K Number
K160153
Device Name
ABL90 FLEX PLUS
Date Cleared
2016-11-04

(287 days)

Regulation Number
862.1120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood.

The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.

It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order.

Bilirubin measurements on the ABL90 FLEX PLUS analyzer are intended to aid in assessing the risk of kernicterus in neonates.

pH, pO2 and pCO2: pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa+); sodium measurements are used in the diagnosis and treatment of aldosteronism. diabetes insipidus, adrenal hypertension, Addison's disease, dehydration,inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCl-): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): qlucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoqlycemia, and of pancreatic islet cell carcinoma.

Lactate (cLac): The lactate measurements measure the concentration of lactate in plasma. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.)

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoqlobin and oxyhemoglobin plus reduced hemoqlobin.

FO2Hb: oxyhemoqlobin as a fraction of total hemoqlobin.

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoqlobin as a fraction of total hemoglobin.

Fraction of Fetal Hemoglobin (FHbF): FHbF indicates the amount of fetal hemoglobin. FHbF is seldom used clinically.

Device Description

The ABL90 FLEX PLUS is a portable, automated system intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, chloride, glucose, lactate, neonatal bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO-Hb, FCOHb, FMetHb, FHHb and FHbF).

The manufacturer of the ABL90 FLEX PLUS is Radiometer Medical ApS.

The ABL90 FLEX PLUS consists of an instrument with a sensor cassette and a solution pack as the main accessories. Multiple models of sensor cassettes are available.

The various sensor cassette models for different parameter combinations. For each parameter combination, models allowing for different test load are available. The solution pack is available in two models differing in the number of tests available.

AI/ML Overview

The provided text describes the ABL90 FLEX PLUS analyzer, an in vitro diagnostic device. The submission is for a design change to an existing device, the ABL90 FLEX, with the introduction of the ABL90 FLEX PLUS which includes a mechanized inlet module (AutoInlet) and a Short Probe Mode.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The core of the performance evaluation is a "Method comparison of ABL90 FLEX PLUS Short Probe mode versus ABL90 FLEX syringe mode with inlet clip" and "Imprecision" studies. The reported device performance is that all acceptance criteria were met.

Method Comparison Acceptance Criteria & Performance:

ParameterAcceptance Criteria (Linear Regression)Reported Performance
SlopeBetween 0.95 and 1.05Slopes were between 0.95 and 1.05.
Coefficient of Determination (R²)> 0.97Coefficients of determination R² were > 0.97.
InterceptspH: ±0.75Intercepts were within acceptance criteria for all parameters.
pO2: ±11 mmHg
pCO2: ±4.5 mmHg
Cl-: ±11 mM
Na+: ±15 mM
K+: ±0.5 mM
Ca2+: ±0.5 mM
Glucose: ±0.6 mmol/L
Lactate: ±0.4 mmol/L
tHb: ±1.5 g/dL
sO2: ±10%
FO2Hb: ±10%
FCOHb: ±1%
FMetHb: ±1%
FHHb: ±2.4%
FHbF: ±21%
Neonatal bilirubin: ±28 μmol/L

Imprecision Acceptance Criteria & Performance:

ParameterAcceptance CriteriaReported Performance
Clinical PrecisionThe same or better clinical precision than originally determined for ABL90 FLEX (K092686 and K132691).All within-run and total imprecisions were within the acceptance criteria.
Within-run (Sr)Pooled across sites must be the same or better than originally determined for ABL90 FLEX (K092686 and K132691) at a 95% confidence level using a Chi-square test. Specific values are itemized in tables for capillary, syringe, and short probe modes.All within-run and total imprecisions were within the acceptance criteria.
Total Imprecision (ST)Pooled across sites must be the same or better than originally determined for ABL90 FLEX (K092686 and K132691) at a 95% confidence level using a Chi-square test. Specific values are itemized in tables for capillary, syringe, and short probe modes.All within-run and total imprecisions were within the acceptance criteria.

2. Sample size used for the test set and the data provenance

  • Method Comparison Test Set: "more than 40 samples (N) per parameter"
  • Data Provenance: Samples were "heparinized, leftover whole blood samples (analyzed 2-3 hours post draw)." The specific country of origin is not explicitly stated, but the submission is from Radiometer Medical ApS in Denmark, suggesting the study likely occurred in a European context or by their internal methods. The study is retrospective as it uses "leftover whole blood samples".

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable to this type of in vitro diagnostic device and study. The ground truth for this device is based on measurements from a predicate device (ABL90 FLEX) or a reference instrument, not expert consensus.

4. Adjudication method for the test set

Not applicable. The study compares quantitative measurements between two devices, not subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in vitro diagnostic device for quantitative measurements, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the performance studies described (Method Comparison and Imprecision) evaluate the standalone analytical performance of the ABL90 FLEX PLUS device (ABL90 FLEX PLUS Short Probe mode and ABL90 FLEX PLUS for imprecision). The comparisons are against a predicate device or reference instrument, not involving human interpretation.

7. The type of ground truth used

  • Method Comparison: The predicate device, ABL90 FLEX syringe mode with inlet clip, served as the comparative "truth". The study assessed if the new ABL90 FLEX PLUS Short Probe mode yields equivalent results to this established method.
  • Imprecision: "The total imprecision for all parameters except neonatal bilirubin was calculated as the imprecision of the bias towards a reference value determined for each sample on an ABL90 FLEX reference instrument." For neonatal bilirubin, the ground truth source is not explicitly defined beyond "aqueous solutions," but implies a known concentration.

8. The sample size for the training set

Not explicitly stated. For in vitro diagnostic devices, "training set" is not a standard term as it is in machine learning. The studies described are performance verification studies for a medical device. If there was an internal development phase for calibration or algorithm adjustment, that data is not detailed here.

9. How the ground truth for the training set was established

Not applicable directly as this is not an ML/AI model with a "training set" in the conventional sense. For the performance studies, ground truth (or reference values) for comparison were established by:

  • Method Comparison: Measurements from the predicate device (ABL90 FLEX syringe mode with inlet clip).
  • Imprecision: Measurements on an "ABL90 FLEX reference instrument" for most parameters, and "aqueous solutions" for neonatal bilirubin (implying known concentrations).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol, composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

RADIOMETER MEDICAL APS SOEREN BØEGESTRAND SENIOR REGULATORY AFFAIRS SPECIALIST AAKANDEVEJ 21 BROENSHOEJ DK-2700, DENMARK

November 4, 2016

Re: K160153

Trade/Device Name: ABL 90 FLEX PLUS Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases (PCO2, PO2) and blood pH test system Regulatory Class: II Product Code: CHL, CEM, JGS, CGZ, CGA, KHP, GHS, GKR, KQI, JIX, JJY, MQM, JFP Dated: October 05, 2016 Received: October 07, 2016

Dear Soeren Boegestrand:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K160153

Device Name ABL90 FLEX PLUS

Indications for Use (Describe)

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood.

The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.

It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order.

Bilirubin measurements on the ABL90 FLEX PLUS analyzer are intended to aid in assessing the risk of kernicterus in neonates.

pH, pO2 and pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acidbase disturbances.

Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa+): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCl-): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Lactate (cLac): The lactate measure the concentration of lactate in plasma. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.)

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin plus reduced hemoglobin.

FO2Hb: oxyhemoglobin as a fraction of total hemoglobin.

{3}------------------------------------------------

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoglobin as a fraction of total hemoglobin.

Fraction of Fetal Hemoglobin (FHbF): FHbF indicates the amount of fetal hemoglobin. FHbF is seldom used clinically.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

1. Submitter and contact information

Submitter
Company Name:Radiometer Medical ApS
ER Number:3002807968
Address:Aakandevej 212700 BroenshoejDenmark
Phone:+45 3827 3827
Fax:+45 3827 2727
Contact person:Soeren Boegestrand
E-mail:soren.bogestrand@radiometer.dk

Contact Person

Name:Soeren Boegestrand
Function:Senior Regulatory Affairs Specialist
E-mail:soren.bogestrand@radiometer.dk
Phone:+45 3827 3852
Fax:+45 3827 2727
Date prepared
Date:November 4, 2021

2. a. Device Information

Device Name:ABL90 FLEX PLUS
Common Name:Blood gases (pCO2, pO2) and blood pH test system.

6

Classification:

Classification nameCFR SectionDevice ClassProduct Code
Electrode measurement, blood-gases(pCO2, pO2) and blood pH862.1120IICHL
Potassium test system862.1600IICEM
Sodium test system862.1665IIJGS
Chloride test system862.1170IICGZ
Glucose test system862.1345IICGA
Carboxyhemoglobin assay864.7425IIGHS
Automated hemoglobin system864.5620IIGKR
Fetal hemoglobin assay864.7455IIKQI
Calibrator862.1150IIJIX
Calcium test system862.1145IIJFP

{5}------------------------------------------------

Quality control material (assayed andunassayed)862.1660IJJY
Bilirubin (total and unbound) in theneonate test system862.1113IMQM
Lactic acid test system862.1450IKHP

{6}------------------------------------------------

2. b. Device Description

Instrument name, manufacturer, models and accessories

The ABL90 FLEX PLUS is a portable, automated system intended for in vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, potassium, sodium, chloride, glucose, lactate, neonatal bilirubin, and co-oximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions FO-Hb, FCOHb, FMetHb, FHHb and FHbF).

The manufacturer of the ABL90 FLEX PLUS is Radiometer Medical ApS.

The ABL90 FLEX PLUS consists of an instrument with a sensor cassette and a solution pack as the main accessories. Multiple models of sensor cassettes are available.

The various sensor cassette models for different parameter combinations. For each parameter combination, models allowing for different test load are available. The solution pack is available in two models differing in the number of tests available.

2. c. Purpose of submission

The purpose of this submission is to seek clearance for a design change to the existing ABL90 FLEX analyzer. The design change will result in the introduction of an additional variant of the ABL90 FLEX analyzer called ABL90 FLEX PLUS. To accommodate this, the ABL90 FLEX analyzer has been changed as outlined below:

  • . The inlet module on the ABL90 FLEX PLUS analyzer has been mechanized, the mechanized inlet module is in multiple places in this 510(k) referred to as the AutoInlet. The opening and closing of the inlet is either button activated using the touch screen or in certain cases automatic when appropriate. The mechanized inlet module subsequently calls for the following derived changes listed below.
  • Addition of the Short Probe Mode measuring mode on the ABL90 FLEX PLUS. In ABL90 . FLEX, a little plastic clip (ABL90 FLEX Inlet Clip) component to be placed on the ABL90 FLEX inlet when using non-Radiometer samplers is incorporated. The purpose of the inlet clip is to shorten the maximum travel of the inlet into the sampler and thereby reduce the sample volume required. The use of an inlet clip for blood measurements is no longer relevant with the introduction of the Short Probe Mode. On the ABL90 FLEX PLUS, the short probe position has been built directly into the analyzer to replace the use of the inlet clip allowing the user the option of the standard syringe mode or the Short Probe Mode.
  • . Modifications to software to support both ABL90 FLEX and ABL90 FLEX PLUS. The modifications in the software relevant to this application are the support of the mechanized inlet module in ABL90 FLEX PLUS (including the Short Probe Mode), as well as the ability to distinquish and detect the type of inlet (manual or mechanized) an analyzer is fitted with.
  • Instructions for Use for ABL90 FLEX PLUS have been established. .

Except for the above identified modifications and minor differences in visual appearance the ABL90 FLEX and ABL90 FLEX PLUS analyzers are identical. Both analyzers use the same software version and have the same accessories.

{7}------------------------------------------------

3. Intended Use/Indications for use

The ABL90 FLEX PLUS analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood qases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood.

The ABL90 FLEX PLUS analyzer is intended for use by trained technologists, nurses, physicians and therapists.

It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order.

Bilirubin measurements on the ABL90 FLEX PLUS analyzer are intended to aid in assessing the risk of kernicterus in neonates.

pH, pO2 and pCO2: pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Potassium (cK+): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa+); sodium measurements are used in the diagnosis and treatment of aldosteronism. diabetes insipidus, adrenal hypertension, Addison's disease, dehydration,inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCl-): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): qlucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus and idiopathic hypoqlycemia, and of pancreatic islet cell carcinoma.

Lactate (cLac): The lactate measurements measure the concentration of lactate in plasma. Lactate measurements are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood.)

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sO2: oxygen saturation, more specifically the ratio between the concentration of oxyhemoqlobin and oxyhemoglobin plus reduced hemoqlobin.

FO2Hb: oxyhemoqlobin as a fraction of total hemoqlobin.

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoqlobin as a fraction of total hemoglobin.

Fraction of Fetal Hemoglobin (FHbF): FHbF indicates the amount of fetal hemoglobin. FHbF is seldom used clinically.

{8}------------------------------------------------

4. Predicate device: ABL90 FLEX (K132691) Substantial Equivalence

The ABL90 FLEX PLUS analyzer is substantially equivalent in Intended Use, fundamental scientific technology, features, and characteristics to the predicate devices:

Predicate Device: 510(k) Number, Name, Device Manufacturer:

K132691, ABL90 FLEX, Radiometer Medical ApS K131988, ABL90 FLEX, Radiometer Medical ApS

Similarities
IssueSE DevicePredicate Devices (K132691, K131988)
ParametersSamepH, pO2, pCO2, sodium, potassium, calcium, chloride, glucose, lactate, sO2, FO2Hb, FCOHb, FMetHb, FHHB, FHbF, neonatal bilirubin.
Measuring methodSamePotentiometry, Amperometry, Optical pO2 ,Spectrophotometry
Calibration MethodSameTwo-point liquid calibration
Intended UseSameThe ABL90 FLEX analyzer is an in vitro diagnostic, portable, automated analyser that quantitatively measures, pH, blood gases, electrolytes, glucose, lactate and oximetry in heparinized whole blood, and neonatal bilirubin in heparinized capillary whole blood.The ABL90 FLEX analyzer is intended for use by trained technologists, nurses, physicians and therapists.It is intended for use in a laboratory environment, near patient or point-of-care setting.These tests are only performed under a physician's order.Bilirubin measurements on the ABL90 FLEX analyzer are intended to aid in assessing the risk of kernicterus in neonates.
Intended use siteSameLaboratory and point-of-care.
Differences
IssueSE DevicePredicate Devices (K132691, K131988)
InletMechanized inletManual inlet
Short probe modeYesNo

{9}------------------------------------------------

5. Performance Characteristics

No performance data are affected by the change. The existing performance data still apply.

Verification of the ABL90 FLEX PLUS showed that the precision performance of all three modes of the ABL90 FLEX PLUS is equivalent to the precision performance of ABL90 FLEX for the 17 parameters available on the analyzers: pH, pCO2, pO2, cCa2+, cCli, cK*, cGlu, cLac, ctHb, sO3, FO3Hb, FHHb, FMetHb, FCOHb, FHbF, and neonatal bilirubin.

The existing clinical precision performance claims in the manual for ABL90 FLEX are also applicable to ABL90 FLEX PLUS.

Verification of the ABL90 FLEX PLUS Short Probe mode showed that measurements performed with ABL90 FLEX PLUS in Short Probe mode are equivalent to measurements performed with ABL90 FLEX in syringe mode with inlet clip for the 17 parameters available on the analyzers: pH, pCO₂, pO₂, cCa²†, cCl˙, cK†, cNa†, cGlu, cLac, ctHb, sO₂, FO₂Hb, FHHb, FMetHb, FCOHb, FHbF, and neonatal bilirubin.

The existing clinical method comparison claims in the manual for ABL90 FLEX syringe mode are applicable to ABL90 FLEX PLUS Short Probe mode.

{10}------------------------------------------------

6. Summary of Design Control activities

Risk Assessment:

We conducted an FMEA risk analysis and mitigated all identified hazards to As Low As Reasonably Practicable (ALARP) per ISO 14971, and verified mitigations by using test protocols. Results met predefined acceptance criteria

Performance testing:

IssueAcceptancecriteriaVerificationmethodResult and Pass/FailComments
Methodcomparison ofABL90 FLEXPLUS ShortProbe modeversus ABL90FLEX syringemode with inletclipData analysis by linearregression must result in:Slope between 0.95and 1.05 Coefficient ofdeterminationR2>0.97 Intercepts: pH:±0.75pO2:±11 mmHgpCO2:±4.5 mmHgCl-:±11 mMNa+:±15 mMK+:±0.5 mMCa2+:±0.5 mMGlucose: ±0.6 mmol/LLactate: ±0.4 mmol/LtHb:±1.5 g/dLsO2:±10%FO2Hb:±10%FCOHb:±1%FMetHb:±1%FHHb:±2.4%FHbF:±21%Neonatalbilirubin: ±28 μmol/LIn-house method comparisonstudy of ABL90 FLEX PLUSShort Probe mode versusABL90 FLEX syringe mode withinlet clip with more than 40samples (N) per parameterspanning the measuring range.Each sample was analyzed insinglicate. For each parameter,less than 20% of the sampleswere spiked.Samples were heparinized,leftover whole blood samples(analyzed 2-3 hours postdraw).Data analysis by linearregression showed:Slopes werebetween 0.95 and1.05. Coefficients ofdetermination R2were >0.97. Intercepts werewithin acceptancecriteria. Pass.Existing methodcomparisonclaims for ABL90FLEX syringemode apply toABL90 FLEXPLUS ShortProbe mode.
IssueAcceptancecriteriaVerificationmethodResult and Pass/FailComments
ImprecisionThe same or betterclinical precision thanoriginally determined forABL90 FLEX (K092686and K132691).The within-run ( $S_r$ ) andtotal imprecision ( $S_T$ )pooled across sites mustbe the same or betterthan originally determinedfor ABL90 FLEX (K092686and K132691) at a 95%confidence level using aChi-square test.The acceptance criteriahave been itemized in thebelow table.Point-of-care precision studieson ABL90 FLEX PLUS syringe,short probe, and capillarymode were performed at 3point-of care sites over at least20 days, with 2 runs per dayand 2 replicates per run.Samples were spiked wholeblood except for neonatalbilirubin for which sampleswere aqueous solutions. ForFHbF, the sample material wascord blood.Fresh whole blood sampleswere prepared each test day.To eliminate the sample-to-sample imprecisioncontribution, the totalimprecision for all parametersexcept neonatal bilirubin wascalculated as the imprecision ofthe bias towards a referencevalue determined for eachsample on an ABL90 FLEXreference instrument.All within-run and totalimprecisions werewithin the acceptancecriteria.PassExisting clinicalprecision claimsfor ABL90 FLEXapply to ABL90FLEX PLUS.

{11}------------------------------------------------

{12}------------------------------------------------

Imprecision acceptance criteria

AnalyteLevelSr (within-run)%CV (within-run)St (total)%CV (total)
Capillary mode
7.270.00300.040.00560.08
pH7.380.00280.040.00650.09
7.470.00320.040.00670.09
280.712.540.863.07
pCO₂ (mmHg)420.761.811.132.69
611.041.701.512.48
450.561.241.212.69
pO₂ (mmHg)770.941.221.862.42
2052.451.204.822.35
0.580.0213.620.0325.52
cCa²⁺ (mmol/L)1.210.0090.740.0292.40
1.890.0361.900.0703.70
940.380.401.411.50
cCl⁻ (mmol/L)1080.430.401.831.69
1220.370.301.691.39
4.80.051.040.122.50
cK⁺ (mmol/L)6.60.040.610.091.36
1200.360.300.900.75
cNa⁺ (mmol/L)1410.480.341.150.82
1510.480.321.150.76
0.780.045.130.1114.10
cGlu (mmol/L)5.50.091.640.234.18
15.10.231.520.775.10
2.10.094.290.178.10
cLac (mmol/L)16.10.442.731.569.69
3.70.071.890.092.43
ctHb (g/dL)150.151.000.332.20
21.60.200.930.281.30
760.300.390.610.80
sO₂ (%)930.200.220.450.48
1000.110.110.200.20
3.20.041.250.299.06
FCOHb (%)37.30.040.110.320.86
3.40.113.240.339.71
FMetHb (%)10.00.131.300.292.90
6.60.263.940.507.58
FHHb (%)23.80.261.090.602.52
FO₂Hb (%)740.300.410.670.91
890.280.310.440.49
970.200.210.290.30
FHbF (%)422.866.817.0716.83
642.894.527.0811.06
360.922.561.774.92
Neonatal bilirubin(µmol/L)1271.851.462.552.01
5237.071.359.411.80
AnalyteLevelSr (within-run)%CV (within-run)ST (total)%CV (total)
Syringe mode
pH7.270.00230.030.00470.06
7.370.00150.020.00450.06
7.470.00130.020.00600.08
pCO2 (mmHg)280.662.360.863.07
420.771.831.102.62
601.151.921.833.05
pO2 (mmHg)450.270.600.902.00
760.841.111.431.88
2041.820.893.541.74
cCa2+ (mmol/L)0.580.0030.520.0223.79
1.210.0090.740.0262.15
1.890.0090.480.0422.22
cCl (mmol/L)940.250.271.381.47
1080.420.391.821.69
1220.240.201.321.08
cK+ (mmol/L)4.80.040.830.081.67
6.60.030.450.081.21
1190.360.300.770.65
cNa+ (mmol/L)1410.480.340.900.64
1510.360.240.900.60
0.730.045.480.0912.33
cGlu (mmol/L)5.50.050.910.234.18
14.70.181.220.644.35
2.20.073.180.188.18
cLac (mmol/L)16.20.513.151.9411.98
3.80.071.840.123.16
ctHb (g/dL)150.110.730.171.13
21.70.140.650.200.92
750.290.390.490.65
sO2 (%)930.170.180.440.47
1000.110.110.220.22
3.20.061.880.288.75
FCOHb (%)38.00.080.210.320.84
3.40.092.650.236.76
FMetHb (%)10.10.080.790.302.97
6.70.172.540.405.97
FHHb (%)24.30.251.030.411.69
720.280.390.590.82
FO2Hb (%)910.320.350.450.49
970.180.190.260.27
FHbF (%)412.325.666.3715.54
644.567.138.7613.69
Neonatal bilirubin(µmol/L)361.193.312.125.89
1271.351.062.481.95
5254.460.859.151.74
AnalyteLevelSr (within-run)%CV (within-run)ST (total)%CV (total)
Short probe mode
pH7.270.0020.030.0050.06
7.370.0020.020.0050.06
7.470.0010.020.0060.08
pCO2 (mmHg)280.662.360.863.07
420.771.831.102.62
601.151.921.833.05
pO2 (mmHg)450.270.600.902.00
760.841.111.431.88
2041.820.893.531.73
cCa²+ (mmol/L)0.580.0030.520.023.79
1.210.0090.740.032.15
1.890.0090.480.042.22
cCl⁻ (mmol/L)940.250.271.381.47
1080.420.391.821.69
1220.240.201.321.08
cK⁺ (mmol/L)4.80.040.830.081.67
6.60.030.450.081.21
cNa⁺ (mmol/L)1190.360.300.770.65
1410.480.340.900.64
1510.360.240.900.60
cGlu (mmol/L)0.730.045.480.0912.33
5.50.050.910.234.18
14.70.181.220.654.42
cLac (mmol/L)2.20.073.180.188.18
16.20.513.151.9411.98
ctHb (g/dL)3.80.071.840.123.16
150.110.730.171.13
21.70.140.650.200.92
sO2 (%)750.290.390.490.65
930.170.180.440.47
1000.110.110.220.22
FCOHb (%)3.20.061.880.288.75
38.00.080.210.320.84
FMetHb (%)3.40.092.650.226.47
10.10.080.790.302.97
FHHb (%)6.70.172.540.405.97
24.30.251.030.411.69
FO₂Hb (%)720.280.390.590.82
910.320.350.450.49
970.180.190.260.27
FHbF (%)412.325.666.3715.54
644.567.138.7613.69
Neonatal bilirubin(µmol/L)361.193.312.125.89
1271.351.062.481.95
5254.460.850.141.74

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

7. Conclusion

Based on the substantial equivalence comparison and the results of the conducted performance evaluations it has been concluded that the ABL90 FLEX PLUS is as safe and effective as the predicate device.

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.