Search Results
Found 2 results
510(k) Data Aggregation
(273 days)
Radiometer Medicals ApS
The ABL90 FLEX PLUS System is an in vitro diagnostic, portable, automated analyzer that quantitatively measures electrolytes (cK+, cNa+, cCa2+), glucose, and lactate in heparinized arterial and venous whole blood.
The ABL90 FLEX PLUS System is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient, or point-of-care setting. These tests are only performed under a physician's order.
Potassium (cK+): Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.
Sodium (cNa+): Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, delydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.
Calcium (cCa2+): Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
Glucose (cGlu): Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Lactate (cLac): The lactate measure the concentration of lactate. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidity of the blood).
The ABL90 FLEX PLUS System consists of the ABL90 FLEX PLUS analyzer, sensor cassette and solution pack consumables, and related accessories for the analyzers. The ABL90 FLEX PLUS is a portable, automated system intended for in vitro testing of samples of balanced heparinized whole blood for electrolytes (cK+, cNa*, cCa²), glucose, and lactate. The ABL90 FLEX PLUS System has an automated sample inlet mechanism, which can collect blood through two different measuring modes: the S65 syringe mode and the SP65 short probe mode.
The provided text is a 510(k) Summary for the ABL90 FLEX PLUS System, an in vitro diagnostic device. This document focuses on demonstrating substantial equivalence to a legally marketed predicate device (ABL90 FLEX) rather than proving the device meets specific acceptance criteria as might be defined for a novel AI/ML device.
Therefore, much of the requested information regarding acceptance criteria for AI/ML performance, study design (test set, ground truth establishment, expert adjudication, MRMC studies, standalone performance, training set details) is not applicable to this type of device and its regulatory submission.
The document primarily proves the analytical performance of the new device is comparable to the predicate device through various analytical studies.
Here's a breakdown of the applicable information based on the provided text, and an explanation of why other requested information is not present:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present "acceptance criteria" in a pass/fail table for each performance metric in the way it might for a novel AI/ML device. Instead, it presents analytical performance data (linearity, precision, detection, method comparison, interference) which is implicitly compared against pre-defined internal specifications or what is considered acceptable for the similar predicate device. The goal is to show the new device performs equivalently to the predicate.
Below is a summary of the reported device performance from the tables in the document. The "Acceptance Criteria" column cannot be fully populated as precise numerical thresholds are not explicitly stated as "acceptance criteria" in this 510(k) summary, but are rather implied by the successful demonstration of performance often within CLSI guidelines and comparable to the predicate.
Parameter (Unit) | Test Category | Reported Performance (Range / Values) | Implicit Acceptance Criteria (based on predicate equivalence and CLSI) |
---|---|---|---|
cCa2+ (mg/dL) | Linearity | Slope: 0.883, Intercept: 0.445, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
LoQ | 1.26 | Established lower limit of reliable quantitation. | |
Precision (QC) | Repeatability SD: 0.003-0.014, CV%: 0.1-0.3 | Low SD and CV%, demonstrating consistent results. | |
Precision (Blood) | Repeatability SD: 0.003-0.022, CV%: 0.06-0.45 | Low SD and CV%, demonstrating consistent results within biological samples. | |
Method Comp. (Bias at MD) | S65: 0.001-0.003, SP65: 0.003-0.009 | Low bias compared to the predicate device, indicating equivalent measurements. | |
cK+ (mEq/L) | Linearity | Slope: 1.001, Intercept: 0.027, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
LoQ | 1.6 | Established lower limit of reliable quantitation. | |
Precision (QC) | Repeatability SD: 0.00-0.01, CV%: 0.1-0.2 | Low SD and CV%, demonstrating consistent results. | |
Precision (Blood) | Repeatability SD: 0.007-0.026, CV%: 0.14-0.96 | Low SD and CV%, demonstrating consistent results within biological samples. | |
Method Comp. (Bias at MD) | S65: 0.002-0.004, SP65: 0.004-0.008 | Low bias compared to the predicate device, indicating equivalent measurements. | |
cNa+ (mEq/L) | Linearity | Slope: 1.001, Intercept: -0.642, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
LoQ | 99 | Established lower limit of reliable quantitation. | |
Precision (QC) | Repeatability SD: 0.1-0.2, CV%: 0.1 | Low SD and CV%, demonstrating consistent results. | |
Precision (Blood) | Repeatability SD: 0.061-0.194, CV%: 0.05-0.14 | Low SD and CV%, demonstrating consistent results within biological samples. | |
Method Comp. (Bias at MD) | S65: 0.265-0.290, SP65: 0.221-0.259 | Low bias compared to the predicate device, indicating equivalent measurements. | |
cGlu (mg/dL) | Linearity | Slope: 1.032, Intercept: -1.073, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
LoD/LoQ | LoD: 5, LoQ: 5 | Established lower limits of detection and reliable quantitation. | |
Precision (QC) | Repeatability SD: 0.3-1.3, CV%: 0.5-1.1 | Low SD and CV%, demonstrating consistent results. | |
Precision (Blood) | Repeatability SD: 0.207-2.221, CV%: 0.35-0.85 | Low SD and CV%, demonstrating consistent results within biological samples. | |
Method Comp. (Bias at MD) | S65: -0.460 to -2.028, SP65: -0.663 to -2.045 | Low bias compared to the predicate device, indicating equivalent measurements. | |
cLac (mg/dL) | Linearity | Slope: 0.971, Intercept: -0.433, R^2: 1.000 | R^2 near 1.0, slope near 1.0, intercept near 0, demonstrating linearity over the reportable range. |
LoD/LoQ | LoD: -0.3, LoQ: 2 | Established lower limits of detection and reliable quantitation. (Note: Negative LoD likely a calculation artifact near zero) | |
Precision (QC) | Repeatability SD: 0.2-0.3, CV%: 0.3-1.1 | Low SD and CV%, demonstrating consistent results. | |
Precision (Blood) | Repeatability SD: 0.177-0.379, CV%: 0.75-2.25 | Low SD and CV%, demonstrating consistent results within biological samples. | |
Method Comp. (Bias at MD) | S65: -0.116 to 0.013, SP65: -0.156 to -0.169 | Low bias compared to the predicate device, indicating equivalent measurements. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set (for performance validation):
- Linearity: The specific number of samples tested for linearity is not explicitly stated as 'N' values in Table 1 but ranges presented (e.g., 1.896-11.146 for cCa2+) imply a sufficient number of points across the range were used.
- Detection (LoB, LoD, LoQ): Not explicitly stated as 'N' values in Table 2.
- Precision (using stable, aqueous ampoule-based QC material): Varies per parameter/level, but generally 243-244 replicates (N) per parameter/level.
- Precision (using blood): Varies per parameter/mode/interval, ranging from 2 to 202 replicates (N).
- Method Comparison:
- Arterial blood (S65 mode): 221-225 samples (N) across parameters.
- Arterial blood (SP65 mode): 214-218 samples (N) across parameters.
- Venous blood (S65 mode): 231-234 samples (N) across parameters.
- Venous blood (SP65 mode): 219-225 samples (N) across parameters.
- Combined (S65 mode): 436-441 samples (N) for combined arterial/venous.
- Combined (SP65 mode): 420-425 samples (N) for combined arterial/venous.
- Interference: "Large panel of likely interferents" for paired-difference study; dose-response studies for significant interferents. Specific sample sizes for each interferent are not detailed in the summary.
- Data Provenance: The document states that precision studies using QC material were conducted at "three external sites." Method comparison and precision studies using blood were conducted using both arterial and venous blood, and in both sample collection modes. The country of origin for the data (patients or samples) is not specified in this summary. The studies are described as "analytical performance testing," implying they are prospective or controlled laboratory studies rather than retrospective analysis of existing clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable: This device is an in vitro diagnostic (IVD) analyzer that quantitatively measures analytes. Its performance is evaluated against reference measurement procedures or highly controlled materials, not by expert interpretation of images or clinical cases requiring expert consensus or qualifications. Ground truth is established by the reference method itself or the known concentration of QC materials.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable: As this is an IVD device measuring quantitative analytes, there is no expert adjudication process in this context, unlike an AI/ML device interpreting medical images. Performance is determined by comparison to reference methods or statistical analysis against known values.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable: This is an IVD analyzer, not an AI/ML device that assists human readers. Therefore, an MRMC study is not relevant to its regulatory approval process.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Partially Applicable (in a different sense): The ABL90 FLEX PLUS System is a standalone automated analyzer. Its performance is measured directly (algorithm only, if you consider the device's internal measurement algorithm) against reference methods or known concentrations, without a human-in-the-loop interpretation being the primary output that's being evaluated for accuracy. The results presented (linearity, precision, method comparison) are representative of its standalone performance.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Quantitative Reference Methods / Known Concentrations:
- Linearity/Detection: Ground truth is established by preparing samples with known, precise concentrations across the measurement range, or by the inherent properties of the measurement system for LoB/LoD/LoQ.
- Precision: Ground truth is the expected value of the quality control (QC) materials or the prepared blood samples, or simply the reproducibility of measurements on the same sample.
- Method Comparison: Ground truth is the measurement from the legally marketed predicate device (ABL90 FLEX, specifically "ABL90 FLEX PLUS analyzer as it was designed at the time of the clearance of K160153") that the new device is being compared against. This device itself serves as the "reference method" for substantial equivalence.
- Interference: Ground truth is the expected measurement of known samples, with and without the interferent, using a reference method, to identify if the interferent causes a clinically significant deviation.
8. The sample size for the training set
- Not Applicable (in the AI/ML sense): This document describes the analytical validation of a traditional IVD device, not an AI/ML algorithm. There is no "training set" in the machine learning sense for this type of submission. The device is a physical instrument with established chemical/electrochemical measurement principles.
9. How the ground truth for the training set was established
- Not Applicable: As there is no "training set" in the AI/ML context, this question is not relevant. The device's internal parameters and calibration would be established through a manufacturing and calibration process, not through a "training" phase with a ground truth dataset in the way an AI model is trained.
Ask a specific question about this device
(246 days)
Radiometer Medicals ApS
The ABL90 FLEX PLUS System is an in vitro diagnostic, portable, automated analyzer that quantitatively measures pH, blood gas (p02), Oximetry (s02, ctHb, FCOHb, FCOHb, FMetHb, and FHHb), in heparinized arterial and venous whole blood.
The ABL90 FLEX PLUS System is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient, or point-of-care setting. These tests are only performed under a physician's order.
pH and pO2: pH and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.
sO2: Oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin plus reduced hemoglobin.
ctHb (Total Hemoglobin): Total hemoglobin measure the hemoglobin content of whole blood for the detection of anemia.
FO2Hb: Oxyhemoglobin as a fraction of total hemoglobin.
FCOHb: Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.
FMetHb: Methemoglobin as a fraction of total hemoglobin.
FHHb: Reduced hemoglobin as a fraction of total hemoglobin.
The ABL90 FLEX PLUS System consists of the ABL90 FLEX PLUS analyzer, sensor cassette and solution pack consumables, and related accessories for the analyzers. The sensor cassettes, solution packs and related accessories are compatible with both analyzers. Multiple versions of the sensor cassettes are available. The sensor cassette versions vary in the maximum number of tests and availability of sensors for use. The solution pack is available in two versions, differing in the number of activities available.
The FDA 510(k) summary for the Radiometer ABL90 FLEX PLUS System provides detailed information about the device's analytical performance testing to demonstrate its substantial equivalence to its predicate device.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the reported performance metrics (linearity, detection limits, precision, bias) and the comparison to the predicate device. The goal is to show that the ABL90 FLEX PLUS System performs comparably to the predicate (ABL90 FLEX PLUS, K160153) and adheres to recognized standards (CLSI guidelines).
The summary does not explicitly present a table of "acceptance criteria" and "reported performance" side-by-side in a single formatted table. However, the various tables throughout the "Analytical Performance Testing Summary" section serve this purpose by presenting the measured performance of the ABL90 FLEX PLUS System against unstated, but implied, acceptable ranges or a comparison to the predicate.
Here's a synthesized representation of the reported device performance for key analytical parameters, effectively serving as the "reported device performance":
Parameter | Performance Aspect | Reported Value/Range (ABL90 FLEX PLUS System) | Implied Acceptance Criteria (via comparison to predicate and CLSI guidelines) |
---|---|---|---|
pH | Linearity Interval | 6.605-7.997 | Consistent with clinical requirements and predicate's performance. |
Lower LoQ | 6.754 | Detectable and quantifiable at clinically relevant low levels. | |
Upper LoQ | 7.843 | Detectable and quantifiable at clinically relevant high levels. | |
Precision (Repeatability SD) | 0.001-0.003 (blood) | Low variability, suitable for clinical use. | |
Bias (Method comparison to predicate) | -0.003 | Minimal bias from predicate, within clinical acceptable limits. | |
pO2 | Linearity Interval | 0.81-75.41 kPa (or mmHg equivalent) | Consistent with clinical requirements and predicate's performance. |
LoQ | 7.7 mmHg (1.02 kPa) | Detectable and quantifiable at clinically relevant low levels. | |
Precision (Repeatability SD) | 0.197-1.91 (blood/QC) | Low variability, suitable for clinical use. | |
Bias (Method comparison to predicate) | -0.454 to 0.344 | Minimal bias from predicate, within clinical acceptable limits. | |
ctHb | Linearity Interval | 0.068-27.660 g/dL | Consistent with clinical requirements and predicate's performance. |
LoQ | 0.09 g/dL | Detectable and quantifiable at clinically relevant low levels. | |
Precision (Repeatability SD) | 0.01-0.091 (blood/QC) | Low variability, suitable for clinical use. | |
Bias (Method comparison to predicate) | 0.015-0.126 | Minimal bias from predicate, within clinical acceptable limits. | |
Oximetry (sO2, FO2Hb, FCOHb, FMetHb, FHHb) | Linearity Interval | Ranges provided for each (e.g., sO2: 2.18-100.22%) | Consistent with clinical requirements and predicate's performance. |
LoQ | Ranges provided for each (e.g., sO2: 1.4%) | Detectable and quantifiable at clinically relevant low levels. | |
Precision (Repeatability SD) | Low variability reported across all oximetry parameters (blood/QC) | Low variability, suitable for clinical use. | |
Bias (Method comparison to predicate) | Minimal bias reported across all oximetry parameters | Minimal bias from predicate, within clinical acceptable limits. | |
Interference | Various interferents (Intralipid, Bilirubin, etc.) | Reported impact on results, indicating levels where interference was not significant or error messages occurred. | Acceptable performance with common interferents, or clear warnings for known interferences. |
Key takeaway for "Acceptance Criteria": The general acceptance criterion for this 510(k) submission is to demonstrate "substantial equivalence" to the predicate device (ABL90 FLEX PLUS, K160153). While explicit numerical acceptance criteria are not presented in this summary document, the testing aims to show that the new device's performance (linearity, detection, precision, bias, interference) is comparable to the predicate and/or meets recognized clinical and analytical standards as outlined in CLSI guidelines.
2. Sample Sizes Used for the Test Set and Data Provenance
- Linearity Testing: Numbers of samples are not explicitly stated for linearity testing, but it was conducted "in general accordance with CLSI EP06... and EP39."
- Detection Capability (LoB, LoD, LoQ): Numbers of samples are not explicitly stated.
- Precision (using stable, aqueous ampoule-based QC material):
- N = 243 or 244 for each QC ampoule level and parameter.
- Data Provenance: Testing occurred at three external sites. The specific countries of origin are not mentioned, but "external sites" suggests a multi-site study. This appears to be a prospective study, as it's part of the premarket submission.
- Precision (using blood):
- N varies by parameter and test interval, ranging from 4 to 188 samples.
- Data Provenance: Not explicitly stated, but implies collected from blood samples (human derived). Likely prospective data collected for the study.
- Method Comparison (Bias):
- N varies by parameter, blood type, and mode (S65/SP65), ranging from 26 to 235 samples (arterial/venous blood).
- Data Provenance: Not explicitly stated, but implies collected from patient blood samples. This would be prospective data collected specifically for the method comparison study.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This device (ABL90 FLEX PLUS System) is an in vitro diagnostic (IVD) analytical instrument. The ground truth for its performance is established through reference methods, defined concentrations of analytes in quality control materials, and comparison to a legally marketed predicate device, not typically through human expert adjudication of images or clinical outcomes that require multiple medical professionals.
Therefore, the concept of "experts establishing ground truth" in the way it might apply to an AI imaging device (e.g., radiologists reviewing scans) is not directly applicable here. The "experts" would be the laboratory personnel and analytical chemists who perform the testing and ensure adherence to CLSI guidelines. Their qualifications are implicitly assumed to be appropriate for performing such technical laboratory studies.
4. Adjudication Method for the Test Set
Not applicable for an IVD analytical instrument. Ground truth is established by reference methods, certified materials, and comparison with a predicate device, not by expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
Not applicable. MRMC studies are typically performed for imaging devices or diagnostic aids where human interpretation is a key component, often comparing AI-assisted vs. unassisted human performance. This device is an automated, quantitative analytical instrument.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the performance data presented (linearity, detection/quantitation, precision, bias, interference) are all measures of the standalone analytical performance of the ABL90 FLEX PLUS System. The device provides quantitative measurements autonomously without continuous human interpretation required for each result.
7. The Type of Ground Truth Used
The ground truth used for this device's analytical performance studies are:
- Reference materials/Certified Analytes: For linearity, detection, and precision testing. These are materials with known, precisely measured concentrations of the analytes (pH, pO2, ctHb, sO2, etc.).
- Predicate Device Measurements: For method comparison/bias studies, the measurements from the legally marketed ABL90 FLEX PLUS (K160153) served as the comparator (or "ground truth" to determine bias relative to the predicate).
- CLSI Guidelines: The studies adhere to relevant Clinical and Laboratory Standards Institute (CLSI) guidelines (e.g., EP06, EP39, EP17-A2, EP05-A3, EP09c, EP07, EP37), which define accepted methodologies and performance characteristics for IVD devices.
8. The Sample Size for the Training Set
Not applicable. This document describes the performance testing for a finished IVD product, not the development or training of a machine learning model. IVD devices like the ABL90 FLEX PLUS System are based on established analytical principles (potentiometry, optical, spectrophotometry) and calibrated using defined reference materials, not "trained" on a dataset in the AI sense.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of this device's analytical principles. Ground truth for calibration and development of such instruments is established through rigorous analytical chemistry methods using highly purified and characterized reference standards.
Ask a specific question about this device
Page 1 of 1