(125 days)
No
The description mentions digital image processing but does not include any terms or descriptions indicative of AI or ML capabilities.
Yes
The "Intended Use / Indications for Use" section explicitly states that the system is intended for "therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions."
Yes
The "Intended Use / Indications for Use" section explicitly states that theAURORA Surgiscope System is intended "for visualization, diagnostic and/or therapeutic procedures".
No
The device description clearly outlines both hardware components (sheath with LEDs and camera, control unit) and software components (image processing, user interface). It is a system with both physical and digital elements.
Based on the provided information, the AURORA Surgiscope System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use clearly states the system is for "visualization, diagnostic and/or therapeutic procedures" performed in vivo (within the body) during neurosurgery and neuroendoscopy. IVD devices are used to examine specimens in vitro (outside the body), such as blood, urine, or tissue samples, to provide information for diagnosis.
- Device Description: The description details a system for direct visualization of the surgical field within the body using a camera and illumination. This is characteristic of surgical imaging devices, not IVD devices.
- Lack of IVD-Specific Information: The document does not mention any analysis of biological samples, reagents, or laboratory procedures, which are core components of IVD devices.
The AURORA Surgiscope System is a surgical imaging and access device used during procedures performed directly on the patient.
N/A
Intended Use / Indications for Use
The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
Product codes (comma separated list FDA assigned to the subject device)
GWG, GZT
Device Description
The Aurora Surgiscope System consists of two components: (1) a sterile, single use, sheath with integrated illumination LEDs and camera, with an obturator, and (2) a non-sterile, reusable control unit, Image Control Box (ICB).
The sheath is intended to provide access to the surgical site by acting as the insertable portion of the device, as well as the instrument channel to accommodate other surgical tools. Depth markers are present along the length of the sheath for user reference.
At the proximal end of the sheath is the imager, which comprises the following components: LEDs (light emitting diodes), camera (and optical components), and focus knob.
- . The LEDs provide illumination to the surgical field by directing light down the sheath, along the instrument channel.
- The camera captures video image of the surgical field.
The proximal end of the sheath also contains a tab, which serves as the location for the fixation arm to hold/fix the device. To facilitate insertion of the surgical site, an obturator is provided with the device. During insertion, the obturator is fully inserted into the entire unit is advanced to the desired location. The distal end of the obturator is conical in shape to minimize tissue damage. In addition, the proximal handle of the obturator is designed to accommodate various stereotactic instruments for neuronavigation. Once inserted, the obturator is removed.
The ICB is a non-sterile device that provides three main functions in the AURORA Surgiscope System:
- . To power the Surgiscope LEDs and camera
- . To relay the video feed captured by the Surgiscope camera to a display monitor for real-time image visualization
- . To allow the user to make adjustments to the displayed video feed (e.g., contrast, brightness), as well as vary the LED light output.
The user interface is a membrane keypad with buttons located on the ICB that can be depressed for image adjustment, such as zoom, contrast, brightness, and orientation. The ICB is supplied with two cables: A power cable for connection to an AC wall outlet, and a display cable for connection to a high definition surgical monitor.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Video
Anatomical Site
brain, ventricles
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following testing was conducted to demonstrate the safe and substantial equivalence to the predicate device.
- (1) Biocompatibility testing
The AURORA Surgiscope System is intended to come into direct contact with human tissue during use and was evaluated and tested per the requirements and recommendations of the FDA Guidance, "Use of ISO-10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process." The battery of testing included the following tests:
-Cytotoxicity (MEM Elution) -Sensitization (Kligman Maximization) -Irritation (Intracutaneous Injection) -Systemic Toxicity (Systemic Injection) -Hemolysis (In direct) -Materials Mediated Pyrogenicity
-
(2) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the AURORA Surgiscope System consisting of both components, sheath with LEDs and camera, and the ICB. The system complied with the IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-18. -
(3) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by the FDA Guidance. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a major level of concern, since a failure or latent flow in the software could directly result in a serious injury or death to the patient or operator. -
(4) Mechanical and other testing
- Bench testing dimensional, imaging (visualization, resolution, etc.), illumination, tensile . strength between components
- Simulated use testing – clinician evaluation
- . Compatibility - instrument use in instrument channel, use with other devices (i.e., bipolar electrosurgical)
- Particulate Testing per USP .
- Sterilization per ISO 11135-1 to validate a SAL of 10-6 .
- . Packaging and Shelf-life per ISTA 2A and ASTM F1980
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 882.1480 Neurological endoscope.
(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 4, 2020
Rebound Therapeutics Naomi Gong VP of Regulatory Affairs 13900 Alton Parkway Suite 120 Irvine, California 92618
Re: K201840
Trade/Device Name: Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box) Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG, GZT Dated: July 1, 2020 Received: July 2, 2020
Dear Naomi Gong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
1
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
2
Indications for Use
510(k) Number (if known) K201840
Device Name AURORA Surgiscope System
Indications for Use (Describe)
The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) |
---|
☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary (K201840)
SUBMITTER
Rebound Therapeutics 13900 Alton Parkway, Suite 120 Irvine, CA 92618
| Contact Person: | Naomi Gong, RAC
Vice President of Regulatory Affairs |
|-----------------|---------------------------------------------------------|
| Telephone: | (949) 523-6510 |
| Email: | naomi.gong@integralife.com |
| Date Prepared: | November 4, 2020 |
DEVICE
Name of Device: | AURORA Surgiscope System |
---|---|
Regulation Number: | 21 CFR 882.1480, 21 CFR 882.4800 |
Regulation Name: | Neurological endoscope, Self-retaining retractor for neurosurger |
Regulatory Class: | II |
Product Code: | GWG, GZT |
PREDICATE DEVICE
AURORA Surgiscope System, Rebound Therapeutics, K191861 NICO BrainPath, NICO Corp, K172433 (Reference device)
DEVICE DESCRIPTION
The Aurora Surgiscope System consists of two components: (1) a sterile, single use, sheath with integrated illumination LEDs and camera, with an obturator, and (2) a non-sterile, reusable control unit, Image Control Box (ICB).
The sheath is intended to provide access to the surgical site by acting as the insertable portion of the device, as well as the instrument channel to accommodate other surgical tools. Depth markers are present along the length of the sheath for user reference.
At the proximal end of the sheath is the imager, which comprises the following components: LEDs (light emitting diodes), camera (and optical components), and focus knob.
- . The LEDs provide illumination to the surgical field by directing light down the sheath, along the instrument channel.
- The camera captures video image of the surgical field.
The proximal end of the sheath also contains a tab, which serves as the location for the fixation arm to hold/fix the device. To facilitate insertion of the surgical site, an obturator is provided with the device. During insertion, the obturator is fully inserted into the entire unit is advanced to the desired location. The distal end of the obturator is conical in shape to minimize tissue damage. In addition, the proximal handle of the obturator is designed to accommodate various stereotactic instruments for neuronavigation. Once inserted, the obturator is removed.
The ICB is a non-sterile device that provides three main functions in the AURORA Surgiscope System:
4
- . To power the Surgiscope LEDs and camera
- . To relay the video feed captured by the Surgiscope camera to a display monitor for real-time image visualization
- . To allow the user to make adjustments to the displayed video feed (e.g., contrast, brightness), as well as vary the LED light output.
The user interface is a membrane keypad with buttons located on the ICB that can be depressed for image adjustment, such as zoom, contrast, brightness, and orientation. The ICB is supplied with two cables: A power cable for connection to an AC wall outlet, and a display cable for connection to a high definition surgical monitor.
INDICATIONS FOR USE
The Aurora Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
SUMMARY OF NON-CLINICAL TESTING
The following testing was conducted to demonstrate the safe and substantial equivalence to the predicate device.
- (1) Biocompatibility testing
The AURORA Surgiscope System is intended to come into direct contact with human tissue during use and was evaluated and tested per the requirements and recommendations of the FDA Guidance, "Use of ISO-10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process." The battery of testing included the following tests:
-Cytotoxicity (MEM Elution) -Sensitization (Kligman Maximization) -Irritation (Intracutaneous Injection) -Systemic Toxicity (Systemic Injection) -Hemolysis (In direct) -Materials Mediated Pyrogenicity
The sheath with LEDs and camera portion is considered a tissue contacting and externally communicating for a duration of less than 24 hours. The ICB is not intended to have direct or indirect patient contact and therefore biocompatibility for it was not evaluated.
-
(2) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the AURORA Surgiscope System consisting of both components, sheath with LEDs and camera, and the ICB. The system complied with the IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-18. -
(3) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by the FDA Guidance. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a major level of concern, since a failure or latent flow in the software could directly result in a serious injury or death to the patient or operator.
5
- (4) Mechanical and other testing
- Bench testing dimensional, imaging (visualization, resolution, etc.), illumination, tensile . strength between components
- Simulated use testing – clinician evaluation
- . Compatibility - instrument use in instrument channel, use with other devices (i.e., bipolar electrosurgical)
- Particulate Testing per USP .
- Sterilization per ISO 11135-1 to validate a SAL of 10-6 .
- . Packaging and Shelf-life per ISTA 2A and ASTM F1980 _
COMPARISON OF TECHNICAL CHARACTERISTICS TO PREDICATE DEVICE
PREDICATE Device | SUBJECT Device | |
---|---|---|
Aurora Surgiscope System (K191861) | Aurora Surgiscope System | |
INDICATIONS | ||
FOR USE | The Aurora Surgiscope System is intended for use in | |
endoscopic neurosurgery and pure neuroendoscopy | ||
(i.e. ventriculoscopy) for visualization, diagnostic | ||
and/or therapeutic procedures such as | ||
ventriculostomies, biopsies and removal of cysts, | ||
tumors and other obstructions. | The Aurora Surgiscope System is intended for use in | |
neurosurgery, and endoscopic neurosurgery and | ||
pure neuroendoscopy (i.e. ventriculoscopy) for | ||
visualization, diagnostic and/or therapeutic | ||
procedures such as ventriculostomies, biopsies and | ||
removal of cysts, tumors and other obstructions. | ||
Sheath | ■ OD ≤ 11.5mm, ID = 8 mm | |
■ Lengths = 70, 100, 130 mm | ||
■ Incremental depth markings | ||
■ Single working channel | ||
■ With obturator (conical shape, rounded tip) | ■ OD = 16.2 mm, ID = 11 mm | |
■ Lengths = 60 and 80 mm | ||
■ Same | ||
■ Same | ||
■ Same | ||
Materials | Sheath: ABS, PET, adhesives, white ink depth marker | |
Obturator: Polycarbonate, 304 SS, ABS, adhesives | Sheath: Aluminum, adhesives | |
Obturator: Same | ||
Imager | The device incorporates an imaging system with | |
camera and optics (lens, prism): | ||
Direction of View = 0° | ||
Depth of field = 0 to 3 cm | Same | |
Light Source | ||
Camera | 6 LED incorporated into inner dia. of sheath near | |
distal end - fixed intensity level | ||
Integrated CMOS camera and electronics | 4 LEDs incorporated into Imager (proximal end of | |
sheath) - user can vary intensity level | ||
Same | ||
Image Control Box | ||
(control unit) | ■ Electronics (Circuit boards, CPU with software | |
control) | ||
■ Keypad buttons for image adjustment by user | ■ Equivalent | |
■ Same | ||
Other (accessories) | Power supply and cable | |
Display cable | Same | |
Display Monitor | Not supplied | Same |
Access to the | ||
surgical site | Sheath and Obturator used to access the surgical site. | |
The Obturator extends 10mm beyond distal end of | ||
sheath | Same | |
Single working | ||
channel | Removal of obturator reveals working channel which | |
provides visual access for camera and working access | ||
for instruments including suction, bipolar, and | ||
irrigation. | Same | |
Image acquisition | Image acquisition is achieved through an integrated | |
camera | Same | |
Image processing | Image is digitally processed by control unit | Same |
Image display | External display monitor connection | Same |
Illumination | Illumination is achieved via direct LED light sources | |
incorporated into the device. | Same | |
Visualization | CMOS, color, video, camera (with software) | |
incorporated at proximal end of the device and | ||
controlled via control unit. | Equivalent |
6
PREDICATE Device | SUBJECT Device | |
---|---|---|
Aurora Surgiscope System (K191861) | Aurora Surgiscope System | |
Biocompatibility | Surgiscope: Demonstrated based on externally | |
communicating device in direct contact with | ||
tissue/bone/dentin for a limited duration | Same | |
Use and how | ||
supplied | Surgiscope: single use, sterile | |
ICB (control unit): reusable, non-sterile | Same | |
Sterilization | ||
Method | Surgiscope: Ethylene oxide gas | Same |
The NICO BrainPath device (K172433) is included as a reference device because its sheath diameter is comparable and its sheath/obturator has the same intended use to access the surgical field during neurosurgery.
CONCLUSION
Based upon the performance data provided in this submission and comparing indications for use, design, materials, principle of operation and overall technological characteristics, the AURORA Surgiscope System has been determined to be substantially equivalent to the predicate device.