Search Filters

Search Results

Found 12 results

510(k) Data Aggregation

    K Number
    K212579
    Manufacturer
    Date Cleared
    2023-05-02

    (624 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    Organogenesis Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FortiShield is intended for use as a temporary wound covering, and to provide a moist wound healing environment on cleanly debrided wounds after hemostasis has been established.

    FortiShield is indicated for: partial thickness wounds, pressure sores, venous ulcers, diabetic ulcers, chronic vascular ulcers, surgical wounds (donor sites/grafts, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, partial thickness burns, and skin tears) and draining wounds.

    Device Description

    FortiShield is a sterile, translucent biosynthetic wound matrix made from a knitted tri-filament nylon fabric that is mechanically bonded to a semi-permeable silicone membrane. Denatured porcine dermal collagen is bonded to the nylon/silicone bi-layer membrane to provide a dressing that is designed to adhere to the application site, provide a barrier to the external environment, and allow for excess exudate drainage.

    FortiShield is not made with natural rubber latex. FortiShield is non-pyrogenic.

    AI/ML Overview

    This is a 510(k) premarket notification for the FortiShield device, a biosynthetic wound matrix. The document asserts the device's substantial equivalence to predicate devices based on technological characteristics and performance data.

    Here's an analysis of the provided information regarding acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly define formal "acceptance criteria" in a quantitative manner as one might expect for a diagnostic AI device (e.g., minimum sensitivity/specificity). Instead, the assessment focuses on demonstrating equivalence to predicate devices. The "acceptance criteria" are implicitly met by showing comparable performance against the predicate devices across several characteristics.

    CharacteristicAcceptance Criteria (Implicit: Equivalent to Predicate)Reported Device Performance
    Formulation/MaterialsPorous Silicone and Nylon Knit Bi-layer Scaffold with Collagen CoatingPorous Silicone and Nylon Knit Bi-layer Scaffold with Cross-Linked Collagen Coating
    Principles of OperationDevice covers the wound to provide a moist wound environment and allows the passage of exudate.Device covers the wound to provide a moist wound environment and allows the passage of exudate.
    Indications for UseEquivalent to PermeaDerm CW predicate for various wound types.FortiShield is indicated for: partial thickness wounds, pressure sores, venous ulcers, diabetic ulcers, chronic vascular ulcers, surgical wounds, trauma wounds, and draining wounds.
    Device DimensionsRange of sizes within the maximum offered for predicate devices.2" x 2" through 15" x 30"
    SterilitySterile, SAL 10-6Sterile, SAL 10-6
    PyrogenicityNon-pyrogenicNon-pyrogenic
    BiocompatibilityBiocompatible based on ISO 10993 testing for intended nature and duration of contact.Biocompatible based on ISO 10993-1 and FDA General Guidance evaluation, including various biological endpoints.
    Tensile ElongationEquivalent to AWBAT predicate device.Demonstrated as equivalent.
    Fatigue PropertiesEquivalent to AWBAT predicate device.Demonstrated as equivalent.
    Water Vapor Transmission RateEquivalent to AWBAT predicate device.Demonstrated as equivalent.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not describe a "test set" in the context of a typical AI/software performance study with a specific number of cases. The "study" here is a series of non-clinical mechanical and biocompatibility tests on the device itself and direct comparison to predicate devices, not an evaluation of diagnostic performance on patient data.

    • Sample Size for Mechanical Testing: Not explicitly stated, but these would be a number of FortiShield device samples and predicate device samples.
    • Sample Size for Biocompatibility Testing: Not explicitly stated, but this refers to samples of the FortiShield device tested according to ISO 10993 protocols.
    • Data Provenance: Not applicable in the sense of patient data. The data originates from laboratory testing of the physical devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. "Ground truth" in the context of this device is established through:
    * Material characterization and manufacturing specifications.
    * Standard ISO and FDA-recognized mechanical testing protocols.
    * Standard ISO and FDA-recognized biocompatibility testing protocols.
    * Comparison to the known characteristics of legally marketed predicate devices.

    4. Adjudication Method for the Test Set:

    Not applicable. There is no human adjudication process described here for establishing ground truth, as it's a conformity assessment against physical and material properties and a comparison to known predicate devices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This device is a wound dressing, not a diagnostic imaging AI algorithm. There is no human interpretation component where AI assistance would be evaluated.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    Not applicable. This is not an AI algorithm. The performance evaluation is for the physical device itself.

    7. Type of Ground Truth Used:

    The "ground truth" for this device's performance is established based on:
    * Material and Device Specifications: Characteristics like composition, structure, dimensions, and manufacturing standards.
    * Validated Laboratory Testing: Results from mechanical tests (tensile elongation, fatigue, water vapor transmission rate) and biocompatibility tests (ISO 10993 endpoints like irritation, toxicity, cytotoxicity, etc.).
    * Predicate Device Characteristics: The known, legally marketed properties and performance of the substantially equivalent predicate devices.

    8. Sample Size for the Training Set:

    Not applicable. This is not an AI device or a machine learning model, so there is no training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220317
    Manufacturer
    Date Cleared
    2022-07-01

    (148 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Organogenesis Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PuraPly® Micronized Wound Matrix (PuraPly® MZ) is intended for the management of wounds that include:

    • · Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • · Diabetic ulcers
    • Chronic vascular ulcers
    • · Tunneled/undermined wounds
    • Surgical wounds (e.g., donor sites/grafts, post-Mohs' surgery, Podiatric wound, and wound dehiscence).
    • · Trauma wounds (e.g., abrasions, lacerations, and skin tears)
    • Partial thickness burns
    • Draining wounds
      The device is intended for single patient use only.
    Device Description

    PuraPly® Micronized Wound Matrix (PuraPly® MZ) consists of micronized porcine collagen intended for the management of wounds. PuraPly® Micronized Wound Matrix is supplied as a dry powder of particle size of ≤ 1000μm. The device is sterile and packaged in a vial sealed in a single pouch.

    AI/ML Overview

    I am sorry, but the provided text does not contain information about acceptance criteria and a study proving a device meets them. The document is an FDA 510(k) clearance letter for a wound matrix device (PuraPly® Micronized Wound Matrix) and focuses on demonstrating substantial equivalence to predicate devices, not on specific performance acceptance criteria or a study designed to meet them.

    The "Performance Data Non-Clinical Test" section mentions tests performed on the device (e.g., color, particle size, absorption capacity, endotoxin levels, pH) and biocompatibility tests (e.g., cytotoxicity, sensitization, irritation, pyrogenicity, various systemic toxicities, genotoxicity, and a large animal wound healing study). However, it does not provide:

    • A table of acceptance criteria with corresponding reported device performance values.
    • Details on the study design, sample size, or data provenance for these tests.
    • Information about experts, ground truth, adjudication methods, or MRMC studies.
    • Specific details about the training set for any AI/ML model, as this device does not appear to be an AI-powered device.

    Therefore, I cannot fulfill your request based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071555
    Date Cleared
    2007-08-30

    (84 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FortaGen Oral Membrane is intended to be used during guided bone regeneration (GBR) and guided tissue regeneration (GTR) procedures as a biodegradable barrier membrane for:

    • Ridge augmentation for later implant insertions;
    • . Simultaneous ridge augmentation and implant insertions;
    • . Ridge augmentation around implants inserted in delayed extraction sites;
    • . Ridge augmentation around implants inserted in immediate extraction sites;
    • . Alveolar ridge preservation consequent to tooth (teeth) extraction(s);
    • Over the window in lateral window sinus elevation procedures; .
    • . In implants with vertical bone loss due to infection, only in cases where satisfactory debridement and implant surface disinfection can be achieved;
    • . In intra bony defects around teeth;
    • . For treatment of recession defects, together with coronally positioned flap;
    • . In furcation defects in multi-rooted teeth.
    Device Description

    FortaGen Oral Membrane consists of a multi-laminate sheet predominantly of Type I porcine collagen. The device is supplied hydrated in sheet form in sizes ranging from 13 mm x 25 mm to 40 mm x 50 mm in sterile double layer peelable packaging.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the FortaGen Oral Membrane. This document focuses on demonstrating substantial equivalence to pre-existing devices rather than outlining specific acceptance criteria and a detailed study proving performance against those criteria.

    Therefore, many of the requested sections (Table of acceptance criteria, sample size for test set, number of experts for ground truth, adjudication method, MRMC study, sample size for training set, ground truth for training set) cannot be fully addressed from the provided text.

    Here's an analysis based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state a table of acceptance criteria with numerical targets. Instead, it relies on demonstrating "substantial equivalence" to predicate devices based on technological characteristics and performance data. The reported "performance" is qualitative, asserting similarity to existing devices.

    Acceptance Criteria CategoryReported Device Performance (FortaGen Oral Membrane)
    Biocompatibility"FortaGen Oral Membrane has been shown to be biocompatible."
    Resorption Profile"designed to resorb within six months. Its handling characteristics and resorption profile are substantially equivalent to those of its predicate devices."
    Handling Characteristics"Its handling characteristics and resorption profile are substantially equivalent to those of its predicate devices."
    Equivalency to Predicates"should function equivalently to Ossix-Plus and Bio-Gide when used as intended for GTR and GBR procedures."
    Material/Composition"consists of a multi-laminate sheet predominantly of Type I porcine collagen." (Matches predicate's collagen-based nature).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a specific "test set" in the context of a clinical study with a defined sample size. The substantial equivalence argument appears to be based on pre-clinical data (biocompatibility) and material characteristics, implicitly compared against established knowledge of the predicate devices. No information on data provenance (country of origin, retrospective/prospective) is provided in relation to a formal test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The document does not describe a clinical study with a "test set" requiring expert-established ground truth. The demonstration of equivalence primarily relies on material science and manufacturing process attributes, and general biological function.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no described test set requiring adjudication is present.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a resorbable dental barrier membrane, not an AI software or diagnostic imaging device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the FortaGen Oral Membrane appears to be based on:

    • Material properties and composition: Verified through standard material characterization tests to confirm it's primarily Type I porcine collagen.
    • Biocompatibility: Demonstrated through standard biocompatibility testing methods (likely in-vitro and/or in-vivo animal studies, although specific details are not provided in this summary).
    • Resorption profile: Assessed through in-vitro or in-vivo studies (likely animal models) to determine the timeframe for resorption.
    • Handling characteristics: Likely evaluated during manufacturing and potentially in pre-clinical settings.
      The "truth" is that these characteristics are "substantially equivalent" to predicate devices, which are already accepted as safe and effective.

    8. The sample size for the training set

    Not applicable. This refers to training data for algorithms, which is not relevant for this physical device.

    9. How the ground truth for the training set was established

    Not applicable. This refers to ground truth for training data for algorithms, which is not relevant for this physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051647
    Date Cleared
    2005-11-08

    (140 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FortaDerm™ Antimicrobial PHMB Wound Dressing is intended for the management of wounds and as an effective barrier to resist microbial colonization within the dressing and reduce microbes penetrating through the dressing. FortaDerm™ Antimicrobial PHMB Wound Dressing may be used for the management of: partial and full thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, wound dehiscence), trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) and draining wounds.

    The device is intended for one-time use.

    Device Description

    FortaDerm™ Antimicrobial PHMB Wound Dressing consists of two layers, crosslinked sheet of fenestrated sheet of porcine intestinal collagen coated with 0.1 % Polyhexamethylene Biguanide hydrochloride (PHMB). FortaDerm Wound Dressing is supplied dry in sheet form in sizes ranging from 4 x 4 cm to 12 x 36 cm. The device is packaged in sterile, sealed single pouches.

    AI/ML Overview

    This 510(k) summary (K051647) describes the FortaDerm™ Antimicrobial PHMB Wound Dressing. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than independent performance claims against specific acceptance criteria. Therefore, much of the requested information regarding a formal study with detailed acceptance criteria, sample sizes, expert involvement, and statistical analysis is not provided in this document.

    However, based on the available text, here's what can be extracted:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityPassed the requirements
    PerformancePassed the requirements
    Specific quantitative criteria (e.g., microbial reduction percentage, barrier effectiveness)Not specified in this document.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified.
      • Data Provenance: Not specified, but likely from in-vitro and/or in-vivo laboratory tests given the nature of the device and testing categories (biocompatibility, performance). It is not stated if this was retrospective or prospective data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Number of Experts: Not applicable/not specified. The testing described (biocompatibility, performance) implies laboratory assessments against predetermined standards or methods rather than expert consensus on a 'ground truth' in a clinical diagnostic sense.
      • Qualifications of Experts: Not applicable/not specified.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Adjudication Method: Not applicable. This type of adjudication is typically used in clinical studies involving interpretation of data by multiple readers, which is not described here.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • MRMC Study: No. This device is a wound dressing, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Standalone Performance: Not applicable. This device is a physical wound dressing, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Type of Ground Truth: For "biocompatibility" and "performance" tests, the "ground truth" would be the predefined pass/fail criteria or standards for those specific tests (e.g., ISO standards for biocompatibility, or specific laboratory protocols for antimicrobial efficacy or barrier properties). The document does not detail these specific standards.
    7. The sample size for the training set:

      • Sample Size: Not applicable. This is not a machine learning device that requires a training set. The "testing" likely refers to validation against established physical and biological performance characteristics.
    8. How the ground truth for the training set was established:

      • Ground Truth Establishment: Not applicable, as there is no training set for this device.

    Summary based on the provided text:

    The 510(k) submission for the FortaDerm™ Antimicrobial PHMB Wound Dressing primarily relies on demonstrating substantial equivalence to existing predicate devices (FortaDerm™ Wound Dressing K011026 and Xcell Antimicrobial Cellulose Wound Dressing K024054) in terms of intended use, technological characteristics, materials, and physical construction.

    The only "performance data" mentioned is that the device underwent "a number of tests to assess biocompatibility and performance" and "passed the requirements of all tests." However, specific acceptance criteria for these tests, the methodology, sample sizes, or detailed results are not provided in this summary. This is typical for a 510(k) where detailed study reports are usually referenced but not fully included in the public-facing summary. The evaluation focuses on meeting general safety and performance standards for wound dressings, rather than specific quantitative metrics against a defined "ground truth" involving expert interpretation (as would be the case for a diagnostic AI).

    Ask a Question

    Ask a specific question about this device

    K Number
    K042809
    Device Name
    CUFFPATCH
    Date Cleared
    2004-11-02

    (21 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CuffPatch™ surgical mesh is intended for the reinforcement of soft tissues repaired by sutures or suture anchors, during tendon repair surgery including reinforcement of rotator cuff, patellar, Achilles, biceps, quadriceps or other tendons.

    CuffPatch™ surgical mesh is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps quadriceps or other tendons. Sutures, used to repair the tear, and sutures or bone anchors, used to attach the tissue to the bone, provide biomechanical strength for the tendon repair. CuffPatch™ surgical mesh reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue.

    Device Description

    CuffPatch™ consists of laminated sheets of porcine intestinal collagen matrix is primarily Type I porcine collagen, and is free of cells and cell remnants. The product is supplied sterile in doublelayered peelable packaging.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called CuffPatch™ Surgical Mesh. It does not contain information about acceptance criteria for a study or a study that proves the device meets such criteria in the way typically expected for an AI/ML device.

    Instead, the document details the device's intended use, classification, description, and claims of substantial equivalence to legally marketed predicate devices. The "Performance Data" section briefly states that "Evaluation of the device's materials, physical and performance characteristics revealed that CuffPatch™ is substantially equivalent to the predicate devices, and is suitable for its intended clinical applications." This indicates an assessment was made, but the specific acceptance criteria and the detailed study results are not provided.

    Therefore, most of the requested information cannot be extracted from this document, as it outlines a regulatory submission based on substantial equivalence, not a detailed performance study with explicit acceptance criteria.

    However, I can extract what is implied about the performance assessment rather than explicit acceptance criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Primary: Substantial equivalence to predicate devices in:CuffPatch™ is substantially equivalent to:
    • FortaFlex™ Surgical Mesh (CuffPatch™), K020049 and K011025
    • Restore® Orthobiologic Soft Tissue Implant, K031969 |
      | - Materials | Evaluation of the device's materials revealed substantial equivalence. CuffPatch™ consists of laminated sheets of porcine intestinal collagen matrix (primarily Type I porcine collagen, free of cells and cell remnants). This is similar to predicate devices which are also surgical meshes. |
      | - Design | Evaluation of the device's design revealed substantial equivalence. |
      | - Physical characteristics | Evaluation of the device's physical characteristics revealed substantial equivalence. |
      | - Performance characteristics | Evaluation of the device's performance characteristics revealed substantial equivalence. The device's intended use is for reinforcement of soft tissues during tendon repair and provides a resorbable scaffold. This is similar to predicate devices. The document explicitly states it "is suitable for its intended clinical applications." |
      | - Biological attributes | Evaluation of the device's biological attributes revealed substantial equivalence. The device is a resorbable scaffold replaced by the patient's own soft tissue. |
      | Secondary (Suitability for Intended Use): Function as intended. | The device's "Evaluation... revealed that CuffPatch™ is substantially equivalent to the predicate devices, and is suitable for its intended clinical applications."
      Intended Use: Reinforcement of soft tissues repaired by sutures or suture anchors, during various tendon repair surgeries. It reinforces soft tissue and provides a resorbable scaffold. It is not intended to replace normal body structure or provide full mechanical strength; sutures provide biomechanical strength. |

    Given the nature of the document as a 510(k) summary for a conventional medical device (surgical mesh), it does not address AI/ML-specific testing methodologies. Therefore, the following points cannot be answered:

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Not applicable/Not provided. The document describes a "Performance Data" section but does not detail a study involving test sets in the context of data-driven performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. No ground truth establishment by experts for a test set is mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not provided. This is a conventional medical device, not an AI/ML diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not provided. This is a conventional medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable/Not provided. The performance assessment was based on comparing material, physical, performance, and biological characteristics to existing predicate devices, rather than a clinical ground truth in the context of an AI/ML model.

    8. The sample size for the training set

    • Not applicable/Not provided. No training set is relevant for this conventional device.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. No training set is relevant for this conventional device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K021107
    Date Cleared
    2002-05-15

    (40 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FortaPerm™ is intended for implantation to reinforce and support soft tissue where weakness exists including, but not limited to the following procedures: pubourethral support, prolapse repair (urethral, vaginal, rectal and colon), reconstruction of the pelvic floor, bladder support, sacrocolposuspension, reconstructive procedures and tissue repair. By providing pubourethral support, FortaPerm may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency. The device is intended for one-time use.

    Device Description

    Forta Perm consists of a multi-laminate sheet predominantly of Type I porcine collagen. The device is supplied in sheet form in sizes ranging from 2 x 5 cm to 12 x 36 cm in sterile double laye packaging.

    AI/ML Overview

    The submitter, Organogenesis Inc., provided limited information regarding the performance data for FortaPerm™ Surgical Mesh. The application states, "FortaPerm was subjected to a panel of tests to assess biocompatibility, integrity, and performance. The device passed the requirements of all tests."

    However, specific acceptance criteria and detailed quantitative results of these tests are not explicitly stated in the provided documents. The information is very general and does not allow for a comprehensive breakdown as requested.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    (Not Specified)Passed all requirements

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified.
    • Data Provenance: Not specified (e.g. country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable as this is a medical device (surgical mesh), not an AI/imaging device requiring expert ground truth for interpretation of results. The "ground truth" would likely be laboratory or animal model assessments of material performance and biocompatibility.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable for this type of device performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for this device would be established by standard scientific and engineering methodologies for evaluating biocompatibility, integrity, and performance of surgical mesh biomaterials (e.g., in vitro mechanical testing, biocompatibility assays, animal implantation studies). Specific details are not provided in the document.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device requiring a training set.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI/machine learning device.

    Summary of Study (Based on Provided Information):

    The submission states that FortaPerm™ was subjected to a "panel of tests to assess biocompatibility, integrity, and performance" and "The device passed the requirements of all tests." No further details on the specific tests, their methodologies, sample sizes, or quantitative results are provided in the excerpt. The substantial equivalence claim is based on its similarity to predicate devices (GraftPatch and Surgisis Sling) in terms of intended use, technological characteristics, materials, physical construction, and performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K021105
    Date Cleared
    2002-05-10

    (35 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FortaGen is intended to be used for implantation to reinforce soft tissue including, but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, hernias, suture-line reinforcement and reconstructive procedures. The device is intended for one-time use.

    Device Description

    FortaGen consists of a multi-laminate sheet predominantly of Type I porcine collagen. The device is supplied in sheet form in sizes ranging from 5 x 5 cm to 12 x 36 cm in sterile double layer peelable packaging.

    AI/ML Overview

    The provided text describes a medical device called FortaGen™ Surgical Mesh and its clearance by the FDA based on substantial equivalence to predicate devices. However, the information provided does not contain details about specific acceptance criteria, a study proving the device meets those criteria, or a detailed breakdown of performance data against specific metrics.

    The document states:

    • "FortaGen was subjected to a panel of tests to assess biocompatibility, integrity, and performance. The device passed the requirements of all tests." This is a general statement that tests were conducted and passed, but no specific criteria, results, or study details are provided.
    • The FDA letter confirms that the device is "substantially equivalent" to legally marketed predicate devices, which implies that it meets the safety and efficacy standards demonstrated by those predicates. However, it doesn't detail what specific performance metrics constitute substantial equivalence or how FortaGen™'s performance was directly compared against them in a structured study report.

    Therefore, I cannot populate the requested table or answer most of the specific questions about the study design as the information is not present in the provided text.

    Based on the available information, here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance:

      Acceptance CriteriaReported Device Performance
      Biocompatibility requirementsDevice passed all tests
      Integrity requirementsDevice passed all tests
      Performance requirementsDevice passed all tests
      Substantial Equivalence to predicate devices (GraftPatch (K970561) and Surgisis (K980431))Determined to be substantially equivalent by FDA
    2. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified.
      • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The document only states "FortaGen was subjected to a panel of tests."
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not specified. The document does not describe a process involving experts establishing ground truth for a test set. This type of information is typically relevant for diagnostic devices or AI-assisted systems, which this surgical mesh is not.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable/Not specified. Adjudication methods are typically used in clinical studies with human interpretation, which is not described here for the device's technical assessment.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a surgical mesh, not an AI-assisted diagnostic or imaging device. Therefore, an MRMC study comparing human readers with/without AI assistance would not be performed for this product.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical surgical mesh, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the device's performance tests (biocompatibility, integrity, performance), the ground truth would be established by validated test methods and established standards/specifications for medical devices, rather than expert consensus on diagnostic images or pathology. The document does not specify the exact methods or standards used for these tests.
    8. The sample size for the training set:

      • Not applicable. This is a physical surgical mesh, not an AI model that requires a training set.
    9. How the ground truth for the training set was established:

      • Not applicable. This is a physical surgical mesh, not an AI model.

    In summary, the provided text indicates that the FortaGen™ Surgical Mesh passed a panel of tests for biocompatibility, integrity, and performance and was deemed substantially equivalent to predicate devices by the FDA. However, it does not offer the granular detail requested about specific acceptance criteria values, study designs, sample sizes, or methods for establishing ground truth, which are more typically provided for diagnostic devices or those involving interpretation of data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K020049
    Date Cleared
    2002-03-18

    (70 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FortaFlex™ Surgical Mesh is intended to be used for implantation to reinforce soft tissue including, but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, rectal and vaginal prolapse, reconstruction of the pelvic floor, hernias, suture-line reinforcement and reconstructive procedures. The device is also intended for reinforcement of the soft tissues which are repaired by suture or sutere anchors, limited to the supraspinatus, during rotator cuff repair surgery. The device is intended for one-time use.

    Device Description

    FortaFlex Surgical Mesh consists of a multi-laminate sheet predominantly of Type I porcine collagen. The device is supplied in sheet form in sizes ranging from 5 x 5 cm to 12 x 36 cm in sterile double layer peelable packaging.

    AI/ML Overview

    The provided text describes the FortaFlex™ Surgical Mesh, a Class II device intended for soft tissue reinforcement. The submission is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to legally marketed predicate devices, rather than proving safety and effectiveness through clinical trials with specific acceptance criteria and performance metrics.

    Therefore, the document does not contain the information requested regarding acceptance criteria, a study proving device meets acceptance criteria, sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.

    The "Performance Data" section merely states: "FortaFlex Surgical Mesh was subjected to a panel of tests to assess biocompatibility, integrity, and performance. The device passed the requirements of all tests." This is a high-level statement without any specific details of the tests, their acceptance criteria, or the actual results.

    The 510(k) process focuses on demonstrating similarity to an already approved device. The FDA's letter confirms substantial equivalence, which means they believe the new device is as safe and effective as the predicate device(s) for the stated indications, based on similar technological characteristics, materials, and intended use as presented in the submission. It does not involve proving the device meets specific new performance acceptance criteria through the kind of study described in the prompt.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011025
    Date Cleared
    2001-08-24

    (141 days)

    Product Code
    Regulation Number
    878.3300
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K011026
    Date Cleared
    2001-06-13

    (70 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    ORGANOGENESIS, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2