Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    K Number
    K163326
    Device Name
    SmartEP
    Date Cleared
    2017-09-29

    (305 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SmartEP is an evoked response testing and diagnostic device, that is capable of eliciting, acquiring, and measuring auditory, somatosensory, visual, and vestibular evoked myogenic potential data, as well as providing nerve stimulation and monitoring.

    The intended use of the SmartEP device is to objectively record evoked responses from patients of all ages upon the presentation of sensory stimuli. The product is indicated for use as a diagnostic aid and adjunctive tool in sensory related disorders (i.e., auditory, somatosensory, visual, and vestibular) and in surgical procedures for inter-operative nerve monitoring.

    The SmartEP system is intended to be used by trained personnel in a hospital, nursery, clinic, audiologist's, EP technologist's, surgeon's, or physician's office, operating room, or other appropriate setting.

    Device Description

    The SmartEP device records evoked potentials by using delivery of auditory, somatosensory, visual, or nerve sensory stimuli and using signal averaging techniques to extract the evoked potential from the uncorrelated electrical activity of the brain (electroencephalography or EEG) and muscles (electromyography or EMG). The device has options for Auditory Evoked Potentials (AEPs), Somatosensory Evoked Potentials (SEPs), Visual Evoked Potentials (VEPs), Vestibular Evoked Myogenic Potentials (VEMPs), and nerve stimulation and monitoring. The SEP, VEP, and nerve stimulation and monitoring functionality. operating principles, and intended uses are the same as on the predicate SmartEP device. On the SmartEP device with VEMP modality, the AEP modality has been modified to facilitate VEMP recording and analysis with optional biofeedback. The VEMP features added are comparable to those found in the ICS Chartr 200 predicate device. The VEMP modality does not provide a diagnosis. Diagnosis is made by a medical professional.

    The SmartEP device is a Windows OS personal computer (PC) based system composed of software modules, an external main hardware unit, an optional biofeedback box, and peripheral stimulus delivery and recording components and accessories. The biofeedback box, stimulation, and recording devices are connected to the main hardware unit which is connected to the PC via a Universal Serial Bus cable. Software on the computer is used for the user interface to facilitate test parameter specification and for data display and analysis purposes.

    The SmartEP with VEMP device has an optional biofeedback hardware accessory (VEMP feedback box) or uses a computer monitor for indicating EMG levels during VEMP testing. The VEMP feedback box has LEDs that indicate that the measured EMG level is either below the minimum value set by the user (Low - orange LED), or is between the minimum and maximum values set by the user (Satisfactory green LED), or is above the maximum value as set by the user (High – orange LED). The computer monitor displays a bar graph and pictorial face that indicates that the measured EMG level is either below the minimum value set by the user (Low - small pink bar and sad face), or is between the minimum and maximum values set by the user (Satisfactory - medium green bar and happy face), or is above the maximum value as set by the user (High - large pink bar and sad face). Recording of VEMPs can be set to occur when the EMG level is within the user programmed satisfactory range.

    AI/ML Overview

    The provided document describes the substantial equivalence determination for the SmartEP device, focusing on its Vestibular Evoked Myogenic Potential (VEMP) testing modality compared to a predicate device. The information primarily addresses performance testing related to VEMP repeatability and reproducibility, rather than a typical AI model's acceptance criteria based on sensitivity/specificity/accuracy or an MRMC study.

    Therefore, many of the requested fields cannot be directly answered as they pertain to AI/machine learning evaluation paradigms which are not the subject of this 510(k) summary. However, I will answer all applicable points based on the provided text.


    Acceptance Criteria and Device Performance (VEMP Modality Only)

    The document does not explicitly state "acceptance criteria" in terms of pre-defined numerical thresholds for performance metrics like sensitivity, specificity, or accuracy, as this is not an AI/diagnostic algorithm in the conventional sense. Instead, the performance is demonstrated through comparisons of repeatability and reproducibility metrics (specifically, correlation values of VEMP waveforms) to a legally marketed predicate device. The general acceptance criterion is demonstrating that the SmartEP device's VEMP performance is "similar or higher" than the predicate device.

    Table of Performance for VEMP Waveform Correlation (SmartEP vs. Predicate)

    Performance MetricSmartEP Device (Range of Mean Correlation)Predicate Device (Mean Correlation)Acceptance Criterion (Implicit)
    cVEMP Waveform Test-Retest Repeatability (Session 1 Mean Correlation)0.911 - 0.915Not explicitly stated for this metric in predicate data.Similar or higher than predicate (if available for comparison) and demonstrating high repeatability.
    cVEMP Waveform Test-Retest Repeatability (Session 2 Mean Correlation)0.921 - 0.941Not explicitly stated for this metric in predicate data.Similar or higher than predicate (if available for comparison) and demonstrating high repeatability.
    oVEMP Waveform Test-Retest Repeatability (Session 1 Mean Correlation)0.944 - 0.956Not explicitly stated for this metric in predicate data.Similar or higher than predicate (if available for comparison) and demonstrating high repeatability.
    oVEMP Waveform Test-Retest Repeatability (Session 2 Mean Correlation)0.951 - 0.955Not explicitly stated for this metric in predicate data.Similar or higher than predicate (if available for comparison) and demonstrating high repeatability.
    cVEMP Waveform Reproducibility (Session 1 vs. Session 2 Mean Correlation)0.847 - 0.8680.915 (Right), 0.916 (Left)Similar or higher than predicate.
    oVEMP Waveform Reproducibility (Session 1 vs. Session 2 Mean Correlation)0.902 - 0.9400.926 (Right), 0.93 (Left)Similar or higher than predicate.

    Reported Device Performance:
    The document states: "The mean correlation values for both cVEMP and oVEMP recordings obtained the SmartEP with VEMP device for both test-retest repeatability and reproducibility are similar or higher than those reported for the ICS Chartr EP 200 with VEMP device." This generally indicates that the SmartEP device met the implicit acceptance criterion of performing comparably or better than the predicate device for VEMP testing.


    Study Details

    1. Sample sizes used for the test set and the data provenance:

      • Study 1 (P1 latency from cVEMP): 215 normal subjects.
      • Study 2 (cVEMP and oVEMP waveforms): 10 adult normal subjects.
      • Data Provenance: The document does not explicitly state the country of origin. It describes the studies as "performed with using the SmartEP with VEMP device," suggesting they were conducted by or for the manufacturer. The studies appear to be prospective to evaluate the device's performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This device is an evoked potential system used to record physiological responses, not an AI diagnostic algorithm that produces an interpretation requiring expert ground truth for classification/detection tasks. The "ground truth" here is the physiological response itself, as measured by the device. Therefore, this question is not directly applicable to the type of device and study described.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      Not applicable, as the "ground truth" is the recorded physiological signal, not an interpretation requiring human adjudication.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, an MRMC comparative effectiveness study was not done. This is a medical device for measuring physiological responses, not an AI-assisted diagnostic tool that humans would use to interpret images or signals. The study focuses on the device's measurement consistency.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      This device is a standalone measurement system. The "performance testing data" section describes the device's inherent ability to record VEMP waveforms repeatably and reproducibly, independent of a specific human interpretation loop for a diagnostic task.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
      The "ground truth" is the physiological VEMP response itself, measured by the device. The studies characterize the device's ability to consistently and accurately capture these responses (repeatability and reproducibility). There isn't a "diagnostic ground truth" in the sense of a disease state confirmed by a gold standard like pathology, as the device is for recording diagnostic data, not making the diagnosis itself. It is stated that "The VEMP modality does not provide a diagnosis. Diagnosis is made by a medical professional."

    7. The sample size for the training set:
      Not applicable. This is not an AI/machine learning device that requires a "training set" to learn from data. It's a measurement instrument.

    8. How the ground truth for the training set was established:
      Not applicable, as there is no "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K090579
    Date Cleared
    2009-07-07

    (126 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartScreener-Plus2 device is indicated for use in the recording and automated analysis of human physiological data (auditory brainstem responses and/or otoacoustic emissions) necessary for the diagnosis of auditory and hearingrelated disorders. The device is especially indicated for use in the screening of infants to determine hearing loss.

    Sites appropriate for use include the well-baby nursery, neonatal intensive care unit (NICU), mother's bedside, outpatient clinic, audiologist's office, or doctor's office. The device is simple to operate. It does not require special technical skills or interpretation of results by the device operator. Basic training with the device is sufficient to learn how to operate it.

    The SmartScreener-Plus2 device can be used for patients of all ages, from children to adults, including infants and geriatric patients. It is especially indicated for use in testing individuals for whom behavioral audiometric results are deemed unreliable, such as infants, young children, and cognitively impaired or uncooperative adults.

    The possible anatomical sites of contact are the patient's ear, ear canal, and/or external skin surrounding the ear (with the contact object being a sound delivery eartip, headphone cushion, or an acoustic ear coupler such as a circumaural ear cup, intra-aural ear bud, or supra-aural ear patch), the patient's head (with the contact object being a bone vibrator), and the patient's head, nape of the neck and shoulder (with the contact object being electrodes that are capable of measuring biopotentials).

    The SmartScreener-Plus2 device is for prescription use.

    Device Description

    The SmartScreener-Plus2 is a noninvasive device used to test for hearing and related neurological abnormalities. The device is especially indicated for use in the screening of infants to determine hearing loss.

    The SmartScreener-Plus2 device records and detects auditory evoked potentials (AEPs) and otoacoustic emissions (OAEs) which are generated in response to a series of auditory stimuli delivered to the patient. AEPs are electrical signals produced by the patient's nourological activity and are picked up by electrodes attached to the patient's skin. The portion of the AEP that originates from the anatomical structures of the inner ear to the auditory brainstem is commonly called the Auditory Brainstem Response (ABR). Otoacoustic emissions are acoustic signals produced by the patient's inner ear and are picked up by a sensitive microphone located in an OAE probe. The SmartScreener-Plus2 device is capable of recording Transient Evoked OAEs (TEOAEs or TrOAEs), and Distortion Product OAEs (DPOAEs)

    The SmartScreener-Plus2 uses advanced signal processing technology to separate AEPs and OAEs from background noise and other physiological activity. The device uses proven statistical algorithms to automatically determine if there is an AEP or OAE response present to the auditory stimuli delivered to the patient. A user, with device administrator approved password security clearance, is able to set the output level of the stimulus. AEP stimulus levels between 30 to 40 dB nHL, and TEOAE stimulus level of 85 dB SPL, and DPOAE stimulus levels of 65 and 55 dB SPL are commonly used for newborn hearing screening applications.

    The SmartScreen-Plus2 device is simple to operate. It does not require any special technical skills or interpretation of results. Basic training is sufficient to learn how to use the device.

    AI/ML Overview

    Here's an analysis of the provided text regarding the SmartScreener-Plus2 device, focusing on acceptance criteria and supporting studies.

    Important Note: The provided document is a 510(k) summary for a medical device submitted to the FDA in 2009. These summaries are designed to establish substantial equivalence to predicate devices, not necessarily to detail comprehensive clinical trial results in the same way a new drug application or a more recent medical device submission might. Therefore, some of the requested information, particularly regarding detailed study methodologies, sample sizes, and expert qualifications for ground truth establishment, is not explicitly present in this type of document.


    Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state quantitative acceptance criteria in terms of sensitivity, specificity, accuracy, or other performance metrics. Instead, "acceptance" is framed around demonstrating substantial equivalence to existing, legally marketed predicate devices. The core acceptance criterion for this 510(k) submission is that the SmartScreener-Plus2 device is as safe and effective as its predicate devices for its intended use.

    The "reported device performance" is largely presented through a comparative table with predicate devices, highlighting similarities in design, materials, indications for use, and compliance with safety standards.

    Acceptance Criterion (Implied)Reported Device Performance
    Safety- Meets Electrical Safety Standards: IEC60601-1, UL2601-1, CSA-C22.2 No. 601.1, AAMI.ES1, EN60601-1, EN60601-1-2, EN60601-2-40, EN55011, EN61000-3-2, EN61000-3-3.
    • Biocompatible (materials are completely biocompatible).
    • No chemicals involved.
    • No mechanical parts in contact with patient. |
      | Effectiveness (Functional Equivalence) | - Records and detects Auditory Evoked Potentials (AEPs) and Otoacoustic Emissions (OAEs) for diagnosis of auditory/hearing-related disorders and infant hearing screening.
    • Uses advanced signal processing and proven statistical algorithms for automated response determination.
    • Operates simply, without special technical skills or result interpretation.
    • Stimulus levels (AEP: 30-40 dB nHL, TEOAE: 85 dB SPL, DPOAE: 65 and 55 dB SPL) are commonly used for newborn screening. |
      | Substantial Equivalence to Predicate Devices | - Target Population: Same (all ages).
    • Where Used: Same (clinical setting).
    • Anatomical Sites: Same (scalp, ear, and other skin sites).
    • Sterility: Same (none required).
    • Biocompatibility: Same (completely biocompatible).
    • Design: Same (main hardware unit connected to patient cables/transducers and portable computer with software).
    • Components/Materials: Same (main hardware unit, transducers, acoustic couplers, electrodes, cables, computer, printer, cart).
    • Human Factors: Same (simple, easy-to-follow instructions and software user interface).
    • Standards: Same (none known explicitly for performance, but electrical/thermal safety standards met).
    • Energy Used/Delivered: Same (stimulation of auditory system).
    • Chemical Safety: Same (no chemicals).
    • Mechanical Safety: Same (no mechanical parts in contact with patient).
    • Thermal Safety: Same (device not thermal in nature). |

    Study Details:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • The document states: "The SmartScreener-Plus2 device specifications were verified and validated."
      • However, no specific details about a clinical test set sample size, data provenance, or whether the study was retrospective or prospective are provided in this 510(k) summary. This type of detail is often summarized or cross-referenced to internal validation reports, which are not publicly disclosed in the 510(k) summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not specified. The document focuses on the device's technical specifications and substantial equivalence, not a detailed clinical outcomes study involving expert interpretation for ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not specified. Since no explicit clinical test set with human ground truth establishment is detailed, no adjudication method is mentioned.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done or described. This device is an automated screening tool, not an AI-assisted interpretation tool for human readers in the traditional sense. It automates the analysis of physiological data (AEPs and OAEs) directly, with a "pass/refer" outcome, meaning it eliminates the need for complex interpretation by the device operator. The text explicitly states: "It does not require any special technical skills or interpretation of results."
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. The device is designed for standalone automated analysis. The "Performance Testing" section states, "The SmartScreener-Plus2 device specifications were verified and validated." While specifics are absent, this verification and validation would inherently test the algorithm's performance in automatically determining the presence or absence of AEP or OAE responses. The device is marketed as not requiring "interpretation of results by the device operator," indicating its standalone analytical function.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The document mentions the device uses "proven statistical algorithms to automatically determine if there is an AEP or OAE response present." For this type of device, ground truth would typically be established by:
        • Physiological standards/biomarkers: The presence or absence of a physiological AEP or OAE response based on established audiology and neurophysiology principles, often confirmed by expert audiologists or neurophysiologists using gold-standard diagnostic equipment.
        • Clinical diagnosis/outcomes: Correlation of screening results with definitive diagnostic tests (e.g., full audiometry, follow-up examinations) to determine actual hearing status.
      • The specific method for establishing ground truth for validation is not detailed in this summary.
    7. The sample size for the training set:

      • Not specified. This pertains to the internal development and calibration of the "advanced signal processing technology" and "proven statistical algorithms." The 510(k) summary does not include details on the training set for these algorithms.
    8. How the ground truth for the training set was established:

      • Not specified. Similar to the training set size, the method for establishing ground truth for the training of the device's algorithms is not disclosed in this regulatory summary. It would typically involve large datasets of AEP and OAE recordings from individuals with known hearing status, often categorized by expert review or correlation with comprehensive diagnostic audiometric evaluations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K073384
    Date Cleared
    2008-02-28

    (87 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Smart Coupler is a passive, ear-coupling, sound-delivery device accessory intended for short-term, non-invasive use to support auditory evaluation studies (including use with audiometers, auditory evoked potential and otoacoustic emissions devices) on patients of all ages. The Smart Coupler is a biocompatible, non-sterile, single patient use, disposable device.

    Device Description

    The Smart Coupler is a non-invasive, cutaneous, passive, sound-delivery device accessory that provides a coupling interface between the patient's ear and the auditory stimulator used during audiometric testing. The device is a biocompatible, non-sterile, single patient use, disposable device.

    The anatomical sites of contact for the Smart Coupler are the external skin surrounding the ear (circumaural model), the external skin of the outer ear (supra-aural model), or the external skin of the ear canal (intra-aural model) with the contact object being a thin layer of medical-grade, biocompatible adhesive hydrogel, used for adhering the Smart Coupler device to the patient.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary for the "Smart Coupler" device, focusing on acceptance criteria and the study to prove it, as requested.

    It's important to note that the provided document is a 510(k) summary for a device accessory (an acoustic coupler), not a complex AI-powered diagnostic device. Therefore, many of the requested elements like "multi reader multi case (MRMC) comparative effectiveness study," "standalone algorithm performance," "number of experts," and detailed "training set" information are not applicable or described in the context of this type of submission. The focus for this device is on safety, biocompatibility, and functional equivalence to a predicate device, rather than diagnostic accuracy.


    Acceptance Criteria and Device Performance for Smart Coupler (K073384)

    The "Smart Coupler" is a passive device accessory. The "acceptance criteria" for a device like this primarily revolve around demonstrating substantial equivalence to a predicate device in terms of safety and effectiveness. This is achieved by showing similar performance characteristics and demonstrating that any new features (like an additional model or specific materials) do not raise new questions of safety or effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit from Submission)Reported Device Performance (Smart Coupler)
    BiocompatibilityMeet ISO 10993 standards for short-term skin contact (cytotoxicity, skin irritation, skin sensitization).All materials in contact with the patient successfully passed biocompatibility testing criteria specified in ISO 10993 for short-term skin contact.
    Functional EquivalenceProvide a coupling interface for auditory stimulators for audiometric testing (same intended use as predicate).Supports auditory evaluation studies (audiometers, auditory evoked potential, otoacoustic emissions devices). Same intended use as the predicate device.
    Material CompositionBiocompatible, non-sterile, single-patient use, disposable. Materials similar to predicate (foam, flexible elastomeric material, adhesive).Biocompatible, non-sterile, single-patient use, disposable. Similar materials (flexible elastomeric material, adhesive hydrogel) used, with the note "expect no foam" compared to predicate.
    Safety (General)No new issues of safety or effectiveness compared to predicate.Device is passive, does not consume/produce energy, no chemicals involved, no mechanical parts, not electrical/thermal in nature. Results of performance testing were as expected, raising no new issues of safety or effectiveness.
    DesignProvide coupling for circumaural or intra-aural application (at least, similar to predicate).Available in circumaural, intra-aural, and supra-aural models. Offers an additional supra-aural model compared to the predicate's two models.
    Human FactorsSimple, easy-to-follow instructions.Same as predicate (implied to be simple, easy-to-follow instructions).

    2. Sample Size Used for the Test Set and Data Provenance

    The submission primarily refers to "performance testing of prototype devices to specifications."

    • Test Sample Size: Not explicitly stated as a number of devices or subjects. The context suggests that the testing was primarily focused on material properties (biocompatibility) and functional validation, not clinical trials with a defined "test set" sample size in the traditional sense for diagnostic accuracy.
    • Data Provenance: Not specified regarding country of origin or retrospective/prospective. Given the nature of the device (acoustic coupler), it is highly likely that the biocompatibility testing was conducted in a laboratory setting. Functional testing would have involved prototypes.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable in the context of this device. The "Smart Coupler" is an accessory; its "ground truth" is its ability to meet material safety standards and functionally couple an auditory stimulator. This is assessed via technical standards (ISO 10993) and engineering verification, not by expert interpretation of output.

    4. Adjudication Method for the Test Set

    • Not applicable. There is no "adjudication" necessary for this type of device's performance assessment.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, Effect Size

    • No. An MRMC study is relevant for AI-powered diagnostic tools where human readers are interpreting images or data. This device is a passive accessory and does not involve human interpretation of its output in a diagnostic capacity.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This device does not contain an algorithm. It is a passive, mechanical accessory.

    7. The Type of Ground Truth Used

    The "ground truth" for this device relates to its fundamental safety and functional attributes:

    • Biocompatibility: Established by adherence to ISO 10993 standards (e.g., cell viability assays for cytotoxicity, visual assessment for skin irritation/sensitization).
    • Functional Performance: Established by engineering specifications and physical testing to confirm its ability to serve as a coupling interface for auditory stimulators, consistent with the intended use and predicate device. This would likely involve acoustic measurements and fit testing.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not use an algorithm or machine learning, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As there is no training set, there is no ground truth establishment for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K070608
    Date Cleared
    2007-07-18

    (135 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SmartEP is an evoked response testing and diagnostic device, that is capable of eliciting, acquiring, and measuring auditory, somatosensory, and visual evoked potential data, as well as providing nerve stimulation and monitoring.

    The intended use of the SmartEP device is to objectively record evoked responses from patients of all ages upon the presentation of sensory stimuli. The product is indicated for use as a diagnostic aid and adjunctive tool in sensory related disorders (i.e., auditory, somatosensory, visual) and in surgical procedures for inter-operative nerve monitoring.

    Device Description

    SmartEP is an evoked response system that is capable of eliciting, acquiring, and measuring auditory, somatosensory, and visual evoked potential data, as well as providing nerve stimulation/monitoring.
    The feature modifications described in this 510(k) are to incorporate additional/expanded indications to the SmartEP device for somatosensory evoked potential, visual evoked potential, and nerve stimulation testing to the previously FDA cleared indications for auditory evoked potential testing. Specifically, additional hardware and software features have been added to the original SmartEP device system to objectively acquire somatosensory/visual evoked potentials upon the presentation of a somatosensory/visual stimulus, and to provide nerve stimulation/monitoring functionality.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the SmartEP device, seeking expanded indications for Somatosensory Evoked Potential (SEP), Visual Evoked Potential (VEP), and Nerve Stimulation/Monitoring. The core of the information regarding acceptance criteria and supporting studies is based on demonstrating substantial equivalence to predicate devices.

    Here's a breakdown of the requested information:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the equivalence of key specifications and intended use to legally marketed predicate devices. The "reported device performance" is a demonstration that the SmartEP's characteristics meet or are comparable to those of the predicate devices. The tables provided (Table E-1 and Table E-2) serve as the primary evidence for this.

    Table 1: Acceptance Criteria and Reported Device Performance for SEP & VEP (Summarized from Table E-1)

    ParameterAcceptance Criteria (Predicate Device: Nicolet Viking II K890495)Reported SmartEP Performance (Device Under Review)
    Intended UseSEP: Stimulate, record, and process somatosensory evoked potentials. VEP: Stimulate, record, and process visual evoked potentials.Same
    Indications for UseRecording and analysis of physiological data necessary for the diagnosis of somatosensory and visual related disorders.Same
    Target PopulationAll AgesSame
    DesignExternal box housing circuitry connected to CPU via RS232 connection.External box housing circuitry connected to personal computer via a USB connection. (Note: This is a design difference, but implied functionally equivalent).
    MaterialsAssorted electrical components, circuit boards, SEP electrical stimulation box, SEP electrical stimulator probe, video monitor, and electrodes.Assorted electrical components, circuit boards, SEP electrical stimulation box, SEP electrical stimulator probe, visual stimulation LED array, and electrodes.
    SterilityNone RequiredSame
    BiocompatibilityCompletely BiocompatibleSame (specifically for SEP Stimulator Probe: Type-302 stainless steel, meets ISO-10993)
    Anatomical SitesSEP: Upper/lower limbs and head. VEP: Scalp.Same
    Energy DeliverySEP: Stimulation of upper or lower limbs with surface electrical signals. VEP: Stimulation of eyes with visual light patterns.Same
    Where UsedClinical SettingSame
    SafetyConforms to UL544 & IEC 60601-1Meets EN 60601-1
    Patient IsolationType BF (IEC 60601-1) (for data), Fiber Optic Signal Link.Type BF (IEC 60601-1) (for data), Fiber Optic Signal Link (for amp), Medical-grade power supplies, 4000Vdc, 10 MegaOhms (for power), 3500Vdc, 10 TeraOhms (for probe), Isolation transformers, Isolation amplifiers.
    Somatosensory Stimuli - TypeConstant Current or VoltageConstant Current
    Somatosensory Stimuli - Current Levels0 - 100 mA, 0 - 400V (into 4000 Ohms load)0 - 100 mA (400 Volt maximum) (into 4000 Ohms load)
    Visual Stimuli - Output IlluminationNone SpecifiedRed LED Light (Class I AEL Level), Peak Wavelength: 625 nm, Max Accessible Power:
    Ask a Question

    Ask a specific question about this device

    K Number
    K063402
    Date Cleared
    2006-12-11

    (32 days)

    Product Code
    Regulation Number
    882.1320
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IHS TM-Wick Electrode is intended for non-invasive use in acquiring electrocochleography (ECochG) data on patients of all ages, for the purpose of monitoring and recording in the diagnosis of cochlear and auditory disorders. The electrodes are biocompatible, non-sterile, single patient use, disposable devices.

    Device Description

    IHS TM-Wick Electrodes are non-invasive, cutaneous devices used in the acquisition of electrocochleography (ECochG) signals from the tympanic membrane for the purpose of monitoring and recording. These extra-tympanic electrodes are inserted into the ear canal to make contact with the tympanic membrane for acquiring the ECochG data. The electrodes are biocompatible, non-sterile, single patient use, disposable devices.

    The IHS TM-Wick Electrode is a non-invasive, TM ECochG electrode which is biocompatible, prewired, non-sterile, for single patient use, and disposable. The electrode features a fine Silver/Silver Chloride (Ag/AgCl) wire attached to a soft cotton tip (i.e., wick) pre-soaked in a highly conductive, low impedance electrode gel/saline solution, which is hypoallergenic, non-irritating, bacteriostatic, water soluble, non-staining, and non-gritty. The conductive gel reduces the impedance of the contact between the electrode surface and the skin. The Ag/AgCl wire is encased inside of a thin, soft, flexible silicone tubing which is resilient, stretchable, non-corrosive, and inert. This tubing's small size and characteristics provide for ease of insertion and removal, as well as patient comfort and safety.

    The other end of the Ag/AgCl wire is permanently adhered to an electrode cable lead wire. The solder joint is then covered by a heat shrink tubing so as not to allow exposure of the lead wire, as well as to provide strain relief. The electrode cable lead is terminated on the opposite end using two industry standard 1.5mm (0.06in) molded DIN safety socket connectors, which are subsequently used to interface to the monitoring device. For electrical safety, these connectors do not allow accidental connection to an A/C wall electrical outlet.

    AI/ML Overview

    The provided 510(k) summary for the IHS TM-Wick Electrode does not describe a clinical study in the traditional sense, with specific acceptance criteria related to accuracy, sensitivity, or specificity of diagnostic performance for the device itself. Instead, it relies on demonstrating substantial equivalence to a predicate device (Bio-Logic TM-ECochGtrode, K944314) by comparing technological characteristics and ensuring compliance with established safety standards.

    Therefore, the "acceptance criteria" and "device performance" in this context refer to the device meeting the requirements established for substantial equivalence through a comparison of its specifications and materials with those of the predicate device, and demonstrating compliance with relevant biocompatibility and electrical safety standards.

    Here's an breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a substantial equivalence submission for an electrode, the "acceptance criteria" are predominantly about meeting safety and functional equivalence to the predicate device, rather than diagnostic accuracy metrics.

    Acceptance Criteria (Implied from comparison and standards)Reported Device Performance (IHS TM-Wick Electrode)
    Intended Use (Same as predicate)Non-invasive acquisition of ECochG data on patients of all ages, for monitoring & recording in the diagnosis of cochlear and auditory disorders. (Same as predicate)
    Indications for Use (Same as predicate)Monitoring & recording in the diagnosis of cochlear and auditory disorders. (Same as predicate)
    Target Population (Same as predicate)All Ages (Same as predicate)
    Design Equivalence/FunctionalityConstruction: Thin, bare, conducting Ag/AgCl wire insulated inside flexible silicone tubing, with end attached to a soft cotton tip pre-soaked in conductive gel/saline solution. Pre-wired/Assembled: Fully assembled and pre-wired in individual packaging. Connectivity: Electrode cable lead terminated with two industry standard 1.5mm molded DIN safety socket connectors for interface to monitoring device. Size: Silicone tubing is approximately 1/3 smaller in size diameter than the predicate's tubing (implying enhanced ease of insertion/removal and patient comfort).
    Materials Equivalence/BiocompatibilitySilicone tubing, cotton wick tip, conductive gel/saline, Silver/Silver-Chloride wire, electrode cable lead, molded DIN safety socket connectors. Biocompatibility: All patient-contacting components meet ISO 10993-1, -5 (Cytotoxicity), -10 (Skin irritation & sensitization) standards.
    Sterility (Same as predicate)None Required (Non-sterile). (Same as predicate)
    Safety - ElectricalElectrode lead wire assembly is in compliance with FDA 21 CFR 898 (Performance Standard for Electrode Lead Wires and Patient Cables). DIN safety connectors prevent accidental connection to A/C electrical outlets.
    Safety - Patient ComfortSilicone tubing's small size and characteristics provide for ease of insertion and removal, as well as patient comfort and safety. Conductive gel is hypoallergenic, non-irritating, bacteriostatic, water soluble, non-staining, and non-gritty.
    Disposable/Single Use (Same as predicate)Single patient use, disposable devices. (Same as predicate)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not describe a test set or data provenance in the way one would for a clinical performance study involving patient data. The basis for safety and effectiveness is a comparison to a predicate device and compliance with established standards (ISO 10993, 21 CFR 898). No human subject data (prospective or retrospective) is mentioned in relation to proving the performance of the device for ECochG signal acquisition in this summary.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. No clinical test set with a ground truth established by experts is described in this 510(k) summary.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or adjudication method is described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No. The document does not describe a MRMC study.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    Not applicable. This is a physical electrode, not a software algorithm.

    7. The Type of Ground Truth Used

    Not applicable in the context of diagnostic accuracy. The "ground truth" here is the established safety and performance profile of the predicate device and the biocompatibility and electrical safety standards. The new device is shown to be equivalent or compliant with these.

    8. The Sample Size for the Training Set

    Not applicable. No training set is described as this is not an AI/ML device related to diagnostic accuracy based on a dataset.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. No training set is described.

    In summary, this 510(k) submission for the TM-Wick Electrode focuses on demonstrating substantial equivalence to a legally marketed predicate device (Bio-Logic TM-ECochGtrode) by comparing their intended use, indications, target population, design, materials, and safety characteristics. The "study" involves a detailed comparison of specifications and confirmation of compliance with relevant engineering and biocompatibility standards, rather than a clinical trial with performance metrics like sensitivity and specificity.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061443
    Date Cleared
    2006-06-23

    (30 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Smart USBLite device system is for the recording of auditory evoked potential, otoacoustic emissions, & auditory steady-state evoked potential data. The product is intended to be used as a diagnostic aid in auditory and hearing related disorders, as an objective measure of cochlear function, and as an adjunctive tool in the estimation of behavioral hearing thresholds on patients of all ages.

    Device Description

    Smart USBLite is an auditory evoked potential and otoacoustic emission system that is capable of recording and measuring auditory evoked potential, otoacoustic emission, and auditory steady-state evoked potential data. The Smart USBLite system combines the following FDA 510(k) previously cleared devices into a repackaged, smaller-sized, single integrated unit: SmartEP (K904926), SmartScreener (K925648), SmartOAE (K964426), SmartTrOAE (K023859), SmartEP-ASSR (K031051), and Opti-Amp DC-Powered (K052060).

    AI/ML Overview

    The Smart USBLite is an auditory evoked potential and otoacoustic emission system that records and measures auditory evoked potential, otoacoustic emission, and auditory steady-state evoked potential data. It combines several previously cleared predicate devices into a repackaged, smaller-sized, single integrated unit. The device is intended as a diagnostic aid for auditory and hearing-related disorders, an objective measure of cochlear function, and an adjunctive tool for estimating behavioral hearing thresholds across all patient ages.

    The study presented does not include specific acceptance criteria with quantifiable metrics for performance, nor does it detail a standalone study proving the device meets particular performance thresholds. Instead, the submission relies on demonstrating substantial equivalence to predicate devices through a comparison of technological characteristics and specifications. The "acceptance criteria" appear to be satisfied by the device's adherence to the same intended use, methodologies, and performance specifications as the established predicate devices, coupled with certification for electrical safety and electromagnetic compatibility.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit quantifiable acceptance criteria with corresponding reported performance metrics are not provided in the document. The submission focuses on claiming substantial equivalence to predicate devices. The "performance" is implicitly accepted if the new device's specifications are the same as the predicate devices or within acceptable limits for a repackaged unit.

    However, the closest representation of "acceptance criteria" and "reported device performance" can be extrapolated from the "Specifications" table by comparing the "Predicate Devices" column to the "Device Under Current 510(k) Review (Smart USBLite)" column. The acceptance criterion for each parameter is implicitly the specification of the predicate device, and the reported performance is the "Same" or adjusted value for the Smart USBLite.

    ParameterAcceptance Criteria (Predicate Device Specification)Reported Device Performance (Smart USBLite)
    Intended UseStimulate, record, & process auditory evoked potentials, auditory evoked potential screening, elicitation and measurement of otoacoustic emissions, and auditory steady-state evoked potentials. Diagnostic aid for auditory and hearing related disorders, objective measure of cochlear function, adjunctive tool for estimation of behavioral hearing thresholds.Same
    Indications for UseRecording & analysis of physiological data for diagnosis of auditory/hearing-related disorders; determination of cochlear function.Same
    Target PopulationAll AgesSame
    DesignExternal box housing circuitry connected to personal computer via USB.Same
    MaterialsAssorted electrical components, circuit boards, sound transducers, microphone, electrodes, disposable probe ear tips.Same
    SterilityNone requiredSame
    BiocompatibilityCompletely BiocompatibleSame
    Anatomical SiteEar canal, arms, and headSame
    Energy DeliveryStimulation of the ear with sound signalsSame
    Where UsedClinical SettingSame
    SafetyMeets EN 60601-1Same
    Operating Temperature15°C to 40°CSame
    Storage Temperature5°C to 50°CSame
    Relative Humidity15% to 90% (at 40°C Non-Condensing)Same
    Atmospheric PressureNone SpecifiedSame
    Mode of OperationContinuousSame
    Type of ProtectionClass I (IEC 601)Same
    Degree of ProtectionType B (IEC 601)Same
    Protection Against FluidsIPX0 - Ordinary Equipment (IEC 601)Same
    Degree of MobilityPortable EquipmentSame
    Vibration and ShockN/ASame
    Expected Lifetime10 Years from date of manufactureSame
    Power SupplyExternal medical grade power supply (3 Outputs: +15V, -15V, +5V) OR Internal medical grade power supply (3 Outputs: +15V, -15V, +5V)External medical grade power supply (1 Output: +5V (2A))
    Rated Input Current1.3 Amps or 2.0 Amps0.5 Amps
    Rated Frequency50-60 HzSame
    Rated Input Voltage100-240 VoltsACSame
    Rated Max Input Power40 Watts10 Watts
    Fuse TypeN/A (External Power Supply) or Time Lag Fuse (IEC 60127-2 compliant)N/A (External Power Supply)
    Fuse RatingN/A (External Power Supply) or 1.6 Amps, 250 VoltsACN/A (External Power Supply)
    Patient Isolation4000Vdc, 10 Mega OhmsSame
    A/D Resolution16 bitSame
    Artifact RejectionProgrammableSame
    Input ChannelsSmartEP, SmartScreener, SmartEP-ASSR: 1-4 Channels; SmartOAE, SmartTrOAE: 1-2 ChannelsSmart USBLite (EP, Screener, ASSR): 1 Channel; Smart USBLite (DPOAE, TrOAE): 1 Channel
    GainSmartEP, SmartScreener, SmartEP-ASSR: Variable (30k-300k); SmartOAE: Fixed (100); SmartTrOAE: Fixed (100)Smart USBLite (EP, Screener, ASSR): Fixed (100k); Smart USBLite (DPOAE): Same; Smart USBLite (TrOAE): Same
    Lowpass FilterSmartEP, SmartScreener: Variable (30-5000 Hz); SmartOAE: 100 kHz; SmartTrOAE: 5000 Hz; SmartEP-ASSR: Variable (30-5000 Hz)Smart USBLite (EP, Screener): Fixed (1500 Hz); Smart USBLite (DPOAE): Same; Smart USBLite (TrOAE): Same; Smart USBLite (ASSR): Fixed (300 Hz)
    Highpass FilterSmartEP, SmartScreener, SmartEP-ASSR: Variable (1-500 Hz); SmartOAE: 200 Hz; SmartTrOAE: 450 HzSmart USBLite (EP, Screener, ASSR): Fixed (30 Hz); Smart USBLite (DPOAE): Same; Smart USBLite (TrOAE): Same
    Filter SlopeSmartEP, SmartScreener, SmartEP-ASSR: 6 dB/octave; SmartTrOAE: 24 dB/octaveSmart USBLite (EP, Screener, ASSR): 12 dB/octave; Smart USBLite (TrOAE): Same
    Notch Filter50/60HzSame
    Noise LevelSmartEP, SmartScreener, SmartEP-ASSR: 0.33 μV RMS (1-3000 Hz), 0.12 μVpp for 1024 averaged sweeps; SmartOAE, SmartTrOAE: 40 μV RMSSmart USBLite (EP, Screener, ASSR): 0.13 μV RMS (30-1500 Hz), 0.1 μVpp for 1024 averaged sweeps; Smart USBLite (DPOAE, TrOAE): Same
    Input ImpedanceOpti-Amp DC Powered: 5 MegaOhmsSame
    CMR RatioOpti-Amp DC Powered: 117 dB at 60Hz, 110 dB at 1kHzSame
    Auditory Stimuli TypesClicks, Pure Tones, Multifrequency Stimuli, Tone BurstsSame
    Duration25-5000 µsec or ContinuousSame
    EnvelopesLinear, Blackman, Gaussian, Hanning, Rectangular, Triangular, Trapezoidal, Exact Blackman, Cosine, Cosine Squared, Cosine CubedSame
    Intensity0-125 dB SPL (EP/Screener), 55-80 dB SPL (OAE), 60-90 dB SPL (TrOAE), 0-80 dB SPL (ASSR)Same
    Repetition Rate1-100 Hz or ContinuousSame
    Test Frequencies500-16,000 HzSame
    PresentationMonaural or Binaural (EP/Screener/ASSR); Monaural (OAE/TrOAE)Same
    MaskingWhite Noise ProgrammableSame
    TransducersInsert Earphones, Bone Vibrator, Headphones, Sound Field, Ear ProbeInsert Earphones, Headphones, Ear Probe
    Analysis WindowVariable (up to 2.5 msec) or Fixed (25.6 msec, 102.4 msec, 2.5-22.5 msec)Same
    Artifact Rejection ThresholdUser Selectable or Automated/FixedSame
    Measured ValuesResponse Level, Noise Level, Signal to Noise Ratio, Response Latency, FrequencySame
    Computer TypePersonal ComputerSame
    Operating SystemMicrosoft Windows 2000 or XPSame
    Interface ConnectionUSB (Universal Serial Bus)Same

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a "test set" in the context of clinical data or patient samples. The study is a technical comparison for substantial equivalence. The "test" in this context refers to engineering verification and validation activities to ensure the new device meets its design specifications and electrical safety standards; these do not typically involve patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. No "test set" involving expert-established ground truth for clinical performance is described. The comparison is based on technical specifications and adherence to safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no clinical test set requiring adjudication of ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a measurement system, not an AI-assisted diagnostic tool for human readers, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not an algorithm-only device. It is a hardware and software system for audiological measurements. The "study" for this submission is a declaration of conformance to design controls and safety standards, and a comparison of technical specifications to predicate devices. It states that "all verification and validation activities have been performed for the Smart USBLite device," and that "The results have demonstrated that all predetermined acceptance criteria have been met." However, no details of these specific verification and validation activities are provided in the summary beyond adherence to general standards (e.g., EN60601-1).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the clinical sense. The "ground truth" for the device's technical specifications and safety would be established by engineering standards, testing protocols, and comparison to the predicate devices' established performance. For example, the noise level specifications are a technical measurement, not a clinical "ground truth."

    8. The sample size for the training set

    Not applicable. This device is not an AI/ML device that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable. This device is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052060
    Date Cleared
    2005-08-31

    (33 days)

    Product Code
    Regulation Number
    882.1835
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Opti-Amp DC-Powered is as a physiological signal amplifier. It is intended to be used by trained personnel in a hospital, nursery, clinic, audiologist's office or other appropriate setting.

    Device Description

    Opti-Amp DC-Powered is a bio-amplifier testing device (made up of a transmitter and receiver) that is capable of acquiring evoked potentials.

    AI/ML Overview

    The provided text is a 510(k) summary for the Opti-Amp DC-Powered device, which is an evoked potential amplifier. It describes the device, its intended use, and its technological characteristics as compared to a predicate device.

    However, the document does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies.

    The summary focuses on demonstrating substantial equivalence to a predicate device (Opti-Amp K914876) based on intended use and technological characteristics, specifically noting a change in power supply from batteries to a medical-grade power supply with additional protection. It does not include the type of performance data and studies requested in your prompt.

    Therefore, I cannot provide the requested table and study details.

    Ask a Question

    Ask a specific question about this device

    K Number
    K031051
    Date Cleared
    2004-02-20

    (324 days)

    Product Code
    Regulation Number
    882.1900
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the SmartEP-ASSR product is for the recording of auditory steadystate evoked potential data. The product is intended to be used as a diagnostic aid in auditory and hearing related disorders and as an adjunctive tool in the estimation of behavioral hearing thresholds.

    Device Description

    SmartEP-ASSR is a auditory evoked potential testing device that is capable of measuring auditory steady state responses (ASSR).

    AI/ML Overview

    The provided text is a 510(k) summary for the SmartEP-ASSR device, which is an auditory evoked potential testing device. The summary primarily focuses on demonstrating substantial equivalence to a predicate device (Bio-logic MASTER Evoked Response System, K021895) through a comparison of technical specifications.

    Based on the provided information, there is no detailed study described that establishes explicit acceptance criteria and proves the device meets those criteria in a quantitative performance study. The submission relies on demonstrating substantial equivalence by comparing technical parameters to a legally marketed predicate device. This type of submission usually implies that if the technical specifications are comparable or superior, and there are no new questions of safety or effectiveness, then the device is considered to meet the requirements for marketing.

    Here’s a breakdown of the requested information, indicating where the text provides details and where it does not:


    1. Table of Acceptance Criteria and Reported Device Performance

    As noted above, no explicit acceptance criteria for diagnostic accuracy or performance metrics (like sensitivity, specificity, or agreement rates) are stated, nor is there a study detailing such performance. The provided table focuses on technical specifications for comparing the SmartEP-ASSR to its predicate device. This comparison is the primary method used to argue for "substantial equivalence."

    ParameterPredicate Device (Bio-Logic Master, K021895)SmartEP-ASSR (Device Under Review)Acceptance Criterion/Performance (Implied)
    Data Acquisition
    A/D Resolution16 bit16 bitAt least 16 bit
    Artifact RejectionProgrammableProgrammableProgrammable
    Amplifiers
    Channels1 Channel Optically Isolated1-4 Channels Optically IsolatedAt least 1 channel optically isolated
    Gain10kVariable (30k-300k)Comparable or improved range of gain
    Filters
    Slope12 dB/octave6 dB/octaveSpecified filter slope
    LP (Low Pass)1.5k, 3k, 10k, 20k Hz30, 100, 300, 500, 1k, 1.5k, 3k, 5k HzComparable or wider range of LP filters
    HP (High Pass)0.1, 0.3, 1, 3, 10, 30, 100, 300Hz1, 10, 30, 50, 100, 150, 300, 500 HzComparable or wider range of HP filters
    Notch50/60Hz50/60Hz50/60Hz
    Digital Filters1-200HzUser Selectable (1 -5k Hz)Comparable or wider range of digital filters
    Noise Level0.45 μV RMS (10-3k Hz)0.33 μV RMS (1-3k Hz)Noise level ≤ 0.45 μV RMS (or better)
    Input Impedance100M Ohms5M OhmsSpecified input impedance
    CMR Ratio110 dB at 50/60 Hz117 dB at 60 Hz and 110 dB at 1k HzCMR Ratio ≥ 110 dB
    Impedance Test
    Signal20 Hz Sinewave1k Hz SinewaveSpecified impedance test signal
    Auditory Stimuli
    PresentationMonaural or BinauralMonaural or BinauralMonaural or Binaural
    Number of Frequencies1-41-8At least 1-4 frequencies
    Test Frequencies500, 750, 1k, 1.5k, 2k, 3k, 4k, 6k, 8k HzUser Selectable: Clicks, Pure Tones and Multifrequency Stimuli (500Hz-8k Hz)Comparable range of test frequencies
    TypesSinewaveSinewave, Tone BurstAt least Sinewave
    EnvelopesLinear, Blackman, Gaussian, Hanning EnvelopesLinear, Blackman, Gaussian, Hanning, Rectangular, Triangular, Trapezoidal, Exact Blackman, Cosine, Cosine Squared, Cosine CubedComparable or wider range of envelopes
    Intensity0-125 dB SPL (132 dB Optional)0-125 dB SPL0-125 dB SPL
    MaskingWhite Noise ProgrammableWhite Noise ProgrammableWhite Noise Programmable
    TransducersTDH Earphones, Insert Earphones, Bone Conduction, Sound FieldTDH Earphones, Insert Earphones, Bone Conduction, Sound Field, OAE ProbeComparable range of transducers
    Computer Requirements
    Computer TypePersonal ComputerPersonal ComputerPersonal Computer
    Operating SystemMicrosoft WindowsMicrosoft Windows 98SE, ME, 2000, XPMicrosoft Windows (compatible)
    Interface ConnectionSerialUSB (Universal Serial Bus)Specified interface connection

    The "acceptance criteria" can be inferred as the SmartEP-ASSR meeting or exceeding the technical specifications of the predicate device, thereby demonstrating substantial equivalence for its intended use. For instance, the noise level being 0.33 μV RMS is "better" than the predicate's 0.45 μV RMS, which would contribute to showing equivalence or improvement.


    2. Sample size used for the test set and the data provenance
    The document does not describe any clinical test set or data derived from study participants for performance evaluation. The submission relies solely on comparing technical specifications to a predicate device.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable, as no clinical test set with ground truth established by experts is described.


    4. Adjudication method for the test set
    Not applicable, as no clinical test set requiring adjudication is described.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an auditory evoked potential testing device, not an AI-assisted diagnostic imaging or interpretation tool for human readers. No MRMC study is mentioned.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. The SmartEP-ASSR is a medical device for measuring physiological responses, not an algorithm being tested for standalone diagnostic performance in the absence of human interpretation. Its function is to acquire and process evoked potential data for trained personnel to interpret.


    7. The type of ground truth used
    Not applicable, as no clinical performance study requiring ground truth is described. The "ground truth" equivalent for this type of submission is the established safety and effectiveness of the predicate device and the technical measurements demonstrating the new device performs similarly.


    8. The sample size for the training set
    Not applicable, as this device does not utilize a machine learning algorithm that requires a training set in the context of the provided document.


    9. How the ground truth for the training set was established
    Not applicable, as no training set for a machine learning algorithm is discussed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023859
    Device Name
    SMARTTROAE
    Date Cleared
    2003-01-08

    (49 days)

    Product Code
    Regulation Number
    874.1050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the SmartTrOAE device is to measure otoacoustic emissions, which allow the operator to get information on cochlear function without requiring a subjective response from the person being tested with the device. It is intended to be used by trained personnel in a hospital, nursery, clinic, audiologist's office or other appropriate setting.

    Device Description

    SmartTrOAE is an otoacoustic emissions testing device that is capable of measuring transient and spontaneous otoacoustic emissions produced by the inner ear.

    AI/ML Overview

    This 510(k) summary does not contain the detailed information necessary to fully answer all aspects of your request. The document focuses on demonstrating substantial equivalence to a predicate device rather than providing a comprehensive study protocol and results for the SmartTrOAE device independently.

    However, I can extract the available information regarding acceptance criteria and device performance based on similarity to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" in terms of specific thresholds for performance metrics. Instead, it demonstrates substantial equivalence by comparing the technological characteristics of the SmartTrOAE to the predicate device, ILO88 (K890124). The implicit acceptance criterion is that the SmartTrOAE's performance in these listed parameters is similar enough to the predicate to be considered substantially equivalent.

    ParameterAcceptance Criteria (Predicate Device ILO88 - K890124)Reported Device Performance (SmartTrOAE)
    Stimulus TypeClickClick or Tone
    Stimulus Duration100 µsec25 - 5000 µsec
    Stimulus Intensity90 dB SPL60 - 90 dB SPL
    Repetition Rate50 Hz1 - 50 Hz
    MicrophoneOtodynamicsEtymotic Research ER 10B (K930553) or Etymotic Research ER 10D (K011114)
    Measured ValuesStimulus Level (dB SPL), Response Level (dB SPL), Noise Level (dB SPL), Signal to Noise Ratio (dB SPL)Stimulus Level (dB SPL), Response Level (dB SPL), Noise Level (dB SPL), Signal to Noise Ratio (dB SPL)
    Bandlimit Filter500 - 5000 Hz500 - 5000 Hz
    Analysis Window2.5 to 20 msec post stimulus2.5 - 22.5 msec post stimulus
    Freq. Spectrum0 - 6000 Hz0 - 6000 Hz
    Sweeps1 - 2601 - 260
    Rejection ThresholdUser SelectableUser Selectable

    2. Sample size used for the test set and the data provenance:

    This document does not report on a specific test set, its sample size, or data provenance (country of origin, retrospective/prospective). The substantial equivalence is based on comparing technical specifications and intended use with a predicate device. There is no mention of a clinical study with a test set for the SmartTrOAE's performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This document does not report on any ground truth establishment by experts for a test set.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This document does not report on an adjudication method for a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This document does not report on an MRMC comparative effectiveness study. The SmartTrOAE is an otoacoustic emissions testing device, which typically provides objective measurements (otoacoustic emissions) and does not involve "human readers" in the interpretation of complex images or data that an AI might assist with.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This document does not explicitly detail a standalone performance study. The device itself is an "otoacoustic emissions testing device," implying it takes objective measurements. The intended use "without requiring a subjective response from the person being tested" suggests a standalone measurement capability. However, detailed performance metrics from such a study are not provided. The comparison is solely based on technical specifications.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This document does not describe a type of ground truth used as it does not present a formal clinical study for the SmartTrOAE. The basis for equivalence is the similarity in technical characteristics and intended use to a legally marketed predicate device.

    8. The sample size for the training set:

    This document does not report on a training set sample size. This type of device (otoacoustic emissions) typically uses established physiological principles and signal processing, rather than machine learning models that require a "training set" in the conventional sense of AI/ML devices.

    9. How the ground truth for the training set was established:

    This document does not report on the establishment of ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964426
    Device Name
    SMARTOAE
    Date Cleared
    1997-02-03

    (90 days)

    Product Code
    Regulation Number
    874.1050
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    INTELLIGENT HEARING SYSTEMS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SmartOAE device is intended to be used for the purpose of determining the distortion product response of the human cochlea. The device is intended to be used by hearing health professionals to determine cochlear function. The test administered by the device does not require any patient behavioral response, such as pushing a button switch. The device delivers sound signals into the outer ear and subsequently records the sound in the outer ear which contain sounds produced by normal cochleas. The device also measures the level of sound near the distortion product frequency to estimate the noise level present while the test was performed.

    Device Description

    Not Found

    AI/ML Overview

    This looks like a 510(k) clearance letter from the FDA for a device called "SmartOAE." While it describes the device's intended use, it does not contain the detailed information about acceptance criteria, study design, or performance metrics that you've requested.

    The document primarily states that the device is substantially equivalent to legally marketed predicate devices and is subject to general controls. It doesn't include a study summary or the specific data you're looking for.

    Therefore, I cannot provide the requested information based on the provided text. To get this information, you would typically need to refer to the original 510(k) submission documentation, which is usually more extensive and includes the technical details and study reports.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1