510(k) Data Aggregation
(132 days)
S.I.N. Implant System LTDA.
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. dental implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
S.I.N. Dental Implant System implants with lengths of 18, 20, 22 or 24mm may be tilted up to 30º. When used in the mandible or maxilla with implants with lengths of 18, 20, 22 or 24 mm at an angulation of 30º, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22 or 24 mm at angulations between 0º and less than 30º, the S.I.N. Dental Implant System implants are only indicated for multiple unit restoration in splinted applications that utilize at least two implants.
The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes various endosseous implants with 16° Morse tapers interface connection. This submission adds to the S.I.N. Dental Implants System, the Versalis S implants line.
The Versalis S implant line has the Versalis S and Versalis S Plus implants, and both have a Morse taper connection with an internal 16° cone taper. The Versalis S implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment. The Versalis S Plus implants have implant body lengths from 8.5 to 24mm and body diameter Ø from 3.5 to 7.0mm., they are manufactured from unalloyed titanium conforming to ASTM F67, with a double acid-etched surface treatment plus hydroxyapatite surface coating.
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700), Grade 4. The material used to manufacture the dental implants in this submission (conforming to ASTM F67) is identical to the material used to manufacture the device implants cleared in primary predicate K170392 and in additional predicate K222231.
The provided text is a 510(k) clearance letter for a dental implant system. It does not contain any information about acceptance criteria or a study proving that an AI/ML device meets acceptance criteria. The "Performance Data" section discusses non-clinical testing for the dental implant itself, such as sterilization validation, shelf-life testing, and biocompatibility, as well as fatigue testing and MRI environment compatibility. This is for a physical medical device, not a software device that relies on AI/ML.
Therefore, I cannot provide the requested information as it is not present in the given document.
To reiterate, the document is about a physical dental implant system and the performance data pertains to its physical properties, sterility, and biocompatibility, not to the performance of an AI/ML algorithm.