Search Results
Found 15 results
510(k) Data Aggregation
(117 days)
Ask a specific question about this device
(29 days)
Ask a specific question about this device
(90 days)
Ask a specific question about this device
(30 days)
Ask a specific question about this device
(90 days)
Ask a specific question about this device
(29 days)
Ask a specific question about this device
(28 days)
Ask a specific question about this device
(76 days)
The IntraStent™ DoubleStrut™ LD Biliary Stent is indicated as a palliative treatment for malignant neoplasms in the biliary tree.
The IntraStent™ DoubleStrut™ LD is a balloon expandable stainless steel stent with an open cell design. The stent is provided unmounted, to be manually crimped onto an open lumen PTA balloon catheter for biliary stent expansion of choice by the physician. Upon balloon inflation the crimped stent expands to conform to the duct inner luminal surface and retains the expanded state upon balloon deflation.
The provided text describes a 510(k) submission for the IntraStent™ DoubleStrut™ LD Biliary Stent, focusing on demonstrating substantial equivalence to predicate devices, rather than a study proving the device meets acceptance criteria from a clinical outcome perspective.
Here's an analysis based on the information provided, specifically addressing the requested points:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not define specific clinical acceptance criteria in terms of performance metrics (e.g., patency rates, complication rates). Instead, the "acceptance criteria" are implied as demonstrating substantial equivalence through bench tests to established predicate devices.
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Meet same performance characteristics as predicate IntraStent™ DoubleStrut™ (K991929) | Bench tests were performed to verify this. |
| Non-pyrogenic | Pyrogenicity testing performed on a lot-to-lot basis supports this claim. |
| Identical materials to predicate devices | Stated as identical to currently marketed IntraStent™ DoubleStrut™ and Cordis Corporation/Johnson & Johnson PALMAZ™ and PALMAZ-SCHATZ™ stents. |
| Identical indication for use to predicate devices | Stated as identical: "palliative treatment for malignant strictures of the biliary tree." |
| Identical technological characteristics to predicate devices | Stated as identical, providing a larger version of IntraStent™ DoubleStrut™ (9-12 mm expanded diameter), fabricated by laser-cutting 316L stainless steel hypotube. |
2. Sample Size Used for the Test Set and Data Provenance
The submission describes bench tests and pyrogenicity testing, not clinical trials with patient data.
- Test Set Sample Size: Not explicitly stated for bench tests (e.g., number of stents tested). For pyrogenicity, it's mentioned as "lot to lot basis," implying samples from manufacturing lots.
- Data Provenance: Not applicable in the context of clinical data (no country of origin or retrospective/prospective clinical data mentioned). The data provenance is from internal bench testing conducted by the manufacturer.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The ground truth for bench tests is the engineering specification and performance of the predicate device. There is no mention of experts establishing a "ground truth" for the test set in the sense of clinical diagnoses or interpretations.
4. Adjudication Method for the Test Set
Not applicable. This concept primarily applies to human reader studies or clinical trials where expert consensus is needed to resolve discrepancies. Bench tests rely on objective physical measurements and compliance with engineering specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance?
Not applicable. This is a medical device (stent), not an AI-powered diagnostic or decision-support system. Therefore, MRMC studies comparing human readers with and without AI assistance are not relevant to this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a medical device (stent), not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for this 510(k) submission is the performance and characteristics of the legally marketed predicate devices (IntraStent™ DoubleStrut™ K991929 and Cordis Corporation PALMAZ™ and PALMAZ-SCHATZ™ stents). The testing (bench tests, pyrogenicity) aimed to demonstrate that the new device's performance aligns with these established predicates.
8. The Sample Size for the Training Set
Not applicable. This is not an AI or machine learning device requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, this is not an AI or machine learning device.
Summary of the Study Proving Acceptance Criteria:
The study proving the device meets the "acceptance criteria" (understood as demonstrating substantial equivalence to predicate devices) primarily involved:
- Bench Tests: These were conducted to verify that the IntraStent™ DoubleStrut™ LD met the "same performance characteristics" as the predicate IntraStent™ DoubleStrut™ (K991929). The submission states "Performance testing (bench) further supports a substantial equivalence claim." No specific details about the types of bench tests (e.g., radial force, fatigue, crimp retention) or their results are provided in this summary.
- Pyrogenicity Testing: This was performed on a "lot to lot basis" to support the claim that the device is non-pyrogenic.
The conclusion is that "The collective evidence therefore provides assurance that the IntraStent™ meets the requirements that are considered acceptable for the intended use" by demonstrating its "identical" nature in materials, indication for use, and technological characteristics, and by showing similar performance in bench tests, to its predicate devices.
Ask a specific question about this device
(56 days)
The IntraStent™ Biliary Stent is indicated as a palliative treatment for malignant neoplasms in the biliary tree.
The IntraStent™ is a balloon expandable stainless steel stent with an open lattice design. The device is provided unmounted, to be manually crimped onto a noncompliant PTA balloon catheter for biliary stent expansion of choice by the physician. Upon balloon inflation the crimped stent expands to conform to the duct inner luminal surface and retains the expanded state upon balloon deflation.
The provided document describes a 510(k) premarket notification for a medical device, specifically the IntraStent™ Biliary Stent. This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than proving its safety and effectiveness through extensive clinical trials.
Therefore, the study described is not a typical clinical study with acceptance criteria in the usual sense (e.g., sensitivity, specificity, accuracy against a gold standard). Instead, the "acceptance criteria" here relate to demonstrating that the new device performs equivalently to existing devices through bench testing.
Here's a breakdown based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Meet same performance characteristics as predicate ITI Stent (K980290 & K991929) | Bench tests were performed to verify this. (Specific performance characteristics are not detailed in the summary, but implied to be equivalent). |
| Non-pyrogenic | Pyrogenicity testing performed on a lot-to-lot basis supports the claim. |
| Identical materials to predicate stents | Stated as identical materials (balloon expandable stainless steel stent with open lattice design, 316L stainless steel hypotube). |
| Identical indication for use as predicate stents | Stated as "as a palliative treatment for malignant neoplasms in the biliary tree," which is the same as the predicate devices. |
| Identical technological characteristics to predicate stents | Stated as having identical technological characteristics (e.g., laser-cut slots/apertures, cleaning, electro-polishing, packaging, sterilization method for the new device – ethylene oxide vs. CO60 for predicate Palmaz). Exception: new device is a smaller version allowing 4-7mm expanded diameter. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: Not explicitly stated as a numerical sample size. The testing involves "bench tests" and "pyrogenicity testing on a lot to lot basis." For bench tests of medical devices like stents, this typically means a representative number of units from manufacturing lots are tested.
- Data Provenance: Not specified. Bench testing is typically performed in a laboratory setting, and pyrogenicity testing would be performed on manufactured product lots. There is no mention of human clinical data or geographical origin.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This type of information is not applicable to this submission. The "ground truth" for substantial equivalence is the performance of the predicate devices. The validation is against engineering and material specifications and the performance of existing, approved devices, not against expert interpretation of data.
4. Adjudication Method for the Test Set
This is not applicable. There is no "test set" in the sense of clinical cases requiring adjudication by experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. This submission is for a physical medical device (biliary stent), not an AI-powered diagnostic or interpretive tool. Therefore, MRMC studies and AI assistance are irrelevant.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used
- The "ground truth" for this submission is the established performance and characteristics of the legally marketed predicate devices (IntraTherapeutics, Inc. ITI Stent (K980290 & K991929) and Cordis Corporation PALMAZ™, PALMAZ-SCHATZ™, and PERFILEX™ Balloon-Expandable Biliary Stent (K905720, K911581, K964688, K980653)). The new device's performance is compared to these benchmarks through bench testing.
8. The Sample Size for the Training Set
This is not applicable. There is no "training set" as this is not a machine learning or AI-driven device.
9. How the Ground Truth for the Training Set was Established
This is not applicable.
Ask a specific question about this device
(55 days)
The IntraMax™ guide catheter is intended for intravascular introduction of diagnostic, therapeutic and interventional* devices into the extracranial vasculature.
The IntraMax™ is single lumen devices designed to aid the physician in the access of distal vasculature with the aid of a quidewire. The IntraMax™ is a family of catheters currently manufactured in three French sizes (6, 7 and 8 FR), multiple distal stem configurations, lengths and side holes ("style features"). Per French size, catheters will be offered with differing handling characteristics to allow flexibility in meeting physician preferences within the indication for use.
The provided text describes the 510(k) summary for the IntraMax™ Guide Catheter, focusing on its substantial equivalence to a predicate device rather than a study demonstrating the device meets specific acceptance criteria in the manner you've requested (e.g., performance metrics against established thresholds).
Therefore, much of the information you've asked for (e.g., detailed acceptance criteria table with performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth establishment) is not available in the provided document.
The document indicates that "functional testing was performed to ensure the IntraMax™ would function according to its intended use instructions" and "All testing conducted confirmed the acceptability of the IntraMax™ for the intended purpose." However, specific quantitative acceptance criteria or detailed results of these functional tests are not provided. The focus is on demonstrating substantial equivalence to a previously cleared device (IntraTherapeutics, Inc. IntraMax™ Guide Catheter, K981191) based on identical intended use, equivalent materials, and technological characteristics.
Here's a summary of the available information based on your request, highlighting what is not explicitly stated in the provided text:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| Function according to intended use instructions. | "functional testing was performed to ensure the IntraMax™ would function according to its intended use instructions." and "All testing conducted confirmed the acceptability of the IntraMax™ for the intended purpose." (No specific quantitative performance metrics are provided, but the statement implies successful completion of functional tests.) |
| Biocompatibility in compliance with ISO 10993 and GLP requirements. | "Biocompatibility tests per ISO 10993 were also performed and were in compliance with ISO and GLP requirements." (This indicates successful compliance with specific standards, but details of the tests and results are not given.) |
| Substantial equivalence to predicate device (K981191) in materials. | "The IntraMax™ is identical for indication for use and equivalent in materials and technological characteristics." (No specific material composition details are provided, but equivalence to the predicate is asserted.) |
| Substantial equivalence to predicate device (K981191) in technological characteristics. | "The IntraMax™ is identical for indication for use and equivalent in materials and technological characteristics. Performance functional testing (bench) further supports a substantial equivalence claim." (Details of the specific technological characteristics and the comparative "bench" tests are not provided, only the assertion of equivalence.) |
2. Sample size used for the test set and the data provenance
- Sample size for test set: Not explicitly stated. The document refers to "functional testing" but does not provide details on the number of devices tested.
- Data provenance: The tests described (functional and biocompatibility) are laboratory/bench tests and not human clinical data. Their origin is internal to the manufacturer (IntraTherapeutics, Inc.).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This information is not applicable to the type of testing described (functional and biocompatibility bench tests) for establishing substantial equivalence. No human "ground truth" was established in this context.
4. Adjudication method for the test set
- Not applicable, as it was not a human-reader study.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC study was not done. This device is a guide catheter and not an AI-powered diagnostic device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable, as this is a medical device (catheter) and not an algorithm.
7. The type of ground truth used
- For functional testing, the "ground truth" would be engineering specifications and functional requirements (e.g., flow rate, flexibility, tensile strength, pushability, torqueability, lubricity, etc.), though these specific metrics are not detailed.
- For biocompatibility testing, the "ground truth" is compliance with ISO 10993 standards and Good Laboratory Practice (GLP) requirements, as determined by laboratory analysis.
8. The sample size for the training set
- Not applicable. This is not an AI/machine learning device that would require a "training set."
9. How the ground truth for the training set was established
- Not applicable.
In summary: The provided 510(k) summary focuses on demonstrating "substantial equivalence" of the IntraMax™ Guide Catheter to a predicate device through functional and biocompatibility testing. It does not contain the detailed quantitative acceptance criteria, specific performance metrics, or study design information (like sample sizes, expert qualifications, or ground truth methods) that would be present in a submission for a novel device or a device requiring clinical performance evaluation with human data.
Ask a specific question about this device
Page 1 of 2