K Number
K250470

Validate with FDA (Live)

Date Cleared
2025-06-23

(125 days)

Product Code
Regulation Number
880.2910
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.

Device Description

The ear thermometer is a handheld device that displays the temperature of the measured patient by measuring the thermal radiation of the eardrum. Measurement unit: °C or °F. The results can be displayed on LCD. The thermometers are powered by 1.5V×2 (AAA or AA) alkaline batteries, which can be used for people over three months. A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the eardrum, which is converted into temperature measurement with the unit of °C or °F. All the models share the same critical components, intended use, working principle and similar product design, and compose of a sensor, PCB, buttons, LCD display and housing. Functions include temperature measurement, memory reading recall, unit switch, low battery detection and high temperature indicator.

AI/ML Overview

The FDA 510(k) clearance letter and accompanying 510(k) Summary for the Ear Thermometers (EAR-E101, EAR-E102, EAR-E103) provide information on acceptance criteria and supporting studies.

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on demonstrating substantial equivalence to a predicate device (Jiangsu Yuyue Medical Equipment & Supply Co., Ltd, Infrared Ear Thermometers YHT101 and YHT200, K203583). Thus, the "acceptance criteria" are largely framed as demonstrating equivalence or adherence to relevant standards. The performance data is presented in comparison to the predicate and overall compliance with standards.

Acceptance Criteria (Stated or Implied by Comparison)Reported Device Performance (Subject Device)
Accuracy (Clinical)Measurement Range: 32.0℃~42.9℃ (89.6°F-109.2°F) Accuracy: ±0.2℃/±0.4°F (for 32.0℃ to 42.9℃)
BiocompatibilityPassed Cytotoxicity, Sensitization, and Irritation tests per ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23
Electrical SafetyCompliant with IEC 60601-1, IEC 60601-1-11, ISO 80601-2-56
Electromagnetic Compatibility (EMC)Compliant with IEC 60601-1-2
Operating EnvironmentsTemperature: 15°C~ 40°C Humidity: ≤ 85% RH, non-condensing Atmospheric pressure: 70–106 kPa
Transport and Storage EnvironmentsAmbient Temperature: -20°C to 55°C Relative Humidity: ≤ 93% RH, non-condensing Atmospheric pressure: 50 kPa to 106 kPa
Regulatory ComplianceCompliant with 21 CFR 880.2910 (Clinical Electronic Thermometer)
Indications for Use (Patient Population)Measures human body temperature of people over three months from surface of eardrum. (Narrower than predicate, excluding preterm and newborns)

2. Sample size used for the test set and the data provenance:

  • Non-clinical Data (Bench Testing): The document does not specify a "sample size" in terms of number of devices for bench testing. It states that "The device has been tested according to the following standards," implying tests were performed on representative samples to ensure compliance.
  • Clinical Data: The summary states, "The clinical testing has been conducted per ISO 80601-2-56 Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement." However, the specific sample size (number of patients/measurements) and data provenance (e.g., country of origin, retrospective/prospective) for this clinical study are NOT provided in the given document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Clinical Data: The document does not specify the number of experts or their qualifications for establishing ground truth in the clinical study. As it's an ear thermometer, the "ground truth" would typically come from a reference temperature measurement method (e.g., rectal thermometry) rather than expert interpretation of an image or signal.

4. Adjudication method for the test set:

  • Clinical Data: Given that the device measures a quantitative value (temperature), an adjudication method in the context of expert consensus (like multiple readers for an image) is generally not applicable. The comparison would be between the device's reading and the reference method's reading. The document does not describe any specific adjudication method for the clinical test set.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was NOT done. This type of study (MRMC) is relevant for diagnostic imaging devices where human interpretation plays a significant role, sometimes aided by AI. This document is for an ear thermometer, a direct measurement device, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance was done for the ear thermometer. The device itself is designed to provide temperature readings directly, without requiring human interpretation or input to calculate the core measurement. The bench and clinical testing described aim to quantify this standalone performance in terms of accuracy and adherence to standards.

7. The type of ground truth used:

  • Clinical Data: While not explicitly stated, for a clinical thermometer, the ground truth for performance testing is typically established by comparison against a standardized, highly accurate reference thermometer (e.g., a rectal thermometer or an oral thermometer with known accuracy) or a specialized blackbody calibrator in a controlled environment as specified by standards like ISO 80601-2-56. The document mentions "clinical accuracy test" and "performance test," strongly implying such a comparative methodology.

8. The sample size for the training set:

  • Not applicable / Not provided. The device is an ear thermometer, which is a sensor-based measurement tool, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. The device's performance is determined by its design, calibration, and adherence to physical principles and engineering specifications, which are validated through bench and clinical testing.

9. How the ground truth for the training set was established:

  • Not applicable. As noted above, the concept of a "training set" and its associated ground truth establishment is not relevant for this type of medical device submission.

FDA 510(k) Clearance Letter - Ear Thermometers

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID # 04017.07.05
Silver Spring, MD 20993
www.fda.gov

June 23, 2025

Shenzhen AOJ Medical Technology Co., Ltd.
Jack Wang
Deputy Chief
Rm 301&4F, Block A, Bldg A, Jingfa Intelligent Manufacturing Park,Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an Dis
Shenzhen, Guangdong 518126
China

Re: K250470
Trade/Device Name: Ear Thermometer (EAR-E101); Ear Thermometer (EAR-E102); Ear Thermometer (EAR-E103)
Regulation Number: 21 CFR 880.2910
Regulation Name: Clinical Electronic Thermometer
Regulatory Class: Class II
Product Code: FLL
Dated: May 23, 2025
Received: May 23, 2025

Dear Jack Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.


U.S. FOOD & DRUG ADMINISTRATION

June 23, 2025

Shenzhen AOJ Medical Technology Co., Ltd.
Jack Wang
Deputy Chief
Rm 301&4F, Block A, Bldg A, Jingfa Intelligent Manufacturing Park,Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an Dis
Shenzhen, Guangdong 518126
China

Re: K250470
Trade/Device Name: Ear Thermometer (EAR-E101); Ear Thermometer (EAR-E102); Ear Thermometer (EAR-E103)
Regulation Number: 21 CFR 880.2910
Regulation Name: Clinical Electronic Thermometer
Regulatory Class: Class II
Product Code: FLL
Dated: May 23, 2025
Received: May 23, 2025

Dear Jack Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Page 2

K250470 - Jack Wang Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K250470 - Jack Wang Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

PORSCHE P. BENNETT -S

Porsche Bennett
For David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K250470

Device Name
Ear Thermometer (EAR-E101);
Ear Thermometer (EAR-E102);
Ear Thermometer (EAR-E103)

Indications for Use (Describe)
The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

510(k) Number (if known)
K250470

Device Name
Ear Thermometer (EAR-E101);
Ear Thermometer (EAR-E102);
Ear Thermometer (EAR-E103)

Indications for Use (Describe)
The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2025/06/20

1. Submission sponsor

Name: Shenzhen AOJ Medical Technology Co., Ltd.
Address: Room 301&4F, Block A, Building A, Jingfa Intelligent Manufacturing Park, Xiaweiyuan, Gushu Community, Xixiang Street, Bao'an District, 518126 Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Contact person: Jack Wang
Title: Deputy Chief
TEL: 86 755-27786026

2. Subject Device Information

Trade/Device NameEar Thermometer (EAR-E101); Ear Thermometer (EAR-E102); Ear Thermometer (EAR-E103)
Common NameClinical Electronic Thermometer
Modelmodels EAR-E101, EAR-E102, EAR-E103
Regulation Number21 CFR 880.2910
Regulatory ClassClass II
Product CodeFLL
Submission typeTraditional 510(K)

3. Predicate Device

Manufacturer: Jiangsu Yuyue Medical Equipment & Supply Co., Ltd
Device name: Infrared Ear Thermometer, models YHT101 and YHT200.
510(K) Number: K203583.

4. Device Description

The ear thermometer is a handheld device that displays the temperature of the measured patient by measuring the thermal radiation of the eardrum. Measurement unit: ℃ or ℉.The results can be displayed on LCD.

The thermometers are powered by 1.5V×2 ( AAA or AA) alkaline batteries, which can be used for people over three months.

A thermopile sensor is employed to detect or monitor the infrared thermal energy emitted from the eardrum, which is converted into temperature measurement with the unit of °C or °F.

All the models share the same critical components, intended use, working principle and similar product design, and compose of a sensor, PCB, buttons, LCD display and housing. Functions include

Page 6

temperature measurement, memory reading recall, unit switch, low battery detection and high temperature indicator.

The major difference is the mechanical and the corresponding hardware adjustment.

5. Intended use & Indication for use

The Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.

6. Comparison to the Predicate Device

ITEMSubject Device (K250470)Predicate Device (K203583)Comparison Result
Common NameEar ThermometerInfrared Ear Thermometer--
ModelEAR-E101, EAR-E102, EAR-E103YHT101, YHT200--
ManufacturerShenzhen AOJ Medical Technology Co., Ltd.Jiangsu Yuyue Medical Equipment & Supply Co., Ltd--
Product CodeFLLFLLSame
Regulation No.21 CFR 880.291021 CFR 880.2910Same
ClassificationIIIISame
Indications for Use / Intended UseThe Ear Thermometer is intended to measure human body temperature of people over three months from surface of eardrum. It applies to both professional use and home use.The YUWELL® Infrared ear thermometer is a non-sterile, reusable clinical thermometer. The device is to display the body temperature in the ear cavity by thermal radiation for people of all ages except preterm babies and newborns (1-29 days old).Note 1
Patients populationpeople over three monthsPeople of all ages except preterm babies and newborns (1-29days old)Note 2
OperationHand held-Manually operatedHand held-Manually operatedSame
SensorInfraredInfraredSame
Rx/OTC UseOTCOTCSame
Measurement TechnologyThermopileThermopileSame
Operating ModeAdjusted modeAdjusted modeSame
Measurement SiteeardrumearSame
Measurement Range32.0℃~42.9℃ (89.6°F-109.2°F)34.0 °C to 42.2 °C (93.2℉~108.0℉)Note 3
Accuracy±0.2℃/±0.4°F34.0℃~34.9℃(93.2℉~94.8℉), ±0.3℃(±0.5℉) 35.0℃~42.0℃(95.0℉~107.6℉), ±0.2℃(±0.4℉) 42.1℃-42.2℃(107.7℉~108.0℉), ±0.3℃(±0.5℉)Note 4
DisplayLCDLCDSame
Memory40 valuesYHT101: 7 groups of measuredNote 5

Page 7

ITEMSubject Device (K250470)Predicate Device (K203583)Comparison Result
functionvalues can be memorized YHT200: 10 groups of measured values can be memorized
Scale°C/°F°C/°FSame
Resolution of display0.1 °C/°F0.1 °C/°FSame
Automatic power off timeWithin 30s60s±10sNote 6
Operating EnvironmentsTemperature: 15°C~ 40°C Humidity:≤ 85% RH, non-condensing Atmospheric pressure: 70–106 kPaTemperature: 10 °C~ 40 °C (50℉-104℉) Humidity: 15%–90% (Non-Condensing) Atmospheric pressure: 70–106 kPaNote 7
Transport and Storage EnvironmentsAmbient Temperature: -20C to 55C Relative Humidity:≤ 93% RH, non-condensing Atmospheric pressure: 50 kPa to 106 kPaAmbient Temperature: -20C to +55C (-4℉-131℉) Relative Humidity:15%–90% (Non-Condensing) Atmospheric pressure: 70 kPa - 106 kPaNote 8
Power supply2x1.5 V AA Alkaline battery (EAR-E101,EAR-E102) 2x1.5 V AAA Alkaline battery (EAR-E103)Two (2) AAA alkaline batteriesNote 9
Dimensions (mm×mm×mm)EAR-E101: 149mm x 80.4mm x 38.2mm EAR-E102: 152.3mm x 55.6mm x 36mm EAR-E103: 140 mm x 35mm x 40mmYHT101: 145mm x 37mm x 57mm YHT200: 143mm x 39 mm x 57mmNote 10
Electrical SafetyIEC 60601-1 IEC 60601-11 ISO 80601-2-56IEC 60601-1 IEC 60601-11 ISO 80601-2-56Same
EMCIEC 60601-1-2IEC 60601-1-2Same
Patient Contacting MaterialsABS, PP, LLDPSYHT101:Shell: ABS material Button: PC material YHT200: Shell: ABS material Button: ABS materialNote 11 Both were validated for cytotoxicity, irritation and sensitization
BiocompatibilityISO 10993-1 ISO 10993-5, ISO 10993-10, ISO 10993-23ISO 10993-1 ISO 10993-5, ISO 10993-10Equivalent

Page 8

Justification for the differences:

No.DifferenceJustification
Note 1 Note 2Indications for Use / Intended Use & Patients populationThe scope of the applicable group of people has been narrowed down to exclude pediatric patients under 3 months (predicate device applicable population is also included). The subject device has a narrower intended population and it does not raise any new or different questions of safety or effectiveness.
Note 3 Note 4Measurement range & accuracyThe measurement range of the predicate device is different from that of the subject device. The scale of clinical thermometer like Mercury thermometer reads temperature from 35℃ to 42℃. The subject device measurement range is 32.0℃ to 42.9℃ meet the normal clinical use. For the normal clinical measurement range of 32.0℃ to 42.9℃, the accuracy of the subject device and the predicate device are same or even stricter than the predicate device. And the performance test and clinical accuracy test showed that the subject device do not raise any new questions of safety and effectiveness.
Note 5Memory functionIn terms of the memory function, it is to help patients remember and check the previous results, so this minor difference would not raise any safety and effectiveness questions.
Note 6Automatic power off timeThe design purpose is to give the user time to review or record the data if needed. The difference will not raise any new safety or effectiveness questions.
Note 7 Note 8Operation Environments & Storage EnvironmentsAlthough the operation and storage conditions of the subject device are slightly different from the predicate device, they meet the same standard requirements of ISO 80601-2-56 and IEC 60601-1-11.Therefore, the operating and storage conditions do not raise new questions of safety and effectiveness.
Note 9Power supplyThe subject device powered by internal battery (AA or AAA) ,which both belong to alkaline battery. The test results have been verified by electrical safety and EMC testing. It does not affect the safety and effectiveness of subject device.
Note 10DimensionsThe subject and the predicate are has slight difference in dimensions due to different outlook. Moreover, such engineering design has been verified against international standards, so such minor differences will not raise any safety and effectiveness questions.
Note 11Patient Contacting MaterialsThe materials used for subject device and predicate device are similar, and it is validated for cytotoxicity per ISO10993-5 and irritation as well as sensitization per ISO 10993-10&10993-23, and it is demonstrated that the subject device does not raise any new questions of safety and effectiveness.

As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The differences between the subject and predicate devices do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the bench testing, the different technological characteristics do not affect the safety and effectiveness of the subject device.

Page 9

7. Non-clinical Data

Biocompatibility testing

The biocompatibility evaluation for the device was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

  • Cytotoxicity
  • Sensitization
  • Irritation

Bench testing

The device has been tested according to the following standards:

  • IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
  • IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
  • ISO 80601-2-56 Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
  • IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
  • FDA Guidance on the Content of Premarket Notification [510(K)] Submissions for Clinical Electronic Thermometers

8. Clinical data

The clinical testing has been conducted per ISO 80601-2-56 Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

9. Conclusion

It is concluded from the non-clinical and clinical tests that the subject devices are substantially equivalent to the legally marketed predicate device identified above.

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.