Intended Use

The Vado Steerable Sheath is indicated for introducing various catheters into the vasculature and into the chambers of the heart including the left side of the heart through the interatrial septum.

Device Description

The Vado Steerable Sheath consists of an 8.8F (ID) sheath and dilator, which is designed to provide flexible catheter positioning in the cardiac anatomy. The steerable introducer includes a hemostasis valve to minimize blood loss during catheter exchange. A sideport with three-way stopcock is provided for aspiration, fluid infusion, blood sampling and pressure monitoring. A handle equipped with a deflection knob, deflects the sheath tip. The steerable introducer features distal vent holes for aspiration and radiopaque markers to facilitate visualization under fluoroscopy.

AI/ML Overview

The provided text describes the regulatory clearance of a medical device, the "Vado Steerable Sheath," and details the non-clinical testing performed to demonstrate its substantial equivalence to a predicate device. It does not describe an AI/ML powered device, nor does it contain information about acceptance criteria or supporting studies that are typically associated with performance metrics of AI/ML models (e.g., sensitivity, specificity, AUC).

Therefore, I cannot provide the information requested in the prompt, as the document does not contain:

  1. A table of acceptance criteria and reported device performance in the context of AI/ML metrics.
  2. Sample size used for the test set or data provenance for an AI/ML model.
  3. Number of experts or their qualifications for establishing ground truth for AI/ML validation.
  4. Adjudication method for AI/ML ground truth establishment.
  5. Information on a multi-reader, multi-case (MRMC) comparative effectiveness study for AI assistance.
  6. Standalone performance (algorithm only) for an AI/ML model.
  7. Type of ground truth used (expert consensus, pathology, outcomes data) for AI/ML validation.
  8. Sample size for the training set of an AI/ML model.
  9. How the ground truth for the training set was established for an AI/ML model.

The document focuses on the mechanical and material performance of a medical device (a catheter introducer), including:

  • Biocompatibility tests: To ensure the device is safe for human contact.
  • Bench performance tests: To assess physical characteristics like dimensions, leakage resistance, deflection, aspiration, and strength.
  • Packaging integrity tests: To ensure sterility and product protection.

These tests are standard for a traditional medical device and are aimed at demonstrating that the device meets its functional and performance characteristics and is equivalent to the predicate device in terms of safety and effectiveness, not that it performs an interpretive or diagnostic function with AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 7, 2017

Kalila Medical, Inc. Carrie Neuberger Regulatory Affairs 745 Camden Ave., Suite A Campbell CA, 95008

Re: K160721

Trade/Device Name: Vado Steerable Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: January 29, 2017 Received: January 31, 2017

Dear Carrie Neuberger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando Aguel Fernando Aguel -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Change Control Table, Change History

Change Control Table

VersionDocument AuthorDocument ApproverDate Approved
1.00Name, Title, OfficeName, Title, OfficeMM/DD/YYYY

Complete Change Control Table (all versions) retained in SWIFT Docs.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Kalila Medical. The logo consists of an orange and gray abstract shape resembling two intertwined letter 'S' shapes, followed by the text "Kalila Medical" in a combination of orange and gray colors. The word "Kalila" is in orange, while "Medical" is in gray.

Indications for Use

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K160721

Device Name

Vado Steerable Sheath

Indications for Use (Describe)

The Vado Steerable Sheath is indicated for introducing various catheters into the vasculature and into the chambers of the heart including the left side of the heart through the interatrial septum.

Type of Use (Select one or both, as applicable)

Z Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Kalila Medical. The logo consists of an orange and black abstract shape on the left, followed by the words "Kalila Medical" in orange and gray. The word "Kalila" is in orange, while the word "Medical" is in gray.

Section 5510(k) Summary
510(k) NumberK160721
Submitter Name and Address
Name:Kalila Medical, Inc.
Address:745 Camden Avenue, Suite A
Campbell, CA 95008
Telephone:408-637-2150
Fax:408-370-6325
Contact:Carrie Neuberger
Date Prepared:June 13, 2016
General Device Information
Device Name:Vado Steerable Sheath
Common Name:Catheter introducer
Classification:21 CFR 870.1340A catheter introducer is a sheath used tofacilitate placing a catheter through the skininto a vein or artery.
Device Class:Class II
Product Code:DYB
Predicate Device
ManufacturerDevice Name510(k) Number
Kalila MedicalVado Steerable SheathK140420
Device Description
The Vado Steerable Sheath consists of an 8.8F (ID) sheath and dilator, which isdesigned to provide flexible catheter positioning in the cardiac anatomy. Thesteerable introducer includes a hemostasis valve to minimize blood loss duringcatheter exchange. A sideport with three-way stopcock is provided for aspiration,fluid infusion, blood sampling and pressure monitoring. A handle equipped with adeflection knob, deflects the sheath tip. The steerable introducer features distal ventholes for aspiration and radiopaque markers to facilitate visualization underfluoroscopy.
Intended Use (Indications)
The Vado Steerable Sheath is indicated for introducing various cardiovascularcatheters into the vasculature and into the chambers of the heart including theleft side of the heart through the interatrial septum.
Comparison to the Predicate Device
The Vado Steerable Sheath has the same intended use and fundamentalscientific technology as the predicate device, including biocompatibility,packaging, sterilization, and labeling. Where dimensional differences existbetween the subject device and the predicate device, performance testingdemonstrates that these differences do not adversely affect safety and

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Kalila Medical. The logo consists of an orange and black abstract shape resembling a stylized "S" or a winding path, followed by the words "Kalila" in orange and "Medical" in dark gray. The text is positioned to the right of the abstract shape.

effectiveness.

This submission supports the position the modified Vado Steerable Sheath, FG1_007, is substantially equivalent to the Vado Steerable Sheath, FG1_005 (K140420).

Summary of Non-Clinical and Clinical Testing

The 510(k) notice contains summaries of bench performance tests, packaging performance tests, and biocompatibility studies to evaluate the performance characteristics of the modified Vado Steerable Sheath, FG1_007. The data presented demonstrate that the modified Vado Steerable Sheath meets its functional and performance characteristics in accordance with applicable industry standards and is equivalent to the predicate device.

To verify that the modified Vado Steerable Sheath met its functional and performance requirements, representative sterilized samples of the device underwent sterilization, biocompatibility, bench testing, packaging integrity, and shelf life testing.

GLP Biocompatibility Tests:

    1. Cytotoxicity (MEM Elution)
    1. Sensitization (Guinea Pig Maximization Sensitization)
    1. Irritation (Intracutaneous Reactivity Irritation in Rabbits)
    1. Acute Systemic Toxicity
    1. Pyrogen (Materials Mediated in Rabbits)
    1. Hemolysis (Direct Contact)
    1. Hemolysis (Indirect Contact)

Bench Performance Tests:

    1. Steerable Sheath and Dilator Visual Inspection
    1. Steerable Sheath Dimension Inspection
    1. Steerable Sheath Marker Band Location
    1. Dilator Dimensional Inspection
    1. Valve Leakage Resistance at 40kPa
    1. Leakage Resistance at 300kPa
    1. Device Preparation
    1. Dilator Snap Disengagement
    1. Dilator Insertion and Retraction
    1. Shaft Deflection
    1. Curvature Dimensions
    1. Aspiration
    1. Catheter Insertion Cycling and Flexion Cycling
    1. Bend Radius to Kink
    1. Corrosion Resistance
    1. Junction Strengths
    1. Torque and Turns to Failure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Kalila Medical. The logo consists of an orange square with a stylized "S" shape inside, with the text "Kalila Medical" to the right of the square. The word "Kalila" is in orange, while the word "Medical" is in gray.

  1. Radiopacity

Packaging Integrity Tests:

    1. Pouch Seal Strength
    1. Gross Leak Detection

Pre-Clinical testing is not provided in this submission.

Clinical testing is not provided in this submission.

The fundamental scientific technology and technological characteristics of the modified Vado Steerable Sheath are the same as the Vado Steerable Sheath (K140420) predicate device including mechanism of action, packaging, biocompatibility, sterilization, and labeling. Through bench performance testing it was demonstrated that the modified Vado Steerable Sheath is as safe, and is as effective as the predicate device, and does not raise any new questions of safety and effectiveness.

Statement of Equivalence

The Vado Steerable Sheath has the same indications for use and technological characteristics as the predicate device. Based on this and the data provided in this pre-market notification, the subject device and the predicate device have been shown to be substantially equivalent.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).