(25 days)
No
The document describes a standard electrosurgical generator and does not mention any AI or ML capabilities.
Yes.
The device is used for cutting and coagulating tissue, sealing vessels, and cardiac tissue ablation, which are all therapeutic actions.
No
The device is described as an electrosurgical generator used for cutting, coagulating, sealing, and ablating tissue. Its primary functions involve delivering energy for surgical procedures, not for diagnosing conditions.
No
The device description clearly states it is a "high frequency electrosurgical generator" and a "combination of a full-featured general-surgery electrosurgical unit, a bipolar vessel sealing system, and an ablation unit," which are hardware components that deliver RF energy. While it contains software, it is not solely software.
Based on the provided information, the Valleylab FT10 Energy Platform is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use clearly describes the device as a high frequency electrosurgical generator used for cutting, coagulating, sealing, and ablating tissue within the body during surgical procedures. This is an in vivo application.
- Device Description: The description reinforces its use in surgical procedures in hospitals and healthcare facilities, providing energy for various surgical applications.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (like blood, tissue samples, etc.) outside the body to provide information for diagnosis, monitoring, or treatment.
Therefore, the Valleylab FT10 Energy Platform is a surgical device used directly on patients, not an IVD.
N/A
Intended Use / Indications for Use
The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. When used with compatible ablation devices, it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.
Product codes
GEI
Device Description
The Valleylab™ FT10 Energy Platform (VLFT10GEN) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, tissue-fusion and vessel-sealing applications (LigaSure vessel-sealing function), and cardiac ablation. It is a combination of a full-featured general-surgery electrosurgical unit, a bipolar vessel sealing system, and an ablation unit. The monopolar and bipolar sections, including the LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure vessel-sealing section of the system provides power for vessel sealing. The cardiac ablation feature of the device is only available when used with compatible Cardioblate™ instruments.
The VLFT10GEN is used in hospitals and other health care facilities where surgical procedures are carried out.
The VLFT10GEN can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The VLFT10GEN connects to electrical mains and operates at an input line frequency of 47-63 Hz.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Verification and validation testing confirmed that the modifications to the Valleylab FT10 Energy Platform do not raise different questions of safety and effectiveness. The testing confirmed the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:
- Ex vivo testing using porcine tissue showed comparable performance with regards to thermal spread
- . System verification showed the VLFT10GEN possesses all required functionality and meets system specifications
- . Software verification and validation testing was conducted in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices [2005]and IEC 62304 Medical device software – Software life cycle processes
- . Confirmatory electrical safety testing showed the VLFT10GEN complies with the 2017 version of IEC 60601-2-2 Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories - Edition 6.0
- Validation testing was conducted in accordance with FDA Guidance Applying Human ● Factors and Usability Engineering to Medical Devices [2016] and IEC 62366-1:2015 Medical Devices – Application of Usability Engineering to Medical Devices
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
July 12, 2019
Covidien Ms. Jennie van Diemen Regulatory Affairs Specialist 5920 Longbow Drive Boulder, Colorado 80301
Re: K191601
Trade/Device Name: Vallevlab FT10 Energy Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: June 13, 2019 Received: June 17, 2019
Dear Ms. Jennie van Diemen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K191601
Device Name Valleylab FT10 Energy Platfom
Indications for Use (Describe)
The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. When used with compatible ablation devices, it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use this function for these procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Image /page/3/Picture/2 description: The image contains the logo for Covidien. The logo consists of a blue square with a lighter blue vertical line and a white square in the center on the left. To the right of the square is the word "COVIDIEN" in a dark blue sans-serif font.
510(k) Summary
Date summary prepared: July 5, 2019
510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
Contact:
Ms. Jennie van Diemen Regulatory Affairs Specialist Telephone: 303-581-7037 Fax: 303-530-6313 Email: jennie.vandiemen@medtronic.com
Name of Device
Trade Name: | Valleylab FT10 Energy Platform |
---|---|
Catalog Numbers: | VLFT10GEN |
Common Name: | Electrosurgical Generator |
Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400, Class II, GEI, OCL) |
Predicate Device
Trade Name: | Valleylab FT10 Energy Platform |
---|---|
Catalog Number: | VLFT10GEN |
Common Name: | Electrosurgical Generator |
510(k) Number: | K182610 (cleared 12/19/2018) |
Manufacturer: | Covidien |
Recalls: | This device has not been subject to a design-related recall |
Reference Device
Trade Name: | Valleylab FX8 Energy Platform |
---|---|
Catalog Number: | VLFX8GEN |
Common Name: | Electrosurgical Generator |
510(k) Number: | K181389 (cleared 06/25/2018) |
Manufacturer: | Covidien |
Recalls: | This device has not been subject to a design-related recall |
4
Device Description
The Valleylab™ FT10 Energy Platform (VLFT10GEN) provides radio frequency (RF) energy for monopolar and bipolar surgical applications, tissue-fusion and vessel-sealing applications (LigaSure vessel-sealing function), and cardiac ablation. It is a combination of a full-featured general-surgery electrosurgical unit, a bipolar vessel sealing system, and an ablation unit. The monopolar and bipolar sections, including the LigaSure section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery. The LigaSure vessel-sealing section of the system provides power for vessel sealing. The cardiac ablation feature of the device is only available when used with compatible Cardioblate™ instruments.
The VLFT10GEN is used in hospitals and other health care facilities where surgical procedures are carried out.
The VLFT10GEN can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The VLFT10GEN connects to electrical mains and operates at an input line frequency of 47-63 Hz.
Indications for Use
The Valleylab FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics. When used with compatible ablation devices, it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium.
The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use this function for these procedures.
Contraindications
The cardiac tissue ablation feature is contraindicated for patients that have active endocarditis at the time of surgery.
Ablation in a pool of blood is contraindicated (for example, through a purse string suture).
Comparison of Technological Characteristics with the Predicate Device
The Valleylab FT10 Energy Platform with the proposed software modifications has similar technological and performance characteristics as the predicate Valleylab FT10 Energy Platform cleared in K182610. Both versions of the energy platform are a combination of a full-featured general-surgery electrosurgical system, a vessel sealing system, and a cardiac ablation system. The Valleylab FX8 Energy Platform, cleared by K181389, serves as a reference device and test control device for the adoption of bipolar modes by the proposed device. The proposed Valleylab FT10 Energy Platform provides increments over the predicate while maintaining the same basic functionality and intended use. These improvements are summarized below.
- Addition of bipolar modes Precise, Standard, and Macro
- Improved Bipolar Resection Cut mode initiation speed
- . Feature allowing the user to save settings and access recent settings
- Bug fixes associated with non-safety related issues
5
Performance Characteristics
Verification and validation testing confirmed that the modifications to the Valleylab FT10 Energy Platform do not raise different questions of safety and effectiveness. The testing confirmed the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:
- Ex vivo testing using porcine tissue showed comparable performance with regards to thermal spread
- . System verification showed the VLFT10GEN possesses all required functionality and meets system specifications
- . Software verification and validation testing was conducted in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices [2005]and IEC 62304 Medical device software – Software life cycle processes
- . Confirmatory electrical safety testing showed the VLFT10GEN complies with the 2017 version of IEC 60601-2-2 Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories - Edition 6.0
- Validation testing was conducted in accordance with FDA Guidance Applying Human ● Factors and Usability Engineering to Medical Devices [2016] and IEC 62366-1:2015 Medical Devices – Application of Usability Engineering to Medical Devices
Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
Conclusions
Verification and validation activities demonstrate that the Valleylab FT10 Energy Platform is substantially equivalent to the predicate Valleylab FT10 Energy Platform cleared in K182610. The intended use of the Valleylab FT10 Energy Platform has not changed. Moreover, the results of testing demonstrate that the software modifications do not affect the safety or performance of the energy platform.