Search Results
Found 105 results
510(k) Data Aggregation
(90 days)
The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
This assay is intended to aid in the differential diagnosis of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.
The Panther Fusion GI Expanded Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Yersinia enterocolitica, Vibrio (V. parahaemolyticus, V. vulnificus, V. cholerae), Escherichia coli O157, and Plesiomonas shigelloides. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.
Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.
Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).
Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction. Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.
The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:
| Analyte | Gene Targeted | Instrument Channel |
|---|---|---|
| Yersinia enterocolitica | InvA (Invasive antigen A) | FAM |
| Vibrio parahaemolyticus | gyrB (Gyrase B) | HEX |
| Vibrio vulnificus | gyrB (Gyrase B) | HEX |
| Vibrio cholerae | ompW (Outer Membrane Protein W) | HEX |
| Escherichia coli O157 | rfbE (Perosamine synthase-O-antigen) | ROX |
| Plesiomonas shigelloides | hugA (Heme utilization gene A) | RED647 |
| Internal Control | Not Applicable | RED677 |
Assay Components: The assay components configuration for the Panther Fusion GI Expanded Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Expanded Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Expanded Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:
- Aptima Assay Fluids Kit (303014)
- Aptima Multitest Swab Specimen Collection Kit (PRD-03546)
Table 1: Reagents Required to Perform the Panther Fusion GI Expanded Bacterial Assay
| Box | Components Description |
|---|---|
| Refrigerated Box | Panther Fusion GI Expanded Bacterial Assay Cartridges |
| Room Temperature Box | Panther Fusion Extraction Reagent-B• Panther Fusion Capture Reagent-B bottles• Panther Fusion Enhancer Reagent-B bottles |
| Refrigerated Box | Panther Fusion Internal Control-B |
| Room Temperature Box | Panther Fusion Reconstitution Buffer I |
| Room Temperature Box | Panther Fusion Elution Buffer |
| Room Temperature Box | Panther Fusion Oil |
| Refrigerated Box | Panther Fusion GI Expanded Bacterial Assay Controls• Panther Fusion GI Expanded Bacterial Positive Control• Panther Fusion Negative Control |
Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Expanded Bacterial Assay
| Aptima Assay Fluids Kit |
|---|
| Aptima Multitest Swab Specimen Collection Kit |
Instrumentation: The Panther Fusion GI Expanded Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Expanded Bacterial Assay.
N/A
Ask a specific question about this device
(99 days)
The Panther Fusion GI Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test for the rapid and qualitative detection and differentiation of Salmonella, Shigella/Enteroinvasive Escherichia coli (EIEC), Campylobacter (C. coli, C. jejuni) nucleic acids and Shiga-toxin producing Escherichia coli Shiga toxins 1 and 2 (undifferentiated) genes. Nucleic acids are isolated and purified from Cary-Blair preserved stool specimens collected from individuals exhibiting signs and symptoms of gastroenteritis.
This assay is intended to aid in the differential diagnosis of Salmonella, Campylobacter, Shigella/Enteroinvasive E. coli (EIEC) and Shigatoxigenic Escherichia coli (STEC) infections. The results of this assay should be used in conjunction with clinical presentation, laboratory findings, and epidemiological information and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out co-infection with other organisms that are not detected by this test and may not be the sole or definitive cause of patient illness. Negative results in the setting of clinical illness compatible with gastroenteritis may be due to infection by pathogens that are not detected by this test, or non-infectious causes such as ulcerative colitis, irritable bowel syndrome, or Crohn's disease. This assay is designed for use on the Panther Fusion System.
The Panther Fusion GI Bacterial Assay is a multiplex real-time PCR in vitro diagnostic test developed for use on the fully automated Panther Fusion system to detect nucleic acids from Salmonella, Shigella/Enteroinvasive Escherichia coli (EIEC), Campylobacter (C. coli, C. jejuni) and Shiga-toxin producing Escherichia coli Shiga toxins 1 and 2 (undifferentiated) genes.
The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Bacterial Assay. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion System. The eluate is transferred to the Panther Fusion System reaction tube containing the assay reagents. Multiplex real-time PCR is then performed for the eluted nucleic acid on the Panther Fusion System.
Sample processing: Prior to processing and testing on the Panther Fusion System, specimens are transferred to an Aptima Multitest tube containing specimen transport media (STM) that lyses the cells, releases target nucleic acid, and protects them from degradation during storage.
Nucleic acid capture and elution: An internal control (IC-B) is added automatically to each specimen via the working Panther Fusion Capture Reagent-B (wFCR-B) to monitor for interference during specimen processing, amplification, and detection caused by reagent failure or inhibitory substances. Specimens are first incubated in an alkaline reagent (FER-B) to enable cell lysis. Nucleic acid released during the lysis step hybridizes to magnetic particles in the wFCR-B. The capture particles are then separated from residual specimen matrix in a magnetic field by a series of wash steps with a mild detergent. The captured nucleic acid is then eluted from the magnetic particles with a reagent of low ionic strength (Panther Fusion Elution Buffer).
Multiplex PCR amplification and fluorescence detection: Lyophilized single unit dose reaction master mix is reconstituted with the Panther Fusion Reconstitution Buffer I and then combined with the eluted nucleic acid into a reaction tube. Panther Fusion Oil reagent is added to prevent evaporation during the PCR reaction. Target-specific primers and probes then amplify targets via polymerase chain reaction while simultaneously measuring fluorescence of the multiplexed targets. The Panther Fusion System compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of each analyte.
The analytes and the channel used for their detection on the Panther Fusion System are summarized in the table below:
| Analyte | Gene Targeted | Instrument Channel |
|---|---|---|
| Salmonella | InvA (Invasive antigen A) | FAM |
| Campylobacter | glyA (serine hydroxymethyl transferase)/cadF (outer membrane fibronectin-binding protein) | HEX |
| Shigella/EIEC | ipaH (Invasion plasmid antigen H) | ROX |
| STEC | stx1 (Shigatoxin 1)/stx2 (Shigatoxin 2) | RED647 |
| Internal Control | Not Applicable | RED677 |
Assay Components
The assay components configuration for the Panther Fusion GI Bacterial Assay is analogous to the Panther Fusion Respiratory Assays. The reagents required to perform the Panther Fusion GI Bacterial Assay are packaged and sold separately. There are 7 boxes containing 9 reagents which are required for sample processing. The Panther Fusion GI Bacterial Assay requires one ancillary kit and one specimen collection kit, neither of which are provided with the assay and can be acquired separately:
- Aptima Assay Fluids Kit (303014)
- Aptima Multitest Swab Specimen Collection Kit (PRD-03546)
Table 1: Reagents Required to Perform the Panther Fusion GI Bacterial Assay
| Box | Components Description |
|---|---|
| Refrigerated Box | Panther Fusion GI Bacterial Assay Cartridges |
| Room Temperature Box | Panther Fusion Extraction Reagent-B |
| Refrigerated Box | Panther Fusion Internal Control-B |
| Room Temperature Box | Panther Fusion Reconstitution Buffer I |
| Room Temperature Box | Panther Fusion Elution Buffer |
| Room Temperature Box | Panther Fusion Oil |
| Refrigerated Box | Panther Fusion GI Bacterial Assay Controls |
Table 2: Ancillary and Collection Kits Required to Perform the Panther Fusion GI Bacterial Assay
| Kit |
|---|
| Aptima Assay Fluids Kit |
| Aptima Multitest Swab Specimen Collection Kit |
Instrumentation
The Panther Fusion GI Bacterial Assay has been designed for and validated on the Panther Fusion system. The Panther Fusion System fully automates specimen processing, including sample lysis, nucleic acid capture, amplification, and detection for the Panther Fusion GI Bacterial Assay.
N/A
Ask a specific question about this device
(279 days)
Genius AI Detection is a computer-aided detection and diagnosis (CADe/CADx) software device intended to be used with compatible digital breast tomosynthesis (DBT) systems to identify and mark regions of interest including soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in DBT exams from compatible DBT systems and provide confidence scores that offer assessment for Certainty of Findings and a Case Score.
The device intends to aid in the interpretation of digital breast tomosynthesis exams in a concurrent fashion, where the interpreting physician confirms or dismisses the findings during the reading of the exam.
Genius AI Detection 2.0 is a software device intended to identify potential abnormalities in breast tomosynthesis images. Genius AI Detection 2.0 analyzes each standard mammographic view in a digital breast tomosynthesis examination using deep learning networks. For each detected lesion, Genius AI Detection 2.0 produces CAD results that include:
- the location of the lesion;
- an outline of the lesion;
- a confidence score for the lesion
- Genius AI Detection 2.0 also produces a case score for the entire breast tomosynthesis exam.
Genius AI Detection 2.0 packages all CAD findings derived from the corresponding analysis of a tomosynthesis exam into a DICOM Mammography CAD SR object and distributes it for display on DICOM compliant review workstations. The interpreting physician will have access to the CAD findings concurrently to the reading of the tomosynthesis exam. In addition, a combination of peripheral information such as number of marks and case scores may be used on the review workstation to enhance the interpreting physician's workflow by offering a better organization of the patient worklist.
Here's a breakdown of the acceptance criteria and study details for Genius AI Detection 2.0, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Device Performance for Genius AI Detection 2.0
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes a non-inferiority study to demonstrate that the performance of Genius AI Detection 2.0 on Envision (ENV) images is equivalent to its performance on the predicate's Standard of Care (SOC) images (Hologic's Selenia Dimensions systems). The primary acceptance criterion was non-inferiority of the Area Under the Curve (AUC) of the ROC curve, with a 5% margin. Secondary metrics included sensitivity, specificity, and false marker rate per view.
| Acceptance Criteria Category | Specific Metric | Predicate Device Performance (SOC Images) | Subject Device Performance (ENV Images) | Acceptance Criteria Met? |
|---|---|---|---|---|
| Primary Endpoint (Non-Inferiority) | AUC of ROC Curve (ENV-SOC) | N/A (Comparison study) | -0.0017 (95% CI -0.023 - 0.020) | Yes (p-value for difference = 0.87, indicating no significant difference, and within 5% non-inferiority margin) |
| Secondary Metrics | Sensitivity | N/A (Comparison study) | No significant difference reported between modalities | Yes |
| Specificity | N/A (Comparison study) | No significant difference reported between modalities | Yes | |
| False Marker Rate per View | N/A (Comparison study) | No significant difference reported between modalities | Yes | |
| CC-MLO Correlation | Accuracy on Malignant Lesions | N/A | 90% | Yes (Considered accurate) |
| Accuracy on Negative Cases (Correlated pairs) | N/A | 73% | Yes (Considered accurate) | |
| Implant Cases | Location-specific cancer detection sensitivity | N/A | 76% (CI 68%~84%) | Yes (Considered acceptable based on confidence intervals) |
| Specificity | N/A | 67% (CI 62%~72%) | Yes (Considered acceptable based on confidence intervals) |
(Note: The document focuses on demonstrating equivalence to the predicate's performance on a new platform rather than absolute performance against a fixed threshold for all metrics, except for the implant case where specific CIs are given and deemed acceptable.)
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Main Comparison Study): 1475 subjects
- 200 biopsy-proven cancer subjects
- 275 biopsy-proven benign subjects
- 78 BI-RADS 3 subjects (considered BI-RADS 1 or 2 upon diagnostic workup)
- 922 BI-RADS 1 and 2 subjects (at screening)
- Implant Case Test Set: 480 subjects
- 132 biopsy-proven cancer subjects
- 348 negative subjects (119 biopsy-proven benign, 229 screening negative)
- Data Provenance:
- Country of Origin: Not explicitly stated, but collected from a "national multi-center breast imaging network" within the U.S., implying U.S. origin.
- Retrospective or Prospective: The main comparison study data was collected for evaluating the safety and effectiveness of the Envision platform, with an IRB approved protocol. This suggests a retrospective study design, where existing images were gathered for evaluation. The implant cases were collected between 2015 and 2022, also indicating a retrospective approach.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Two
- Qualifications: Both were MQSA-certified radiologists with over 20 years of experience.
4. Adjudication Method for the Test Set
The document explicitly states that the "ground truthing to evaluate performance metrics including the locations of cancer lesions was done by two MQSA-certified radiologists with over 20 years of experience."
- Adjudication Method: It does not specify a particular adjudication method (e.g., 2+1, 3+1). It simply states that ground truthing was done by two experts. This implies either consensus was reached between the two, or potentially an unstated arbitration method if they disagreed, or that their individual findings were used for analysis. Given the phrasing, expert consensus is the most likely implied method, but not explicitly detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was NOT done. The study described is a standalone performance comparison of the AI algorithm on images from different modalities (Envision vs. Standard of Care), not a study involving human readers with and without AI assistance to measure effect size.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone study WAS done. The document explicitly states, "A standalone study was conducted to compare the detection performance of FDA cleared Genius AI Detection 2.0 (K221449) using Standard of Care (SOC) images acquired on the Dimensions systems against images acquired on the FDA approved Envision Mammography Platform (P080003/S009)." This study evaluated the algorithm's performance (fROC, ROC, sensitivity, specificity, false marker rate) directly against the ground truth without human intervention.
7. The Type of Ground Truth Used
- Ground Truth Type: A combination of biopsy-proven cancer and biopsy-proven benign cases, along with BI-RADS diagnostic outcomes (for negative cases). For the cancer cases, the "locations of cancer lesions" were part of the ground truth.
8. The Sample Size for the Training Set
- Not provided. The document states that the test dataset was "sequestered from any training datasets by isolating it on a secured server with controlled access permissions" and that the data for implant cases was "sequestered from the training datasets for Genius AI Detection." However, the actual sample size of the training set is not mentioned.
9. How the Ground Truth for the Training Set Was Established
- Not provided. Since the training set sample size and details are not disclosed, the method for establishing its ground truth is also not mentioned in this document. It is generally assumed that similar rigorous methods (e.g., biopsy-proven truth, expert review) would have been used for training data, but this specific filing does not detail it.
Ask a specific question about this device
(105 days)
The Aptima® SARS-CoV-2 Assay is a nucleic acid amplification in vitro diagnostic test intended for the qualitative detection of RNA from severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) isolated and purified from nasopharyngeal (NP) swab and anterior nasal (AN) swab specimens obtained from patients with signs and symptoms of COVID-19.
Positive results are indicative of the presence of SARS-CoV-2 RNA. The Aptima SARS-CoV-2 Assay is intended for use as an aid in the diagnosis of COVID-19 if used in conjunction with other clinical, epidemiological, and laboratory findings. Clinical correlation with patient history and other diagnostic information is necessary to determine patient infection status. Positive results do not rule out bacterial infection or co-infection with other viruses.
Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information.
The Aptima SARS-CoV-2 Assay is a nucleic acid amplification in vitro diagnostic test developed for use on the fully automated Panther/Panther Fusion system to detect RNA from SARS-CoV-2 isolated and purified from nasopharyngeal and anterior nasal swab specimens collected into UTM/VTM or with the RespDirect Collection Kit.
The Aptima SARS-CoV-2 Assay combines the technologies of target capture, Transcription Mediated Amplification (TMA), and Dual Kinetic Assay (DKA).
Specimens are collected and transferred into their respective specimen transport tubes. The transport solutions in these tubes release the RNA target and protect them from degradation during storage. When the Aptima SARS-CoV-2 Assay is performed in the laboratory, the target RNA molecules are isolated from specimens by use of capture oligomers via target capture that utilizes magnetic microparticles. The capture oligomers contain sequences complementary to specific regions of the target molecules as well as a string of deoxyadenosine residues. A separate capture oligomer is used for each target. During the hybridization step, the sequence specific regions of the capture oligomers bind to specific regions of the target molecules. The capture oligomer:target complex is then captured out of solution by decreasing the temperature of the reaction to room temperature. This temperature reduction allows hybridization to occur between the deoxyadenosine region on the capture oligomer and the poly-deoxythymidine molecules that are covalently attached to the magnetic particles. The microparticles, including the captured target molecules bound to them, are pulled to the side of the reaction vessel using magnets and the supernatant is aspirated. The particles are washed to remove residual specimen matrix that may contain amplification reaction inhibitors. After the target capture steps are completed, the specimens are ready for amplification.
Target amplification assays are based on the ability of complementary oligonucleotide primers to specifically anneal and allow enzymatic amplification of the target nucleic acid strands. The Aptima SARS-CoV-2 Assay replicates specific regions of the RNA from SARS-CoV-2 virus. Detection of the RNA amplification product sequences (amplicon) is achieved using nucleic acid hybridization. Single-stranded chemiluminescent nucleic acid probes, which are unique and complementary to a region of each target amplicon and Internal Control (IC) amplicon, are labeled with different acridinium ester (AE) molecules. The AE-labeled probes combine with the amplicon to form stable hybrids. The Selection Reagent differentiates hybridized from unhybridized probe, eliminating the generation of signal from the unhybridized probe. During the detection step, light emitted from the labeled hybrids is measured as photon signals in a luminometer and are reported as Relative Light Units (RLU). In DKA, differences in the kinetic profiles of the labeled probes allow for the differentiation of signal; kinetic profiles are derived from measurements of photon output during the detection read time.
The Aptima SARS-CoV-2 Assay amplifies and detects 2 conserved regions of the ORF1ab gene in the same reaction, using the "glower" kinetic type. The 2 regions are not differentiated and amplification of either or both regions lead to RLU signal. The assay results are determined by a cut-off based on the total RLU and the kinetic curve type.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the performance metrics presented as evidence of substantial equivalence to the predicate device. The key metrics are Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA). While explicit numerical acceptance thresholds are not provided in the document, the reported performance demonstrates "comparable" results to the predicate device.
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance (NP Swab, Overall) | Reported Device Performance (AN Swab, UTM/VTM) | Reported Device Performance (AN Swab, RespDirect eSTM) |
|---|---|---|---|---|
| Clinical Performance: | ||||
| Positive Percent Agreement (PPA) | High Agreement | 95.4% (90.3-97.9% CI) | 96.5% (92.1-98.5% CI) | 100% (96.6-100% CI) |
| Negative Percent Agreement (NPA) | High Agreement | 99.5% (98.9-99.8% CI) | 97.6% (96.5-98.4% CI) | 98.0% (96.9-98.7% CI) |
| Analytical Performance (Key): | ||||
| Limit of Detection (LoD) - Cultured Virus | Extremely Low Concentration | 0.01 TCID50/mL | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Limit of Detection (LoD) - WHO Standard | Extremely Low Concentration | 87.5 IU/mL | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Reactivity (detection of strains) | 100% Positivity at low concentrations | Demonstrated at 0.03-0.151 TCID50/mL for tested strains; 99.98% by in silico analysis | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Specificity/Microbial Interference | No cross-reactivity/interference | No observed cross-reactivity/interference for 48 tested organisms | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Interfering Substances | No impact on performance | No impact on performance for tested substances | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Carryover Contamination Rate | Low/None | 0% (0/294) | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
| Assay Precision (Total %CV of signal) | Low variability | 2.75% to 3.84% (Negative, Low, Moderate Positive panels) | Not applicable (analytical, not specimen-type specific) | Not applicable (analytical, not specimen-type specific) |
Study Details:
-
Sample Sizes Used for the Test Set and Data Provenance:
- Clinical Study 1 (NP Swab Specimens):
- Total Enrolled: 1646 specimens
- Evaluable (Final Data Set): 1495 NP swab specimens (1195 fresh, 300 frozen)
- Provenance: Prospective multicenter study from four participating US pediatric/adolescent, private and/or university hospitals. Specimens collected between June-July 2020 and January-April 2023. Remnant specimens.
- Clinical Study 2 (Anterior Nasal Swab Specimens):
- Total Enrolled Subjects: 2301
- Evaluable Subjects: 2177 individuals (1159 with evaluable anterior nasal swab specimens in UTM/VTM, and 1018 with evaluable nasal swab specimens in eSTM).
- Provenance: Prospective, multicenter clinical study at nine geographically and ethnically diverse US sites during the 2022-2023 respiratory season.
- Clinical Study 1 (NP Swab Specimens):
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
- The document does not specify the number or qualifications of experts directly establishing the clinical ground truth. Instead, it relies on a Composite Comparator Algorithm (CCA). The CCA consists of "two highly sensitive US FDA EUA SARS-CoV-2 molecular tests" and a "validated PCR followed by bi-directional sequencing (PCR/BDS) assay." This implies that the 'expertise' comes from the validation and regulatory clearance of these comparator assays, rather than individual human experts adjudicating each case.
-
Adjudication Method for the Test Set:
- Composite Comparator Algorithm (CCA): "A final CCA result was assigned when two of the three comparator assay results were in concordance." This serves as the adjudication method for determining the true positive/negative status of the clinical samples.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No. This type of study (MRMC) is typically performed for diagnostic imaging devices where human readers interpret images with and without AI assistance. The Aptima SARS-CoV-2 Assay is an in vitro diagnostic (IVD), a laboratory test that detects nucleic acids, not an imaging device requiring human interpretation of visual data. Therefore, an MRMC study is not applicable here.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Yes, this is effectively a standalone performance study. The Aptima SARS-CoV-2 Assay is an automated nucleic acid amplification test run on the Panther/Panther Fusion system. Its performance is directly compared to the CCA, without human intervention in the interpretation of the device's results.
-
The Type of Ground Truth Used:
- Clinical Studies: Composite Comparator Algorithm (CCA) based on the concordance of results from two highly sensitive FDA EUA SARS-CoV-2 molecular tests and a validated PCR/BDS assay. This is a form of reference standard derived from established and highly sensitive laboratory methods.
- Analytical Studies (e.g., LoD, Reactivity, Specificity): The ground truth was established by known concentrations of SARS-CoV-2 virus strains or other microorganisms/interfering substances, prepared in controlled laboratory settings (e.g., "spiked with inactivated cultured SARS-CoV-2 virus").
-
The Sample Size for the Training Set:
- The document describes performance evaluation studies (analytical and clinical) for market clearance. It does not provide information on the specific training set size used for the development or training of the assay's internal algorithms (e.g., for the kinetic curve analysis or cut-off determination). This information would typically be part of the assay development and validation, not necessarily detailed in a 510(k) summary unless it significantly changed or impacted performance during the clearance process for the specific assay rather than the underlying platform. The focus here is on the performance of the final, already "trained" device.
-
How the Ground Truth for the Training Set Was Established:
- As noted above, details regarding the training set's ground truth establishment are not provided in this 510(k) summary. For IVD devices, ground truth for training internal algorithms typically involves using characterized positive and negative clinical samples, spiked samples with known viral loads, and potentially synthetic data, all carefully confirmed by highly sensitive reference methods or gold standard assays during the R&D phase of the product.
Ask a specific question about this device
(28 days)
The Aptima® CMV Quant Assay is an in vitro nucleic acid amplification test for the quantitation of human cytomegalovirus (CMV) DNA in human EDTA plasma on the fully automated Panther® system.
The Aptima CMV Quant Assay is intended for use to aid in the management of solid-organ transplant patients and hematopoietic stem cell transplant patients receiving anti-CMV therapy, serial DNA measurements can be used to assess viral response to treatment. The results from Aptima CMV Quant assay must be interpreted within the context of all relevant clinical and laboratory findings.
The Aptima CMV Quant Assay is not intended for use as a screening assay for the presence of CMV in blood or blood products.
The Aptima CMV Quant Assay is an in vitro nucleic acid amplification test that uses real-time transcription mediated amplification (TMA) technology on the Panther Fusion system to quantify CMV DNA, genotypes 1, 2, 3, and 4. The primer design targets the highly conserved UL56 gene to ensure accurate quantitation of the CMV DNA. The assay is standardized to the WHO International Standard for human cytomegalovirus. The Aptima CMV Quant Assay is intended for use to aid in the management of solid-organ transplant patients and hematopoietic stem cell transplant patients.
The addition of a protocol to treat specific plasma samples with proteinase K in the assay package insert will allow the retesting of those samples to obtain a valid result.
The provided text describes a 510(k) premarket notification for the "Aptima CMV Quant Assay" and specifically details an update to the assay's package insert to include a protocol for retesting plasma samples that initially yield an "ML2 flag" invalid result. This retesting involves a manual pretreatment with proteinase K. The information provided is for a diagnostic assay, not an AI/ML-based device. Therefore, many of the requested elements pertaining to AI/ML development, such as training sets, ground truth establishment for AI/ML, expert consensus for imaging, MRMC studies, and effect sizes of AI assistance, are not applicable.
However, I can extract information related to the acceptance criteria and the study that proves the device (specifically, the proteinase K pretreatment protocol) meets the acceptance criteria within the context of a diagnostic assay's performance and validation.
Acceptance Criteria and Study for the Proteinase K Pretreatment Protocol
The primary acceptance criterion for this change is to demonstrate that the proteinase K pretreatment protocol effectively resolves the "ML2 flag" invalid results and does not negatively impact the accuracy of the CMV quantification.
1. Table of Acceptance Criteria and the Reported Device Performance:
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Reduction/Elimination of ML2 Flag Invalid Results: The protocol should effectively enable valid results for samples initially flagged with ML2. | The prevalence of specimens invalidated with ML2 flags at the NIH decreased from 0.87% (on testing 1039 plasma specimens without proteinase K treatment) to 0% (on testing 4098 specimens with proteinase K treatment) when the proteinase K protocol was used. |
| Maintenance of Accuracy: The proteinase K pretreatment should not negatively impact the accuracy of CMV quantification. | The NIH also demonstrated that using proteinase K for pretreatment of specimens did not impact the accuracy of CMV quantification in the Aptima Assay (as referenced by Youn et al. 2024). The Hologic risk assessment concluded that the benefits outweigh the risks and the "assay performance characteristics including accuracy are maintained with incorporation of this modification." |
| Feasibility and Safety of Protocol Implementation: The protocol should be practical for use in clinical laboratories and not introduce new safety concerns for the assay. | The protocol is described with clear steps (temperature, volume, timing). The risk assessment stated that "the benefits provided by the updates to the Hologic specimen handling workflow at customer sites outweigh the risks." |
2. Sample Size Used for the Test Set and the Data Provenance:
- Test Set Sample Sizes:
- Without Proteinase K: 1039 plasma specimens
- With Proteinase K: 4098 plasma specimens
- Data Provenance: The plasma specimens were collected from the transplant patient population by the National Institute of Health (NIH). This suggests a retrospective analysis of archived samples or a prospective study conducted at the NIH. The location is the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
- Not applicable as this is a quantitative diagnostic assay for DNA detection, not an image-based AI/ML device requiring expert interpretation of images for ground truth. The "ground truth" here is the actual CMV DNA concentration, which the assay is designed to measure.
4. Adjudication Method for the Test Set:
- Not applicable for a quantitative diagnostic assay. The performance is assessed against the ability to produce a valid quantitative result for CMV DNA and maintain accuracy, not against human consensus.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, this is not an MRMC study. MRMC studies are typically performed for visual diagnostic aids (e.g., medical imaging AI) to compare human reader performance with and without AI assistance. This device is a direct quantitative assay.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Yes, in a sense. The study validates the protocol (which includes a manual pretreatment step) and its effect on the automated assay system's ability to produce a valid result. The "performance" refers to the assay's output after the sample preparation, not an AI algorithm's independent decision. The core assay itself operates in a "standalone" fashion once the sample is loaded onto the Panther system.
7. The type of ground truth used:
- The implicit ground truth is the actual CMV DNA concentration in patient samples (or spiked controls used in accuracy assessments), as determined by a reference method or known concentrations. The study's focus, however, is on resolving an assay technical issue (ML2 flag) and demonstrating that the resolution method maintains the accuracy of CMV quantification. The reference to the NIH demonstrating that "using proteinase K for pretreatment of specimens did not impact the accuracy of CMV quantification" confirms that accurate quantitation (the ground truth outcome) was assessed.
8. The Sample Size for the Training Set:
- Not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The "training" for such an assay would be its initial development and optimization, which isn't described in terms of a specific "training set" size for this modification. The data provided refers to validation/test sets for the new protocol.
9. How the ground truth for the training set was established:
- Not applicable, as it's not an AI/ML device with a training set for model development.
Ask a specific question about this device
(28 days)
Aptima BV Assay: The Aptima BV Assay is an in vitro nucleic acid amplification test that utilizes real time Transcription-Mediated Amplification (TMA) technology for detection and quantitation of ribosomal RNA from bacterial vaginosis (BV), including Lactobacillus (L. gasseri, L. crispatus, and L. jenseni), Gardnerella vaginalis), and Atopobium vaginae (A. vaginae). The assay reports a qualitative result for BV and does not report results for individual organisms. The assay is intended to aid in the diagnosis of BV on the automated Panther System using clinician-collected and patient-collected vaginal swab specimens from females with a clinical presentation consistent with vaginitis and/or vaginosis.
Aptima CV/TV Assay: The Aptima CV/TV Assay is an in vitro nucleic acid amplification of RNA from microorganisms associated with vulvovaginal candidiasis and trichomoniasis. The assay utilizes real time Transcription-Mediated Amplification (TMA) technology to detect and qualitatively report results for the following organisms: Candida species group (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis), Candida glabrata (C. glabrata), Trichomonas vaginalis (TV). The assay differentiates between C. glabrata and the Candida species group (C spp) by targeting the RNA component of RNAse P ribonucleoprotein; the assay does not differentiate among C spp. For TV, the assay targets ribosomal RNA (rRNA) and differentiates the results for C. glabrata and C spp. The assy is intended to aid in the diagnosis of vulvovaginal candidiasis and trichomoniasis on the automated Panther System using clinician-collected and patientcollected vaginal swab specimens from females with a clinical presentation consistent with vagintis.
Aptima BV Assay: Like the Aptima BV assay 100 test kit, the Aptima BV assay 250 test kit is an in vitro nucleic acid amplification test for the detection and quantitation of rRNA from bacteria associated with bacterial vaginosis in women with a clinical presentation consistent with vaginitis/vaginosis. The Aptima BV assay utilizes the automated Panther system to provide qualitative results to aid in the diagnosis of bacterial vaginosis.
Aptima CV/TV Assay: Like the Aptima CV/TV assay 100 test kit, the Aptima CV/TV assay 250 test kit is an in vitro nucleic acid amplification test for the detection and quantitation of RNA from microorganisms associated with vaginitis, trichomoniasis, and vulvovaginitis, in women with a clinical presentation consistent with vaginitis, trichomoniasis, and vulvovaginitis. The Aptima CV/TV assay utilizes the automated Panther system to provide qualitative results to aid in the diagnosis of vulvovaginal candidiasis and trichomoniasis.
The provided document, a 510(k) summary for the Hologic Aptima BV Assay and Aptima CV/TV Assay, describes the device's technical specifications and a general statement regarding performance data for substantial equivalence. However, it does not contain the detailed acceptance criteria and a comprehensive study report with specific performance metrics (like sensitivity, specificity, PPV, NPV against a clinical gold standard) that would typically be expected for demonstrating a device meets acceptance criteria in a clinical validation context.
Specifically, the document focuses on demonstrating substantial equivalence of a new kit configuration (250 tests) to an already cleared kit configuration (100 tests) by showing similar analytical performance, particularly Limit of Detection (LoD). It does not present a de novo clinical study with pre-defined acceptance criteria for diagnostic accuracy against a true clinical ground truth.
Therefore, many of the requested items (e.g., number of experts, adjudication method, MRMC studies, training set details) are not applicable or not provided in this specific type of submission, which is for a modification to an already cleared device, not a novel device requiring full clinical validation from scratch.
Here's an attempt to answer the questions based only on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria and reported device performance in terms of diagnostic accuracy (e.g., sensitivity, specificity) against a clinical reference for the 250-test kit. Instead, it asserts equivalence by confirming L.O.D. for the new kit configuration.
The primary acceptance criteria for the new 250-test kit configuration, as implied by the performance data section, is that it must meet the established Limit of Detection (LoD) of the previously cleared 100-test kit configuration.
Aptima BV Assay (250 Test Kit):
| Acceptance Criteria (LoD of 100-Test Kit) | Reported Performance (LoD Confirmation for 250-Test Kit) |
|---|---|
| L. crispatus (LC): 143 CFU/mL | LC: 143 CFU/mL |
| L. gasseri (LG): 2,207 CFU/mL | LG: 2,207 CFU/mL |
| L. jenseni (LJ): 10 CFU/mL | LJ: 10 CFU/mL |
| A. vaginae (AV) C95: 5.10 log CFU/mL | AV C95: 5.10 log CFU/mL (128,397 CFU/mL) |
| G. vaginalis (GV) C95: 4.86 log CFU/mL | GV C95: 4.86 log CFU/mL (72,836 CFU/mL) |
Aptima CV/TV Assay (250 Test Kit):
| Acceptance Criteria (LoD of 100-Test Kit) | Reported Performance (LoD Confirmation for 250-Test Kit) |
|---|---|
| C. albicans C95 (LoD): 4439 CFU/mL | C. albicans C95 (LoD): 4439 CFU/mL |
| C. parapsilosis C95 (LoD): 9416 CFU/mL | C. parapsilosis C95 (LoD): 9416 CFU/mL |
| C. tropicalis C95 (LoD): 811 CFU/mL | C. tropicalis C95 (LoD): 811 CFU/mL |
| C. dubliniensis C95 (LoD): 1176 CFU/mL | C. dubliniensis C95 (LoD): 1176 CFU/mL |
| C. glabrata LoD: 41 CFU/mL | C. glabrata LoD: 41 CFU/mL |
| T. vaginalis LoD: 0.0024 cells/mL | T. vaginalis LoD: 0.0024 cells/mL |
The conclusion states: "The performance study results demonstrate that the Aptima BV assay 250 Test Kit on the Panther system performs comparably to the predicate device that is currently marketed for the same intended use." and "The Aptima BV 100 Test Kit LoD was confirmed in the Aptima BV 250 Test Kit configuration." (Similar statements are made for the CV/TV assay).
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: For the LoD confirmation study for both assays, at least 20 replicates per concentration, per reagent lot, using three lots were tested. This totals to at least 60 replicates per strain (for BV) or per organism (for CV/TV).
- Data Provenance: The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It describes analytical sensitivity studies using prepared dilutions of cell lysates/suspensions. For the CV/TV assay, the use of "Natural Vaginal Swab Matrix (NVSM)" and "Simulated Vaginal Swab Matrix (SVSM)" is mentioned for dilutions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided and is not applicable to the type of analytical study performed. The LoD confirmation uses known concentrations of target organisms, not clinical samples requiring expert interpretation for ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided and is not applicable to an analytical LoD confirmation study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not provided and is not applicable. The device is a diagnostic assay (in vitro nucleic acid amplification test), not an AI-assisted imaging device that would involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The device itself is a standalone diagnostic assay (an in vitro nucleic acid amplification test) run on an automated system (Panther system). Its performance (LoD) was confirmed without human interpretation of raw signals, as the Panther system software interprets the amplification signal emergence times to generate results.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth used for this specific study (LoD confirmation) was known concentrations of purified target organisms/cell lysates. This is an analytical ground truth, not a clinical ground truth derived from expert consensus, pathology, or outcomes data.
8. The sample size for the training set
This information is not provided and is not applicable. This is not an AI/ML device that requires a training set in the conventional sense. The "training" for the assay involves internal optimization and validation during development, but the document does not detail these earlier stages.
9. How the ground truth for the training set was established
This information is not provided and is not applicable, as it's not an AI/ML device with a distinct training set and ground truth establishment methodology in the context of this submission.
Ask a specific question about this device
(87 days)
The Panther Fusion® SARS-CoV-2/Flu A/B/RSV assay is a fully automated multiplexed real-time polymerase chain reaction (RT-PCR) in vitro diagnostic test intended for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus (Flu A), influenza B virus (Flu B), and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab specimens and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This assay is intended to aid in the differential diagnosis of SARS-CoV-2, Flu A, Flu B, and RSV infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
Positive results do not rule out coinfection with other organism(s) detected by the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. This assay is designed for use on the Panther Fusion system.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is classified as a Class II in vitro diagnostic device per 21 CFR 866.3981 and has product code QOF. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is designed for use on the fully automated Panther Fusion System.
The Panther Fusion system is a class II exempt device under 21CFR 862.2570 that has product code OOI.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a multiplex real-time reverse transcriptase PCR (RT-PCR) in vitro diagnostic test developed for use on the fully automated Panther Fusion system to detect and differentiate SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV) directly from nasopharyngeal and anterior nasal swab specimens, from individuals exhibiting signs and symptoms of a respiratory tract infection.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay involves the following steps: sample lysis, nucleic acid capture and elution transfer, and multiplex RT-PCR where analytes (when present) are simultaneously amplified, detected and differentiated. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion system. The eluate is transferred to the Panther Fusion system reaction tube containing the assay reagents. Multiplex RT-PCR is then performed for the eluted nucleic acid on the Panther Fusion system.
Sample lysis, nucleic acid capture, and elution: Prior to processing and testing on the Panther Fusion system, specimens are transferred to a Specimen Lysis Tube containing specimen transport media (STM). Alternatively, samples can be collected with the RespDirect Collection kit which contains enhanced specimen transport media (eSTM). STM and eSTM lyse the cells, release target nucleic acid and protect them from degradation during storage. The Internal Control-S (IC-S) is added to each test specimen and controls via the working Panther Fusion Capture Reagent-S (wFCR-S). The IC-S is the reagent used to monitor specimen processing, amplification and detection. Magnetic particles with covalently bound oligonucleotides mediate the nucleic acid capture. Capture oligonucleotides hybridize to nucleic acid in the test specimen. Hybridized nucleic acid is then separated from the lysed specimen in a magnetic field. Wash steps remove extraneous components from the reaction tube. The elution step elutes purified nucleic acid. During the nucleic acid capture and elution step, total nucleic acid is isolated from specimens.
Elution transfer and RT-PCR: During the elution transfer step, eluted nucleic acid is transferred to a Panther Fusion tube already containing oil and reconstituted mastermix. Target amplification occurs via RT-PCR. A reverse transcriptase generates a DNA copy of the target sequence. Target specific forward and reverse primers and probes then amplify targets while simultaneously detecting and discriminating multiple target types via multiplex RT-PCR. The Panther Fusion system compares the fluorescence signal to a predetermined cut-off to produce a qualitative result for the presence or absence of the analyte. The positive result for each analyte will be accompanied by the cycle threshold (Ct value).
Here's a breakdown of the acceptance criteria and the study details for the Hologic Panther Fusion SARS-CoV-2/Flu A/B/RSV assay, based on the provided document:
Acceptance Criteria and Device Performance
The document describes the performance of the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay in detecting SARS-CoV-2, Flu A, Flu B, and RSV in prospectively collected Anterior Nasal (AN) swab specimens in RespDirect eSTM.
Table of Acceptance Criteria (Implicit) and Reported Device Performance
While explicit acceptance criteria (e.g., minimum PPA/NPA values) are not stated, the document presents the observed Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with their 95% Confidence Intervals. These values are implicitly the performance targets for the assay.
| Target Virus | Prevalence (%) | PPA (% (95% CI)) | NPA (% (95% CI)) |
|---|---|---|---|
| Prospective Study (n=1021 evaluable AN swab specimens) | |||
| SARS-CoV-2 | 10.8 | 99.1 (95.0, 99.8) | 99.0 (98.1, 99.5) |
| Flu A | 1.1 | 100 (74.1-100) | 99.9 (99.4, 100) |
| Flu B | 0.6 | 83.3 (43.6, 97.0) | 99.8 (99.3, 99.9) |
| RSV | 0.1 | 100 (20.7, 100) | 99.9 (99.4, 100) |
| Enrichment Study (n=205 evaluable AN swab specimens) | |||
| Flu A | N/A | 97.2 (90.3-99.2) | 93.2 (87.6-96.4) |
| Flu B | N/A | 97.8 (88.4-99.6) | 99.4 (96.5-99.9) |
| RSV | N/A | 98.4 (91.3-99.7) | 95.8 (91.2-98.1) |
Study Information
2. Sample Size Used for the Test Set and Data Provenance
- Prospective Study:
- Sample Size: 1021 evaluable anterior nasal (AN) swab specimens.
- Data Provenance: Prospective, multicenter study conducted in the US between January 2023 and May 2023 at nine participating medical facilities.
- Enrichment Study (Supplemental Clinical Data for Low Prevalence Analytes):
- Sample Size: 205 evaluable anterior nasal (AN) swab specimens.
- Data Provenance: Enrichment phase of the study, conducted in the US between October 2023 and February 2024 at six participating medical facilities. The specimens were from individuals with a positive standard of care (SOC) test result for Flu A, Flu B, and/or RSV. All were frozen prior to testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts used to establish the ground truth. Instead, it refers to:
- For SARS-CoV-2: A composite comparator algorithm (CCA) consisting of up to three highly sensitive US FDA EUA SARS-CoV-2 molecular tests.
- For Flu A, Flu B, and RSV: A US FDA-cleared molecular Flu A/B/RSV assay (in the prospective study) and a U.S. FDA-cleared molecular Flu A/B/RSV assay (in the enrichment study).
- For the enrichment study, initial selection was based on a positive standard of care (SOC) test result for Flu A, Flu B, and/or RSV.
4. Adjudication Method for the Test Set
- For SARS-CoV-2 in the prospective study: A Composite Comparator Algorithm (CCA) was used. A final CCA result was assigned when two of the three composite comparator assays were in concordance.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This study is for an in vitro diagnostic (IVD) assay (a molecular test), not an AI-powered diagnostic imaging device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, this study represents a standalone performance evaluation of the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay. The assay itself is an automated molecular diagnostic test on the Panther Fusion system, and its performance was evaluated against comparator methods (other molecular tests or algorithms), not against human interpretation of its outputs in a clinical setting for diagnostic aid.
7. The Type of Ground Truth Used
- SARS-CoV-2: Composite comparator algorithm (CCA) based on multiple FDA EUA molecular tests.
- Flu A/B/RSV: US FDA-cleared molecular Flu A/B/RSV assay.
- Enrichment Study (Flu A/B/RSV): US FDA-cleared molecular Flu A/B/RSV assay, and initial selection based on positive Standard of Care (SOC) test results.
This represents a reference standard method ground truth, where another established diagnostic test is used as the benchmark.
8. The Sample Size for the Training Set
The document does not provide information on a specific "training set" sample size for algorithm development. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a real-time RT-PCR in vitro diagnostic test, not a machine learning or AI algorithm in the context of typical training/validation splits. Its development would involve analytical studies and design verification, but not typically a "training set" in the same sense as an AI model. The provided clinical data are for performance evaluation.
9. How the Ground Truth for the Training Set Was Established
As noted in point 8, the concept of a "training set" with established ground truth in the context of AI models is not directly applicable here. The assay's performance characteristics (e.g., limit of detection, analytical reactivity) are established through various analytical studies (bench studies), not through a "training set" with a clinical ground truth. The clinical studies described are for validation and performance assessment, not for training an algorithm.
Ask a specific question about this device
(115 days)
The Fluent Pro Fluid Management System is intended to provide liquid distension of the uterus during diagnostic and operative hysteroscopy, and to monitor the volume differential between the irrigation liquid flowing into and out of the uterus while providing drive, control and suction for hysteroscopic morcellators.
The Fluent Pro Fluid Management System is a hysteroscopic fluid management system and drive console for hysteroscopic morcellators. The Fluent Pro Fluid Management System consists of a console and single-use procedure kit. The single-use procedure kit consists of sterile inflow (In-FloPak) and outflow (Out-FloPak) tube sets, and a non-sterile waste bag. The FloPaks and waste bag connect for use in hysteroscopic procedures. The console includes motors that control fluid inflow and outflow for hysteroscopic insufflation, as well as a pressure sensor monitoring system and graphical user interface. In addition, the console includes a connection for MyoSure Tissue Removal Devices and a pneumatic foot pedal to perform tissue removal procedures. The Fluent Pro Fluid Management System is compatible with currently marketed MyoSure Tissue Removal Devices.
This FDA 510(k) summary describes a medical device, the Fluent Pro Fluid Management System, for which substantial equivalence is claimed against a predicate device. The document focuses on regulatory compliance and device characteristics rather than a clinical study evaluating diagnostic or treatment efficacy.
Therefore, the requested information regarding acceptance criteria, reported device performance, sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, multi-reader multi-case studies, and standalone algorithm performance cannot be fully extracted or accurately described from the provided text. These types of details are typically found in clinical study reports, which are not included here.
However, based on the provided text, I can infer and extract some information related to performance testing and acceptance criteria as described in the context of device safety and functional equivalence:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria here are functional and safety performance, demonstrated through various engineering and bench tests, rather than clinical efficacy metrics.
| Performance Metric | Acceptance Criteria (Implied/Directly Stated) | Reported Device Performance (Summary) |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993-1, 5, 10, 11, 23 (cytotoxicity, sensitization, systemic toxicity, irritation) | Patient contacting materials are biocompatible and comply with ISO 10993-1:2018 requirements. |
| Sterilization | Compliance with ISO 11135:2014 | Sterilization validation conducted in accordance with ISO 11135:2014. |
| Shelf-Life | Compliance with ASTM F1980 (36 months shelf life) | Passed shelf-life testing per ASTM F1980, achieving a 36-month shelf life. |
| Electrical Safety & EMC | Compliance with IEC 60601-1 and IEC 60601-1-2 | Underwent electrical safety and EMC testing in accordance with IEC 60601-1 and IEC 60601-1-2. |
| Intrauterine Pressure Control & Fluid Deficit Accuracy | Deficit Accuracy: $\pm$ 50 mL under normal use (specified for predicate and proposed device) | Passed Intrauterine Pressure Control and Fluid Deficit Bench Testing. (Implied: met $\pm$ 50 mL accuracy) |
| System Mechanical, Hardware, Electrical | Not explicitly stated but implied to meet design specifications | Passed System Mechanical, Hardware, and Electrical Bench Testing. |
| Software | Not explicitly stated but implied to meet design specifications | Passed Software Unit, Integration, and System Testing. |
| Disposables (Pull Force, Flow Rate, Duty Cycle) | Not explicitly stated but implied to meet design specifications | Passed Disposables Pull Force, Flow Rate, and Duty Cycle Bench Testing. |
| Usability | Compliance with IEC 62366-1:2015 + A1:2020 and FDA Guidance | Passed simulated use testing in accordance with IEC 62366-1 and FDA Guidance. |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for the various performance tests (e.g., number of units tested for electrical safety, number of sterility tests etc.). The provenance of the "data" is from internal company testing and validation activities as part of the 510(k) submission process. These are all prospective tests conducted on the device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable in the context of this 510(k) submission. The "ground truth" for these engineering and performance tests is defined by adherence to international standards and design specifications, not expert consensus on medical images or patient outcomes.
4. Adjudication method
Not applicable. Testing involves engineering verification and validation against objective standards and specifications, not subjective expert adjudication.
5. Multi-reader multi-case (MRMC) comparative effectiveness study
No, an MRMC comparative effectiveness study was not done or reported in this 510(k) summary. This type of study is relevant for diagnostic devices where human readers interpret medical data (e.g., images) with and without AI assistance to assess improved accuracy or efficiency. The Fluent Pro Fluid Management System is a hysteroscopic insufflator and morcellator control system; its functions are mechanical and fluidic control, not diagnostic interpretation.
6. Standalone (i.e. algorithm only without human-in-the loop performance) was done
This refers to an AI algorithm's performance without human interaction. This is not applicable to the Fluent Pro Fluid Management System, as it is a physical medical device with software components that control mechanical actions, not an AI-driven image analysis or diagnostic algorithm. Its "performance" is its functional operation in conjunction with a human operator during a hysteroscopic procedure.
7. The type of ground truth used
For the performance testing mentioned, the "ground truth" is defined by:
- International Standards: e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, IEC 60601-1/1-2 for electrical safety/EMC, IEC 62366-1 for usability.
- Device Design Specifications: For metrics like Intrauterine Pressure Control, Fluid Deficit Accuracy ($\pm$ 50 mL), Maximum Inflow Rate, etc. These are objective engineering parameters.
8. The sample size for the training set
Not applicable. The Fluent Pro Fluid Management System is not an AI/ML device that requires a "training set" in the conventional sense for learning and model development. The software within the device undergoes testing (unit, integration, system testing) to ensure it performs according to its predefined logical and control functions, not to train a learning algorithm.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set" for an AI/ML algorithm.
Ask a specific question about this device
(299 days)
The Aptima HCV Quant Dx assay is a real-time transcription-mediated amplification (TMA) test used for both detection and quantitation of hepatitis C virus (HCV) RNA in fresh and frozen human serum and plasma from HCV-infected individuals.
Plasma may be prepared in ethylenediaminetetraacetic acid (EDTA), anticoagulant citrate dextrose (ACD) solution, and plasma preparation tubes (PPT). Serum may be prepared in serum separator tubes (SST). Specimens are tested using the Panther system for automated specimen processing, and quantitation. Specimens containing HCV genotypes 1 to 6 are validated for detection and quantitation in the assay.
The Aptima HCV Quant Dx assay is indicated for use as an aid in the diagnosis of active HCV infection in the following populations: individuals with antibody evidence of HCV infection with evidence of liver disease of to be actively infected with HCV antibody evidence, and individuals at risk for HCV infection with antibodies to HCV. Detection of HCV RNA indicates that the virus is replicating and, therefore, is evidence of active infection of HCV RNA does not discriminate between acute and chronic states of infection.
The Aptima HCV Quant Dx assay is also indicated for use as an aid in the management of HCV infected patients undergoing HCV antiviral drug therapy. The assay can be used to measure HCV RNA levels periodically prior to, during, and after treatment to determine sustained virological response (SVR) or nonsustained virological response (NSVR), Assay performance characteristics have been established for individuals infected with HCV and treated with certain direct-acting antiviral agents (DAA) regimens. No information is available on the assay's predictive value when other therapies are used. The results from the Aptima HCV Quant Dx assay must be interpreted within the context of all relevant clinical and laboratory findings.
The Aptima HCV Quant Dx assay is not approved for use as a screening test for the presence of HCV RNA in blood or blood products.
Clearance of this pre-market application will 1) enable additional reagent on-board / off-board cycling, allowing operators to load reagents onto the Panther System 8 times instead of 5 times; and 2) update the Assay Definition Module (ADM) to detect, flag and invalidate results impacted by a faulty or flickering LED, using estimated background minimum limits. These changes do not introduce any changes to the original design, method of manufacture, assay procedure, principle of operation, mechanism of action, conditions of use or hardware of the Panther instrument, or to the results interpretation for the cleared assay.
The Aptima HCV Quant Dx assay is a nucleic acid amplification test that uses real-time TMA technology to detect and quantitate HCV RNA for aiding diagnosis or to establish baseline viral load, as well as to measure on-treatment and post-treatment responses. The assay targets a conserved region of the HCV genome, detecting and quantitating genotypes 1, 2, 3, 4, 5, and 6. The assay is standardized against the 2nd WHO International Standard for Hepatitis C Virus (NIBSC Code 96/798).
The Aptima HCV Quant Dx assay involves three main steps, which all take place in a single tube on the Panther system: target capture, target amplification by TMA, and detection of the amplification products (amplicon) by the fluorescent labeled probes (torches).
During target capture, viral RNA is isolated from specimens. The specimen is treated with a detergent to solubilize the viral envelope, denature proteins, and release viral genomic RNA. Capture oligonucleotides hybridize to highly conserved regions of HCV RNA, if present, in the test specimen. The hybridized target is then captured onto magnetic microparticles that are separated from the specimen in a magnetic field. Wash steps remove extraneous components from the reaction tube.
Target amplification occurs via TMA, which is a transcription-mediated nucleic acid amplification method that utilizes two enzymes, Moloney murine leukemia virus (MMLV) reverse transcriptase and T7 RNA polymerase. The reverse transcriptase is used to generate a DNA copy (containing a promoter sequence for T7 RNA polymerase) of the target sequence. T7 RNA polymerase produces multiple copies of RNA amplicon from the DNA copy template. The Aptima HCV Quant Dx assay utilizes the TMA method to amplify a portion of the 5' UTR of the HCV genome. Amplification of this region is achieved using specific primers which are designed to amplify HCV genotypes 1, 2, 3, 4, 5, and 6.
Detection is achieved using single-stranded nucleic acid torches that are present during the amplification of the target and that hybridize specifically to the amplicon in real time. Each torch has a fluorophore and a quencher. When the torch is not hybridized to the amplicon, the quencher is in close proximity of the fluorophore and suppresses the fluorescence. When the torch binds to the amplicon, the quencher is moved farther away from the fluorophore and it will emit a signal at a specific wavelength when excited by a light source. As more torches hybridize to amplicon a higher fluorescent signal is generated. The time taken for the fluorescent signal to reach a specified threshold is proportional to the starting HCV concentration. Each reaction has an internal calibrator/internal control (IC) that controls for variations in specimen processing, amplification, and detection. The concentration of a sample is determined by the Panther system software using the HCV and IC signals for each reaction and comparing them to calibration information.
The Aptima HCV Quant Dx assay has been designed for and validated on the Panther system. The Panther system is an integrated hardware and software system that together with the Aptima HCV Quant Dx assay fully automates all the steps necessary to perform the assay from sample preparation through amplification of nucleic acid, detection, data reduction and amplicon inactivation.
The provided document describes the acceptance criteria and a study for the Aptima HCV Quant Dx Assay.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria | Reported Device Performance | Met? |
|---|---|---|---|
| Intended Use Testing (Software 7.2.9 vs 7.2.7) | For software 7.2.9 vs 7.2.7 (reported in PMA annual report for P160023/R004): Assay performance is equivalent to prior comparators. | Yes | |
| QC Control A (Negative) | Not Detected. Max 1 false positive. | 1 false positive. | Yes |
| QC Control B (2.16 log IU/mL) | ±0.5 logs from Value Assignment (or Target Conc if Value). Max 1/30 outliers. Std Dev ≤ 0.25 log copy. | 100% Positivity, Ave diff between expected and recovered is -0.05 log IU/mL, largest SD is 0.07 log IU/mL. | Yes |
| QC Control D (6.52 log IU/mL) | ±0.5 logs from Value Assignment (or Target Conc if Value). Max 0/25 outliers. Std Dev ≤ 0.25 log copy. | 100% Positivity, Ave diff between expected and recovered is -0.12 log IU/mL, largest SD is 0.03 log IU/mL. | Yes |
| QC Panel S (1.13 log IU/mL) | 95% Positivity | 100% Positivity. | Yes |
| Assay Verification (5 to 8 reagent reload change) | |||
| HCV 3.3 log IU/mL (5 Cycles) | Not explicitly stated for each concentration, but implicitly refers to maintaining performance within acceptable ranges. | Cycle Avg LogIU/mL: 3.22, Cycle Std Dev: 0.05, Baseline Avg LogIU/mL: 3.20, Diff (Cycle-Baseline): 0.02. | Yes |
| HCV 3.3 log IU/mL (10 Cycles) | Not explicitly stated for each concentration, but implicitly refers to maintaining performance within acceptable ranges. | Cycle Avg LogIU/mL: 3.28, Cycle Std Dev: 0.06, Baseline Avg LogIU/mL: 3.10, Diff (Cycle-Baseline): 0.18. | Yes |
| HCV 7.0 log IU/mL (5 Cycles) | Not explicitly stated for each concentration, but implicitly refers to maintaining performance within acceptable ranges. | Cycle Avg LogIU/mL: 7.45, Cycle Std Dev: 0.15, Baseline Avg LogIU/mL: 7.56, Diff (Cycle-Baseline): -0.11. | Yes |
| HCV 7.0 log IU/mL (10 Cycles) | Not explicitly stated for each concentration, but implicitly refers to maintaining performance within acceptable ranges. | Cycle Avg LogIU/mL: 7.30, Cycle Std Dev: 0.07, Baseline Avg LogIU/mL: 7.56, Diff (Cycle-Baseline): -0.26. | Yes |
| Assay Verification (Flickering LED) | New estimated background limits will increase fault detection for these assays. | One test order flagged by FAM estimated background limit was also flagged by QNS and considered a true fault. This analysis supports the implementation of proposed Estimated Background minimum limits. | Yes |
| Software V&V (PRM, PRD, SRS, PSD) | All requirements pass. | PRM: 23 Pass, 0 Fail; PRD: 29 Pass, 0 Fail; SRS: 210 Pass, 0 Fail; PSD: 4 Pass, 0 Fail. | Yes |
| Software V&V (Regression Testing) | All requirements pass. | 12 Pass, 0 Fail. | Yes |
| Software V&V (Anomalies) | No known software anomalies exceeding "Negligible" severity. | 12 anomalies, all rated "Negligible." | Yes |
| Additional Intended Use Testing (HCV Assay Software 5.2.5.1 vs 5.3.5.1) | All acceptance criteria met for low level clinical panel. | Aptima HCV Assay Software versions 5.2.5.1 and 5.3.5.1 both resulted in 100% not detected for the negative clinical matrix panel. Both resulted in 100% positivity for the 30 IU/mL clinical positive panel. | Yes |
| Additional On-Board Stability (OBS) and Reagent Cycling | All acceptance criteria met (for 5, 8, and 10 cycles with 72 hours on board stability). | Aptima reagents successfully completed five, eight, and ten on and off on-board cycles with a cumulative 72 hours on board stability and met all acceptance criteria. | Yes |
(*) Note: The document states "may be removed from calculation of Standard Deviatio logs from Target Concentration if Value Assignm" and "±0.50 logs from Target Concentration if Value", which appear to be incomplete sentences or truncated notes in the original document.
2. Sample Sizes Used for the Test Set and Data Provenance
- Intended Use Testing (Software 7.2.9 vs 7.2.7):
- QC Control A (Negative): 120 samples.
- QC Control B (2.16 log IU/mL): 40 samples.
- QC Control D (6.52 log IU/mL): 40 samples.
- QC Panel S (1.13 log IU/mL): 80 samples.
- Data Provenance: Not explicitly stated beyond "previous software version, 7.2.7 (reported to the FDA in the 2022 PMA annual report for the Aptima HCV Quant Dx Assay, P160023/R004)".
- Assay Verification (5 to 8 reagent reload change):
- HCV 3.3 log IU/mL: Not explicitly stated, but "5-10 cycles of reagent storage" and performance evaluated.
- HCV 7.0 log IU/mL: Not explicitly stated, but "5-10 cycles of reagent storage" and performance evaluated.
- Data Provenance: "prepared virus and armored RNA (aRNA) spiked into negative human serum matrix (BI0052)".
- Assay Verification (Flickering LED process control change):
- 381 field instruments and 445,910 unique test orders were analyzed from March 2018 to March 2022.
- Data Provenance: "field instruments" indicates real-world data, likely retrospective. Country of origin not specified.
- Additional Intended Use Testing (HCV Assay Software 5.2.5.1 vs 5.3.5.1):
- Negative clinical specimen: 1 (tested with 10 replicates per run).
- HCV RNA concentration around 3xLLoQ (30 IU/ml): 3 specimens (tested with 10 replicates of each per run).
- Total replicates per software version: (1 negative specimen * 10 replicates) + (3 positive specimens * 10 replicates) = 40 replicates. This was done for both software versions.
- Data Provenance: "clinical specimens", "negative clinical matrix panel", "low positive clinical positive panel". Samples prepared by diluting high positive sample in negative clinical matrix.
- Additional On-Board Stability (OBS) and Reagent Cycling:
- Minimum of one negative sample, a sample around 3xLLoQ (30 IU/ml), and a moderate positive or high positive sample.
- Minimum of 5 replicates per sample for each timepoint, with more replicates at baseline.
- Data Provenance: Not explicitly stated, but implies laboratory-controlled testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This device is a diagnostic assay for HCV RNA detection and quantitation, not an imaging device that requires interpretation by human experts. The ground truth would be based on the quantitative/qualitative results of the assay itself and reference standards/controls.
4. Adjudication Method for the Test Set
Not applicable. As noted above, this refers to a diagnostic assay. The "results" are numerical or qualitative (detected/not detected) from an automated system compared against established performance characteristics and controls.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This information is for an in-vitro diagnostic (IVD) assay, not an AI-powered medical imaging or decision support system that typically involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies described are essentially standalone performance evaluations of the assay system (Aptima HCV Quant Dx Assay on the Panther system software). The device operates automatically from sample preparation through amplification, detection, and quantitation. The performance checks compare the output of the system against expected values, referent methods, or established criteria.
7. The Type of Ground Truth Used
The ground truth for the performance evaluations appears to be based on:
- Reference standards and controls: For quantitative and qualitative measurements (e.g., QC Control A, B, D, S, prepared virus and armored RNA, 3xLLoQ and negative samples).
- Previous established performance: Comparison of new software/hardware changes against the predicate device's established performance (PMA P160023).
- Root-cause analysis: For the flickering LED study, specific root-cause analysis was used to identify true faults.
8. The Sample Size for the Training Set
No training set information is provided as this is a diagnostic assay product, not a machine learning algorithm that requires a distinct training phase. The "development" or "calibration" of the assay would typically involve internal studies not detailed here, but these are not referred to as "training sets" in the context of conventional IVD development.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set is described.
Ask a specific question about this device
(76 days)
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a fully automated multiplexed real-time polymerase chain reaction (RT-PCR) in vitro diagnostic test intended for the qualitative detection and differentiation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), influenza A virus (Flu A), influenza B virus (Flu B), and respiratory syncytial virus (RSV). Nucleic acids are isolated and purified from nasopharyngeal (NP) swab specimens and anterior nasal (AN) swab specimens obtained from individuals exhibiting signs and symptoms of a respiratory tract infection. Clinical signs and symptoms of respiratory viral infection due to SARS-CoV-2, influenza, and RSV can be similar. This assay is intended to aid in the differential diagnosis of SARS-CoV-2, Flu A, Flu B, and RSV infections in humans and is not intended to detect influenza C virus infections.
Nucleic acids from the viral organisms identified by this test are generally detectable in NP and AN swab specimens during the acute phase of infection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory tract infective of the presence of the identified virus and aids in diagnosis if used in conjunction with other clinical and epidemiological information, and laboratory findings. The results of this test should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results do not rule out coinfection with other organism(s) detected by the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay may not be the definite cause of disease. Negative results do not preclude SARS-CoV-2, influenza A virus, influenza B virus, or RSV infections. This assay is designed for use on the Panther Fusion system.
The Hologic RespDirect Collection Kit is cleared for NP swab specimens only for testing with the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is classified as a Class II in vitro diagnostic device per 21 CFR 866.3981 and has product code QOF. The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is designed for use on the fully automated Panther Fusion System.
The Panther Fusion system is a class II exempt device under 21CFR 862.2570 that has product code OOI.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay is a multiplex real-time reverse transcriptase PCR (RT-PCR) in vitro diagnostic test developed for use on the fully automated Panther Fusion system to detect and differentiate SARS-CoV-2, influenza A, influenza B, and respiratory syncytial virus (RSV) directly from nasopharyngeal and anterior nasal swab specimens, from individuals exhibiting signs and symptoms of a respiratory tract infection.
The Panther Fusion SARS-CoV-2/Flu A/B/RSV assay involves the following steps: sample lysis, nucleic acid capture and elution transfer, and multiplex RT-PCR where analytes (when present) are simultaneously amplified, detected and differentiated. Nucleic acid capture and elution takes place in a single tube on the Panther Fusion system. The eluate is transferred to the Panther Fusion system reaction tube containing the assay reagents. Multiplex RT-PCR is then performed for the eluted nucleic acid on the Panther Fusion system.
Here's a breakdown of the acceptance criteria and the study details for the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Target Virus | Acceptance Criteria (Implicit for New Specimen Type) | Reported Performance (Prospective AN Swab) PPA % (95% CI) [n/N] | Reported Performance (Prospective AN Swab) NPA % (95% CI) [n/N] | Reported Performance (Retrospective AN Swab) PPA % (95% CI) [n/N] | Reported Performance (Retrospective AN Swab) NPA % (95% CI) [n/N] |
|---|---|---|---|---|---|
| SARS-CoV-2 | Comparable to predicate device; high PPA/NPA expected for molecular tests. | 94.8 (90.1, 97.3) [146/154] | 98.8 (98.0, 99.3) [1023/1035] | N/A (Not reported separately) | N/A (Not reported separately) |
| Flu A | Comparable to predicate device; high PPA/NPA expected for molecular tests. | 91.8 (80.8, 96.8) [45/49] | 99.6 (99.0, 99.8) [1135/1140] | 97.9 (89.1, 99.6) [47/48] | 98.4 (94.4, 99.6) [123/125] |
| Flu B | Comparable to predicate device; high PPA/NPA expected for molecular tests. | 66.7 (20.8, 93.9) [2/3] | 99.7 (99.3, 99.9) [1183/1186] | 97.2 (90.3, 99.2) [69/71] | 100 (96.4, 100) [102/102] |
| RSV | Comparable to predicate device; high PPA/NPA expected for molecular tests. | 94.4 (74.2, 99.0) [17/18] | 99.8 (99.4, 100) [1169/1171] | 98.0 (89.5, 99.6) [49/50] | 99.2 (95.5, 99.9) [122/123] |
Note on Acceptance Criteria: The document explicitly states the goal is to demonstrate "comparable" performance to the predicate device (which uses an NP swab) for the new anterior nasal (AN) swab specimen type. While specific numerical acceptance criteria (e.g., "PPA must be >X%") are not provided in this summary, the reported performance metrics (PPA, NPA, and their confidence intervals) are presented to demonstrate this comparability and overall effectiveness for diagnostic use, as typical for molecular assays expanding their indications. The high NPA values for all targets indicate good specificity, and generally high PPA values indicate good sensitivity, especially when considering the confidence intervals. The lower PPA for Flu B in the prospective study was addressed and supported by the retrospective study, which showed strong performance.
2. Sample Sizes Used for the Test Set and Data Provenance
-
Prospective Clinical Study (Anterior Nasal Swab):
- Sample Size: 1,268 individuals were enrolled. 1,230 AN swab specimens were tested, resulting in 1,189 evaluable specimens for analysis (not all evaluable for all analytes).
- Data Provenance: Prospective, multicenter study conducted in the US. Specimens were collected from individuals attending nine participating medical facilities in the US during the 2022-2023 respiratory infection season.
-
Retrospective Clinical Study (Anterior Nasal Swab - Supplement for Flu B and RSV):
- Sample Size: 175 preselected retrospective specimens.
- Data Provenance: Retrospective. The origin country is not explicitly stated but is implied to be related to the US context from the prospective study.
-
Analytical Studies (Reprocessing from K222736):
- Sample Size: Not explicitly stated as "sample size" for each analytical study detail, but refers to the number of replicates/conditions tested for LOD, inclusivity, exclusivity, etc. For example, Flu B LoD confirmation involved 28 replicates per panel.
- Data Provenance: Already existing data from the previous 510(k) submission (K222736) using Nasopharyngeal (NP) swabs, reprocessed with the updated software.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not mention the use of experts to establish ground truth for the clinical test sets. Instead, the ground truth was established by comparing the candidate device's results to reference molecular assays:
- SARS-CoV-2: A composite comparator algorithm (CCA) consisting of up to three highly sensitive US FDA EUA SARS-CoV-2 molecular tests. A final CCA result was assigned when two of the three comparator assays were in concordance.
- Flu A, Flu B, and RSV: A US FDA-cleared molecular Flu A/B/RSV assay.
4. Adjudication Method for the Test Set
- SARS-CoV-2 Ground Truth: For SARS-CoV-2, a composite comparator algorithm (CCA) was used. A final CCA result was assigned when "two of the three composite comparator assays were in concordance." This effectively acts as an adjudication method where agreement among a majority of reference tests determines the ground truth.
- Flu A, Flu B, RSV Ground Truth: For Flu A, Flu B, and RSV, a single "US FDA-cleared molecular Flu A/B/RSV assay" was used as the comparator method. There's no mention of an adjudication process among multiple comparators for these targets.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or conducted. This type of study is more common for imaging devices or diagnostics that involve subjective interpretation by human readers. This device is an automated molecular diagnostic assay, where the output is objective (positive/negative, Ct value).
6. Standalone (Algorithm Only) Performance Study
- Yes, the performance data presented (PPA, NPA) directly reflects the standalone performance of the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay. It is compared to a ground truth established by other molecular tests, without human interpretation of the assay's output influencing the performance metrics. The assay itself is automated, so its performance is inherently "algorithm only" in the context of generating results.
7. Type of Ground Truth Used
- Molecular Comparator Assays: The ground truth for the clinical test sets (both prospective and retrospective) was established using other legally marketed and highly sensitive molecular diagnostic tests.
- For SARS-CoV-2: A composite of up to three US FDA EUA SARS-CoV-2 molecular tests.
- For Flu A, Flu B, RSV: A US FDA-cleared molecular Flu A/B/RSV assay.
8. Sample Size for the Training Set
- The document does not explicitly state the sample size used for the training set. This submission focuses on the validation of an existing assay with a new specimen type (anterior nasal swab) and a software update (Adaptive Crosstalk Correction). The original development and training data for the core assay (K222736) would have been described in that prior submission, but are not detailed here for this specific submission's scope. The mention of "reprocessed with validated Results Processor Tool" for previously submitted analytical studies also suggests reuse of data rather than new training.
9. How the Ground Truth for the Training Set Was Established
- Since the document doesn't explicitly detail a new "training set" for this specific submission, it also doesn't describe how ground truth for such a set was established. It's likely that the original assay development followed similar methods of using well-characterized samples or comparator assays for internal development and optimization.
- Regarding the Adaptive Crosstalk Correction (ACC) factor, the validation involved "testing of SARS-CoV-2 positive clinical specimens that are representative of those that yielded false positive Flu B results in the field." This indicates that the problem the ACC addresses arose from real-world performance, and the resolution was validated against specific clinical scenarios rather than a formal "training set" in the machine learning sense.
Ask a specific question about this device
Page 1 of 11