Search Results
Found 103 results
510(k) Data Aggregation
(267 days)
Integra LifeSciences Production Corporation
For use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.
The Libertís™ Shunt Catheter includes a ventricular and distal (peritoneal) drainage catheter that are used as part of a cerebrospinal fluid (CSF) shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system. The ventricular catheter diverts excess CSF from the ventricles of the brain through the valve. After passing through the valve, the excess CSF is drained through the distal catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters have Endexo® polymer, a surface modifying macromolecule, blended into their base silicone. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. The Libertís™ Shunt Catheter has been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have ink markings on the silicone tubing to aid in positioning of the catheter.
The catheters are packaged with two accessories, a stylet and a right-angle adapter. The stylet is used to help introduce the ventricular catheter into the brain's ventricles. The right-angle adapter is used to fixate the ventricular catheter to the cranium.
This document is a 510(k) clearance letter for a medical device, specifically a shunt catheter. It does not describe an AI medical device or software. Therefore, most of the requested information regarding AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, or standalone algorithm performance cannot be extracted from the provided text.
The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing and biocompatibility testing, rather than performance studies of an AI algorithm.
However, I can provide the acceptance criteria and performance information that is present in the document, which pertains to the physical and biological characteristics of the shunt catheter.
Acceptance Criteria and Device Performance (for the physical device, not an AI algorithm):
Since the provided document is for a medical device (shunt catheter) and not an AI/software device, the questions related to AI-specific criteria (e.g., sample size for test/training sets of AI, expert adjudication, MRMC studies, standalone algorithm performance, type of ground truth for AI) are not applicable.
The acceptance criteria here refer to the performance of the physical device as demonstrated through various bench and biocompatibility tests. The document states that all tests were "Pass," implying that the device met the pre-defined acceptance criteria for each test.
1. Table of Acceptance Criteria and Reported Device Performance
Test | Acceptance Criteria (Implied by "Pass") | Reported Device Performance |
---|---|---|
Evaluation of Safety and Performance per ISO 7197 and ASTM F647-94 | Complies with the safety and performance requirements of these standards. | Pass |
Functional/Mechanical Performance per ASTM F647-94 and ASTM D412-16 | Complies with the functional and mechanical performance requirements of these standards. | Pass |
MRI Compatibility per ASTM F2053-20 | Meets MRI compatibility standards (specifically, "MR-safe" designation). | Pass |
Antimicrobial Efficacy – Zone of Inhibition (ZOI) per internal methods | Demonstrates adequate zone of inhibition as per internal methods. | Pass |
Antibiotic Content Analysis using High-performance Liquid Chromatography (HPLC), per internal methods | Meets specified antibiotic content levels as per internal methods. | Pass |
Residual Solvents Testing via Headspace Gas Chromatography – Flame Ionization Detection (HS GC-FID), per FDA guidance "Q3C – Tables and List Guidance for Industry" | Meets specified limits for residual solvents. | Pass |
Usability Evaluation and Design Validation per IEC 62366-1 | Complies with usability and design validation requirements of IEC 62366-1. | Pass |
In vitro flow loop thrombogenicity test | Observed significantly less thrombus deposition compared to the predicate device. | Observed significantly less thrombus deposition on Codman Libertis Distal Shunt Catheter with Bactiseal and Endexo Technology compared to predicate device |
Cytotoxicity per ISO 10993-5:2009 | No cytotoxic effects. | Pass |
Sensitization per ISO 10993-10:2021 | No sensitization reactions. | Pass |
Irritation per ISO 10993-10:2021 and ISO 10993-23:2021 | No irritation. | Pass |
Acute Systemic Toxicity per ISO 10993-11:2017 | No acute systemic toxicity. | Pass |
Pyrogenicity per ISO 10993-11:2017 | No pyrogenic effects. | Pass |
Subacute/Subchronic Toxicity per ISO 10993-6:2016 and ISO 10993-11:2017 | No subacute/subchronic toxicity. | Pass |
Genotoxicity per ISO 10993-3:2014 | No genotoxic effects. | Pass |
Implantation per ISO 10993-6:2016 | Acceptable tissue response upon implantation. | Pass |
Chronic Toxicity per ISO 10993-11:2017 | No chronic toxicity. | Pass |
Hemocompatibility per ISO 10993-4:2017 | Hemocompatible. | Pass |
Carcinogenicity per ISO 10993-3:2014 | No carcinogenic effects. | Pass |
Sterilization Validation (E-beam) per ISO 11137-1, ISO 11137-2, ISO 11737-1, and ISO 11737-2 | Achieves a sterility assurance level (SAL) of 10-6. | Achieved SAL of 10-6 |
Shelf Life | At least 1 year. | 1 year |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify general "sample sizes" in terms of patient data or typical "test sets" as one would for an AI model. Instead, it refers to "production equivalent devices" used for the various bench and biocompatibility tests. The specific number of devices or replicates for each test is not detailed.
- Data Provenance: The data is generated from laboratory studies and bench testing performed on the physical device. This is not clinical patient data. The provenance for this type of testing is typically the manufacturer's internal testing facilities or contract labs. No country of origin for data or retrospective/prospective nature is applicable as this is not patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Not applicable. This pertains to an AI/software device. The "ground truth" for a physical device is established by objective measurements and standardized test methods, not expert consensus on data interpretation.
4. Adjudication Method (e.g. 2+1, 3+1, none) for the Test Set
- Not applicable. This pertains to an AI/software device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, not applicable. This pertains to an AI/software device evaluating human reader performance.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No, not applicable. This pertains to an AI algorithm; the device is a physical shunt catheter.
7. The Type of Ground Truth Used
- For the performance bench tests and biocompatibility tests, the "ground truth" is defined by established international (ISO, ASTM) and FDA-recognized consensus standards, as well as pre-defined internal test methods and specifications. For instance, a "Pass" for biocompatibility means the device met the criteria specified in the relevant ISO 10993 standards. For thrombogenicity, the ground truth was a statistically significant reduction in thrombus deposition compared to the predicate device.
8. The Sample Size for the Training Set
- Not applicable. This pertains to an AI/software device. No training set is involved for validating a physical medical device.
9. How the Ground Truth for the Training Set was Established
- Not applicable. This pertains to an AI/software device.
Ask a specific question about this device
(264 days)
Integra LifeSciences Production Corporation
The Codman Libertís™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.
The Codman LibertísTM EVD Catheter with Bactiseal® and Endexo® Technology (LibertísTM EVD Catheters) include a ventricular catheter that is supplied with component accessories that facilitate placement and use of the catheter for reducing and controlling intracranial pressure due to excess cerebrospinal fluid (CSF). The ventricular catheter is subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. Laboratory studies show Bactiseal treated catheters reduce the colonization of gram-positive bacteria on the tubing surface. Additionally, LibertísTM EVD Catheters contain an Endexo® polymer additive; a surface modifying macromolecule (SMM) polymer blended into the catheter's base silicone. The ventricular catheter is placed in the ventricles of the brain and CSF enters the fluid conduit through the inlet holes near the tip of the catheter and drains into the external drainage system connected to the catheter. The catheter contains a barium sulfate stripe for radiopacity and includes numerical depth markings and circumferential bands, made of ink, from the proximal tip.
The provided document is a 510(k) clearance letter from the FDA for a medical device (Codman Libertís™ EVD Catheter). It describes the device, its intended use, a comparison to a predicate device, and summaries of nonclinical testing.
However, the question asks for details related to acceptance criteria and studies that prove a device meets acceptance criteria, specifically those typically associated with AI/software medical devices. This document does not describe an AI/software medical device; it describes a physical medical device (a catheter for draining CSF). Therefore, much of the information requested in the prompt, such as "number of experts used to establish ground truth," "adjudication method," "multi-reader multi-case (MRMC) study," "standalone performance," "training set size," and "how ground truth for training set was established," are not applicable to this type of device and are not present in the provided text.
The document focuses on demonstrating substantial equivalence to a predicate device through bench testing, biocompatibility, and sterilization validation, which are standard for physical medical devices.
Here's an attempt to answer the applicable parts of your request based on the provided text, and explicitly state where information is not available or not applicable.
Description of Acceptance Criteria and Proving Device Meets Criteria
The Codman Libertís™ EVD Catheter is a physical medical device. Its acceptance criteria are primarily demonstrated through a battery of nonclinical performance bench tests, biocompatibility testing, and sterilization validation, to show that it is substantially equivalent to a legally marketed predicate device. The performance criteria are generally considered a "Pass" or "Fail" based on established standards and internal methods.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly list quantitative acceptance criteria values but rather states a "Pass" or "Fail" for each test. The acceptance criterion for each test is implicitly that the device performs as expected according to the specified standard or internal method.
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Evaluation of Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 | Meets specifications of ASTM F647 and ISO 7197 for EVD catheters. | Pass |
Evaluation of Aged Codman Libertís EVD Catheters per ASTM F647 and ISO 7197 (T = 12 months) | Meets specifications of ASTM F647 and ISO 7197 for EVD catheters after 12 months simulated aging. | Pass |
Functional/Mechanical Performance of Codman Libertís EVD Catheters | Meets predefined functional and mechanical performance specifications. | Pass |
Functional/Mechanical Performance of Aged Codman Libertís EVD Catheters (T = 12 months) | Meets predefined functional and mechanical performance specifications after 12 months simulated aging. | Pass |
MRI Compatibility Assessment of Codman Libertís EVD Catheters per ASTM F2503 | Meets specifications of ASTM F2503 for MRI compatibility (Note: Catheter is MR Conditional, Trocar and Stylet are MR-Unsafe, as noted in labeling). | Pass |
Antimicrobial Efficacy of Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use | Demonstrates sufficient Zone of Inhibition against specified bacteria after 28 days of simulated use to show antimicrobial efficacy. | Pass |
Antimicrobial Efficacy of Aged Codman Libertís EVD Catheters – Zone of Inhibition (ZOI) after 28 Days of Simulated Use (T = 12 months) | Demonstrates sufficient Zone of Inhibition after 12 months simulated aging and 28 days of simulated use. | Pass |
Antibiotic Content Testing of Codman Libertís EVD Catheters | Contains the specified levels of antibiotics (rifampicin and clindamycin hydrochloride). | Pass |
Residual Solvents Testing of Codman Libertís EVD Catheters | Levels of residual solvents are within acceptable limits. | Pass |
Libertís EVD Catheters with Bactiseal® and Endexo® Technology – Summative Usability Study | Meets user-related safety and effectiveness requirements, indicating the device can be used safely and effectively by its intended users. | Pass |
Product Shelf Life Testing per ASTM F1980 and ISO 11607-1 | Maintains integrity and performance over its stated shelf life (12 months) as per ASTM F1980 and ISO 11607-1. | Pass |
Biocompatibility Testing per ISO 10993-1 | Demonstrates biocompatibility in accordance with ISO 10993-1 for medical devices in contact with the body. | Pass |
Sterilization Validation (Electron Beam) | Achieves a sterility assurance level (SAL) of 10-6. | Pass |
2. Sample size used for the test set and the data provenance
The document states, "All testing was performed on production equivalent devices." However, it does not specify the sample size for any of the individual bench tests.
Data provenance (e.g., country of origin of the data, retrospective or prospective) is not applicable in the traditional sense for these bench tests, which are conducted in a laboratory setting to evaluate physical properties and performance characteristics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This concept is not applicable to this type of medical device (physical catheter). "Ground truth" in this context would refer to established engineering standards, material specifications, and validated test methods, rather than expert interpretation of medical images or clinical outcomes.
4. Adjudication method for the test set
This concept is not applicable to this type of medical device. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple readers (e.g., radiologists) to resolve disagreements in interpretation, which is not relevant for bench testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done, as this is a physical medical device (catheter), not an AI/software device. Therefore, this question is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable since the device is a physical catheter, not a software algorithm.
7. The type of ground truth used
The "ground truth" for the performance tests of this physical device consists of:
- Established industry standards (e.g., ASTM F647, ISO 7197, ASTM F2503, ISO 10993-1, ASTM F1980, ISO 11607-1).
- Internal predefined functional and mechanical specifications.
- Validated laboratory methods for chemical content, antimicrobial efficacy (Zone of Inhibition), and sterility assurance (SAL 10-6).
- Usability requirements for the Summative Usability Study.
8. The sample size for the training set
This concept is not applicable as this is a physical medical device, not an AI/software device that would involve a "training set."
9. How the ground truth for the training set was established
This concept is not applicable for the same reason as above.
Ask a specific question about this device
(28 days)
Integra LifeSciences Corporation
The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic, and/or therapeutic procedures, such as ventriculostomies, biopsies and removal of cysts, tumors, and other obstructions.
The Aurora Surgiscope System consists of two components: (1) a sterile, single use, Sheath with integrated illumination LEDs and camera, with an Obturator, and (2) a non-sterile, reusable control unit, Image Control Box (ICB).
The Sheath is intended to provide access to the surgical site by acting as the insertable portion of the device, as well as the instrument channel to accommodate other surgical tools. Depth markers are present along the length of the Sheath for user reference. The proximal end of the Sheath also incorporates a Tab, which serves as the location for fixation arm to hold the device.
At the proximal end of the Sheath is the Imager, which comprises the following components: LEDs (light emitting diodes), camera (and optical components), and focus knob.
- The LEDs provide illumination to the surgical field by directing light down the Sheath, along the working channel.
- The camera captures videos of the surgical field.
- The focus knob allows the user to adjust the focus of the camera to obtain the desired image quality.
To facilitate insertion of the Sheath into the surgical site, an Obturator is provided with the device. During device insertion, the Obturator is fully inserted into the Sheath, and the entire AURORA Surgiscope is advanced to the desired surgical location. The distal end of the Obturator is conical in shape to minimize tissue damage during device insertion. In addition, the proximal handle of the Obturator is designed to accommodate various stereotactic instruments for neuronavigation, which can further aid in device placement. The Obturator is removed after insertion.
The ICB is a non-sterile device that provides three main functions in the AURORA Surgiscope System:
- To power the LEDs and camera of the AURORA Surgiscope.
- To relay the video feed captured by the AURORA Surgiscope camera to a connected Medical Grade Surgical Monitor for real-time image visualization.
- To allow the user to make adjustments to the displayed video feed (e.g., contrast, brightness), and to vary the light output of the LEDs.
The user interface is a membrane keypad with buttons located on the ICB that can be depressed for image adjustment, such as zoom, contrast, and brightness. The connection ports to the AURORA Surgiscope, Medical Grade Surgical Display Monitor, and Power are located on the side of the ICB, along with the ON/OFF switch.
The provided FDA 510(k) clearance letter for the AURORA Surgiscope System (K250752) does not contain the detailed information necessary to fully answer all the questions regarding acceptance criteria and the study that proves the device meets them.
The document primarily focuses on demonstrating substantial equivalence to a predicate device (K201840) based on technological characteristics and functional requirements. It explicitly states that "No clinical test/studies were required or performed as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate device (K201840)."
Therefore, for many of the requested points, the answer will be that the information is not available in the provided text.
Here's a breakdown of what can and cannot be answered based on the input:
1. A table of acceptance criteria and the reported device performance
The document mentions "functional requirements" and "performance tests" but does not detail specific acceptance criteria or quantitative performance results. It only states that the device meets these requirements after sterilization, environmental, and transit conditioning, and equivalent to a 1-year shelf-life.
Acceptance Criterion | Reported Device Performance |
---|---|
Functional requirements after 2X EO sterilization | Device meets functional requirements |
Functional requirements after environmental and transit conditioning | Device meets functional requirements |
Functional requirements after equivalent of 1-year claimed shelf-life | Device meets functional requirements |
Obturator handle strength (improved connection) | Met (due to design modification with two bridge features) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample size for test set: Not specified. The document states "non-clinical testing was performed" but does not detail the number of units tested.
- Data provenance: Not specified. The nature of the testing (functional performance, sterilization effects) suggests it would be laboratory testing rather than patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not specified. Since no clinical studies were performed, there was no need for expert review of clinical data to establish ground truth. The "ground truth" here would be the successful function of the device in engineering tests.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not specified. No clinical data was being adjudicated.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The document explicitly states: "No clinical test/studies were required or performed". This device is a surgical endoscope, not an AI-assisted diagnostic tool, so an MRMC study comparing human readers with and without AI assistance is not relevant to its clearance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical medical device (endoscope), not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the non-clinical performance tests mentioned, the "ground truth" would be the engineering specifications and functional integrity of the device. This is typically verified through direct measurement, visual inspection, and functional tests (e.g., image quality assessment, illumination intensity, camera function, mechanical integrity) against predefined specifications. It is not based on clinical "ground truth" like pathology or expert consensus.
8. The sample size for the training set
- Not applicable. This notice does not describe an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. No training set was involved for this device.
Ask a specific question about this device
(142 days)
Integra LifeSciences Corporation
The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:
Plastic and Reconstructive surgery, Orthopedic Surgery and Thoracic Surgery and the following specific uses:
Neurosurgery - including removal of primary and secondary malignant and spinal tumors, including but not limited to meningiomas and gliomas
Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy
Gynecological Surgery - including removal of dysplastic genital epithelial tissue including vulvar and vaginal intraepithelial neoplasia, removal of condyloma, debulking of metastatic uterine, ovarian, fallopian tube or primary peritoneal carcinoma, and open or laparoscopic excision of tissue and adhesions associated with endometriosis
The CUSA Clarity Ultrasonic Surgical Aspirator System (CUSA Clarity) is an ultrasonically vibrating surgical device which, in combination with irrigation and aspiration, fragments, emulsifies and removes unwanted tissue. It allows the selective dissection of target tissue while preserving vessels, ducts and other delicate structures. The CUSA Clarity consists of a console that provides power and control of the ultrasonic, aspiration and irrigation functions, two surgical handpieces that provide ultrasonic mechanical energy (23 kHz and 36 kHz), a footswitch to allow user control over the ultrasonics, titanium surgical tips (variety of models), irrigation flues, suction/irrigation system (manifold tubing and vacuum canister) and accessories used for assembly/disassembly and reprocessing. The CUSA Clarity may also be optionally used with the CUSA Electrosurgical Modules which provide optional electrosurgical capability.
The information provided describes a 510(k) premarket notification for the CUSA® Clarity Ultrasonic Surgical Aspirator System, primarily focusing on a modification to its Indications for Use to include more specific gynecological surgery indications. The submission claims substantial equivalence to a predicate device (K182809) and did not involve any design or technological changes to the device itself. Therefore, the "acceptance criteria" and "device performance" in the traditional sense of a new or modified device's functional characteristics are not directly presented as a series of quantitative benchmarks met by a specific study on the device's technical performance. Instead, the "acceptance criteria" here refer to demonstrating that the expanded indications for use for gynecological surgery are safe and effective, supported by existing literature and showing no new safety or effectiveness concerns compared to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied for Expanded Indications) | Reported Device Performance (Based on Literature Review) |
---|---|
For Dysplastic Genital/Perianal Epithelial Tissue (e.g., VIN, Condyloma): | CUSA was used to treat 272 patients with dysplasia or condyloma. Successful achievement of desired treatment outcomes was reported, and no significant postoperative complications or scarring were noted in the clinical literature for condyloma cases. |
For Debulking Metastatic Uterine, Ovarian, Fallopian Tube, or Primary Peritoneal Carcinoma: | The provided articles demonstrated the safe and effective use of CUSA in debulking ovarian cancer, as well as metastases from endometrial cancer, tubal adenocarcinoma, and peritoneal tumors for 630 patients requiring debulking procedures. Articles reported both increased and decreased procedure durations when CUSA was used. One article noted risk of DIC (Disseminated Intravascular Coagulation), but no other studies found such a risk in a CUSA population of 1,465 patients. |
For Excision of Tissue and Adhesions Associated with Endometriosis: | In cases where endometrial tissue was removed (18 patients), CUSA effectively removed the endometrial tissue while preserving vessels and nerves. |
Overall Safety and Effectiveness for Gynecological Applications: | The literature review, encompassing 1,465 patients across 54 articles, concluded that CUSA is safe and effective in these gynecological applications, demonstrating substantial equivalence to the predicate device. |
2. Sample size used for the test set and the data provenance
The "test set" in this context refers to the body of clinical evidence reviewed to support the expanded indications.
- Sample Size: 1,465 patients. This total is comprised of:
- 272 patients for dysplasia or condyloma.
- 630 patients for debulking procedures (cancers).
- 18 patients for endometriosis.
- The remaining patients likely fall into categories not specifically broken down or represent overlaps in articles.
- Data Provenance: The data was derived from an "analysis of peer-reviewed articles." This indicates a retrospective review of published clinical literature. The country of origin of the data is not specified but would likely span various international institutions given the nature of peer-reviewed publications over an extended period.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable in the typical sense. This was a literature review, not a study where experts established ground truth for a novel dataset. The "ground truth" is implicitly established by the reported outcomes in the published clinical studies themselves, which are conducted by various medical professionals (surgeons, pathologists, etc.) in their respective fields. The submission does not specify an independent panel of experts reviewing the collected literature to establish a "ground truth" for the purpose of this 510(k).
4. Adjudication method for the test set
Not applicable. There was no explicit adjudication method for a test set as this was a review of existing peer-reviewed literature. The outcomes reported in the individual studies within the literature served as the evidence.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is an ultrasonic surgical aspirator system, not an AI or imaging diagnostic device that involves human "readers" interpreting output. This submission focuses on the safety and effectiveness of the existing device for expanded (gynecological) indications.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The CUSA Clarity is a surgical instrument used by a surgeon, not a standalone algorithm.
7. The type of ground truth used
The "ground truth"-like evidence used was outcomes data and reported effectiveness/safety from published clinical studies. This includes:
- Achievement of desired treatment outcomes (e.g., successful removal of tissue).
- Reported presence or absence of postoperative complications.
- Preservation of vessels and nerves.
- Comparison of procedure durations.
8. The sample size for the training set
Not applicable. This was a 510(k) submission for an existing device with expanded indications, not an AI/algorithm development where a training set would be used. The "evidence" supporting the expanded indications comes from published literature.
9. How the ground truth for the training set was established
Not applicable. As above, there was no training set for an algorithm.
Ask a specific question about this device
(139 days)
Integra LifeSciences Production Corporation
The ICP Monitor is intended for use as an interface between compatible strain gauge type pressure transducers and standard physiological pressure monitoring systems. The ICP Monitor is also intended for use as an independent pressure monitor for displaying the mean, systolic and diastolic values of a physiologic pressure waveform in the absence of an external patient monitor. The CereLink ICP Extension cable is intended for use as a connecting cable between the ICP input channel of the CereLink ICP Monitor and a CereLink ICP Sensor.
The CereLink ICP Monitor is indicated for use in the ICU or Operating Room (OR) environment for monitoring intracranial pressure (ICP) via a solid-state sensor placed directly in parenchymal tissue or integrated into an external ventricular drainage catheter placed in the ventricle. In addition to monitoring ICP and activating alarms when the intracranial pressure is outside user-set limits, the device performs these functions:
- Displays ICP Waveform
- Displays Mean ICP numeric
- Displays the historic mean pressure as a trend
- Displays trend statistics (Pressure Time Dosage (PTD), time above threshold, boxplot, histogram)
- Stores 14-days' worth of mean ICP values
- Stores 24 hours of pressure waveform
- Can capture and store screen-shots
- Can download various data to a USB device for printing or analysis
- Real-time data streaming of mean ICP and waveform via USB connection
- Connect to external patient monitor
The CereLink ICP Monitor can be transported with the hospital to continuously record data. The monitor includes a 7" color touch screen that is compatible with the use of gloves. The monitor is provided to the user with a CereLink ICP Extension Cable, external power supply, and comes equipped with an internal rechargeable battery. The monitor has one output channel to transfer physiological data to a compatible Patient Monitor, as well as one input channel to receive ICP readings from the implanted CereLink ICP sensor. The implanted sensor is connected to the CereLink ICP Monitor by way of the CereLink ICP Extension Cable; the CereLink ICP Monitor connects to compatible patient monitors through the patient monitor interface cables.
The provided text describes a 510(k) premarket notification for a medical device, the CereLink ICP Monitor and CereLink ICP Extension Cable, which are intended for intracranial pressure monitoring. The document outlines acceptance criteria and performance testing for these devices, particularly focusing on modifications made to the extension cable.
1. Table of Acceptance Criteria and Reported Device Performance:
Test | Acceptance Criteria | Reported Device Performance/Conclusion |
---|---|---|
Mechanical Tests | Not explicitly stated | Pass |
Electrical Testing (Impedance Measurement) | Not explicitly stated | Pass |
Out of Range Failure Test (using electrical stress setup) | Recovery within specified time after correct connection of patient lead | All CereLink Systems recovered from failure within specified amount of time. Pass |
30 Day ICP Drift Test (using electrical stress setup) | Meet specification | ICP drift met the specification. Pass |
Mean Time Between Failure Calculation Test | Not explicitly stated | Pass |
Simulated Environment Validation Test (13 days) | Not explicitly stated | Pass |
Label Verification Test | Not explicitly stated | Pass |
Drawing Verification Test | Not explicitly stated | Pass |
Summative Usability Report | Not explicitly stated | Pass |
Biocompatibility: ISO 10993-5:2009 | Not explicitly stated | Pass |
Biocompatibility: ISO 10993-10:2021 | Not explicitly stated | Pass |
Biocompatibility: ISO 10993-23:2021 | Not explicitly stated | Pass |
Electrical Safety: IEC 60601-1: 2005+AMD1:2012+AMD2:2020 | Not explicitly stated | Pass |
Electrical Safety: IEC 60601-1-6: 2010+AMD1:2013+AMD2:2020 | Not explicitly stated | Pass |
Electrical Safety: IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 | Not explicitly stated | Pass |
Electrical Safety: IEC 60601-1-2:2014+AMD1:2020 | Not explicitly stated | Pass |
Electrical Safety: IEC 62366-1:2015+AMD1:2020 | Not explicitly stated | Pass |
2. Sample Size Used for the Test Set and Data Provenance:
The document refers to "the CereLink Systems" being exposed to the electrical stress setup. It does not provide a specific number for the sample size (e.g., number of devices or test units) used in these tests.
The data provenance is from bench testing as explicitly stated ("Performance Bench Test Results"). This indicates controlled laboratory conditions. There is no mention of country of origin for the data, and the nature of the tests (bench) makes the retrospective/prospective distinction less relevant in the typical clinical study sense. However, the tests are for design verification and validation, implying they were conducted prospectively to assess the new design.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
Not applicable. This submission pertains to a hardware device (ICP monitor and cable) and its electrical and mechanical performance, not an AI/software device that requires human expert review to establish ground truth for image or diagnostic interpretation. The testing relies on established engineering and medical device standards.
4. Adjudication Method for the Test Set:
Not applicable. As noted above, this is not an AI/software device requiring subjective interpretation or adjudication by experts. The tests are objective performance evaluations against specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No, an MRMC comparative effectiveness study was not done. The device is a hardware ICP monitor and cable, not an AI-assisted diagnostic tool for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an algorithm-based device. The "performance" refers to the physical and electrical characteristics of the monitor and cable. The tests described (ICP Drift, Out-of-Range Failure, Electrical Safety, etc.) are essentially "standalone" performance evaluations of the device itself.
7. The Type of Ground Truth Used:
The ground truth for the device's performance is based on engineering specifications and internationally recognized standards (e.g., ISO, IEC). For the specific "Out-of-Range Failure Test," the ground truth for "failure" is when the device enters the out-of-range state, and the "recovery" is when it exits that state within a specified time, which are objective, measurable outcomes in a controlled "electrical stress setup" designed to reproduce these failures.
8. The Sample Size for the Training Set:
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the Ground Truth for the Training Set was Established:
Not applicable, as there is no training set for this hardware device.
Ask a specific question about this device
(90 days)
Integra LifeSciences Production Corporation
The Bactiseal Catheters are indicated for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid (CSF) is indicated.
The Bactiseal Barium Striped Catheters are indicated for use in the treatment of hydrocephalus as a component of a shunt system when draining of cerebrospinal fluid (CSF) is indicated.
The Bactiseal Endoscopic Ventricular Catheter is designed for use in the treatment of hydrocephalus when shunting cerebrospinal fluid (CSF) from the ventricles of the brain.
The Bactiseal Catheters, Bactiseal Barium Striped Catheters and Bactiseal Endoscopic Ventricular Catheter include a ventricular and/or distal (peritoneal) drainage catheter that are used as part of a CSF shunting system to treat hydrocephalus. Both catheters are attached to the valve portion of a shunting system, which is then implanted in the patient's brain. The ventricular catheter diverts the excessive CSF from the ventricles of the brain through the valve. After passing through the valve, the excessive CSF is drained through the distal (peritoneal) drainage catheter into another part of the body, such as the peritoneal cavity, where it is reabsorbed into the bloodstream. The catheters are subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. Bactiseal silicone catheters have been shown in laboratory studies to reduce the colonization of gram-positive bacteria on the tubing surface. The catheters contain barium sulfate for radiopacity and have tantalum "dots" incorporated onto the silicone tubing to aid in positioning of the catheter. The Bactiseal Catheters and Bactiseal Endoscopic Ventricular Catheter are made of radiopaque silicone tubing, and the Bactiseal Barium Striped Catheters are made of clear silicone tubing with radiopaque striping. The Bactiseal Endoscopic Ventricular Catheter has a slit in the tip of the ventricular catheter in order for the catheter to be placed with the use of an endoscope.
This document is a 510(k) summary for modifications made to existing Bactiseal Catheters, Bactiseal Barium Striped Catheters, and Bactiseal Endoscopic Ventricular Catheters. The modifications primarily involve updates to MRI labeling and a change in the supplier of clindamycin hydrochloride.
Therefore, the submission focuses on demonstrating that these modifications do not introduce new questions of safety or effectiveness, rather than proving the initial efficacy of an entirely new device. This means that a conventional study with specific acceptance criteria, test sets, expert adjudication, and detailed ground truth establishment as typically seen for entirely new AI/CADe devices, is not applicable in this context. The document relies on bench testing and an equivalency assessment to the predicate devices.
Here's a breakdown of the requested information based on the provided text, with significant portions noted as "Not applicable" due to the nature of this 510(k) submission:
1. A table of acceptance criteria and the reported device performance
Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
MRI Safety Testing (ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119) | Device meets established MRI safety standards for MR Conditional. | Pass |
Drug Equivalency Testing (USP standards, USP Monograph for Clindamycin Hydrochloride) | Clindamycin hydrochloride from new supplier is equivalent to current supplier in identity, formulation, concentration, application method, and drug release. | Pass |
Drug Effectiveness Testing (USP and internal test methods) | Device continues to demonstrate the intended antimicrobial effectiveness. | Pass |
Sterilization Equivalency Assessment | Sterilization process remains effective with the new clindamycin hydrochloride supplier. | Acceptable |
Biocompatibility Assessment | New clindamycin hydrochloride supplier does not introduce new biocompatibility issues. | Determined no new issues |
Explanation of Implied Acceptance Criteria: The document states that the testing "utilized well-established methods, including those from FDA consensus standards." For a "Pass" result in such tests, the device must meet the specific criteria outlined in those standards. For drug equivalency and effectiveness, the stated goal is to confirm the new supplier's clindamycin hydrochloride is "equivalent" and "continues to meet the same drug specifications" and efficacy. The biocompatibility assessment "determined that the introduction of the new supplier for clindamycin hydrochloride does not introduce any new issues."
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified. The document indicates "All testing was performed on production equivalent devices," but the number of devices or units tested for each benchmark is not provided.
- Data Provenance: Not applicable in the context of clinical data. The tests are benchtop performance tests. The specific labs or countries where these bench tests were conducted are not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. This submission concerns bench testing and equivalency assessment of device modifications, not clinical performance requiring expert-established ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. This submission concerns bench testing and equivalency assessment of device modifications, not clinical performance requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/CADe device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/CADe device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable in the conventional sense. The "ground truth" for the bench tests would be the established scientific and engineering principles, and the specifications of the predicate device/original drug, against which the modified device's performance is compared. For example, the ground truth for MRI safety is defined by the ASTM standards.
8. The sample size for the training set
- Not applicable. This is not an AI/CADe device, and no training set is mentioned or implied for its development or evaluation.
9. How the ground truth for the training set was established
- Not applicable. As there is no training set, there is no ground truth to establish for it.
Ask a specific question about this device
(60 days)
Integra LifeSciences Production Corporation
The Bactiseal EVD Catheter and Bactiseal Clear EVD Catheter sets are indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.
The Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set include a ventricular catheter that is supplied with component accessories that facilitate placement and use of the catheter for reducing and controlling intracranial pressure due to excess cerebrospinal fluid. The ventricular catheter is subjected to a treatment process by which the silicone is impregnated with two antimicrobials, rifampicin and clindamycin hydrochloride. Laboratory studies show Bactiseal treated catheters reduce the colonization of gram-positive bacteria on the tubing surface. The ventricular catheter is placed in the ventricles of the brain and CSF enters the fluid conduit through the inlet holes near the tip of the catheter and drains into the external drainage system connected to the catheter. The catheter contains barium sulfate for radiopacity and includes numerical depth markings and circumferential bands, made of ink, from the proximal tip.
The provided document, K233448, is an FDA 510(k) clearance letter for the Integra LifeSciences Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set. Based on the content of this letter, the device in question is a physical medical device (catheter) and not an AI/software device.
Therefore, the request for information related to acceptance criteria and studies proving the device meets those criteria, specifically concerning AI/ML performance metrics (such as multi-reader multi-case studies, ground truth establishment for training/test sets, sample sizes for image data, and expert adjudication), is not applicable to this document.
The document describes the changes made to an existing device (Bactiseal EVD Catheter Sets) as:
- Labeling changes: Updating MRI labeling to state MR Safe and administrative updates/harmonized symbols.
- New supplier for clindamycin hydrochloride: The antimicrobial impregnated into the catheter.
The studies mentioned are non-clinical (bench testing and engineering analysis) to demonstrate equivalence to the predicate device despite these changes.
Here's a summary of the relevant "acceptance criteria" and "study that proves the device meets the acceptance criteria" as described in the document for this non-AI/software device:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Test/Evaluation | Reported Device Performance/Conclusion |
---|---|---|
Drug Equivalency | Drug Equivalency Testing per USP standards and USP Monograph for clindamycin hydrochloride | Pass |
Drug Effectiveness | Drug Effectiveness Testing per USP and internal test method | Pass |
Sterilization | Sterilization equivalency assessment (comparing predicate to proposed device with new supplier clindamycin hydrochloride) | Deemed acceptable |
Shelf Life | Assessment for changes in shelf life due to proposed modifications | No changes |
Biocompatibility | Assessment (due to new clindamycin supplier) | No new issues; additional testing not necessary |
Overall Equivalence | Comparison to predicate device | Substantially equivalent |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated in terms of number of units tested, but it refers to "production equivalent devices" for performance testing.
- Data Provenance: Not applicable in the context of clinical data. The tests are bench tests performed by the manufacturer.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
- Not applicable. This pertains to an AI/software device requiring human expert annotation/ground truth. For this catheter, the "ground truth" is adherence to established industry standards (USP) and internal test methods for drug performance and other material characteristics.
4. Adjudication Method for the Test Set:
- Not applicable. This pertains to clinical data review for AI/software devices. Testing was based on established laboratory methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, not applicable. This type of study is for evaluating observer performance with and without AI assistance for clinical interpretations. The current device is a physical catheter, not an AI diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, not applicable. This applies to AI/software performance.
7. The Type of Ground Truth Used:
- For drug equivalency and effectiveness: USP standards and monograph, internal test methods.
- For biocompatibility: Assessment against known materials and established literature/standards.
- For physical characteristics: Engineering analysis and comparison to predicate device specifications.
8. The Sample Size for the Training Set:
- Not applicable. There is no "training set" as this is not an AI/ML product.
9. How the Ground Truth for the Training Set was Established:
- Not applicable. There is no "training set" or need for ground truth establishment in this context.
Ask a specific question about this device
Integra Lifesciences Corporation
The CUSA® Clarity Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable.
The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in:
Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses:
Neurosurgery - including removal of primary and secondary malignant and benign brain and spinal tumors, including but not limited to meningiomas and gliomas
Gastrointestinal and Affiliated Organ Surgery - including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy
Urological surgery- including removal of renal parenchyma during nephrectomy or partial nephrectomy
General Surgery - including removal of benign or malignant tumors or other unwanted soft or hard tissue in open or minimally invasive general surgical procedures
Laparoscopic Surgery - including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy
The device within the scope of this premarket notification is the optional CUSA® Electrosurgery Module (CEM) accessory that is intended to be used with the 23 kHz components of the CUSA® Clarity Ultrasonic Surgical Aspirator System.
The purpose of this submission is to modify the CEM nosecone accessory currently offered with CUSA Clarity to allow for connection with additional electrosurgical generators, to continue to provide electrosurgical capabilities to the user. The additional electrosurgical generators that the modified CEM Nosecone may be used with include the Medtronic FT10 (K191601), Medtronic FX8 (K181389), Erbe VIO 300D (K083452), and Erbe VIO 3 (K190823). Compatibility with the Medtronic Force FX (K143161) will be maintained as well.
The CUSA Clarity 23kHz Expanded CEM Nosecone has the same intended use and technological characteristics as the predicate CUSA Clarity 23 kHz CEM Nosecone (K190180). The subject CEM nosecone will continue to allow the surgeon to apply immediate electrosurgical coagulation to bleeding tissue at the surgical site, with the same handpiece assembly that is removing unwanted tissue.
The provided text describes the CUSA Clarity Ultrasonic Surgical Aspirator System, specifically focusing on a modified CEM nosecone accessory for the 23 kHz components. This submission is a 510(k) premarket notification claiming substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present acceptance criteria in a quantitative table format with corresponding device performance values. Instead, it describes various non-clinical tests undertaken to ensure the safety and efficacy of the device and its substantial equivalence to the predicate.
The reported device performance is broadly stated as:
- "Testing was determined successful and supports the conclusion that all product specifications and design inputs have been met."
- "The results of the non-clinical testing indicate that the intended use of the device, fundamental scientific technology, and performance of the CUSA Clarity 23 kHz Expanded CEM Nosecone is substantially equivalent to the predicate device."
Therefore, a table of quantitative acceptance criteria and specific reported device performance cannot be generated from the provided text. The information indicates that all tests were passed and specifications met, implying the device performed within acceptable limits.
2. Sample Size Used for the Test Set and Data Provenance
The document describes non-clinical testing but does not specify a "test set" in terms of subject or patient data. The tests are focused on device characteristics rather than clinical outcomes with a patient population.
- Sample size for test set: Not applicable in the context of device performance testing described. The tests are on the device itself (e.g., handpiece life, functionality, mechanical properties).
- Data provenance: Not applicable in the context of patient data. The provenance is from internal testing conducted by Integra LifeSciences Corporation. There is no mention of country of origin for data related to clinical or patient studies, as none were performed. The tests are prospective as they were conducted as part of the submission process.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of experts: Not applicable. The "ground truth" for the non-clinical tests is established by engineering specifications, recognized standards (FDA guidance documents, ISO standards for biocompatibility, EMC, and electrical safety), and comparison to the predicate device's established performance.
- Qualifications of experts: Not specified as a separate set of experts for ground truth. However, the development and testing would have been overseen by Integra LifeSciences Corporation's engineering and regulatory teams.
4. Adjudication Method for the Test Set
- Adjudication method: Not applicable. The testing described is objective device performance (e.g., sterilization, biocompatibility, electrical safety, mechanical, thermal effects). Success or failure is determined by meeting predefined engineering specifications and regulatory standards, not by an adjudication process between human experts on a specific outcome.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC study: No. The document explicitly states: "No clinical studies were performed or required as all conducted performance tests appropriately support a determination of substantial equivalence compared with the predicate device."
- Effect size of human readers with/without AI assistance: Not applicable, as no MRMC study or study involving human readers with or without AI assistance was conducted. The device is a surgical aspirator system, not an AI diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone study: Not applicable. The CUSA Clarity Ultrasonic Surgical Aspirator System is a physical medical device, not an algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant.
7. The Type of Ground Truth Used
The "ground truth" for the device's acceptable performance is based on:
- Established engineering specifications and design inputs: Ensuring the device functions as intended.
- Compliance with FDA guidance documents and recognized standards: Such as those for sterilization, biocompatibility, EMC, electrical safety, thermal effects, and capacitive coupling.
- Substantial equivalence to the legally marketed predicate device (CUSA® Clarity Ultrasonic Surgical Aspirator System K190180): This implies that the predicate's established safety and effectiveness profile serves as a benchmark for the new component.
8. The Sample Size for the Training Set
- Sample size for training set: Not applicable. This device is not an AI/ML algorithm that requires a training set. The performance is assessed through traditional engineering and regulatory compliance testing.
9. How the Ground Truth for the Training Set Was Established
- How ground truth for training set was established: Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(93 days)
Integra LifeSciences Production Corporation
The Certas Plus Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
The Certas Plus Programmable Valves are implantable, sterile, single use devices that provide constant intraventricular pressure and drainage of cerebrospinal fluid (CSF) for the management of hydrocephalus. Hydrocephalus is a condition caused by excessive accumulation of CSF in the ventricles of the brain due to a disturbance of CSF secretion, flow, or absorption, which causes a rise in intracranial pressure (ICP). To relieve ICP, CSF can be diverted through a shunting device, such as a Certas Plus Programmable Valve, to another body cavity where it is subsequently absorbed. The Certas Plus Programmable Valves can be set to eight different performance settings for intraventricular pressure and drainage of CSF. The performance settings of the valves can be set preoperatively and can also be noninvasively changed postimplantation by using the Certas Tool Kits. The Certas Tool Kits employ magnetic force to select one of eight settings.
The provided document is a 510(k) summary for the Certas Plus Programmable Valves, which are medical devices used for the management of hydrocephalus. This specific document outlines the regulatory review of a proposed modification to an already legally marketed device.
The document does not describe a study that proves the device meets clinical acceptance criteria in the way typically seen for novel medical imaging or diagnostic AI devices. Instead, it focuses on demonstrating substantial equivalence to existing predicate devices after minor modifications (material changes and dimensional/tolerance changes). Therefore, much of the requested information regarding clinical study design, ground truth establishment, expert adjudication, or MRMC studies will not be present in this type of submission.
Here's a breakdown based on the information available and what is not applicable given the nature of the submission:
1. A table of acceptance criteria and the reported device performance:
The acceptance criteria and performance are reported for bench and biocompatibility testing, not clinical performance in patients.
Test | Acceptance Criteria (Implicit: "Pass") | Reported Performance |
---|---|---|
Accelerated Aging for Shelf Life Testing per ASTM F1980 | Device maintains integrity and functionality over its claimed shelf life. | Pass |
Functional Testing per ISO 7197 | Device performs its intended function (e.g., constant intraventricular pressure, CSF drainage) within specified parameters. | Pass |
Functional Testing per ASTM F647 | Device performs its intended function within specified parameters. | Pass |
MRI Testing per ISO/TS 10974 | Device is safe and performs as intended in an MRI environment (e.g., MR conditional). | Pass |
X-Ray Testing per ASTM F640 | Device is visible and its characteristics can be assessed via X-ray. | Pass |
MTS Cytotoxicity per ISO 10993-5 | No cytotoxic effects observed. | Pass |
Guinea Pig Maximization Sensitization Study per ISO 10993-10 | No sensitization observed. | Pass |
Intracutaneous Irritation Study in Rabbits per ISO 10993-10 | No irritation observed. | Pass |
Acute Systemic Toxicity Study in mice per ISO 10993-11 | No acute systemic toxicity observed. | Pass |
Rabbit Pyrogen Study per United States Pharmacopeia 42 – NF 37 | No pyrogenic response observed. | Pass |
Subcutaneous Implantation Studies in Rabbits, 1 and 4 weeks per ISO 10993-6 | No adverse tissue reactions at the implantation site. | Pass |
Systemic Toxicity and Local Effects Study in Rabbits following Subcutaneous Implantation 13 weeks per ISO 10993-6 and ISO 10993-11 | No systemic toxicity or significant local effects observed over 13 weeks. | Pass |
Bacterial Reverse Mutation Study per ISO 10993-3 and ISO/TR 10993-33 | No mutagenic effects observed. | Pass |
In Vitro Mouse Lymphoma Study per ISO 10993-3 and ISO/TR 10993-33 | No genotoxic effects observed. | Pass |
Hemolysis on Extract Study per ISO 10993-4 and ASTM F756 | No significant hemolysis observed. | Pass |
2. Sample sizes used for the test set and the data provenance:
- Test set sample size: Not explicitly stated in terms of number of devices. The document mentions "production equivalent devices" were used for testing, implying multiple units were tested for each performance and biocompatibility assessment. Specific quantities are not provided, as this is a summary.
- Data provenance: Not applicable in the context of clinical patient data. The data comes from laboratory bench tests and biocompatibility studies on the device itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable. Ground truth, expert consensus, and expert qualifications are relevant for studies involving the interpretation of medical images or diagnostic outputs by human experts. This submission focuses on the physical and functional properties of an implantable device through bench and biocompatibility testing. The "ground truth" for these tests would be the established scientific and engineering principles and the predefined pass/fail criteria of the specified ASTM/ISO standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not applicable as there are no human interpretations or diagnostic outcomes to adjudicate. Test results are objective (e.g., chemical analysis, physical measurements, functional outputs).
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. MRMC studies are used for evaluating diagnostic AI tools where human readers (e.g., radiologists) interpret cases with and without AI assistance to assess the AI's impact on their performance. The Certas Plus Programmable Valves are an implantable device, not a diagnostic AI tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This is not applicable. This device is not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for the performance and biocompatibility testing is based on:
- Standardized test methods: Adherence to established ASTM, ISO, and USP standards for device function, material properties, and biological safety.
- Predefined pass/fail criteria: Each test has specific criteria that the device must meet to be deemed compliant.
8. The sample size for the training set:
This is not applicable. This is a physical medical device, not an AI model or algorithm that requires a training set.
9. How the ground truth for the training set was established:
This is not applicable for the same reason as above.
Ask a specific question about this device
(112 days)
Integra LifeSciences Production Corporation
The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
The Codman Hakim Programmable Valves Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
The Codman Hakim® Precision Fixed Pressure and Programmable Valves are implantable, sterile, single use devices that provide constant intraventricular pressure and drainage of cerebrospinal fluid (CSF) for the management of hydrocephalus. Hydrocephalus is a condition caused by excessive accumulation of CSF in the ventricles of the brain due to a disturbance of CSF secretion, flow, or absorption, which causes a rise in intracranial pressure (ICP). To relieve ICP, CSF can be diverted through a shunting device, such as a Hakim Precision Valve or Hakim Programmable Valve, to another body cavity where it is a subsequently absorbed. Both the Codman Hakim Precision Fixed Pressure and Programmable Valves are pressure regulating valves which maintain intraventricular pressure at a constant level. The Hakim Precision valves are fixed pressure valves and are available in 5 different opening pressure ranges. The Codman Hakim Programmable Valves, not having fixed pressures, permit non-invasive adjustment of the valve opening pressure. The Codman Hakim Programmable Valves can be adjusted to 18 different opening pressure settings.
The provided text describes a 510(k) premarket notification for Hakim Programmable Valves and Hakim Precision Fixed Pressure Valves. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove acceptance criteria for a new device's performance against specific targets.
Therefore, many of the requested details about acceptance criteria, specific device performance, sample sizes, expert involvement, and ground truth establishment are not applicable or not explicitly stated in the provided document. The submission is for minor changes (material and dimensional) to an existing device, and relies on bench testing and an equivalency assessment.
Here's a breakdown of the information that can be extracted and a clear indication of what is not available from the given FDA letter:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative "acceptance criteria" for the device's clinical performance. Instead, it demonstrates that the device passed a series of performance and biocompatibility tests, indicating compliance with established standards. The acceptance criteria for the tests are implied by the "Pass" outcome, meaning they met the requirements of the specified standards (e.g., ISO 7197, ASTM F1980).
Test | Acceptance Criteria (Implied by standard) | Reported Device Performance |
---|---|---|
Product Shelf Life Testing per ASTM F1980 | Met requirements of ASTM F1980 | Pass |
Functional Testing per ISO 7197 | Met requirements of ISO 7197 | Pass |
Functional Testing per ASTM F647 | Met requirements of ASTM F647 | Pass |
MRI Spatial Gradient Testing per ASTM F2052 | Met requirements of ASTM F2052 | Pass |
X-Ray Testing per ASTM F640 | Met requirements of ASTM F640 | Pass |
MTT and MTS Cytotoxicity Studies per ISO 10993-5 | Met requirements of ISO 10993-5 | Pass |
Guinea Pig Maximization Sensitization Studies per ISO 10993-10 | Met requirements of ISO 10993-10 | Pass |
Intracutaneous Irritation Studies in Rabbits per ISO 10993-10 | Met requirements of ISO 10993-10 | Pass |
Acute Systemic Toxicity Study in Mice per ISO 10993-11 | Met requirements of ISO 10993-11 | Pass |
Rabbit Pyrogen Studies per USP 42 – NF 37 | Met requirements of USP 42 – NF 37 | Pass |
Subcutaneous Implantation Studies in Rabbits, 1 Week and 4 Weeks per ISO 10993-6 | Met requirements of ISO 10993-6 | Pass |
Systemic Toxicity and Local Effects Study in Rabbits Following Subcutaneous Implantation, 13 Weeks per ISO 10993-6 and ISO 10993-11 | Met requirements of ISO 10993-6 and ISO 10993-11 | Pass |
Bacterial Reverse Mutation Studies per ISO 10993-3 and ISO/TR 10993-33 | Met requirements of ISO 10993-3 and ISO/TR 10993-33 | Pass |
In Vitro Mouse Lymphoma Studies per ISO 10993-3 and ISO/TR 10993-33 | Met requirements of ISO 10993-3 and ISO/TR 10993-33 | Pass |
Hemolysis on Extract Studies per ISO 10993-4 and ASTM F756 | Met requirements of ISO 10993-4 and ASTM F756 | Pass |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified for each individual test. The document states "All testing was performed on production equivalent devices," implying a sufficient number of units to conduct the referenced tests, but the exact count is not given.
- Data Provenance: Not explicitly stated (e.g., country of origin). The testing is described as "performance bench test results" and "biocompatibility testing results," indicating lab-based testing rather than patient data. These appear to be prospective tests conducted for the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This submission does not involve clinical studies requiring expert ground truth for a test set. The tests are bench and biocompatibility studies against established standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As there are no clinical trials or interpretations of complex data like medical images by multiple experts, no adjudication method is cited.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study was not conducted and is not relevant to this device submission, which involves physical CSF shunting devices, not AI-assisted reading.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a medical device, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the tests performed is the compliance with the specified international and national standards (e.g., ISO 7197, ASTM F1980, ISO 10993 series, USP). The "Pass" outcome indicates the device met the objective criteria defined within these standards.
8. The sample size for the training set
- Not Applicable. This refers to a medical device, not a machine learning model, so there is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
- Not Applicable. See point 8.
Ask a specific question about this device
Page 1 of 11