K Number
K213861
Device Name
CER-S
Manufacturer
Date Cleared
2022-04-25

(136 days)

Product Code
Regulation Number
870.1425
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CER-S is a stand-alone software medical device intended to analyze, edit, review and report digital continuous ECG recordings.

CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device.

The device provides the clinician an additional tool to aid in the interpretation of ECG data for diagnosis of heart rhythm disorders in adult and pediatric patients.

Note: This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

CER-S analysis provides indications for evaluation of:

  • Patients with rhythm disturbances (cardiac arrhythmias),
  • Patients with transient myocardial ischemia,
  • Patients with pacemaker (only if pacing detection is available from the input recording),
  • Patients needing HRV evaluation,
  • Newborn patients limited to QRS detection.
Device Description

CER-S is a tool, designed to offer a framework for the interaction of different software-modules, providing advanced solutions for Continuous ECG Recording (CER). Different modules provide: ECG Beat detection and classification Analysis of ECG rhythm, arrhythmia detection interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotations and noise Windows interactive Continuous ECG Viewer interactive display/management of ECG Templates Holter-like report for analyzed Continuous ECG records, record exportation, in ISHNE format generation of aECG FDA HL7 XML (v. 2), for the submission of continuous ECG recording to the FDA ECG Warehouse. Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel (dry and dry/metal electrodes are not intended to be used) placed on standard location in compatible formats from any device used for the arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary. Moreover, CER-S allows the automatic analysis for the following patch location: two-electrode patches positioned in the left upper chest area at a roughly inclined angle three-electrode triangular shape patches positioned on the patient's left upper chest area below the 1st rib, at an inclined angle three-electrode T-shaped patches positioned in the center-thoracic position between the upper part of the chest (manubrium) and the sternum In all cases, patch placement must strictly follow the indication provided by the manufacturer of the certified-patch device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the CER-S device, based on the provided text:

Acceptance Criteria and Device Performance

The provided document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a quantitative format (e.g., sensitivity, specificity, accuracy thresholds). Instead, it states that "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." The performance data section focuses on adherence to recognized standards and the demonstration of substantial equivalence to a predicate device.

However, the document lists improvements in performance over the predicate device, which can be interpreted as target performance goals for these specific aspects:

Acceptance Criteria (Implied Improvements)Reported Device Performance (Relative to Predicate Device)
Beat detection in recordings with low amplitude and noise (Improved performance expected)The subject device improves the performance in beat detection in recordings with low amplitude and with noise.
Beat measures in recordings with higher heart rate (Improved performance expected)The subject device improves the performance of the beat measures in recording with higher heart rate.
Ability to perform HRV measurements in the frequency domain (New capability/performance)The subject device allows HRV measurements in the frequency domain.
Ability to detect Sinus Tachycardia (New capability/performance)The subject device allows Sinus Tachycardia detection.
Usability of the software and graphic interface for editing arrhythmic events (Improved user experience/efficiency)Further minor changes in subject device are developed to enhance the usability of the software and the graphic interface of some its components, which provides a smooth and faster editing of automatically detected arrhythmic events.

Study Details

The submission explicitly states: "No clinical tests were performed to demonstrate the substantial equivalence of CER-S." Therefore, the following sections will indicate "Not applicable" or detail the nature of the non-clinical tests conducted.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable (No clinical test set was used).
  • Data Provenance: Not applicable (No clinical test set was used). Non-clinical tests were based on simulations and adherence to specified test standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable (No clinical test set and thus no ground truth established by experts for performance evaluation).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable (No clinical test set requiring adjudication).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. A MRMC comparative effectiveness study was not performed. The device is described as a "stand-alone software medical device" that aids in interpretation and is "not intended for use as the only diagnostic tool."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes. The improvements described, such as "beat detection in recordings with low amplitude and with noise" and "beat measures in recording with higher heart rate," relate to the standalone performance of the algorithm. These improvements were demonstrated through non-clinical tests by adhering to standards such as ANSI/AAMI EC57:1998/(R)2003 and IEC 60601-2-47:2012.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • For the non-clinical performance evaluation, the "ground truth" was established based on simulated data and standardized test methodologies defined by the referenced ECG performance testing standards (ANSI/AAMI EC57, IEC 60601-2-47, IEC 60601-2-25) to verify the algorithm's accuracy against known, controlled inputs.

8. The sample size for the training set:

  • Not specified within the provided text. The document focuses on regulatory submission and does not detail the specifics of machine learning model training data.

9. How the ground truth for the training set was established:

  • Not specified within the provided text.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

April 25, 2022

CardioCalm Srl Fabio Badilini President Via Martiri della Liberta, nr. 40 Montichiari, Brescia 25018 Italy

Re: K213861

Trade/Device Name: CER-S Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: March 24, 2022 Received: March 29, 2022

Dear Fabio Badilini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213861

Device Name CER-S

Indications for Use (Describe)

CER-S is a stand-alone software medical device intended to analyze, edit, review and report digital continuous ECG recordings.

CER-S is intended for use in clinical and hospital settings only by qualified medical personnel or adequately trained professional personnel working under the supervision and responsibility of a clinician. Users must undergo a thorough software training before using the medical device.

The device provides the clinician an additional tool to aid in the interpretation of ECG data for diagnosis of heart thythm disorders in adult and pediatric patients.

Note: This device does not provide an automated interpretation and is not intended for use as the only diagnostic tool.

CER-S analysis provides indications for evaluation of:

  • · Patients with rhythm disturbances (cardiac arrhythmias),
  • · Patients with transient myocardial ischemia,
  • · Patients with pacemaker (only if pacing detection is available from the input recording),
  • · Patients needing HRV evaluation,
  • · Newborn patients limited to ORS detection.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary
required by 21 CFR 807.92
(as required by 21 CFR 807.92)
----------------------------------
SubmitterCardioCalm SrlCorso Martiri della Libertà, nr. 4025018 Montichiari (BS)
Telephone+ 39 030 9650782
Fax+ 39 030 9650572
Contact PersonFabio BadiliniPresident & Legal Representative
AddressCorso Martiri della Libertà, nr. 40
Telephone+ 39 030 9650782
Fax+ 39 030 9650572
Emailbadilini@cardiocalm.com
Date Prepared17/11/2021
Trade NameCER-S
Common NameContinuous ECG Recording - Suite
Panel CodeCardiovascular/74
Classification NameComputer diagnostic, programmable
ClassClass II
Regulation Number21 CFR 870.1425
Regulation DescriptionProgrammable diagnostic computer
Product CodeDQK
Name of Predicate Device510(k) #Manufacturer
Name of Predicate Device510(k) #Manufacturer
CER-SK193177CardioCalm Srl
Reason forSubmissionIntention to introduce into commercial distribution a device, CER-S, that has undergone significantchanges that could affect the substantial equivalence with the predicate device.
DeviceDescriptionCER-S is a tool, designed to offer a framework for the interaction of different software-modules,providing advanced solutions for Continuous ECG Recording (CER).Different modules provide:ECG Beat detection and classification Analysis of ECG rhythm, arrhythmia detection interactive Viewer and set of tools to perform editing of ECG beats, Rhythm annotationsand noise Windows interactive Continuous ECG Viewer interactive display/management of ECG Templates Holter-like report for analyzed Continuous ECG records, record exportation, in ISHNEformat generation of aECG FDA HL7 XML (v. 2), for the submission of continuous ECGrecording to the FDA ECG Warehouse. Note: The automatic analysis is limited to data acquired from electrodes with conductive paste/gel(dry and dry/metal electrodes are not intended to be used) placed on standard location incompatible formats from any device used for the arrhythmia diagnostics such as Holter, eventrecorder, 12-lead ambulatory ECG devices when assessment of the rhythm is necessary.Moreover, CER-S allows the automatic analysis for the following patch location: two-electrode patches positioned in the left upper chest area at a roughly inclined angle three-electrode triangular shape patches positioned on the patient's left upper chest areabelow the 1st rib, at an inclined angle three-electrode T-shaped patches positioned in the center-thoracic position between theupper part of the chest (manubrium) and the sternum In all cases, patch placement must strictly follow the indication provided by the manufacturer of thecertified-patch device.
Indications foruseCER-S is a stand-alone software medical device intended to analyze, edit, review and report digitalcontinuous ECG recordings.CER-S is intended for use in clinical and hospital settings only by qualified medical personnel oradequately trained professional personnel working under the supervision and responsibility of a

{4}------------------------------------------------

Traditional 510(k) Premarket Notification CER-S

clinician. Users must undergo a thorough software training before using the medical device.The device provides the clinician an additional tool to aid in the interpretation of ECG data fordiagnosis of heart rhythm disorders in adult and pediatric patients.Note: This device does not provide an automated interpretation and is not intended for use as theonly diagnostic tool.CER-S analysis provides indications for evaluation of:Patients with rhythm disturbances (cardiac arrhythmias), Patients with transient myocardial ischemia, Patients with pacemaker (only if pacing detection is available from the input recording), Patients needing HRV evaluation, Newborn patients limited to QRS detection.
TechnologicalCharacteristicsand SubstantialEquivalenceThe fundamental scientific principles and technological characteristic, including the indications foruse and general design are equivalent to the predicate device, as summarized hereafter:Software type, Holter Analyzer, working in Windows Operative System, as the predicate; Patient population, adults and pediatrics; User groups: qualified medical personnel or adequately trained professional personnelworking under the supervision and responsibility of a clinician., equivalent to predicatedevice; Indications for use, equivalent to the predicate; Environmental use, equivalent to the predicate. The technological characteristics of the subject device and the predicate device are substantiallyequivalent to the predicate.As for the predicate device described below, the Subject device is a PC based SaMD, working inWindows Operative System.The following significant differences exist between the subject and predicate devices:The subject device improves the performance in beat detection in recordings with lowamplitude and with noise; The subject device improves the performance of the beat measures in recording withhigher heart rate; The subject device allows HRV measurements in the frequency domain and SinusTachycardia detection; Further minor changes in subject device are developed to enhance the usability of the softwareand the graphic interface of some its components, which provides a smooth and faster editing ofautomatically detected arrhythmic events.
PerformanceDataNon-clinical tests have been conducted on the Subject Device:ECG performance testing, according to:ANSI/AAMI EC57:1998/(R)2003 and IEC 60601-2-47:2012 (RecognitionNumber 3-155) IEC 60601-2-25:2011 (Recognition Number 3-105) Software development life cycle, according to:IEC 62304:2006+A1:2015 (Recognition Number 13-79) Usability engineering process, according to:IEC 62366-1:2015+AMD1:2020 (Recognition Number 5-129) Any potential hazards have been evaluated and controlled through Risk Managementactivities, according to:ISO 14971:2019 (Recognition Number 5-125) The testing demonstrated that CER-S is substantially equivalent to the predicate for the proposedintended use.No clinical tests were performed to demonstrate the substantial equivalence of CER-S.
ConclusionBased on the indications for use, technological characteristics, and comparison to predicate devices, the Subject device CER-S has been shown to be substantially equivalent to the legally marketed predicate devices.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).