K240609Device Name
S.I.N. Dental Implant System
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2024-07-17
(135 days)
Product Code
DZE, NHA
Regulation Number
872.3640Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Description
The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants and mating abutments for screw-retained, multi-unit prostheses.
The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to approximately 2.3 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5. 35. 37.5. 40. 42.5.45. 47.5. 50. 52.5.57.5. and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).
The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 1.5, 2.0, and 2.5 mm. The prosthetic post is angled 52° or 60° to the long axis of the implant. These components are used with previously cleared abutment screws.
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA100 surface treatment, identical to that cleared in K231127. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants and abutments are provided sterile to the end user.
Ask a Question
K Number
K241378Device Name
S.I.N. Instrument Kits
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2024-07-12
(58 days)
Product Code
KCT
Regulation Number
880.6850Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:
Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.
S.I.N. Instrument Kits are intended for sterilization of non-porous loads.
S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
The combined weight of the Epikut PTG Surgical Kit and associated instruments is 632 grams.
The weight of the empty Epikut PTG Surgical Kit is 353 grams.
Device Description
The subject device is a reusable rigid container, comprising a base (bottom), a removable inner tray, and a lid (cover). The subject device tray is to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The tray is perforated to allow for penetration of the sterilant, is to be used with moist heat (steam) and requires the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instruments in the kits are manufactured from silicone.
Ask a Question
K Number
K232821Device Name
S.I.N. Dental Implant System
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2024-05-31
(261 days)
Product Code
DZE
Regulation Number
872.3640Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.
Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.
Device Description
The purpose of this submission is to add labeling claims for a surface treatment designated HA™®, which is applied to implants of the S.I.N. Dental Implant System, as previously cleared in K170392. K193096. K200992, K211921, K221453, and K222231. The new labeling claims and the information supporting the claims are shown below in Performance Data.
The following information is a summary of the designs of the subject device implants, as they have been previously cleared with the HA"" surface treatment. Note that there are no changes to any aspect of the design, nor to the HA" surface treatment, as described in the respective 510(k) Premarket Notifications referenced above.
The previously cleared implants that are the subject of this submission includes eleven (11) product lines from six (6) 510(k)s. Implant body diameters, platform diameters and lengths for each product line are shown below in Table 10.1 Summary of Subject Device Designs and Sizes.
K170392: Unitite Slim, Unitite, Unitite Compact K193096: Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus K200992: Strong SW Plus K211921: Epikut Plus CM, Epikut Plus HE K221453: Epikut Plus CM (additional lengths) K22231: Epikut S Plus
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The HA1000 surface treatment is identical to that cleared in K211921.
Ask a Question
K Number
K231127Device Name
S.I.N. Dental Implant System
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2023-12-07
(231 days)
Product Code
DZE, NHA
Regulation Number
872.3640Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Dental Implant System Zygomatic implants are intended for placement in the maxillary arch to provide support for fixed or removable dental prostheses in patients with partially or fully edentulous maxillae. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System Zygomatic implants are intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Description
The purpose of this submission is to add components to the S.I.N. Dental Implant System. This submission includes zygomatic dental implants, mating abutments for screw-retained, multi-unit prostheses, and the corresponding abutment screws.
The subject device Zygomatic Plus Morse Taper implants have a body diameter of 4.0 mm that tapers to 1.65 mm at the apex. The implants have a Morse taper ("CM") abutment interface connection, with an internal 16° cone taper. The abutment platform is 4.0 mm. All implants are threaded starting at the apex and extending approximately 17 mm in the coronal direction, with a major thread diameter of 4 mm; the thread tapers over the final 6 mm of the implant to the apex. The implants are provided with overall lengths of 30, 32.5, 35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 57.5, and 60 mm. The surface of all subject device implants is acid-etched from the implant platform to the apex, followed by application of a hydroxyapatite coating (HA nano).
The subject device abutments are multi-unit, indexed abutments for use only with the subject device Zygomatic Plus Morse Taper implants for maxillary full-arch restorations. The subject device abutments have a prosthetic platform diameter of 4.8 mm, and gingival heights of 2 mm and 3 mm. The prosthetic post is angled 45° to the long axis of the implant. These components are used with subject device abutment and prosthetic screws, or previously cleared screws.
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. All implants have an acid-etched HA1200 surface treatment, identical to that cleared in K200992. All subject device abutments are manufactured from titanium alloy (Ti-6A1-4V) conforming to ASTM F136. All subject device implants, abutments, and abutment screws are provided sterile to the end user.
Ask a Question
K Number
K222231Device Name
S.I.N. Dental Implant System
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2022-10-27
(94 days)
Product Code
DZE, NHA
Regulation Number
872.3640Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
S.I.N. Dental Implants with lengths of 18, 20, 22, or 24 mm be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Device Description
The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K211921, K200992, and K170392.
This submission includes dental implants Epikut S with a Morse taper (CM) abutment interface and an acid-etched endosseous surface, and Epikut S Plus implants with an endosseous surface produced by acidetching followed by application of a hydroxyapatite coating (HA™M). The implant design and endosseous surfaces are nearly identical to the Epikut Plus CM implants cleared in K211921, with the exception of the implant-abutment connection (subject device 16° Morse taper, K211921 11.5° Morse taper), the additional body/platform diameter (4.0 mm), and the longer lengths (18, 20, 22, and 24 mm).
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K211921, and the HA100 surface treatment is identical to that cleared in K211921.
Ask a Question
K Number
K222005Device Name
S.I.N. Instrument Kits
Manufacturer
S.I.N. - Sistema De Implante Nacional S.A.
Date Cleared
2022-09-29
(84 days)
Product Code
KCT
Regulation Number
880.6850Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema De Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.
The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:
Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.
S.I.N. Instrument Kits are intended for sterilization of non-porous loads.
S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
Device Description
The subject device includes a total of three (3) instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instruments in the kits are manufactured from silicone. The subject device includes two (2) sizes (lid, base, and enclosed volume), and a total of three (3) tray configurations.
Ask a Question
K Number
K212404Device Name
S.I.N. Instrument Kits
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2021-11-03
(93 days)
Product Code
KCT
Regulation Number
880.6850Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
Device Description
The subject device includes a total of 4 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner travs, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 3 sizes (same lid, base, and enclosed volume), and a total of 4 tray configurations.
Ask a Question
K Number
K211921Device Name
S.I.N. Dental Implant System
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2021-10-20
(121 days)
Product Code
DZE
Regulation Number
872.3640Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Description
The purpose of this submission is to add components to the S.I.N. Dental Implant System components cleared in K200992, K193096, and K170398.
This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and a double acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA™Y). These implant endosseous surfaces are identical to those cleared in K200992.
This submission also includes dental implants Epikut HE with an external hexagon (HE) abutment interface and a double acid-etched endosseous surface, and Epikut Plus HE implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA1000). These implant endosseous surfaces also are identical to those cleared in K200992.
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K200992. K193096, and K170398. The Epikut Plus CM and Epikut Plus HE implants have the identical HA™ surface treatment cleared in K200992 and K193096.
The subject device dental implants are compatible with abutments and prosthetic components cleared previously in K200992. K193096. K170398. and K051859. Epikut HE and Epikut Plus HE dental implants are not for use with angled abutments.
Ask a Question
K Number
K201688Device Name
S.I.N. Instrument Kits
Manufacturer
S.I.N. - Sistema de Implante Nacional S.A.
Date Cleared
2020-10-21
(121 days)
Product Code
KCT
Regulation Number
880.6850Why did this record match?
Applicant Name (Manufacturer) :
S.I.N. - Sistema de Implante Nacional S.A.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
S.I.N. Instrument Kits are intended to enclose other medical devices that are to be sterlized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated vycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
Device Description
The subject device includes a total of 14 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner trays, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 5 sizes (same lid, base, and enclosed volume), and a total of 14 tray configurations.