K Number
K242419

Validate with FDA (Live)

Date Cleared
2024-11-15

(92 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) of the peripheral vasculature in the iliac, femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Description

The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is a coaxial catheter with a semi compliant balloon near the distal tip. It is an Over-the-Wire (OTW) Percutaneous Transluminal Anqioplasty (PTA) device with various shaft lengths. The balloon has two radiopaque markers that aid in the placement of the balloon within the stenosis. The clearance between the inner and outer shafts acts as the passage for the inflation medium for balloon expansion. The proximal end of the catheter has a bifurcated manifold & strain relief that allows for the 0.018" guidewire and the attachment of a balloon inflation device via a standard luer connector. The inflation device is used to inflate and deflate the balloon with a contrast medium. The device is used by positioning the balloon catheter over a guidewire. The balloon is aligned under fluoroscopy in the diseased vessel at the area to be treated. The balloon is then inflation media to pressures ranging between the nominal and the rated burst pressure to dilate the occluded area. On completion the balloon is then deflated under vacuum and removed from the patient. The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is to be provided sterile (via ethylene oxide, EtQ) and is intended for single use only.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter and its associated summary for a medical device: the "Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter." This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving the safety and effectiveness of a novel device through clinical trials and rigorous statistical validation against predefined acceptance criteria.

Therefore, the information regarding acceptance criteria, study design for proving device performance, sample sizes for test sets, establishment of ground truth, expert adjudication, MRMC studies, standalone performance, training set details, and ground truth for training data is not present in the provided text.

The document primarily outlines:

  • Acceptance Criteria (Implicit via Substantial Equivalence): The implicit acceptance criterion for this 510(k) submission is that the device must be shown to be "substantially equivalent" to a legally marketed predicate device. This means it has the same intended use, similar technological characteristics, and any differences do not raise new questions of safety or effectiveness.
  • Reported Device Performance: Instead of a detailed table of acceptance criteria and performance against them, the document lists bench tests performed to demonstrate that the small changes (expanded matrix of balloon sizes) compared to the predicate device do not compromise safety or effectiveness.

Here's a breakdown of what is available and what is missing from your request based on the provided text:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as pass/fail thresholds for specific performance metrics in a table. The overall acceptance criterion is "substantial equivalence" to the predicate.

  • Reported Device Performance: The document lists bench tests performed, indicating that the results "demonstrate that the technological differences identified do not raise new questions of safety or effectiveness." Specific quantitative results or acceptance criteria for these bench tests are not provided.

    Performed Bench Tests (as listed):

    • Rated burst pressure (RBP)
    • Inflation & deflation time
    • Balloon diameters at nominal pressure to RBP
    • Simulated use - Pushability & Trackability & Sheath Compatibility
    • Coating integrity (b)
    • Multiple inflation/fatigue & leak test
    • Tensile strength Balloon to Shaft (proximal bond)
    • Particulate
    • Balloon Length & Marker Band Position (Leveraged from predicate)
    • Radiopacity (Leveraged from predicate)
    • Ancillary Tool Compatibility (Guidewire) (Leveraged from predicate)
    • Catheter Effective Length (Leveraged from predicate)
    • Tensile Strength Manifold to Shaft (Manifold Bond) (Leveraged from predicate)
    • Tensile Strength (Distal / Tip Bond) (Leveraged from predicate)
    • Tip Profile (Geometry of the catheter most distal tip) (Leveraged from predicate)
    • Simulated Use, Push & Track (Leveraged from predicate)
    • Flexibility & Kink (Leveraged from predicate)
    • Coating Lubricity (Leveraged from predicate)
    • Coating Integrity (a) (Leveraged from predicate)
    • Torque Strength (Leveraged from predicate)

2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Sizes: Not specified. This document refers to bench testing of a medical device, which typically involves a number of units tested to engineering specifications, but these numbers are not disclosed.
  • Data Provenance: The tests are bench tests, not clinical data, so country of origin/retrospective/prospective doesn't apply in the same way. The manufacturer is based in Ireland (Creagh Medical Ltd. dba Surmodics, Inc.).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • This is not part of a bench test evaluation for a substantial equivalence determination. Experts are not typically used to establish "ground truth" for physical device performance metrics like burst pressure.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable to bench testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was done. This device is a physical catheter, not an AI/software device that assists human readers.
  • The document explicitly states: "No clinical studies were required for the Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable as it's not an AI/software device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not applicable. Performance is measured against engineering specifications and industry standards for physical device properties, not against "ground truth" derived from clinical data or expert consensus.

8. The sample size for the training set

  • Not applicable; there is no training set for a physical medical device. (This is relevant for AI models).

9. How the ground truth for the training set was established

  • Not applicable; there is no training set.

In summary, the provided FDA document is a 510(k) clearance letter for a physical medical device, based on demonstrated "substantial equivalence" through bench testing, not clinical studies or AI model validation. Therefore, most of the detailed information requested regarding clinical study design, data characteristics, expert involvement, and AI model evaluation (points 2-9) is not found in this type of submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2024

Creagh Medical Ltd. dba Surmodics, Inc. Anne-Marie Keenan Senior Regulatory Affairs Specialist IDA Business Park, Ballinasloe Co. Galway Ballinasloe, H53 K8P4 Ireland

Re: K242419

Trade/Device Name: Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: September 19, 2024 Received: September 19, 2024

Dear Anne-Marie Keenan:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ariel G.Ash-shakoor -SDigitally signed byAriel G. Ash-shakoor-SDate: 2024.11.1519:09:54 -05'00'
---------------------------------------------------------------------------------------------------------------------------

For

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242419

Device Name

Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter

Indications for Use (Describe)

The Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter for Percutaneous Transluminal Angioplasty (PTA) of the peripheral vasculature in the iliac, femoral, popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "SURMODICS" in all capital letters. To the left of the word is a blue and green graphic. The graphic is a stylized shape that looks like a ribbon or folded piece of paper. The word "SURMODICS" is in a sans-serif font and is black.

510k Summary

Date Prepared: November 14, 2024

Submitters Name/Contact Person

510K Submitter and Contact for Routine Correspondence

Anne-Marie Keenan Senior Regulatory Affairs Specialist Creagh Medical Ltd. (Surmodics Inc.) Ballinasloe Co. Galwav H53 K8P4 IRELAND

Email: akeenan@surmodics.com

510k Submitter Establishment Registration Number

3005994106

General Information
Trade Name:Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter
Common / Usual Name:PTA Balloon Dilatation Catheter
Classification NameCatheter, Angioplasty, Peripheral, Transluminal
Regulation/Product Code21 CFR 870.1250
Device PanelCardiovascular
Regulatory Classification:Class II
Product Code:LIT
Predicate Device:018 Hydrophilic Coated OTW PTA Balloon Dilatation Catheter (K180007)
Reference DeviceSterling™ Over-the-Wire (OTW) PTA Balloon Dilatation Catheter (K132430)

1.1 Device Description

The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is a coaxial catheter with a semi compliant balloon near the distal tip. It is an Over-the-Wire (OTW) Percutaneous Transluminal Anqioplasty (PTA) device with various shaft lengths. The balloon has two radiopaque markers that aid in the placement of the balloon within the stenosis. The clearance between the inner and outer shafts acts as the passage for the inflation medium for balloon expansion. The proximal end of the catheter has a bifurcated manifold & strain relief that allows for the 0.018" guidewire and the attachment of a balloon inflation device via a standard luer connector. The inflation device is used to inflate and deflate the balloon with a contrast medium. The device is used by positioning the balloon catheter over a guidewire. The balloon is aligned under fluoroscopy in the diseased vessel at the area to be treated. The balloon is then inflation media to pressures ranging between the nominal and the rated burst pressure to dilate the occluded area. On completion the balloon is then deflated under vacuum and removed from the patient. The Advance Serenity™ 18 Hydrophilic PTA Balloon Dilatation Catheter is to be provided sterile (via ethylene oxide, EtQ) and is intended for single use only.

Surmodics, Inc. 9924 West 74th Street Eden Prairie, MN 55344 USA Phone 952-500-7001 www.surmodics.com Creagh Medical Ltd. & Vetex Medical Ltd are now part of Surmodics, Inc. IDA Business Park Ballinasloe, Co. Galway Ireland Phone +353 90 9646300

{5}------------------------------------------------

1.2 Indications for Use

The Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) of the peripheral vasculature in the iliac, femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

1.3 Comparison of Technological Characteristics

The subject Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter is substantially equivalent to the predicate 018 Hydrophilic Coated OTW PTA Balloon Dilatation Catheter (K180007) in design, intended use, principles of use, biocompatibility, sterility, and labeling. Changes to the predicate device that have led to the submission of this new 510(k), are the expansion of the matrix of dilatation balloons available for peripheral balloon angioplasty. All characteristics that were not identical to the predicate device were verified through performance bench testing and determined to be substantially equivalent.

CharacteristicSubject Device: K242419Advance Serenity 18Hydrophilic PTA BalloonDilatation CatheterPredicate Device: K180007018 Hydrophilic Coated OTWPTA Balloon DilatationCatheterComparison
Indications for UseThe Advance Serenity 18Hydrophilic PTA BalloonDilatation Catheter is indicatedfor Percutaneous TransluminalAngioplasty (PTA) of theperipheral vasculature in theiliac, femoral, popliteal, infra-popliteal, and renal arteries,and for the treatment ofobstructive lesions of native orsynthetic arteriovenous dialysisfistulae.The 018 Hydrophilic CoatedOTW PTA Balloon DilatationCatheter is indicated forPercutaneous TransluminalAngioplasty (PTA) of theperipheral vasculature in theiliac, femoral, popliteal, infra-popliteal, and renal arteries,and for the treatment ofobstructive lesions of native orsynthetic arteriovenous dialysisfistulae.Identical
Principles ofOperationInflation of semi-compliant balloon for dilatationIdentical
GuidewireCompatibility018"Identical
Catheter ShaftLength (cm)40, 90, 120, 13540, 90, 120, 135, 150Identical(40, 90, 120, 135cm)
SheathCompatibility5Fr4Fr, 5FrIdentical(5Fr)
Catheter ShaftOuter Diameter(mm)5Fr: ≤1.6674Fr: ≤1.3335Fr: ≤1.667Identical(5Fr)
Balloon Length(mm)22015, 20, 40, 60, 80, 100, 120,150, 220Identical(220mm)
Balloon Diameters(mm)5.0, 6.0 & 7.02.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0,5.5, 6.0, 6.5, 7.0, 8.0, 9.0, 10.0Identical(5.0, 6.0 & 7.0mm)
Nominal Pressure8 ATM8 ATMIdentical
Rated BurstPressure14 & 16 ATM10, 12, 14, 16 ATMIdentical(14 & 16 ATM)
Balloon MaterialSemi-CompliantIdentical
Catheter ShaftMaterialsPolymers, all materials are biocompatibleSubstantiallyEquivalent

{6}------------------------------------------------

CharacteristicSubject Device: K242419Advance Serenity 18Hydrophilic PTA BalloonDilatation CatheterPredicate Device: K180007018 Hydrophilic Coated OTWPTA Balloon DilatationCatheterComparison
Radiopaque MarkerBandsTwo radiopaque marker bandsIdentical
Coating MaterialHydrophilic CoatingIdentical
PackagingConfigurationSame function and designIdentical
Sterilization MethodEthylene OxideIdentical
BiocompatibilityExternally communicating device in contact with circulating bloodfor limited exposure duration.SubstantiallyEquivalent
Single Use/ReprocessSingle use onlyIdentical

1.4 Performance Bench Testing

Results of design verification testing demonstrate that the technological differences identified do not raise new questions of safety or effectiveness compared to the predicate device. The device has been evaluated through the following tests:

  • Rated burst pressure (RBP) -
  • Inflation & deflation time -
  • -Balloon diameters at nominal pressure to RBP
  • -Simulated use - Pushability & Trackability & Sheath Compatibility
  • -Coating integrity (b)
  • Multiple inflation/fatique & leak test -
  • Tensile strength Balloon to Shaft (proximal bond) -
  • -Particulate

Performance Bench Tests have been leveraged from predicate device for the following:

  • Balloon Length & Marker Band Position -
  • -Radiopacity
  • -Ancillary Tool Compatibility (Guidewire)
  • Catheter Effective Length -
  • Tensile Strength Manifold to Shaft (Manifold Bond) -
  • -Tensile Strength (Distal / Tip Bond)
  • Tip Profile (Geometry of the catheter most distal tip) -
  • -Simulated Use, Push & Track
  • -Flexibility & Kink
  • Coating Lubricity -
  • -Coating Integrity (a)
  • -Torque Strength

1.5 Clinical Studies and Testing

No clinical studies were required for the Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter.

1.6 Conclusion

Based on the device description, materials, technological characteristics, and accompanying performance data, it can be concluded that the device modifications made to the Advance Serenity 18 Hydrophilic PTA Balloon Dilatation Catheter are substantially equivalent to the predicate device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).