K Number
K063096
Device Name
PRECISION OXYGEN MONITOR
Date Cleared
2007-02-27

(140 days)

Product Code
Regulation Number
868.1720
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Precision Medical Oxygen Monitor is intended to measure the concentration of oxygen being delivered to the patient. The oxygen monitor is not intended as a life supporting device.
Device Description
The Precision Medical, Inc. Oxygen monitor is made of two components, the monitor, and the sensor, attached by a coiled cable. The monitor houses the LCD screen, circuit board, key pad, and batteries. The sensor is a galvanic oxygen sensor that produces a millivolt signal that is proportional to partial pressure of oxygen being measured. The monitor converts the millivolt signal sent from the sensor to a percent oxygen concentration displayed on the LCD screen.
More Information

No
The description details a simple sensor and display system with no mention of AI or ML algorithms for data processing or interpretation.

No.
The device's intended use is to measure oxygen concentration, not to provide therapy or directly treat a disease or condition, and it is explicitly stated that it is "not intended as a life supporting device."

No.
The Precision Medical Oxygen Monitor is intended to measure the concentration of oxygen being delivered to the patient. It provides a measurement of oxygen concentration, which is a vital sign, but it does not diagnose a medical condition or disease.

No

The device description explicitly states it is made of two components, a monitor and a sensor, which are hardware components.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to "measure the concentration of oxygen being delivered to the patient." This is a measurement of a gas being delivered externally to the patient, not a test performed on a biological sample taken from the patient.
  • Device Description: The device measures the partial pressure of oxygen in a gas stream using a sensor and displays it as a percentage. This is consistent with measuring the composition of a gas, not analyzing a biological sample.
  • Lack of Biological Sample Analysis: There is no mention of collecting or analyzing any biological samples (blood, urine, tissue, etc.). IVD devices are specifically designed to perform tests on such samples.

Therefore, the Precision Medical Oxygen Monitor, as described, falls under the category of a medical device used for monitoring a physiological parameter (oxygen concentration in delivered gas) rather than an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The oxygen monitor provides continuous, direct monitoring of oxygen mixtures in a wide variety of medical applications such as anesthesiology [e.g. anesthesia machines], respiratory devices [e.g. respirators, ventilators, pediatric incubators], and oxygen therapy [e.g., oxygen tents]. The oxygen monitor is to be used by trained healthcare professionals under the supervision, or on the order, of a physician in a hospital [ or other clinical setting. The Precision Medical, Inc. oxygen monitor is not intended for transport use, This device is not an oxygen supply source.

The Precision Medical Oxygen Monitor is intended to measure the concentration of oxygen being delivered to the patient. The oxygen monitor is not intended as a life supporting device.

Product codes (comma separated list FDA assigned to the subject device)

CCL

Device Description

The Precision Medical, Inc. Oxygen monitor is a restricted medical device intended for use by qualified and trained personnel under the direction of a physician in institutional environments where monitoring of oxygen levels are required.

The Precision Medical, Inc. Oxygen monitor is made of two components, the monitor, and the sensor, attached by a coiled cable. The monitor houses the LCD screen, circuit board, key pad, and batteries. The sensor is a galvanic oxygen sensor that produces a millivolt signal that is proportional to partial pressure of oxygen being measured. The monitor converts the millivolt signal sent from the sensor to a percent oxygen concentration displayed on the LCD screen.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained healthcare professionals under the supervision, or on the order, of a physician in a hospital [ or other clinical setting.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The Precision Medical, Inc. Oxygen Monitor has successfully pass tests in the following areas:

Mechanical / Climatic
Operating, Temperature, and Humidity range of the PM5900 Oxygen Monitor Precision Medical Test # 699-4

Device Performance: Performance testing, Accuracy: +/- 1% of full scale @ constant temperature and pressure when calibrated at full scale. Precision Medical Test # 699-1

Electromagnetic Emissions & Immunity Tests per IEC 60601-1-2:2004 Performed by Retlif Testing Laboratories for Precision Medical, Inc. Test Report No. R-2915P

Soft ware validation: Following guidance document: General Principles of Software Validation, Final Guidance for Industry and FDA Staff

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Accuracy: ±1% of full scale at constant temperature, R.H. and pressure when calibrated at full scale.
Total Accuracy: ±3% Actual Oxygen Level over full operating temperature range
Response Time: 90% of final value in approximately 12 seconds at 25°C
Expected Sensor Life: >1,000,000% O2 Hours
Alarm Accuracy: exact to displayed alarm value

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K911344

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K002384

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.1720 Oxygen gas analyzer.

(a)
Identification. An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.(b)
Classification. Class II (performance standards).

0

K063096

510(k) Summary

FEB 27 2007

Precision Medical, Inc. Oxygen Monitor

Submitter Information

| Submitter | Precision Medical, Inc.
300 Held Drive
Northampton, Pa.
18067 |
|-----------------------------|------------------------------------------------------------------------|
| Contact | James Parker
Quality Assurance Manager |
| Tel: | (610)-262-6090 Extensions 228 |
| Fax: | (610)-262-6080 |
| Preparation Date: | October 2, 2006 |
| Device Name | |
| Proprietary Name: | Precision Oxygen Monitor |
| Common Name: | Oxygen Analyzer |
| Classification Name: | Analyzer, Gas Oxygen as per CFR 868.1720 |
| Classification Product Code | CCL |
| Device Classification | Class II |

Predicate Device Equivalence

Precision Medical, Inc. is claiming substantial equivalence to the Cerion Hand Held Oxygen Analyzer. Now (Maxtec) 510K number K911344

1

Device Description

The Precision Medical, Inc. Oxygen monitor is a restricted medical device intended for use by qualified and trained personnel under the direction of a physician in institutional environments where monitoring of oxygen levels are required. Additional information:

The Precision Medical, Inc. Oxygen monitor is made of two components, the monitor, and the sensor, attached by a coiled cable. The monitor houses the LCD screen, circuit board, key pad, and batteries. The sensor is a galvanic oxygen sensor that produces a millivolt signal that is proportional to partial pressure of oxygen being measured. The monitor converts the millivolt signal sent from the sensor to a percent oxygen concentration displayed on the LCD screen.

Precision Medical, Inc. will purchase the Sensor from IT Dr. Gambert Gmbh Hinter dem Chor 21 23966 Wismar Germany 510K number K002384

Intended Use

The oxygen monitor provides continuous, direct monitoring of oxygen mixtures in a wide variety of medical applications such as anesthesiology [e.g. anesthesia machines], respiratory devices [e.g. respirators, ventilators, pediatric incubators], and oxygen therapy [e.g., oxygen tents]. The oxygen monitor is to be used by trained healthcare professionals under the supervision, or on the order, of a physician in a hospital [ or other clinical setting. The Precision Medical, Inc. oxygen monitor is not intended for transport use, This device is not an oxygen supply source.

1

2

02 MonitorPredicate (Maxtec)Precision Medical
Specifications
Measurement0.0 - 100%0.0 - 100%
Range
Resolution0.10%0.10%
Accuracy &±1% of full scale at constant±1% of full scale at constant
Linearitytemperature, R.H. and pressure when
calibrated at full scale.temperature, R.H. and pressure when
calibrated at full scale.
Total Accuracy±3% Actual Oxygen Level over full
operating temperature range±3% Actual Oxygen Level over full
operating temperature range
Response Time90% of final value in approximately 15
seconds at 23°C90% of final value in approximately 12
seconds at 25°C
Warm-up Timenone requirednone required
Operating
Temperature15°C - 40°C [59°F - 104°F]10°C - 45°C [50°F - 113°F
Storage
Temperature-15°C - 50°C [5°F - 122°F]-15°C - 50°C [5°F - 122°F]
Humidity0 - 95% [non-condensing]0 - 95% [non-condensing]
Power
Requirements2, AA Alkaline Batteries
[2 x 1.5 Volts]4, AA Alkaline Batteries
[4 x 1.5 Volts]
Battery Lifeapproximately 3000 hours in typical useapproximately 1500 hours in continuous
use.
Non alarm condition
Low Battery
Indication"LOW BAT" icon displayed on screen"LOW BAT" icon displayed on screen
Sensor TypeMaxtec MAX-250E galvanic fuel cellPrecision Medical 504877
Galvanic fuel cell
Expected Sensor
Life>900,000% O2 Hours>1,000,000% O2 Hours
Alarm Systemhigh/low alarms, flashing red LEDs,
audible alarm beeperhigh/low alarms, flashing red LEDs,
audible alarm beeper
Low Alarm Range18% - 99%18% - 99%
High Alarm Range19% - 99%19% - 99%
Alarm Accuracyexact to displayed alarm valueexact to displayed alarm value
Dimensions3.5"[W] x 5.5"[H] x 1.5"[D] [89mm x
140mm x 38mm]3.6"[W] x 5.4"[H] x 1.5"[D] [89mm x
140mm x 38mm]
Weightapproximately 0.92 lbs. [417g]approximately 1.11 lbs. [430g]
Cable Length10 ft. [3m]10 ft. [3m]
Diverter Fittingfits industry standard, 15 mm "T" adapterfits industry standard, 15 mm "T" adapter

11

. ..

2

.

3

Summary of Performance Testing

The Precision Medical, Inc. Oxygen Monitor has successfully pass tests in the following areas:

Mechanical / Climatic

Operating, Temperature, and Humidity range of the PM5900 Oxygen Monitor Precision Medical Test # 699-4

Device Performance: Performance testing, Accuracy: +/- 1% of full scale @ constant temperature and pressure when calibrated at full scale. Precision Medical Test # 699-1

Electromagnetic Emissions & Immunity Tests per IEC 60601-1-2:2004 Performed by Retlif Testing Laboratories for Precision Medical, Inc. Test Report No. R-2915P

Soft ware validation: Following guidance document: General Principles of Software Validation, Final Guidance for Industry and FDA Staff,

Conclusions

In Summary, Precision Medical, Inc. has demonstrated that the Precision Medical, Inc. oxygen Monitor is safe and effective. The combined testing and analysis of results provides assurance that the device meets the specifications and is safe and effective for the intended use.

Risk Analysis

Precision Medical, Inc. is using International Standard ISO 14971 Medical Devices Application of Risk management to medical devices Risk analysis will be completed for this device, before the final release. This document will be updated and released as part of design control.

Design control

Precision Medical, Inc. is in compliance with 21 CFR 820.30 for design control. The design inputs/outputs matrix has been developed using form PMF 202. These documents will be completed before the release of the product.

4

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. James Parker Quality Assurance Manager Precision Medical, Incorporated 300 Held Drive Northampton, Pennsylvania 18067

FEB 2 7 2007

Re: K063096

Trade/Device Name: Precision Oxygen Monitor Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: II Product Code: CCL Dated: January 29, 2007 Received: February 1, 2007

Dear Mr. Parker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

5

Page 2 -- Mr. Parker

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Clues

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

Indications for Use

510(k) Number : K063096

Device Name: Precision Medical, Inc. Oxygen Monitor

Indications for Use:

The Precision Medical Oxygen Monitor is intended to measure the concentration of oxygen being delivered to the patient. The oxygen monitor is not intended as a life supporting device.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

.

. lll. hhh

K06 3094