. Users only need 10 seconds for quick reading, and will display the temperature value on the screen.

    AI/ML Overview

    The provided FDA 510(k) K213485 submission describes the acceptance criteria and the study that proves the Digital Thermometer, Models: AOJ-25A, AOJ-25B, AOJ-25C, AOJ-25D and AOJ-25E meets these criteria.

    Here's a breakdown of the requested information:

    1. Table of acceptance criteria and the reported device performance

    The submission primarily focuses on the clinical accuracy of the thermometer as a key performance criterion, referencing the ISO 80601-2-56 standard.

    Acceptance Criteria (from ISO 80601-2-56)Reported Device Performance
    Clinical accuracy of the subject device is at least the same as that of the reference device.Demonstrated to be accurate within ±0.1°C as claimed in the manual.
    No single measurement error exceeding the allowable limit.Achieved with 99% confidence. Includes a reference device comparison.

    Note: The document also mentions other performance aspects like electrical safety, EMC, and biocompatibility, which are met through compliance with relevant IEC and ISO standards. However, specific "acceptance criteria" and "reported performance" are not detailed in a table format for these aspects within the provided text, beyond stating compliance.

    2. Sample size used for the test set and the data provenance

    • Sample Size: A minimum of 90 subjects were used for the clinical accuracy study.
    • Data Provenance: The document does not explicitly state the country of origin. It indicates a "clinical study" was conducted, referring to it as a "randomization, simple blind homologous control, pairing design of clinical investigation." It is implied to be a prospective study due to the nature of a clinical investigation outlined by the standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish the ground truth. It states that a "reference device is introduced as required by the standard," implying that the ground truth for temperature measurements during the clinical study was established by this validated reference device, not by human expert interpretation.

    4. Adjudication method for the test set

    The document mentions a "simple blind homologous control, pairing design of clinical investigation." This suggests a comparative design with a reference device. It does not describe an explicit adjudication method involving multiple human readers for differing interpretations, as the ground truth is based on the objective measurement of temperature by a reference device according to the standard.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This device is a digital thermometer, not an imaging device requiring human interpretation of clinical images. The study focuses on the accuracy of the temperature measurement itself, not on human readers' interpretation of output.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the clinical accuracy study for the digital thermometer is essentially a standalone performance evaluation of the device in measuring temperature. While the device is used by a human, the performance measured is the accuracy of the device's output against a reference, not its performance in assisting human decision-making or interpretation.

    7. The type of ground truth used

    The ground truth for the clinical accuracy study was established by a reference device as required by the ISO 80601-2-56 standard. This reference device is used to measure the body temperature against which the subject device's measurements are compared.

    8. The sample size for the training set

    The document does not provide information about a separate "training set" or its sample size. For medical devices like digital thermometers, the development process typically involves calibration and validation against known standards, which is distinct from the machine learning "training set" concept often associated with AI/ML devices. The "clinical study" described serves as the validation of the final algorithm's accuracy.

    9. How the ground truth for the training set was established

    As there is no mention of a traditional machine learning training set, the concept of establishing ground truth for it is not applicable here. The device's underlying "algorithm" (firmware) related to temperature measurement would be calibrated and developed based on the physical principles of thermistor resistance and validated through bench testing and the described clinical study against a precise reference thermometer.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213503
    Date Cleared
    2022-02-18

    (109 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Wrist Blood Pressure Monitor is intended to measure and diastolic pressure and diastolic pressure, as well as the pulse rate of adult person via non-invasive oscillometric technique by an inflatable cuff wrapped around the wrist at medical facilities or at home.

    Device Description

    AOJ-35 series wrist blood pressure monitor is designed as a battery driven automatic on-invasive blood pressure monitor. It can automatically complete the inflation, deflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult person at wrist within its claimed range and accuracy via the oscillometric technique. The result will be displayed in the international unit mmHg. The device also has low voltage indication, which will be triggered when the battery is low. All the models included in this submission follow the same software, same measurement principle and same specifications. The main differences are color of the face shell and keys, which will not affect the safety and effectiveness of the device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (from ISO 80601-2-30 & ISO 81060-2:2013)Reported Device Performance (as per Clinical Data)
    Blood Pressure Accuracy: Mean value of difference between device and reference standard should be less than 5.0 mmHg.Mean value is less than 5.0 mmHg
    Blood Pressure Accuracy: Standard deviation of difference between device and reference standard should be less than 8.0 mmHg.Standard deviation is less than 8.0 mmHg
    Cuff Size Distribution Requirement (ISO 81060-2:2013 clause 5.1.4):
    - Upper half of the cuff: At least 40%54% (Complied)
    - Lower half of the cuff: At least 40%46% (Complied)
    - Upper quarter of the cuff: At least 20%29% (Complied)
    - Lower quarter of the cuff: At least 20%24% (Complied)
    - Upper octal of the cuff: At least 10%11% (Complied)
    - Lower octal of the cuff: At least 10%10% (Complied)
    Blood Pressure Distribution Requirement (ISO 81060-2:2013 clause 5.1.5):
    - Systolic Blood Pressure ≤100 mmHg: At least 5%12% (Complied)
    - Systolic Blood Pressure ≥160 mmHg: At least 5%13% (Complied)
    - Systolic Blood Pressure ≥140 mmHg: At least 20%33% (Complied)
    - Diastolic Blood Pressure ≥100 mmHg: At least 5%16% (Complied)
    - Diastolic Blood Pressure ≤60 mmHg: At least 5%10% (Complied)
    - Diastolic Blood Pressure ≥85 mmHg: At least 20%38% (Complied)
    Non-Invasive Sphygmomanometer Performance EffectivenessComplies with ISO 80601-2-30
    Electrical SafetyComplies with IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2
    BiocompatibilityComplies with ISO 10993-1, ISO 10993-5, ISO 10993-10 (Cytotoxicity, Skin Sensitization, Skin Irritation)
    Software Verification and ValidationPassed according to FDA Guidance for "major" level of concern

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: 100 subjects (49 female, 51 male).
    • Data Provenance: The document does not explicitly state the country of origin of the data. It is a prospective clinical study, as it was conducted "per ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    The document mentions that the clinical testing was conducted "per ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type." This standard typically outlines the requirements for validation of automated non-invasive sphygmomanometers against a reference standard. While the document doesn't explicitly state the number of experts, ISO 81060-2 typically requires measurements by trained observers using a mercury-column or auscultatory device as the reference ground truth. The qualifications of these observers would be that they are trained to accurately perform manual blood pressure measurements according to the standard. No specific details about the number or precise qualifications (e.g., "radiologist with 10 years of experience") are provided in this summary, but the reference to the ISO standard implies adherence to its requirements for ground truth establishment.

    4. Adjudication Method for the Test Set:

    The document does not explicitly mention an adjudication method like 2+1 or 3+1. For blood pressure validation, ISO 81060-2 typically involves simultaneous or closely timed measurements by multiple trained observers (often two) and the device under test. The differences between the device readings and the average or adjudicated readings of the observers (or a specific protocol for handling discrepancies) form the basis of the accuracy assessment.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted for this device. The study described focuses on the standalone accuracy of the blood pressure monitor against a reference standard, not its impact on human reader performance or diagnostic accuracy.

    6. Standalone Performance Study:

    Yes, a standalone study (algorithm only without human-in-the-loop performance) was done. The clinical data section explicitly details the performance of the "monitor" (the device) in measuring blood pressure and pulse rate against a reference standard. The "Bench Testing" section also describes tests conducted on the device's technical performance.

    7. Type of Ground Truth Used:

    The ground truth used for the clinical study was established according to ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type. This standard mandates a clinical validation process where the automated device's measurements are compared to measurements taken by trained observers using a validated reference method (e.g., auscultation with a mercury sphygmomanometer). This constitutes an expert-based reference standard rather than pathology or outcomes data.

    8. Sample Size for the Training Set:

    The document does not provide information on the sample size for the training set. This is a premarket notification for a medical device, and the focus of the provided summary is on the clinical validation (testing) and non-clinical data, not on the development or training of an AI algorithm (though the device itself likely contains algorithms, their training data is not detailed here).

    9. How the Ground Truth for the Training Set was Established:

    As no training set information is provided, how its ground truth was established is not detailed in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202173
    Device Name
    Pulse Oximeter
    Date Cleared
    2020-12-07

    (126 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The pulse oximeter is a reusable device and intended for spot-checking of oxygen saturation and pulse rate for use with the finger of adult patients in healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.

    Device Description

    The oximeter consists of probe, electronic circuits, and display and plastic enclosures. And one side of probe is designed to locate light emitting diodes and a light detector (called a photo-detector). Red and Infrared lights are shone through the tissues from one side of the probe to the other. Then parts of the light emitted absorbed by blood and tissues. The light absorbed by the blood varies with the oxygen saturation of haemoglobin. After that, the photo-detector detects the light volume transmitted through the tissues which depends on blood pulse, Hereafter, the microprocessor calculates a value for the oxygen saturation (SpO2). The subject device is a reusable device, and need to reprocess as suggested in the user manual after each use. And the device is intended to be used on the finger, and powered by 2*1.5V AAA battery.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on what's available for this pulse oximeter:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (from Predicate Device/Standards)Reported Device Performance (from Subject Device)
    SpO2 Range0% ~ 100%0% ~ 100%
    SpO2 Resolution1%1%
    SpO2 Accuracy70% ~ 100%: ±2%70% ~ 100%: ±2%
    0% ~ 69%: unspecified0% ~ 69%: unspecified
    Pulse Rate Range25 bpm ~ 250 bpm25 bpm ~ 250 bpm
    Pulse Rate Accuracy±2 bpm±2 bpm
    Pulse Rate Resolution1 bpm1 bpm
    BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10 compliantCompliant (Cytotoxicity, Skin Sensitization, Skin Irritation passed)
    Electrical SafetyIEC 60601-1 compliantCompliant
    Electromagnetic Compatibility (EMC)IEC 60601-1-2 compliantCompliant
    General PerformanceISO 80601-2-61 compliantCompliant

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly mentions a "Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment."

    However, the specific sample size for this clinical test set is not provided in the given text.

    The data provenance (country of origin, retrospective/prospective) is not explicitly stated beyond stating it was "clinical testing... in human subjects."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For pulse oximetry, the "ground truth" (reference SpO2) is typically established by arterial blood gas analysis performed by trained medical personnel, but the specifics of who performed it or how many were involved in establishing the ground truth are not detailed.

    4. Adjudication Method for the Test Set

    This information is not provided in the document.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not conducted or reported. This type of study is more relevant for diagnostic imaging interpretation where human readers are involved. For a pulse oximeter, the primary performance evaluation is a direct comparison to a reference standard (arterial blood gas) rather than comparing human interpretation with and without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was done. The entire premise of a pulse oximeter's performance testing (as described by ISO 80601-2-61) is to assess the device's accuracy in measuring SpO2 and pulse rate independently. The "clinical testing... in human subjects" is a standalone performance evaluation against a reference standard.

    7. The Type of Ground Truth Used

    The ground truth used for the clinical testing of SpO2 accuracy is arterial blood gas analysis (ABG). This is inferred from the reference to "evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61," as this standard mandates ABG as the reference method for determining true arterial oxygen saturation for pulse oximeter accuracy studies.

    8. The Sample Size for the Training Set

    This information is not applicable and therefore not provided. Pulse oximeters are based on established biophysical principles (spectrophotometry) and do not typically use AI/machine learning models that require a "training set" in the conventional sense for their core SpO2 and pulse rate algorithms. The device's internal algorithms are based on physics and calibrated using empirical data, but this is distinct from "training data" for a machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as outlined in point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2