K Number
K123409
Date Cleared
2012-11-20

(15 days)

Product Code
Regulation Number
878.4493
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Quill™ Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

Device Description

The Quill TM Monoderm TM Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic absorbable tissue-closure device that is intended for use in the closure of soft tissue. It is comprised of a copolymer of glycolide and e-caprolactone, undyed, or dyed with D&C Violet No. 2. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Sizes 2-0 to 0 in various lengths affixed to various needle types.

AI/ML Overview

This document describes the 510(k) submission for the Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design. The device is an absorbable surgical suture. The submission focuses on demonstrating substantial equivalence to predicate devices through performance testing.

Acceptance Criteria and Device Performance Study Information:

This 510(k) summary does not include detailed acceptance criteria or a "performance table" in the typical sense of a diagnostic device's metrics (e.g., sensitivity, specificity). Instead, the "acceptance criteria" are implied by the requirements of the USP monograph for absorbable sutures and FDA's Class II Special Controls Guidance Document for Surgical Sutures. The device's performance is demonstrated by meeting these standards.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Tensile StrengthConforms to USP monograph for absorbable sutures.Testing demonstrates conformity to USP monograph.
Needle AttachmentConforms to USP monograph for absorbable sutures.Testing demonstrates conformity to USP monograph.
In vitro post-hydrolysis tensile testingDemonstrates substantial equivalence to predicate devices.Testing conducted to demonstrate substantial equivalence.
Material CompositionIdentical to Quill™ Monoderm™ predicates.Identical in material composition.
Size RangeIdentical to Quill™ Monoderm™ predicates.Identical in size range.
DesignIdentical to Quill™ Monoderm™ Knotless Tissue-Closure Device and Quill™ PDO Knotless Tissue-Closure Device predicates.Identical in design.
Intended UseSame as all three predicate devices.Has the same intended use.

2. Sample Size Used for the Test Set and Data Provenance:

The document states "Non-clinical laboratory performance testing was conducted." However, it does not specify the sample size used for the various tests (tensile strength, needle attachment, in vitro post-hydrolysis tensile testing).

The data provenance is non-clinical laboratory testing. The document does not provide details on the country of origin of the data or whether it was retrospective or prospective, as these concepts are generally not applicable to laboratory performance testing of this nature.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This type of information is not applicable to this device's submission. The "ground truth" for a surgical suture is defined by established engineering and material science standards (e.g., USP monographs, FDA guidance). Expert review, as in clinical or imaging studies, is not part of establishing the performance for these specific non-clinical tests.

4. Adjudication Method for the Test Set:

This information is not applicable. Adjudication methods are typically relevant for clinical studies where multiple reviewers assess outcomes or image interpretations. For laboratory performance testing, the results are objectively measured against established standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret data, often with and without AI assistance, to assess the impact of AI on their performance. The Quill™ Monoderm™ device is a surgical suture, not a diagnostic tool requiring human interpretation in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

This information is not applicable. The device is a physical surgical suture, not an algorithm or AI system. Therefore, standalone algorithm performance is not a relevant concept.

7. The Type of Ground Truth Used:

The ground truth for the non-clinical performance testing is based on established industry standards and regulatory requirements. Specifically:

  • USP monograph for absorbable sutures for tensile strength and needle attachment.
  • FDA's Class II Special Controls Guidance Document: Surgical Sutures, Issued June 3, 2003.
  • The characteristics and performance of the legally marketed predicate devices to which equivalence is claimed.

8. The Sample Size for the Training Set:

This information is not applicable. The device is a surgical suture, not a machine learning model that requires a training set. The performance testing described is for product verification and validation, not for algorithm training.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no training set for this type of medical device submission.

{0}------------------------------------------------

K123409 page 1/2

510(k) Summary

NOV 2 0 2012

Date Prepared:October 24, 2012
Company:Angiotech
100 Dennis Dr.
Reading, PA 19606
Contact:Kirsten Stowell
Regulatory Affairs Manager
Phone:610-404-3367
Fax:610-404-3924
Email:kstowell@angio.com
Device trade name:Quill TM Monoderm Knotless-Tissue Closure Device, Variable Loop Design
Device Common Name:Absorbable poly(glycolide/l-lactide) Surgical Suture
Device classification:Absorbable poly(glycolide/l-lactide) Surgical Suture
Product code, GAM
21 CFR 878.4493
Class II
Legally marketed devices to which the device is substantially equivalent:K122898 Quill TM Monoderm TM Knotless Tissue-Closure Device, Variable Loop Design, Size 2-0
K113744: Quill TM PDO Knotless Tissue-Closure Device (Polydioxanone)
K072028: Quill TM Self-Retaining System (SRS) comprised of Monoderm TM
Description of the device:The Quill TM Monoderm TM Knotless Tissue-Closure Device, Variable Loop Design is a sterile, synthetic absorbable tissue-closure device that is intended for use in the closure of soft tissue. It is comprised of a copolymer of glycolide and e-caprolactone, undyed, or dyed with D&C Violet No. 2. The device is designed with small uni-directional barbs along the long axis of the suture monofilament which contains a welded primary loop and secondary loop design at the distal end. It is available in diameter Sizes 2-0 to 0 in various lengths affixed to various needle types.

.

.

.

{1}------------------------------------------------

K123409 page 2/2

  • QuillTM Indications for Use: Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.
    The QuillTM Monoderm™ Knotless Tissue-Closure device, Substantial Variable Loop Design is identical in material composition and Equivalence: size range as the Quili™ Monoderm™ predicates. The proposed device is identical in design to the Quill™ Monoderm™ Knotless Tissue-Closure Device and Quill™ PDO Knotless Tissue-Closure Device predicate devices. In addition, the proposed device has the same intended use as all three predicate devices.

  • Performance tests: Non-clinical laboratory performance testing was conducted to confirm that the Quill™ Monoderm™ Knotless Tissue-Closure device, Variable Loop Design, conforms to the USP monograph for absorbable sutures for tensile strength (as applicable) and needle attachment. This testing was performed in accordance with FDA's Class II Special Controls Guidance Document: Surgical Sutures, Issued June 3, 2003. Additional performance testing was conducted in order to demonstrate substantial equivalence to the predicate devices including in vitro posthydrolysis tensile testing.
    The results of this testing demonstrates that the QuillTM Monoderm™ Knotless Tissue-Closure device, Variable Loop Design, is substantially equivalent to the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle, with its wings forming three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Angiotech % Ms. Kirsten Stowell Regulatory Affairs Manager 100 Dennis Avenue Reading, Pennsylvania 19606

Novermber 20, 2012

Re: K123409

Trade/Device Name: Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design

Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly (glycolide/l-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: October 24, 2012 Received: November 5, 2012

Dear Ms. Stowell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Kirsten Stowell

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Peter D. Rumm -S

Mark N. Melkerson' Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4 - Indications for Use Statement

K123409 510k number if known:

Device Name: Quill™ Monoderm™ Knotless Tissue-Closure Device, Variable Loop Design

Indications for Use:

Quill™ Knotless Tissue-Closure Device comprised of Monoderm™ is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krause 2012.11.19 16:09:25 -05'00'

(Division Sign-Off) Division of Surgical Devices K123409 510 (k) Number

CONFIDENTIAL

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.