Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K212584
    Date Cleared
    2022-05-28

    (285 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Nitrile Examination Gloves Powder Free is disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Nitrile Examination Gloves Powder Free are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250). The gloves are made from acrylonitrilebutadiene copolymer dispersion. These gloves are blue in color and are powder free.

    AI/ML Overview

    The provided text is a 510(k) summary for "Blue Nitrile Examination Gloves Powder Free". It details the device's characteristics, intended use, and a comparison to a predicate device. The document explicitly states that clinical testing is not applicable and that clinical data is not needed for gloves or for most devices cleared by the 510(K) process.

    Therefore, there is no information in the provided text regarding a study that proves the device meets acceptance criteria related to AI or human-in-the-loop performance, as this is a physical medical device (examination gloves), not an AI-powered diagnostic or assistive tool.

    However, I can extract the acceptance criteria and reported performance for the non-clinical tests conducted on the device, which are relevant for demonstrating its substantial equivalence to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    CharacteristicStandardAcceptance CriteriaReported Device Performance
    DimensionsASTM D6319-2019
    LengthMin 230 mm for all sizesX-Small: 303 mmSmall: 304 mmMedium: 305mmLarge: 306mmX-Large: 307mm
    Width (for specific sizes)X-small: 70+/-10 mmSmall: 80+/-10 mmMedium: 95+/-10mmLarge: 110+/-10 mmX-Large: 120+/-10 mmX-small: 74 mmSmall: 84 mmMedium: 94 mmLarge: 104 mmX-Large: 115 mm
    Physical PropertiesASTM D6319-2019
    Before Aging:
    Tensile Strength14 Mpa Min for all sizesX-Small: 18.44 MpaSmall: 18.49 MpaMedium: 18.56 MpaLarge: 18.62 MpaX-Large: 18.64 Mpa
    Ultimate Elongation500% Min for all sizesX-Small: 683%Small: 687%Medium: 692%Large: 695%X-Large: 698%
    After Aging:
    Tensile Strength14 Mpa Min for all sizesX-Small: 17.49 MpaSmall: 17.53 MpaMedium: 17.65 MpaLarge: 17.71 MpaX-Large: 17.76 Mpa
    Ultimate Elongation400% Min for all sizesX-Small: 656%Small: 660%Medium: 663%Large: 666%X-Large: 669%
    ThicknessASTM D6319-2019Palm 0.05 mm minFinger 0.05 mm minPalm: 0.10 mm for all sizesFinger: 0.12 mm for all sizes
    Water TightnessASTM D5151-2019AQL-1.5Passes AQL 1.5
    Powder ResidueASTM D6124-06 (Reapproved 2017)2 mg/glove MaxX-Small: 0.21 mg/gloveSmall: 0.21 mg/gloveMedium: 0.22 mg/gloveLarge: 0.22 mg/gloveX-Large: 0.22 mg/glove
    Biocompatibility
    Primary Skin IrritationISO 10993-10Under the condition of study not an irritantUnder the condition of study not an irritant
    Dermal SensitizationISO 10993-10Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    In vitro CytotoxicityISO 10993-5:2009Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic.
    Acute Systemic ToxicityISO 10993-11:2017Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
    Material Mediated PyrogenicityISO 10993-11:2017 / USP 41 <151>Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.

    Regarding the other requested information (2-9), as a physical medical device (gloves) and not an AI/software device, these questions are not applicable to the context of the provided document.

    • 2. Sample size used for the test set and the data provenance: Not applicable. (These are physical tests on glove samples, not a test set for AI performance).
    • 3. Number of experts and qualifications: Not applicable. (These are standardized physical and biological tests).
    • 4. Adjudication method: Not applicable. (These are standardized physical and biological tests).
    • 5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not applicable. This is for AI-assisted image interpretation or similar.
    • 6. Standalone (algorithm only without human-in-the-loop performance): Not applicable. This is for AI algorithms.
    • 7. Type of ground truth used: Not applicable in the AI sense. The "ground truth" for these tests comes from established industry standards (ASTM, ISO, USP) for physical and chemical properties and biological responses.
    • 8. Sample size for the training set: Not applicable. Applies to AI/machine learning.
    • 9. How the ground truth for the training set was established: Not applicable. Applies to AI/machine learning.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212597
    Date Cleared
    2022-05-28

    (285 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Amazing+ Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Amazing Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device: "Amazing+ Latex Examination Powder Free Gloves." This document describes the device and demonstrates its substantial equivalence to a legally marketed predicate device.

    However, the request asks for information related to a study that proves a device meets acceptance criteria, specifically in the context of an AI-powered medical device (e.g., sample size for test set, data provenance, expert adjudication, MRMC study, standalone performance, ground truth establishment for training and test sets).

    The provided 510(k) Summary does not describe an AI medical device. It describes physical specifications and performance of latex examination gloves. Therefore, many of the requested fields are not applicable to this type of device.

    Despite this, I will extract and present the relevant information that is available in the document, and clearly state when information is not applicable to this physical device.


    Device: Amazing+ Latex Examination Powder Free Gloves
    K Number: K212597
    Predicate Device: JR MEDIC Blue Latex Examination Powder Free Gloves (K192329)


    1. A table of acceptance criteria and the reported device performance

    CharacteristicStandard/Test MethodAcceptance CriteriaReported Device PerformanceComparison to Predicate
    Dimensions - LengthASTM D3578-2019Min 230 mm for all sizesX-Small 305 mm, Small 306 mm, Medium 307 mm, Large 308 mm, X-Large 310 mmSimilar
    Dimensions - WidthASTM D3578-2019X-Small 70+/-10 mm, Small 80+/-10mm, Medium 95+/-10 mm, Large 111+/-10 mm, X-Large 115+/-10 mmX-Small 76 mm, Small 85 mm, Medium 96 mm, Large 106 mm, X-Large 116 mmSimilar
    Dimensions - ThicknessASTM D3578-2019Palm > 0.08 mm min, Finger > 0.08 mm min for all sizesPalm: 0.16mm (all sizes), Finger: 0.22mm (all sizes)Similar
    Physical Properties - Tensile StrengthASTM D3578-2019Before Ageing: > 18 Mpa Min for all sizes. After Ageing: > 14 Mpa Min for all sizes.Before Ageing: X-Small 22.07 Mpa, Small 22.15 Mpa, Medium 22.22 Mpa, Large 22.30 Mpa, X-Large 22.32 Mpa. After Ageing: X-Small 18.49 Mpa, Small 18.56 Mpa, Medium 18.67 Mpa, Large 18.74 Mpa, X-Large 18.76 Mpa.Similar
    Physical Properties - Ultimate ElongationASTM D3578-2019Before Ageing: > 650% Min for all sizes. After Ageing: > 500% Min for all sizes.Before Ageing: X-Small 858%, Small 869%, Medium 874%, Large 880%, X-Large 882%. After Ageing: X-Small 841%, Small 848%, Medium 854%, Large 860%, X-Large 862%.Similar
    Physical Properties - Stress at 500% ElongationASTM D3578-2019Before Ageing: 5.5 Mpa Max for all sizes.Before Ageing: X-Small 5.1 Mpa, Small 5.1 Mpa, Medium 5.2 Mpa, Large 5.2 Mpa, X-Large 5.2 Mpa.N/A
    Water Tightness (Holes)ASTM D5151-2019AQL 2.5Gloves Pass AQL 1.5Same
    Powder Free ResidueASTM D6214<2 mg/glove MaxX-Small 0.21 mg/glove, Small 0.21 mg/glove, Medium 0.22 mg/glove, Large 0.22 mg/glove, X-Large 0.22 mg/gloveSimilar
    Biocompatibility - Primary Skin IrritationISO 10993-10:2010(E)Under the condition of study, not an irritantUnder the condition of study, not an irritantSame
    Biocompatibility - Dermal SensitizationISO 10993-10:2010(E)Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
    Biocompatibility - In vitro CytotoxicityISO 10993-5:2009(E)Under the conditions of the study, cytotoxicUnder the conditions of the study, cytotoxic (Note: This might seem counter-intuitive, but for certain test types, "cytotoxic" as a test result may simply indicate a reaction within the test conditions, not necessarily a failure from a biocompatibility standpoint for its intended use, especially if it's consistent with a cleared predicate. The key is that it's "Same" as the predicate's finding.)Same
    Biocompatibility - Material mediated PyrogenicityISO 10993-11:2017(E) / USP 41<151>Under the conditions of the study non pyrogenicUnder the conditions of the study, non-pyrogenicSame
    Biocompatibility - Acute Systemic Toxicity TestISO 10993-11:2017(E)Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concernSame
    Extractable ProteinASTM D 5712-95 (Re approved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber200 µg/ dm² Max for all sizes43.65 µg/dm² (all sizes)N/A (Not compared to predicate in table, but meets criteria)

    Information Not Applicable to this Device (as it's a physical product, not an AI/software device):

    Given that the device is "Amazing+ Latex Examination Powder Free Gloves," the following sections of your request are not applicable as they pertain to AI/software functionality and clinical studies that are not performed for basic Class I examination gloves:

    • 2. Sample size used for the test set and the data provenance: This device is a physical product. Testing involves laboratory measurements on physical samples (e.g., gloves from a production lot) to assess properties like dimensions, strength, and integrity, not dataset sizes. The provenance is from "Amazing Rubber Products Pvt. Ltd." in Cochin, Kerala, India. The tests are non-clinical, likely performed according to ISO/ASTM standards.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical properties is established by standardized measurement methods (e.g., ASTM standards), not expert opinion or consensus.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Physical measurements do not require adjudication.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI device.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. Ground truth for device performance is based on the specified ASTM and ISO standards for physical and chemical properties and biocompatibility testing.
    • 8. The sample size for the training set: Not applicable. There is no "training set" for physical examination gloves; manufacturing process controls ensure consistent product.
    • 9. How the ground truth for the training set was established: Not applicable. See above.

    Summary of Non-Clinical Performance Data (Reiterated from document):

    The performance test data of the non-clinical tests meet the following standards:

    • ASTM D 3578-2019 Standard Specification for Rubber Examination Gloves
    • ASTM D 5151-2019 Standard Test Method for Detection of Holes in Medical Gloves
    • ASTM D 6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
    • ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber

    Clinical Testing Summary:

    • "This section is not applicable because clinical data not needed for gloves." (As stated in the document).
    Ask a Question

    Ask a specific question about this device

    K Number
    K212596
    Date Cleared
    2022-03-01

    (197 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Sterile Latex Surgical Gloves Powder Free is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

    Device Description

    The proposed device, Sterile Latex Surgical Gloves Powder Free with protein content labeling claim of 50 µg/ dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Sterile Latex Surgical Gloves Powder Free, and variants of different sizes. All variants share the same natural color (No Color is added). The proposed device is sterilized using Gamma and ETO method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 5years.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves the device meets those criteria for Sterile Latex Surgical Gloves Powder Free (K212596), intended to be worn in operating rooms. The study is a non-clinical performance (bench testing) study demonstrating substantial equivalence to a predicate device (K192328).

    Here's the breakdown of the information requested, based on the provided document:


    Acceptance Criteria and Device Performance Study for Sterile Latex Surgical Gloves Powder Free (K212596)

    The study performed is a non-clinical performance (bench testing) study, as explicitly stated under "SUMMARY OF NONCLINICAL PERFORMANCE DATA" and "SUMMARY OF CLINICAL PERFORMANCE DATA: Clinical data was not required for this submission." This type of study focuses on technical and physical characteristics rather than performance in a clinical setting with human participants.

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria are generally based on ASTM D3577-2019, Standard Specification for Rubber Surgical Gloves, and other specific standards for each test.

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance (Average Value of Subject Device)Met Acceptance Criteria?
    DimensionsASTM D3577-2019
    Length (All sizes)Min 265 mmVaries by size (e.g., Size 7: 486 mm, Size 9: 488 mm)Yes
    Width (e.g., Size 7)89 ± 6 mm91 mmYes
    Finger Thickness (All sizes)Min 0.10 mm0.33 mmYes
    Palm Thickness (All sizes)Min 0.10 mm0.21 mmYes
    Cuff Thickness (All sizes)Min 0.10 mm0.16 mmYes
    Tensile StrengthASTM D3577-2019
    Before aging (All sizes)24 MPa minimum28.61 MPaYes
    After aging (All sizes)18 MPa minimum24.12 MPaYes
    Ultimate ElongationASTM D3577-2019
    Before aging (All sizes)750% minimum871%Yes
    After aging (All sizes)560% minimum732%Yes
    Stress at 500% before aging5.5 MPa Max5.1 MPaYes
    Pinhole AQLASTM D3577-2019, ASTM D5151-2019
    Before aging (All sizes)Max 1.51.0Yes
    After aging (All sizes)Max 1.51.0Yes
    Freedom from Holes AQLAQL 1.0 (ASTM D3577-2019, ASTM D5151-2019)1.0Yes
    Powder Residue< 2 mg/Glove; (ASTM D3577-2019, ASTM D6124-06)0.38 mg/GloveYes
    Protein Content< 50 µg/dm² (ASTM D3577-2019, ASTM D5712-15)46.40 µg/dm²Yes
    Biocompatibility
    Skin Irritation & SensitizationNon-irritant and Non-Sensitizer (ISO 10993-10)Non-irritant and Non-SensitizerYes
    In vitro CytotoxicityNon-cytotoxic (ISO 10993-5)Cytotoxic (Note: While reported as 'cytotoxic', this appears to be a misstatement or misinterpretation in the table, as the comparison table above it shows 'non-cytotoxic' for the predicate. However, in the detailed table, it is reported as 'cytotoxic', matching what is written for the predicate. This requires clarification to understand why a 'cytotoxic' result is deemed acceptable if the acceptance criteria is 'non-cytotoxic'.)Check with manufacturer/FDA guidance
    Material Mediated PyrogenicityNon-pyrogenic (ISO 10993-11, USP 41 <151>)Non-pyrogenicYes
    Systemic ToxicityDevice extracts do not pose a systemic toxicity concern (ISO 10993-11)Device extracts do not pose a systemic toxicity concernYes
    Bacterial Endotoxin<20 EU/pair of gloves (USP 42 <85>)<20 EU/pair of glovesYes

    Note on Cytotoxicity: The specific call out for "in vitro cytotoxicity" in the detailed results table (page 9) states "Under the conditions of the study cytotoxic." However, the acceptance criteria listed immediately above it is "Under the conditions of study non cytotoxic." And in the "Technological Characteristics Comparison" table (page 7), both the subject and predicate devices are listed as "cytotoxic" with an acceptance criterion of "cytotoxic" but then the standard states "non-cytotoxic." This is a discrepancy that would typically require clarification in a regulatory submission. Assuming the manufacturer has demonstrated that this level of cytotoxicity is acceptable for the intended use and does not pose a safety risk, or that the "acceptance criteria" presented in the detailed result is actually just reporting the finding, it passed. However, simply stating "cytotoxic" when the standard implies "non-cytotoxic" needs careful review. For the purpose of this summary, it's listed as reported.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not explicitly state the exact sample sizes (number of gloves) used for each bench test. Standards like ASTM D3577 typically specify sampling plans (e.g., AQL levels for defect testing).
    • Data Provenance: The tests were conducted internally by the manufacturer or by contracted labs. The manufacturer is Amazing Rubber Products Pvt. Ltd. located in Kerala, India. The data is retrospective in the sense that it was collected as part of the device development and verification process to support the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable as the study is a bench test verification of physical and chemical properties against established industry standards (e.g., ASTM, ISO, USP). Ground truth in this context is defined by these measurable standards and the associated testing methodologies, not by expert interpretation of, for example, medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not applicable as the study is a bench test verification. Adjudication methods are typically used in clinical studies, especially those involving human interpretation (e.g., radiology reads) to resolve discrepancies.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic AI devices where human performance is being evaluated or improved with AI assistance. The device in question is a surgical glove, which is a physical product, not a diagnostic tool requiring human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, this is not applicable. The device is a physical product (surgical glove), not an algorithm or AI system. Its performance is assessed through bench testing against established performance standards.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The "ground truth" for this device is defined by established international and national standards and specifications for surgical gloves (e.g., ASTM D3577-2019, ISO 10993 series, USP monographs). These standards specify the acceptable physical, chemical, and biological properties through defined test methods.

    8. The sample size for the training set

    • Not applicable. This question pertains to machine learning models. The device is a physical product, not an AI/ML model that requires training.

    9. How the ground truth for the training set was established

    • Not applicable. As above, this pertains to machine learning models.
    Ask a Question

    Ask a specific question about this device

    K Number
    K212586
    Device Name
    Amazing+
    Date Cleared
    2022-02-17

    (185 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Nitrile Patient Examination Gloves Powder Free Blue Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Sterile Nitrile Patient Examination Gloves Powder Free Blue Color are Class I patient examination gloves bearing the product code Nitrile - LZA (21CFR880.6250).The gloves are made from acrylonitrile-butadiene copolymer dispersion. These gloves are blue in color and are powder free. Both Gamma & ETO sterilized gloves are available in X-Small, Small, Medium, Large & X-Large sizes.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device: "Sterile Nitrile Patient Examination Gloves Powder Free Blue Color". This summary focuses on demonstrating substantial equivalence to a predicate device, rather than proving that the device meets acceptance criteria through a specific study with AI or human readers. Therefore, many of the requested items (2, 3, 4, 5, 6, 7, 8, 9) are not applicable or cannot be extracted from this document as it does not describe an AI/software study.

    However, the document does contain "Acceptance Criteria" and "Result" for various non-clinical performance and biocompatibility tests.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test Method (Characteristic)PurposeAcceptance CriteriaReported Device Performance (Result)
    ASTM D6319-2019 (Length)To determine the length of the glovesMin 220 mm for X-small and Small sizes; Min 230 mm for other sizesX-Small: 303 mm; Small: 304 mm; Medium: 305mm; Large: 306mm; X-Large: 307mm
    ASTM D6319-2019 (Width)To determine the width of the glovesX-small: 70±10 mm; Small: 80±10 mm; Medium: 95±10 mm; Large: 110±10 mm; X-Large: 120±10 mmX-small: 74 mm; Small: 84 mm; Medium: 94 mm; Large: 104 mm; X-Large: 115 mm
    ASTM D6319-2019 (Thickness)To determine the thickness of the glovesPalm 0.05 mm min; Finger 0.05 mm min for all sizesAll sizes: Palm 0.10mm; Finger 0.12mm
    ASTM D6319-2019 (Tensile Strength - Before Ageing)To Determine the physical properties- Tensile strengthBefore Ageing Tensile Strength 14Mpa Min for all sizesX-Small: 18.44Mpa; Small: 18.49Mpa; Medium: 18.56Mpa; Large: 18.62Mpa; X-Large: 18.64Mpa
    ASTM D6319-2019 (Tensile Strength - After Ageing)To Determine the physical properties- Tensile strengthAfter Ageing Tensile Strength 14Mpa Min for all sizesX-Small: 17.49Mpa; Small: 17.53Mpa; Medium: 17.65Mpa; Large: 17.71Mpa; X-Large: 17.76Mpa
    ASTM D6319-2019 (Ultimate Elongation - Before Ageing)To Determine the physical properties- Ultimate ElongationBefore Ageing Ultimate Elongation 500% Min for all sizesX-Small: 683%; Small: 687%; Medium: 692%; Large: 695%; X-Large: 698%
    ASTM D6319-2019 (Ultimate Elongation - After Ageing)To Determine the physical properties- Ultimate ElongationAfter Ageing Ultimate Elongation 400% Min for all sizesX-Small: 656%; Small: 660%; Medium: 663%; Large: 666%; X-Large: 669%
    ASTM D5151-2019 (Water Tight / Holes)To determine the holes in the glovesAQL 2.5Passes AQL 1.5
    ASTM D6124-06 (Reapproved 2017) (Residual Powder)To determine the residual powder in the gloves2 Mg/Glove MaxAll sizes: 0.21mg/glove (X-small, Small), 0.22 mg/glove (Medium, Large, X-Large)
    ISO 10993-10 (Skin Irritation)To determine the potential of the material under test to produce dermal irritation in RabbitsUnder the condition of study not an irritantUnder the condition of study not an irritant
    ISO 10993-10 (Skin Sensitization)To determine the skin sensitization potential of the material both in terms of induction and elicitation in Guinea Pig.Under the conditions of the study not a sensitizerUnder the conditions of the study not a sensitizer
    ISO 10993-5:2009 (In vitro cytotoxicity)To evaluate the in vitro cytotoxic potential of the test item (both inner and outer surface) Extracts in L-929 mouse fibroblasts cells using elution method.Under the conditions of study non cytotoxicUnder the conditions of the study cytotoxic. (Note: This result does not meet the acceptance criteria. The document states "Under the conditions of the study cytotoxic" against an acceptance criteria of "Under the conditions of study non cytotoxic". This could be a typo in the provided text as later in the conclusion it says it performs "as well as or better than" in some cases. However, strictly based on the table, it appears unmet.)
    ISO 10993-11:2017 (Acute Systemic Toxicity)To determine the acute systemic toxicity potential of the test item extracts (both inside and outer surfaces) in Swiss Albino mice.Under the conditions of study the device extracts do not pose a systemic toxicity concernUnder the conditions of study the device extracts do not pose a systemic toxicity concern
    Material Mediated Pyrogenicity ISO 10993-11:2017(E) / USP 41<151>To determine the pyrogenic potential of the test item extract following intravenous injection in New Zealand white Rabbits.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.Under the conditions of the study, the device did not demonstrate a material mediated pyrogenicity response.
    Bacterial Endotoxin test USP 42<85>To determine the bacterial Endotoxin limit in the gloveNMT 20 EU/pair of gloves<20 EU/pair of gloves

    2. Sample size used for the test set and the data provenance: Not applicable. This document describes physical and biocompatibility testing, not an AI/software study with a test set of data. The "test sets" here would refer to the number of gloves tested per batch or samples for biocompatibility - which is not specified in detail.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" here is established through standardized laboratory testing methods (e.g., ASTM D6319, ISO 10993 series) and their respective criteria. For instance, for physical properties, it's objective measurements based on the standard. For biocompatibility, it's the observed biological response in animal models or in vitro assays compared to established acceptable limits.

    8. The sample size for the training set: Not applicable.

    9. How the ground truth for the training set was established: Not applicable.

    Summary of the study that proves the device meets the acceptance criteria:

    The device "Sterile Nitrile Patient Examination Gloves Powder Free Blue Color" underwent a series of non-clinical performance and biocompatibility tests according to recognized international and national standards (ASTM D6319, ASTM D5151, ASTM D6124, ISO 10993-10, ISO 10993-5, ISO 10993-11, USP 41<151>, USP 42<85>). The purpose of these tests was to demonstrate the device's physical properties (length, width, thickness, tensile strength, elongation), integrity (water tightness/holes), chemical residues (powder content), and biological safety (skin irritation, skin sensitization, cytotoxicity, systemic toxicity, pyrogenicity, bacterial endotoxin).

    The results of these tests, as presented in the "Device Performance" columns, were compared against the "Acceptance Criteria" specified by the respective standards. The document states that the device met these criteria for most tests, demonstrating its safety and effectiveness. Notably, for "In vitro cytotoxicity," the device's result was "cytotoxic" while the acceptance criteria was "non cytotoxic." Despite this, the overall conclusion states that the device is "as safe, as effective, and performs as well as or better than the legally marketed predicate device K171093." This suggests either an interpretation allowing for some cytotoxicity in the context of the device's use and predicate equivalence, or a potential inconsistency in the provided information.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1