Search Filters

Search Results

Found 88 results

510(k) Data Aggregation

    K Number
    K241085
    Manufacturer
    Date Cleared
    2024-07-12

    (84 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien llc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VLFT10GEN: The Valleylab FT10 energy platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7mm thick tissue (tissue bundles), and lymphatics. When used with compatible ablation devices it is indicated for cardiac tissue ablation. The generator can also be used with compatible resectoscopes for endoscopically controlled removal or coagulation of tissue using 0.9% NaCl solution as the irrigation medium. The tissue fusion function has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures.

    Valleylab ™FT10 Electrosurgical Generator (VLFT10FXGEN): The VLFT10FXGEN is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.

    Valleylab ™FT10 Vessel Sealing Generator (VLFT10LSGEN): The VLFT10LSGEN is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics.

    Device Description

    Valleylab™ FT10 Energy Platform (VLFT10GEN): The Valleylab™ FT10 Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue. The generator provides radio frequency (RF) energy for monopolar and bipolar surgical applications, tissue-fusion, and vessel-sealing applications (LigaSure/vessel sealing function) and cardiac applications. It is a combination of a full-featured general surgery electrosurgical unit and a bipolar vessel sealing system. The monopolar and bipolar sections, including the LigaSure/Bipolar section of the system, are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery.

    Valleylab™ FT10 Electrosurgical Generator (VLFT10FXGEN): The Valleylab™ FT10 Electrosurgical Generator is a high frequency electrosurgical only version of the VLFT10GEN Energy Platform. It is intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.

    Valleylab™ FT10 Vessel Sealing Generator (VLFT10LSGEN): The Valleylab™ FT10 Vessel Sealing Generator is the vessel sealing only version of the VLFT10GEN Energy Platform. It is a high frequency electrosurgical generator. When used with compatible sealing devices, it is indicated for sealing vessels up to and including 7 mm, tissue bundles, and lymphatics.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device, the Valleylab™ FT10 Energy Platform Software Version 5.0. It describes the device, its intended use, a comparison to a predicate device, and performance data. However, this document does not contain details about a study evaluating a device based on AI/ML. Instead, it describes an electrosurgical generator and highlights performance data related to electrical safety, EMC, software verification/validation, and mechanical testing.

    Therefore, I cannot extract information pertaining to:

    1. A table of acceptance criteria and the reported device performance regarding AI/ML performance metrics (e.g., sensitivity, specificity, AUC). The document refers to "acceptance criteria" for engineering and safety tests, not for AI performance.
    2. Sample sizes used for the test set and the data provenance for AI/ML evaluation.
    3. Number of experts used to establish the ground truth for an AI/ML test set and their qualifications.
    4. Adjudication method for an AI/ML test set.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done for AI assistance.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done for an AI/ML algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for AI/ML.
    8. The sample size for the training set for an AI/ML algorithm.
    9. How the ground truth for the training set was established for an AI/ML algorithm.

    The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence." This means that no clinical studies (which would typically involve human readers for an MRMC study or large datasets for standalone AI performance) were conducted or included in this 510(k) submission. The "Software verification and validation testing" mentioned is general software quality assurance, not specific to AI/ML model performance.

    In summary, the provided text does not contain the information requested about AI/ML device performance and testing.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Valleylab™ SM Smoke Management Pencil and accessories are designed for general electrosurqical applications, including cutting and coagulation, and for removing surgical smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.

    Device Description

    The Valleylab™ SM Smoke Management Pencils are monopolar electrosurgical smoke evacuation pencils intended for cutting and coagulation of tissue while simultaneously removing surgical smoke. The pencils are designed to capture surgical smoke and improve visibility to target tissues, while reducing staff and patient exposure to the hazards of surgical plume. The Valleylab™ SM Smoke Management Extended Nozzle accessories are for use with longer electrodes for deeper access procedures. The predicate device is the Valleylab™ Smoke Evacuation Rocker Switch Pencil cleared under K182772.

    The proposed devices are compatible with Covidien electrosurgical generators at a maximum peak voltage 4500 Vpk having a 3-prong connector and smoke evacuators having a 3/8" port. Electrode compatibility includes use with standard 3/32" diameter hex and non-hex electrodes. The pencils are available in 10' and 15' tubing/cable lengths models.

    AI/ML Overview

    This document pertains to the 510(k) premarket notification for the Valleylab™ SM Smoke Management Pencil and accessories. It's important to note that this device is a physical electrosurgical tool and does not contain an AI component. Therefore, much of the requested information regarding acceptance criteria and studies for AI-driven devices (like sample sizes for test sets, ground truth establishment for AI, MRMC studies, standalone algorithm performance, and training set details) is not applicable here.

    However, I can extract and present the acceptance criteria and performance testing conducted for this device, which are relevant to its substantial equivalence determination.


    Acceptance Criteria and Reported Device Performance

    The substantial equivalence determination for the Valleylab™ SM Smoke Management Pencil and accessories is based on meeting various performance standards and demonstrating safety and efficacy comparable to its predicate device (Valleylab™ Smoke Evacuation Rocker Switch Pencil cleared under K182772).

    Here's a summary of the performance testing and their acceptance criteria, as reported in the 510(k) summary:

    CategoryAcceptance Criteria (Met Standards/Requirements)Reported Device Performance
    Sterilization & Shelf-LifeSterilization by ethylene oxide (EO) to sterility assurance level 10-6 in accordance with ISO 11135, ISO 11737-1, ISO 11737-2, and ISO 10993-7. Product packaging designed and validated per ISO 11607-1, ISO 11607-2, ASTM D4169. Shelf-life determination based on accelerated aging methods per ASTM F1980.The Valleylab™ SM Smoke Management Pencil and accessories met the acceptance criteria for sterilization by ethylene oxide (EO) to sterility assurance level 10-6 in accordance with the applicable validation standards. Product packaging was designed and validated in accordance with packaging standards. The Valleylab™ SM Smoke Management Pencils are labeled for a 4-year shelf-life, and the Valleylab™ SM Smoke Management Extended Nozzles are labeled for a 5-year shelf-life in accordance with accelerated aging methods.
    BiocompatibilityRequirements of biocompatibility standard ISO 10993-1 for the following endpoints: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.The Valleylab™ SM Smoke Management Pencil and accessories met the requirements of biocompatibility standard ISO 10993-1 for the specified endpoints.
    SoftwareN/A (The device is not a software-driven device).The Valleylab™ SM Smoke Management Pencil does not contain software.
    Electromagnetic Compatibility & Electrical SafetyApplicable clauses of electromagnetic compatibility standard IEC 60601-1-2 and electrical safety standards IEC 60601-1, and IEC 60601-2-2.The Valleylab™ SM Smoke Management Pencil met the applicable clauses of electromagnetic compatibility standard IEC 60601-1-2 and electrical safety standards IEC 60601-1, and IEC 60601-2-2.
    Performance Testing - BenchDevice requirements for mechanical, functional, reliability, and ex vivo thermal tissue effect testing. (Specific quantitative acceptance criteria are not detailed in this summary document but are implied to be met through internal validation.)The Valleylab™ SM Smoke Management Pencil and accessories met the device requirements for mechanical, functional, reliability, and ex vivo thermal tissue effect testing.
    Usability/Human FactorsDemonstration of safe and effective use for the intended users, intended use, and use environments according to methods in IEC 60601-1-6 and IEC 62366-1.The Valleylab™ SM Smoke Management Pencil and accessories demonstrated safe and effective use for the intended users, intended use, and use environments according to methods in IEC 60601-1-6 and IEC 62366-1.

    Study Details Regarding Acceptance Criteria (as applicable to a non-AI medical device)

    Given that this is a 510(k) submission for an electrosurgical pencil and accessories, and not an AI-driven device, the detailed questions regarding AI study methodology are largely not applicable. However, I will address what is known from the provided text for each point:

    1. A table of acceptance criteria and the reported device performance: See the table above.
    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      • The document does not specify exact sample sizes for each test (e.g., number of units tested for sterilization, biocompatibility, electrical safety, or bench performance).
      • The data provenance is not explicitly stated in terms of country of origin but is from Covidien LLC, a Medtronic company. The tests would be prospective in nature, as they involve testing the actual device prototypes or production samples against established standards.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable for this type of device testing. Ground truth (e.g., sterility, electrical safety, mechanical integrity) is established by adherence to recognized international and national standards, not expert consensus in the diagnostic sense.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Device performance against technical standards is typically assessed through objective measurements, not human adjudication.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For this device, "ground truth" refers to the objective, measurable parameters defined by recognized international and national standards (e.g., ISO for sterilization and biocompatibility, IEC for electrical safety and EMC, ASTM for packaging and shelf-life).
    8. The sample size for the training set: Not applicable, as this is not an AI device.
    9. How the ground truth for the training set was established: Not applicable, as this is not an AI device.

    In conclusion, the Valleylab™ SM Smoke Management Pencil and accessories have undergone rigorous non-clinical performance evaluations against established medical device standards to demonstrate substantial equivalence to a predicate device. The nature of this product as a physical electrosurgical tool means that many of the AI-specific questions are not relevant.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shiley™ Oral/Nasal Endotracheal Tube Intermediate Cuff, Non-DEHP is intended for oral or nasal intubation of the trachea.
    The Shiley™ Oral/Nasal Endotracheal Tube Intermediate Cuff, Non DEHP is intended for use in oral or nasal intubation of the trachea for anesthesia and is indicated for airway management.

    Device Description

    The subject device is an oral/nasal endotracheal tube intermediate cuff. The translucent tube incorporates a Magill curve and features a radiopaque line. The tube features a thin wall, polyvinyl chloride (PVC) high pressure, low volume cuff with two different cuff shapes. An inflation system consisting of an inflation line, pilot balloon, and inflation valve allows inflation and deflation of the cuff. The subject device is manufactured from materials without latex or DEHP.

    AI/ML Overview

    This document is a 510(k) Summary for the Shiley™ Oral/Nasal Endotracheal Tube Intermediate Cuff, Non-DEHP, and it details the device's substantial equivalence to predicate devices, supported by performance data.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    Acceptance Criteria (Standard)Test PerformedReported Device Performance (Met Acceptance Criteria?)
    ISO 5361:2016 (FDA 1-118)Connector Pull TestMet
    ISO 5361:2016 (FDA 1-118)Cuff PerformanceMet
    ISO 5361:2016 (FDA 1-118)Inflation Line Pull TestMet
    ISO 5361:2016 (FDA 1-118)Fluid Seal TestMet
    ISO 5361:2016 (FDA 1-118)Kink ResistanceMet
    ISO 5361:2016 (FDA 1-118)Print AdherenceMet
    ISO 5361:2016Dimensions (e.g., bevel angle, radius of curvature, cuff diameter, connectors)Met
    ISO 5356ConnectorsMet (implies compliance with applicable parts of the standard)
    ISO 18190(Not explicitly detailed, but stated as being complied with)Met (implied)
    ISO 10993-1:2018 (FDA 2-258)CytotoxicityMet
    ISO 10993-1:2018 (FDA 2-258)ImplantationMet
    ISO 10993-1:2018 (FDA 2-258)SensitizationMet
    ISO 10993-1:2018 (FDA 2-258)Irritation/Intracutaneous reactivityMet
    ISO 10993-1:2018 (FDA 2-258)Material Mediated PyrogenicityMet
    EN 62366-1:2015 + A1:2020Human Factors/Usability StudyConformed
    (Not specified standard)Shelf Life (5 years)Met
    (Ethylene Oxide Sterilization Standard)Sterilization ValidationValidated

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the exact sample sizes for each bench test conducted. It states that "terminally sterilized samples" were tested. For biocompatibility, it mentions "testing was performed on final finished subject device" for sizes 8.5mm to 10.0mm, and testing from the predicate device was leveraged for sizes 3.0mm to 8.0mm.
    • Data Provenance: The document does not provide details on data provenance such as country of origin or whether the studies were retrospective or prospective. It describes laboratory bench testing and biocompatibility testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This section is not applicable as the document describes bench testing and biocompatibility assessments, which do not typically involve human experts establishing a "ground truth" in the way clinical studies or diagnostic AI studies do. The "ground truth" for these tests is defined by the objective pass/fail criteria of the specified international standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As stated above, this pertains to objective bench and biocompatibility testing following established international standards, not clinical adjudication by experts.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. A MRMC comparative effectiveness study was not performed. This document describes the clearance of a physical medical device (an endotracheal tube), not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an endotracheal tube; it does not involve an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance data in this submission consists of the acceptance criteria defined by international consensus standards (ISO 5361, ISO 5356, ISO 18190, ISO 10993-1) and specific test protocols for mechanical properties, biocompatibility, and sterilization, along with human factors standards (EN 62366-1).

    8. The sample size for the training set

    Not applicable. This document describes the clearance of a physical medical device (an endotracheal tube), not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230693
    Manufacturer
    Date Cleared
    2024-02-26

    (350 days)

    Product Code
    Regulation Number
    882.1400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BISTM Advance Monitoring System is intended for monitoring the state of the brain by data acquisition of EEG signals under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BISTM Advance Monitoring System, and all its associated parameters, is intended for use on adults and pediatric patients (4 years old and above) within a hospital or medical facility.

    For Adult patients, the BIS™ Index, one of the BIS™ Advance Monitoring System's output parameters, may be used to guide anesthetic administration of desflurane, propofol and sevoflurane with balanced anesthetic techniques in order to monitor the anesthetic effects on the brain.

    The use of the BIS™ Index for monitoring may be associated with the following when used with propofol anesthesia: reduction in primary anesthetic use; reduction in emergence and recovery time; and reduction in incidence of awareness with recall.

    For pediatric patients, ages 4 and above, the BIS™ Index, one of the BIS™ Advance Monitoring System's output parameters, may be used to guide anesthetic administration of sevoffurane with balanced anesthetic techniques in order to monitor the anesthetic effects on the brain.

    The use of the BIS™ Index in pediatric patients, when used with sevothurane anesthesia, has demonstrated a reduction in primary anesthetic use.

    Device Description

    The BIS™ Advance Monitoring System is a user-configurable patient monitoring system designed to monitor the hypnotic state of the brain based on acquisition and processing of EEG signals. It processes raw EEG signals to produce a single number, called the Bispectral Index, or BIS value, which correlates with the patient's level of hypnosis.

    The BIS™ Advance Monitoring system is comprised of the following components: BIS™ Advance Monitor, BIS™ Advance Docking Station, BIS™ Advance Adapter Cable, GCX Mounting Accessory, BISx/BISx4 Module, Patient Interface Cable (PIC) and Monitor Interface Cable (MIC).

    The BIS™ Advance Monitor displays:

    • The current BIS™ number .
    • . Raw EEG waveforms in real time
    • . Various signal quality indicators (EMG. SQI)
    • 트 Trend graphs of processed EEG parameters (including various options)
    • I Processed EEG variables:
      • . Electromyography (EMG)
      • Signal Quality Index (SQI)
      • . Suppression Ratio (SR)
      • . Burst Count (BURST) (for Extend Sensor and four-channel monitoring only)
      • . Suppression Time (ST)
      • I Spectral Edge Frequency (SEF)
      • Median Frequency (MF)
      • EEG Power Asymmetry Index (ASYM) (for four-channel monitoring only)
    • I Alarm Indicator and Messages

    The BIS™ Advance Monitor displays 2 channels of EEG when connected to the BISx module and a unilateral BIS sensor (BIS™ Extend Sensor, BIS™ Pediatric Sensor and BIS™ Quatro Sensor) and displays 4 channels of EEG, two from each side of the brain, when connected to the BISx4 module and BIS™ Bilateral Sensor.

    For both the 2-channel and the 4-channel systems, BIS monitoring is implemented as follows:

    A sensor placed on the patient's head transmits EEG signals to the BISx module. The BISx module filters the data, analyzes it for artifacts and processes it using digital signal processing techniques, then sends the data to the monitor for display. The purpose of processing the EEG waveform data is to extract characteristic features from the complex signal in order to provide easier pattern recognition of changes over time during the recording.

    AI/ML Overview

    The acceptance criteria for the BIS™ Advance Monitoring System are primarily related to its proposed changes: a narrowed indication for use and a new monitor design. The submission aims to demonstrate substantial equivalence to the predicate device (BIS EEG Vista Monitor System and BISX, K072286).

    Here's an analysis of the acceptance criteria and the studies presented:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in a pass/fail format. Instead, it relies on demonstrating that the new device's performance is equivalent or better than the predicate, especially for the new monitor design, and that the narrowed indications for use are clinically supported.

    However, based on the comparative effectiveness study for the Indications for Use, we can infer some performance aspects:

    Acceptance Criteria (Inferred from Indications for Use)Reported Device Performance (from Clinical Studies)
    For Adult patients: BIS™ Index guides anesthetic administration (desflurane, propofol, sevoflurane with balanced anesthetic techniques) to monitor anesthetic effects on the brain.OLIVER Study: BIS™ values were highly correlated with the level of sedation. Prediction probabilities for correctly predicting unresponsiveness to a verbal command were very high, suggesting BIS™ is an excellent predictor for the level of consciousness for sevoflurane, sevoflurane with remifentanil, sevoflurane with fentanyl, desflurane, or isoflurane. (This supports the broader claim for adults using these agents).
    For Adult patients with propofol anesthesia: Reduction in primary anesthetic use, reduction in emergence and recovery time, reduction in incidence of awareness with recall.OLIVER Study: While it demonstrated high correlation for BIS™ with level of sedation for various anesthetics, the document does not explicitly state the direct measured reduction in primary anesthetic use, emergence/recovery time, or awareness with recall specifically for propofol in the context of Oliver study. It states that "The use of the BIS™ Index for monitoring may be associated with the following when used with propofol anesthesia." This is a cautious statement.
    For Pediatric patients (4 years and above) with sevoflurane anesthesia: BIS™ Index guides anesthetic administration to monitor the anesthetic effects on the brain.BTIGER Study: Provided clinical evidence that the BIS™ index can be used to guide the anesthetic administration of sevoflurane in pediatric patients aged 4-18 years.
    For Pediatric patients with sevoflurane anesthesia: Demonstrated reduction in primary anesthetic use.BTIGER Study: Mean end-tidal sevoflurane (ETSevo) administration was statistically significantly lower in the BIS™ guided treatment group compared to the standard practice group.
    General Device Performance (New Monitor Design): Equivalence in fundamental technology, principle of operation, safety, and performance with predicate monitor, as well as compliance with relevant electrical safety, EMC, and software standards.Various Bench Tests: Electrical safety and EMC testing compliant with standards (ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 80601-2-26). All performance bench testing complied with requirements. Software complies with all software requirements and design specifications. Hazard analysis concluded residual risks were acceptable.

    2. Sample Size for the Test Set and Data Provenance

    • OLIVER Study (Adults):
      • Sample Size: 143 subjects
      • Data Provenance: Multicenter, prospective study conducted across 3 sites in the United States.
    • BTIGER Study (Pediatrics):
      • Sample Size: 170 subjects
      • Data Provenance: Multicenter, prospective, randomized control study conducted across 8 sites in the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document does not detail the number or qualifications of experts used to establish ground truth for the clinical studies. However, for studies involving "level of consciousness" and "unresponsiveness to verbal command" in the OLIVER study, and "anesthetic administration" in the BTIGER study, licensed healthcare practitioners would have been involved in assessing these clinical endpoints. The studies were conducted under the "direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use," as stated in the Indications for Use.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method (e.g., 2+1, 3+1) for establishing ground truth in the clinical studies.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? The document describes two clinical studies (OLIVER and BTIGER) that compare BIS™ guidance to standard practice or assess the correlation of BIS™ with clinical endpoints. While these are comparative studies, they are not framed as "multi-reader multi-case" studies in the typical sense of AI-assisted image interpretation. The comparison is between a device-guided approach and non-device-guided approaches or correlation with physiological states, rather than human readers interpreting cases with and without AI assistance.
    • Effect size of human reader improvement with AI vs. without AI assistance: Not applicable in the traditional MRMC context, as the studies are not designed to measure improved human reader performance with AI assistance for interpretation. Instead, the BTIGER study showed that the BIS™ guided treatment group achieved statistically significantly lower end-tidal sevoflurane administration compared to the standard practice group, indicating a direct effect on patient management rather than an improvement in human interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The core of the BIS™ Advance Monitoring System is mentioned as processing raw EEG signals to produce a single number, the Bispectral Index (BIS value), which correlates with the patient's level of hypnosis. The device also displays raw EEG waveforms, signal quality indicators, and processed EEG parameters. The clinical studies (OLIVER and BTIGER) inherently evaluate the performance of this algorithm output (BIS Index) in a clinical setting, effectively testing its "standalone" diagnostic utility in influencing anesthetic administration and correlating with consciousness levels. The output (BIS index) as a predictor of consciousness or a guide for anesthetic use is evaluated.

    7. Type of Ground Truth Used

    • OLIVER Study: The ground truth for correlating BIS™ with the level of sedation appears to be based on clinical assessment of patient responsiveness, specifically "response to a verbal command."
    • BTIGER Study: The ground truth for evaluating the impact of BIS™ guidance was the clinical outcome of "mean end-tidal sevoflurane (ETSevo) administration" during the maintenance phase of anesthesia, and the clinical decision-making by practitioners.

    8. Sample Size for the Training Set

    The document focuses on the performance of the current BIS™ algorithm and its new monitor. It states that "no software changes were performed to the BISx/BISx4 module (the unit that performs the computation for EEG acquisition), the BIS algorithm nor to the algorithm database structure." This suggests the core BIS algorithm was developed and trained prior to this submission. The document does not provide information on the sample size for the training set of the BIS algorithm itself. It only provides information for the clinical validation studies (test sets).

    9. How the Ground Truth for the Training Set Was Established

    As with the training set sample size, the document does not provide information on how the ground truth was established for the training set of the BIS algorithm, as the algorithm itself was not modified in this submission. The core algorithm's development and training would have occurred during the development of earlier versions of the BIS monitoring system (e.g., the predicate device K072286).

    Ask a Question

    Ask a specific question about this device

    K Number
    K231861
    Manufacturer
    Date Cleared
    2023-07-21

    (28 days)

    Product Code
    Regulation Number
    876.1725
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endoflip™ 300 System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters in adults and to measure pressure and dimensions in the esoplagus, in patients from five years of age. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.

    Device Description

    The subject Endoflip™ 300 System is equivalent to the predicate Endoflip™ 300 System (K223705) except for an update to the Endoflip™ 300 System software to remove the analysis episode feature of the predicate device. Other software updates include minor feature enhancements and bug fixes. Changes to device labeling rrom the software updates and other minor changes for clarification purposes were also addressed.

    The Endoflip™ 300 System software is supplied pre-installed in the Endoflip™ 300 Display System. It is also the software used for the Endoflip™ 300 Reader in reader mode only. There is no change to the display system hardware or to how the reader is supplied (USB stick).

    No changes were made to any of the other components that comprise the Endoflip™ 300 System when compared to the predicate.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Endoflip™ 300 System, which involves software updates to an existing device. This entire document is a 510(k) submission meant to demonstrate substantial equivalence to a predicate device, not a study specifically designed to establish acceptance criteria for a new device.

    Therefore, many of the requested details regarding a standalone study, multi-reader multi-case study, and detailed ground truth methodologies for a specific study proving acceptance criteria are not present in this type of regulatory document.

    However, based on the information provided, here's what can be extracted:

    Acceptance Criteria and Device Performance:

    The document's primary objective is to demonstrate substantial equivalence to the predicate device (Endoflip™ 300 System K223705) after software updates. Therefore, the "acceptance criteria" are implicitly that the updated device performs equivalently to the predicate device and does not introduce new safety or effectiveness concerns.

    Specific numerical acceptance criteria and reported device performance from a clinical trial or performance study are not explicitly provided in terms of metrics like sensitivity, specificity, or error rates. Instead, the performance is evaluated through software verification.

    Table of Acceptance Criteria and Reported Device Performance:

    Since explicit quantitative acceptance criteria for a new device's performance are not given (as this is a 510(k) for an updated software version of an existing device), we can infer the acceptance criteria for the software updates.

    Acceptance Criteria (Inferred for Software Updates)Reported Device Performance (as per submission)
    Device operates as intended with software updates.Software verification testing confirmed proper operation.
    No new safety concerns introduced by software changes.Software changes do not raise any new questions of safety and effectiveness; supported by verification activities.
    No adverse impact on existing functionality.Software updates include minor feature enhancements and bug fixes; changes addressed for clarification in labeling. Functionality maintained.
    Removal of "analysis episode feature" functions as intended without negative impact.The analysis episode feature was successfully removed from the software.
    Software complies with relevant standards and guidance documents.Software verification performed per IEC 62304 and FDA guidance for software in medical devices (May 2005) and 510(k) evaluations (July 28, 2014).
    Software maintains "moderate" level of concern and associated mitigation.Software was considered "moderate" level of concern (prior to mitigation, failure could result in minor injury or erroneous diagnosis/delay in care leading to minor injury). Verification activities support this.

    Study Details:

    The document describes software verification testing as the primary study type to establish substantial equivalence for the software updates.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not specified. Software verification typically involves testing against a range of inputs and scenarios, but a "sample size" in the context of clinical data for performance metrics is not applicable here as no clinical performance data is presented.
      • Data Provenance: Not specified, as it's software verification, not a clinical data study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This was software verification, not a study requiring expert-established ground truth on clinical data. The "ground truth" for software testing would be the expected behavior of the software according to its design specifications.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable. This was software verification.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a multi-reader multi-case comparative effectiveness study was not done. The device (Endoflip™ 300 System) is a "Gastrointestinal Motility Monitoring System" that measures pressure and dimensions; it's not described as an AI-powered diagnostic tool for interpretation, but rather a direct measurement device with software for processing and displaying those measurements. The updates were minor software enhancements and bug fixes.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, in a sense. The "study" was software verification, which inherently evaluates the algorithm's performance on its own against specifications, without human interpretation in the loop for diagnostic accuracy. The device itself is standalone in its measurement function, providing data to clinicians for their interpretation.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For software verification, the "ground truth" would be the software's design specifications and requirements. Each test case has an expected output or behavior based on these specifications, and the software's actual output is compared against this expected behavior. This is not clinical ground truth like pathology or expert consensus.
    7. The sample size for the training set:

      • Not applicable. This was software verification testing for an updated version of an existing medical device, not a machine learning model involving a training set.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223705
    Device Name
    EndoflipTM 300
    Manufacturer
    Date Cleared
    2023-04-14

    (123 days)

    Product Code
    Regulation Number
    876.1725
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Endoflip™ 300 System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters in adults and to measure pressure and dimensions in the esophagus, in patients from 5 years of age. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.

    Device Description

    The Endoflip™ 300 System is the next generation of the predicate Endoflip™ System consisting of design changes to the device hardware and software components. The design changes improve device usability when compared to the predicate. Changes were made to the platform components only; no design changes were made to the system catheters (Endoflip™ or Esoflip™) except for labeling changes (not related to indications for use). The system is comprised of a pump, display, cart and accessories, including pre-use tube and balloon catheters.

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter and summary for the Endoflip™ 300 System. It describes the device, its intended use, and the types of testing performed to demonstrate substantial equivalence to a predicate device.

    However, the document does not contain information on acceptance criteria for a performance study evaluating the device's diagnostic accuracy or effectiveness against a ground truth, nor does it describe a study specifically designed to prove the device meets such criteria in terms of clinical performance.

    The performance data summarized in section VII focuses on engineering, safety, and usability aspects of the device's hardware and software components, rather than its clinical diagnostic or treatment efficacy. Specifically, the document states:

    • "Clinical studies were not required to demonstrate the safety and performance of the Endoflip TM 300 System."

    Therefore, I cannot provide the requested information about acceptance criteria for clinical performance and a study proving those criteria are met. The document indicates that for this particular 510(k) submission, clinical studies to demonstrate safety and performance (in the sense of diagnostic or treatment efficacy) were not deemed necessary, likely due to the device being a "next generation" of an already cleared predicate with no change to the indications for use and improvements primarily in usability and platform components.

    To answer your specific points based on the provided text:

    1. A table of acceptance criteria and the reported device performance: Not applicable. The document describes engineering, safety, and usability testing, not clinical performance against specific diagnostic or treatment acceptance criteria.
    2. Sample sized used for the test set and the data provenance: Not applicable for clinical performance. For engineering, safety, and usability tests, the sample sizes and data provenance are not specified in this summary, but would be part of the underlying test reports.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth was established for the performance studies presented.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a gastrointestinal motility monitoring system, not an AI-assisted diagnostic tool for image interpretation by human readers.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable in the context of diagnostic performance. The device itself is a measurement system; its "standalone" performance relates to its ability to accurately measure pressure and dimensions, which would be verified through mechanical and software testing.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable for clinical performance evaluation. The "ground truth" for the tested engineering aspects would be established against calibrated standards or specifications.
    8. The sample size for the training set: Not applicable. The document does not describe a machine learning or AI algorithm that would require a training set for diagnostic purposes.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document focuses on the engineering and design controls demonstrating the safety and effectiveness for a device that is essentially an updated version of a previously cleared system, rather than a novel diagnostic or therapeutic device requiring extensive clinical performance studies to establish efficacy against a clinical ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223158
    Manufacturer
    Date Cleared
    2023-01-23

    (108 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien, llc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LigaSure XP Maryland Jaw Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure XP Maryland Jaw Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as colorectal, bariatric, urologic, vascular, thoracic, and gynecologic. These may include, but are not limited to, such procedures as Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, and so forth.

    The LigaSure XP Maryland Jaw Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure XP Maryland Jaw Sealer/Divider for these procedures.

    Device Description

    The proposed LigaSure™ XP Maryland Jaw Sealer/Divider is a sterile, single-use bipolar vessel sealer. It is labeled as prescription use only. The device connects to the Valleylab™ FT10 Electrosurgical Platform generator for tissue ligation. Energy is applied to tissue interposed between the nano-coated Maryland-style jaws creating a ligation. The jaws contain an independent cutting blade for division of tissue. A distinctive characteristic is the new continuous rotation capability of the jaws and shaft. The shaft is a common 5 mm diameter available in three lengths (23 cm, 37 cm, 44 cm) for various general surgical procedures in both open and minimally invasive (laparoscopic) approaches. The instrument body is a pistol grip design which can be used by right or left-handed users to access the controls. The new device is offered with two handle body designs, One-Step Sealing (LXMJ23S, LXMJ37S, LXMJ44S) or Latching Handle (LXMJ23L, LXMJ37L, LXMJ44).

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and study proving a digital health device's performance. Instead, it describes a substantial equivalence determination for a surgical instrument, the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Therefore, many of the requested categories in your prompt, such as "number of experts used to establish the ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "training set size," and "ground truth for the training set," are not applicable to this medical device submission, which focuses on hardware and electrosurgical performance rather than an AI-powered diagnostic or assistive tool.

    However, I can extract the relevant performance testing and acceptance criteria information that is present in the document for the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Here's the information structured to the best of my ability given the provided document:


    Acceptance Criteria and Study for LigaSure™ XP Maryland Jaw Sealer/Divider

    This submission details the substantial equivalence of an electrosurgical cutting and coagulation device, the LigaSure™ XP Maryland Jaw Sealer/Divider. The performance testing described focuses on various engineering, biological, and functional aspects of the surgical instrument, not on the performance of a digital health or AI device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Predetermined Device Requirements)Reported Device Performance
    SterilizationCompliance with ISO 11135, ISO 11737-1, ISO 11737-2, ISO 10993-7Met acceptance criteria for sterilization by ethylene oxide (EO).
    Shelf LifeCompliance with ISO 11607-1, ISO 11607-2, ASTM D4169, ASTM F1980Packaging and product integrity acceptable for five-year shelf life.
    BiocompatibilityCompliance with ISO 10993-1 for: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.Met requirements for all specified endpoints.
    SoftwareN/A (Device does not contain software)N/A
    Electromagnetic Compatibility (EMC) & Electrical SafetyCompliance with IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18Met applicable clauses of standards.
    Bench PerformancePredetermined device requirements for: mechanical/functional performance, visual inspection, device reliability, ex vivo vessel sealing/burst, and ex vivo lymphatic sealing/burst testing.Met all predetermined device requirements.
    Animal PerformancePredetermined device requirements for: in vivo acute hemostasis, acute thermal spread, and chronic hemostasis tissue testing.Met all predetermined device requirements.
    Clinical PerformanceSufficiency of clinical literature to support safety and effectiveness for stated indications.Clinical literature studies were evaluated to further support safety and effectiveness.
    Technological CharacteristicsComparison to predicate devices for aspects like indications for use, sterility, packaging, instrument design, lever latching mechanism, energy activation, shaft/jaws rotation, jaw type/shape, electrical characteristics, compatible energy platform, operating amperage/voltage, direct tissue contacting materials, patient contacting materials, jaw coating, and shaft material.Differences (e.g., continuous 360° rotation, removal of heat shrink on shaft) were deemed not to affect safety and effectiveness and not to raise different questions of safety and effectiveness as demonstrated through performance testing.

    2. Sample size used for the test set and the data provenance
    The document does not specify exact sample sizes for each test set (e.g., number of vessels, lymphatic segments, or animals).

    • Provenance: The tests described (bench, animal) are laboratory-based performance studies, likely conducted by the manufacturer Covidien, llc. The document does not mention data provenance in terms of country of origin or whether human data was retrospective or prospective, as clinical trials for efficacy/safety in humans were not the primary evaluation method shown (literature review was used for clinical support).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable and not provided. The "ground truth" for this device's performance would be objective measurements in engineering, biological models, and animal studies (e.g., burst pressure, hemostasis, thermal spread), not expert consensus on interpretations.

    4. Adjudication method for the test set
    This is not applicable and not provided. Performance outcomes are likely based on direct measurement against predefined physical/biological thresholds.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This is not applicable. This device is a surgical instrument, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This is not applicable. This device is a surgical instrument, not an algorithm.

    7. The type of ground truth used
    The "ground truth" for this device's performance is based on:

    • Engineering standards (e.g., ISO, ASTM, IEC for sterilization, packaging, electrical safety).
    • Biocompatibility standards (ISO 10993-1).
    • Predefined physical and biological thresholds/measurements from ex vivo (vessel/lymphatic sealing/burst) and in vivo (hemostasis, thermal spread) studies.
    • "Clinical literature studies were evaluated to further support the safety and effectiveness use of the proposed LigaSure™ XP Maryland Jaw Sealer/Divider indications for use." This implies that existing clinical evidence for similar devices and principles contributes to the overall justification, but no new clinical study data from the sponsor is explicitly referenced.

    8. The sample size for the training set
    This is not applicable. This is a hardware device; there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established
    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213911
    Manufacturer
    Date Cleared
    2022-07-29

    (226 days)

    Product Code
    Regulation Number
    868.1400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Microstream™ CO2 NanoPod, when connected to the host monitor, is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the respiration cycle, EtCO2, FiCO2, and CO2-based respiration rate for adult, pediatric and neonatal patients, utilizing Microstream™ Advance CO2 sampling lines.

    The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/ hr) and oxygen desaturation index (ODI) values. IPI is intended for pediatric and adult patients only. A/hr and ODI are intended for age 22 and up.

    IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SPO2 data to the NanoPod in order to calculate these parameters.

    The Microstream™ CO2 NanoPod is intended for use in hospitals, hospital-type facilities, during intra-hospital transport between and within areas of care.

    Device Description

    The Microstream™ CO2 NanoPod is a finished medical device that when connected to a host monitor is intended to provide continuous, non-invasive measurement and monitoring of carbon dioxide (CO2) during the respiration cycle, including End-tidal CO2 (EtCO2), Fractional Concentration of Inspired CO2 (FiCO2), and CO2 based respiration rate, for adult, pediatric, and neonatal patients utilizing Microstream™CO2 or Microstream™ Advance CO2 sampling lines.

    The Microstream™ CO2 NanoPod also provides the clinician with integrated pulmonary index (IPI), apnea per hour (A/hr), and oxygen desaturation index (ODI) values. The IPI measurement is intended for use in pediatric and adult patients only, and A/hr and ODI measurements are intended for age 22 and up. Both IPI and ODI values can be calculated and displayed only if the host monitor collects SpO2 data and provides this SpO2 data to the NanoPod to calculate these parameters.

    The Microstream™ CO2 NanoPod encloses a NanoMediCO2 capnography module, which measures inspired/expired carbon dioxide and respiration rate. The NanoPod is powered by the host monitor on a DC supply.

    The Microstream™ CO2 NanoPod system includes the following sub-systems:

    • . NanoPod is the medical device enclosure that encloses the NanoMediCO2 module. The NanoPod has a connected active cable that has a LEMO connector end, which connects to a host monitoring system.
    • . NanoMediCO2 capnography module is enclosed in the NanoPod. The NanoMediCO2 module measures inspired and expired carbon dioxide and respiration rate. The active cable is what provides bidirectional communication between the Host monitor and the NanoMediCO2 module enclosed inside the NanoPod.
    • . Interface Board is enclosed inside NanoPod which consists of an LED exposed on the outside of the NanoPod to provide the functional status of the subject device to the user.
    • l Cradle is a plastic holder used to connect the NanoPod with a GCX clamp or other mounting solution, which can then be used to mount the NanoPod on a pole or bedrail. There are no sharp edges on the Microstream™ CO2 NanoPod, which enables the medical device to be held in the hand at times when it is removed from its mounting cradle.

    The Microstream™ and or Microstream™ Advance CO2 sampling lines can attach to the NanoPod CO2 port on the NanoPod enclosure. Microstream™ Capnography parameters are displayed the Microstream™ and or Microstream™ Advance CO2 sampling line is connected to the NanoPod CO2 Port and the Microstream™ CO2 NanoPod is connected via LEMO connector to a host monitoring system. Please note that the Microstream™ and Microstream™ Advance sampling lines have been cleared in their own respective 510K submissions.

    AI/ML Overview

    Due to the nature of the provided document (a 510(k) summary for a medical device), the information typically presented as "acceptance criteria" and "study proving it meets criteria" from the perspective of an AI/software device is not directly applicable. This document describes a hardware medical device (CO2 monitor) and its regulatory clearance process, which relies on demonstrating substantial equivalence to a predicate device rather than meeting specific performance metrics against a clinical ground truth via a study.

    Therefore, many of the requested fields (sample size, expert qualifications, adjudication, MRMC, standalone performance, etc.) are not available in this document because they pertain to clinical studies or AI/software validation, which were explicitly stated as not required for this device's clearance.

    However, I can extract the information relevant to this medical device's clearance process as best as possible, interpreting "acceptance criteria" as the performance and safety requirements it needed to meet for its 510(k) clearance, and "study" as the testing performed.

    Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (Required Performance/Safety)Reported Device Performance (Summary of Testing)
    Compliance with Performance Standards (ISO 80601-2-55 for CO2 monitors)The Microstream™ CO2 NanoPod has "successfully undergone performance, safety, electromagnetic and environmental testing to ensure compliance with expected performance standards," including ISO 80601-2-55.
    Compliance with Main Safety Standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 80601-2-49 via section 202.8.102, Figure 202 104)The device underwent testing to ensure compliance with these electrical safety, EMC, and usability standards.
    Software Validation (to ensure performance and substantial equivalence)"Software testing was performed to validate the performance of the Microstream™ CO2 NanoPod and its substantial equivalence to the predicate device."
    Hazard Analysis (ISO 14971:2019) demonstrating acceptable residual risks"A hazard analysis was carried out on the Microstream™ CO2 NanoPod in compliance with ISO 14971:2019. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the device."
    Substantial Equivalence to Predicate Device (Capnostream™35 Portable Respiratory Monitor, K150272) in intended use, technological characteristics, and performanceThe document repeatedly states that the device is "substantially equivalent" to the predicate, with differences detailed (external module vs. standalone monitor, no SpO2 in subject device). "Based on the results of the verification and validation studies, Covidien has established that the subject device, the Microstream™ CO2 NanoPod, is substantially equivalent to the predicate device." "The subject device, Microstream™ CO2 NanoPod is equivalent in intended use, technological characteristics, and performance to the existing legally marketed device. Testing did not raise any issue of safety and effectiveness."

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not applicable. This was a hardware device undergoing bench and software testing, not a clinical study with patient data.
    • Data Provenance: Not applicable. The testing was laboratory-based, focused on device performance according to engineering and regulatory standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. "Ground truth" in the clinical sense (e.g., diagnosis by experts) was not established. The "ground truth" for the device's performance would be the reference measurements from calibrated equipment against which the device was tested.

    4. Adjudication method for the test set:

    • Not applicable. There was no clinical test set requiring adjudication of findings.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a measurement tool (CO2 monitor), not an AI-assisted diagnostic device, and no MRMC study was performed or required.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a hardware device; the concept of a "standalone algorithm" is not relevant here. The device itself performs the measurements.

    7. The type of ground truth used:

    • The "ground truth" for the device's performance would be the established reference standards and calibrated instrumentation used in the bench performance testing to verify the accuracy of CO2 measurements, respiration rate, etc. This is typical for the validation of medical measurement devices.

    8. The sample size for the training set:

    • Not applicable. This device does not use a "training set" in the machine learning sense. Its operation is based on established physical principles (Non-dispersive infrared spectroscopy) and predefined algorithms.

    9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN220006
    Device Name
    ProdiGI
    Manufacturer
    Date Cleared
    2022-06-13

    (150 days)

    Product Code
    Regulation Number
    876.4410
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ProdiGI Traction Wire: The Medtronic ProdiGI Traction Wire is indicated to grasp tissue within the esophagus, stomach, and colon of adults during an Endoscopic Submucosal Dissection (ESD) procedure.

    ProdiGI Traction Magnet: The Medtronic ProdiGI Traction Magnet is indicated to grasp tissue within the stomach and colon of adults during an Endoscopic Submucosal Dissection (ESD) procedure.

    Device Description

    The ProdiGI Traction System includes the Traction Wire and Traction Magnet devices. The Traction Wire and Traction Magnet are not intended to be used together. Both devices are used in adults only and are used to provide improved visualization of the submucosal space during an endoscopic submucosal dissection (ESD) procedure.

    ProdiGI Traction Wire Device: The Traction Wire consists of two graspers: a primary tissue grasper with traction wire attached (ERD-TW20 and ERD-TW35), and a secondary tissue grasper (ERD-TWSG) without a wire. The secondary tissue grasper is used to secure the distal end of the traction wire. The traction wire is a nitinol shape-memory loop (2.0 cm or 3.5 cm in length) that provides tension to the attached tissue after deployment. The nitinol wire is attached to the grasper with a stainless-steel crimp.

    ProdiGI Traction Magnet Device: The traction magnet (ERD-TMST and ERD-TMLG) consists of two identical tissue graspers with a permanent neodymium magnet attached to the grasper via a suture. The sutured magnet provides tension to the attached tissue after deployment.

    AI/ML Overview

    Here's an analysis of the provided text to describe the acceptance criteria and the study proving the device meets them, based on your requested format:

    Acceptance Criteria and Device Performance for ProdiGI

    The ProdiGI device underwent non-clinical/bench studies and usability testing to demonstrate its safety and effectiveness. Importantly, no clinical data was provided. The primary studies proving the device meets acceptance criteria are summarized in the "Performance Testing - Bench", "Usability Testing", and "Performance Testing - Animal" sections of the provided document.

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document presents acceptance criteria and reported performance in a narrative and tabular format for both bench and usability testing. I will consolidate the key performance criteria for each device type into a single table. Note that specific numerical results (e.g., 'b)(4)') were often redacted in the original text, so I will indicate where this occurs.

    Table 1: Consolidated Acceptance Criteria and Reported Device Performance (Bench and Usability)

    Test CategorySpecific Test / Needs IDAcceptance CriteriaReported Device Performance
    BiocompatibilityGeneralConformance to ISO 10993-1:2003 for cytotoxicity, sensitization, irritation, material-mediated pyrogenicity, and acute systemic toxicity.PASS: "The results supported the biocompatibility of the ProdiGI device." (Tests performed on final finished devices.)
    Shelf Life/SterilitySterilizationConformance to ANSI/AAMI/ISO 11135:2014, ensuring a sterility assurance level of 10^-6.PASS: "Sterilization was evaluated for conformance..." (Implied successful validation as no issues reported and device granted.)
    Accelerated AgingSupport 1.5-year shelf life per ASTM F1980-16, demonstrating packaging integrity (gross leak, pouch seal strength).PASS: "Accelerated aging to support a 1.5-year shelf life was performed... The shelf life was verified by demonstrating packaging integrity..." (Implied successful validation.)
    Bench/Mechanical (Traction Wire)Tensile Test (Grasper to Wire)Grasper to wire tensile force should be above (b)(4) lbf.PASS: "The device passed all the tests in Tables 1 and 2." (Table 1 states "The grasper to wire tensile force should be above (b)(4)" for both, implying the criteria was met.)
    Tensile Test (Crimp to Wire)Crimp to wire tensile test must be above (b)(4). lbf.PASS: "The device passed all the tests in Tables 1 and 2." (Table 1 states "The crimp to wire tensile test must be above (b)(4)" for both, implying the criteria was met.)
    Torque TestTorque to rotate device (b)(4) lbf; 2nd grasper > (b)(4) lbf; 2nd grasper arms not come out of capsule; force to remove 2nd graspers from tissue (b)(4) lbf and = (b)(4) lbf; force to remove magnet from suture >= (b)(4) lbf; force to separate crimp from suture >= (b)(4) lbf; force to remove magnet from grasper body (b)(4) lbf; arms not come out of capsule during detachment; force to remove grasper from tissue = (b)(4) prior and after being opened/closed (b)(4).PASS: "The device passed all the tests in Tables 1 and 2." (Implied compliance with all criteria based on "passed all tests" statement.)
    Deployment ForceForce to deploy by squeezing handle: > (b)(4) lbf and (b)(4) lbf (lower); = (b)(4).PASS: "The device passed all the tests in Tables 1 and 2." (Implied compliance with all criteria based on "passed all tests" statement.)
    Usability (Traction Wire)N1.1 (Visualization)Average visualization score (dissection with traction wire) >= average visualization score (without traction wire). (3-point scale: 1=unacceptable, 2=acceptable, 3=exceeds expectations)PASS: Test: 15/24, Control: (b)(4) (Specific average values are redacted, but "PASS" indicates the criterion was met. This implies the average score with the device was at least as good as without.)
    N1.2 (Procedure Time)Average procedure time (with traction device) = average visualization score (without traction magnet). (5-point scale: 1=unacceptable, 2=poor, 3=clinically acceptable, 4=good, 5=exceeds expectations)PASS: (Specific score values redacted, but "PASS" indicates the criterion was met. This implies the average score with the device was at least as good as without.)
    NL2 (Procedure Time)Average procedure time (with traction device) 5 years of experience and three or more ESD cases per month.
    *   For the **Traction Wire**, "skilled by moderate and provice physicians participated."
    *   For the **Traction Magnet**, "two skilled, two moderate and two novice physicians participated."
    *   The observations and ratings from these physicians formed the basis of the usability ground truth.
    
    • Animal Testing: Bleeding, perforation, mucosal laceration, visualization, ease of removal, and ease of procedure were assessed by "study physicians intraoperatively via endoscopic visualization." No specific number of physicians is given for the animal studies, but they are implied to be experts in therapeutic endoscopy and ESD.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method (e.g., 2+1, 3+1) for the usability or animal studies. It appears that the expert physicians directly provided the ratings and assessments which were then used to calculate means and comparisons. For perforation in the animal study, it notes a singular assessment that was then reviewed for causality ("unrelated to the device use or design or the control procedure").

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    While not explicitly termed an "MRMC study," the usability testing and animal (in vivo) testing were designed as comparative effectiveness studies where human readers (physicians/experts) performed procedures with and without AI (device) assistance. The "AI" here refers to the device itself as providing assistance, not an AI algorithm in the traditional sense of image analysis.

    • Effect Size of Human Readers (with device vs. without):
      • Visualization: For both Traction Wire and Traction Magnet, the device showed PASS for visualization, meaning the average visualization score with the device was equal to or better than without. This indicates a positive effect on visualization compared to unassisted procedures. For the animal study, Visualization (Start, Mid, End) was consistently rated as Traction better than Control for both devices.
      • Procedure Time: For the Traction Wire, procedure time was PASS (device time
    Ask a Question

    Ask a specific question about this device

    K Number
    K212555
    Manufacturer
    Date Cleared
    2022-05-06

    (266 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Covidien llc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nellcor™ OxySoftN single patient use sensor is indicated for use with neonatal, infant, pediatric and adult patients during both no motion and motion conditions and for patients who are either well or poorly perfused, in hospitals, hospital-type facilities, intra-hospit, mobile emergency medical applications including both ground and air transport and home environments. Transport environments include intra-hospital transport and air emergency transport: road ambulances and fixed-wing aircraft and helicopters. The Nellcor™ OxySoft SpO2 Sensor works in conjunction with monitoring systems that use OxiMax™ and Nellcor compatible pulse oximetry systems to facilitate spot checking or continuous monitoring of a patient's arterial hemoglobin (SpO2) and pulse rate via topical application of the sensor over a pulsating arteriolar vascular bed, such as a finger, foot or hand. This device is for prescription use only.

    Device Description

    The Nellcor™ OxySoft™ Neonatal/Adult SpO2 Sensor, OxySoftNHC monitors the amount of oxygen in the patient's blood (oxygen saturation) or SpO2 and pulse rate after applying the sensor on the finger, foot, or hand. The sensor utilizes a pair of light emitting diodes (LEDs) employing light at two wavelengths, and the time varying absorbance of the tissue due to the pulsatile blood signal is obtained, amplified from a photodiode to determine functional arterial oxygen saturation. A clinician will prescribe the sensor for use as spot check or continuous use in hospitals, hospital-type facilities, intra-hospital transport, mobile emergency medical applications including both ground and air transport and home environments. The OxySoftNHC sensor can be used with all Nellcor™ compatible systems. The sensor consists of a sensor assembly and cable assembly and is provided sterile (EtO) and is meant for single use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies that prove the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance

    Device Name: Nellcor™ OxySoft™ Neonatal-Adult SpO2 Sensor, OxySoftN/OxySoftNHC

    Acceptance CriteriaReported Device Performance
    Pulse rate accuracy+/- 3 BPM RMS (non-motion)
    +/- 5 BPM RMS (motion, 70-100% SpO2 range, using N-600x)
    SpO2 Accuracy (Adult and Neonate)70% to 100% (+/- 2% RMS) (non-motion)
    70% to 100% (+/- 3% RMS) (motion)
    SpO2 Accuracy (Low Saturation)60% to 80% (+/- 3% RMS)
    SpO2 Accuracy (Low Perfusion)70% to 100% (+/- 2% RMS)
    Sensor Off Feature FunctionalityDisplays "Sensor Off" message within 1 minute at least 90% of the time (Measured at 99.6% with 95% lower bound of 98.3%)
    Gentleness on Fragile SkinAdhesive removed significantly less cells than predicate and caused minimal increase in TEWL.
    Adhesion after RepositioningRemains adhered to the skin after 18th repositioning.
    BiocompatibilityMeets ISO 10993-1 and FDA Biocompatibility Guidance (Cytotoxicity, Sensitization, Intracutaneous reactivity, Chemical characterization and toxicological risk assessment passed).
    Electrical SafetyComplies with IEC 60601-1.
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2.
    Mechanical, Use Life, Light Shunting, Storage, Ambient Light, Ingress Protection, Shock & Vibration, Time to Post, Adhesive Liner Card and Initial Peel Force, Nellcor Functional Compatibility, Surface Temperature Rise, Operational Altitude & Temperature & Humidity, Transient Temperature & Humidity, Human FactorsMet for all conditions of use.

    Study Details

    The provided text describes several clinical and non-clinical studies to demonstrate the device's performance.

    1. "MDT20002VALNIC: Performance evaluation of the OxySoftN neonatal sensor in the NICU compared to SaO2"

    • Sample size: 27 patients (neonates)
    • Data Provenance: Clinical study, prospective (data collected to evaluate sensor in NICU). No country of origin is explicitly stated, but clinical studies for FDA clearance are typically conducted in the US or in countries adhering to similar ethical and regulatory standards.
    • Experts for ground truth: Not applicable – ground truth for SpO2 was established through CO-oximeter analysis of simultaneously drawn arterial blood samples, which is an objective measurement.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This was a standalone performance evaluation.
    • Standalone performance: Yes, this study directly evaluated the device's accuracy against a reference standard.
    • Type of ground truth: Objective measurement (SaO2 values obtained from CO-oximeter analysis of arterial blood).
    • Training set sample size: Not specified as this is a clinical validation study, not an algorithm training study.
    • Training set ground truth: Not applicable.

    2. "MDT20037CYBOXY: Validation of the OxySoftN to Determine the Gentleness on Fragile Skin"

    • Sample size: Not explicitly stated but implied through "healthy elderly subjects."
    • Data Provenance: Clinical study, prospective. No country of origin is explicitly stated.
    • Experts for ground truth: Not applicable – ground truth involved objective measurements like evaporative water loss (TEWL) and skin protein disruption.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This study compared the subject device to the predicate device regarding skin gentleness.
    • Standalone performance: Yes, it directly measured the device's effect on skin.
    • Type of ground truth: Objective measurements (TEWL, amount of skin protein disruption).
    • Training set sample size: Not applicable.
    • Training set ground truth: Not applicable.

    3. "MDT20047OXSNOF: Sensor-Off Validation of the OxySoftN sensor in Healthy Adults"

    • Sample size: Not explicitly stated but implied through "a sequence of steps for different sensor off configurations was performed per subject."
    • Data Provenance: Clinical study, prospective. No country of origin is explicitly stated.
    • Experts for ground truth: Not applicable – ground truth was the observation of whether a "Sensor Off" message appeared.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This was a standalone functional validation.
    • Standalone performance: Yes, it evaluated a specific safety feature of the device.
    • Type of ground truth: Objective observation (presence/absence of "Sensor Off" message within 1 minute).
    • Training set sample size: Not applicable.
    • Training set ground truth: Not applicable.

    4. "MDT20038CYBADH: Multiple Reposition Peel Adhesion on Human Skin Validation, OxySoftN"

    • Sample size: Not explicitly stated but implied through "volar forearm of healthy adults."
    • Data Provenance: Clinical study, prospective. No country of origin is explicitly stated.
    • Experts for ground truth: Not applicable – ground truth was the objective measurement of peel adhesion.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This was a standalone functional validation.
    • Standalone performance: Yes, it directly measured the adhesion performance.
    • Type of ground truth: Objective measurement (peel adhesion after repositioning).
    • Training set sample size: Not applicable.
    • Training set ground truth: Not applicable.

    5. "MDT20028OXYLOV: SpO2 Accuracy Validation of the OxySoftN sensor via Reference CO-Oximetry in Healthy, Well-Perfused Subject"

    • Sample size: "healthy adult subjects" - specific number not provided.
    • Data Provenance: Clinical study, prospective. No country of origin is explicitly stated.
    • Experts for ground truth: Not applicable – ground truth for SpO2 was established through CO-oximeter analysis of simultaneously drawn arterial blood samples, and for pulse rate, via 3-lead ECG.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This was a standalone performance evaluation.
    • Standalone performance: Yes, this study directly evaluated the device's accuracy against a reference standard.
    • Type of ground truth: Objective measurement (SaO2 values from CO-oximeter; 3-lead ECG for pulse rate).
    • Training set sample size: Not applicable.
    • Training set ground truth: Not applicable.

    6. "MDT20006OXYVMT: SpO2 Accuracy Validation of the OxySoft sensor via Reference CO-Oximetry Motion Study"

    • Sample size: "diverse subject population" - specific number not provided.
    • Data Provenance: Clinical study, prospective. No country of origin is explicitly stated.
    • Experts for ground truth: Not applicable – ground truth for SpO2 was established through CO-oximeter analysis.
    • Adjudication method: Not applicable.
    • MRMC comparative effectiveness study: No. This was a standalone performance evaluation.
    • Standalone performance: Yes, this study directly evaluated the device's accuracy against a reference standard under motion and non-motion conditions.
    • Type of ground truth: Objective measurement (SaO2 values from CO-oximeter).
    • Training set sample size: Not applicable.
    • Training set ground truth: Not applicable.

    General Notes on Training Sets:
    For medical devices like oximeters, "training set" and "ground truth for training set" are typically relevant to machine learning algorithms. Pulse oximeters, as described, use a fundamental technology based on spectrophotometry and plethysmography, which are physics-based principles, not typically machine learning algorithms requiring explicit training data in the same way an AI diagnostic tool would. The performance data presented here are for validation and verification of the physical sensor and its underlying technology against established physiological standards and safety requirements.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 9