Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243312
    Device Name
    CER-S
    Manufacturer
    Date Cleared
    2025-02-20

    (121 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    CardioCalm Srl

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CER-S is a medical device stand-alone software intended for the analysis, editing, review and reporting of continuous ECG digital recordings, to support the clinician in the interpretation of the trace and the diagnosis of cardiac rhythm disorders in adult and pediatric patients.

    CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device. This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

    CER-S analysis provides indications for evaluation of:

    • Patients with rhythm disturbances (cardiac arrhythmias),
    • Patients with transient myocardial ischemia,
    • Patients with pacemaker (only if pacing detection is available from the input recording),
    • Patients needing HRV evaluation.
    • Newborn patients limited to QRS detection.
    Device Description

    CER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER).
    Different modules provide:

    • ECG Beat detection and classification,
    • analysis of ECG rhythm, arrhythmia detection,
    • interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows,
    • interactive Continuous ECG Viewer,
    • interactive display/management of ECG Templates,
    • holter-like report for analyzed Continuous ECG records,
    • record exportation, in ISHNE format,
    • generation of aECG FDA HL7 XML (v. 2).

    Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary.
    Moreover, CER-S allows the automatic analysis for the following patch location:

    • two-electrode patches positioned in the left upper chest area at a roughly inclined angle
    • three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle
    • three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternum
      In all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the CER-S device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document (a 510(k) Premarket Notification) does not explicitly state acceptance criteria in terms of specific performance metrics with numerical thresholds (e.g., sensitivity > X%, specificity > Y%). Instead, it outlines the scope of performance testing and concludes that the device is "substantially equivalent" to its predicate and meets the relevant standards.

    The document focuses on demonstrating that the device adheres to recognized industry standards and its performance is comparable to the predicate.

    Acceptance Criteria (Implied from the document)Reported Device Performance
    Compliance with ECG performance testing standards.Non-clinical tests were conducted according to:
    • IEC 60601-2-47:2012 (Recognition Number 3-155)
    • IEC 60601-2-25:2011 (Recognition Number 3-105)
    The testing demonstrated that CER-S is substantially equivalent to the predicate for the proposed intended use.
    Compliance with software development lifecycle standards.Software development life cycle was conducted according to:
    • IEC 62304:2015 (Recognition Number 13-79)
    Compliance with usability engineering process standards.Usability engineering process was conducted according to:
    • IEC 62366-1:2020 (Recognition Number 5-129)
    Compliance with risk management standards.Potential hazards were evaluated and controlled through Risk Management activities, according to:
    • ISO 14971:2019 (Recognition Number 5-125)
    Equivalence in intended use, technological characteristics, and safety/effectiveness vs. predicate.The subject device has the same intended use as the predicate device (K213861). Any differences in technological characteristics (Enhanced software usability, Option to activate and run pacemaker analysis, Overall optimization of embedded algorithms for beat and noise detection, rhythm analysis and beat measurements, Strengthened cybersecurity measures) do not raise different questions of safety or effectiveness. The device has been shown to be substantially equivalent to the legally marketed predicate devices based on indications for use, technological characteristics, and comparison.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." This implies that there isn't a "test set" in the traditional sense of a clinical dataset used to evaluate the device's diagnostic performance against a ground truth.

    The performance evaluation relied on non-clinical tests against recognized standards rather than a clinical dataset with individual patient records. Therefore, information about sample size, country of origin, or retrospective/prospective nature of a clinical test set is not available in this document.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    As there were no clinical tests performed with a ground truth established by experts, this information is not applicable/not available from the provided text.

    4. Adjudication Method for the Test Set

    Since no clinical test set was used with expert review, an adjudication method is not applicable/not available from the provided text.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Improvement with AI vs. Without AI Assistance

    The document explicitly states, "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." Therefore, no MRMC comparative effectiveness study was conducted, and information about human reader improvement with AI assistance is not available. The CER-S is a standalone software, not an AI-assisted reading tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was done through non-clinical testing against recognized standards. The device itself is described as a "stand-alone software" intended to "support the clinician," not replace them, and explicitly states it "does not provide an automated interpretation and is not intended for use as the only diagnostic tool." The reported performance focused on compliance with ECG performance standards and software/risk management standards, indicating the algorithm's functionality and safety were evaluated.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance evaluation appears to be the requirements and specifications within the referenced industry standards (IEC 60601-2-47, IEC 60601-2-25 for ECG performance; IEC 62304 for software; IEC 62366-1 for usability; ISO 14971 for risk management). The device's algorithms and functions were tested to ensure they met the technical specifications and safety/performance criteria outlined by these standards, rather than a clinical ground truth like pathology or expert consensus on patient cases.

    8. The Sample Size for the Training Set

    The document does not provide information about the training set size for the CER-S algorithms. It mentions "Overall optimization of embedded algorithms for beat and noise detection, rhythm analysis and beat measurements," which implies an underlying development and training process, but no details are given.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213861
    Device Name
    CER-S
    Manufacturer
    Date Cleared
    2022-04-25

    (136 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    CardioCalm Srl

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CER-S is a stand-alone software medical device intended to analyze, edit, review and report digital continuous ECG recordings.

    CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device.

    The device provides the clinician an additional tool to aid in the interpretation of ECG data for diagnosis of heart rhythm disorders in adult and pediatric patients.

    Note: This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

    CER-S analysis provides indications for evaluation of:

    • Patients with rhythm disturbances (cardiac arrhythmias),
    • Patients with transient myocardial ischemia,
    • Patients with pacemaker (only if pacing detection is available from the input recording),
    • Patients needing HRV evaluation,
    • Newborn patients limited to QRS detection.
    Device Description

    CER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER). Different modules provide: ECG Beat detection and classification Analysis of ECG rhythm, arrhythmia detection interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows interactive Continuous ECG Viewer interactive display/management of ECG Templates Holter-like report for analyzed Continuous ECG records, record exportation, in ISHNE format generation of aECG FDA HL7 XML (v. 2), for the submission of continuous ECG recording to the FDA ECG Warehouse. Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. Moreover, CER-S allows the automatic analysis for the following patch location: two-electrode patches positioned in the left upper chest area at a roughly inclined angle three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternum In all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the CER-S device, based on the provided text:

    Acceptance Criteria and Device Performance

    The provided document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a quantitative format (e.g., sensitivity, specificity, accuracy thresholds). Instead, it states that "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." The performance data section focuses on adherence to recognized standards and the demonstration of substantial equivalence to a predicate device.

    However, the document lists improvements in performance over the predicate device, which can be interpreted as target performance goals for these specific aspects:

    Acceptance Criteria (Implied Improvements)Reported Device Performance (Relative to Predicate Device)
    Beat detection in recordings with low amplitude and noise (Improved performance expected)The subject device improves the performance in beat detection in recordings with low amplitude and with noise.
    Beat measures in recordings with higher heart rate (Improved performance expected)The subject device improves the performance of the beat measures in recording with higher heart rate.
    Ability to perform HRV measurements in the frequency domain (New capability/performance)The subject device allows HRV measurements in the frequency domain.
    Ability to detect Sinus Tachycardia (New capability/performance)The subject device allows Sinus Tachycardia detection.
    Usability of the software and graphic interface for editing arrhythmic events (Improved user experience/efficiency)Further minor changes in subject device are developed to enhance the usability of the software and the graphic interface of some its components, which provides a smooth and faster editing of automatically detected arrhythmic events.

    Study Details

    The submission explicitly states: "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." Therefore, the following sections will indicate "Not applicable" or detail the nature of the non-clinical tests conducted.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable (No clinical test set was used).
    • Data Provenance: Not applicable (No clinical test set was used). Non-clinical tests were based on simulations and adherence to specified test standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable (No clinical test set and thus no ground truth established by experts for performance evaluation).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable (No clinical test set requiring adjudication).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. A MRMC comparative effectiveness study was not performed. The device is described as a "stand-alone software medical device" that aids in interpretation and is "not intended for use as the only diagnostic tool."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes. The improvements described, such as "beat detection in recordings with low amplitude and with noise" and "beat measures in recording with higher heart rate," relate to the standalone performance of the algorithm. These improvements were demonstrated through non-clinical tests by adhering to standards such as ANSI/AAMI EC57:1998/(R)2003 and IEC 60601-2-47:2012.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical performance evaluation, the "ground truth" was established based on simulated data and standardized test methodologies defined by the referenced ECG performance testing standards (ANSI/AAMI EC57, IEC 60601-2-47, IEC 60601-2-25) to verify the algorithm's accuracy against known, controlled inputs.

    8. The sample size for the training set:

    • Not specified within the provided text. The document focuses on regulatory submission and does not detail the specifics of machine learning model training data.

    9. How the ground truth for the training set was established:

    • Not specified within the provided text.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193177
    Device Name
    CER-S
    Manufacturer
    Date Cleared
    2020-08-26

    (282 days)

    Product Code
    Regulation Number
    870.2800
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    CardioCalm

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CER-S is a stand-alone software medical device intended to analyze, edit, review and report digital continuous ECG recordings.

    CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device.

    The device provides the clinician an additional tool to aid in the interpretation of ECG data for diagnosis of heart rhythm disorders in adult and pediatric patients.

    Note: This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

    CER-S analysis provides indications for, but not limited to, evaluation of:

    • Patients with rhythm disturbances (cardiac arrhythmias),
    • Patients with transient myocardial ischemia,
    • Patients with pacemaker (only if pacing detection is available from the input recording),
    • Patients needing HRV evaluation,
    • Newborn patients limited to QRS detection.
    Device Description

    CER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER).
    Different modules provide:
    ECG Beat detection and classification Analysis of ECG rhythm, arrhythmia detection interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows interactive Continuous ECG Viewer interactive display/management of ECG Templates Holter-like report for analyzed Continuous ECG records, record exportation, in ISHNE format generation of aECG FDA HL7 XML (v. 2), for the submission of continuous ECG recording to the FDA ECG Warehouse.
    Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary.
    Moreover, CER-S allows the automatic analysis for the following patch location:
    two-electrode patches positioned in the left upper chest area at a roughly inclined angle three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternum
    In all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.

    AI/ML Overview

    The provided text does not contain the detailed information required to fill out all sections of your request regarding acceptance criteria and a study proving device performance. The document is a 510(k) summary for the CER-S device, which outlines its purpose, regulatory information, and a high-level overview of some testing. However, it explicitly states: "No clinical tests were performed to demonstrate safety and effectiveness of CER-S". This means there was no clinical study with human participants to measure the device's performance against specific acceptance criteria in a real-world setting.

    Instead, the submission relies on non-clinical tests (performance testing against existing standards and software development life cycle adherence) and substantial equivalence to a predicate device.

    Therefore, I can only provide information based on what is available in the text. Many of your requested points will be answered with "Not applicable" or "Information not provided in the document."

    Here is the breakdown based on the provided text:


    Acceptance Criteria and Device Performance Study (CER-S)

    As stated in the 510(k) Summary, no clinical tests were performed to demonstrate the safety and effectiveness of CER-S. The submission relies on non-clinical performance testing and software development life cycle adherence to established standards, as well as arguments for substantial equivalence to a predicate device. Therefore, a study demonstrating the device meets performance acceptance criteria via a clinical trial with human-in-the-loop or standalone AI performance metrics is not detailed in this document.

    The document mentions compliance with several non-clinical standards for ECG performance and software development. These standards inherently have their own acceptance criteria, but the specific performance results against these criteria are not numerical values reported in a table within this summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that no clinical study measuring specific performance metrics against clinical acceptance criteria was conducted and detailed in this summary, a table of this nature cannot be generated from the provided text. The document states:

    • "The following non-clinical tests have been conducted on the Subject Device:
      • ECG performance testing, according to: ANSI/AAMI EC57:1998/(R)2003 and IEC 60601-2-47:2012 (Recognition Number 3-155) IEC 60601-2-25:2011 (Recognition Number 3-105)
      • Software development life cycle, according to: ANSI AAMI IEC 62304:2006 (Recognition Number 13-32)
      • Usability engineering process, according to: IEC 62366:2007"
    • "The testing demonstrated that CER-S is safe and effectiveness for the propose intended use."

    While these standards contain acceptance criteria for the device's functions (e.g., ECG signal processing accuracy, software robustness), the specific quantitative results are not reported in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): Not applicable, as no clinical test set/study is detailed. The non-clinical tests would have used synthetic or recorded ECG data, but the "sample size" of such data is not specified.
    • Data Provenance: Not applicable for a clinical test set. The document refers to "data acquired from electrodes with conductive paste/gel... from any device used for the arrhythmia diagnostics." The origin (country, retrospective/prospective) of this data, if used for internal validation, is not specified.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable, as no clinical test set/study requiring expert ground truth is detailed.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable, as no clinical test set/study is detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed or detailed for AI assistance versus human-alone performance. The device is described as an "additional tool to aid in the interpretation of ECG data" and explicitly states it "does not provide an automated interpretation and is not intended for use as the only diagnostic tool." This suggests it's an analysis and review tool, not a standalone AI diagnostic.
    • Effect Size: Not applicable.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop)

    • Standalone Performance Study: No, a standalone (algorithm only) performance study in the sense of an AI diagnostic evaluating cases without human intervention is not described. The device is a software suite for analyzing, editing, reviewing, and reporting, implying human interaction. It also explicitly states it "does not provide an automated interpretation."

    7. Type of Ground Truth Used

    • Type of Ground Truth: Not applicable for a clinical study. For the non-clinical ECG performance testing, the ground truth would be defined by the known characteristics of the ECG signals and simulated arrhythmias as per the testing standards (e.g., ANSI/AAMI EC57).

    8. Sample Size for the Training Set

    • Sample Size (Training Set): Not provided. The document describes the device as a "tool, designed to offer a framework for the interaction of different software-modules," some of which perform "ECG Beat detection and classification Analysis of ECG rhythm, arrhythmia detection." While these functions imply some form of algorithm or model development, the document does not specify if or how an explicit "training set" (in the machine learning sense) was used, nor its size.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth Establishment (Training Set): Not provided. If algorithms within the software were developed using training data, the method for establishing ground truth for such data is not described in this summary.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1