Search Results
Found 2 results
510(k) Data Aggregation
(83 days)
The Medical Nitrile Examination Glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.
The subject device is powder free nitrile examination gloves. The subject device is blue. It can be available in six specifications: S,M,L,XL. The subject device is non-sterile.
The provided text describes the 510(k) summary for the "Medical Nitrile Examination Gloves" (K220066). This document is for a Class I medical device (non-powdered patient examination glove), which is a low-risk device. For such devices, clinical testing is typically not required, and the focus of the submission is on non-clinical performance and substantial equivalence to a predicate device.
Therefore, the requested information regarding AI device performance, sample sizes for test sets, ground truth establishment, expert adjudication, MRMC studies, and standalone algorithm performance does not apply to this specific submission. This document details the testing of physical properties and biocompatibility of examination gloves, not an AI or imaging device.
However, I can extract the acceptance criteria and performance data for the glove product based on the provided non-clinical testing summary.
Here's the relevant information based on the document:
1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical, for Medical Nitrile Examination Gloves)
| Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
|---|---|---|---|
| ASTM D6319 - Physical Dimensions Test | |||
| Length (mm) | S: ≥220; M/L/XL: ≥230 | S: ≥220; Pass M/L/XL: ≥230; Pass | |
| Width (mm) | S: 80±10; M: 95±10; L: 110±10; XL: 120±10 | S: 83-88 /Pass M: 93-96/ Pass L: 104-109/ Pass XL: 110-116/ Pass | |
| Thickness (mm) | Finger: ≥0.05; Palm: ≥0.05 | Finger: 0.102-0.117/Pass Palm: 0.068-0.073/Pass | |
| ASTM D5151 - Watertightness Test | For Detection of Holes | Meet the requirements of ASTM D5151 AQL 2.5 | 0/125/Pass (This implies 0 failures out of 125 samples tested) |
| ASTM D6124 - Powder Content | Meet the requirements of ASTM D6124 < 2.0mg | 0.22mg/Pass | |
| ASTM D412 - Physical Properties | Before Aging | Tensile Strength: ≥14MPa; Ultimate Elongation: ≥500% | Tensile Strength: 22.1-35.9 MPa/Pass; Ultimate Elongation: 505-588%/Pass |
| After Aging | Tensile Strength: ≥14MPa; Ultimate Elongation: ≥400% | Tensile Strength: 21.4-44.3MPa/Pass; Ultimate Elongation: 452-540%/Pass | |
| ISO 10993-5 - Cytotoxicity | Device extract is cytotoxic (This seems to be a reporting of a finding, not an "acceptance criteria" for passing submission, but rather a characteristic. Given no other remarks, it likely indicates the test was performed, and the output confirmed, but not necessarily a "pass/fail" in the typical sense for this specific line. However, the ISO 10993-11 for systemic toxicity and ISO 10993-10 for irritation/sensitization are clearly "pass" for safety.) | Under conditions of the study, device extract is cytotoxic. (Note: For medical devices, this often means the extract was confirmed to cause a cytotoxic effect in vitro, and further evaluation/mitigation or context like the specific application/contact duration, and other biocompatibility tests like irritation and sensitization, inform the overall safety profile.) | |
| ISO 10993-11 - Systemic Toxicity | Non-acute systemic toxicity | N/A (implicit acceptance is non-toxicity) | Under conditions of the study, did not show acute systemic toxicity in vivo / Pass |
| ISO 10993-10 - Irritation | Non-irritating | N/A (implicit acceptance is non-irritating) | Under the conditions of the study, not an irritant/ Pass |
| ISO 10993-10 - Sensitization | Non-sensitizing | N/A (implicit acceptance is non-sensitizing) | Under conditions of the study, not a sensitizer./ Pass |
Points 2-9 are not applicable to this submission as it is not for an AI / imaging device.
Regarding the "study that proves the device meets the acceptance criteria," the "Summary of Non-clinical Testing" (Section 8.0) and "Table 2 - Summary of non-clinical performance testing" provide the evidence.
-
Sample sizes used for the test set and data provenance:
- For the Watertightness Test (ASTM D5151), the result "0/125/Pass" implies a sample size of 125 gloves were tested.
- Other tests (e.g., physical dimensions, strength, thickness, powder content, biocompatibility) would have specific sample sizes per the ASTM/ISO standards, but these are not explicitly stated in the summary table beyond the result itself (e.g., "S: 83-88 /Pass").
- Data provenance: Not explicitly stated, but typically non-clinical testing for medical devices is performed by accredited labs, often in the country of manufacture (China, in this case, as per the submitter's address). This is retrospective for the purpose of the submission.
-
Number of experts used to establish the ground truth for the test set and qualifications of those experts: Not applicable. These are laboratory tests measuring physical and chemical properties, not human interpretation of images or clinical outcomes.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, not applicable.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, not applicable.
-
The type of ground truth used:
- For physical/chemical tests: The "ground truth" is defined by the technical specifications and methodologies outlined in the referenced ASTM and ISO standards (e.g., specific measurement techniques, chemical analyses, biological assays).
- For biocompatibility tests: The "ground truth" is based on the biological responses observed in the in vitro and in vivo models as per ISO 10993 standards.
-
The sample size for the training set: Not applicable (this is not an AI/ML device).
-
How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(86 days)
Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.
The Medical Surgical Mask is single use, three-layer, flat-pleated style with ear loops and nose piece. The Medical Surgical Mask is manufactured with three layers, the inner and outer layers are made of polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the lavers of facemask is to allow the user to fit the facemask around their nose. which is made of Galvanized iron wire. The Medical Surgical Mask will be provided in blue. The Medical Surgical Mask is sold as sterile and are intended to be single use, disposable devices.
Here's a breakdown of the acceptance criteria and study information for the "Medical Surgical Mask" based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Item | Purpose | Acceptance Criteria | Reported Device Performance (Result) |
|---|---|---|---|
| Fluid Resistance Performance ASTM F1862 | Evaluate the Fluid Resistance | 29 Out of 32 pass at 120 mmHg (21.3 kPa) | Pass - Lot 1: 31 Out of 32 pass at 160mmHg (21.3 kPa) - Lot 2: 31 Out of 32 pass at 160mmHg (21.3 kPa) - Lot 3: 32 Out of 32 pass at 160mmHg (21.3 kPa) |
| Particulate Filtration Efficiency ASTM F2299 | Evaluate the Particulate Filtration Efficiency | ≥ 98% | Pass - Lot 1: 99.02%-99.58% - Lot 2: 99.05%-99.59% - Lot 3: 99.10%-99.68% |
| Bacterial Filtration Efficiency ASTM F2101 | Evaluate the Bacterial Filtration Efficiency | ≥ 98% | Pass - Lot 1: 98.8%-99.7% - Lot 2: 98.7%-99.5% - Lot 3: 98.9%-99.5% |
| Differential Pressure (Delta P) MIL-M-36954C | Evaluate the Differential Pressure (Delta P) | < 6.0 mmH2O/cm² | Pass - Lot 1: 1.66-4.44 mmH2O/cm² - Lot 2: 1.63-4.24 mmH2O/cm² - Lot 3: 1.60-4.43 mmH2O/cm² |
| Flammability 16 CFR 1610 | Evaluate the Flammability | Class 1 | Pass, Class 1 - Lot 1: Class I - Lot 2: Class I - Lot 3: Class I |
| Biocompatibility: Cytotoxicity | Evaluate the Cytotoxicity endpoint | Not explicitly stated as a numerical criterion, but implied as "non-cytotoxic" | Pass (Under the conditions of the study, the subject device was non-cytotoxic) |
| Biocompatibility: Irritation | Evaluate the Irritation endpoint | Not explicitly stated as a numerical criterion, but implied as "non-irritating" | Pass (Under the conditions of the study, the subject device was non-irritating) |
| Biocompatibility: Sensitization | Evaluate the Sensitization endpoint | Not explicitly stated as a numerical criterion, but implied as "non-sensitizing" | Pass (Under the conditions of the study, the subject device was non-sensitizing) |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the total sample size for the entire test set in a singular number. However, it indicates the sample size for specific tests per lot:
- Fluid Resistance: 32 samples per lot (tested against 3 lots).
- Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability: The results are presented for 3 distinct 'Lots' (Lot 1, Lot 2, Lot 3). While the exact number of samples per lot for these tests isn't always explicitly stated (except for fluid resistance), the presentation implies multiple measurements per lot.
- Biocompatibility: The sample size for biocompatibility tests (cytotoxicity, irritation, sensitization) is not specified.
Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that HUNAN ZHENHEYIKANG MEDICAL INSTRUMENT CO., LTD is located in China, it's highly probable the testing was conducted there.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not applicable and not provided in the document. The tests performed are objective, quantitative laboratory measurements (e.g., filtration efficiency, fluid resistance) based on established ASTM and military standards, not subjective assessments requiring expert consensus on a "ground truth" (like image interpretation).
4. Adjudication Method for the Test Set
This information is not applicable and not provided. As explained above, the tests are objective laboratory measurements, not subjective assessments requiring an adjudication method among experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There was no MRMC comparative effectiveness study and no AI component mentioned. This document describes the performance testing of a physical medical device (a surgical mask), not an AI-powered diagnostic tool. Therefore, effects on human reader performance with or without AI assistance are not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
There was no standalone (algorithm only) performance study done, as this product is a physical medical device and does not involve an algorithm or AI.
7. The Type of Ground Truth Used
The "ground truth" for the performance tests directly stems from standardized laboratory test results against predefined acceptance criteria. For example:
- For filtration efficiency, the ground truth is the measured percentage of particles/bacteria filtered.
- For fluid resistance, the ground truth is whether the sample passed or failed at a specific pressure.
- For biocompatibility, the ground truth is the biological response observed in the in vitro or in vivo tests.
This is based on objective measurements and established scientific methods, not expert consensus, pathology, or outcomes data in the traditional sense of clinical studies.
8. The Sample Size for the Training Set
This information is not applicable and not provided. There is no "training set" in the context of testing a physical medical device like a surgical mask, as there is no algorithm or model being trained.
9. How the Ground Truth for the Training Set was Established
This information is not applicable and not provided, as there is no training set for this type of device submission.
Ask a specific question about this device
Page 1 of 1