Search Results
Found 2 results
510(k) Data Aggregation
(29 days)
The Rist™ 071 Radial Access Guide Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
The Rist™ 071 Radial Access Guide Catheter is a single lumen, variable stiffness catheter with a stainless steel and nitinol reinforced shaft to provide support. The embedded stainless-steel flat wire cross coil is in the proximal section of the catheter, which transitions to a nitinol round wire single coil in the distal end. The catheter has a radiopaque platinum/iridium marker band on the distal end to aid in visualization. The distal 25 cm of the Rist™ 071 Radial Access Guide Catheter has a hydrophilic coating which reduces the insertion force and allows the catheter to traverse the vasculature more easily. The catheter has a nominal outer diameter of 0.084 inches and a nominal inner diameter of 0.071 inches. It is available in three working lengths: 95 cm, 100 cm, and 105 cm. The Rist™ 071 Radial Access Guide Catheter has a PTFE-lined lumen to reduce friction with other devices introduced through the lumen. It is intended to provide access to the target site via transradial access and, once in place, provides a reinforcing conduit for other intravascular devices. A radial access dilator is included as an accessory. The Rist™ 071 Radial Access Guide Catheter is supplied sterile, non-pyrogenic, and intended for single use only.
The Rist™ 071 Radial Access Guide Catheter is a medical device intended for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The acceptance criteria and supporting studies are detailed below.
1. Table of Acceptance Criteria and Reported Device Performance
The device underwent extensive testing, and the results consistently indicated that "All units met the acceptance criteria" or similar statements. Specific numerical acceptance criteria were not explicitly provided in the document for most tests, but the qualitative statement of meeting criteria is present.
| Test Performed | Acceptance Criteria (Explicit or Qualitative) | Reported Device Performance |
|---|---|---|
| Biocompatibility | ||
| Cytotoxicity | Non-cytotoxicity, no abnormal events (pH change, debris) | The test article is considered non-cytotoxic... No abnormal events such as pH change or debris were noted. |
| Sensitization | No sensitization response | The test article did not elicit a sensitization response. |
| Irritation | No significant dermal reactions | No significant dermal reactions were observed. |
| Acute Systemic Toxicity | No abnormal clinical signs indicative of toxicity | None of the study subjects were observed with abnormal clinical signs indicative of toxicity. |
| Pyrogenicity | No pyrogenic response; all validity criteria met | The test article extracts did not cause a pyrogenic response and all validity criteria were met. |
| Hemocompatibility (Complement Activation) | Not statistically significant (p>0.05) to reference material/comparison article | The test article results... were not statistically significant (p>0.05). |
| Hemocompatibility (Hemolysis) | Blank corrected percent hemolysis above negative control of 0.0% | The test article returned a blank corrected percent hemolysis above the negative control of 0.0%. |
| Hemocompatibility (Platelet & Leukocyte Count) | Platelet count within specification | The platelet count was within specification. |
| Hemocompatibility (Partial Thromboplastin Time) | No more material mediated coagulation abnormalities than predicate in intrinsic pathway | The test article did not create any more material mediated coagulation abnormalities... when compared to the predicate. |
| Sterilization Validation | Product sterility per ANSI/AAMI/ISO 11135:2014 and AAMI TIR 28:2016 | Product was sterile. |
| Packaging | Sterile barrier not compromised | All units met all the packaging acceptance criteria. |
| Catheter and Dilator Tensile Strength | Met tensile strength acceptance criteria per ISO 10555-1:2013/Amd. 1:2017(E) | All units met the tensile strength acceptance criteria. |
| Catheter PTFE Delamination | Met PTFE delamination acceptance criteria | All units met the catheter PTFE delamination acceptance criteria. |
| Torque | Met torque acceptance criteria | All units met the torque acceptance criteria. |
| Catheter Burst Pressure | Met burst pressure acceptance criteria per ISO 10555-1:2013/Amd. 1:2017(E) | All units met the catheter burst pressure acceptance criteria. |
| Particulate Testing | Met particulate acceptance criteria per USP <788> | All units met the particulate testing acceptance criteria. |
| Coating Integrity | Met coating integrity acceptance criteria (no surface irregularities after conditioning and tortuous path) | All units met the coating integrity acceptance criteria. |
| Leak (Liquid) | No leakage per ISO 10555-1:2013/Amd. 1:2017(E) | All units met the liquid leak test acceptance criteria. |
| Leak (Air) | No air leaks per ISO 10555-1:2013/Amd. 1:2017(E) | All units met the air leak test acceptance criteria. |
| Catheter Dimensional Inspection | ID, OD, and working length met acceptance criteria | All units met the catheter dimensional inspection acceptance criteria. |
| Dilator Dimensional Inspection | ID and OD met acceptance criteria | All units met the dilator dimensional inspection acceptance criteria. |
| Chemical Compatibility | No signs of degradation, no obstruction of ID after exposure to saline, dextrose, heparin, radiocontrast | All units met the chemical compatibility acceptance criteria. |
| Visual Inspection of Catheter and Dilator | Smooth tip transition points, no surface defects | All units met the visual inspection acceptance criteria. |
| Catheter Coating Length | Length of hydrophilic coating met specification | All units met the catheter coating length acceptance criteria. |
| Kink Resistance | Met kink resistance acceptance criteria | All units met the kink resistance acceptance criteria. |
| In vitro Simulated Use Study | Product performance verified by physicians | All acceptance criteria were met. |
| Corrosion (from predicate) | Met corrosion acceptance criteria per ISO 10555-1:2013/Amd. 1:2017(E) | All units met the corrosion acceptance criteria. |
| Hub Compatibility (from predicate) | Met hub compatibility acceptance criteria per ISO 594-1:1986 and ISO 594-2:1998 | All units met the hub compatibility acceptance criteria. |
| Radiopacity (from predicate) | Physicians identified distal tip location, catheter shape, and curve locations on fluoroscopic images | All acceptance criteria were met. |
| Label Content (from predicate) | Information included is accurate | All units met the acceptance criteria for label content. |
| Label Legibility (from predicate) | Labeling remained legible after transportation and conditioning | All units met the acceptance criteria for label legibility. |
| Barcode (from predicate) | Scanned barcode matched appropriate information | All units met the acceptance criteria for barcode. |
| Dilator Working Length (from predicate) | Dilator length met specification | All units met the acceptance criteria for dilator working length. |
| Useability (from predicate) | Evaluators confirmed usability as per Instructions for Use | All acceptance criteria were met. |
| Shelf-Life Testing | Product and packaging met all acceptance criteria for 6-month shelf life | All acceptance criteria were met. |
2. Sample Size for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test in the "Test Set." However, it repeatedly mentions "All units" or "All units met" indicating that a sufficient number of samples were tested to achieve statistical significance or to demonstrate compliance.
The data provenance is not specified in terms of country of origin. The studies are described as "Design verification and validation" (in-house testing) and based on applicable ISO/ASTM/USP standards, suggesting a laboratory-based, prospective testing approach. No mention of retrospective or prospective clinical data for the test set is present, as this is a pre-market submission focused on device performance rather than clinical outcomes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
For the "In vitro Simulated Use Study," it states that "A simulated interventional procedure was performed by physicians." The exact number of physicians or their specific qualifications (e.g., "radiologist with 10 years of experience") are not provided.
For the "Radiopacity" test (from the predicate device), "Physicians were shown fluoroscopic images." Again, the specific number and qualifications of these physicians are not detailed.
4. Adjudication Method for the Test Set
The document does not describe any formal adjudication method (e.g., 2+1, 3+1) for the tests. The results indicate a direct pass/fail or "met acceptance criteria" outcome, implying consensus on the directly measurable or observable results rather than a subjective assessment requiring adjudication. For physician-evaluated tests, the document indicates a general 'acceptance criteria met' without detailing disagreement resolution.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was mentioned. The studies focused on device performance characteristics rather than comparing human reader performance with or without AI assistance. The device itself is a physical catheter, not an AI-powered diagnostic tool.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
No standalone algorithm performance study was done. This is consistent with the nature of the device as a physical medical instrument, not a software algorithm.
7. Type of Ground Truth Used
The ground truth for most performance tests was established by adherence to recognized international standards (ISO, ASTM, USP) and predefined internal specifications for various physical, chemical, and biological properties. This includes:
- Physical Measurements: Dimensional accuracy (ID, OD, length), tensile strength metrics, burst pressure values, torque limits, kink resistance thresholds.
- Chemical/Biological Properties: Biocompatibility assays (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility), chemical compatibility.
- Functional Performance: Leak testing, coating integrity, particulate count, simulated use performance, radiopacity (as confirmed by physicians).
- Sterility and Packaging Integrity: Demonstrated through sterilization validation and packaging tests.
8. Sample Size for the Training Set
No training set is applicable or mentioned. This device is a physical medical device, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable as there is no training set for a physical device.
Ask a specific question about this device
(129 days)
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The Riptide™ Aspiration System is designed to restore blood flow in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease. The Riptide™ Aspiration System is designed for use within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries. The Riptide™ Aspiration System is composed of the following components:
- Arc™ Catheter .
- Riptide™ Aspiration Tubing .
- Riptide™ Aspiration Pump ●
- Riptide™ Collection Canister with Intermediate Tubing .
The Arc™ Catheter is introduced into the vasculature through the Split-Y Introducer Sheath. A lubricous, tapered liner is used to create a structure that has both proximal stiffness and distal flexibility. The Arc™ Catheter has a radiopaque marker band encapsulated at the distal tip for visualization under fluoroscopy. The Arc™ Catheter is navigated to the intended treatment site and positioned proximal to the site of occlusion. The Arc™ Catheter is the only component of the Riptide™ Aspiration System that is used intravascularly.
The Riptide™ Aspiration Tubing serves as a conduit to supply vacuum from the Riptide™ Aspiration Pump to the distal tip of the Arc™ Catheter. The Riptide™ Aspiration Tubing provides a connection between the sterile and non-sterile environments. The proximal end of the Riptide™ Aspiration Tubing is connected to the Riptide™ Collection Canister (outside of the sterile environment) while the distal end of the Riptide™ Aspiration Tubing is connected to the Arc™ Catheter (inside the sterile environment). The Riptide™ Collection Canister is connected to the Riptide™ Aspiration Pump (also outside of the sterile environment) via the Intermediate Tubing.
The Riptide™ Aspiration Pump is designed to generate vacuum for the Riptide™ Aspiration System. The vacuum pressure of the Riptide™ Aspiration Pump is set by turning the vacuum control valve until the vacuum gauge reads a minimum of 20inHg but not exceeding 25inHg. The Riptide™ Aspiration Pump is reusable, non-sterile, and intended to be utilized outside of the sterile environment.
The Riptide™ Collection Canister is provided non-sterile and is pre-assembled with the Intermediate Tubing. The Riptide™ Collection Canister with Intermediate Tubing is single-use and the repository for aspirated material. The Riptide™ Collection Canister is placed into the receptacle of the Riptide™ Aspiration Pump while the Intermediate Tubing is connected to the vacuum inlet port.
The provided text describes the acceptance criteria and the studies conducted to demonstrate the substantial equivalence of the Riptide™ Aspiration System to its predicate device, the Penumbra System® and Penumbra Pump MAX™.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
The document lists numerous tests conducted on various components of the Riptide™ Aspiration System, with an explicit statement that the device met the acceptance criteria for each. The specific acceptance criteria values are generally not quantified in the "Results" column, but rather stated qualitatively (e.g., "met the acceptance criteria").
Table 1: Acceptance Criteria and Reported Device Performance (Summary)
| Component | Test Category | Individual Test | Acceptance Criteria Statement (from document) |
|---|---|---|---|
| Arc™ Catheter | Biocompatibility | USP Physicochemical | Meets USP Physicochemical (Pass) |
| ISO MEM Elution Using L-929 Mouse Fibroblast Cells | Non-cytotoxic | ||
| ISO Guinea Pig Maximization Sensitization Test | Non-sensitizer | ||
| ISO Intracutaneous Irritation Test | Non-irritant | ||
| ISO Acute Systemic Injection Test | Non-cytotoxic | ||
| ISO Materials Mediated Rabbit Pyrogen | Non-pyrogenic | ||
| ASTM Hemolysis Assay – Direct Contact and Extract Method | Non-hemolytic | ||
| Complement activation C3a and SC5b-9 Assay | The compliment activation of the C3a and SC5b assays were similar for test and comparison articles. | ||
| Thromboresistance Evaluation | The test and control articles exhibited similar thromboresistant characteristics. | ||
| ISO Bacterial Mutagenicity Test - Ames Assay | Non-mutagenic | ||
| ISO in vitro Mouse Lymphoma with Extended Treatment | Non-mutagenic (non-genotoxic and non-clastogenic) | ||
| ISO in vivo Mouse Micronucleus Assay | Non-mutagenic | ||
| Partial Thromboplastin Time | Non-activator (of intrinsic coagulation pathway) | ||
| in vitro Hemocompatibility Assay | No adverse effect on leukocyte or platelet counts. | ||
| Arc™ Catheter | Bench - Microbial | Ethylene Oxide (EO) Residual | Met the acceptance criteria for EO Residual. |
| Ethylene Chlorohydrin (ECH) | Met the acceptance criteria for ECH. | ||
| Bacterial Endotoxin | Met the acceptance criteria for Bacterial Endotoxin. | ||
| Bench - Packaging | Visual Inspection | Met the acceptance criteria for Visual Inspection. | |
| Bubble Leak | Met the acceptance criteria for Bubble Leak. | ||
| Seal Strength | Met the acceptance criteria for Seal Strength. | ||
| Bench - Performance | Lumen Patency | Met the acceptance criteria for Lumen Patency. | |
| Dimensional Inspection | Met the acceptance criteria for Dimensional Inspection. | ||
| Tip Buckling | Met the acceptance criteria for Tip Buckling. | ||
| Injection Flow Rate | Met the acceptance criteria for Injection Flow Rate. | ||
| Suction Flow Rate | Met the acceptance criteria for Suction Flow Rate. | ||
| Vacuum Resistance | Met the acceptance criteria for Vacuum Resistance. | ||
| Air Aspiration Leak | Met the acceptance criteria for Air Aspiration Leak. | ||
| Hub/Shaft Peak Tensile Force | Met the acceptance criteria for Hub/Shaft Peak Tensile Force. | ||
| Coating Integrity | Characterization only. | ||
| Coating Lubricity/Durability | Met the acceptance criteria for Coating Lubricity/Durability. | ||
| Particulate | Met the acceptance criteria for Particulate. | ||
| Kink Resistance | Met the acceptance criteria for Kink Resistance. | ||
| Liquid Leak | Met the acceptance criteria for Liquid Leak. | ||
| Static/Dynamic Burst | Met the acceptance criteria for Static/Dynamic Burst. | ||
| Corrosion Resistance | Met the acceptance criteria for Corrosion Resistance. | ||
| Torque to Failure | Characterization only. | ||
| Riptide™ Aspiration Tubing | Bench - Microbial | Ethylene Oxide (EO) Residual | Met the acceptance criteria for EO Residual. |
| Ethylene Chlorohydrin (ECH) | Met the acceptance criteria for ECH. | ||
| Bioburden Recovery | Met the acceptance criteria for Bioburden Recovery. | ||
| Bioburden | Met the acceptance criteria for Bioburden. | ||
| Bacterial Endotoxin | Met the acceptance criteria for Bacterial Endotoxin. | ||
| Bench - Packaging | Aseptic Presentation | Met the acceptance criteria for Aseptic Presentation. | |
| Seal Width | Met the acceptance criteria for Seal Width. | ||
| Dye Leak | Met the acceptance criteria for Dye Leak. | ||
| Visual Inspection | Met the acceptance criteria for Visual Inspection. | ||
| Bubble Leak | Met the acceptance criteria for Bubble Leak. | ||
| Legibility | Met the acceptance criteria for Legibility. | ||
| Foreign Material | Met the acceptance criteria for Foreign Material. | ||
| Seal Strength | Met the acceptance criteria for Seal Strength. | ||
| Bench - Performance | Dimensional Inspection | Met the acceptance criteria for Dimensional Inspection. | |
| Identification of Heat Shrink Bands | Met the acceptance criteria for Identification of the Heat Shrink Bands. | ||
| Identification of Flow Switch | Met the acceptance criteria for Identification of the Flow Switch. | ||
| Joint Tensile Strength | Met the acceptance criteria for Joint Tensile Strength. | ||
| Leak (Vacuum and Pressure Decay) | Met the acceptance criteria for Leak. | ||
| Degree of Collapse | Met the acceptance criteria for Degree of Collapse. | ||
| Compatibility (Arc™ Catheter & Collection Canister) | Met the acceptance criteria for Compatibility to the Arc™ Catheter and Riptide™ Collection Canister. | ||
| Riptide™ Aspiration Pump | Bench - Performance | Degrees of Tilt | Met the acceptance criteria for Degrees of Tilt. |
| Durability | Met the acceptance criteria for Durability. | ||
| Maximum Vacuum | Met the acceptance criteria for Maximum Vacuum. | ||
| Flow Capacity | Met the acceptance criteria for Flow Capacity. | ||
| Electrical Safety | Met the acceptance criteria for Electrical Safety. | ||
| Electromagnetic Compatibility | Met the acceptance criteria for Electromagnetic Compatibility. | ||
| Riptide™ Collection Canister with Intermediate Tubing | Bench - Packaging | Packaging Integrity | Met the acceptance criteria for Packaging Integrity. |
| Packaging Labeling | Met the acceptance criteria for Packaging Labeling. | ||
| Bench - Performance | Volume Capacity | Met the acceptance criteria for Volume Capacity. | |
| Volume Marking | Met the acceptance criteria for Volume Marking. | ||
| Canister Strength | Met the acceptance criteria for Canister Strength. | ||
| Overfill Protection | Met the acceptance criteria for Overfill Protection. | ||
| Dimensional Inspection | Met the acceptance criteria for Dimensional Inspection. | ||
| Degree of Collapse | Met the acceptance criteria for Degree of Collapse. | ||
| Port Identification | Met the acceptance criteria for Port Identification. | ||
| Compatibility (Riptide™ Aspiration Pump) | Met the acceptance criteria for Compatibility. | ||
| Riptide™ Aspiration System | Bench - System Performance | Simulated Clot Retrieval | Met the acceptance criteria for Simulated Clot Retrieval. |
| Vacuum Pressure | Met the acceptance criteria for Vacuum Pressure. | ||
| Flow Rate | Met the acceptance criteria for Flow Rate. | ||
| Lumen Collapse | The Arc™ Catheter is resistant to Lumen Collapse. | ||
| Usability | Met the acceptance criteria for Usability. |
Study Information:
The provided document describes non-clinical bench testing and non-clinical animal testing. It explicitly states "Not Applicable" for clinical performance data. Therefore, questions related to human reader studies, ground truth establishment for a test set of clinical images, and training set details are not directly addressed in this document.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size:
- Bench Testing: The document does not specify the exact sample sizes (number of units tested) for each individual bench test. The nature of these tests often involves a batch of devices or components.
- Animal Testing: The document mentions "a porcine model" but does not specify the number of animals used for the animal testing.
- Data Provenance:
- Bench Testing: Conducted by the manufacturer, Micro Therapeutics, Inc. d/b/a ev3 Neurovascular, as part of their R&D and regulatory submission process. Location not specified, but the company is based in Irvine, California, USA.
- Animal Testing: Conducted "in a porcine model." Location not specified.
- Retrospective/Prospective: These are non-clinical studies (bench and animal), so the terms retrospective/prospective in the context of human data acquisition do not directly apply. They are inherently prospective in the sense that the tests were designed and executed to evaluate this specific device for its regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. The studies described are non-clinical (bench and animal). There is no mention of human expert-established ground truth for a test set of clinical images or data. The "ground truth" for these tests comes from objective measurements against defined standards (e.g., ISO, ASTM, USP standards, or internal specifications) and observations in animal models.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As no human expert evaluation of clinical data for ground truth establishment is described, adjudication methods are not relevant to this submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The document explicitly states "Performance Data - Clinical: Not Applicable." This means no human reader studies (MRMC or otherwise) were conducted or submitted as part of this 510(k). The device is not an AI-powered system that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- No. This device is a medical instrument (aspiration system for stroke), not an AI algorithm. Therefore, a standalone algorithm performance study is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Bench Testing: Ground truth is established by objective measurements against pre-defined engineering specifications, international standards (ISO, ASTM, USP), and validated test methods. For example, "The Arc™ Catheter met the acceptance criteria for EO Residual" means the measured residual levels were below the established safe limits per ISO 10993-7.
- Animal Testing: Ground truth is established through direct observation and measurement in a controlled porcine model, in accordance with "21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory Studies". The goal was to evaluate "safety, efficacy, and usability" and compare it to the predicate device.
8. The sample size for the training set
- Not Applicable. No AI/machine learning component is described for this device, so there is no concept of a "training set" in the context of algorithm development.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set for an AI algorithm mentioned, this question is not relevant.
Ask a specific question about this device
Page 1 of 1