Found 7 results

510(k) Data Aggregation

    K Number
    K133420
    Date Cleared
    2013-11-29

    (21 days)

    Product Code
    Regulation Number
    878.4840
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TranQuill barbed device comprised of polydioxanone, is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

    Device Description

    The TranQuill barbed device is a sterile, synthetic absorbable device that is intended for use in the approximation of soft tissue. It is comprised of polyester [poly (p-dioxanone)], dyed with D&C Violet No. 2. The device is designed with small bi-directional barbs along the long axis of the suture monofilament. It is available in diameter Size 0 through 2-0 in various lengths affixed to various needle types.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called the TranQuill barbed device, an absorbable surgical suture. The focus of the documentation is to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study based on the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Acceptance Criteria (Inferred from testing)Reported Device Performance
    Material CompositionComprised of Polydioxanone (PDO)Comprised of Polydioxanone (PDO), dyed with D&C Violet No. 2
    DesignDesigned with small bi-directional barbs along the long axis of the suture monofilamentDesigned with small bi-directional barbs along the long axis of the suture monofilament
    Physical DimensionsAvailable in diameter Size 0 through 2-0 in various lengthsAvailable in diameter Size 0 through 2-0 in various lengths
    SterilitySterileSterile
    Tensile StrengthConforms to USP monograph for absorbable suturesConforms to USP monograph for absorbable sutures
    In vitro Barb Holding StrengthPerformance comparable to predicate deviceDemonstrated substantial equivalence to predicate device in barb holding strength
    Post-Hydrolysis Tensile TestingPerformance comparable to predicate deviceDemonstrated substantial equivalence to predicate device in post-hydrolysis tensile strength
    Intended UseSoft tissue approximation where use of absorbable sutures is appropriateIndicated for soft tissue approximation where use of absorbable sutures is appropriate
    BiocompatibilityNot explicitly stated as a test, but implied by regulatory requirements for implantable materialsNot explicitly stated, but implied by use of PDO, a known biocompatible material

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated. The documentation mentions "Non-clinical laboratory performance testing," but does not provide specific sample numbers for each test (e.g., number of sutures tested for tensile strength).
    • Data Provenance: The testing was "Non-clinical laboratory performance testing," indicating it was conducted in a controlled lab environment. There is no information about the country of origin of the data or whether it was retrospective or prospective, as it pertains to benchtop testing rather than clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable. This device is a surgical suture, and the evaluation focused on non-clinical laboratory performance testing to establish substantial equivalence to a predicate device. It did not involve a test set requiring expert interpretation or ground truth establishment in the context of diagnostic or screening performance.

    4. Adjudication Method for the Test Set:

    • Not applicable. As the study involved non-clinical laboratory performance testing of physical characteristics, there was no need for an adjudication method as would be used in studies involving subjective expert opinions or diagnostic interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC study was not done. This type of study is typically performed for diagnostic or screening devices to assess human reader performance with and without AI assistance. The TranQuill device is a surgical product, not a diagnostic tool requiring interpretation by multiple readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

    • Not applicable. The device is a physical surgical suture, not a software algorithm. Therefore, there is no "standalone performance" in the context of an algorithm.

    7. Type of Ground Truth Used:

    • Objective Measurement and Standards Compliance: The "ground truth" for this evaluation was based on objective measurements and adherence to established standards and predicate device performance.
      • USP monograph for absorbable sutures: This serves as a "ground truth" or standard for tensile strength.
      • Predicate device characteristics: The performance of the predicate device (Quill Synthetic Absorbable Barbed Suture, Quill™ Self-Retaining System (SRS) comprised of PDO) served as the comparative "ground truth" for barb holding strength and post-hydrolysis tensile testing.

    8. Sample Size for the Training Set:

    • Not applicable. This device is a physical medical device (suture), not an AI/ML algorithm that requires a training set. The evaluation focuses on physical and mechanical properties.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As there is no training set for a physical device, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Polypropylene suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
    Quill™ Knotless Tissue-Closure Device comprised of Polypropylene is indicated for soft tissue approximation excluding closure of the epidermis.

    Device Description

    The Polypropylene Surgical Suture and Quill™ Polypropylene Tissue-Closure Devices are sterile, synthetic Knotless nonabsorbable surgical suturing devices that are intended for use in the closure of soft tissue. They are comprised of high molecular weight, isotactic polypropylene, undyed, or dyed with Phthalocyaninato (2-) Copper. The proposed blue modification is a change in supplier of the polypropylene resin raw material. The devices are supplied either single or doublearmed with various needle types, and are available in USP diameter Sizes 2 through 11-0, as applicable.
    The Quill™ Polypropylene devices are designed with either small opposing bi-directional or uni-directional barbs along the long axis of the suture monofilament. The uni-directional products contain a welded primary loop and secondary loop design at the distal end, opposite the attached needle.

    AI/ML Overview

    The provided text describes the performance tests conducted for the Polypropylene Surgical Suture and Quill™ Polypropylene Knotless Tissue-Closure Devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document states that the devices conform to the USP (United States Pharmacopeia) monograph for nonabsorbable sutures regarding diameter and tensile strength.

    Acceptance Criteria (from USP Monograph for Nonabsorbable Sutures)Reported Device Performance (from Non-clinical Laboratory Testing)
    Conformity to specified diameter requirementsDevices conform to specified diameter requirements
    Conformity to specified tensile strength requirementsDevices conform to specified tensile strength requirements
    Chemical equivalency to predicate devicesChemical characterization and biological risk assessment demonstrates chemical equivalency

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample sizes used for the non-clinical laboratory performance testing. It only mentions that "Non-clinical laboratory performance testing was conducted." The data provenance is implied to be retrospective as it involves testing of manufactured devices to confirm adherence to existing standards and comparison to predicate devices, rather than a prospective study to generate new clinical data. The country of origin for the data is not explicitly stated but is implied to be in the USA given the FDA context.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the ground truth for these non-clinical performance tests is established by adherence to a standardized monograph (USP) and direct physical/chemical testing, not by expert interpretation or consensus.

    4. Adjudication Method for the Test Set:

    This information is not applicable for non-clinical laboratory performance testing. Adjudication methods are typically used in clinical studies or studies involving human judgment (e.g., image interpretation).

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, and the effect size:

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. The study described is a non-clinical laboratory performance study, not a clinical study involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This question is not applicable as the device is a physical surgical suture, not a software algorithm or AI-driven system.

    7. The Type of Ground Truth Used:

    The ground truth used for these non-clinical performance tests is based on established industry standards and physical/chemical properties, specifically:

    • The USP (United States Pharmacopeia) monograph for nonabsorbable sutures for diameter and tensile strength.
    • Chemical characterization and biological risk assessment to evaluate chemical equivalency.

    8. The Sample Size for the Training Set:

    This information is not applicable as the study described is a non-clinical performance test for a physical medical device, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as above (no training set for a physical medical device).

    Ask a Question

    Ask a specific question about this device

    K Number
    K130078
    Date Cleared
    2013-02-28

    (45 days)

    Product Code
    Regulation Number
    878.5010
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quill™ Knotless Tissue-Closure Device comprised of Polypropylene is indicated for soft tissue approximation excluding closure of the epidermis.

    Device Description

    The Quill™ Polypropylene Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic nonabsorbable tissue-closure device that is intended for use in the closure of soft tissue, excluding closure of the epidermis. It is comprised of high molecular weight, isotactic polypropylene, undyed, or dyed blue with Phthalocyaninato (2-) Copper. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Size 0 in various lengths affixed to various needle types.

    AI/ML Overview

    The provided text describes a medical device, the Quill™ Polypropylene Knotless Tissue-Closure Device, and its regulatory submission. It does not contain information about a study with acceptance criteria, reported device performance, sample sizes for test or training sets, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies.

    The document is a 510(k) summary for a substantial equivalence determination by the FDA, primarily relying on performance testing against existing standards and comparison to predicate devices, rather than a clinical study evaluating diagnostic or therapeutic efficacy with defined acceptance criteria for such metrics.

    Therefore, I cannot provide the requested information based on the given text. The document focuses on:

    • Device Description: A sterile, synthetic nonabsorbable tissue-closure device with barbs and a loop design, made of polypropylene.
    • Indications for Use: Soft tissue approximation, excluding epidermis closure.
    • Substantial Equivalence Claims: Based on material composition and intended use being identical to a predicate barbed suture, and design being identical to other knotless tissue-closure devices.
    • Performance Tests: Non-clinical laboratory testing to confirm compliance with USP monograph for nonabsorbable sutures regarding tensile strength and needle attachment, performed according to FDA's Class II Special Controls Guidance Document: Surgical Sutures.

    Since the device is a surgical suture, the "acceptance criteria" discussed are related to physical and mechanical properties (tensile strength, needle attachment) rather than diagnostic performance metrics (e.g., sensitivity, specificity) expected in studies involving AI algorithms or human reader performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123877
    Date Cleared
    2013-01-02

    (16 days)

    Product Code
    Regulation Number
    878.4840
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    QuillTM PDO Knotless Tissue-Closure Device comprised of Polydioxanone is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

    Device Description

    The QuillTM PDO Knotless Tissue-Closure Device, Variable Loop Design (Polydioxanone) is a sterile, synthetic absorbable tissue- closure device that is intended for use in the closure of soft tissue. It is comprised of polyester [poly (p-dioxanone)], dyed with D&C Violet No. 2. The instrument is designed with small uni- directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Size 0, 2-0 and 3-0 in various lengths affixed to various needle types.

    AI/ML Overview

    This 510(k) submission describes a Quill™ PDO Knotless-Tissue Closure Device, Variable Loop Design (Polydioxanone), which is an absorbable surgical suture. The device is a sterile, synthetic absorbable tissue-closure device comprised of polyester [poly (p-dioxanone)], with uni-directional barbs and a welded primary and secondary loop design at the distal end. It is available in diameter Size 0, 2-0, and 3-0 in various lengths affixed to various needle types.

    The submission focuses on demonstrating substantial equivalence to a predicate device (K113744 Quill™ PDO Knotless Tissue-Closure Device, Variable Loop Design, Size -0-) by confirming that the new device, which differs only in suture diameter, meets established performance standards for absorbable sutures.

    Here's an analysis of the provided information, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from USP Monograph for Absorbable Sutures and FDA Guidance)Reported Device Performance (from Non-clinical Laboratory Testing)
    Conformance to USP monograph for absorbable sutures for tensile strength.Device conforms to USP monograph for absorbable sutures for tensile strength.
    Conformance to USP monograph for absorbable sutures for needle attachment.Device conforms to USP monograph for absorbable sutures for needle attachment.
    Performance demonstrated as substantially equivalent to the predicate device in in vitro post-hydrolysis tensile testing.In vitro post-hydrolysis tensile testing demonstrated substantial equivalence to the predicate device.

    Note: The document explicitly states the new device has "the same design and materials as the Quill™ PDO Knotless Tissue-Closure Device predicate device, including the same intended use and technological characteristics as the predicate device. The only difference between the proposed and predicate device is the suture diameter." This implies the acceptance criteria for these performance aspects are to meet or exceed the performance of the predicate device and the general USP standards.

    2. Sample size used for the test set and the data provenance

    The document specifies "Non-clinical laboratory performance testing was conducted." It also mentions "Additional performance testing was conducted in order to demonstrate substantial equivalence to the predicate device including in vitro post-hydrolysis tensile testing."

    • Sample Size: The document does not explicitly state the sample size used for the non-clinical laboratory performance testing or the in vitro post-hydrolysis tensile testing.
    • Data Provenance: The tests are described as "non-clinical laboratory performance testing" and "in vitro," indicating they were conducted in a laboratory setting, likely by the manufacturer, Angiotech. There is no information regarding the country of origin of the data or whether the data was retrospective or prospective in the context of animal or human studies, as these were material and mechanical property tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is not applicable to this submission. The "ground truth" for this medical device (surgical suture) is established by adherence to recognized standards for material properties (e.g., USP monograph) and performance measures like tensile strength and needle attachment, rather than expert interpretation of images or clinical outcomes that require expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving expert review of ambiguous cases (e.g., diagnostic imaging). For non-clinical, laboratory-based material property testing, the "ground truth" is determined by objective measurements and reference to established standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This submission is for a surgical suture, not an AI-assisted diagnostic device. Therefore, MRMC studies and the concept of human readers improving with AI assistance are irrelevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This submission is for a surgical suture, not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is based on:

    • Established Standards: Primarily the USP monograph for absorbable sutures.
    • Mechanical and Material Properties: Objective measurements of tensile strength, needle attachment, and in vitro post-hydrolysis tensile strength.
    • Predicate Device Performance: The new device's performance was compared to that of the legally marketed predicate device (K113744) to demonstrate substantial equivalence.

    8. The sample size for the training set

    Not applicable. "Training set" refers to data used to train machine learning models. This submission is for a surgical suture, not an AI or machine learning device.

    9. How the ground truth for the training set was established

    Not applicable. As stated above, there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123836
    Date Cleared
    2012-12-21

    (8 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quill™ Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

    Device Description

    The QuillTM MonodermTM Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic absorbable tissue-closure device that is intended for use in the closure of soft tissue. It is comprised of a copolymer of glycolide and e-caprolactone, undyed, or dyed with D&C Violet No. 2. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Sizes 1 to 3-0, in various lengths affixed to various needle types.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for the Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Metric)Reported Device Performance
    USP Monograph for Absorbable Sutures:
    - Tensile StrengthConforms to USP monograph
    - Needle AttachmentConforms to USP monograph
    Substantial Equivalence to Predicate Devices:
    - In vitro post-hydrolysis tensile testingDemonstrates substantial equivalence to predicate devices

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a specific sample size for the "test set" in the context of human data or clinical study. The performance testing described is non-clinical laboratory performance testing. Therefore, the concept of data provenance (e.g., country of origin, retrospective/prospective) related to human subjects is not applicable here.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this device's performance is established through non-clinical laboratory testing according to established standards (USP monograph) and direct comparison to predicate devices, not through expert consensus on a test set of images or clinical cases.

    4. Adjudication Method for the Test Set

    Not applicable. There is no mention of a test set requiring adjudication in the context of this device's performance evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a surgical suture, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical surgical suture, not an algorithm or software.

    7. The Type of Ground Truth Used

    The ground truth used for this device includes:

    • USP Monograph Standards: For tensile strength and needle attachment. These are well-defined, established standards for absorbable sutures.
    • Predicate Device Performance Data: For in vitro post-hydrolysis tensile testing, the performance of the predicate devices serves as the benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device (suture) and does not involve AI or machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123409
    Date Cleared
    2012-11-20

    (15 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quill™ Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

    Device Description

    The Quill TM Monoderm TM Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic absorbable tissue-closure device that is intended for use in the closure of soft tissue. It is comprised of a copolymer of glycolide and e-caprolactone, undyed, or dyed with D&C Violet No. 2. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Sizes 2-0 to 0 in various lengths affixed to various needle types.

    AI/ML Overview

    This document describes the 510(k) submission for the Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design. The device is an absorbable surgical suture. The submission focuses on demonstrating substantial equivalence to predicate devices through performance testing.

    Acceptance Criteria and Device Performance Study Information:

    This 510(k) summary does not include detailed acceptance criteria or a "performance table" in the typical sense of a diagnostic device's metrics (e.g., sensitivity, specificity). Instead, the "acceptance criteria" are implied by the requirements of the USP monograph for absorbable sutures and FDA's Class II Special Controls Guidance Document for Surgical Sutures. The device's performance is demonstrated by meeting these standards.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Tensile StrengthConforms to USP monograph for absorbable sutures.Testing demonstrates conformity to USP monograph.
    Needle AttachmentConforms to USP monograph for absorbable sutures.Testing demonstrates conformity to USP monograph.
    In vitro post-hydrolysis tensile testingDemonstrates substantial equivalence to predicate devices.Testing conducted to demonstrate substantial equivalence.
    Material CompositionIdentical to Quill™ Monoderm™ predicates.Identical in material composition.
    Size RangeIdentical to Quill™ Monoderm™ predicates.Identical in size range.
    DesignIdentical to Quill™ Monoderm™ Knotless Tissue-Closure Device and Quill™ PDO Knotless Tissue-Closure Device predicates.Identical in design.
    Intended UseSame as all three predicate devices.Has the same intended use.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states "Non-clinical laboratory performance testing was conducted." However, it does not specify the sample size used for the various tests (tensile strength, needle attachment, in vitro post-hydrolysis tensile testing).

    The data provenance is non-clinical laboratory testing. The document does not provide details on the country of origin of the data or whether it was retrospective or prospective, as these concepts are generally not applicable to laboratory performance testing of this nature.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This type of information is not applicable to this device's submission. The "ground truth" for a surgical suture is defined by established engineering and material science standards (e.g., USP monographs, FDA guidance). Expert review, as in clinical or imaging studies, is not part of establishing the performance for these specific non-clinical tests.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods are typically relevant for clinical studies where multiple reviewers assess outcomes or image interpretations. For laboratory performance testing, the results are objectively measured against established standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret data, often with and without AI assistance, to assess the impact of AI on their performance. The Quill™ Monoderm™ device is a surgical suture, not a diagnostic tool requiring human interpretation in this context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    This information is not applicable. The device is a physical surgical suture, not an algorithm or AI system. Therefore, standalone algorithm performance is not a relevant concept.

    7. The Type of Ground Truth Used:

    The ground truth for the non-clinical performance testing is based on established industry standards and regulatory requirements. Specifically:

    • USP monograph for absorbable sutures for tensile strength and needle attachment.
    • FDA's Class II Special Controls Guidance Document: Surgical Sutures, Issued June 3, 2003.
    • The characteristics and performance of the legally marketed predicate devices to which equivalence is claimed.

    8. The Sample Size for the Training Set:

    This information is not applicable. The device is a surgical suture, not a machine learning model that requires a training set. The performance testing described is for product verification and validation, not for algorithm training.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122898
    Date Cleared
    2012-10-22

    (31 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SURGICAL SPECIALTIES CORP. DBA ANGIOTECH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Quill™ Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

    Device Description

    The Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic absorbable tissue-closure device that is intended for use in the closure of soft tissue. It is comprised of a copolymer of glycolide and e-caprolactone. undyed, or dyed with D&C Violet No. 2. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Size 2-0 in various lengths affixed to various needle types.

    AI/ML Overview

    The provided document is a 510(k) summary for the Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design. It describes the device, its indication for use, and a general statement about performance testing. However, it does not contain specific acceptance criteria, detailed study results, or the other requested information to fully address all parts of your prompt.

    Here's a breakdown of what can and cannot be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance

    • Acceptance Criteria: Not explicitly stated in terms of quantitative values or pass/fail thresholds. The document generally refers to conforming to the USP monograph for absorbable sutures and demonstrating substantial equivalence.
    • Reported Device Performance: The document states that "The results of this testing demonstrates that the Quill™ Monoderm™ Knotless Tissue-Closure device, Variable Loop Design, is substantially equivalent to the predicate devices." No specific performance metrics (e.g., actual tensile strength values, degradation rates) are provided.
    Acceptance Criteria (Inferred from text)Reported Device Performance
    Conformance to USP monograph for absorbable sutures for tensile strength and needle attachment.Conforms
    Substantial equivalence to predicate devices (Quill™ Monoderm™, Quill™ PDO Knotless Tissue-Closure Device, V-Loc™ 90 Absorbable Wound Closure Device) through simulated use tensile testing in porcine tissue and in vitro post-hydrolysis tensile testing.Demonstrated substantially equivalent to predicate devices

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified.
    • Data Provenance: The document only mentions "testing" and "simulated use tensile testing in porcine tissue and in vitro post-hydrolysis tensile testing." It does not specify country of origin, retrospective or prospective nature, or where the testing was conducted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This device is a surgical suture, and the performance tests described are mechanical and material characterization studies, not image interpretation or diagnostic studies that would require expert consensus for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI/imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No. This is not an AI/imaging device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • For the non-clinical laboratory performance testing (tensile strength, needle attachment), the "ground truth" would be established by standardized measurement techniques and USP monograph specifications.
    • For the simulated use tensile testing in porcine tissue and in vitro post-hydrolysis tensile testing, the "ground truth" would be the measured physical properties of the device and the predicate devices, compared against each other.

    8. The sample size for the training set

    • Not applicable. This is not a machine learning/AI device that requires a training set. The performance tests are experimental studies, not algorithm training.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.

    In summary: The provided 510(k) summary offers a high-level overview of the device and the types of performance testing conducted to support substantial equivalence. It lacks the detailed quantitative data, sample sizes, and specific methodologies that would be required to fully answer most of your detailed questions, particularly those related to clinical studies, AI performance, or expert-adjudicated ground truth, as those concepts do not apply to this type of device and submission.

    Ask a Question

    Ask a specific question about this device