Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242419
    Device Name
    018 Hydrophilic Coated OTW PTA Balloon Dilatation Catheter
    Manufacturer
    Creagh Medical Ltd. dba Surmodics, Inc.
    Date Cleared
    2024-11-15

    (92 days)

    Product Code
    LIT
    Regulation Number
    870.1250
    Why did this record match?
    Applicant Name (Manufacturer) :

    Creagh Medical Ltd. dba Surmodics, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) of the peripheral vasculature in the iliac, femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
    Device Description
    The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is a coaxial catheter with a semi compliant balloon near the distal tip. It is an Over-the-Wire (OTW) Percutaneous Transluminal Anqioplasty (PTA) device with various shaft lengths. The balloon has two radiopaque markers that aid in the placement of the balloon within the stenosis. The clearance between the inner and outer shafts acts as the passage for the inflation medium for balloon expansion. The proximal end of the catheter has a bifurcated manifold & strain relief that allows for the 0.018" guidewire and the attachment of a balloon inflation device via a standard luer connector. The inflation device is used to inflate and deflate the balloon with a contrast medium. The device is used by positioning the balloon catheter over a guidewire. The balloon is aligned under fluoroscopy in the diseased vessel at the area to be treated. The balloon is then inflation media to pressures ranging between the nominal and the rated burst pressure to dilate the occluded area. On completion the balloon is then deflated under vacuum and removed from the patient. The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is to be provided sterile (via ethylene oxide, EtQ) and is intended for single use only.
    Ask a Question

    Page 1 of 1