Search Results
Found 9 results
510(k) Data Aggregation
(439 days)
JKH Health Co., Ltd.
Apply stationary use of ultrasound to: Generate deep heat within body tissues for the treatment of selected medical conditions such as the relief of pain, the relief of muscle spasms, the treatment of joint contractures, and the local increase in circulation.
Apply continuous movement of ultrasound for:
- Pain.
- Pain relief, muscle spasms, and joint contractures.
- Relief of pain, muscle spasms, and joint contractures that may be associated with:
- Adhesive capsulitis,
- Bursitis with slight calcification,
- Myositis,
- Soft tissue injuries, and
- Shortened tendons due to past injuries and scar tissues.
- Relief of pain, muscle spasms, and joint contractures resulting from:
- Capsular tightness, and
- Capsular scarring.
- Localized increase in blood flow.
- Increased range of motion of contracted joint using heat and stretch techniques.
As a portable and rechargeable prescriptive device in the home or clinical/hospital setting, the subject device is intended to apply ultrasound to the patient's treatment site. It is demonstrated ultrasound can be used to apply therapeutic deep heat for the treatment of medical conditions including relief of pain, muscle spasms, and joint contractures.
It is intended to be used for 30 minutes per cycle (up to 4 cycles) in the home or clinical/hospital setting by or under the direction of a medical professional to apply ultrasound to the patient's treatment site. The device automatically alerts the patient in case of improper application or performance. All the control components and the rechargeable battery of the device are protectively housed, and the device is connected to one or two applicators of ultrasound.
The subject device consists of the main device, the applicator, the charger/adapter, the liquid and/or solid ultrasound gel, and the sticky pad. For the best result, use the ultrasound gel and the sticky pad supplied.
This looks like a fascinating case study. Based on the provided FDA 510(k) clearance letter and summary, here's an analysis of the acceptance criteria and the study that proves the device meets them.
It's important to note that this document is for an Ultrasound Stimulator (Ultrasonic Diathermy), which delivers therapeutic ultrasound, not an AI/Software as a Medical Device (SaMD) that typically involves image analysis or diagnostic support. Therefore, many of the typical acceptance criteria and study designs related to diagnostic accuracy, MRMC studies, and nuanced ground truth establishment for AI models (as implied by the original request's detailed points) are not applicable to this type of physical therapy device.
The "acceptance criteria" here are primarily based on demonstrating substantial equivalence to predicate devices through engineering specifications, performance testing, and safety profiles, rather than clinical efficacy studies in the way an AI diagnostic tool would require.
Acceptance Criteria and Study Proving Device Meets Criteria
Given that the device is an Ultrasound Stimulator aiming for 510(k) clearance based on substantial equivalence, the "acceptance criteria" are not framed in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) but rather in terms of demonstrating that its technical characteristics, performance, and safety are equivalent to legally marketed predicate devices. The "study" proving it meets these criteria is a comprehensive set of non-clinical, bench performance tests and comparisons.
1. Table of Acceptance Criteria and Reported Device Performance
For an Ultrasound Stimulator, the "acceptance criteria" are primarily a demonstration of equivalence to the predicate devices across various technical and functional parameters, and compliance with relevant safety standards. The "reported device performance" are the results of the non-clinical tests and direct comparisons to the predicate devices.
Parameter/Acceptance Criteria Category | Specific Acceptance Criteria (Demonstrated Equivalence/Compliance) | Reported Device Performance (Subject Device) |
---|---|---|
Intended Use | Identical or substantially equivalent indications to predicate device. | Apply stationary/continuous ultrasound for deep heat, pain relief, muscle spasms, joint contractures, increased circulation, range of motion (identical to primary predicate, similar to other predicate). |
Product Code | Identical to predicate device. | IMI, PFW (Identical to predicate ManaSport+, and PFW to sam 2.0) |
Prescription Status | Identical (Prescription Use). | Prescription Use (Identical). |
Power Source | Similar, safe power delivery. | 100~240 Vac with 5Vdc Input and rechargeable 3.7Vdc battery (Identical to primary predicate, similar to other predicate). |
Number of Outputs | 1 or 2 outputs. | 1 or 2 (Identical to sam 2.0, similar to ManaSport+ which is 1). |
Software/Microprocessor Control | Presence and safe operation. | Yes (Identical). |
Safety Features | Automatic No-Load Trip, Automatic Shut Off, User Override Control, Indicator Display (On/Off, Low Battery). | All 'Yes' for these features (Identical where predicate information available). |
Timer Range | Functionally acceptable and safe timer range. | 30 minutes (per cycle), max 4 cycles (Longer than primary predicate (20min), shorter than predicate (240min), deemed safe by temperature control). |
Compliance with Voluntary Standards | Compliance with relevant IEC/ISO electrical safety, EM compatibility, and biocompatibility standards. | Yes (Complies with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-5, ISO 10993-5, ISO 10993-10, IEC 60601-1-6, IEC 60601-1-11). |
Biocompatibility | Non-toxic, non-irritating materials. | Yes (Identical). |
Sterility | Non-sterile device. | Non-Sterile (Identical). |
Housing Construction Material | Compatible material. | ABS (Identical). |
Physical Dimensions/Weight | Similar physical characteristics not affecting safety/effectiveness. | Console/Generator: 138×62×22 mm, 0.18 kg (Similar to predicates). Treatment Head: 61 x 45 x 12 mm, 0.047 kg (Similar to predicates). |
Functions and Design | Ultrasound generation. | Ultrasound (Identical). |
Frequency | 1.5 MHz or similar. | 1.5 MHz ± 10% (Identical to primary predicate, similar to other predicate's 3MHz). |
Leakage Current | Below maximum limit. |
Ask a specific question about this device
(21 days)
JKH Health Co., Ltd.
Cold Compression is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
Cold Compression is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.
The subject device is an AC powered, software-controlled multimodality device, and is intended to be used by or on the order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings.
The subject device features cold therapy and compression therapy. Cold Compression is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated.
The device has a power/time button, temperature adjustment button, pressure setting button; Patients can adjust the temperature, compression pressure, and duration of treatment according to their needs.
The device has a bucket body and a lid. The ice and water can be added by opening the lid, and the loading level of the ice and water can be indicated/seen.
The subject device and its packaging/clothing are clean and non-sterile. A connecting tube is used to connect the subject device to the wraps come with a variety of reusable cold treatment inflators for cold and compression treatment of the universal, back, shoulder, ankle/foot, hip, and knee.
The provided text is a 510(k) Summary for a medical device called "Cold Compression". This document aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than proving the device meets specific acceptance criteria based on a study of its performance in a clinical or analytical setting.
Therefore, the input document does not contain the information requested regarding:
- A table of acceptance criteria and reported device performance.
- Sample size used for a test set and data provenance.
- Number of experts and their qualifications for establishing ground truth.
- Adjudication method for a test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
- Standalone (algorithm only) performance study.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document states that "Clinical testing was not needed in support of this 510(k) application." (Section 8) and focuses on demonstrating substantial equivalence through comparison with predicate devices and non-clinical tests for electrical safety and electromagnetic compatibility.
Instead, the document provides:
Non-Clinical Tests Performed:
- Tests: Non-clinical tests were performed to validate the design and assure conformance with voluntary design standards related to medical device electrical safety and electromagnetic compatibility.
- Standards:
- IEC 60601-1 "Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance (IEC 60601-1)".
- IEC 60601-1-2 "Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral standard: Electromagnetic Compatibility - Requirements and Tests".
Comparison Table (Table 1):
This table compares the subject device ("Cold Compression") with two predicate devices (K223541 and K183702) across various technical characteristics and specifications to establish substantial equivalence. While it doesn't present "acceptance criteria" in the traditional sense of a performance study, it details the characteristics where the device's performance parameters are deemed acceptable due to their similarity to a legally marketed device.
Characteristic | Subject Device Performance | Predicate Device Performance (relevant to equivalence) | Equivalence Conclusion |
---|---|---|---|
Product Codes | IRP and ILO | Primary Predicate: IRP, ILO, and JOW | |
Predicate: IRP and ILO | The subject device has the product codes covered by the primary predicate device, and also identical to the predicate device. | ||
Indications for Use | To treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. Used by or on order of licensed healthcare professionals in rehabilitation facilities/hospitals, outpatient clinics, athletic training settings, and home settings. | Primary Predicate: Combines cold, heat, contrast, and compression therapy. Intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain. Also provides DVT therapy. Used by or on order of licensed healthcare professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings. | |
Predicate: To treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. Used by or on order of licensed healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings. | The subject device has the Indications for Use covered by the primary predicate device, and also substantially equivalent to the predicate device. | ||
Intended Users | Health Care Professionals and lay users (under prescription), with adult use; children and elderly under direct medical professional supervision. | Primary Predicate and Predicate: Identical. | Identical |
Prescription or OTC | Prescription | Primary Predicate and Predicate: Prescription | Identical |
Power Source(s) | 100-240 VAC, 50/60 Hz | Primary Predicate and Predicate: 100-240 VAC, 50/60 Hz | Identical |
Compliance with Voluntary Standards | Yes | Primary Predicate and Predicate: Yes | Identical |
Safety (Electrical, Mechanical, Chemical, Thermal, Radiation) | Yes | Primary Predicate and Predicate: Yes | Identical |
Therapy | Cold and Compression. Work together or independently. | Primary Predicate: Cold, Heat, Contrast and Compression. | |
Predicate: Cold and Compression. Work together or independently. | The subject device has the therapy covered by the primary predicate device, and also identical to the predicate device. The difference between the subject device and the primary predicate device is the former does not provide heat or contrast therapy. | ||
Cold Therapy Temperature | 45°F - 55°F | Primary Predicate: Default: 50°F; Custom: 41°F-55°F | |
Predicate: 45°F - 60°F | Equivalent Note 1: The temperature range of cold therapy for the subject device is within that of the primary predicate device and the predicate device. Adjustable by healthcare professionals. Insignificant difference. | ||
Heat Therapy | Not Available | Primary Predicate: Default: 105°F; Custom: 105°F-109°F | |
Predicate: Not Available | Identical to the predicate device. The difference from the primary predicate device (lack of heat/contrast therapy) is noted but does not raise new safety/effectiveness issues for this specific device. | ||
Compression Setting | 0 - 60 mmHg | Primary Predicate: Alternate mode: 15 -75 mm Hg; Continuous mode: 15 -75 mm Hg | |
Predicate: Low (0-25 mmHg), Regular (0-50 mmHg) | Equivalent Note 2: The pressure range for the subject device is within that of the primary predicate device and slightly higher than the predicate device. Adjustable by healthcare professionals. Insignificant difference. | ||
Reservoir Fluid Capacity | 620 mL | Primary Predicate: 350 mL | |
Predicate: Not Available | Different, but it will not raise any new issue of safety or effectiveness. | ||
Recommended Coolant | Tap Water | Primary Predicate: 90% Distilled Water, 10% Isopropyl Alcohol | |
Predicate: Tap Water | Identical to the predicate device. The difference from the primary predicate is noted but does not raise new safety/effectiveness issues. | ||
Water Cooling Source | Ice | Primary Predicate: Vapor compression | |
Predicate: Ice | Identical to the predicate device. The difference from the primary predicate is noted but does not raise new safety/effectiveness issues. | ||
User Interface | Keypad with indicator lights or LCD screen | Primary Predicate: Touch Screen | |
Predicate: Keypad with indicator lights | Different from the primary predicate device and similar to the predicate device. It will not raise any new issue of safety or effectiveness. | ||
Dimensions | L235xW235xH280mm | Primary Predicate: L295xW285xH295 mm | |
Predicate: Not Available | Different. The difference of dimensions does not change the product performance or parameters, which will not raise any new issue of the safety or effectiveness. | ||
Weight Approx. | 2.4kg | Primary Predicate: 8.2kg | |
Predicate: Not Available | Different. The difference of weight does not change the product performance or parameters, which will not raise any new issue of the safety or effectiveness. | ||
Types of Garments | Various anatomical thermal garments for: Back, Elbow/Universal, Shoulder, Knee, Ankle, Hip. | Primary Predicate: Various anatomical thermal garments for: Back, Elbow, Shoulder, Knee, Ankle, Hip. DVT Garments: Calf and Foot. | |
Predicate: Various anatomical thermal garments for: Back, Universal, Shoulder, Knee, Foot/Ankle, Hip. | The subject device has the garments covered by the primary predicate device, and also substantially equivalent to the predicate device. | ||
Biocompatibility of Patient Contacting Material | Biocompatible | Primary Predicate and Predicate: Biocompatible | Identical |
Sterile/Non-Sterile | Non-sterile | Primary Predicate and Predicate: Non-sterile | Identical |
Cleaning Disinfection Validation of Labeling | Yes | Primary Predicate and Predicate: Yes | Identical |
Operational and Environmental Conditions | Normal working ambient temperature: 5 | Primary Predicate: Normal working ambient temperature: 16 | |
Predicate: Normal working ambient temperature: 5 | Identical to the predicate device. The difference between the subject device and the primary predicate device is the former does not provide heat or contrast therapy. (Note: The operational and environmental conditions are listed as identical to the second predicate device, not the primary one, despite the "difference between the subject device and the primary predicate device" wording). |
Ask a specific question about this device
(52 days)
JKH Health Co., Ltd.
The Lymphedema Compression is comprised of a gradient compression sleeve and a portable intermittent pump to provide graduated compression in both sustained and intermittent settings for use in both the hospital and outpatient setting. It is pre-set to the default setting of 50 mmHg that cannot be adjusted or can be adjusted by the physician to a pressure within the specified range. It is intended for use in:
- Treatment of lymphedema
- Treatment of chronic venous insufficiency
- Treatment and promotion of healing of stasis dermatitis and venous stasis ulcers
- Reducing venous leg ulcer healing time
- Reducing edema due to venous stasis
- Enhancing venous return
The device is intended for home, and hospital use.
As a portable and rechargeable prescriptive device, the Lymphedema Compression is intended to be used in the home or clinical/hospital setting by or under the direction of a medical professional. It applies pressure to treat lymphedema and other edematous conditions.
Supplied clean and non-sterile, the Lymphedema Compression utilizes the pneumatically controlled cuff and is actuated by an electronically controlled air pump unit. All pump, battery and control components are protectively housed in a plastic case of the control unit, which is attached to an inflatable cuff/sleeve. The Lymphedema Compression contains an ON/OFF button, a SET button, and a display for user interface. In addition, there is also a port on the control unit for connecting the battery charger/AC adapter plug. The cuff component consists of multiple air chambers/bladders encased inside a soft nylon fabric or equivalent biocompatible fabric for increased patient comfort and biocompatibility compliance.
In operation, the user simply press the ON/OFF button to turn on the control unit, to which a cuff containing air chambers/bladders is connected. And the control unit then inflates the cuff to the default pre-determined pressure (50 mmHg) that cannot be adjusted to a pressure within the specified range via the SET button. Once the cuff pressure of the multiple air chambers/bladders reaches the pre-determined or adjusted level, the pump is turned off, and the cuff deflates to the ambient pressure through a valve inside the control unit. After the rest period, the cycle of inflation and deflation repeats until the device is turned off automatically or manually.
The provided FDA 510(k) summary (K240011) describes the "Lymphedema Compression" device and its substantial equivalence to a predicate device, ManaFlow (K200353).
1. A table of acceptance criteria and the reported device performance
The provided document doesn't explicitly list "acceptance criteria" in the typical sense of numerical thresholds for performance metrics. Instead, the substantial equivalence relies on demonstrating that the subject device's characteristics and performance are identical to or similar to the predicate device. The primary "acceptance criterion" is functional and technical equivalence to the predicate.
Here's a table based on the comparison provided in the 510(k) summary:
Feature/Criterion | Acceptance Criteria (Based on Predicate Device: ManaFlow K200353) | Reported Subject Device Performance (Lymphedema Compression) | Equivalence |
---|---|---|---|
Intended Use | To provide graduated compression in both sustained and intermittent settings for: Treatment of lymphedema, chronic venous insufficiency, stasis dermatitis/venous stasis ulcers (healing & reducing healing time), reducing edema due to venous stasis, enhancing venous return. For home and hospital use. | Identical to Predicate: To provide graduated compression in both sustained and intermittent settings for: Treatment of lymphedema, chronic venous insufficiency, stasis dermatitis/venous stasis ulcers (healing & reducing healing time), reducing edema due to venous stasis, enhancing venous return. For home and hospital use. | Identical |
Pressure Setting | Pre-set to 50 mmHg (cannot be adjusted) or adjustable (20-80 mmHg) by physician. | Pre-set to 50 mmHg (cannot be adjusted) or adjustable (20-80 mmHg) by physician. | Identical |
Prescription/OTC | Prescription | Prescription | Identical |
Power Source | 5V DC supply (100-240 VAC input) and 3.7V rechargeable battery | 5V DC supply (100-240 VAC input) and 3.7V rechargeable battery | Identical |
Battery Charge Time | Approx. 4 hours from depleted state | Approx. 4 hours from depleted state | Identical |
Power Supply (Input/Output) | Input: 100-240 Vac, 50-60 Hz; Output: 5 Vdc @ 2 Amp | Input: 100-240 Vac, 50-60 Hz; Output: 5 Vdc @ 2 Amp | Identical |
Internal Rechargable Batteries | Yes | Yes | Identical |
Compliance with Voluntary Standards | Yes (Electrical Safety, Mechanical Safety, Chemical Safety, Thermal Safety, Radiation Safety) | Yes (ANSI AAMI ES60601-1, IEC 60601-1-2) | Identical |
Functions and Design | Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities. | Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities. | Identical |
Contraindications | MUST NOT be used for: Suspected DVT, ischemic vascular disease, severe arteriosclerosis, pulmonary edema, severe congestive heart failure, thrombophlebitis, active infection; on legs interfering with: vein ligation, gangrene, dermatitis, open wounds, recent skin graft, massive edema, extreme deformity; on any neuropathy; on insentitive extremities; where increased venous/lymphatic return is undesirable. | Identical to Predicate. | Identical |
Target Population / Users | Patients needing venous return and lymphedema treatment | Patients needing venous return and lymphedema treatment | Identical |
Where Used | Home, Hospital, Surgery Center, Altitude travel, areas of limited mobility | Home, Hospital, Surgery Center, Altitude travel, areas of limited mobility | Identical |
Application | Non-invasive / external | Non-invasive / external | Identical |
Portability | Portable, ambulant | Portable, ambulant | Identical |
Basis of operation | Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities. | Aids venous return by cyclic, intermittent, pneumatic pressure application (inflation-deflation) to compress extremities. | Identical |
Anatomical Site | Leg (for predicate) | Extremities | Identical or Similar |
System management | Microprocessor | Microprocessor | Identical |
Pressure Source | Micro pump controlled by microprocessor | Micro pump controlled by microprocessor | Identical |
Operating Modes | Preset and adjustable modes | Preset and adjustable modes | Identical |
Working Pressure | Preset at 50 mmHg and adjustable from 20-80 mmHg | Preset at 50 mmHg and adjustable from 20-80 mmHg | Identical |
Cycle Time | Inflation in sequence (foot-up), then deflation; repeats. | Inflation in sequence (foot-up), then deflation; repeats. | Identical |
System diagnostics | Audible and visual alarms for system faults | Audible and visual alarms for system faults | Identical |
Air delivery | Via flexible plastic tube(s) connected directly to air bladder. | Via flexible plastic tube(s) connected directly to air bladder. | Identical |
Sterility | Clean / non-sterile | Clean / non-sterile | Identical |
Leg cuff usage | Single Patient Use | Single Patient Use | Identical |
Material Used | Air bladder chambers in soft, non-latex medical fabric or equivalent for comfort/biocompatibility. | Air bladder chambers in soft, non-latex medical fabric or equivalent for comfort/biocompatibility. | Identical |
Fastening | Snap and screw | Snap and screw | Identical |
Biocompatibility | Biocompatible | Biocompatible | Identical |
Software | Moderate | Moderate | Identical |
Operating Temperature | +10 ℃ (50 ºF) to +40 ºC (104 ºF) | +10 ℃ (50 ºF) to +40 ºC (104 ºF) | Identical |
Humidity | 30%-75% | 30%-75% | Identical |
Cleaning and Disinfecting | Specific instructions for pump unit and cuffs; air dry. | Specific instructions for pump unit and cuffs; air dry. | Identical |
Disposal | Electromechanical device with PCBs and rechargeable batteries; not for landfill; consult local requirements/recycle center. | Electromechanical device with PCBs and rechargeable batteries; not for landfill; consult local requirements/recycle center. | Identical |
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical testing was not needed in support of this 510(k) application."
Therefore, there was no clinical test set of patient data, no sample size of patients, and no data provenance relevant to clinical performance evaluation. The "test set" for the device's technical specifications consists of the individual components and the assembled device that underwent non-clinical bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Since no clinical testing was performed, there was no ground truth established by experts on patient cases. The evaluation of electrical safety, EMC, and physical/performance requirements relies on adherence to published voluntary standards and internal company specifications, not expert consensus on clinical outcomes.
4. Adjudication method for the test set
Not applicable. As no clinical testing or human reader study was conducted, there was no need for an adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was conducted. This device is a pneumatic compression device, not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a physical medical device (pneumatic compression) and does not involve an algorithm with standalone performance in the context of diagnostic interpretation. Its "performance" is based on mechanical and electrical parameters (e.g., pressure output, cycle time, safety features).
7. The type of ground truth used
The "ground truth" for this submission is adherence to recognized voluntary standards for medical device safety and performance (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2) and the technical specifications and functional characteristics of the legally marketed predicate device. Bench tests validated that the subject device met its own design specifications, which were shown to be identical or very similar to the predicate.
8. The sample size for the training set
Not applicable. This device is not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable. (See #8)
Ask a specific question about this device
(336 days)
JKH Health Co., Ltd.
Cold/Hot Compression combines cold, heat, contrast, and compression therapy. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated.
Cold/Hot Compression also provide DVT therapy. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
ColdHot Compression is intended to be used by, or on the order of, licensed health care professionals in rehabilitation facilities, outpatient clinics, athletic training settings, and home settings.
The subject device is a prescriptive device, which combines cold/hot therapy and air compression. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain for which cold and compression are indicated. It is intended to treat post traumatic and post-surgical medical and/or surgical conditions for which localized thermal therapy (hot or cold) are indicated. It is intended to reduce the risk of the formation of deep venous thrombosis (DVT) by aiding blood flow back to the heart via lower extremity limb compression.
The subject device comes with the thermal-therapy wraps and the DVT wraps. The thermal-therapy wrap is nylon fabric on one side and velvet cloth on the other side; the DVT wrap is velvet on both sides. To avoid any potential adverse skin reactions such as redness, irritation, and cold/hot injury, the sock/clothing should be worn by the patient prior to use.
The subject device includes a main device and an optional DVT device, and has a limited shelf life of 3 years, based on the charge retention characteristics of the device's battery.
This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with acceptance criteria and performance metrics in the typical sense.
Therefore, the information requested might not be fully available or explicitly stated in the provided text as it would be in a clinical study report.
Based on the provided text, here's what can be extracted and inferred:
1. Table of acceptance criteria and the reported device performance
The concept of "acceptance criteria" here is primarily tied to demonstrating equivalence to the predicate device. The performance is assessed by comparing technical specifications.
Feature | Predicate Device Performance / Range (Therm-X) | Subject Device Performance / Range (Cold/Hot Compression) | Equivalence Note |
---|---|---|---|
Cold Therapy (Default) | 34°F, 45°F, 55°F | 50°F | Equivalent (Note 1) |
Cold Therapy (Custom) | 34°F - 55°F | 41°F-55°F | Equivalent (Note 1) |
Cold Therapy (Lowest) | 34°F or 40°F | 41°F | Equivalent (Note 1) |
Heat Therapy (Default) | 105°F, 107°F, 110°F | 105°F | Equivalent (Note 2) |
Heat Therapy (Custom) | 105°F - 110°F | 105°F-109°F | Equivalent (Note 2) |
Heat Therapy (Highest) | 110°F | 109°F | Equivalent (Note 2) |
Edema Pressure (Lowest) | 5 mm Hg or 20 mmHg | 15 mm Hg | Equivalent (Note 3) |
Edema Pressure (Highest) | 70 mmHg | 75 mm Hg | Equivalent (Note 3) |
DVT Pressure (Calf) | 50-70 mmHg | 55 mmHg | Equivalent (Note 4) |
DVT Pressure (Foot) | 90 - 130 mmHg | 130 mmHg | Equivalent (Note 4) |
Cycle Length (Longest) | 40 or 60 minutes | 60 minutes | Equivalent (Note 5) |
Contrast Therapy (Heat) | 105°F | 105°F | Equivalent (Note 6) |
Contrast Therapy (Cold) | 38°F | 49°F | Equivalent (Note 6) |
Contrast Cycle (Heat) | 3-10 minutes | 10 minutes | Equivalent (Note 7) |
Contrast Cycle (Cold) | 3-10 minutes | 20 minutes | Equivalent (Note 7) |
Reservoir Fluid Capacity | 650 mL | 350 mL | Equivalent (Note 8) |
Notes on Equivalence (as explained in the document):
- Note 1 (Cold Therapy): The subject device's cold therapy temperatures are within or similar to the range of the predicate/reference devices, and are adjustable by professionals. Differences are considered insignificant and do not raise new safety/effectiveness concerns.
- Note 2 (Heat Therapy): Similar to cold therapy, the heat therapy temperatures are within or similar to the predicate/reference devices and adjustable. Differences are insignificant.
- Note 3 (Edema Pressure): The subject device's lowest pressure is within the predicate's range and similar to the reference. The highest pressure is slightly higher but still adjustable by professionals. Differences are insignificant.
- Note 4 (DVT Pressure): The subject device's DVT pressures for calf and foot fall within or are slightly higher than the predicate/reference devices. Differences are insignificant.
- Note 5 (Cycle Length): The subject device's longest cycle length is similar to or longer than the predicate's, though shorter than the reference. Cycle length is adjustable. Differences are insignificant.
- Note 6 (Contrast Therapy Temp): The subject device's contrast therapy temperatures are either similar to the predicate or identical to the reference. No new safety/effectiveness issues.
- Note 7 (Contrast Cycle Length): The contrast therapy cycle length is slightly different from the predicate but identical to the reference. No new safety/effectiveness issues.
- Note 8 (Reservoir Fluid Capacity): The subject device's capacity is smaller than the predicate's but larger than the reference's. This difference does not raise new safety/effectiveness issues.
2. Sample size used for the test set and the data provenance
The document explicitly states: "Clinical testing was not needed in support of this 510(k) application."
Therefore, there is no test set in the sense of patient data from a clinical study. The "testing" involved non-clinical bench tests and comparison of specifications to predicate devices. The data provenance is from internal JKH Health Co., Ltd. testing and comparison to publicly available predicate and reference device information.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. As no clinical testing was performed, there was no "test set" requiring expert ground truth establishment for a patient-based diagnostic or therapeutic outcome.
4. Adjudication method for the test set
Not applicable. There was no clinical test set requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical medical device (Cold/Hot Compression unit), not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm-only device.
7. The type of ground truth used
The "ground truth" for the non-clinical tests would be the measured physical parameters of the device (e.g., actual temperature output, actual pressure levels, cycle times) against the device's design specifications and the specifications of the predicate devices. For safety standards, the ground truth is adherence to the requirements of the standards (e.g., ANSI AAMI ES60601-1, IEC 60601-1-2).
8. The sample size for the training set
Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The design was likely based on engineering principles and comparison to existing devices.
9. How the ground truth for the training set was established
Not applicable. No training set was used.
Ask a specific question about this device
(599 days)
JKH HEALTH CO., LTD.
Infrared Thermometer is intended temperature measurement at home or hospital for people of all ages without contact to human body.
Not Found
Here's an analysis of the provided text regarding the acceptance criteria and study for the JKH Health Co., Ltd. Infrared Thermometer:
Based on the provided FDA 510(k) clearance letter and Indications for Use, the document does not contain detailed information about the specific acceptance criteria for performance, the study design, sample sizes, expert qualifications, or ground truth establishment that would typically be found in a clinical study report or detailed submission summary.
The document primarily focuses on the regulatory clearance process, confirming that the device (Infrared Thermometer, Models: JKH-300, JKH-301, JKH-302, JKH-303, JKH-304) is substantially equivalent to legally marketed predicate devices. It states its intended use for non-contact temperature measurement for people of all ages at home or in a hospital setting.
Therefore, many of the requested details cannot be extracted from this particular document. I will highlight what can be inferred or is directly stated, and mark what is not present.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Not specified in document | Not specified in document |
(Typically would include accuracy, precision, bias, measurement range, e.g., "Accuracy +/- 0.2°C at 37.0°C") | (e.g., "Accuracy was measured as +/- 0.15°C for 95% of readings") |
2. Sample sized used for the test set and the data provenance
- Sample size: Not specified.
- Data provenance: Not specified (e.g., country of origin, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: Not specified.
- Qualifications of experts: Not specified.
4. Adjudication method for the test set
- Adjudication method: Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: Not applicable. This device is a standalone thermometer, not an AI-assisted diagnostic tool that would typically involve human readers interpreting results.
- Effect size of human improvement with AI: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone performance: Yes, this is an inference given the device's nature as an infrared thermometer. Its performance would be evaluated as a standalone device without human interpretation of its core function (temperature measurement). However, the specific details of this testing (e.g., how accuracy was validated against a reference standard) are not provided in this document.
7. The type of ground truth used
- Type of ground truth: Not explicitly stated. For a thermometer, the ground truth would typically be a highly accurate reference thermometer measurement (e.g., a laboratory-grade precision thermometer, or an invasive core body temperature measurement in a clinical setting for calibration/validation).
8. The sample size for the training set
- Sample size for training set: Not applicable and not specified. Infrared thermometers typically do not involve "training sets" in the machine learning sense. Their calibration and performance are based on physical principles and manufacturing processes, demonstrated through verification and validation testing.
9. How the ground truth for the training set was established
- Ground truth for training set: Not applicable and not specified. See point 8.
Summary of what's provided for the JKH Health Co., Ltd. Infrared Thermometer:
- Device Name: Infrared Thermometer, Model: JKH-300, JKH-301, JKH-302, JKH-303, JKH-304
- Regulation Number/Name: 21 CFR 880.2910, Clinical Electronic Thermometer
- Regulatory Class: Class II
- Product Code: FLL
- Intended Use: Temperature measurement at home or hospital for people of all ages without contact to the human body.
- Type of Use: Over-The-Counter Use
- Regulatory Status: Cleared via 510(k) (K211050) as substantially equivalent to predicate devices.
What is NOT provided in this FDA 510(k) clearance letter (and would typically be found in a detailed test report or submission summary):
- Specific numerical acceptance criteria for accuracy, precision, or other performance characteristics.
- Detailed results of the device's performance against these criteria.
- Information about the clinical or laboratory studies conducted (e.g., number of subjects, age groups, measurement sites, comparison methods, statistical analysis).
- Any information regarding training data, ground truth establishment, or expert involvement, as these are more pertinent to AI/ML-driven devices.
Ask a specific question about this device
(118 days)
JKH HEALTH CO., LTD.
The Compression Therapy Device is intended to be an easy to use portable by a physician, for use in the home or clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (stimulating muscle contractions). This device can be used to:
- · Aid in the prevention of DVT;
- · Enhance blood circulation;
- · Diminish post-operative pain and swelling;
- · Reduce wound healing time;
- · Aid in the treatment and healing of stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs.
The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.
Compression Therapy Device
Based on the provided text, the document is an FDA 510(k) clearance letter for a "Compression Therapy Device." This letter does not contain the type of detailed study information (acceptance criteria, specific performance metrics, sample sizes, ground truth establishment, expert qualifications, etc.) that would be required to answer your request.
The letter confirms that the device is substantially equivalent to legally marketed predicate devices and lists its indications for use. It discusses regulatory compliance, labeling, and other administrative aspects. It does not provide a technical report of device performance or the study design and results validating its functionality against specific acceptance criteria.
Therefore, I cannot provide the requested information based on the given text.
To answer your request, you would need a different type of document, such as a summary of safety and effectiveness data (SSED), a clinical study report, or a comprehensive technical submission to the FDA.
Ask a specific question about this device
(119 days)
JKH Health Co., LTD
TENS:
PL-029K29, PL-029K30, and PL-029Q are used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
PL-029K29, PL-029K30, and PL-029Q are also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
PMS:
PL-029K29, PL-029K30, and PL-029Q are used to stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and for strengthening muscles in the arms, abdomen. legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
PL-029K29, PL-029K30, and PL-029Q are also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.
Heating:
PL-029K30 and PL-029Q is intended for temporary relief of minor aches and pains.
The subject device delivers electric pulses generated to the user's body areas such as the back and foot through the electrodes. The devices include operating elements, such as the ON/OFF button, intensity increase button, and intensity decrease button, and could be attached to electrodes. The device has multiple program modes of different pulse frequencies, covering TENS and PMS that is also called Electrical Muscle Stimulation (EMS). the device may also provide heat for temporary relief of minor aches and pains.
The device could be easily operated through its buttons to manually realize its functions, such as turning on/off and increasing/decreasing intensity, providing heat/light if needed. The optional wireless control via a remote or Bluetooth APP could provide a secondary operation way to the user, who could be able to wirelessly realize the functions mentioned above.
The electrodes cleared include the electrode patches/pads and electrode garments, which could be packaged together with the 510(k)-cleared devices or packaged separately as the replacement electrodes for 510(k)-cleared devices. The exact percentage of ingredients (such as lidocaine, menthol, and/or cannabidiol) used in the biocompatible electrodes is withheld as the trade secret and may be disclosed as requested.
The provided text is a 510(k) Summary for the StimPlus Patch, a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) with heating function. The document focuses on demonstrating that the subject device is substantially equivalent to legally marketed predicate devices, not on proving that the device meets specific acceptance criteria through a clinical study involving human patients.
Therefore, many of the requested details about acceptance criteria, study design, expert involvement, and ground truth establishment (especially pertaining to clinical performance) cannot be extracted from this document because the submission method relies on demonstrating substantial equivalence to pre-existing, cleared devices rather than a de novo clinical performance study.
However, I can extract the information related to the non-clinical tests and the comparison of the device's technical specifications to those of its predicates, which serve as a form of "acceptance criteria" for demonstrating substantial equivalence.
Here's the breakdown of what can and cannot be answered based on the provided text:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria for this 510(k) submission are based on demonstrating "substantial equivalence" to predicate devices. This is achieved by showing that the subject device has the same intended use and similar technological characteristics, or that any differences do not raise new questions of safety or effectiveness. The core of this is the comparison table of technical parameters.
Table: Comparison of Subject Device (StimPlus Patch) to Predicate Devices
Parameter | Subject Device (K200561) Performance | Primary Predicate (K162517) Performance | Secondary Predicate (K141260) Performance | Equivalence Claim |
---|---|---|---|---|
Intended Use | TENS: Temporary relief of pain, chronic pain, arthritis pain. PMS: Improve muscle performance, tone, strength; temporarily increase local blood circulation. Heating: Temporary relief of minor aches and pains. | TENS: Temporary relief of pain, chronic pain, arthritis pain. PMS: Improve muscle performance, tone, strength; temporarily increase local blood circulation. Heating: Temporary relief of minor aches and pains. | Temporary relief of pain associated with sore and aching muscles. | Identical |
Prescription or OTC | OTC | OTC | OTC | Identical |
Power Source(s) | Rechargeable or non-rechargeable battery | Rechargeable battery | Non-rechargeable battery | Identical |
Method of Line Current Isolation | Battery Supply | Battery Supply | Battery Supply | Identical |
Patient Leakage Current: Normal Condition (μA) | 2.0 | 2.0 | 2.0 | Identical |
Patient Leakage Current: Single Fault Condition (μA) |
Ask a specific question about this device
(217 days)
JKH Health Co., Ltd.
TENS (Modes 1, 2, 4, 5, 6, 8)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
PMS (also called EMS, Modes 1, 3, 7)
To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.
Heating Mode Temporary relief of minor aches and pains.
Electronic Pulse Stimulator delivers electric pulses generated to the user's body areas such as the back neck and foot through the electrodes. The portable and compact device has multiple modes of different pulse frequencies, covering TENS and PMS that is also called Electrical Muscle Stimulation (EMS). It includes operating elements of ON/OFF button, intensity increase button, intensity decrease button, mode selection button, and/or timer selection button, and could be attached to electrodes. In addition, the device may also provide heat/temperature. While used in the heating mode, the device is coupled with electronically controlled electrodes to provide automatic thermal heat to the skin with the maximum temperature of 43 °C.
The device could be easily operated through its buttons to manually realize its functions, such as turning on/off, increasing/decreasing intensity, changing mode/timer, and providing heat/temperature if needed. The optional wireless control via a remote or Bluetooth APP could provide a secondary operation way to the user, who could be able to wirelessly realize the functions mentioned above.
The electrodes cleared include the electrode pads and electrode garments, which could be packaged separately and/or together with the subject device.
This document is a 510(k) summary for an Electronic Pulse Stimulator (device K162517). It describes the device, its intended use, and demonstrates substantial equivalence to predicate devices (K153520 and K070299). Since this is a 510(k) summary for a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulation (PMS) device, the acceptance criteria and study information will focus on its electrical and heating parameters rather than clinical performance metrics typically associated with AI/ML devices (like sensitivity, specificity, AUC).
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in the traditional sense of a predetermined numerical threshold for a clinical outcome (e.g., "sensitivity must be >90%"). Instead, it demonstrates "substantial equivalence" based on various technical specifications and intended use. The "reported device performance" is essentially a comparison of these technical specifications between the subject device and the predicate devices.
The table below summarizes the comparison from Table 1 in the document for the subject device (PL-029K13, as it includes the heating mode) and its primary predicate (PL-029K6, K153520) for electrical stimulation, and the heating predicate (ezFit Digital Heating TENS, K070299) for the heating function. The ranges and specific values are presented as performance data.
Feature | Predicate Device (K153520) Performance (PL-029K6) | Predicate Device (K070299) Performance (ezFit Digital Heating TENS) | Subject Device (K162517) Performance (PL-029K13) | Acceptance Criteria/Basis for Equivalence |
---|---|---|---|---|
TENS Mode | ||||
Intended Use (TENS) | Temporary relief of pain (sore/aching muscles, chronic intractable pain, arthritis) | Symptomatic relief and management of chronic intractable pain. | Temporary relief of pain (sore/aching muscles, chronic intractable pain, arthritis) | Substantially Equivalent to K153520 for TENS. |
Max Output Voltage (500Ω) | Mode 2: 39.2; Mode 3: 64; Mode 4: 56.8; Mode 5: 33.2; Mode 6: 31.6 | Undisclosed | Mode 2: 31.2; Mode 3: 46.0; Mode 4: 42.0; Mode 5: 27.6; Mode 6: 27.6; Mode 7: 40.8; Mode 8: 23.2 | Similar technical characteristics topredicate. The 7th and 8th modes of the subject device are similar to the 6 initial modes of the predicate. |
Max Output Voltage (2kΩ) | Mode 2: 82.4; Mode 3: 84; Mode 4: 79.2; Mode 5: 70.4; Mode 6: 67.2 | Undisclosed | Mode 2: 68.0; Mode 3: 90.4; Mode 4: 68.8; Mode 5: 60.0; Mode 6: 60.0; Mode 7: 84.0; Mode 8: 50.4 | Similar technical characteristics topredicate. |
Max Output Voltage (10kΩ) | Mode 2: 129; Mode 3: 120; Mode 4: 84.8; Mode 5: 121; Mode 6: 124 | Undisclosed | Mode 2: 118; Mode 3: 124; Mode 4: 78.4; Mode 5: 115; Mode 6: 115; Mode 7: 124; Mode 8: 99.2 | Similar technical characteristics topredicate. |
Max Output Current (500Ω) | Mode 2: 78.4; Mode 3: 128; Mode 4: 113.6; Mode 5: 64.4; Mode 6: 63.2 | Undisclosed | Mode 2: 62.4; Mode 3: 92.0; Mode 4: 84.0; Mode 5: 55.2; Mode 6: 55.2; Mode 7: 81.6; Mode 8: 46.4 | Similar technical characteristics topredicate. |
Max Output Current (2kΩ) | Mode 2: 41.2; Mode 3: 42; Mode 4: 39.6; Mode 5: 35.2; Mode 6: 33.6 | Undisclosed | Mode 2: 34.0; Mode 3: 45.2; Mode 4: 34.4; Mode 5: 30.0; Mode 6: 30.0; Mode 7: 42.0; Mode 8: 25.2 | Similar technical characteristics topredicate. |
Max Output Current (10kΩ) | Mode 2: 12.9; Mode 3: 12; Mode 4: 8.5; Mode 5: 12.1; Mode 6: 12.4 | Undisclosed | Mode 2: 11.8; Mode 3: 12.4; Mode 4: 7.84; Mode 5: 11.5; Mode 6: 11.5; Mode 7: 12.4; Mode 8: 9.92 | Similar technical characteristics topredicate. |
Pulse Period (mSec) | 10-833 | Undisclosed | 5.6-806 | Similar technical characteristics topredicate. |
Frequency (Hz) | Mode 2: 69.4; Mode 3: 13.0-52.1; Mode 4: 1.2; Mode 5: 96.2; Mode 6: 96.2 | Undisclosed | Mode 2: 73.5; Mode 3: 13.7-59.5; Mode 4: 1.24; Mode 5: 104.1; Mode 6: 104.1; Mode 7: 20.8; Mode 8: 178.5 | Similar technical characteristics topredicate. |
Max Phase Charge (500Ω) | Mode 2: 15.1; Mode 3: 25.6; Mode 4: 18.2; Mode 5: 12.8; Mode 6: 10.1 | Undisclosed | Mode 2: 11.5; Mode 3: 16.9; Mode 4: 15.5; Mode 5: 10.2; Mode 6: 10.2; Mode 7: 15.0; Mode 8: 8.54 | Similar technical characteristics topredicate. |
Max Current Density (500Ω) | Mode 2: 2.18; Mode 3: 3.56; Mode 4: 3.16; Mode 5: 1.84; Mode 6: 1.76 | Undisclosed | Mode 2: 2.23; Mode 3: 3.29; Mode 4: 3.00; Mode 5: 1.97; Mode 6: 1.97; Mode 7: 2.91; Mode 8: 1.66 | Similar technical characteristics topredicate. |
Max Average Power Density (500Ω) | Mode 4: 0.03; Mode 2: 1.14; Mode 3: 0.64-2.56; Mode 5: 1.13; Mode 6: 0.85 | Undisclosed | Mode 2: 0.92; Mode 3: 0.32-1.37; Mode 4: 0.04; Mode 5: 1.04; Mode 6: 1.04; Mode 7: 0.46; Mode 8: 1.26 | Similar technical characteristics topredicate. |
PMS Mode | ||||
Intended Use (PMS) | Stimulate healthy muscles (improve/facilitate muscle performance, muscle tone/firmness, strengthening, increase local blood circulation) | Not applicable (K070299 only TENS and heating) | Stimulate healthy muscles (improve/facilitate muscle performance, muscle tone/firmness, strengthening, increase local blood circulation) | Substantially Equivalent to K153520 for PMS. |
Heating Mode | ||||
Intended Use (Heating) | Not applicable (K153520 only TENS and PMS) | Temporary relief of minor aches and pains and muscle spasms. | Temporary relief of minor aches and pains. | Substantially Equivalent to K070299. The technology is similar, and the indications are comparable. |
Heating Setting | N/A | Adjustable (36-42 °C) | Low and High | Similar technology (delivery of heat). |
Maximum Temperature Setting | N/A | 42 °C | 43 °C | Similar maximum temperature setting, within acceptable safety margins for human skin. |
Skin Temperature | N/A | N/A (implied safe operation across range) | Reached 41 °C in 20 min and remained stable. | Demonstrates safe temperature limits and stable operation. |
Prescription/OTC | OTC | Prescription | OTC | Transition from prescription to OTC for heating is supported by substantial equivalence in technology and safety. The electrical stimulation aspects are OTC as well (similar to K153520). |
For non-clinical tests, the "acceptance criteria" were compliance with specific voluntary standards:
- IEC 60601-1 (medical electrical equipment safety and essential performance)
- IEC 60601-1-2 (electromagnetic compatibility)
- IEC 60601-2-10 (particular requirements for nerve and muscle stimulators)
- FDA Guidance for Premarket Submissions for Software Contained in Medical Devices.
The reported device performance for these non-clinical tests is that the device "conformed" and "met the requirement of safety" for the standards and guidance, respectively.
2. Sample Size Used for the Test Set and the Data Provenance
The document explicitly states: "The subject device does not conduct, nor rely upon, clinical tests to determine substantial equivalence. Non-clinical tests were performed on the subject device..."
Therefore, there is no clinical test set, human subject data, or data provenance (country of origin, retrospective/prospective) described in this 510(k) summary for the purpose of demonstrating substantial equivalence. The evaluation relies on technical comparisons to predicate devices and adherence to engineering and safety standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
As no clinical test set was used, there was no ground truth established by experts in a clinical context for this submission. The "ground truth" for the non-clinical tests would be the specifications and requirements of the regulatory standards themselves, which are established by expert committees in their respective fields (e.g., electrical engineering, medical device safety).
4. Adjudication Method for the Test Set
Since no clinical test set with human subject outcomes was used for demonstrating substantial equivalence, no adjudication method (like 2+1 or 3+1 consensus) was applied.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed to evaluate the impact of an AI/ML device on human reader performance, which is not applicable to an Electronic Pulse Stimulator submission focused on electrical and heating parameters and substantial equivalence to existing devices.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done
No, a standalone performance study in the context of an AI/ML algorithm was not done. The "performance" in this submission refers to the technical specifications of the device (e.g., output voltage, current, temperature) and its compliance with safety standards.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)
For the technical specifications, the "ground truth" is established by engineering measurements against reference standards (e.g., voltmeters, ammeters, thermometers calibrated to international standards) to verify the device's output. For compliance with voluntary standards (IEC 60601-1, -1-2, -2-10), the "ground truth" is the requirements outlined in those international standards. For software, it's compliance with FDA guidance for software in medical devices.
8. The Sample Size for the Training Set
Since this is not an AI/ML device, there is no training set sample size described. The device's operation is based on fixed electrical circuits and heating elements, not learned models.
9. How the Ground Truth for the Training Set Was Established
As there is no training set, this question is not applicable.
Ask a specific question about this device
(149 days)
JKH HEALTH CO.,LTD.
TENS (Modes 1, 2, 4, 5, 6)
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, arm, and leg, due to strain from exercise or normal household and work activities.
It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
PMS (Modes 1 and 3)
To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or for the treatment of any medical conditions or diseases.
It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.
Not Found
The provided text is a 510(k) premarket notification letter from the FDA regarding an "Electronic Pulse Stimulator." It confirms substantial equivalence to predicate devices for specific Indications for Use. However, this document does not contain any information regarding acceptance criteria, device performance, study details, sample sizes, ground truth establishment, or expert qualifications.
Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided text. The document is primarily a notification of regulatory clearance, not a summary of performance studies.
Ask a specific question about this device
Page 1 of 1