K Number
K192328

Validate with FDA (Live)

Date Cleared
2020-03-10

(196 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

The proposed device, JR Medic Latex Surgeon's Gloves Sterile Powder Free with protein content labeling claim of 50 µg/ dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, as per standard ASTM D35 77 -09(2015), Standard Specification for Rubber Surgical Gloves. The classification is: Type I - gloves compounded primarily from natural rubber latex. The proposed device is Powder Free Latex Surgeon's Gloves, and variants of different sizes. All variants share the same color, creamy, white. The proposed device is sterilized using Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 3 years.

AI/ML Overview

The provided document describes the acceptance criteria and a study proving that JR MEDIC Latex Surgeon's Gloves Sterile Powder Free meet these criteria.

Here's the breakdown of the information requested:

1. A table of acceptance criteria and the reported device performance

CriteriaAcceptance Criteria (Specification per ASTM D3577-09)Reported Device Performance (Average Value)Compliance
Length (for all sizes)Min 265mm305mm - 306mmYes
Width
Size 676+/-6mm78mmYes
Size 6 1/283+/-6mm85mmYes
Size 789+/-6mm88mmYes
Size 7 1/295+/-6mm97mmYes
Size 8102+/-6mm103mmYes
Size 8 1/2108+/-6mm110mmYes
Size 9114+/-6mm116mmYes
Thickness (for all sizes)
Finger ThicknessMin 0.10mm0.21mmYes
Palm ThicknessMin 0.10mm0.18mmYes
Cuff ThicknessMin 0.10mm0.11mmYes
Tensile Strength (Before aging)24Mpa minimum26.0MpaYes
Tensile Strength (After aging Pinhole AQL)18Mpa minimum22.0MpaYes
Ultimate Elongation (Before aging)750% minimum860%Yes
Ultimate Elongation (After aging Pinhole AQL)560% minimum725%Yes
Stress at 500% (before aging)5.5 MPa Max2.7 MpaYes
Pinhole AQL (Before aging)Max 1.51.0Yes
Pinhole AQL (After aging)Max 1.51.0Yes
Powder content< 2 mg/Glove0.34 mg/GloveYes
Protein Content< 50 µg/dm²43 µg/dm²Yes
Biocompatibility - Irritation & SensitizationNon-irritant and Non-SensitizerNon-irritant and Non-SensitizerYes (Same)
Biocompatibility - Cytotoxicity(Implied: Non-cytotoxic, as the predicate is)CytotoxicNo (Different - but accepted by FDA as part of substantial equivalence)
Biocompatibility - Systemic Toxicity(Implied: No systemic toxicity)No systemic toxicity under the conditions of the testYes
Pyrogen TestNon-pyrogenicNon-pyrogenicYes (Same)
Sterility Assurance Level (SAL)10^-610^-6Yes

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the sample size for each specific test within the non-clinical performance data. It indicates "Average value of subject device" for each criterion, implying that multiple units were tested. The data provenance is from JR Engineering & Medical Technologies (M) SDN.BHD., which is located in Malaysia. The study is a non-clinical bench test study to verify that the device meets design specifications, and therefore, the terms retrospective or prospective don't directly apply in the same way they would for a clinical study with human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the study is a non-clinical bench test for physical and chemical properties of gloves. "Experts" in this context would likely be laboratory technicians or engineers performing standardized tests, rather than medical experts establishing a ground truth for diagnostic purposes. The ground truth is established by the specifications in recognized standards like ASTM and ISO.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as the tests are objective measurements against established standards, not subjective assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is a surgical glove, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable for a surgical glove.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth used for the acceptance criteria is based on recognized international standards and specifications for medical devices, specifically:

  • ASTM D3577-09(15): Standard Specification for Rubber Surgical Gloves.
  • ASTM D 5151-06 (2015): Standard Test Method for Detection of Holes in Medical Gloves.
  • ASTM D6124-06 (2017): Standard Test Method for Residual Powder on Medical Gloves.
  • ASTM D5712-15: Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.
  • ASTM D6499-18: Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products.
  • ASTM F 1929-2015: Standard Test Method for Detecting Seal Leaks in Porous Packaging by Dye Penetration.
  • ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
  • ISO 10993-11:2017(E) Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity and Biological Tests.
  • USP 41 <151> Pyrogen Test.
  • ISO 11137-1-2006/ (R) 2010 - validation of sterilization process.
  • ISO 11137-2:2013, sterilization of health care products - radiation - part 2: establishing the sterilization dose.

8. The sample size for the training set

This is not applicable as this is a non-clinical bench test for a physical device, not a machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

March 10, 2020

JR Engineering & Medical Technologies (M) SDN.BHD. % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Dr, STE 114 Aurora, Illinois 60504

Re: K192328

Trade/Device Name: JR MEDIK Latex Surgeon's Gloves Sterile Powder Free Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: February 12, 2020 Received: February 14, 2020

Dear Manoj Zacharias:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192328

Device Name

JR Medic Latex Surgeon's Gloves Sterile Powder Free

Indications for Use (Describe)

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510K SUMMARY

K192328

as required by: 21 CFR § 807.92

A. APPLICANT INFORMATION

Submitter NameJR Engineering & Medical Technologies (M)SDN.BHD.
Date Submitted24th Nov 2019
AddressLot 8 &10, Jalan Zurah 3 & Lot 1 & 3, Jalan Zurah 3A/1Pusat Perindustrian 2, 44200 Rasa,Hulu Selangor, Selangor Darul Ehsan, Malaysia.
Phone+603-60572081
Fax+603-60572181
E-mailganeshjrmt@gmail.com
Contact PersonMr. Ganesan Subramaniam
DesignationManaging Director
Contact Number+6012 224 6677
Contact Emailganeshjrmt@gmail.com

B. US AGENT & CONTACT PERSON INFORMATION

US agent & contact person nameManoj Zacharias
AddressLiberty Management Group Ltd.2871, Coastal Dr. Aurora, IL-60503, USA
Phone(630) 270-2921
Fax(815) 986-2632
E-mailmanoj@libertymanagement.us

C. DEVICE IDENTIFICATION

Common NameSurgeon's Gloves
Device NameSurgeon's Gloves powder free
Product proprietaryor trade nameJR MEDIC LATEX SURGEON'S GLOVES STERILE POWDER FREE
Classification nameSurgeon's Gloves
Device ClassificationI
Product CodeKGO
Regulation Number21 CFR 878.4460
Review PanelGeneral Hospital

{4}------------------------------------------------

D.PREDICATE DEVICE INFORMATION

DeviceName of device510k Number510K Owner
PredicatedevicePristeen Latex Surgeon's Glovespowder free with protein contentlabeling claim of 50 $\mu g/dm^2$ or lessper glove of extractable proteinK172942Beta Healthcare ProductsPvt.Ltd, Plot No 21B,Cochin Special EconmicZone, Kakkanad, Kerala,India-682037.
ReferencedeviceMedismart+ Latex Surgeon'sGloves powder with protein contentlabeling claim of 50$\mu g/dm^2$ or less per glove ofextractable proteinK151114St.Marys Rubbers Pvt.Ltd,Koovappally P.O,Kanjirappally, KottayamDistrict, Kerala State,India-686518

E. DESCRIPTION OF THE DEVICE

The proposed device, JR Medic Latex Surgeon's Gloves Sterile Powder Free with protein content labeling claim of 50 µg/ dm² or less per glove of extractable protein is a sterile and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.

The proposed device is made of natural rubber latex, as per standard ASTM D35 77 -09(2015), Standard Specification for Rubber Surgical Gloves.

The classification is: Type I - gloves compounded primarily from natural rubber latex.

The proposed device is Powder Free Latex Surgeon's Gloves, and variants of different sizes. All variants share the same color, creamy, white.

The proposed device is sterilized using Gamma irradiation method to achieve the Sterility Assurance Level (SAL) of 10 and place in a sterility maintained package to ensure a shelf life of 3 years.

F. INDICATION FOR USE

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

G. TECHNOLOGICAL CHARACTERISTIC COMPARISON TABLE

a. General Characteristics Comparison

CharacteristicSubject deviceK192328Predicate deviceK172942comparison
Product CodeKGOKGOSame
Regulation No.21 CFR 878.446021 CFR 878.4460Same
ClassIISame
Intended UseA latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contaminationA latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contaminationSimilar

{5}------------------------------------------------

Powdered orPowder freepowered freepowered freeSame
CompoundingClassificationType I - gloves compoundedprimarily from natural rubberlatexType I - gloves compoundedprimarily from natural rubber latexSame
SterilizationRadiation,SAL- 10-6ETO/as well asRadiation, SAL- 10-6Different
LabelingMeet FDA's labelRequirementsMeet FDA's labelRequirementsSame
Special labelclaimProtein content labeling claim of50µg/dm² or lessProtein content labeling claim of50µg/dm² or lessSame
Type of useOver the counter useOver the counter useSame

b. Technological Characteristics Comparison

Specification and physical properties per ASTMD3577- 09(Reapproved 2015), Standard Specificationfor Rubber Surgical Gloves.SubjectdeviceK192328PredicatedeviceK172942Remarks
Dimensions Length: Min 265 mm300 mm282 mmSimilar
Width
Size6.0(76+/-6mm)78mm78mmSimilar
6.5(83+/-6mm)85mm84mm
7.0(89+/-6mm)88mm91mm
7.5(95+/-6mm)97mm97mm
8.0(102+/-6mm)103mm103mm
8.5(108+/-6mm)110mm109mm
9(114+/-6mm)116mm115mm
Thickness for all sizes
Cuff, Palm, Finger Tip Min 0.10 mmCuff- 0.11mmCuff- 0.13mmSimilar
Palm- 0.18mmPalm-0.16mm
Finger Tip-0.21mmFinger Tip-0.18mm
Physical Properties Before Ageing -for all sizes
Tensile Strength 24Mpa minimum26.0Mpa28.0MpaSimilar
Ultimate Elongation 750% minimum860%920%
Stress at 500% 5.5 MPa Max2.7 Mpa3 Mpa
Physical Properties After Ageing -for all sizes
Tensile Strength 18Mpa minimum22.0Mpa24.0MpaSimilar
Ultimate Elongation 560% minimum725%750%
Freedom from Holes AQL 1.5, AST M D5151-06,(2015), Standard Test Method for Detection of HolesAQL 1.0AQL 1.0Similar

{6}------------------------------------------------

in Medical Gloves
Powder content < 2 mg/Glove,ASTM D6124-06, (2017), Standard Test Method forResidual Powder on Medical Gloves0.34mg/Glove0.3mg/gloveSimilar
Protein Content < 50 $\mu$ g/ dm²,ASTM D5712-15, Standard Test Method for theAnalysis of Aqueous Extractable Protein in NaturalRubber and Its Products Using the Modified LowryMethod;ASTM D6499-18, Standard Test Method for theImmunological Measurement of Antigenic Protein inNatural Rubber and Its Products43 $\mu$ g/ dm²40 $\mu$ g/ dm²Similar
Biocompatibility
ISO 10993-10, Biological evaluation of medicaldevices - Part 10: Tests for irritation and skinsensitizationNon- irritantand Non-SensitizerNon- irritantand Non-SensitizerSame
ISO 10993-5: 2009 (E), Biological Evaluation ofMedical Devices - Part 5-Tests for in vitro CytotoxicityCytotoxicNon cytotoxicDifferent
ISO 10993-11: 2017(E). Biological Evaluation ofMedical Devices - Part 11, Tests for Systemic Toxicity,No systemictoxicity underthe conditionsof the testNot testedDifferent
USP 41 <151> Pyrogen TestNon pyrogenicNon pyrogenicsame

H. SUMMARY OF NON-CLINICAL PERFORMANCE DATA

Sl.NoCriteriaSpecification as perASTMD3577-09,StandardSpecification forRubber SurgicalGlovesAverage value ofsubject deviceCompliance with theASTMD 3577-09,Standard Specificationfor Rubber SurgicalGloves
1Length
Size 6Min 265mm306mmYes
Size 6'/2Min 265mm306mmYes
Size 7Min 265mm305mmYes
Size 7'/2Min 265mm305mmYes
Size 8Min 265mm305mmYes
Size 8'/2Min 265mm305mmYes
Size 9Min 265mm305mmYes

{7}------------------------------------------------

2Width
Size 676+/-6mm78mmYes
Size 6'⁄283+/-6mm85mmYes
Size 789+/-6mm88mmYes
Size 7'⁄295+/-6mm97mmYes
Size 8102+/-6mm103mmYes
Size 8'⁄2108+/-6mm110mmYes
Size 9114+/-6mm116mmYes
3Finger Thickness(All sizes)Min 0.10mm0.21mmYes
4Palm Thickness(All sizes)Min 0.10mm0.18mmYes
5Cuff Thickness(All sizes)Min 0.10mm0.11mmYes
SL.NOCriteriaSpecification as perASTMD3577-09StandardSpecification forRubber SurgicalGlovesAverage value ofsubject deviceWhether subject devicecomplied with theASTMD3577 -09,Standard Specificationfor Rubber SurgicalGloves
6Tensile Strength
Before aging(All sizes)24Mpa minimum26.0MpaYes
After aging@70°±2C for166±2 hr(All sizes)18Mpa minimum22.0MpaYes
7Ultimate Elongation
Before aging(All sizes)750% minimum860%Yes

{8}------------------------------------------------

After aging@70°±2C for166±2 hr(All sizes)560% minimum725%Yes
8Stress at 500%before ageing(All sizes)5.5 MPa Max2.7 MpaYes
9Pinhole AQL
Before aging(All sizes)Max 1.51.0Yes
After aging@70°C for 7 days(All sizes)Max 1.51.0Yes

Bench tests were conducted to verify that the proposed device met all design specifications or acceptance criteria found in the test method or standard. The test results demonstrated that the proposed device met design specifications or acceptance criteria with the following standards:

ASTM D3577-09(15):- Standard Specification for Rubber Surgical Gloves.

ASTM D 5151-06 (2015):-Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (2017):- Standard Test Method for Residual Powder on Medical Gloves.

ASTM D5712-15:-Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.

ASTM D6499-18:-Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products.

ASTM F 1929-2015:- Standard Test Method for Detecting Seal Leaks in Porous Meckaging by Dye Penetration.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

ISO 10993-11:2017(E) Biological Evaluation of Medical Devices - Part 11, Tests for Systemic Toxicity and Biological Tests

USP 41 <151> Pyrogen Test

ISO 11137-1-2006/ (R) 2010 - validation of sterilization process

ISO 11137-2:2013, sterilization of health care products - radiation - part 2: establishing the sterilization dose

I. CONCLUSION

The conclusion drawn from the nonclinical tests demonstrates that. JR MEDIC Latex Surgeon's Gloves Sterile Powder Free is as safe, as effective, and performs as well as or better than the legally marketed predicated device K172942.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).