K Number
K212404
Date Cleared
2021-11-03

(93 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle: Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time. S.I.N. Instrument Kits are intended for sterilization of non-porous loads. S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

Device Description

The subject device includes a total of 4 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner travs, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 3 sizes (same lid, base, and enclosed volume), and a total of 4 tray configurations.

AI/ML Overview

The document provides information on the nonclinical testing performed for the S.I.N. Instrument Kits to demonstrate they meet acceptance criteria.

1. A table of acceptance criteria and the reported device performance

Test MethodologyAcceptance CriteriaResults
Manual Cleaning ValidationProtein assay Acceptance criterion: No color change = absence of protein residue. Sensitivity = 1 µg protein residue.Total organic carbon assay Acceptance criterion: results < limit of quantification = 0.1 ppm; Assay limit of detection = 0.015 ppm.Pass
Bacterial Endotoxin Testing (USP <85>)Endotoxin results ≤ 20 EU/device.Pass
Sterilization Validation (ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2)3 consecutive half-cycles performed for each of 3 sizes of trays demonstrate complete inactivation of all biologic indicators. A minimum SAL of 10-6 is achieved if the Instructions for Use are followed.Pass
Dry TimeUsing pre-cycle and post-cycle weights, the weight gain after drying will be ≤ 3%.Pass
Life Cycle / Simulated Use Life ValidationVisual inspection, component dimensional fit verification, functional closure/latch verification for 250 use cycles.Pass
Biocompatibility of Subject Device (Cytotoxicity testing - ANSI/AAMI/ISO 10993-5, ANSI/AAMI/ISO 10993-12)Non-cytotoxic if ≤ 50% of L-929 cells exposed to extract of device show rounding and lysis.Pass

2. Sample sizes used for the test set and the data provenance

  • Sterilization Validation: 3 consecutive half-cycles were performed for each of the 3 sizes of trays. This indicates a sample size of at least 9 sterilization cycles in total (3 sizes * 3 half-cycles).
  • Other tests: The exact sample sizes for Manual Cleaning Validation, Bacterial Endotoxin Testing, Dry Time, Life Cycle / Simulated Use Life Validation, and Biocompatibility are not explicitly stated by number, but the results indicate that sufficient samples were tested to meet the acceptance criteria.
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. Given that the manufacturer is S.I.N. - Sistema de Implante Nacional S.A. from Brazil, it is likely the testing was conducted in Brazil or at an international testing facility. The nature of the tests (laboratory testing of medical devices) implies a prospective test setup.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable. The tests performed are laboratory-based performances of a medical device (instrument kits for sterilization) against established regulatory standards and guidance, not diagnostic evaluations requiring expert consensus on "ground truth."

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This section is not applicable for this type of nonclinical, laboratory-based device performance testing. Ground truth is established by the predefined acceptance criteria of the test methodologies.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This type of study is not applicable as the device is a medical instrument kit for sterilization, not an AI-based diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. This concept is not applicable as the device is a physical medical instrument kit.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is based on established scientific and regulatory standards and methods for medical device sterilization, cleaning, biocompatibility, and durability. These include:

  • Predefined acceptance criteria based on standard test methods (e.g., absence of protein residue, endotoxin levels, sterility assurance level (SAL), weight gain after drying, visual inspection for integrity, cytotoxicity).
  • Compliance with recognized standards such as USP <85>, ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO 17665-2, ANSI/AAMI/ISO 10993-5, and ANSI/AAMI/ISO 10993-12.
  • FDA Guidance documents (e.g., "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling").

8. The sample size for the training set

This section is not applicable as the device is a physical medical instrument kit, not an AI/machine learning model that requires training data.

9. How the ground truth for the training set was established

This section is not applicable as the device is a physical medical instrument kit and does not involve a training set or AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

November 2, 2021

S.I.N. - Sistema de Implante Nacional S.A. % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real. Suite 400 San Diego, California 92130

Re: K212404

Trade/Device Name: S.I.N. Instrument Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: August 2, 2021 Received: August 2, 2021

Dear Kevin Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K212404

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Safe Drill Epikut Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill Epikut Kit and the associated instruments is 154 grams.

The weight of the empty Safe Drill Epikut Kit is 138 grams.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K212404

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Unitite Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Kit and the associated instruments is 620 grams.

The weight of the empty Unitite Surgical Kit is 520 grams.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K212404

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Epikut Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Epikut Surgical Kit and the associated instruments is 605 grams.

The weight of the empty Epikut Surgical Kit is 520 grams.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K212404

Device Name

S.I.N. Instrument Kits

Indications for Use (Describe)

Indications for Use for Epikut Surgical Guided Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.

The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 ℃ for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Epikut Surgical Guided Kit and the associated instruments is 808 grams.

The weight of the empty Epikut Surgical Guided Kit is 650 grams.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart G)

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{6}------------------------------------------------

510(k) Summary

K212404

S.I.N. – Sistema de Implante Nacional S.A.

S.I.N. Instrument Kits

October 21, 2021

ADMINISTRATIVE INFORMATION

Manufacturer NameS.I.N. - Sistema de Implante Nacional S.A.
Avenida Vereador Abel Ferreira, 1100
São Paulo, São Paulo
03340-000 Brazil
Telephone+55-11-21693000 ext 3236
Official ContactDenise Domiciano, Quality and Regulatory Manager
Representative/ConsultantKevin A. Thomas, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
12264 El Camino Real, Suite 400
San Diego, CA 92130
Telephone+1 858-792-1235
Fax+1 858-792-1236
Emailkthomas@paxmed.com
flarson@paxmed.com

DEVICE NAME AND CLASSIFICATION

Trade/Proprietary NameS.I.N. Instrument Kits
Common NameInstrument sterilization trays
Regulation Number21 CFR 880.6850
Regulation NameSterilization Wrap Containers, Trays, Cassettes & Other Accessories
Regulatory ClassClass II
Product CodeKCT
Classification PanelGeneral Hospital
Reviewing OfficeOffice of Surgical and Infection Control Devices (OHT4)
Reviewing DivisionDivision of Infection Control and Plastic Surgery Devices (DHT4B)

PREDICATE DEVICE INFORMATION The primary predicate device is: K201688, S.I.N. Instrument Kits, S.I.N. - Sistema de Implante Nacional S.A.

{7}------------------------------------------------

SUBJECT DEVICE INDICATIONS FOR USE STATEMENTS

Indications for Use for Safe Drill Epikut Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Safe Drill Epikut Kit and the associated instruments is 154 grams. The weight of the empty Safe Drill Epikut Kit is 138 grams.

Indications for Use for Unitite Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Unitite Surgical Kit and the associated instruments is 620 grams. The weight of the empty Unitite Surgical Kit is 520 grams.

The weight of the empty Unlite Surgical Kit is 320 grain.

Indications for Use for Epikut Surgical Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Epikut Surgical Kit and the associated instruments is 605 grams.

The weight of the empty Epikut Surgical Kit is 520 grams.

Indications for Use for Epikut Surgical Guided Kit

S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N. Instrument Kits require the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices. The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:

Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.

S.I.N. Instrument Kits are intended for sterilization of non-porous loads.

S.I.N. Instrument Kits are recommended not to be stacked during sterilization.

The combined weight of the Epikut Surgical Guided Kit and the associated instruments is 808 grams.

The weight of the empty Epikut Surgical Guided Kit is 650 grams.

{8}------------------------------------------------

SUBJECT DEVICE DESCRIPTION

The subject device includes a total of 4 instrument trays. The subject device trays are reusable rigid containers, comprising a base (bottom), one or more removable inner travs, and a lid (cover). The subject device trays are to be used to organize and protect the instruments that are sterilized by the healthcare provider. The base, inner tray, and lid components are designed to be integrated into a single unit which contains and protects the interior components during sterilization. The trays are perforated to allow for penetration of the sterilant, are to be used with moist heat (steam), and require the use of an FDA cleared wrap to maintain sterility. The subject device components are manufactured from injection molded polysulfone (PSU), and holders of various geometries to position instrument in the kits are manufactured from silicone. The subject device includes a total of 3 sizes (same lid, base, and enclosed volume), and a total of 4 tray configurations.

TECHNOLOGICAL CHARACTERISTICS COMPARISON TABLE

The comparison of the technological characteristics and Indications for Use Statements for the subject devices and the predicate device are provided at the end of this summary on pages 6-8.

The subject device is provided in 3 sizes and 4 configurations; the predicate device K201688 is provided in 5 sizes and 14 configurations. The subject device and the predicate device have similar overall dimensions, enclose similar volumes, and have similar vent to volume ratios. Differences in the dimensions and vent to volume ratios between the subject device and the predicate device are mitigated by the sterilization validation performed.

SUMMARY OF NONCLINICAL TESTING

Provided below are the nonclinical test methodologies performed to demonstrate the subject devices met the acceptance criteria of the standard.

{9}------------------------------------------------

Summary of Nonclinical Testing Table
Test MethodologyPurposeAcceptance CriteriaResults
Manual Cleaning ValidationFDA GuidanceReprocessing Medical Devices inHealth Care Settings: ValidationMethods and LabelingThe purpose of this test is tovalidate that the cleaninginstructions provided in theInstructions for Use appropriatelyclean the tray, and to ensure theProtein assay Acceptance criterion:No color change = absence of proteinresidueSensitivity = 1 µg protein residuePass
(issued March 2015)Referenced from K201688sterilization cycle will beeffective.Total organic carbon assay Acceptancecriterion:results < limit of quantification = 0.1 ppm;Assay limit of detection = 0.015 ppmPass
Bacterial Endotoxin Testing,USP <85>Referenced from K201688The purpose of this test is tovalidate that the cleaninginstructions provided in theInstructions for Use appropriatelyclean the tray, and to ensure theBET level meets FDA expectation(≤ 20 EU/device)Acceptance criterion:Endotoxin results ≤ 20 EU/devicePass
Sterilization Validationincluding sterilant penetrationand dry time validationANSI/AAMI/ISO 17665-1ANSI/AAMI/ISO 17665-2Referenced from K201688The purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Use appropriatelysterilize the tray and contents.Acceptance criterion:3 consecutive half-cycles performed foreach of 3 sizes of trays demonstratecomplete inactivation of all biologicindicators:A minimum SAL of 10-6 is achieved if theInstructions for Use are followedPass
Dry timeReferenced from K201688The purpose of this test is tovalidate that the sterilizationinstructions listed in theInstructions for Use appropriatelydry the wrapped tray for storage.Acceptance criterion:Using pre-cycle and post-cycle weights,the weight gain after drying will be ≤ 3%Pass
Life Cycle / Simulated Use LifeValidation FDA GuidanceReprocessing Medical Devices inHealth Care Settings: ValidationMethods and Labeling(issued March 2015)Referenced from K201688The purpose of this test is tovalidate the service life of thetrays as stated in the Instructionsfor Use.Acceptance criteria:Visual inspection, component dimensionalfit verification, functional closure/latchverification for 250 use cyclesPass
Biocompatibility of SubjectDeviceCytotoxicity testingANSI/AAMI/ISO 10993-5ANSI/AAMI/ISO 10993-12The purpose of this test is toevaluate the cytotoxicity potentialof the test article using an in vitrocell culture assay.Acceptance criterion:Non-cytotoxic if ≤ 50% of L-929 cellsexposed to extract of device showrounding and lysisPass

In summary, the nonclinical testing provided for these devices met the acceptance criteria for each standard and test methodology used to evaluate the devices as shown in the table above.

No clinical data were included in this submission.

{10}------------------------------------------------

CONCLUSION

The conclusions drawn from the nonclinical tests demonstrate that the subject device in 510(k) submission K212404, S.I.N. Instrument Kits, is as safe, as effective, and perform as well as or better than the legally marketed predicate device cleared under K201688.

{11}------------------------------------------------

Indications for Use Statements

Indications for Use Statements
Subject DeviceK212404S.I.N. Instrument KitsS.I.N. - Sistema de Implante Nacional S.A.Indications for Use for Safe Drill Epikut KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N.use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Safe Drill Epikut Kit and the associated instruments is 154 grams.The weight of the empty Safe Drill Epikut Kit is 138 grams.
Indications for Use for Unitite Surgical KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N.use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Unitite Surgical Kit and the associated instruments is 620 grams.The weight of the empty Unitite Surgical Kit is 520 grams.
Indications for Use for Epikut Surgical KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N.use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Epikut Surgical Kit and the associated instruments is 605 grams.The weight of the empty Epikut Surgical Kit is 520 grams.
Indications for Use for Epikut Surgical Guided KitS.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.N.use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.The combined weight of the Epikut Surgical Guided Kit and the associated instruments is 808 grams.The weight of the empty Epikut Surgical Guided Kit is 650 grams.
Predicate DeviceK201688S.I.N. Instrument KitsS.I.N. - Sistema de Implante Nacional S.A.S.I.N. Instrument Kits are intended to be used to enclose other medical devices that are to be sterilized by a health care provider. S.I.N. Instrument Kits are intended to allow sterilization of the enclosed medical devices. S.I.the use of a wrap that is FDA cleared to maintain the sterility of the enclosed devices.The kits are to be enclosed in a sterilization wrap that is FDA cleared for the indicated cycles, and moist heat (steam) sterilized using the following cycle:Gravity displacement - Exposure at 121 °C for 30 minutes and 30 minutes dry time.S.I.N. Instrument Kits are intended for sterilization of non-porous loads.S.I.N. Instrument Kits are recommended not to be stacked during sterilization.
Safe Drill Unitite KitThe combined weight of the Safe Drill Unitite Kit and the associated instruments is 304 grams.The weight of the empty Safe Drill Unitite Kit is 150 grams.
Safe Drill SW KitThe combined weight of the Safe Drill SW Kit and the associated instruments is 278 grams.The weight of the empty Safe Drill SW Kit is 138 grams.
Prosthetic KitThe combined weight of the Prosthetic Kit and the associated instruments is 332 grams.The weight of the empty Prosthetic Kit is 160 grams.
Rotatory Expanding KitThe combined weight of the Rotatory Expanding Kit and the associated instruments is 276 grams.The weight of the empty Rotatory Expanding Kit is 133 grams.
Bone Expander KitThe combined weight of the Bone Expander Kit and the associated instruments is 974 grams.The weight of the empty Bone Expander Kit is 367 grams.
Indications for Use Statements
Sinus Lift KitThe combined weight of the Sinus Lift Kit and the associated instruments is 808 grams.The weight of the empty Sinus Lift Kit is 370 grams.
Osteotome KitThe combined weight of the Osteotome Kit and the associated instruments is 957 grams.The weight of the empty Osteotome Kit is 350 grams.
Unitite Surgical KitThe combined weight of the Unitite Surgical Kit and the associated instruments is 1126 grams.The weight of the empty Unitite Surgical Kit is 515 grams.
Strong SW Surgical KitThe combined weight of the Strong SW Surgical Kit and the associated instruments is 698 grams.The weight of the empty Strong SW Surgical Kit is 130 grams.
Tryon Surgical Kit KCHE 03The combined weight of the Tryon Surgical Kit KCHE 03 and the associated instruments is 1127 grams.The weight of the empty Tryon Surgical Kit KCHE 03 is 520 grams.
Tryon Surgical Kit KCHE 04The combined weight of the Tryon Surgical Kit KCHE 04 and the associated instruments is 1138 grams.The weight of the empty Tryon Surgical Kit KCHE 04 is 523 grams.
Unitite Surgical Guided KitThe combined weight of the Unitite Surgical Guided Kit and the associated instruments is 1434 grams.The weight of the empty Unitite Surgical Guided Kit is 650 grams.
Strong SW Surgical Guided KitThe combined weight of the Strong SW Surgical Guided Kit and the associated instruments is 1399 grams.The weight of the empty Strong SW Surgical Guided Kit is 647 grams.
Zygomatic Surgical KitThe combined weight of the Zygomatic Surgical Kit and the associated instruments is 1150 grams.The weight of the empty Zygomatic Surgical Kit is 464 grams.

ization of the enclosed medical devices. S.I.N. Instrument Kits require the

ization of the enclosed medical devices. S.I.N. Instrument Kits require the

ization of the enclosed medical devices. S.I.N. Instrument Kits require the

ization of the enclosed medical devices. S.I.N. Instrument Kits require the

rilization of the enclosed medical devices. S.I.N. Instrument Kits require

{12}------------------------------------------------

510(k) Summary Page 7 of 8

S.I.N Dental Implant System K212404

{13}------------------------------------------------

Technological Characteristics

Subject DevicePrimary Predicate DeviceComparison
AttributeK212404S.I.N. Instrument KitsS.I.N. - Sistema de Implante Nacional S.A.K201688S.I.N. Instrument KitsS.I.N. - Sistema de Implante Nacional S.A.
Product CodeKCTKCTSame
DesignRigid polymer base, lid, and removable inner trayRigid polymer base, lid, and removable inner traySame
MaterialsPolysulfone (base, tray, lid)Medical grade silicone (grommets)Polysulfone (base, tray, lid)Medical grade silicone (grommets)Same
Materials Compatible with Sterilization MethodYesYesSame
PerforatedYes; allows moist heat (steam) penetration to achieve sterilizationYes; allows moist heat (steam) penetration to achieve sterilizationSame
ReusableYesYesSame
Number of Sizes35Similar
Total Number of Configurations414Similar
Dimensions andVent to Volume RatioTrayTraySimilar
COESD 02, Safe Drill Epikut TrayLength x Width x Height, mm118.0 x 78.25 x 29.2Vent to Volume Ratio0.0083 cm²/cm³COUSD 02, Safe Drill Unitite TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0089 cm²/cm³
COSU 04, Unitite Surgical TrayLength x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³COWSD 02, Safe Drill SW TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0089 cm²/cm³
COSE 01, Epikut Surgical TrayLength x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³COTMEC, Prosthetic TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0089 cm²/cm³
COSEG 01I, Epikut Surgical Guided TrayLength x Width x Height, mm142 x 206 x 72Vent to Volume Ratio0.0083 cm²/cm³COER, Rotatory Expanding TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0089 cm²/cm³
COEXP, Bone Expander TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0100 cm²/cm³
COLEV, Sinus Lift TrayLength x Width x Height, mm215 x 100 x 33.5Vent to Volume Ratio0.0100 cm²/cm³
COOST, Osteotome TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0100 cm²/cm³
COSU 03, Unitite Surgical TrayLength x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³
COSW 02, Strong SW Surgical TrayLength x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³
COHE 03, Tryon Surgical Tray KCHE 03Length x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³
COHE 04, Tryon Surgical Tray KCHE 04Length x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0086 cm²/cm³
COSUG 02, Unitite Surgical Guided TrayLength x Width x Height, mm142 x 206 x 72Vent to Volume Ratio0.0083 cm²/cm³
COSWG 02, Strong SW Surgical Guided TrayLength x Width x Height, mm165 x 190 x 55Vent to Volume Ratio0.0083 cm²/cm³
COKZ, Zygomatic Surgical TrayLength x Width x Height, mm113.7 x 75.7 x 29.5Vent to Volume Ratio0.0131 cm²/cm³
ReusableYesReusable up to 250 cyclesYesReusable up to 250 cyclesSame
Use Life TestingDisassembled, cleaned, assembled, sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch) verificationDisassembled, cleaned, assembled, sterilizedVisual inspectionComponent dimensional fit verificationFunctional closure (lid-base latch) verificationSame
Sterilization Method
SterilantMoist heat (steam)Moist heat (steam)Same
CyclesWrapped in a sterilizable wrap that is FDA cleared for the indicated cycleGravity displacement30 minute exposure at 121 °C (250 °F)30 minute drying timeGravity displacementSame
Sterile BarrierSterilization wrap, FDA cleared for indicated method and cycleSterilization wrap, FDA cleared for indicated method and cycleSame

S.I.N Dental Implant System

K212404

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).