Search Filters

Search Results

Found 11 results

510(k) Data Aggregation

    K Number
    K951459
    Date Cleared
    1997-07-14

    (837 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate is used to identify and enumerate the percentage of CD4+ human T lymphocytes in whole blood by flow cytometry.

    Identification and enumeration of abnormal levels of CD4lymphocytes may be clinically significant in the prognosis of secondary immunodeficiency diseases (acquired immunodeficiency syndrome [AIDS]).

    AIDS is characterized by a severe reduction in T helper CD3+CD4+) lymphocytes and a decrease in the CD4:CD8 ratio. The CD4:CD8 ratio has been used as a predictor of time to the development of AIDS. CD4 counts should also be determined using a CD3/CD4 combination reagent to eliminate monocyte contamination (CD3-CD4+dim). CD4+lymphocytes (expressed as an absolute number, a percentage of lymphocytes or as a ratio of CD4+ to CD8+ T lymphocytes) represents the best single predictor of the progression of AIDS. CD4+ lymphocyte numbers usually start to decline relatively soon after human immunodeficiency virus (HIV) infection.

    Device Description

    Ortho-mune OKT4A Monoclonal Antibody (Murine) FITC Conjugate contains the purified monoclonal antibody OKT4A conjugated to the fluorochrome fluorescein isothiocyanate.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Ortho-mune™OKT™4A (CD4) FITC Conjugate

    This document outlines the acceptance criteria and performance data for the Ortho-mune™OKT™4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate, as submitted in K951459. The study aimed to demonstrate substantial equivalence to the predicate device, CD4 (Leu™-3a) FITC.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for substantial equivalence are inferred from the performance characteristics presented, specifically demonstrating comparable results to the predicate device and acceptable reproducibility and linearity. Given the nature of a 510(k) submission for substantial equivalence, the "acceptance criteria" are implicitly met by showing that the new device performs "as well as" or "equivalently to" the predicate device on key performance metrics.

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Ortho-mune OKT4A)
    Equivalence to PredicateMean and range of percent CD4+ cells for normal and AIDS/ARC populations should be comparable to the predicate device. Regression analysis (Ortho-mune vs. Predicate) should show a strong correlation (R-value close to 1) and a slope near 1 with a small intercept.Normal Donors (N=206): Mean % CD4+ = 47.0 (Range: 12.4 - 68.7). Predicate (LEU-3a) Mean % CD4+ = 45.1 (Range: 14.6 - 64.6).
    AIDS/ARC Patients (N=88): Mean % CD4+ = 18.3 (Range: 1.0 - 55.1). Predicate (LEU-3A) Mean % CD4+ = 17.2 (Range: 0.6 - 51.6).
    Linear Regression: Y = 0.969 + 1.019(X), R = 0.985 (combining normal and AIDS/ARC populations). This demonstrates strong correlation and a slope very close to 1, indicating equivalence.
    Within-Laboratory ReproducibilityCoefficients of Variation (CV) should be acceptably low across low, normal, and high CD4 levels, and 95% Confidence Intervals (CI) should demonstrate precision.Mean Percent Positive & CV across 3 sites (N=10 per site, 10 replicates each):
    • Low CD4: Mean % (Site 1: 9.60, Site 2: 9.36, Site 3: 8.86); CV (Site 1: 7.60, Site 2: 6.12, Site 3: 5.47)
    • Normal CD4: Mean % (Site 1: 47.38, Site 2: 47.19, Site 3: 48.54); CV (Site 1: 2.33, Site 2: 2.00, Site 3: 2.21)
    • High CD4: Mean % (Site 1: 56.13, Site 2: 55.83, Site 3: 56.50); CV (Site 1: 1.93, Site 2: 1.79, Site 3: 2.09)
      Demonstrated excellent within-laboratory reproducibility. |
      | Between-Laboratory Reproducibility | Coefficients of Variation (CV) should be acceptably low across low, normal, and high CD4 levels, and 95% Confidence Intervals (CI) should demonstrate precision across different laboratories. | Mean Percent Positive & CV across 3 sites (N=10 samples, 10 replicates each):
    • Low CD4: Mean % = 9.25; CV = 4.09; +/- 95% CI = 1.63
    • Normal CD4: Mean % = 47.70; CV = 1.53; +/- 95% CI = 3.14
    • High CD4: Mean % = 56.15; CV = 0.60; +/- 95% CI = 1.46
      Demonstrated excellent between-laboratory reproducibility. |
      | Linearity | Linear regression analysis (observed vs. expected CD4+ counts) should show slopes indistinguishable from 1 and R-values close to 1,000 (R=1). | Slopes and R-values for 4 donors and overall:
    • Donor 1: Slope = 0.999 (CI 0.003), Intercept = 6.099 (CI 5.242), R = 1.000
    • Donor 2: Slope = 1.000 (CI 0.001), Intercept = 0.395 (CI 2.226), R = 1.000
    • Donor 3: Slope = 0.999 (CI 0.003), Intercept = 4.449 (CI 4.176), R = 1.000
    • Donor 4: Slope = 1.000 (CI 0.001), Intercept = 0.285 (CI 1.316), R = 1.000
    • All Donors: Slope = 0.999 (CI 0.001), Intercept = 2.777 (CI 1.655), R = 1.000
      Demonstrates linear performance across 20 to 9000 cells/uL. |

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Equivalence Study:
      • 206 normal donors
      • 88 AIDS/ARC patients
      • Total: 294 whole blood specimens.
    • Sample Size for Reproducibility Studies:
      • 10 normal donors for within-laboratory study (processed to create low, normal, high CD4 populations).
      • 10 normal donors for between-laboratory study (processed to create low, normal, high CD4 populations).
    • Sample Size for Linearity Study:
      • Specimens from 4 normal donors (processed to produce low, normal, high numbers of lymphocyte subsets via concentration/dilution).
    • Data Provenance: The studies were conducted at three external, geographically distinct sites. The data is prospective for the purpose of this submission, though the underlying specimens are from existing populations (normal donors and AIDS/ARC patients). The country of origin of the data is not explicitly stated but is implied to be the United States given the submission to the FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    No external experts were explicitly used to establish ground truth in the traditional sense of consensus reading. The "ground truth" for the comparative effectiveness study was the measurement obtained by the predicate device, CD4 (Leu-3a) FITC, using flow cytometry. For the reproducibility and linearity studies, the ground truth was derived from the measurements of the device itself (Ortho-mune OKT4A) against established laboratory methods (e.g., automated hematology analyzer for total lymphocyte count).

    4. Adjudication Method for the Test Set

    No adjudication method was used for the test set. The study directly compared the measurements obtained by the new device to those obtained by the predicate device for equivalence, and relied on statistical analysis of direct measurements for reproducibility and linearity.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC study was done, as this device is an immunophenotyping reagent for flow cytometry, not an imaging device requiring human reader interpretation in the same manner. The "human readers" (flow cytometry operators) would be operating the instruments rather than interpreting complex images. The focus was on the performance of the reagent itself. Therefore, comparing human reader improvement with and without AI assistance is not applicable here.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies evaluating the performance of the Ortho-mune OKT4A (CD4) FITC Conjugate can be considered "standalone" in the sense that they assess the direct analytical performance of the reagent itself, as measured by a flow cytometer. The flow cytometry process is automated in terms of cell counting and fluorescent signal detection, and the reagent's performance is objectively measured. While human operators are involved in sample preparation and instrument setup, the core performance metrics (staining, enumeration, reproducibility, linearity) characterize the reagent's analytical capability independently of direct "human interpretation" of results in the way an imaging AI might be evaluated.

    7. The Type of Ground Truth Used

    • For Equivalence Study: The predicate device, CD4 (Leu-3a) FITC, was used as the reference standard. The goal was to establish substantial equivalence, meaning the new device performs comparably to a legally marketed device. This is a form of comparative ground truth.
    • For Reproducibility Studies: The measurements from the Ortho-mune OKT4A reagent itself, over multiple replicates and sites, were used to assess the consistency of its own performance. This relies on the inherent precision of the laboratory measurements.
    • For Linearity Study: A combination of an automated hematology analyzer (for total lymphocyte count) and the CYTORONABSOLUTE flow cytometer (with the Ortho-mune reagent for percent positive CDx cells) was used. The "expected" values were calculated based on serial dilutions and established counts, which serves as a form of derived or calculated ground truth.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of machine learning or AI development. This device is a monoclonal antibody reagent, and its development typically involves laboratory-based optimization (e.g., antibody titration, conjugation efficiency) rather than algorithm training with large datasets. The studies described here are for validation of the final product.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, the concept of a "training set" and associated ground truth establishment for AI/machine learning is not applicable to this type of medical device (an immunophenotyping reagent). The development process would have involved standard biological and chemical methods to characterize the antibody and its conjugate, ensuring specificity and reactivity, but not a "training set" with ground truth in the AI context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K951632
    Date Cleared
    1997-03-06

    (699 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OK-COMBO Control is intended for use as negative control for immunophenotying of human lymphocytes in whole blood by flow cytometry using Orthomune OK-COMBO immunophenotyping reagents. Ortho-mune OK-COMBO immunophenotyping reagents are blends of two purified monoclonal antibodies conjugated to fluorescein isothiocyanate and phycoerythrin, respectively. The negative control is used to check for nonspecific background staining and to set the negative/positive rergions of the fluorescent cytograms.

    Device Description

    Ortho-mune OK-COMBO Control IgG2-FITC/IgG2-PE Monoclonal Antibody (Murine) is a blend of murine monoclonal antibodies, not specific for human cellular antigens, conjugated to the fluorochromes, fluorescein isothiocyanate (FITC) and phycoerythrin (PE), respectively.

    AI/ML Overview

    Here's an analysis of the provided text, outlining the acceptance criteria and study details for the Ortho-mune™ OK-COMBO Control IgG2-FITC/IgG2-PE Monoclonal Antibody (Murine):

    Device: Ortho-mune™ OK-COMBO Control IgG2-FITC/IgG2-PE Monoclonal Antibody (Murine)
    Intended Use: Negative control for immunophenotyping of human lymphocytes in whole blood by flow cytometry using Ortho-mune OK-COMBO immunophenotyping reagents. Used to check for non-specific background staining and to set negative/positive regions of fluorescent cytograms.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" with numerical thresholds for performance. Instead, the study aims to demonstrate substantial equivalence to a predicate device (Simultest™ IMK Plus Reagent B-Control) and acceptable reproducibility and linearity. Therefore, the "acceptance criteria" are implied by the comparative performance and good internal consistency.

    Performance MetricAcceptance Criteria (Implied by Equivalence/Acceptability)Reported Device Performance (Ortho-mune OK-COMBO Control)
    Equivalence to Predicate Device
    Mean % Events in Positive Region (Normal Donors)Similar mean % and range as the predicate (Simultest IMK Control), with small ΔMean and CI. (e.g., ΔMean approaching 0, CI covering 0).FITC: Mean 0.32% (SD 0.50, Range 0-1.32). ΔMean vs. predicate: 0.07 (CI 0.07).
    PE: Mean 0.38% (SD 0.78, Range 0-1.95). ΔMean vs. predicate: 0.13 (CI 0.12).
    Mean % Events in Positive Region (AIDS/ARC Patients)Similar mean % and range as the predicate (Simultest IMK Control), with small ΔMean and CI.FITC: Mean 0.43% (SD 0.34, Range 0-1.10). ΔMean vs. predicate: 0.11 (CI 0.08).
    PE: Mean 0.44% (SD 0.56, Range 0-1.55). ΔMean vs. predicate: 0.13 (CI 0.12).
    Reproducibility
    Within Laboratory Reproducibility (CV)Low coefficient of variation (CV) across different sample types, suggesting consistent results within a laboratory. (No specific threshold given, but values generally below ~1% are good in flow cytometry).CD19 Depleted: 0.403%. CD4 Depleted: 0.151%. CD8 Depleted: 0.084%. Normal: 0.215%. (These are "All Sites" CVs based on aggregated variance; individual site CVs are lower).
    Between Laboratory Reproducibility (CV)Low coefficient of variation (CV) across different laboratories, suggesting consistent results between labs. (No specific threshold given, but generally
    Ask a Question

    Ask a specific question about this device

    K Number
    K964754
    Date Cleared
    1997-01-22

    (56 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OKB19A PE Conjugate is intended for use in identification and enumeration of CD19+ human B lymphocytes in whole blood by flow cytometry.

    Device Description

    Ortho-mune OKB19A Monoclonal Antibody (Murine) Phycoerythrin (PE) Conjugate contains the purified monoclonal antibody OKB19A conjugated to the fluorochrome phycoerythrin.

    AI/ML Overview

    The provided document describes the performance characteristics of the Ortho-mune™ OKB™19A (CD19) Monoclonal Antibody (Murine) Phycoerythrin Conjugate, a diagnostic device for identifying and enumerating CD19+ human B lymphocytes.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state numerical acceptance criteria in a formal table. Instead, the acceptance is based on demonstrating "substantial equivalence" to a predicate device (Monoclonal Antibody Test for B Cells [Anti-Leu-12 (CD19) PE (Phycoerythrin)]). The performance is assessed through various studies. Below is a table summarizing the reported device performance and implicitly derived acceptance criteria based on equivalence:

    Performance MetricAcceptance Criteria (Implicit, based on equivalence to predicate)Reported Device Performance (Ortho-mune OKB19A)
    Comparative Performance (Normal Donors)Mean % CD19+ and Range % CD19+ should be comparable to the predicate device.Mean % CD19+: 13.3, Range % CD19+: 1.3 - 31.1
    Comparative Performance (AIDS/ARC Patients)Mean % CD19+ and Range % CD19+ should be comparable to the predicate device.Mean % CD19+: 10.2, Range % CD19+: 0.4 - 41.9
    Linear Regression (Relative to Predicate)High correlation (R value close to 1) and a slope close to 1, indicating similar measurements.R = 0.909, Equation: Y = 0.711 + 0.88(X) (for combined data)
    Within-Laboratory Reproducibility (Normal)Coefficient of Variation (CV) should be acceptable for flow cytometry assays, similar to predicate performance (not explicitly stated, but assumed to be low).Mean % Positive: 15.330, Site A CV: 5.082, Site B CV: 4.635, Site C CV: 4.545
    Within-Laboratory Reproducibility (Low)CV should be acceptable.Mean % Positive: 1.134, Site A CV: 24.182, Site B CV: 20.793, Site C CV: 22.106
    Within-Laboratory Reproducibility (High)CV should be acceptable.Mean % Positive: 26.760, Site A CV: 3.438, Site B CV: 3.401, Site C CV: 3.447
    Between-Laboratory Reproducibility (Normal)Low CV across sites.Overall Across Site CV: 0.976
    Between-Laboratory Reproducibility (Low)Low CV across sites.Overall Across Site CV: 1.466
    Between-Laboratory Reproducibility (High)Low CV across sites.Overall Across Site CV: 0.782
    Linearity (across lymphocyte count range)Slopes indistinguishable from 1 and R values close to 1.000.Slopes between 1.001 and 1.004, R values of 1.000 (for each donor and all combined)

    2. Sample Sizes Used for the Test Set and Data Provenance:

    • Comparative Performance Study (against predicate):
      • Sample Size: 205 normal donors and 87 AIDS/ARC patients.
      • Data Provenance: The study was conducted at "three external, geographically distinct sites." The specific countries are not mentioned, but the distributed nature suggests a multi-center study. The data is prospective as specimens were stained and analyzed during the study.
    • Reproducibility Studies:
      • Sample Size: 10 normal donors for each of the three independent laboratories.
      • Data Provenance: Three independent laboratories (suggesting different locations, likely within the same country as the submitter, USA, but not explicitly stated). The data is prospective as samples were processed and stained for the study.
    • Linearity Study:
      • Sample Size: Four normal donors.
      • Data Provenance: Not explicitly stated where the donors or testing took place, but implied as part of the overall development and testing by the submitter. The data is prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not applicable to this device. The device is a diagnostic reagent for flow cytometry, which measures cell populations based on antibody binding. The "ground truth" for the comparative study is the measurement obtained by the predicate device and the clinical classification of the patients (normal donor vs. AIDS/ARC). For reproducibility and linearity, the ground truth is statistical consistency and expected values determined by dilutions and other analytical methods, not expert consensus on an image or clinical case.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in studies where human readers are making subjective diagnoses or classifications from images/data, and discrepancies need to be resolved. This study involves objective measurements from a flow cytometer.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. MRMC studies are associated with evaluating the effectiveness of diagnostic tools where multiple human readers interpret cases. This study is comparing a new reagent's performance to an existing reagent using instrumental measurements.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Yes, the studies described are essentially a standalone performance evaluation of the Ortho-mune OKB19A reagent. The flow cytometer analyzes the stained cells, and the resulting data (percentage of CD19+ cells) is the output of the "algorithm" (the staining and detection process). Human involvement is in sample preparation, running the instrument, and data analysis, but the core measurement itself is automated and doesn't involve human interpretation of complex patterns in the way an AI algorithm for image analysis would. The performance is assessed purely on the device's ability to accurately identify and enumerate the target cells, comparing its output to a predicate device's output.

    7. Type of Ground Truth Used:

    • Comparative Performance: The ground truth was effectively the measurement obtained by the predicate device (Monoclonal Antibody Test for B Cells [Anti-Leu-12 (CD19) PE (Phycoerythrin)]) on the same samples. The clinical categorisation of "normal donor" and "AIDS/ARC patient" also serves as a clinical context for evaluating performance.
    • Reproducibility Studies: The ground truth was based on the statistical expectation of consistent measurements when the same samples are tested multiple times (within-laboratory) and across different laboratories (between-laboratory). The "true" value for the percentage of CD19+ cells in the prepared samples was the reference point.
    • Linearity Study: The ground truth was the expected values calculated by multiplying serial dilutions by the hematology analyzer-derived buffy coat lymphocyte count and the CYTORONABSOLUTE-derived lymphocyte subset percent positive.

    8. Sample Size for the Training Set:

    This information is not applicable as this is not an AI/Machine Learning device that requires a training set. The device is a monoclonal antibody conjugate; its performance is based on its biochemical properties and interaction with target cells.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the reason stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K963902
    Date Cleared
    1996-11-29

    (63 days)

    Product Code
    Regulation Number
    864.7340
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K950625
    Date Cleared
    1996-05-13

    (458 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OK-COMBO CD4-FITC/CD8-PE is intended for use in identification and enumeration of CD4+ and CD8+ human T lymphocytes in whole blood by flow cvtometry. The intended use is the same as the intended use of the predicate device. Simultest CD4/CD8 (Leu-3a/2a) commercially distributed by Becton Dickinson Immunocytometry Systems.

    Device Description

    Ortho-mune OK-COMBO CD4-FITC/CD8-PE (OKT4A/OKT8) Monoclonal Antibody (Murine) is a blend of the individual purified monoclonal antibodies OKT4A and OKT8 conjugated to the fluorochromes fluorescein isothiocyanate and phycoerythrin respectively.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    EquivalenceDemonstrate equivalence to the predicate device (Simultest CD4/CD8) in identification and enumeration of CD4+ and CD8+ human T lymphocytes in whole blood.Regression Analysis:
    • CD4+: Y = 1.886 + 1.08500 (Ortho-mune) vs. Predicate. Correlation coefficient R = 0.971.
    • CD8+: Y = 0.112 + 0.84400 (Ortho-mune) vs. Predicate. Correlation coefficient R = 0.955.

    Mean % Positive Stained Cells (Normal Donors, N=190):

    • CD4+: Ortho-mune: 47.5% (Range: 11.6-70.8); Predicate: 44.1% (Range: 13.1-61.2)
    • CD8+: Ortho-mune: 27.8% (Range: 10.9-72.8); Predicate: 29.5% (Range: 12.5-69.1)

    Mean % Positive Stained Cells (AIDS/ARC Patients, N=83):

    • CD4+: Ortho-mune: 18.2% (Range: 0.1-60.4); Predicate: 15.8% (Range: 0.2-50.4)
    • CD8+: Ortho-mune: 56.1% (Range: 12.6-81.9); Predicate: 58.9% (Range: 22.5-88.4)

    Conclusion: Performance shown to be equivalent to the predicate device. |
    | Reproducibility| Acceptable within and between laboratory reproducibility for determining total CD4+ and CD8+ lymphocyte percentages. | Within Laboratory Reproducibility (N=11 donors, 3 sites):

    • CD4+ Low (mean 11.232%): CVs from 6.701% to 8.268%
    • CD4+ Normal (mean 49.455%): CVs from 2.443% to 3.118%
    • CD4+ High (mean 67.528%): CVs from 2.093% to 3.360%
    • CD8+ Low (mean 7.521%): CVs from 7.186% to 15.735%
    • CD8+ Normal (mean 32.782%): CVs from 3.102% to 4.244%
    • CD8+ High (mean 56.234%): CVs from 2.462% to 3.392%

    Between Laboratory Reproducibility (N=11 donors, 3 sites):

    • CD4+ Low: CV 5.496%
    • CD4+ Normal: CV 1.189%
    • CD4+ High: CV 1.729%
    • CD8+ Low: CV 5.453%
    • CD8+ Normal: CV 0.427%
    • CD8+ High: CV 1.354%

    Conclusion: Acceptable within and between laboratory reproducibility demonstrated. |
    | Linearity | Demonstrate linear performance for both total CD4+ and CD8+ lymphocyte subsets across a range of lymphocyte counts. | Linear performance demonstrated for both total CD4+ and CD8+ lymphocyte subsets across a lymphocyte count range of 20 cells/ul to 18,676 cells/ul. Slopes were indistinguishable from 1 and R values were 1.000. |

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set for Equivalence (Comparison with Predicate Device):

      • Sample Size: 190 normal donors and 83 AIDS/ARC patients (total 273 specimens).
      • Data Provenance: The study was conducted at "three external, geographically distinct sites." The countries of origin are not specified, but the use of "normal donors" and "AIDS/ARC patients" suggests clinical samples. The data is prospective, collected specifically for this study.
    • Test Set for Reproducibility:

      • Sample Size: 11 normal donors. Samples from these donors were processed to create low, normal, and high relative percent CD4+ and CD8+ cells. Each sample was stained in replicates of 10.
      • Data Provenance: Not explicitly stated, but samples were processed and analyzed at "three independent laboratories." This implies the samples were real human whole blood specimens. The data is prospective.
    • Test Set for Linearity:

      • Sample Size: 4 normal donors. Specimens were processed to produce samples with low, normal, and high numbers of lymphocyte subsets. Each sample was stained in triplicate.
      • Data Provenance: Not explicitly stated, but implies real human whole blood specimens. The data is prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is an in vitro diagnostic reagent for flow cytometry, which measures cell populations. The "ground truth" for this type of device is the measurement obtained from the predicate device and the internal consistency/characteristics (reproducibility, linearity) of the new device itself. There are no human expert adjudicators for establishing a "ground truth" in the way one would for diagnostic imaging. The comparison is against an established method and assessed by statistical measures.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, this is a quantitative measurement device, not an imaging or qualitative diagnostic device requiring expert adjudication. The comparison with the predicate device was based on statistical regression analysis and direct comparison of mean percentages and ranges.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not an AI/imaging device that involves human readers or interpretation of complex cases. It is an in vitro diagnostic reagent that identifies and enumerates specific cell populations using flow cytometry.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The "device" (reagent) is evaluated in a standalone manner by comparing its output (percentage of positive stained cells) directly against that of a predicate device and by assessing its intrinsic performance characteristics (reproducibility, linearity). The analysis is performed by a flow cytometer, and the interpretation of the results from the flow cytometer (i.e., the percentages) is then compared between the new reagent and the predicate reagent. There is no "human-in-the-loop performance" that would alter the quantitative output of the reagent and cytometer system itself.

    7. The Type of Ground Truth Used

    • For Equivalence Study: The "ground truth" was established by the predicate device's measurements (Simultest CD4/CD8). The study aimed to show that the new device's measurements were statistically equivalent to the predicate device's measurements.
    • For Reproducibility and Linearity Studies: The "ground truth" was established based on internal consistency and expected biological behavior. For reproducibility, the ground truth is that repeated measurements of the same sample should yield similar results (low variance). For linearity, the ground truth is that the measured cell counts should scale proportionally with dilution.

    8. The Sample Size for the Training Set

    Not applicable. This device is a monoclonal antibody reagent used in flow cytometry, not a machine learning or AI model. Therefore, there is no "training set" in the context of algorithm development. Its performance is based on the inherent specificity and binding characteristics of the antibodies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K950568
    Date Cleared
    1996-05-13

    (460 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OK-COMBO CD3-FITC/CD4-PE is intended for use in identification and enumeration of CD3+ and CD4+ human T lymphocytes in whole blood by flow cytometry. The intended use is the same as the intended use of the predicate device, Simultest CD3/CD4 (Leu-4/3a) commercially distributed by Becton Dickinson Immunocytometry Systems.

    Device Description

    Ortho-mune OK-COMBO CD3-FITC/CD4-PE (OKT3/OKT4A) Monoclonal Antibody (Murine) is a blend of the individual purified monoclonal antibodies OKT3 and OKT4A conjugated to the fluorochromes fluorescein isothiocyanate and phycoerythrin respectively.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Ortho-mune OK-COMBO CD3-FITC/CD4-PE (OKT3/OKT4A) Monoclonal Antibody (Murine) device, which is intended for the identification and enumeration of CD3+ and CD4+ human T lymphocytes in whole blood by flow cytometry. The study aims to demonstrate substantial equivalence to a predicate device, Simultest™ CD3/CD4 Ortho-mune™ OK-COMBO (Leu™-4 / 3a) CD3-FITC/CD4-PE (OKT™3 /OKT4A) Monoclonal Antibody (Murine).

    Here's a breakdown of the requested information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by achieving a statistically significant correlation between the performance of the new device (Ortho-mune OK-COMBO CD3-FITC/CD4-PE) and the predicate device (Simultest CD3/CD4), as well as between different flow cytometers used with the new device. The benchmark for significance is a p-value (SL) of 0.0001, indicating a very high probability of correlation. The Pearson Product Moment Correlation Coefficient is the metric used to assess this.

    Table 1: Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (Implicit)Reported Device Performance (Correlation Coefficient)
    Normal Donors (N=202)Pearson Correlation significantly high (e.g., SL
    Ask a Question

    Ask a specific question about this device

    K Number
    K951100
    Date Cleared
    1996-05-13

    (430 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K950482
    Date Cleared
    1996-05-10

    (462 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-mune OK-COMBO CD3-FITC/CD8-PE is intended for use in identification and enumeration of CD3+ and CD8+ human T lymphocytes in whole blood by flow cvtometry. The intended use is the same as the intended use of the predicate device(s), Ortho-mune OKT3 Monoclonal Antibody (Murine) FITC Conjugate, and Ortho-mune OKT8 Monoclonal Antibody (Murine) Phycoerythrin conjugate.

    Device Description

    Ortho-mune OK-COMBO CD3-FITC/CD8-PE (OKT3/OKT8) Monoclonal Antibody (Murine) is a blend of the individual purified monoclonal antibodies OKT3 and OKT8 conjugated to the fluorochromes fluorescein isothiccyanate and phycoerythrin respectively.

    AI/ML Overview

    This document describes the performance characteristics of the Ortho-mune™ OK-COMBO CD3-FITC/CD8-PE (OKT3/OKT8) Monoclonal Antibody (Murine) and its equivalence to predicate single-color reagents. Note: This document is from 1996 and uses terminology and study designs common for that era. There are no explicitly stated "acceptance criteria" in the format of pass/fail thresholds in this summary. Instead, the study aims to demonstrate "equivalence" through statistical comparisons and reproducibility. The document also does not describe an AI/algorithm-based device but rather a new diagnostic reagent. Therefore, sections related to AI-specific criteria (e.g., human-in-the-loop, AI vs. without AI assistance, training data) are not applicable.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit, quantitative acceptance criteria with pass/fail thresholds are not stated. Instead, the studies aimed to demonstrate equivalence to predicate devices and acceptable reproducibility and linearity. The reported device performance is presented as statistical comparisons and descriptive statistics.

    Study TypePerformance MetricStated Goal (Implicit Acceptance Criteria)Reported Device Performance and Conclusion
    Equivalence to Predicate Dual-color vs. Single-color ReagentsMean % & Range % for CD3+ and CD8+ cellsEquivalence in percentage detection to predicate single-color reagents.Normal Donors (N=199):
    • CD3+: Dual Color Mean % = 74.5 (Range 45.9-87.4) vs. Single Color Mean % = 75.9 (Range 58.4-90.1)
    • CD8+: Dual Color Mean % = 31.2 (Range 15.0-74.9) vs. Single Color Mean % = 29.2 (Range 13.2-75.5)

    AIDS/ARC Patients (N=77):

    • CD3+: Dual Color Mean % = 73.5 (Range 27.5-90.7) vs. Single Color Mean % = 77.7 (Range 35.6-96.4)
    • CD8+: Dual Color Mean % = 59.1 (Range 25.6-89.8) vs. Single Color Mean % = 60.3 (Range 18.1-92.1)

    Combined Population (Normal & AIDS/ARC):

    • CD3+: Linear regression: Y = -0.359 + 0.976(X), R = 0.928
    • CD8+: Linear regression: Y = 5.547 + 0.882(X), R = 0.969
      Conclusion: Performance is "equivalent" to single-color reagents. |
      | Equivalence to Other Flow Cytometers | Linear regression (Correlation R-value) | Equivalence of ORTHO CYTORONABSOLUTE with recognized flow cytometers (FACScan and EPICS). | CYTORONABSOLUTE vs. FACScan (CD3+CD8+): Y = -0.3519 + 1.03999(x), R = 0.9955
      CYTORONABSOLUTE vs. COULTER EPICS (CD3+CD8+): Y = 2.707 + 0.981(X), R = 0.983
      Conclusion: Demonstrates "equivalent performance" of the ORTHO CYTORONABSOLUTE flow cytometer to Becton Dickinson FACScan and COULTER EPICS. |
      | Within-Laboratory Reproducibility | Coefficient of Variation (CV) | Acceptable within-laboratory reproducibility for various cell percentages (low, normal, high). | TOTAL CD3+ Low/Normal/High: CVs from 1.539% to 3.488% across 3 sites.
      CD3+CD8+ Low/Normal/High: CVs from 2.904% to 11.214% across 3 sites.
      Conclusion: Acceptable within-laboratory reproducibility. |
      | Between-Laboratory Reproducibility | Coefficient of Variation (CV) | Acceptable between-laboratory reproducibility for various cell percentages. | TOTAL CD3+ Low/Normal/High: Across Site CVs from 1.139% to 2.899%.
      CD3+CD8+ Low/Normal/High: Across Site CVs from 1.837% to 12.309%.
      Conclusion: Acceptable between-laboratory reproducibility. |
      | Linearity Study | Slope and R-value of linear regression (Observed vs. Expected) | Linear performance across a wide range of lymphocyte counts. | TOTAL CD3+ (All donors): Slope = 0.999 (CI 0.001), R = 1.000
      CD3+CD8+ (All donors): Slope = 1.000 (CI 0.002), R = 1.000
      Conclusion: Demonstrated linear performance for both subsets across 20 to 18,676 cells/ul with slopes "indistinguishable from 1" and R values of 1.000. (Note: An R-value of exactly 1.000 is highly unusual in biological systems and may suggest data smoothing or specific analytical methods). |

    2. Sample sizes used for the test set and the data provenance

    Equivalence to Predicate Reagents Study:

    • Sample Size: 199 normal donors and 77 AIDS/ARC patients. Total N=276.
    • Data Provenance: Not explicitly stated, but likely from a US-based clinical setting as it's a 510(k) submission to the FDA. The data is prospective, collected for the purpose of this study.

    Equivalence to Other Flow Cytometers Study:

    • Sample Size: Ten normal donors (whole blood, EDTA).
    • Data Provenance: Not explicitly stated, but likely from a US-based clinical setting, prospective.

    Reproducibility Studies:

    • Sample Size: Eleven normal donors (whole blood, EDTA).
    • Data Provenance: Not explicitly stated, but likely from a US-based clinical setting, prospective. Conducted at three independent laboratories.

    Linearity Study:

    • Sample Size: Four normal donors (whole blood, EDTA).
    • Data Provenance: Not explicitly stated, but likely from a US-based clinical setting, prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This device is a reagent for flow cytometry, which directly identifies and enumerates cell populations. The "ground truth" for the test set is established by the measurements from the predicate devices (single-color reagents) which are considered the established method. There are no human experts "establishing ground truth" in the sense of image interpretation or diagnostic classification. The measurements are quantitative outputs from flow cytometers.

    4. Adjudication method for the test set

    Not applicable. This device provides quantitative measurements from a flow cytometer. The "ground truth" is derived from a comparative measurement against established methods (predicate reagents) or from internal consistency (reproducibility, linearity). There is no subjective human assessment requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/algorithm-based diagnostic device where human readers interact with AI. It is a laboratory reagent.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. Its performance is evaluated in a standalone manner against predicate reagents and for its intrinsic measurement characteristics (reproducibility, linearity) when used with a flow cytometer.

    7. The type of ground truth used

    The ground truth used for assessing the new dual-color reagent (Ortho-mune OK-COMBO CD3-FITC/CD8-PE) was primarily:

    • Comparative Measurements: Measurements obtained from the predicate single-color reagents (Ortho-mune OKT3 Monoclonal Antibody FITC Conjugate and Ortho-mune OKT8 Monoclonal Antibody Phycoerythrin conjugate) for CD3+ and CD8+ cell percentages.
    • Internal Consistency: For reproducibility and linearity studies, the ground truth is the expected statistical behavior (e.g., low coefficient of variation for reproducibility, R-value near 1 and slope near 1 for linearity with serially diluted samples or known concentrations).
    • Comparison to other established flow cytometers: For the instrument comparison part of the study, the measurements from the Becton Dickinson FACScan and COULTER EPICS flow cytometers were used as a comparative benchmark for the ORTHO CYTORONABSOLUTE.

    There is no pathology, expert consensus on subjective findings, or outcomes data used as ground truth in this submission for this specific device.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning algorithm. There is no concept of a "training set" for this type of diagnostic reagent.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI/machine learning algorithm, there is no training set or ground truth established for it in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K935720
    Date Cleared
    1996-05-09

    (891 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ortho-Count Calibration Kit for ORTHO CYTORONABSOLUTE Laser Flow Cytometry System is intended to be used to calibrate and verify calibration of the absolute lymphocyte counting function of the ORTHO CYTORONABSOLUTE.

    Device Description

    Ortho-Count Calibration Kit for ORTHO CYTORONABSOLUTE Laser Flow Cytometry System, contains a Calibrator Bead Suspension for absolute lymphocyte count calibration and Verifier Bead Suspensions (I, II and III) to verify accuracy of absolute count calibration. The Calibrator Bead Suspension is manufactured as known "events"/unit volume (i.e. the Calibration Number) in the form of microparticle beads suspended in a proprietary buffer. The size of the microparticles has been selected to emulate the range of size of leukocytes detected by the ORTHO CYTORONABSOLUTE under specified gating conditions. The Verifier Bead Suspensions (I, II and III), are composed of materials identical to the Calibrator Bead Suspension and collectively provide a range of particle bead concentrations. The target values chosen approximate values expected from patients with low (Verifier I), normal (Verifier II) and high (Verifier III) white blood cell counts, The Verifiers are used to verify accuracy of the absolute respectively. lymphocyte count calibration immediately following calibration and are also used as daily quality control verification that the ORTHO CYTORONABSOLUTE is in calibration. A properly calibrated instrument will provide absolute counts within + 10% of the labeled events/uL for each Verifier supsension. In the context of immune status monitoring, absolute cell count is defined as the number of cells per unit volume. The ORTHO CYTORONABSOLUTE uses precision stepper motors to deliver diluted sample to the instrument's flow cell at a constant rate. Since constant flow rate means movement of a fixed volume per unit time, analysis time determines the volume of sample analyzed. Calibration of the ORTHO CYTORONABSOLUTE consists of setting analysis time such that a volume corresponding to 1 uL of undiluted sample is examined. Ortho-Count Calibrator Bead Suspension, representing a known number of particles per unit volume (i.e. events/uL) at the same dilution as used for testing samples, is counted until its "Calibration Number" is reached; the instrument is set to this analysis time. In subsequent analyses events are to passage of the equivalent of 1 uL of undiluted sample the instrument flowcell. As a result of this calibration, sample results can be read in absolute count, i.e. "events"/uL.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    The core acceptance criterion for the Ortho-Count Calibration Kit is that a properly calibrated ORTHO CYTORONABSOLUTE instrument will provide absolute counts within ±10% of the labeled events/uL for each Verifier suspension.

    Acceptance CriteriaReported Device Performance
    Absolute counts within ±10% of labeled events/uL for each Verifier suspension.Verifier I: Mean across sites = 1693, Std. Dev. = 55, % CV = 3.2% (Compared to a target value, which is not explicitly stated but implied to be within 10% of the mean)
    Verifier II: Mean across sites = 7748, Std. Dev. = 193, % CV = 2.5%
    Verifier III: Mean across sites = 15140, Std. Dev. = 409, % CV = 2.7%
    No significant differences in absolute lymphocyte counts attributable to instrument or calibration method when compared to predicate device/method.Table 1: Ratios of Mean Method 1/JT2, Method 2/JT2, and Method 2/Method 1 generally hover around 1.0, with mean standard deviations around 0.04-0.05, indicating no significant differences. For example, the mean ratio of Meth2/Meth1 is 0.99 with a standard deviation of 0.05.
    Inter-laboratory reproducibility.Table 2: Overall Inter-Lab % CV values for Verifiers I, II, and III are 3.2%, 2.5%, and 2.7% respectively, which are comparable to within-laboratory precision, indicating good inter-laboratory reproducibility.
    Strong correlation between direct ORTHO CYTORONABSOLUTE counts and those calculated from combined hematology and other flow cytometer/reagent results.Figure 2: Regression analysis yields a slope of 0.89, a y-intercept of -31 ± 13, and a correlation coefficient (r) = 0.96. This indicates a strong positive correlation.
    Utility as quality control indicators over time (60 days).Figure 3: Shows "tight agreement" in measurements made using each Verifier Bead Suspension over 60 days, confirming suitability as an accurate quality control measure.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Table 1 (Comparison to Predicate Device): 10 donors. Data provenance not explicitly stated (e.g., country of origin), and it's not clear if it was retrospective or prospective.
      • Table 2 (Inter-laboratory Reproducibility): Verifier Bead Suspensions (I, II, and III) were used across four different ORTHO CYTORONABSOLUTE instruments located in different regions of the United States. The frequency or total number of measurements for establishing "Mean" values for each verifier is not specified, but it's likely prospective for this component.
      • Figure 1 (Consistency of Calibration/Cross-Calibration): 41 samples (whole blood panel). Provenance not explicitly stated beyond "different regions of the United States" for the instruments. The samples were handled "overnight at room temperature or shipped for next day delivery to each site," suggesting prospective collection for this study, though the source of the donors/patients is not specified.
      • Figure 2 (Direct vs. Calculated Absolute Counts): Data points for CD19, CD3+CD8+, CD3+CD4+, and Total CD3+. The total number of individual samples is not specified, but the regression analysis implies a substantial number of data points for these subsets. Provenance not explicitly stated.
      • Figure 3 (Quality Control Indicators): Counts/volume were measured on the three verifier suspensions twice per week over a period of 60 days for four ORTHO CYTORONABSOLUTE instruments.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document does not explicitly state the number or qualifications of experts used to establish ground truth for the test sets. Instead, the ground truth is established through:
        • Comparison to a predicate device (Coulter JT2 calibrated with S-CAL Kit) (for Table 1).
        • Labeled 'events'/uL for the Verifier Bead Suspensions themselves, which are manufactured values (for Table 2 and Figure 3).
        • "Calculated from combined hematology and other flow cytometer/reagent results" (for Figure 2's comparative ground truth).
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • No adjudication method involving experts is described. The comparisons are quantitative measurements against established methods or known values.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving human readers (as in interpretation of images or complex data) was conducted. This device is a calibration kit for an automated instrument, not an AI diagnostic tool requiring human interpretation. Therefore, the concept of "human readers improve with AI" is not applicable here.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The study primarily focuses on the standalone performance of the calibration kit in ensuring the accuracy and reproducibility of the ORTHO CYTORONABSOLUTE instrument's absolute counting function. The ORTHO CYTORONABSOLUTE itself is an automated device, so its performance (when calibrated) is inherently "algorithm only" in terms of cell counting. The "human-in-the-loop" would be the technician operating the flow cytometer and applying the calibration kit.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Known Reference Values: For the Verifier Bead Suspensions, the ground truth is the "labeled events/uL," which are manufactured and accurately determined "events"/unit volume.
      • Predicate Device/Method: For the comparison of absolute lymphocyte counts (Table 1), the ground truth is established by a "Coulter JT2 calibrated using S-CAL Kit" which is the predicate device/method.
      • Combined Method: For Figure 2, the ground truth is established by "values calculated from separate flow cytometers and automated hematology analyzers."
    7. The sample size for the training set:

      • The document does not describe a "training set" in the context of machine learning or AI. This is a calibration kit, and its performance is evaluated against established methods and internal consistency, not through an iterative training process for an algorithm.
    8. How the ground truth for the training set was established:

      • As there is no described training set for an algorithm, this question is not applicable. The kit itself is manufactured with known characteristics (e.g., "known 'events'/unit volume" for the Calibrator Bead Suspension).
    Ask a Question

    Ask a specific question about this device

    K Number
    K954570
    Date Cleared
    1996-04-30

    (211 days)

    Product Code
    Regulation Number
    864.5220
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORTHO DIAGNOSTIC SYSTEMS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ORTHO™ ImmunoCount Flow Cytometry System is intended to be used for lymphocyte immunophenotyping. Results may be reported as either percent positive cells, or as absolute counts of lymphocytes and lymphocyte subsets.

    Device Description

    ORTHO ImmunoCount Flow Cytometry System consists of ORTHO CytoronAbsolute Laser Flow Cytometer, ImmunoCount II Software, ORTHO TRIO Monoclonal Antibodies for lymphocyte immunophenotyping, and Ortho-Count Calibration Kit for calibration and verification of system calibration.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and study details for the ORTHO ImmunoCount Flow Cytometry System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to a "predicate method" and the demonstration of linearity and reproducibility. The tables provided (TABLE A, B, C, D, E, F, G, H) present the performance data, which implicitly serves as the "reported device performance." Since explicit numerical acceptance criteria (e.g., "R-value > 0.95") are not stated, I will infer them from the data presented, particularly the R-values, slopes, and intercepts for comparability, and CVs for reproducibility. For comparability, "equivalent" means the values are close to the predicate. For reproducibility, "low CV" is the criterion. For linearity, an R-value close to 1.00 is the criterion.

    Acceptance Criteria CategorySpecific MetricImplied Acceptance Criterion (from Predicate/Reproducibility)Reported Device Performance (ORTHO ImmunoCount)Met/Not Met (Based on typical expectations for medical devices)
    Comparability (Predicate Method)
    Percent Positive Cells (Normal Donors)Mean % difference vs. Predicate (for various subsets like CD3+CD4+, CD16+CD3-, CD19+, Mean CD3+)Mean % difference should be minimal, suggesting equivalence to the predicate. Range overlap.See TABLE A. Differences in mean % are generally small (e.g., CD3+CD4+ ImmunoCount 45.5% vs. Predicate 45.2%; CD16+CD3- ImmunoCount 11.6% vs. Predicate 13.1%). Ranges largely overlap.Met
    Absolute Counts (Normal Donors)Mean (cells/µL) difference vs. Predicate (for various subsets)Mean (cells/µL) difference should be minimal, suggesting equivalence to the predicate. Range overlap.See TABLE A. Differences in mean (cells/µL) are generally small (e.g., CD3+CD4+ ImmunoCount 855 vs. Predicate 900; CD16+CD3- ImmunoCount 221 vs. Predicate 266). Ranges largely overlap.Met
    Percent Positive Cells (HIV+ Donors)Mean % difference vs. Predicate (for various subsets)Mean % difference should be minimal, suggesting equivalence to the predicate. Range overlap.See TABLE B. Differences in mean % are generally small (e.g., CD3+CD4+ ImmunoCount 17.7% vs. Predicate 17.6%; CD16+CD3- ImmunoCount 8.8% vs. Predicate 10.6%). Ranges largely overlap.Met
    Absolute Counts (HIV+ Donors)Mean (cells/µL) difference vs. Predicate (for various subsets)Mean (cells/µL) difference should be minimal, suggesting equivalence to the predicate. Range overlap.See TABLE B. Differences in mean (cells/µL) are generally small (e.g., CD3+CD4+ ImmunoCount 260 vs. Predicate 313; CD16+CD3- ImmunoCount 93 vs. Predicate 144). Ranges largely overlap.Met
    Linear Regression (Normal Donors - Percent)R-value between ImmunoCount and Predicate (for % positive cells)R-value > 0.85 (typical for good correlation), Slope close to 1, Intercept close to 0.See TABLE C. R-values range from 0.83 (CD3+ (4/8/3) vs CD3+ (3/4)) to 1.00 (CD4+ (4/8/3) vs CD4+ (3/4)). Slopes are generally close to 1 (0.87-1.01), intercepts close to 0 (0-12).Met
    Linear Regression (Normal + HIV+ Abs Counts)R-value between ImmunoCount and Predicate (for absolute counts)R-value > 0.85, Slope close to 1, Intercept close to 0.See TABLE D. R-values range from 0.88 (CD3- (16/19/3) vs CD3+ (3/16+56)) to 0.96 (All lymphocyte subsets). Slopes are generally close to 1 (0.76-0.89), intercepts are slightly higher than 0 (3-124).Met
    Reproducibility
    Within-Laboratory ReproducibilityCoefficient of Variation (CV) for absolute lymphocyte counts (Low, Normal, High concentrations)CV
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2