Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203394
    Device Name
    PleuraFlow System with FlowGlide
    Manufacturer
    Date Cleared
    2021-03-02

    (104 days)

    Product Code
    Regulation Number
    878.4780
    Why did this record match?
    Applicant Name (Manufacturer) :

    ClearFlow, Inc,

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    The PleuraFlow® System with FlowGlide® is indicated for use during cardical procedures and chest trauma. Its Active Clearance Technology® proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.
    Device Description
    The PleuraFlow® System with FlowGlide(R) PFFG3 iss a modification to our current PleuraFlow® System with FlowGlide(R) (predicate). The primary components of the System are the Chest Tube and the Clearance Apparatus. The Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister. The Clearance Apparatus that is part of the PleuraFlow System with FlowGlide consists of a Guide Tube with a magnetic shuttle and a PTFE-coated Clearance Wire with a Loop set on its distal end. The Clearance Apparatus is advanced into the PleuraFlow with FlowGlide Chest Tube using a magnetic Shuttle. When indicated, the Clearance Wire and Loop is advanced and retracted within the PleuraFlow with FlowGlide Chest Tube to proactively prevent or break up and clear any tube obstructions or clogging to keep the tube patent. The PleuraFlow System with FlowGlide, PFFG3, has a new shuttle design that includes a spring activated button to allow the user to increase the magnetic force when needed. The range of magnetic force is the same as predicate.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1