(210 days)
The Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stents are indicated for palliation of malignant neoplasms in the billiary tree.
The Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stent is a stainless steel, multi-articulated, slotted tube with nominal lengths of 42mm (PS424), 56mm (PS564), and 78mm (PS784) and an expansion range of 6-10mm. The Stents are intended for palliation of malignant neoplasms in the biliary tree. The Stents are provided unmounted to be crimped onto commercially available Cordis Opta 5™ Catheters which have been qualified by Cordis for stent delivery. The Stent is expanded via inflation of the balloon catheter and deployed. Upon expansion, the Stent forms a strong open lattice in a diamond pattern. The expanded Stent conforms to the inner luminal surface of the bile duct. The ability of the Stent to palliate malignant neoplasms is based upon the ability of the metallic structure of the Stent to plastically deform when the balloon is inflated and maintain its expanded state.
The provided text describes the 510(k) summary for the Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stent. This document details the device, its intended use, technological characteristics, and nonclinical performance testing. It is a premarket notification for a medical device and therefore focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study for a novel AI device.
Given this context, the request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in relation to an AI system, along with specific questions like "Sample sized used for the test set and the data provenance," "Number of experts used to establish the ground truth," and "If a multi reader multi case (MRMC) comparative effectiveness study was done," is not applicable to the provided document. The 510(k) summary describes a physical medical device (a stent), not an AI algorithm or software.
However, I can extract the nonclinical performance parameters and their reported outcomes, which serve as the "acceptance criteria" for this physical device as demonstrated through qualification testing.
Here's a breakdown based on the provided text, adapted to best fit the original request's structure:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Parameter Studied for Qualification) | Reported Device Performance |
|---|---|
| Inflation time | Samples were fully expanded to the nominal diameter within the allotted time period. |
| Balloon burst | Mounted stents did not modify the catheter's labeled maximum burst pressure. |
| Crossing profile | A 7F sheath is required for 6mm diameter balloons with mounted stents and an 8F sheath is required for 8mm and 10mm diameter balloons with mounted stents. |
| Distension | All stents reached their nominal diameter within the recommended pressure. |
| Deflation time/deflatability | All balloons with mounted stents deflated within the allotted time period. The presence of the stent did not cause any interference with balloon deflatability. |
| Stent retention | All samples met or exceeded internal criteria for stent retention. |
| Recoil and Length | The stent remains within its nominal diameter after recoil. Lengths were measured to determine stent length at each expanded diameter (for labeling purposes). |
| Radial Force | The stent will not collapse within the clinically relevant range. |
| Percent Open Area | All stent sizes have an open area of 80-90% in their expanded diameter range. |
| MRI Compatibility | The subject devices were found to be MRI compatible. |
| Fatigue/Corrosion | No cracks or other metallurgical defects resulted from fatigue testing. |
Regarding the other questions, they are not applicable as this document describes a physical medical device, not an AI/software device.
- 2. Sample sized used for the test set and the data provenance: Not applicable. The document mentions "samples" for testing but does not specify sample sizes for each test in a dataset context, nor data provenance like country of origin or retrospective/prospective. These are engineering/material tests.
- 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for these physical characteristics is established through engineering measurements and established material science tests, not expert interpretation in the way it's used for AI.
- 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
- 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
- 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the physical stent's performance is based on direct physical measurements, engineering specifications, and established biocompatibility/material science standards.
- 8. The sample size for the training set: Not applicable. There is no concept of a "training set" for this kind of device.
- 9. How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
JUN 20 1997
Image /page/0/Picture/1 description: The image shows a series of handwritten alphanumeric characters. The characters appear to be "K96 4688" in a bold, slightly irregular font. The numbers are closely spaced and have a handwritten style.
510(k) Summary
Submitter's name:
Cordis Corporation, a Johnson & Johnson Company 40 Technology Drive Warren, New Jersey 07059 908-755-8300 908-412-3915 (fax)
Contact person:
Joan Martin Manager, Regulatory Affairs (908) 412-7250
Date prepared:
Trade name:
Common name:
Classification name:
Predicate devices:
Device description:
Joan Martin
Manager, Regulatory Affairs
(908) 412-7250
Long Medium PALMAZ-SCHATZ™
November 20, 1996
Balloon-Expandable Stent
Biliary Stent
§876.5010 Biliary Catheter and Accessories
PALMAZ™ and PALMAZ-SCHATZ™ Balloon-Expandable Stents
Schneider's Wallstent® Biliary Endoprosthesis
The Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stent is a stainless steel, multi-articulated, slotted tube with nominal lengths of 42mm (PS424), 56mm (PS564), and 78mm (PS784) and an expansion range of 6-10mm. The Stents are intended for palliation of malignant neoplasms in the biliary tree. The Stents are provided unmounted to be crimped onto commercially available Cordis Opta 5™ Catheters which have been qualified by Cordis for stent delivery. The Stent is expanded via inflation of the balloon catheter and deployed. Upon expansion, the Stent forms a strong open lattice in a diamond pattern. The expanded Stent conforms to the inner luminal surface of the bile duct. The ability of the Stent to palliate malignant neoplasms is based upon the ability of the metallic structure of the Stent to plastically deform when the balloon is inflated and maintain its expanded state.
Accessory Devices - The following accessory devices have been slightly modified to accommodate the longer Stent lengths. Shorter versions of these devices are
{1}------------------------------------------------
510(k) Summary (p.2)
marketed for use with our current line of biliary stents, as well as with our iliac stent.
The Crimping Tube is constructed of polyethylene. It is supplied sterile, for one use only and is provided with the unmounted Stent. The Crimping Tube protects the Stent/balloon assembly during the manual crimping process. Crimping tube modifications consist of a longer length to accommodate the longer stents and a slightly decreased wall thickness.
The Crimping Tool (CRT 80) is a hand-held, reusable, metallic instrument which is used to manually crimp the unmounted Stents onto the recommended balloon catheter. The Crimping Tool is provided nonsterile and must be sterilized prior to use by standard autoclave procedures. The tool delivers a radial compressive force to the Stent, thereby securing the Stent to the balloon catheter. The Crimping Tool die has been modified to accommodate the additional stent lengths. Our currently marketed Crimping Tool (CRT40) is used to crimp the 42mm Long Medium Stents.
Intended use statement: The stents are intended for palliation of malignant neoplasms in the biliary tree.
Labeled indication:
The Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stent is indicated for palliation of malignant neoplasms in the biliary tree.
Technological characteristics: The Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stents are substantially equivalent in design, function and intended use to the currently marketed PALMAZ™ and PALMAZ-SCHATZ™ Biliary Stents and to the Schneider Wallstent® Biliary Endoprosthesis.
The subject and predicate devices are constructed of surgical grade stainless steel which is a known biocompatible implant material. The subject and predicate devices all have an open mesh design which is
{2}------------------------------------------------
510(k) Summary (p.3)
intended for palliation of malignant neoplasms in the biliary tree. Both the currently marketed and the modified Cordis stents are constructed from stainless steel tubing. The Wallstent® stents are constructed of braided wire.
The subject device and the currently marketed Cordis biliary stents are balloon-expandable. The Wallstent® stents are self-expanding. All of the devices utilize a catherer to deliver the stent. The range of lengths and diameters of the subject and predicate devices are comparable and are intended to meet clinical need.
Diameter: 6-10mm for the subject devices, 4-12mm for the currently marketed Cordis Stents and predetermined 8-10mm for the Wallstent®.
42mm, 56mm and 78mm for the subject Lenaths: device, 10-39mm for the currently marketed Cordis Stents and 15mm to 150mm for the Wallstent® device.
evices are minor and
the subject devices to p
biliary tree. The desc
The differences between the subject and predicate devices are minor and are not relevant to the ability of the subject devices to palliate malignant neoplasms in the biliary tree. The descriptive characteristics are precise enough to demonstrate equivalence. However, qualification testing was performed to confirm the acceptable performance and properties of the subject devices. The following parameters were studied:
Inflation_time: Samples were fully expanded to the nominal diameter within the allotted time period.
Balloon burst: Mounted stents did not modify the catheters's labeled maximum burst pressure.
Crossing profile: A 7F sheath is required for 6mm diameter balloons with mounted stents and an 8F sheath is required for 8mm and 10mm diameter balloons with mounted stents.
Distension: All stents reached their nominal diameter within the recommended pressure.
Deflation time/deflatability: All balloons with mounted stents deflated within the alotted time period. The
Nonclinical performance:
{3}------------------------------------------------
510(k) Summary (p.4)
presence of the stent did not cause any interference with balloon deflatability.
Stent retention: All samples met or exceeded our internal criteria for stent retention.
Recoil and Length: The stent remains within its nominal diameter after recoil. Lengths were measured to determine stent length at each expanded diameter (for labeling purposes).
Radial Force: The stent will not collapse within the clinically relevant range.
Percent Open Area: All stent sizes have an open area of 80-90% in their expanded diameter range.
MRI Compatibility: The subject devices were found to be MRI compatible.
Fatique/Corrosion: No cracks or other metallurgical defects resulted from fatigue testing.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is depicted in a simple, black and white design.
20 1997 JIIN -
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Joan Martin Manager, Regulatory Affairs Cordis Corporation, - - - - - - - ------a Johnson & Johnson Company 40 Technology Drive Warren, New Jersey 07059
Re: K964688
Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stent ............................ Dated: March 31, 1997 Received: April 1, 1997 Regulatory class: II 21 CFR §876.5010/Product code: 78 FGE
Dear Ms. Martin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D Director, Division of Reproductive. Abdominal, Ear, Nose and Throath and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K964688
Long Medium PALMAZ-SCHATZ†MBalloon Expandable Stent Device Name:
Indications For Use:
The Long Medium PALMAZ-SCHATZ*m Balloon Expandable Stents are indicated for palliation of malignant neoplasms in the billiary" tree.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert E. Sather
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number_
Prescription Use
(Per 21 CFR 801.109) ✓
.
OR
Over-The-Counter Use
(Optional Formai 1-2-96)
§ 876.5010 Biliary catheter and accessories.
(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.