K Number
K213724

Validate with FDA (Live)

Date Cleared
2022-02-20

(86 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

Device Description

The Medical Surgical Mask is single use, three-layer, flat-pleated style with ear loops and nose piece. The Medical Surgical Mask is manufactured with three layers, the inner and outer layers are made of polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the lavers of facemask is to allow the user to fit the facemask around their nose. which is made of Galvanized iron wire. The Medical Surgical Mask will be provided in blue. The Medical Surgical Mask is sold as sterile and are intended to be single use, disposable devices.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Medical Surgical Mask" based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

ItemPurposeAcceptance CriteriaReported Device Performance (Result)
Fluid Resistance Performance ASTM F1862Evaluate the Fluid Resistance29 Out of 32 pass at 120 mmHg (21.3 kPa)Pass - Lot 1: 31 Out of 32 pass at 160mmHg (21.3 kPa) - Lot 2: 31 Out of 32 pass at 160mmHg (21.3 kPa) - Lot 3: 32 Out of 32 pass at 160mmHg (21.3 kPa)
Particulate Filtration Efficiency ASTM F2299Evaluate the Particulate Filtration Efficiency≥ 98%Pass - Lot 1: 99.02%-99.58% - Lot 2: 99.05%-99.59% - Lot 3: 99.10%-99.68%
Bacterial Filtration Efficiency ASTM F2101Evaluate the Bacterial Filtration Efficiency≥ 98%Pass - Lot 1: 98.8%-99.7% - Lot 2: 98.7%-99.5% - Lot 3: 98.9%-99.5%
Differential Pressure (Delta P) MIL-M-36954CEvaluate the Differential Pressure (Delta P)< 6.0 mmH2O/cm²Pass - Lot 1: 1.66-4.44 mmH2O/cm² - Lot 2: 1.63-4.24 mmH2O/cm² - Lot 3: 1.60-4.43 mmH2O/cm²
Flammability 16 CFR 1610Evaluate the FlammabilityClass 1Pass, Class 1 - Lot 1: Class I - Lot 2: Class I - Lot 3: Class I
Biocompatibility: CytotoxicityEvaluate the Cytotoxicity endpointNot explicitly stated as a numerical criterion, but implied as "non-cytotoxic"Pass (Under the conditions of the study, the subject device was non-cytotoxic)
Biocompatibility: IrritationEvaluate the Irritation endpointNot explicitly stated as a numerical criterion, but implied as "non-irritating"Pass (Under the conditions of the study, the subject device was non-irritating)
Biocompatibility: SensitizationEvaluate the Sensitization endpointNot explicitly stated as a numerical criterion, but implied as "non-sensitizing"Pass (Under the conditions of the study, the subject device was non-sensitizing)

2. Sample Size Used for the Test Set and the Data Provenance

The document does not explicitly state the total sample size for the entire test set in a singular number. However, it indicates the sample size for specific tests per lot:

  • Fluid Resistance: 32 samples per lot (tested against 3 lots).
  • Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability: The results are presented for 3 distinct 'Lots' (Lot 1, Lot 2, Lot 3). While the exact number of samples per lot for these tests isn't always explicitly stated (except for fluid resistance), the presentation implies multiple measurements per lot.
  • Biocompatibility: The sample size for biocompatibility tests (cytotoxicity, irritation, sensitization) is not specified.

Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. Given that HUNAN ZHENHEYIKANG MEDICAL INSTRUMENT CO., LTD is located in China, it's highly probable the testing was conducted there.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable and not provided in the document. The tests performed are objective, quantitative laboratory measurements (e.g., filtration efficiency, fluid resistance) based on established ASTM and military standards, not subjective assessments requiring expert consensus on a "ground truth" (like image interpretation).

4. Adjudication Method for the Test Set

This information is not applicable and not provided. As explained above, the tests are objective laboratory measurements, not subjective assessments requiring an adjudication method among experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no AI component mentioned. This document describes the performance testing of a physical medical device (a surgical mask), not an AI-powered diagnostic tool. Therefore, effects on human reader performance with or without AI assistance are not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

There was no standalone (algorithm only) performance study done, as this product is a physical medical device and does not involve an algorithm or AI.

7. The Type of Ground Truth Used

The "ground truth" for the performance tests directly stems from standardized laboratory test results against predefined acceptance criteria. For example:

  • For filtration efficiency, the ground truth is the measured percentage of particles/bacteria filtered.
  • For fluid resistance, the ground truth is whether the sample passed or failed at a specific pressure.
  • For biocompatibility, the ground truth is the biological response observed in the in vitro or in vivo tests.

This is based on objective measurements and established scientific methods, not expert consensus, pathology, or outcomes data in the traditional sense of clinical studies.

8. The Sample Size for the Training Set

This information is not applicable and not provided. There is no "training set" in the context of testing a physical medical device like a surgical mask, as there is no algorithm or model being trained.

9. How the Ground Truth for the Training Set was Established

This information is not applicable and not provided, as there is no training set for this type of device submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 20, 2022

Hunan Zhenheyikang Medical Instrument Co., Ltd % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K213724

Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: February 10, 2022 Received: February 16, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213724

Device Name Medical Surgical Mask

Indications for Use (Describe)

Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary K213724

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 submitter's Information

Name: HUNAN ZHENHEYIKANG MEDICAL INSTRUMENT CO., LTD

Address: No.6 Building Jingxiang Energy,No.55 Xiaguang East Road,Gaoxin District, Xiangtan, Hunan, China

Tel:+86-0731-58262222

Contact: Yilin Yin

Date of Preparation: Nov.12, 2021

Designated Submission Correspondent

Mr. Boyle Wang

Shanghai Truthful Information Technology Co., Ltd.

Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932

Email: Info@truthful.com.cn

2.0 Device Information

Trade name: Medical Surqical Mask Common name: Surgical Face Mask Classification name: Surgical Face Mask Model: Ear-loop Type

{4}------------------------------------------------

3.0 Classification

Production code: FXX Regulation number: 21CFR 878.4040 Classification: Class II

Panel: Surgical Apparel

4.0 Predicate Device Information

Manufacturer: Jiangsu Xingtong Biotechnology Group Co., Ltd. Device: Surgical mask

510(k) number: K211454

5.0 Device Description

The Medical Surgical Mask is single use, three-layer, flat-pleated style with ear loops and nose piece. The Medical Surgical Mask is manufactured with three layers, the inner and outer layers are made of polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the lavers of facemask is to allow the user to fit the facemask around their nose. which is made of Galvanized iron wire. The Medical Surgical Mask will be provided in blue. The Medical Surgical Mask is sold as sterile and are intended to be single use, disposable devices.

6.0 Indication for Use Statement

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided as sterile.

7.0 Technological Characteristic Comparison

{5}------------------------------------------------

ItemSubject DevicePredicate DeviceComparison
K213724K211454
Product NameMedical Surgical MaskSurgical mask--
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassIIIISame
Intended Use&Indications for useMedical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. It is intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s),The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use,Same
provided as sterile.
disposable device,provided as sterile.
Design featuresEar Loops,3 layersEar loops: XT10A1; Tie-on: XT10B1;3 layersDifferentAnalysis 1
Mask StylesFlat pleatedFlat pleatedSame
MaterialOuterfacinglayerPolypropylenePolypropyleneSame
MiddlelayerMelt-blown cloth(polypropylene)Melt-blown cloth(polypropylene)Same
InnerFacinglayerPolypropylenePolypropyleneSame
NosepieceGalvanized iron wirePolyethylene coatedsteel wireDifferentAnalysis 1
Earloops78% Polyester + 22%spandex-Ear loops: Polyester silk &Polyurethane filamentSimilarAnalysis 1
ColorBlueBlueSame
Dimension175mm×95mm±5%Mask body for ear-looptype: 17.5cm×9.5cm &14.5cm×9.5cmMask body for Tie-on type:17.5cm×9.5cmSimilarAnalysis2
OTC useYesYesSame
Shelf life2 years2 yearsSame
Single UseYesYesSame
SterilitySterileSterileSame
Sterilizationmethod andS.A.L.Sterilized by ethylene oxidegas, SAL=10-6Sterilized by ethylene oxidegas, SAL=10-6Same
ASTM F2100 LevelLevel 3Level 3Same
Bacterial filtrationefficiency (BFE) (%)Lot 1# 31 Out of 32 pass at160mmHg (21.3 kPa);Lot 2# 31 Out of 32 pass at160mmHg (21.3 kPa);Lot 3# 32 Out of 32 pass at160mmHg (21.3 kPa).Lot 1# pass at 160mmHg;Lot 2# pass at 160mmHg;Lot 3# pass at 160mmHg.Similar
Different pressure(mmH2O/cm²)Lot1#99.02%-99.58%;Lot2#99.05%-99.59%;Lot3#99.10%-99.68%.Lot 1# 99.7%-99.9%;Lot 2# 99.7%-99.9%;Lot 3# 99.7%-99.9%.Similar
Sub-micron particulatefiltration efficiency at0.1 micron, % (PFE)Lot1#98.8%-99.7%;Lot2#98.7%-99.5%;Lot3#98.9%-99.5%.Lot 1# 99.2%-99.8%;Lot 2# 99.0%-99.7%;Lot 3# 99.3%-99.7%.Similar
Resistance topenetration bysynthetic blood,Minimum pressurein mmHg for passresultLot1#1.66-4.44mmH20/cm²Lot2#1.63-4.24mmH20/cm²Lot3#1.60-4.43mmH20/cm²Lot 1# 2.3-4.6;Lot 2# 2.3-4.5;Lot 3# 2.0-3.9Similar
Flame spreadLot 1# Class I;Lot 2# Class I;Lot 3# Class ILot 1# Class I;Lot 2# Class I;Lot 3# Class ISame
ItemPurposeAcceptanceCriteriaResult
Fluid ResistancePerformance ASTMF1862The purpose of themethodology is toevaluate the FluidResistance29 Out of 32 pass at 120mmHg (21.3 kPa)PassLot 1# 31 Out of 32 passat 160mmHg (21.3 kPa)Lot 2# 31 Out of 32 passat 160mmHg (21.3 kPa)Lot 3# 32 Out of 32 passat 160mmHg (21.3 kPa)
ParticulateFiltration EfficiencyASTM F2299The purpose of themethodology is toevaluate theParticulate FiltrationEfficiency≥ 98%PassLot1#99.02%-99.58%;Lot2#99.05%-99.59%;Lot3#99.10%-99.68%
Bacterial FiltrationEfficiencyASTM F2101The purpose of themethodology is toevaluate the BacterialFiltration Efficiency≥ 98%PassLot1#98.8%-99.7%;Lot2#98.7%-99.5%;Lot3#98.9%-99.5%
DifferentialPressure(Delta P) MILM- 36954CThe purpose of themethodology is toevaluate theDifferential Pressure(Delta P)< 6.0 mmH20/cm²PassLot1#1.66-4.44mmH20/cm²Lot2#1.63-4.24mmH20/cm²Lot3#1.60-4.43mmH20/cm2
Flammability16 CFR 1610The purpose of themethodology is toevaluate theFlammabilityClass 1Pass,Class 1

Table 1 General Comparison

{6}------------------------------------------------

{7}------------------------------------------------

Analysis1: the two devices have some difference in design features and materials, product materials safety is proved by its biocompatibility.

Analysis2: the two devices share same dimensions otherwise the tolerance is different.

Analysis 3: The two devices have some little deviation in product performance, but the difference in the performance test result does not raise additional questions for safety and effectiveness since the two device performance results both meet the requirements of ASTM F2100-19, level 3.

8.0 Non-Clinical Test Conclusion

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Medical face masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.

{8}------------------------------------------------

Table 2 - Performance Testing

Table 3 - Biocompatibility Testing

tomﻟﺘﺎﺭﻳﺦ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪPurposeResult
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{9}------------------------------------------------

CytotoxicityThe purpose of the methodology is toevaluate the Cytotoxicity endpoint.PassUnder the conditions of the study,the subject device was non-cytotoxic
IrritationThe purpose of the methodology is toevaluate the Irritation endpoint.PassUnder the conditions of the study,the subject device was non-irritating
SensitizationThe purpose of the methodology is toevaluate the Sensitization endpoint.PassUnder the conditions of the study,the subject device was non-sensitizing

9.0_Clinical Test Conclusion

No clinical study is included in this submission.

10.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K211454.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.