Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K231895
    Manufacturer
    Date Cleared
    2023-12-04

    (159 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maxtec MaxBlend 2+p is intended to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen and pressure being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e., hospital, subacute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life-supporting device or life sustaining device.

    Device Description

    The Maxtec MaxBlend 2+p is an oxygen delivery device which incorporates an air/oxygen blender, battery powered oxygen and pressure monitor, and an adjustable flowmeter, all in a single assembly. The integral air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also allows the user to monitor pressure simultaneously using adjustable high and low alarm limits.

    AI/ML Overview

    The document describes the Maxtec MaxBlend 2+p, a medical device combining an air/oxygen blender, oxygen monitor, pressure monitor, and flowmeter. The 510(k) submission seeks substantial equivalence to existing predicate devices (Maxtec MaxBlend 2 - K161718 and MaxO2ME+p - K221734), particularly highlighting the addition of a pressure monitoring feature.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for several performance aspects are implicitly derived from the comparative tables (Table 1 and 2) where the subject device's specifications are listed and compared to the predicate devices. The "Reported Device Performance" for the subject device is simply its stated characteristics, which are claimed to meet or be identical to the predicate device specifications, thereby meeting the acceptance criteria based on substantial equivalence.

    Feature / Performance AspectAcceptance Criteria (from predicate devices)Reported Device Performance (Maxtec MaxBlend 2+p)
    Air/Oxygen Mixer Features
    Gas Supply TypeAir / OxygenAir / Oxygen
    Pressure30 to 75 psi30 to 75 psi
    Mixed gas stability± 1% oxygen± 1% oxygen
    Flow range of BlendersLow flow model – 0-30 Lpm; High flow model – 0-100 LpmLow flow model – 0-30 Lpm; High flow model – 0-70 Lpm (Primary Predicate) / 0-100 Lpm (Secondary Predicate)
    Pressure supply differential alarmAir / oxygen pressure < 20 psi an alarm soundsAir / oxygen pressure < 20 psi an alarm sounds
    Pressure Drop< 6 psi @ 50 psi< 6 psi @ 50 psi
    Bleed flow3-13 Lpm at 50 psi depending upon model3-13 Lpm at 50 psi depending upon model
    Oxygen Monitor Features
    Oxygen measurement range0-100%0-100%
    Total Accuracy±3% Actual oxygen level over full operating temperature range±3% Actual oxygen level over full operating temperature range
    Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°C
    Warm-up TimeNone requiredNone required
    High alarm range16 – 100%16 – 100%
    Operating Temperature15°C – 40°C (59°F – 104°F)15°C – 40°C (59°F – 104°F)
    Storage Temperature-15°C – 50°C (5°F – 122°F)-15°C – 50°C (5°F – 122°F)
    Humidity0-95% (non-condensing)0-95% (non-condensing)
    Expected use-life of sensor1,500,000 O2% hours (approx. 2 years)1,500,000 O2% hours (approx. 2 years)
    Flowmeter Features
    Accuracy± 10% of indicated value when inlet pressure is 50 psi± 10% of indicated value when inlet pressure is 50 psi
    Flow meter ranges0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm (some predicates mention 0-100Lpm as well)0 – 3 Lpm, 0 – 15 Lpm, 0 – 30 Lpm, 0 – 70 Lpm
    Pressure Monitor Features(Based on Secondary Predicate MaxO2 ME+p K221734)(Maxtec MaxBlend 2+p)
    TechnologyMicroprocessor controlled deviceMicroprocessor controlled device
    Pressure sensor typeSolid-state pressure transducerSolid-state pressure transducer
    Pressure Measurement Range-15 to +60 cmH2O-15 to +60 cmH2O
    Pressure Resolution1 cmH2O1 cmH2O
    Display resolution0.5 cm H2O0.5 cm H2O
    Total Accuracy± 1 cmH2O± 1 cmH2O
    Low alarm range1 - 30 cmH2O1 - 30 cmH2O
    High alarm range1 - 60 cmH2O1 - 60 cmH2O
    Alarm delay3 seconds (pressure only)3 seconds (pressure only)
    Zero calibrationYesYes
    Operating Temperature15°C - 40°C, 0-95% RH15°C - 40°C, 0-95% RH
    Storage Temperature-15°C - 50°C @ 95% RH-15°C - 50°C @ 95% RH
    Atmospheric Pressure800 – 1013 mBars800 – 1013 mBars
    Power requirements4x - AA alkaline batteries4x - AA alkaline batteries
    Battery Life5000 hours5000 hours

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states:

    • Non-clinical Testing: "We performed a number of tests to demonstrate that the proposed device performed as intended." It then lists several international standards that the device met. It does not specify a sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The testing appears to be primarily related to engineering and performance validation against standards rather than clinical data.
    • Clinical Testing: "No human clinical testing was performed."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since "No human clinical testing was performed," and the primary testing involved adherence to international standards and non-clinical performance, there is no mention of experts establishing ground truth for a clinical test set. The "ground truth" for the non-clinical testing would be the defined requirements and specifications within the cited standards.

    4. Adjudication Method for the Test Set

    Not applicable, as no human clinical test set is mentioned, and the testing focuses on compliance with technical standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical gas mixer and monitor, not an AI-assisted diagnostic or interpretation tool that would involve human readers or image analysis.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a hardware product with integrated monitoring capabilities. Its "standalone" performance is measured through its adherence to the technical specifications and international standards listed under "Non-clinical Testing." The document states: "In all cases the subject device met the standard requirements." This implicitly confirms standalone performance against established benchmarks.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the performance specifications (e.g., oxygen accuracy, pressure accuracy, flow rates), the ground truth is based on metrological standards and calibrated equipment used in accordance with the listed international standards (e.g., ISO 80601-2-55, ISO 11195). For biocompatibility, the ground truth is established by adherence to ISO 10993 and ISO 18562 standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware medical device with integrated sensors and controls, not a machine learning or AI algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221734
    Manufacturer
    Date Cleared
    2023-04-01

    (290 days)

    Product Code
    Regulation Number
    868.1720
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxO2 ME+p is an oxygen monitor with integrated pressure monitoring intended for continuous monitoring of the concentration of oxygen and pressure being delivered to patients ranging from newborns to adults. It can be used in the hospital and sub-acute settings. The MaxO2 ME+p is not intended as a life-supporting device or life sustaining device.

    Device Description

    The MaxO2 ME +p is a battery powered oxygen and pressure monitor in a single assembly. The oxygen monitor measures the oxygen concentration from a gas source, displays these measured concentrations, and provides user selectable high and low oxygen alarms. It also the user to monitor pressure simultaneously and provides user selectable high and low pressure alarms.

    AI/ML Overview

    The MaxO2 ME+p device is an oxygen monitor with integrated pressure monitoring. The acceptance criteria and supporting study details are as follows:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document compares the MaxO2 ME+p (proposed device) to two predicate devices: Maxtec - MaxO2 ME (K153659) for oxygen monitoring and Caradyne - Criterion 40 (K992101) for pressure monitoring.

    AttributeAcceptance Criteria (Predicate MaxO2 ME)Reported Device Performance (MaxO2 ME+p)Acceptance Criteria (Predicate Criterion 40)Reported Device Performance (MaxO2 ME+p)
    Oxygen Monitoring
    Oxygen % Range0.0 to 100%0.0 to 100%N/AN/A (for pressure comparison)
    Oxygen Resolution0.1%0.1%N/AN/A
    Accuracy and Linearity±1% of full scale at constant temperature, RH and pressure when calibrated at fill scale±1% of full scale at constant temperature, RH and pressure when calibrated at fill scaleN/AN/A
    Total Accuracy (Oxygen)±3% Actual oxygen level over full operating temperature range±3% Actual oxygen level over full operating temperature rangeN/AN/A
    Response Time90% of final value in approx. 15 seconds at 23°C90% of final value in approx. 15 seconds at 23°CN/AN/A
    Low Oxygen Alarm Range15%-99% (>1% lower than high alarm)15%-99% (>1% lower than high alarm)N/AN/A
    High Oxygen Alarm Range16%-99% (>1% higher than low alarm)16%-100% (>1% higher than low alarm) (Note: Proposed device allows 100%, predicate 99% - considered acceptable)N/AN/A
    Pressure MonitoringN/A (for oxygen comparison)N/A (for oxygen comparison)
    Pressure Measurement RangeN/AN/AUp to +99 cmH2O-15 to +60 cmH2O (Note: Lower maximum range than predicate, but this difference is discussed and deemed acceptable)
    Pressure ResolutionN/AN/A1 cmH2O1 cmH2O
    Display Resolution (Pressure)N/AN/A0.5 cmH2O0.5 cmH2O
    Total Accuracy (Pressure)N/AN/A± 1 cmH2O± 1 cmH2O
    Low Alarm Range (Pressure)N/AN/A1-20 cmH2O1 - 30 cmH2O (Note: Adjusted for lower max pressure, deemed acceptable)
    High Alarm Range (Pressure)N/AN/A5 - 99 cmH2O1 – 60 cmH2O (Note: Adjusted for lower max pressure, deemed acceptable)
    Alarm Delay (Pressure)N/AN/A1-20 sec3 seconds (Note: Shorter delay than predicate, deemed acceptable)
    Zero Calibration (Pressure)N/AN/AYesYes
    General
    Indications for UseContinuous monitoring of O2, newborns to adults, pre-hospital, hospital, sub-acuteContinuous monitoring of O2 and pressure, newborns to adults, hospital, sub-acuteMeasurement of airway pressure with positive pressure devices, hospital, sub-acute, home careContinuous monitoring of O2 and pressure, newborns to adults, hospital, sub-acute
    Environments of UsePre-hospital, hospital and sub-acute settingsHospital and sub-acute settingsHospital, sub-acute institutions, home careHospital and sub-acute settings (Note: Not for home care like some predicates, deemed acceptable)
    Patient PopulationNewborns to adultsNewborns to adultsNot specified (implied similar)Newborns to adults
    Operating Temperature15°C-40°C15°C-40°C15°C-45°C15°C-40°C (Note: Slightly narrower range than one predicate, deemed acceptable)
    Storage Temperature-15°C-50°C-15°C-50°C-40°C-60°C-15°C-50°C @ 95% RH (Note: Slightly narrower range than one predicate, deemed acceptable)
    Atmospheric Pressure800-1012 mBars800-1013 mBarsNot specified800-1013 m Bars
    Humidity0-95% (non-condensing)0-95% (non-condensing)15-95%RH0-95% RH (Note: Slightly wider range for proposed device, deemed acceptable)
    Power requirements4 – AA Alkaline batteries4 – AA Alkaline batteriesAC / DC4x - AA alkaline batteries
    Battery LifeApprox. 5000 hoursApprox. 5000 hoursUp to 24 hours5000 hours (Note: Significantly longer battery life for proposed device, deemed acceptable advantage)
    StandardsES 60601-1, IEC 60601-1-2, IEC 60601-1-8, ISO 80601-2-55ES 60601-1, IEC 60601-1-2, AIM 7351731:2017, IEC 60601-1-8, ISO 80601-2-55IEC 601-1, IEC 601-1-2ES 60601-1, IEC 60601-1-2, AIM 7351731:2017, IEC 60601-1-8, ISO 80601-2-55

    2. Sample size used for the test set and the data provenance:

    • The document does not specify a sample size for a test set in terms of clinical data or patient samples.
    • The study primarily relies on non-clinical testing (bench testing and adherence to standards) to demonstrate performance.
    • No human clinical testing or animal testing was performed.
    • The data provenance is not applicable as it's not a study involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. Given that the testing was non-clinical (adherence to standards and bench testing), the "ground truth" would be established by the requirements of those standards and the accuracy of reference measurement equipment. Experts involved would be in engineering, quality assurance, and regulatory affairs, but specific numbers and qualifications are not detailed.

    4. Adjudication method for the test set:

    • An adjudication method is not applicable as there was no study involving human readers or interpretation of results that would require consensus among experts. The testing involved verifying the device's performance against defined technical specifications and industry standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This device is an oxygen and pressure monitor, not an AI-powered diagnostic imaging device or a system designed to assist human readers in interpretation. Therefore, this type of study is not relevant to the MaxO2 ME+p.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not directly applicable in the context of this device. The MaxO2 ME+p is a standalone physical device (monitor) that measures oxygen concentration and pressure. Its performance is evaluated through non-clinical bench testing against established standards and specifications, not through an "algorithm only" performance study in the way it might be for a diagnostic AI. The device's functioning is its standalone performance without human input beyond its operation.

    7. The type of ground truth used:

    • The "ground truth" for the performance evaluation was established by technical specifications, measurements from calibrated reference equipment, and compliance with recognized industry standards. These include:
      • AAMI ANSI ES 60601-1: Medical electrical equipment safety and essential performance.
      • IEC 60601-1-2: Electromagnetic Disturbances.
      • AIM Standard 7351731: Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
      • IEC 60601-1-8: Alarm systems.
      • ISO 80601-2-55: Respiratory gas monitors.
      • ISO 10993-5, -10, -11, -18: Biocompatibility.
      • ISO 18562-2, -3: Particulate material and VOCs.

    8. The sample size for the training set:

    • This information is not applicable. This device is a hardware monitor, not a machine learning or AI algorithm that requires a training set of data.

    9. How the ground truth for the training set was established:

    • This information is not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182362
    Manufacturer
    Date Cleared
    2019-05-23

    (266 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Maxtec MaxCap Neo and MaxCap Ped are indicated to provide a semi-quantitative visualization of the CO2 in the patient airvay. It is an adjunct in patient assessment to be used in conjunction with other methods to determine clinical signs and symptoms by or on the order of a physician.

    Patient Population: MaxCap Neo is intended for use with neonates and infants 250g to 6kg. MaxCap Ped is intended for use with pediatric patients 1kg to 15kg

    Environment of Use: Hospital, sub-acute, pre-hospital, transport.

    Device Description

    The proposed MaxCap Neo and MaxCap Ped are comprised of several components:

    • . Housing with standard 15 mm / 22 mm fittings to connect to the ventilator assist device and a face mask or endotracheal tube
    • . Colorimetric litmus media which has been treated with chemical to detect the presence of CO2 by a change in pH
    AI/ML Overview

    The Maxtec MaxCap Neo and MaxCap Ped devices are semi-quantitative CO2 detectors that use a colorimetric, pH-sensitive dye to visualize CO2 in a patient's airway. The devices are intended for use as an adjunct in patient assessment.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" for each performance metric, but rather lists specifications and performance data for the proposed devices and compares them to predicate and reference devices. We can infer the acceptance criteria from the context of these comparisons and discussions of "similar" or "lower" than predicates.

    Feature / Acceptance Criteria (Inferred)Reported Device Performance (MaxCap Neo and MaxCap Ped)
    Color change accuracyTested and performed
    Color change response timeTested and performed
    Drop testTested and performed
    Duration of UseUp to 6 hours
    Flow Resistance / Back PressureMaxCap Neo: < 2.1 cmH2O @ 5 Lpm; < 5.2 cmH2O @ 10 Lpm; < 9.8 cmH2O @ 15 Lpm MaxCap Ped: < 0.2 cmH2O @ 5 Lpm; < 0.5 cmH2O @ 10 Lpm; < 2.8 cmH2O @ 30 Lpm
    Internal Volume (Dead space)MaxCap Neo: 1.0 cc MaxCap Ped: 2.4 cc
    Device weightMaxCap Neo: < 5 g MaxCap Ped: < 4 g
    Shelf-life (aging)1 year
    Detected % CO2 ranges and Colors0% CO2 - Purple 1.0 to 2.0% CO2 - Beige > 5.0% CO2 - Yellow
    Means of communicating meaning of color changesMatching Colored Label on the Outside of the device
    Means of detecting patient exhalationColor change
    Breathes to effect color change< 6 breaths
    Packaging integrity (per ISTA 2A)Performed on reference device (Maxtec FloCap K133540) and applicable due to identical packaging.
    Operational environmentPerformed on reference device (Maxtec FloCap K133540) and applicable due to identical litmus element materials/chemistry.
    Biocompatibility (per ISO 10993-1)Materials similar to reference device (Maxtec FloCap K133540), which met ISO 10993-1 for cytotoxicity, sensitization, and intracutaneous reactivity. Gas emission VOC and PM2.5 testing also performed.

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical performance testing that was conducted.

    • Sample size: Not explicitly stated for each individual test. The statement "We have performed the following performance data and testing for the new models" implies that a sufficient number of devices were tested to evaluate each listed metric.
    • Data provenance: Bench testing results performed by Maxtec, LLC. The document does not specify a country of origin for the data beyond that. It is retrospective in the sense that the testing has already been completed and is being reported as part of the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided as the device is a colorimetric CO2 detector. The "ground truth" for the performance tests (e.g., CO2 concentration, flow rate, internal volume) would be measured by laboratory equipment and calibrated standards, not by human experts interpreting clinical cases.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable as the described studies are non-clinical performance bench tests, not studies involving human interpretation or clinical adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is a physical, colorimetric CO2 indicator, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a physical, colorimetric CO2 indicator, not an algorithm. The reported performance is based on the device's physical and chemical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for the non-clinical performance tests (e.g., color change accuracy, flow resistance, internal volume) was established using calibrated laboratory equipment and controlled gas mixtures to represent specific CO2 concentrations and flow rates. This is an objective measurement rather than expert consensus or pathology.

    8. The sample size for the training set:

    This information is not applicable. The device is a colorimetric indicator and does not involve machine learning or a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161718
    Manufacturer
    Date Cleared
    2016-10-27

    (127 days)

    Product Code
    Regulation Number
    868.5330
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxBlend 2 and MaxBlend Lite are designed to provide a continuous air/oxygen gas mixture and to continuously monitor the concentration of oxygen being delivered to infant, pediatric, and adult patients. It is a restricted medical device intended for use by qualified, trained personnel, under the direction of a physician, in professional healthcare settings, i.e.. hospital, sub-acute, and nursing-care facilities where the delivery and monitoring of air/oxygen mixtures is required. This is not intended as a life supporting device.

    Device Description

    The MaxBlend family of blenders, MaxBlend 2 and MaxBlend Lite, are oxygen delivery devices which incorporate an air/oxygen blender, battery powered oxygen monitor, and an adjustable flowmeter, all in a single assembly. The air/oxygen blender provides precise mixing of medical grade air and oxygen. The flowmeter provides control of the flow rate delivered. The oxygen monitor measures the oxygen concentration from the blender's gas flow, displays these measured concentrations, and provides user selectable high and low oxygen alarms.

    The MaxBlend 2 incorporates the air/oxygen blender within its enclosure. The MaxBlend Lite may be provided with or without the blender component pre-assembled, allowing the user to install the oxygen monitor/flowmeter module, the MaxBlend Lite component, on an existing compatible blender. The addition of the MaxBlend Lite module is intended to improve the safety of an existing air/oxygen blender that is in the user's possession. Both devices use the exact same oxygen monitor sensor and electronics, MaxO2ME (K153659). Both devices use the exact same components/assemblies in the flowmeter and sensor bleed manifold which form the gas pathway to the patient. The only substantive difference is the form of the enclosure for the monitor electronics.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Maxtec, LLC, regarding their MaxBlend 2 and MaxBlend Lite devices. It aims to demonstrate substantial equivalence to a predicate device, the Bird Sentry Blender (K973646), and references other cleared devices (K153659 – Maxtec MaxO2ME, K883038 - CareFusion / Bird - Blender, K925982 - BioMed Devices - Blender).

    Here's an analysis based on your request, highlighting that this document outlines non-clinical testing for a medical device that mixes and monitors breathing gases, not a typically AI-powered diagnostic imaging device. Therefore, many of the requested categories (like MRMC studies, roles of experts for ground truth, sample sizes for training/test sets for AI models, etc.) are not applicable or are addressed differently in the context of hardware device performance testing.

    Acceptance Criteria and Device Performance (Non-Clinical/Hardware)

    The acceptance criteria are primarily based on meeting the performance specifications of the predicate device and established standards for medical devices of this type. The "reported device performance" is generally stated as "passed" or "meets the acceptance criteria" for various tests.

    1. Table of Acceptance Criteria and Reported Device Performance

    AttributeAcceptance Criteria (Predicate / Standards Equivalent)Reported Device Performance (MaxBlend 2 and MaxBlend Lite)
    Indications for UseContinuous air/oxygen gas mixture, monitor O2 concentration for infant, pediatric, adult patients in institutional settings; not life supporting.Continuous air/oxygen gas mixture, continuously monitor O2 concentration for infant, pediatric, adult patients in professional healthcare settings (hospital, sub-acute, nursing-care facilities); not life supporting.
    Environments of UseInstitutional (healthcare settings), Not for MRI.Professional healthcare settings (hospital, sub-acute, nursing-care facilities), Not for MRI.
    Patient PopulationInfant, pediatric, and adult patients.Infant, pediatric, and adult patients.
    Weight2 kg2.4 kg
    Power source of oxygen monitor2 x AA Alkaline batteries4 x AA Alkaline batteries
    Gas Supply TypeAir / Oxygen, 30 to 75 psiAir / Oxygen, 30 to 75 psi
    % Oxygen Control21 - 100%, Accuracy ± 3%21 - 100%, Accuracy ± 3%
    Mixed Gas Stability± 1% oxygen± 1% oxygen
    Flow Range of BlendersLow flow: 0-30 Lpm; High flow: 0-100 Lpm.Low flow: 0-30 Lpm; High flow: 0-100 Lpm.
    Pressure Supply Differential AlarmAir / oxygen pressure < 20 psi alarms sounds.Air / oxygen pressure < 20 psi alarms sounds.
    Pressure Drop≤ 6 psi @ 50 psi≤ 6 psi @ 50 psi
    Bleed Flow3-13 Lpm at 50 psi depending on model.3-13 Lpm at 50 psi depending on model.
    Oxygen Measurement Range0-100%0-100%
    Total Accuracy (Oxygen)+3% Actual oxygen level over full operating temperature range.+3% Actual oxygen level over full operating temperature range.
    Response Time (Oxygen Monitor)< 20 seconds90% of final value in approx. 15 seconds at 23°C (meets/exceeds criterion).
    Warm-up TimeNone required.None required.
    High Alarm Range (Oxygen)18 - 99%16 - 100% (broader range, considered equivalent as it allows broader control and is set by clinician).
    Operating Temperature15°C - 40°C (59°F – 104°F)15°C - 40°C (59°F – 104°F)
    Storage Temperature-15°C - 50°C (5°F – 122°F)-15°C - 50°C (5°F – 122°F)
    Humidity0-100% (non-condensing)0-95% (non-condensing) (slight difference, but likely acceptable for intended environment).
    Expected Use-Life of Sensor750,000 O2% hours1,500,000 O2% hours (approx. 2 years) (superior to predicate, identical to reference device).
    Flowmeter AccuracyTypically ± 10% of indicated value.± 10% of indicated value when inlet pressure is 50 psi.
    Flow Meter Ranges0-30 Lpm for Low flow; 0-100 Lpm for High flow (user discretion).0-3 Lpm, 0-30 Lpm, 0-70 Lpm, 0-100 Lpm (broader range tested, at user's discretion, accuracy is the same).
    Standards ComplianceTested to older versions of equivalent standards (e.g., electrical safety).ANSI/AAMI/ES 60601-1, IEC 60601-1-2, ISO 80601-2-55, IEC 60601-1-8, ISO11195, ISO 15002:2008, ISO15001. (Indicates compliance with current standards).
    Biocompatibility (Materials in Gas Pathway)Materials externally communicating, tissue (permanent), no humidified gases.Externally communicating, tissue (permanent), no humidified gases. VOC and PM2.5 testing performed and passed. Materials identical to cleared Maxtec products and blenders.
    Cleaning Durability[Implied standard durability]Durability of marking tested with standard cleaning wipes/disinfectants. (Passed)
    Packaging Validation[Implied standard validation]Testing per ISTA 2A. (Passed)
    Usability[Implied safe and effective use]Usability testing performed with healthcare professionals. (Passed)

    Overall Conclusion found in the document: "In all cases the proposed device passed or meets the acceptance criteria. One can find the proposed device substantially equivalent to the predicate Bird Sentry Blender (K973646)."


    Regarding the AI/Diagnostic-Specific Questions (Not Directly Applicable to this Hardware Device but Addressed for Completeness):

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable in the AI/ML sense. This submission relates to physical device performance and safety testing. The "test set" would refer to devices manufactured and subjected to various non-clinical engineering and performance tests (e.g., electrical safety, EMC, accuracy, durability, environmental conditions, etc.). The sample size for these tests is not explicitly stated in terms of a "data set" but rather implies a sufficient number of representative devices for engineering validation.
    • Data Provenance: The testing was conducted by or on behalf of Maxtec, LLC, in the USA (Salt Lake City, Utah). It's "prospective" in the sense that the new devices were built and tested to meet pre-defined criteria, not a retrospective analysis of patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Not Applicable in the AI/ML sense. Ground truth for device performance is established by engineering specifications, international and national standards (e.g., ISO, IEC, ANSI/AAMI), and comparison to a legally marketed predicate device.
    • Qualifications of Experts: The experts are likely engineers, quality assurance personnel, and regulatory affairs professionals involved in device design, manufacturing, and testing, potentially with input from clinical users for usability studies. Their qualifications would be in engineering, quality management systems (QMS), and regulatory compliance for medical devices.

    4. Adjudication Method for the Test Set:

    • Not Applicable in the AI/ML sense. Device performance testing involves objective measurements against a standard or specification. There isn't an "adjudication" process like a clinical expert consensus on an image. Test results either pass or fail the predetermined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, not done. This is a hardware device focused on gas mixing and oxygen monitoring, not a diagnostic imaging AI algorithm that assists human readers. Therefore, an MRMC study is not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Not applicable as an "algorithm only" study. The core components (oxygen monitor and blender) are "standalone" in function (they perform their tasks independently) and their performance was evaluated by integrating them into the MaxBlend family. The document states "The major components... have already been cleared for the identical indications for use, population, and environment of use. Integrating them into a single assembly for user convenience does not alter their performance."

    7. The Type of Ground Truth Used:

    • The "ground truth" for this device's performance is based on:
      • Engineering specifications and design requirements.
      • Performance characteristics of the legally marketed predicate device (Bird Sentry Blender - K973646).
      • Compliance with recognized industry standards (e.g., ISO, IEC, ANSI/AAMI).
      • Comparison to previously cleared reference devices (K153659 – Maxtec MaxO2ME, K883038 - CareFusion / Bird - Blender, K925982 - BioMed Devices - Blender) which contribute components to the MaxBlend family.
      • Objective physical measurements and test results (e.g., accuracy of oxygen control, response time, flow rates, pressure, etc.).

    8. The Sample Size for the Training Set:

    • Not Applicable in the AI/ML sense. There is no "training set" for an AI model. This is a hardware device.

    9. How the Ground Truth for the Training Set Was Established:

    • Not Applicable in the AI/ML sense. No training set exists.
    Ask a Question

    Ask a specific question about this device

    K Number
    K153659
    Device Name
    MaxO2ME
    Manufacturer
    Date Cleared
    2016-06-10

    (172 days)

    Product Code
    Regulation Number
    868.1720
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MaxO2ME oxygen monitor is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. It can be used in the pre-hospital, hospital and sub-acute settings. The MaxO2ME is not intended as a life supporting device.

    Device Description

    The MaxO-ME is a handheld oxygen analyzer/monitor capable of measuring the oxygen concentration from 0% to 100% in a sample gas. A MAX-550E oxygen sensor outputs a voltage which is used by the Max O2ME to determine the concentration of oxygen based on a calibration at room air or 100% oxygen. The MaxO2ME contains alarms that can be controlled by the user to set a maximum or minimum allowable oxygen concentration.

    AI/ML Overview

    Below is an analysis of the provided text regarding the MaxO2ME oxygen monitor's acceptance criteria and studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document outlines acceptance criteria implied by the "Performance Specifications" and the reported performance of the MaxO2ME device compared to a predicate device (Precision Medical PM5900 - K063096) and a reference device (MiniOX - K961644). The "Discussion" section for "Non-clinical Testing" states, "In all cases the proposed device passed or meets the acceptance criteria," but does not explicitly list the criteria themselves or the specific values for each test type in the Performance Testing section beyond the initial performance specifications table.

    Based on the "Performance Specifications - The following is a list of the differences between the proposed device and the predicate" and the subsequent "Discussion" for performance, some criteria and performance points can be inferred:

    Performance AttributePredicate Device (PM5900) Acceptance Criteria (or comparable performance)MaxO2ME Reported Device PerformanceReference Device (MiniOX) Performance (for context)
    Measurement Range0.0 to 100%0.0 to 100%(Not explicitly stated for MiniOX)
    Resolution0.1%0.1%(Not explicitly stated for MiniOX)
    Accuracy and Linearity±1% of full scale (constant T, RH, P; calibrated at full scale)±1% of full scale (constant T, RH, P; calibrated at full scale)(Not explicitly stated for MiniOX)
    Total Accuracy±3% Actual oxygen level over full operating temperature range±3% Actual oxygen level over full operating temperature range(Not explicitly stated for MiniOX)
    Response Time90% of final value in 12 seconds at 25°C90% of final value in approx. 15 seconds at 23°C90% in 20 to 30 seconds
    Warm-up TimeNone required (predicate is 3 seconds less than 15s warm-up time of proposed device)None required(Not explicitly stated for MiniOX)
    Operating Temperature10°C – 45°C (50°F – 113°F)15°C – 40°C (59°F – 104°F)(Not explicitly stated for MiniOX)
    Storage Temperature-15°C – 50°C (5°F – 122°F)-15°C – 50°C (5°F – 122°F)(Not explicitly stated for MiniOX)
    Atmospheric PressureUp to 8,000 ft.800 – 1012 mBars(Not explicitly stated for MiniOX)
    Humidity0-95% (non-condensing)0-95% (non-condensing)(Not explicitly stated for MiniOX)
    Battery LifeApprox. 1,500 – 2,000 hours, typical useApprox. 5000 hours, typical use(Not explicitly stated for MiniOX)
    Expected Sensor Life> 1,000,000% O2 Hours> 1,500,000 % O2 Hours, over 2 years typical application(Not explicitly stated for MiniOX)
    Low Oxygen Alarm Range15% - 99% (>1% lower than high alarm)15% - 99% (>1% lower than high alarm)15 – 99%
    High Oxygen Alarm Range18% - 99% (>1% higher than low alarm)16% - 100% (>1% higher than low alarm)16 - 100%
    Alarm Systems (Audible)Not explicitly detailed for audibleNominal 975 Hz audio buzzer (IEC 60601-1-8)Not explicitly detailed for audible

    Study Proving Acceptance Criteria:
    The document states: "We performed a number of tests to demonstrate that the proposed device performed as intended."
    The following non-clinical performance tests were conducted:

    • ISO 80601-2-55 Performance of respiratory gas monitors
    • ISTA2A Shipping Validation Test Report
    • Sensor performance Test Report
    • Gas leakage Test Report
    • Interfering gas effects Test Report
    • Temperature compensation Test Report
    • Drift of measurement accuracy Test Report
    • MaxO2ME Operating and Storage Environment Report
    • Device Cleaning Report and Disinfection Test
    • Measurement accuracy Test Report
    • IEC 60601-1 Electrical safety
    • IEC 60601-1-2 - EMC
    • IEC 60601-1-8 - Alarms
    • Shelf-life / Real-time

    The document concludes: "In all cases the proposed device passed or meets the acceptance criteria."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size for any test set or the data provenance. The tests listed are general performance and safety tests for a medical device, which typically involve device units rather than patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. The tests performed are objective, non-clinical engineering and performance tests on the device itself, not requiring expert ground truth for interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the tests are objective, non-clinical performance and safety tests, not requiring adjudication of results from multiple reviewers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/imaging device. No MRMC study was conducted. The device is an oxygen monitor, and its performance tests are related to its physical and functional specifications.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies conducted were standalone tests of the device itself ("algorithm only" in the sense of the device's functionality) without human-in-the-loop performance being evaluated. The device is an oxygen monitor and its performance is measured against technical specifications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the performance and safety tests would have been established by calibrated reference equipment and standardized testing protocols (e.g., ISO and IEC standards). For example, a gas analyzer would be calibrated against known gas concentrations to determine its accuracy.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI-based device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/machine learning model, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112402
    Manufacturer
    Date Cleared
    2011-12-14

    (114 days)

    Product Code
    Regulation Number
    868.1720
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltraMaxO2 Oxygen Analyzer is a tool used to measure oxygen purity, flow and pressure at the outlet of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer is intended to be used in an environment where oxygen concentrators are being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient Homes, and Respiratory Device Service and Repair Centers.

    Device Description

    The UltraMaxO2 device is used for checking oxygen concentrator performance with the measure of oxygen purity, and flow at the outlet of an oxygen concentrator. The UltraMaxO2 Oxygen Analyzer functions by passing an ultrasonic pulse through the gas sample and measuring the amount of time required for the pulse to transit the sample chamber. The transit time is converted into a gas concentration via calibration data stored in the device. This can be done because the transit time varies according to the molecular mass of the gas in the chamber. Flow is determined using the difference between the ultrasonic pulse traveling against the flow and the ultrasonic pulse traveling with the flow. Pressure is measured using a separate pressure sensor that measures the pressure build-up in the device when the output port is blocked. The pressure is displayed in either kPa or PSI determined by a user operable switch in the battery compartment. The materials of the UltraMaxO2 include ABS plastic for the enclosures and the ultrasonic oxygen sensor tube body, adhesive backed Polycarbonate labels, PVC tubing and nickel plated brass fittings for the flow path, electronic circuitry including transducers/receivers, pressure, temperature and humidity sensors and an LCD screen.

    AI/ML Overview

    The UltraMaxO2 Oxygen Analyzer is a device designed to measure oxygen purity, flow, and pressure from oxygen concentrators.

    Here's a breakdown of its acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the UltraMaxO2 Oxygen Analyzer are implicitly derived from the performance claims and comparison to predicate devices. The study demonstrates that the UltraMaxO2 meets or exceeds the performance of its predicates.

    ParameterAcceptance Criteria (Predicate Performance - typically the most stringent or comparable)Reported UltraMaxO2 Performance
    Oxygen Concentration
    Measurement Range20.8 – 95.7% (DigiFlo), 73 – 95.6% (Check O2 Plus), 20.9 – 100% (Pro2 Check)20.9 – 96%
    Accuracy± 1.8% (DigiFlo), ± 2% Full Scale (Check O2 Plus), ± 2% Full Scale (Pro2 Check)± 1.5% Full Scale (at constant temp. & optimal flow)
    Flow Measurement
    Measurement Range0 – 20 LPM (O2, DigiFlo), 0 – 10 LPM (Air, DigiFlo), 0 – 6 LPM (Check O2 Plus), 0 – 10 LPM (Pro2 Check)0 – 10 LPM
    Accuracy± 0.2 LPM (DigiFlo), ± 0.3 LPM (Check O2 Plus), ± 0.3 LPM (Pro2 Check)± 0.2 LPM
    Pressure Measurement
    Measurement Range0 – 35 PSI (DigiFlo), 0 – 10 PSI (Check O2 Plus), 0 – 10 PSI (Pro2 Check)0.5 – 50 PSI (3.4 – 344 kPa)
    Accuracy± 0.5% (DigiFlo), ± 2% Full Scale (Check O2 Plus), ± 1% Full Scale (Pro2 Check)± 0.5%
    Response Time0.1 sec (DigiFlo), 1 min and 45 sec ± 5 sec (Check O2 Plus), 10 sec (Pro2 Check)Less than or equal to 17 sec.
    Operating Temperature10 – 40 C° (DigiFlo), 15 – 35 C° (Check O2 Plus), 0 – 41 C° (Pro2 Check)15 – 40 C° (59 – 104 F°)
    Other Features
    SensorUltrasonic (All Predicates)Ultrasonic
    Low Battery Alarm or IndicatorLow Battery Indicator (All Predicates)Low Battery Indicator
    Power Source1 Battery: Alkaline 9V (All Predicates)2 Batteries: AA (Alkaline) – 2 x 1.5 V
    DisplayLCD (All Predicates)LCD
    Dimensions9" x 1.5" x 1" (DigiFlo), 3.3" x 7.5" x 1.25" (Check O2 Plus), 3.60" x 5.75" x 1.29" (Pro2 Check)3.16" x 5.10" x 1.04" (80.3mm x 129.5mm x 26.4mm)
    Weight179 g (DigiFlo), 10 oz. (295 g) (Check O2 Plus), 9 oz. (255.15 g) (Pro2 Check)0.4 lbs (181 g)

    Study Proving Device Meets Acceptance Criteria:

    The provided document refers to non-clinical functional and performance tests that were conducted to establish substantial equivalence. These tests aimed to demonstrate that the UltraMaxO2 performs as well as, or better than, the legally marketed predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "non-clinical functional and performance tests" but does not detail the number of units tested or the number of measurements taken.
    • Data Provenance: The tests were non-clinical, likely conducted in a laboratory or manufacturing environment. The country of origin of the data is not specified, but the submitter (Maxtec, LLC) is based in Salt Lake City, Utah, USA, implying the testing likely occurred in the USA. The data is prospective as it was generated specifically for the submission to demonstrate performance.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. This was a non-clinical device performance test, not an evaluation requiring expert interpretation of results for ground truth. The 'ground truth' would have been established by reference standards or calibrated measurement equipment.
    • Qualifications of Experts: N/A

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. As this was a non-clinical performance test comparing measurements against known standards or predicate device specifications, an adjudication method for interpretations is not relevant.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical studies were performed for the UltraMaxO2 as the device represents a well known technology for a recognized indication as evidenced...by comparison to the predicate devices currently cleared for sale in the US market."
    • Effect Size of Human Readers Improve with AI vs. without AI Assistance: Not applicable, as no MRMC or clinical study involving human readers or AI assistance was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Yes, the fundamental performance evaluation was a standalone assessment of the device's accuracy in measuring oxygen purity, flow, and pressure. While not an 'algorithm-only' test in the AI sense, it represents the device's inherent capability independent of human interpretation. The "non-clinical functional and performance tests" assess the device's outputs directly against established standards or predicate device performance.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: For the non-clinical functional and performance tests, the ground truth would have been established using calibrated reference standards or known gas mixtures/flow rates/pressures. For example, a known concentration of oxygen gas would be fed into the device, and the device's reading would be compared to that known concentration. Similarly, known flow rates and pressures would be applied. When comparing to predicate devices, the predicate device's established performance served as a benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. The UltraMaxO2 Oxygen Analyzer is a measurement device, not an AI/ML algorithm that requires a "training set" in the conventional sense. The device uses an ultrasonic pulse measurement method with "calibration data stored in the device." This calibration data would be established during manufacturing and testing, and is not a "training set" like those used for machine learning models.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable in the AI/ML context. The device's internal calibration data is established by feeding known, precise inputs (e.g., specific oxygen concentrations, flow rates, pressures) into the device during its manufacturing and calibration process. The device's internal algorithms are then set to accurately report these known inputs. This would involve highly accurate laboratory equipment as the ground truth.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1