Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K250652
    Manufacturer
    Date Cleared
    2025-07-28

    (146 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ECG-AI LEF 12-Lead algorithm is software intended to aid in earlier detection of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to:

    • patients with cardiomyopathies
    • patients who are post-myocardial infarction
    • patients with aortic stenosis
    • patients with chronic atrial fibrillation
    • patients receiving pharmaceutical therapies that are cardiotoxic, and
    • postpartum women.

    The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.

    A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

    The ECG-AI LEF 12-Lead Algorithm should be applied jointly with clinician judgment.

    Device Description

    The ECG-AI LEF 12-Lead algorithm interprets 12-lead ECG voltage times series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds of a single 12-lead ECG acquisition, and within seconds provides likelihood of LVEF (ejection fraction less than or equal to 40%) to third party software. The results are displayed by the third party software on a device such as a smartphone, tablet, or PC. The ECG-AI LEF 12-Lead algorithm was trained to detect Low LVEF using positive and control cohorts, and the detection of Low LVEF in patients is generated using defined conditions and covariates.

    The ECG-AI LEF 12-Lead algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, frontline clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the ECG-AI LEF 12-Lead algorithm to aid in earlier detection of LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

    The software module can be integrated into a client application to be accessed by clinicians and results viewed through an Electronic Medical Record (EMR) system or an ECG Management System (EMS) accessed via a PC, mobile device, or another medical device. In each case, the physician imports 12-lead ECG data in digital format. The tool analyzes the 10 seconds or longer duration of voltage data collected during a standard 12-lead ECG and outputs a binary result of the likelihood of low ejection fraction as an API result.

    AI/ML Overview

    The provided text is a 510(k) clearance letter and summary for the Anumana, Inc. ECG-AI Low Ejection Fraction (LEF) 12-Lead Algorithm ([K250652](https://510k.innolitics.com/search/K250652)). While it describes the device, its intended use, and substantial equivalence to a predicate device, it does not contain the detailed performance study results, acceptance criteria tables, sample sizes, or ground truth establishment methods that would typically be found in the clinical study section of a full 510(k) submission.

    The document discusses a "Predetermined Change Control Plan (PCCP)" which mentions future performance enhancement validation studies, but it doesn't present the specific results of the validation study that led to this clearance ([K250652](https://510k.innolitics.com/search/K250652)). It only states that "The performance characteristics for the ECG-AI LEF 12-Lead algorithm were evaluated through software verification and labeling verification," which refers to non-clinical data.

    Therefore, many of the requested details cannot be extracted from the provided text. I will populate the table and answer the questions based only on the information available in the given document.


    Acceptance Criteria and Device Performance Study (Extracted from provided 510(k) Summary)

    The provided 510(k) summary (K250652) serves as an update to a previously cleared device (K232699). It focuses on expanding compatibility and minor changes, asserting substantial equivalence based on the predicate's performance rather than detailing a new, comprehensive clinical study for this specific submission. The document emphasizes "software verification and labeling verification" as the evaluation methods for performance characteristics for this particular submission, rather than a clinical performance study with specific metrics for acceptance criteria.

    The Predetermined Change Control Plan (PCCP) section alludes to future performance enhancements and their validation, stating: "To be implemented, a modified version must demonstrate improved performance by meeting pre-specified acceptance criteria. These criteria require the new version's sensitivity and specificity point estimates to be greater than or equal to the previous version, with an improvement shown by either an increased point estimate or a tighter confidence interval lower bound for at least one of these metrics." However, these are future criteria for updates, not the current acceptance criteria for the clearance of K250652 based on a new clinical study.

    Therefore, the specific quantitative acceptance criteria and reported device performance for the clinical study supporting the K250652 clearance are not explicitly stated in the provided text. The clearance is largely based on demonstrating substantial equivalence to the predicate (K232699) and software/labeling verification.

    Based on the provided text, the specific details regarding the clinical performance study (including acceptance criteria, reported performance values, sample sizes, expert details, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for the test set) are NOT available.


    1. A table of acceptance criteria and the reported device performance

    As noted above, the provided text does not contain a table of explicit quantitative acceptance criteria or reported device performance metrics (e.g., sensitivity, specificity, AUC) from a clinical study for K250652. The document claims "The performance characteristics for the ECG-AI LEF 12-Lead algorithm were evaluated through software verification and labeling verification" for this submission, indicating that a new, detailed clinical performance study with such metrics was not the basis for this specific clearance. The PCCP section specifies criteria for future updates, but not for this clearance.

    MetricAcceptance CriteriaReported Device Performance
    Quantitative Performance Metrics (e.g., Sensitivity, Specificity, AUC)Not specified in the provided document for this clearance (K250652). The PCCP mentions that future updates must show sensitivity and specificity point estimates $\ge$ previous version, or improved confidence interval.Not specified in the provided document for this clearance (K250652). The clearance is based on substantial equivalence to a predicate and non-clinical verification.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided document.
    • Data Provenance: Not specified in the provided document. The PCCP mentions "multi-center retrospective clinical study" for future validations, but this isn't linked to the original clearance's test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not specified in the provided document.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not specified in the provided document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not specified in the provided document. The current indication is "to aid in earlier detection" and "applied jointly with clinician judgment," which implies human-in-the-loop, but an MRMC study comparing performance with and without AI assistance is not detailed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • The document states: "The ECG-AI LEF 12-Lead algorithm is not intended to be a stand-alone diagnostic device for cardiac conditions," and "should be applied jointly with clinician judgment." This implies the device is not intended for standalone use in practice. However, whether a standalone performance study was conducted to assess its raw diagnostic capability (e.g., area under the curve) is not explicitly stated. The statement "outputs a binary result of the likelihood of low ejection fraction as an API result" suggests a standalone algorithm output, but the FDA's clearance is for an "aid," not a primary diagnostic tool.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The document mentions the device "was trained to detect Low LVEF using positive and control cohorts." For LVEF, the common ground truth is often echocardiography (measuring ejection fraction), but the specific method used for ground truth (e.g., echocardiography, MRI, or a combination/adjudication) is not specified.

    8. The sample size for the training set

    • Not specified in the provided document.

    9. How the ground truth for the training set was established

    • The document states the device "was trained to detect Low LVEF using positive and control cohorts," but it does not describe how the ground truth was established for these training cohorts (e.g., type of diagnostic test, clinical adjudication process).
    Ask a Question

    Ask a specific question about this device

    K Number
    K233160
    Manufacturer
    Date Cleared
    2023-11-25

    (59 days)

    Product Code
    Regulation Number
    870.1425
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeuTrace EP Mapping System v2.1 is indicated for catheter-based cardiac electrophysiological (EP) procedures. The NeuTrace EP Mapping System v2.1 provides information about the electrical activity of the heart and about catheter location during the procedure. The system can be used on patients who are eligible for a conventional electrophysiological procedure in the right atrium. The system has no special contraindications.

    Device Description

    The NeuTrace EP Mapping System v2.1 (NeuTrace System) is an advanced 3D electroanatomical mapping (EAM) and analysis system capable of:

    • Displaying catheter location during electrophysiology mapping procedures
    • Displaving 3D images of cardiac structures
    • Displaying cardiac activity signals as waveforms (ECGs and EGMs)
    • Displaying derived voltage and time metric overlays over cardiac models -. including Peak-to-Peak Voltage, Local Activation Time (LAT), Fractionation, and Minimum dV/dt

    The NeuTrace System comprises the following software and hardware components:

    • NeuTrace Workstation
    • Window field generator
    • System control unit
    • System interface unit
    • Interface switches and connection cables
    • NeuTrace Software Application v2.1
    • NeuTrace Streaming Application Software v2.0 0

    The NeuTrace System is used together with compatible recording systems and compatible catheters listed in the device labeling to perform its intended use to support electrophysiology procedures in the right atrium.

    AI/ML Overview

    Here's an analysis of the provided text, extracting information related to the acceptance criteria and the study proving the device meets those criteria:

    Device Name: NeuTrace EP Mapping System v.2.1
    Product Code: DQK
    Regulatory Class: Class II


    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly present a formal "acceptance criteria" table with pre-defined thresholds. Instead, it describes performance metrics that were measured and the results achieved, which implicitly serve as the "met criteria." The "Comments" column indicates whether the observed performance meets or aligns with expectations for substantial equivalence.

    Performance MetricAcceptance Criteria (Implicit)Reported Device PerformanceComments
    Ground-truth AccuracyExpected to be low (e.g., < 1mm) and comparable to predicate/reference devices.Statistically demonstrated as < 1mmMeets implicit criterion. Demonstrates high accuracy essential for EP mapping systems.
    ShiftExpected to be low (e.g., < 1mm).< 1mmMeets implicit criterion. Indicates stability of localization.
    DriftExpected to be low (e.g., < 2mm).< 2mmMeets implicit criterion. Indicates stability of localization over time.
    3D Geometry GenerationEquivalent generation of 3D geometries compared with the predicate (CARTO 3 System).Equivalent generation of 3D geometries compared with CARTO 3 SystemDemonstrated equivalence to the predicate, indicating comparable functionality for creating anatomical models.
    Map Generation (Voltage, LAT, Fractionation)Equivalent performance compared with the reference (EnSite X System) for relevant maps.Equivalent accuracy and performance compared with EnSite X System (using FlexAbility™ Ablation Cathether, Sensor Enabled™) for peak-to-peak voltage maps, Local Activation Time (LAT) maps, and fractionation maps.Demonstrated equivalence to the reference device for key electrophysiological maps, ensuring comparable diagnostic information. Note: NeuTrace does not include impedance maps, which were present in reference devices, but this is stated as not required for their intended use.
    System Performance (Overall)Meets all specifications and user requirements. Operates as intended for catheter-based cardiac EP procedures.Passed all testing and met all design specifications and user requirements.Overall system performance verified through comprehensive testing.
    Software Verification & ValidationCompliant with software development standards and functional requirements.PerformedAssures software quality and reliability.
    Hardware VerificationMeets specifications for shift, drift, and intrinsic time delay.Includes 4-hour shift and drift testing and intrinsic time delay testing.Ensures hardware robustness and accurate timing.
    Cybersecurity Risk Management & TestingCompliant with cybersecurity standards for medical devices.PerformedAddresses potential cybersecurity vulnerabilities.
    EMC/EMI TestingCompliant with IEC 60601-1 and IEC 60601-1-2 standards.Performed per IEC 60601-1 and IEC 60601-1-2Ensures electromagnetic compatibility and safety.
    NeuTrace-EnSite accuracy and equivalency testingAccuracy and performance equivalent to EnSite X System.Performed; demonstrated equivalent accuracy and performance.Confirms performance parity with a key reference device.
    NeuTrace-CARTO geometry and mapping equivalency analysesGeometry and mapping generation equivalent to CARTO 3 System.Performed; demonstrated equivalent generation of 3D geometries and maps.Confirms performance parity with the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact sample size (e.g., number of cases, number of data points) used for the bench testing or the GLP animal study. It only states that a GLP animal study was performed.
    • Data Provenance:
      • Country of Origin: Not specified.
      • Retrospective or Prospective: Not specified, but a "GLP animal study" implies prospective data collection under Good Laboratory Practice principles. Bench testing is inherently prospective.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The document does not provide any information on the number of experts used or their qualifications for establishing ground truth, especially for the animal study.
    • The "ground-truth accuracy" results (e.g., < 1mm) suggest that a highly precise physical measurement system was used as the ground truth, rather than human expert interpretation of images or data.

    4. Adjudication Method for the Test Set

    • The document does not describe any adjudication method (e.g., 2+1, 3+1 consensus) for the test set. Given the context of a 3D electroanatomical mapping system and the nature of the performance metrics (accuracy, shift, drift, geometric equivalence), the ground truth seems to be derived from physical measurements or direct comparisons to established systems/methods rather than interpretations requiring human consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done involving human readers improving with AI vs. without AI assistance.
    • This device is an EP mapping system, providing objective electrical and anatomical data, not an AI-assisted diagnostic imaging interpretation tool that would typically undergo MRMC studies. Its function is to provide information for the clinician, not to make a diagnosis or interpretation in place of one.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • The performance data presented (accuracy, shift, drift, geometry, and map generation) are inherently standalone performance of the system's ability to localize and generate data. The system itself generates the maps and measurements; it's not a human-in-the-loop scenario where a human is assisted by AI to perform a task. The NeuTrace system provides the foundational spatial and electrical data directly.

    7. Type of Ground Truth Used

    • The type of ground truth used appears to be based on highly precise physical measurements and direct comparisons to established, validated predicate/reference systems.
      • For accuracy, shift, and drift: Likely precise physical measurement systems (e.g., optical tracking, high-precision fiducial markers) were used in conjunction with the device's own measurements.
      • For 3D geometry and map generation: The ground truth was established by direct comparison to the performance of the predicate (CARTO 3 System) and reference (EnSite X System) devices, which are already considered established and validated standards in the field.

    8. Sample Size for the Training Set

    • The document does not specify any sample size for a training set. This makes sense given that the NeuTrace EP Mapping System v.2.1 is described as an "advanced 3D electroanatomical mapping (EAM) and analysis system" that relies on "Magnetic-based localization" and "derived voltage and time metric overlays." This suggests an algorithmic or physics-based system, rather than a machine learning/AI system that requires a dedicated training set to learn from data.

    9. How Ground Truth for the Training Set Was Established

    • Since no training set is mentioned (implying a non-ML/AI driven system), there is no information provided on how ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232699
    Manufacturer
    Date Cleared
    2023-09-28

    (23 days)

    Product Code
    Regulation Number
    870.2380
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Anumana Low Ejection Fraction AI-ECG Algorithm is software intended to aid in screening for Left Ventricular Ejection Fraction (LVEF) less than or equal to 40% in adults at risk for heart failure. This population includes, but is not limited to:
    · patients with cardiomyopathies

    • patients who are post-myocardial infarction
    • · patients with aortic stenosis
    • · patients with chronic atrial fibrillation
    • · patients receiving pharmaceutical therapies that are cardiotoxic, and
      • postpartum women.

    Anumana Low Ejection Fraction Al-ECG Algorthm is not intended to be a stand-alone diagnostic device for cardiac conditions, should not be used for monitoring of patients, and should not be used on ECGs with a paced rhythm.

    A positive result may suggest the need for further clinical evaluation in order to establish a diagnosis of Left Ventricular Ejection Fraction (LVEF) less than or equal to 40%. Additionally, if the patient is at high risk for the cardiac condition, a negative result should not rule out further non-invasive evaluation.

    The Anumana Low Ejection Fraction AI-ECG Algorithm should be applied jointly with clinician judgment.

    Device Description

    The Low Ejection Fraction AI-ECG Algorithm interprets 12-lead ECG voltage times series data using an artificial intelligence-based algorithm. The device analyzes 10 seconds of a single 12lead ECG acquisition, and within seconds provides a prediction of likelihood of LVEF (ejection fraction less than or equal to 40%) to third party software. The results are displayed by the third-party software on a device such as a smartphone, tablet, or PC. The Low Ejection Fraction AI-ECG Algorithm was trained to predict Low LVEF using positive and control cohorts, and the prediction of Low LVEF in patients is generated using defined conditions and covariates. The Low Ejection Fraction AI-ECG Algorithm device is intended to address the unmet need for a point-of-care screen for LVEF less than or equal to 40% and is expected to be used by cardiologists, front-line clinicians at primary care, urgent care, and emergency care settings, where cardiac imaging may not be available or may be difficult or unreliable for clinicians to operate. Clinicians will use the Low Eiection Fraction AI-ECG Algorithm to aid in screening for LVEF less than or equal to 40% and making a decision for further cardiac evaluation.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided FDA 510(k) clearance letter for the Low Ejection Fraction AI-ECG Algorithm:


    Low Ejection Fraction AI-ECG Algorithm: Acceptance Criteria and Performance Study

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance (95% CI)
    Sensitivity80% or higher84.5% (82.2% to 86.6%)
    Specificity80% or higher83.6% (82.9% to 84.2%)
    Positive Predictive Value (PPV)Not specified (derived metric)30.5% (28.8% to 32.1%)
    Negative Predictive Value (NPV)Not specified (derived metric)98.4% (98.2% to 98.7%)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The clinical validation study included 16,000 patient records initially, though 2,040 records were excluded due to quality checks, resulting in a final analysis sample of 13,960 patient-ECG pairs.
    • Data Provenance: The data was retrospective, collected from 4 health systems across the United States.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts or their qualifications used to establish the ground truth for the clinical validation test set. The ground truth (LVEF <= 40% or > 40%) was derived from transthoracic echocardiogram (TTE) measurements. While TTE interpretation requires expertise, the document doesn't detail the method of expert review or consensus for these TTE results themselves for the test set.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method (e.g., 2+1, 3+1) for the ground truth for the test set. The ground truth was established by TTE measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the AI algorithm against a ground truth without human readers in the loop.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, a standalone performance study was done. The reported sensitivity and specificity values are for the algorithm's performance alone in detecting low LVEF.

    7. The Type of Ground Truth Used

    The type of ground truth used for both training and validation was objective clinical measurements from Transthoracic Echocardiogram (TTE), specifically the Left Ventricular Ejection Fraction (LVEF) measurement. An LVEF of $\le$ 40% was defined as the disease cohort, and > 40% as the control cohort.

    8. The Sample Size for the Training Set

    The training set for the algorithm development consisted of 93,722 patients with an ECG and TTE performed within a 2-week interval. These were split into:

    • Training dataset: 50% of the 93,722 patients.
    • Tuning dataset: 20% of the 93,722 patients.
    • Set-aside testing dataset: 30% of the 93,722 patients (used for internal validation during development, distinct from the independent clinical validation study).

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the training set was established using LVEF measurements obtained from transthoracic echocardiograms (TTE). Specifically, for each patient, the LVEF measurement from the earliest TTE within a 2-week interval of an ECG was paired with the closest ECG recording. LVEF $\le$ 40% defined the disease cohort, and LVEF > 40% defined the control cohort. This data was identified from a research-use authorized clinical database from Mayo Clinic.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1