Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K160721
    Date Cleared
    2017-03-07

    (356 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vado Steerable Sheath is indicated for introducing various catheters into the vasculature and into the chambers of the heart including the left side of the heart through the interatrial septum.

    Device Description

    The Vado Steerable Sheath consists of an 8.8F (ID) sheath and dilator, which is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter exchange. A sideport with three-way stopcock is provided for aspiration, fluid infusion, blood sampling and pressure monitoring. A handle equipped with a deflection knob, deflects the sheath tip. The steerable introducer features distal vent holes for aspiration and radiopaque markers to facilitate visualization under fluoroscopy.

    AI/ML Overview

    The provided text describes the regulatory clearance of a medical device, the "Vado Steerable Sheath," and details the non-clinical testing performed to demonstrate its substantial equivalence to a predicate device. It does not describe an AI/ML powered device, nor does it contain information about acceptance criteria or supporting studies that are typically associated with performance metrics of AI/ML models (e.g., sensitivity, specificity, AUC).

    Therefore, I cannot provide the information requested in the prompt, as the document does not contain:

    1. A table of acceptance criteria and reported device performance in the context of AI/ML metrics.
    2. Sample size used for the test set or data provenance for an AI/ML model.
    3. Number of experts or their qualifications for establishing ground truth for AI/ML validation.
    4. Adjudication method for AI/ML ground truth establishment.
    5. Information on a multi-reader, multi-case (MRMC) comparative effectiveness study for AI assistance.
    6. Standalone performance (algorithm only) for an AI/ML model.
    7. Type of ground truth used (expert consensus, pathology, outcomes data) for AI/ML validation.
    8. Sample size for the training set of an AI/ML model.
    9. How the ground truth for the training set was established for an AI/ML model.

    The document focuses on the mechanical and material performance of a medical device (a catheter introducer), including:

    • Biocompatibility tests: To ensure the device is safe for human contact.
    • Bench performance tests: To assess physical characteristics like dimensions, leakage resistance, deflection, aspiration, and strength.
    • Packaging integrity tests: To ensure sterility and product protection.

    These tests are standard for a traditional medical device and are aimed at demonstrating that the device meets its functional and performance characteristics and is equivalent to the predicate device in terms of safety and effectiveness, not that it performs an interpretive or diagnostic function with AI/ML.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162427
    Date Cleared
    2017-01-27

    (150 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vado® Bi-Directional Steerable Sheath is introducing various cardiovascular catheters into the vasculature and into the chambers of the heart including the heart through the interatrial septum.

    Device Description

    The Vado® Bi-Directional Steerable Sheath consists of a dilator and steerable sheath which are designed to provide intracardiac access and flexible, stable catheter positioning in the cardiac anatomy. A hemostasis valve in the handle minimizes blood loss during catheter introduction and/or exchange. A side port with 3-way stopcock is provided for air or blood aspiration, fluid infusion, blood sampling and pressure monitoring. A deflection knob on the handle deflects the tip of the sheath clockwise ≥ 180° and counterclockwise ≥ 90°. Distal tip vent holes facilitate aspiration and a radiopaque marker permits visualization of the sheath tip under fluoroscopy.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for the Kalila Medical Vado® Bi-Directional Steerable Sheath (K162427). This notification is for a medical device (catheter introducer), not an AI/ML powered device, therefore much of the requested information (e.g., AI/ML specific sections like sample size for test/training sets, number of experts for ground truth, MRMC study, standalone performance) is not applicable or available in this document.

    However, I can provide information based on the device's technical and performance characteristics as described in the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document details various performance tests conducted to demonstrate the device's functional and safety characteristics, in comparison to predicate devices, rather than explicit acceptance criteria with numerical performance targets that are then met. The general acceptance criterion for all tests is that the device "met its functional and performance characteristics in accordance with applicable industry standards and is equivalent to the predicate devices" and that differences "do not adversely affect safety and effectiveness."

    Test CategorySpecific Performance TestReported Device Performance
    Biocompatibility1. Cytotoxicity (MEM Elution)Met requirements
    2. Sensitization (Guinea Pig Maximization)Met requirements
    3. Irritation (Intracutaneous Reactivity)Met requirements
    4. Acute Systemic ToxicityMet requirements
    5. Pyrogen (Materials Mediated in Rabbits)Met requirements
    6. Hemolysis (Direct Contact)Met requirements
    7. Hemolysis (Indirect Contact)Met requirements
    8. Complement ActivationMet requirements
    9. ThrombosisMet requirements
    Bench Performance1. RadiopacityMet requirements for visualization under fluoroscopy
    2. Corrosion ResistanceMet requirements
    3. Sheath and Dilator Visual InspectionMet requirements
    4. Sheath Dimension InspectionMet requirements (specific dimensions not provided, but implicitly comparable to predicate)
    5. Dilator Dimension InspectionMet requirements (specific dimensions not provided, but implicitly comparable to predicate)
    6. Sheath DeflectionMet requirements (Bi-Directional, 180° clockwise / 90° counterclockwise, similar to predicate K061363, an improvement over K140420)
    7. Sheath Curvature DimensionsMet requirements (specific dimensions not provided, but implicitly comparable to predicate for proper catheter positioning)
    8. Insertion and Deflection CyclingMet requirements for durability and function over repeated use
    9. Aspiration Air IngressMet requirements for minimizing air ingress
    10. Valve Leakage ResistanceMet requirements for minimizing blood loss and maintaining seal during catheter introduction/exchange
    11. Sheath Leakage ResistanceMet requirements for maintaining sheath integrity
    12. Device PreparationMet requirements (implies ease of preparation for use)
    13. Valve Integrity and FunctionalityMet requirements (implies effective hemostasis and function of 3-way stopcock)
    14. Shaft Torque StrengthMet requirements for effective steering and tip deflection response
    15. Kink ResistanceMet requirements for maintaining patency and facilitating catheter advancement
    16. Marker Band LocationMet requirements for accurate visualization and positioning under fluoroscopy
    17. Tensile StrengthMet requirements for material integrity and resistance to breakage during use
    Packaging Integrity1. Pouch Seal StrengthMet requirements for maintaining sterility and product integrity during storage and transport
    2. Gross Leak DetectionMet requirements for maintaining sterility and product integrity during storage and transport
    Shelf LifeVarious tests (details not explicitly itemized)Met requirements (data demonstrated maintenance of functional and performance characteristics over specified shelf life)
    SterilizationVarious tests (details not explicitly itemized)Met requirements for effective sterilization (EO gas)

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes used for each individual test. It states "representative sterilized samples of the device underwent sterilization, biocompatibility, bench testing, packaging integrity, and shelf life testing." This suggests a sufficient number of samples were tested to gain confidence in the results according to relevant standards. The data provenance is internal to Kalila Medical, Inc., as these are tests conducted by the manufacturer, or by laboratories contracted by the manufacturer, to support the 510(k) submission. These tests are inherently prospective in nature as they evaluate newly manufactured devices. There is no mention of country of origin of the data beyond the manufacturer's location in Campbell, California, USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical medical instrument, not an AI/ML diagnostic tool requiring expert interpretation for ground truth. Its performance is evaluated through engineering and biological testing methods.

    4. Adjudication method for the test set

    Not applicable for a physical device. Performance is evaluated against objective engineering specifications and biological standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI-powered device.

    7. The type of ground truth used

    For this medical device, the "ground truth" is established by:

    • Engineering specifications and industry standards: Performance characteristics (e.g., dimensions, deflection angles, tensile strength, leak resistance, radiopacity) are measured against predefined engineering tolerances and relevant international/consensus standards for catheter introducers.
    • Biocompatibility standards: Tests (e.g., cytotoxicity, sensitization, irritation) are conducted according to ISO 10993 series standards, where the "ground truth" is compliance with established biological safety thresholds.
    • Sterilization efficacy standards: Validation against established methods for ethylene oxide (EO) gas sterilization.
    • Packaging integrity standards: Evaluation against packaging standards to ensure sterility and physical protection.

    8. The sample size for the training set

    Not applicable. This is not an AI-powered device and therefore does not have a training set in the AI/ML sense.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI-powered device, there is no training set or associated ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140420
    Date Cleared
    2014-04-01

    (42 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vado Steerable Sheath is indicated for introducing various cardiovascular catheters into the vasculature and into the chambers of the heart including the left side of the heart through the interatrial septum.

    Device Description

    The Kalila Vado Steerable Sheath consists of an 8.8F (ID) sheath and dilator, which is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter exchange. A sideport with three-way stopcock is provided for aspiration, fluid infusion, blood sampling and pressure monitoring. A handle equipped with a deflection knob, deflects the tip 140°. The steerable introducer features distal vent holes for aspiration and radiopaque markers to facilitate visualization under fluoroscopy.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Kalila Medical Vado Steerable Sheath:

    Acceptance Criteria and Device Performance for Kalila Medical Vado Steerable Sheath (K140420)

    Based on the provided 510(k) summary, the device's acceptance criteria are derived from its functional and performance requirements, as demonstrated through a series of tests to establish substantial equivalence to the predicate device. The study primarily relies on non-clinical (bench and biocompatibility) testing, as no clinical testing was performed for this submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes categories of tests performed and states that the device "met its functional and performance characteristics in accordance with applicable industry standards and compares favorably to the predicate device." However, specific numerical acceptance criteria (e.g., minimum force, maximum leakage rate) and precise quantitative results are not explicitly detailed in the summary. The "reported device performance" is a general statement of compliance rather than specific measurements against a defined threshold.

    Category of TestSpecific Tests PerformedAcceptance Criteria (Conceptual)Reported Device Performance
    Biocompatibility1. Cytotoxicity
    1. Acute Systemic Toxicity
    2. Hemocompatibility (Direct and Indirect)
    3. Thrombosis
    4. Sensitization
    5. Irritation / Intracutaneous
    6. Complement Activation
    7. Pyrogenicity (rabbit pyrogen and bacterial endotoxin (LAL)) | No adverse biological reactions; compliance with relevant ISO standards for medical devices in contact with blood. | Met all biocompatibility requirements. |
      | Bench Tests | 1. Steerable Sheath and Dilator Visual Inspection
    8. Steerable Sheath Dimension Inspection
    9. Steerable Sheath Marker Band Location Measurement
    10. Dilator Dimensional Inspection
    11. Valve Leakage Resistance at 40kPa
    12. Leakage Resistance at 300kPa
    13. Device Preparation
    14. Dilator Snap Disengagement Force
    15. Dilator Insertion and Retraction Force
    16. Shaft Deflection
    17. Curvature Dimensions
    18. Aspiration
    19. Catheter Insertion Cycling and Flexion Cycling, with Flush
    20. Bend Radius to Kink
    21. Corrosion Resistance
    22. Junction Strengths
    23. Torque and Turns to Failure
    24. Steerable Sheath Radiopacity Study | Device conforms to design specifications; functions as intended; withstands specified pressures; forces within acceptable ranges; maintains integrity under cycling/stress; visible under fluoroscopy. | Met all functional and performance requirements; comparable to predicate device. |
      | Packaging Integrity | 1. Pouch Seal Strength
    25. Gross Leak Detection | Packaging maintains sterility and integrity of the device until use. | Met all packaging integrity requirements. |
      | Shelf Life | (Implicitly part of overall testing) | Device maintains performance and integrity over its stated shelf life. | (No specific details, but implied compliance) |
      | Sterilization | (Implicitly part of overall testing) | Device is adequately sterilized and maintains sterility. | (No specific details, but implied compliance) |

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "representative sterilized samples of the device underwent sterilization, biocompatibility, bench testing, packaging integrity, and shelf life testing." However, no specific sample sizes for any of these tests are provided or mentioned in the summary.

    Data Provenance: The data is entirely non-clinical (bench and laboratory-based). There is no information regarding the country of origin of the data, as it pertains to internal testing conducted by Kalila Medical to support market clearance. The data is prospective in the sense that the tests were designed and executed specifically for this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the study did not involve human interpretation or a "test set" in the context of image analysis or diagnostic performance. The "ground truth" here is objective measurement against engineering specifications and industry standards.

    4. Adjudication Method for the Test Set

    Not applicable. As the study comprised bench and biocompatibility testing, there was no need for an adjudication method as would be used in expert consensus for clinical data. Performance was determined by direct measurement and observation against established criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a MRMC comparative effectiveness study was NOT done. The submission explicitly states: "Clinical testing is not provided in this submission." Therefore, there is no information on how human readers would improve with or without AI assistance, as AI is not a component of this device, and no human-reader study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    Not applicable. The Vado Steerable Sheath is a physical medical device (a catheter introducer), not an algorithm or an AI-powered system. Therefore, standalone algorithm performance is not relevant.

    7. The Type of Ground Truth Used

    The ground truth used for this submission is based on:

    • Engineering specifications and design requirements: For dimensional accuracy, deflection, forces, etc.
    • Industry standards (e.g., ISO for biocompatibility): For biological safety and material compatibility.
    • Predicate device characteristics: For establishing substantial equivalence in performance and functionality.

    Essentially, the ground truth is objective, measurable performance against predefined and accepted engineering and safety standards.

    8. The Sample Size for the Training Set

    Not applicable. This submission is for a physical medical device, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant. The ground truth for the device's development would have been established through a combination of engineering design, materials science, and functional requirements based on the intended use and predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1