Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K212629
    Date Cleared
    2022-05-26

    (280 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PP CARE Nitrile Examination Gloves is a disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

    Device Description

    PP CARE Nitrile Examination Gloves are Class I patient examination gloves bearing the product code LZA (21CFR880.6250). The gloves are made from acrylonitrile butadiene rubber. These gloves are blue in color and are powder free.

    AI/ML Overview

    The document provided focuses on the substantial equivalence determination for PP CARE Nitrile Examination Gloves (K212629) based on non-clinical testing. It does not describe a clinical study or AI/algorithm performance. Therefore, many of the requested points are not applicable.

    Here's the information that can be extracted from the provided text regarding the non-clinical testing:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance CriteriaReported Device Performance (PP CARE Nitrile Examination Gloves)Standard / Test Method
    DimensionsASTM D 6319-19
    Length (Size L)Min 230 mm240 mmASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Width (Size L)110 +/- 10 mm105 mmASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    ThicknessPalm: 0.05 mm min; Finger: 0.05 mm minPalm: 0.10 mm; Finger: 0.15 mmASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Physical PropertiesASTM D 6319-19
    Tensile Strength - Before Aging (Size L)14 MPa Minimal17 MPaASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Tensile Strength - After Aging (Size L)14 MPa Minimal20 MPaASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Ultimate Elongation - Before Aging (Size L)500% Min520%ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Ultimate Elongation - After Aging (Size L)400% Min512%ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Powder Residue2 mg/glove Max0.35 mg/gloveASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application
    Water TightnessNumber of samples tested: 200; Number of Leaks allowed: 7 (Implies AQL 1.5 pass rate)Passed (AQL 1.5)The Testing of unreferenced nitrile gloves in accordance with EN 455-1:2000
    BiocompatibilityUnder the condition of study not an irritant/sensitizer/non-cytotoxic (for predicate)
    Skin IrritationNot an irritantUnder the condition of study not an irritant[As per ISO 10993-10:2010(E)] - SKIN IRRITATION TEST OF NITRILE EXAMINATION GLOVE, NON STERILE IN NEW ZEALAND WHITE RABBITS
    Skin SensitizationNot a sensitizerUnder the conditions of the study, not a sensitizer[As per ISO 10993-10:2010(e)] - SKIN SENSITIZATION STUDY OF NITRILE EXAMINATION GLOVE, NON STERILE IN GUINEA PIGS BY MAXIMIZATION TEST
    In Vitro CytotoxicityNon-cytotoxicThe sample was 100% cytotoxic (but acceptable due to systemic test)[As per ISO 10993-5:2009(E)] - IN VITRO CYTOTOXICITY STUDY OF NITRILE EXAMINATION GLOVE, NON STERILE BY ELUTION METHOD
    Acute Systemic ToxicityN/A (Predicate did not perform this test)Under the conditions of the study, the device was non-toxic[As per ISO 10993-11:2017] - ACUTE SYSTEMIC TOXICITY STUDY OF NITRILE EXAMINATION GLOVE, NON STERILE IN SWISS ALBINO MICE

    2. Sample size used for the test set and the data provenance:

    • Water Tightness Test: 200 samples were tested.
    • Data Provenance: The tests described are non-clinical (laboratory/materials testing) and were performed by EG Group Product and Service Co., Ltd. in Thailand. The data is thus prospective for the purpose of this submission and originates from Thailand.
    • For other tests like Dimensions, Physical Properties, Powder Residue, the sample sizes are not explicitly stated but are implied to be sufficient for demonstrating compliance with the respective ASTM standards.
    • For biocompatibility tests (Skin Irritation, Sensitization, Cytotoxicity, Systemic Toxicity), specific animal models are mentioned (New Zealand White Rabbits, Guinea Pigs, Swiss Albino Mice), indicating laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This document describes non-clinical performance and biocompatibility testing of medical gloves, not a study requiring expert clinical judgment for ground truth. The "ground truth" here is compliance with established international standards (ASTM, ISO, EN).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. This is not a study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not a study involving human readers or AI.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This document is about medical gloves, not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the device's performance against its acceptance criteria is defined by compliance with established international standards (ASTM D6319-19, EN 455-1:2000) and ISO 10993 series for biocompatibility. These standards set the benchmarks for physical properties, barrier integrity, and biological safety.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning study.

    9. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/machine learning study.

    Summary of the Study that Proves Device Meets Acceptance Criteria:

    The device's compliance with acceptance criteria is demonstrated through a series of non-clinical tests performed in accordance with recognized international standards. These tests assessed:

    • Dimensional properties: Length, width, and thickness as per ASTM D6319-19.
    • Physical properties: Tensile strength and ultimate elongation (before and after aging) as per ASTM D6319-19.
    • Powder Residue: Measurement of powder content as per ASTM D6319-19.
    • Water Tightness: To detect holes, conducted in accordance with EN 455-1:2000.
    • Biocompatibility:
      • Skin Irritation (ISO 10993-10:2010(E) using New Zealand White Rabbits)
      • Skin Sensitization (ISO 10993-10:2010(e) using Guinea Pigs)
      • In Vitro Cytotoxicity (ISO 10993-5:2009(E) by elution method)
      • Acute Systemic Toxicity (ISO 10993-11:2017 using Swiss Albino Mice)

    The results show that the PP CARE Nitrile Examination Gloves meet or exceed all specified acceptance criteria derived from these standards for physical and barrier properties, and demonstrate acceptable biocompatibility (despite an "100% cytotoxic" in vitro result, which was mitigated by a successful acute systemic toxicity test, deeming the device non-toxic under study conditions). These non-clinical tests collectively provide the evidence that the device meets its performance requirements and is substantially equivalent to the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K973276
    Manufacturer
    Date Cleared
    1997-11-20

    (79 days)

    Product Code
    Regulation Number
    862.1690
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This kit is intended for the quantitative determination of thyrotropin in blood specimens dried on filter paper as an aid in screening newborns for congenital (neonatal) hypothyroidism, using the 1235 AutoDELFIA™ automatic immunoassay system.

    Device Description

    The AutoDELFIA™ Neonatal hTSH L assay is a solid phase, two-site fluoroimmunometric assay based on the direct sandwich technique in which two monoclonal antibodies are directed against separate antigenic determinants on the hTSH molecule. Standards, controls and test specimen containing hTSH are reacted simultaneously with immobilised monoclonal antibodies in assay buffer. The assay buffer elutes hTSH from the dried blood spots on the filter paper discs. The complete assay requires only one incubation step. Enhancement solution dissociates europium ions from the labelled antibody into solution where they form highly fluorescent chelates with components of the enhancement solution. The fluorescence in each well is then measured. The fluorescence of each sample is proportional to the concentration of hTSH in the sample.

    AI/ML Overview

    The provided text describes the 510(k) summary for the AutoDELFIA™ Neonatal hTSH L Kit, which is intended for screening newborns for congenital hypothyroidism. The study that proves the device meets acceptance criteria is primarily a comparison to a predicate device.

    Here's an analysis based on the information provided:

    Acceptance Criteria and Device Performance

    The document describes a comparison study against a predicate device, the AutoDELFIA™ Neonatal hTSH kit, to demonstrate substantial equivalence. The key performance metrics are the reference ranges and mean values of hTSH concentrations.

    Acceptance CriteriaReported Device Performance (AutoDELFIA™ Neonatal hTSH L kit)Predicate Device (AutoDELFIA™ Neonatal hTSH kit)
    Reference Range0.69 - 12.7 µU/mL blood0.55 - 14.9 µU/mL blood
    Mean Value4.13 µU/mL blood3.96 µU/mL blood

    The qualitative acceptance criteria is "substantially equivalent" to the predicate device. The performance data presented (reference ranges and mean values) are used to support this claim of substantial equivalence. The frequency distributions are also presented to visually demonstrate the similarity.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 860 newborns.
      • Data Provenance: Not explicitly stated, but given the context of U.S. regulatory submission by a U.S.-based company, it's highly likely the data is from the U.S. The study is retrospective, as the hTSH concentrations were "measured in 860 newborns" using both kits.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This study does not involve expert interpretation to establish ground truth in the typical sense of diagnostic imaging or clinical assessment. It's a quantitative assay where the "ground truth" for the comparison is the measurement obtained by the predicate device and the new device. Therefore, no experts were explicitly used to establish a ground truth for the test set as would be the case in subjective diagnostic tasks. The ground truth for hTSH levels is assumed to be the quantitative measurement itself.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • There was no adjudication method described. This is a quantitative assay comparison, not a subjective diagnostic task requiring expert consensus.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This is an in vitro diagnostic device (IVD) for laboratory analysis, not a device that involves human readers interpreting outputs or AI assistance.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, this study represents a standalone performance evaluation of the AutoDELFIA™ Neonatal hTSH L kit in comparison to a predicate device. Both devices are automated immunoassays, meaning they function independently to produce a quantitative result without direct human interpretation in each measurement step.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" used for this comparative study is the quantitative hTSH concentration measured by the predicate device (AutoDELFIA™ Neonatal hTSH kit). The new device's performance is gauged against this established method to demonstrate substantial equivalence.
    7. The sample size for the training set:

      • The document does not provide information about a "training set." This type of in vitro diagnostic device study typically focuses on validation and comparison to a predicate, rather than an AI/ML development pipeline with distinct training and test sets.
    8. How the ground truth for the training set was established:

      • As there's no mention of a training set, the establishment of ground truth for it is not applicable based on the provided text.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1