Search Filters

Search Results

Found 21 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Direct Total Iron-Binding Capacity (TIBC) Reagent is intended for the quantitative determination of total iron-binding capacity in serum using the ACE Alera Clinical Chemistry System. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE Total Iron Reagent is intended for the quantitative determination of iron in serum using the ACE Alera Clinical Chemistry System. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE LDH-L Reagent is intended for the quantitative determination of lactate dehydrogenase activity in serum using the ACE Alera Clinical Chemistry System. Lactate dehydrogenase measurements are used in the diagnosis and treatment of liver diseases such as acute viral hepatitis, cirrhosis, and metastatic carcinoma of the liver, cardiac diseases such as myocardial infarction and tumors of the lung or kidneys. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Direct Total Iron-Binding Capacity (TIBC) Reagent assay, Direct TIBC Color Reagent, an acidic buffer containing an iron-binding dye and ferric chloride, is added to the serum sample. The low pH of Direct TIBC Color Reagent releases iron from transferrin. The iron then forms a colored complex with the dye. The colored complex at the end of the first step represents both the serum iron and excess iron already present in Direct TIBC Color Reagent. Direct TIBC Buffer, a neutral buffer, is then added, shifting the pH and resulting in a large increase in the affinity of transferrin for iron. The serum transferrin rapidly binds the iron by abstracting it from the dye-iron complex. The observed decrease in absorbance of the colored dye-iron complex is directly proportional to the total iron-binding capacity of the serum sample. The absorbance is measured at 647 nm.

    In the ACE Total Iron Reagent assay, transferrin-bound iron in serum is released at an acidic pH and reduced from ferric to ferrous ions. These ions react with ferrozine to form a violet colored complex, which is measured bichromatically at 554 nm/692 nm. The intensity of color produced is directly proportional to the serum iron concentration.

    In the ACE LDH-L Reagent assay, lactate dehydrogenase catalyzes the conversion of L-lactate to pyruvate. Nicotinamide adenine dinucleotide (NAD+) acts as an acceptor for the hydrogen ions released from the L-lactate and is converted to reduced nicotinamide adenine dinucleotide (NADH). NADH absorbs strongly at 340 nm whereas NAD+ does not. Therefore, the rate of conversion of NAD+ to NADH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/647 nm. This rate of conversion from NAD+ to NADH is directly proportional to the lactate dehydrogenase activity in the sample.

    AI/ML Overview

    The provided document describes in vitro diagnostic (IVD) reagents (ACE Direct Total Iron-Binding Capacity (TIBC) Reagent, ACE Total Iron Reagent, and ACE LDH-L Reagent) for use on the ACE Alera Clinical Chemistry System. The acceptance criteria and performance data presented relate to the analytical performance of these reagents/systems, specifically their ability to accurately and precisely measure analytes in serum samples.

    Crucially, this is not a study about an AI/ML powered medical device. Therefore, many of the typical acceptance criteria and study aspects requested in your prompt regarding AI/ML (e.g., ground truth established by experts, multi-reader multi-case studies, human-in-the-loop performance, training/test set sample sizes for AI, adjudication methods) are not applicable to this type of device and submission.

    The "study" described here is a series of analytical performance tests (linearity, precision, method comparison, detection limits, interference) to demonstrate that the new device (ACE Alera system with these reagents) performs comparably to the predicate device (ACE Clinical Chemistry System with the same reagents) and meets established analytical performance specifications for clinical chemistry assays.

    Here's a breakdown of the relevant information from the document in the format you requested, with an explanation of why certain AI/ML-centric points are not applicable:


    Device: ACE Direct Total Iron-Binding Capacity (TIBC) Reagent, ACE Total Iron Reagent, ACE LDH-L Reagent (for use on ACE Alera Clinical Chemistry System)

    1. Table of acceptance criteria and reported device performance:

    Since the document does not explicitly present "acceptance criteria" alongside "reported performance" in a single table, I will infer the acceptance criteria from the context of method comparison, linearity, and precision studies, which are standard for IVD device validation, often aiming for performance comparable to predicate devices or within clinically acceptable limits. The reported performance is directly extracted from the tables provided.

    Interference:
    The acceptance criterion for interference studies in IVD assays is typically that the interferent, up to a specified concentration, does not cause a "significant interference" (e.g., a bias exceeding a defined clinical or analytical threshold). The document lists the concentrations at which no significant interference was observed.

    InterferentNo Significant Interference at or below (Acceptance Criteria Implicit)Reported Device Performance (Concentration where no significant interference was observed)
    TIBC
    IcterusAssumes standard limits for non-interference59 mg/dL
    HemolysisAssumes standard limits for non-interference188 mg/dL*
    LipemiaAssumes standard limits for non-interference1000 mg/dL
    Ascorbic AcidAssumes standard limits for non-interference3 mg/dL
    Iron
    IcterusAssumes standard limits for non-interference59 mg/dL
    HemolysisAssumes standard limits for non-interference125 mg/dL*
    LipemiaAssumes standard limits for non-interference125 mg/dL
    Ascorbic AcidAssumes standard limits for non-interference6 mg/dL
    LDH-L
    IcterusAssumes standard limits for non-interference50 mg/dL
    HemolysisAssumes standard limits for non-interference0.99), ideally with the confidence intervals for slope encompassing 1 and for intercept encompassing 0. This indicates analytical equivalence between the two systems.
    Reagentn (samples)Range (of samples tested)Reported SlopeReported InterceptReported Correlation CoefficientCI SlopeCI Intercept
    TIBC5059 to 676 µg/dL0.9873.60.99600.962 to 1.013-7.2 to 14.4
    Iron4813 to 549 µg/dL0.9930.90.99950.984 to 1.003-0.6 to 2.3
    LDH-L5820 to 799 U/L0.997-3.60.99910.985 to 1.008-6.1 to -1.1

    Precision (POL - Point of Care/Physician Office Lab):
    Similar to in-house precision, specific %CV or SD limits would be the acceptance criteria. The data shows results from 3 POLs compared to in-house.

    ReagentLabSample LevelMeanWithin-Run SD, %CVTotal SD, %CV
    Direct TIBCIn-House13305.1, 1.5%5.8, 1.8%
    POL 112848.3, 2.9%9.6, 3.4%
    POL 212595.6, 2.2%8.5, 3.3%
    POL 312769.1, 3.3%16.7, 6.0%
    In-House24504.9, 1.1%6.8, 1.5%
    POL 124646.3, 1.4%6.6, 1.4%
    POL 224444.2, 1.0%5.4, 1.2%
    POL 324533.2, 0.7%15.5, 3.4%
    In-House35309.4, 1.8%10.8, 2.0%
    POL 135448.2, 1.5%8.3, 1.5%
    POL 235205.0, 1.0%9.0, 1.7%
    POL 3353312.6, 2.4%20.2, 3.8%
    Total IronIn-House11191.8, 1.5%2.5, 2.1%
    POL 111192.7, 2.3%3.2, 2.7%
    POL 211223.1, 2.6%3.1, 2.6%
    POL 311163.2, 2.8%3.4, 3.0%
    In-House22223.8, 1.7%5.1, 2.3%
    POL 122292.0, 0.9%2.5, 1.1%
    POL 222352.3, 1.0%2.4, 1.0%
    POL 322293.4, 1.5%3.9, 1.7%
    In-House34125.2, 1.3%5.7, 1.4%
    POL 134244.0, 0.9%4.6, 1.1%
    POL 234352.4, 0.5%5.3, 1.2%
    POL 3342811.1, 2.6%11.1, 2.6%
    LDH-LIn-House11182.9, 2.4%5.7, 4.8%
    POL 111161.7, 1.5%4.9, 4.3%
    POL 211183.0, 2.5%5.1, 4.3%
    POL 311243.4, 2.7%4.7, 3.8%
    In-House24334.7, 1.1%6.5, 1.5%
    POL 124372.9, 0.7%5.8, 1.3%
    POL 224493.7, 0.8%5.2, 1.2%
    POL 324465.8, 1.3%6.6, 1.5%
    In-House36995.3, 0.8%8.5, 1.2%
    POL 136988.6, 1.2%11.5, 1.6%
    POL 237265.4, 0.8%10.0, 1.4%
    POL 3371614.3, 2.0%16.9, 2.4%

    Method Comparison (POLs vs. In-House (ACE Alera (x) vs. POL ACE Alera (y))):
    Similar to the in-house method comparison, the acceptance criteria are for slopes to be near 1, intercepts near 0, and high correlation coefficients (e.g., >0.99), indicating consistent performance across different lab environments.

    ReagentLab Comparisonn (samples)RangeReported RegressionReported CorrelationCI SlopeCI Intercept
    TIBCIn-House vs. POL 15059 to 676y = 0.994x + 12.40.99340.961 to 1.027-1.7 to 26.5
    In-House vs. POL 25059 to 676y = 0.973x + 0.10.99540.946 to 1.001-11.4 to 11.6
    In-House vs. POL 35059 to 676y = 1.005x + 9.00.98980.963 to 1.047-8.7 to 26.6
    IronIn-House vs. POL 14813 to 549y = 0.976x + 1.00.99860.960 to 0.991-1.4 to 3.3
    In-House vs. POL 24813 to 549y = 0.976x + 2.30.99810.959 to 0.994-0.4 to 5.0
    In-House vs. POL 34813 to 549y = 0.951x + 0.80.99660.927 to 0.974-2.7 to 4.4
    LDH-LIn-House vs. POL 15174 to 799y = 0.992x + 3.50.99860.977 to 1.008-0.1 to 7.1
    In-House vs. POL 25174 to 799y = 1.027x + 3.40.99891.013 to 1.0410.2 to 6.7
    In-House vs. POL 35174 to 799y = 1.010x + 2.50.99840.994 to 1.026-1.3 to 6.2

    2. Sample sizes used for the test set and the data provenance:

    • Sample Sizes for analytical performance studies (Test Set):

      • Method Comparison:
        • TIBC: 50 samples
        • Iron: 48 samples
        • LDH-L: 58 (in-house comparison) / 51 (POL comparison) samples
      • Linearity: The number of samples/levels for linearity is not explicitly stated as 'n', but standard practice involves multiple levels (typically 5-7) prepared from diluted/spiked samples.
      • Precision: Standard runs (e.g., 2 runs per day for 20 days for total precision, with replicates per run for within-run precision) would involve a substantial number of measurements (e.g., 20 days x 2 runs/day x 2 replicates/run = 80 measurements per level). The POL precision data shows n=20, likely referring to 20 days of testing.
      • Interference: The number of samples used for interference studies is not explicitly stated.
    • Data Provenance: "In-House" and "POL" (Physician Office Laboratories). The specific country of origin is not explicitly stated, but given the company's location (New Jersey, USA) and FDA 510(k) submission, it's highly likely to be United States. The studies are prospective analytical validation studies, meaning the data was collected specifically to demonstrate the performance of the device.

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    Not applicable. This is an in vitro diagnostic (IVD) chemistry analyzer and reagent system. "Ground truth" for IVD analytical performance is established by reference methods, certified reference materials, or highly accurate comparative methods, not by human expert consensus or radiologists. The performance is assessed against quantitative values, not qualitative interpretations requiring expert review.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., imaging studies where radiologists disagree). For analytical performance of a chemistry analyzer, the "ground truth" is typically the quantitative value obtained from a reference method or the predicate device, and differences are assessed statistically (e.g., bias, correlation).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. MRMC studies are specific to evaluating the impact of a device on human readers' performance, typically in diagnostic imaging with AI assistance. This device is an automated chemistry analyzer, not an AI-assisted diagnostic imaging tool. There are no human "readers" in the context of this device's operation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Yes, in essence. The performance data provided (linearity, precision, detection limits, interference, method comparison) represents the "standalone" analytical performance of the automated chemistry system (ACE Alera with the new reagents) in measuring the target analytes in patient samples. There isn't an "algorithm only" in the AI sense, but the chemical reactions and photometric measurements are entirely automated by the device. The data shown is the raw analytical output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for these analytical studies is primarily:

    • Highly characterized samples: For linearity, samples with known, precise concentrations (often prepared by dilution of high-concentration materials or spiking low-concentration materials).
    • Comparative method/Predicate device: For method comparison, the results generated by the predicate device (ACE Clinical Chemistry System) are treated as the reference or comparative method against which the new ACE Alera system's results are compared. This is a common and accepted "ground truth" for chemical analyzers seeking substantial equivalence.
    • Reference materials/controls: For precision and detection limits, control materials with established target values are used.

    8. The sample size for the training set:

    Not applicable. This is a traditional IVD device (chemical reagents and analyzer), not an AI/ML device that requires a "training set" in the context of machine learning model development. The reagents perform chemical reactions, and the analyzer reads photometric changes; it does not "learn" from data.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set in the AI/ML sense for this device.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Carbon Dioxide (CO2-LC) Reagent is intended for the quantitative determination of carbon dioxide concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Direct Bilirubin Reagent is intended for the quantitative determination of direct bilirubin concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological and metabolic disorders, including hepatitis and gall bladder block. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Total Bilirubin Reagent is intended for the quantitative determination of total bilirubin concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry System. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological and metabolic disorders, including hepatitis and gall bladder block. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Magnesium Reagent is intended for the quantitative determination of magnesium in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Carbon Dioxide (CO2-LC) Reagent assay, serum carbon dioxide (in the form of bicarbonate) reacts with phosphoenolpyruvate in the presence of phosphoenolpyruvate carboxylase and magnesium to yield oxaloacetic acid and phosphate. In the presence of malate dehydrogenase, the reduced cofactor is oxidized by oxaloacetic acid. The reduced cofactor absorbs strongly at 408 nm whereas its oxidized form does not. The rate of decrease in absorbance, monitored bichromatically at 408 nm/692 nm, is proportional to the carbon dioxide content of the sample.

    In the ACE Direct Bilirubin Reagent assay, sodium nitrite added to sulfanilic acid forms diazotized sulfanilic acid. Bilirubin glucuronide in serum reacts with diazotized sulfanilic acid to form azobilirubin, which absorbs strongly at 554 nm. The increase in absorbance, measured bichromatically at 554 nm/692 nm, one minute after sample addition, is directly proportional to the direct bilirubin concentration.

    In the ACE Total Bilirubin Reagent assay, sodium nitrite, when added to sulfanilic acid, forms diazotized sulfanilic acid. Bilirubin in serum reacts with diazotized sulfanilic acid to form azobilirubin, which absorbs strongly at 554 nm. The inclusion of dimethyl sulfoxide (DMSO) in the reagent as an accelerator causes both direct and indirect bilirubin to react rapidly. The increase in absorbance, measured bichromatically at 554 nm/692 nm, is directly proportional to the total bilirubin concentration in the sample.

    Magnesium ions in serum react with Xylidyl blue-1 in an alkaline medium to produce a red complex which is measured bichromatically at 525 nm/692 nm. The intensity of color produced is directly proportional to the magnesium concentration in the sample. EGTA prevents calcium interference by preferential chelation of calcium present in the sample. A surfactant system is included to remove protein interference.

    AI/ML Overview

    The provided text describes several in vitro diagnostic reagents (ACE Carbon Dioxide (CO2-LC) Reagent, ACE Direct Bilirubin Reagent, ACE Total Bilirubin Reagent, and ACE Magnesium Reagent) and their associated performance data. There isn't information about an AI-powered device or software. Therefore, questions related to AI aspects like multi-reader multi-case studies, effect size of AI assistance, or standalone algorithm performance are not applicable.

    The acceptance criteria are not explicitly stated as clear thresholds in the provided document; rather, the document presents detailed performance data (precision, linearity, interference, and method comparison) that demonstrates the device's capability to perform as intended and to be substantially equivalent to its predicate devices. The "reported device performance" is presented directly through tables and statistical analyses for each reagent.

    Here's an attempt to structure the available information based on the request, interpreting "acceptance criteria" as the performance demonstrated to support substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since explicit "acceptance criteria" (i.e., predefined thresholds for performance metrics) are not provided in the document, the "Reported Device Performance" below represents the data presented that presumably met the internal criteria for demonstrating substantial equivalence. The document primarily focuses on precision, linearity, interference, and method comparison with predicate devices and between different systems (ACE, ACE Alera, ACE Axcel).

    ACE Carbon Dioxide (CO2-LC) Reagent

    MetricAcceptance Criteria (Inferred from study design and historical data, not explicitly stated values)Reported Device Performance (Summary of results across systems/sites)
    Precision (In-House)(Implied to be comparable to or better than predicate and acceptable for clinical use)Serum:
    Low: ~1.1-2.5% CV (Within-Run), ~5.6-7.5% CV (Total)
    Mid: ~1.2-1.5% CV (Within-Run), ~3.3-3.7% CV (Total)
    High: ~0.6-2.8% CV (Within-Run), ~2.6-3.2% CV (Total)
    Plasma:
    Low: ~1.3-3.0% CV (Within-Run), ~3.8-6.1% CV (Total)
    Mid: ~0.7-1.2% CV (Within-Run), ~5.0-5.5% CV (Total)
    High: ~1.0% CV (Within-Run), ~2.3-2.5% CV (Total)
    Precision (POL sites)(Implied to be comparable to in-house and acceptable for clinical use)ACE:
    Low (Sample 1): ~1.6-3.3% CV (Within-Run), ~3.0-4.3% CV (Total)
    Mid (Sample 2): ~1.7-3.1% CV (Within-Run), ~2.7-7.4% CV (Total)
    High (Sample 3): ~1.8-2.4% CV (Within-Run), ~2.4-6.4% CV (Total)
    ACE Alera:
    Low (Sample 1): ~1.3-2.0% CV (Within-Run), ~3.0-6.7% CV (Total)
    Mid (Sample 2): ~0.9-1.7% CV (Within-Run), ~2.4-3.9% CV (Total)
    High (Sample 3): ~1.0-1.6% CV (Within-Run), ~3.1-5.8% CV (Total)
    Method Comparison (Serum vs. Plasma)(Slope near 1, intercept near 0, high correlation)ACE: Slope: 1.031, Intercept: -1.03, Correlation: 0.9922
    ACE Alera: Slope: 1.000, Intercept: -0.09, Correlation: 0.9955
    ACE Axcel: Slope: 0.988, Intercept: -0.35, Correlation: 0.9889
    Method Comparison (POL vs. In-House)(Slope near 1, intercept near 0, high correlation)ACE (POL 1-3 vs. In-House ACE): Slopes: 0.963-0.984, Intercepts: -0.71-1.29, Correlations: 0.9530-0.9908
    ACE Alera (POL 1-3 vs. In-House ACE): Slopes: 0.972-0.987, Intercepts: 0.10-0.57, Correlations: 0.9767-0.9903
    Detection Limits (ACE Alera)(Appropriate for clinical use)LoB: 1.27 mEq/L, LoD: 1.97 mEq/L, LoQ: 3.03 mEq/L
    Linearity (ACE Alera)(Linearity up to/beyond desired measuring range)Linear to: 50 mEq/L (Equation: y=1.006x + 0.01)
    Interferences (ACE Alera)(No significant interference from common interferents)No significant interference at or below Icterus 58.8 mg/dL, Hemolysis 250 mg/dL, Lipemia 2388 mg/dL, Ascorbic Acid 6 mg/dL

    ACE Direct Bilirubin Reagent

    MetricAcceptance Criteria (Inferred from study design and historical data, not explicitly stated values)Reported Device Performance (Summary of results across systems/sites)
    Precision (In-House)(Implied to be comparable to or better than predicate and acceptable for clinical use)Serum:
    Low: ~12.5-24.5% CV (Within-Run), ~14.0-30.0% CV (Total)
    Mid: ~0.9-1.6% CV (Within-Run), ~1.2-2.2% CV (Total)
    High: ~0.6-1.5% CV (Within-Run), ~1.1-1.7% CV (Total)
    Plasma:
    Low: ~16.6-26.6% CV (Within-Run), ~19.7-35.4% CV (Total)
    Mid: ~0.8-2.4% CV (Within-Run), ~1.1-2.8% CV (Total)
    High: ~0.7-1.9% CV (Within-Run), ~1.1-2.3% CV (Total)
    Precision (POL sites)(Implied to be comparable to in-house and acceptable for clinical use)ACE:
    Low (Sample 1): ~2.9-4.2% CV (Within-Run), ~2.9-4.9% CV (Total)
    Mid (Sample 2): ~1.0-1.8% CV (Within-Run), ~1.3-2.1% CV (Total)
    High (Sample 3): ~1.3-2.3% CV (Within-Run), ~2.0-2.3% CV (Total)
    ACE Alera:
    Low (Sample 1): ~2.5-5.1% CV (Within-Run), ~2.5-5.4% CV (Total)
    Mid (Sample 2): ~1.0-1.5% CV (Within-Run), ~1.0-1.9% CV (Total)
    High (Sample 3): ~0.6-2.6% CV (Within-Run), ~1.3-2.6% CV (Total)
    Method Comparison (Serum vs. Plasma)(Slope near 1, intercept near 0, high correlation)ACE: Slope: 1.021, Intercept: 0.00, Correlation: 0.9982
    ACE Alera: Slope: 1.005, Intercept: 0.01, Correlation: 0.9978
    ACE Axcel: Slope: 1.004, Intercept: 0.00, Correlation: 0.9983
    Method Comparison (POL vs. In-House)(Slope near 1, intercept near 0, high correlation)ACE (POL 1-3 vs. In-House ACE): Slopes: 1.003-1.022, Intercepts: 0.04-0.11, Correlations: 0.9984-0.9986
    ACE Alera (POL 1-3 vs. In-House ACE): Slopes: 0.969-0.995, Intercepts: 0.09-0.11, Correlations: 0.9984-0.9991
    Detection Limits (ACE Alera)(Appropriate for clinical use)LoB: 0.06 mg/dL, LoD: 0.08 mg/dL, LoQ: 0.12 mg/dL
    Linearity (ACE Alera)(Linearity up to/beyond desired measuring range)Linear to: 14.0 mg/dL (Equation: y=1.015x + 0.16)
    Interferences (ACE Alera)(No significant interference from common interferents)Not Applicable (Icterus), No significant interference at or below Hemolysis 62.5 mg/dL, Lipemia 782 mg/dL, Ascorbic Acid 6 mg/dL

    ACE Total Bilirubin Reagent

    MetricAcceptance Criteria (Inferred from study design and historical data, not explicitly stated values)Reported Device Performance (Summary of results across systems/sites)
    Precision (In-House)(Implied to be comparable to or better than predicate and acceptable for clinical use)Serum:
    Low: ~11.0-21.3% CV (Within-Run), ~13.9-21.3% CV (Total)
    Mid: ~1.0-1.1% CV (Within-Run), ~1.0-1.1% CV (Total)
    High: ~0.4-0.7% CV (Within-Run), ~0.5-0.8% CV (Total)
    Plasma:
    Low: ~20.3-23.7% CV (Within-Run), ~21.3-29.4% CV (Total)
    Mid: ~0.5-1.0% CV (Within-Run), ~0.5-1.1% CV (Total)
    High: ~0.5-0.6% CV (Within-Run), ~0.5-0.7% CV (Total)
    Precision (POL sites)(Implied to be comparable to in-house and acceptable for clinical use)ACE:
    Low (Sample 1): ~3.4-5.5% CV (Within-Run), ~3.7-5.8% CV (Total)
    Mid (Sample 2): ~0.5-1.7% CV (Within-Run), ~1.3-3.7% CV (Total)
    High (Sample 3): ~1.0-1.2% CV (Within-Run), ~1.2-2.1% CV (Total)
    ACE Alera:
    Low (Sample 1): ~4.2-4.9% CV (Within-Run), ~4.5-5.2% CV (Total)
    Mid (Sample 2): ~0.7-2.0% CV (Within-Run), ~0.8-2.1% CV (Total)
    High (Sample 3): ~0.5-1.4% CV (Within-Run), ~0.6-1.7% CV (Total)
    Method Comparison (Serum vs. Plasma)(Slope near 1, intercept near 0, high correlation)ACE: Slope: 1.017, Intercept: 0.01, Correlation: 0.9996
    ACE Alera: Slope: 1.020, Intercept: 0.00, Correlation: 0.9993
    ACE Axcel: Slope: 1.008, Intercept: 0.00, Correlation: 0.9995
    Method Comparison (POL vs. In-House)(Slope near 1, intercept near 0, high correlation)ACE (POL 1-3 vs. In-House ACE): Slopes: 0.979-1.000, Intercepts: 0.00-0.04, Correlations: 0.9995-0.9998
    ACE Alera (POL 1-3 vs. In-House ACE): Slopes: 0.957-1.020, Intercepts: 0.01-0.07, Correlations: 0.9991-0.9998
    Detection Limits (ACE Alera)(Appropriate for clinical use)LoB: 0.11 mg/dL, LoD: 0.14 mg/dL, LoQ: 0.14 mg/dL
    Linearity (ACE Alera)(Linearity up to/beyond desired measuring range)Linear to: 40.0 mg/dL (Equation: y=1.004x + 0.03)
    Interferences (ACE Alera)(No significant interference from common interferents)Not Applicable (Icterus), No significant interference at or below Hemolysis 62.5 mg/dL, Lipemia 951 mg/dL, Ascorbic Acid 6 mg/dL

    ACE Magnesium Reagent

    MetricAcceptance Criteria (Inferred from study design and historical data, not explicitly stated values)Reported Device Performance (Summary of results across systems/sites)
    Precision (In-House)(Implied to be comparable to or better than predicate and acceptable for clinical use)Serum:
    Low: ~3.1-5.1% CV (Within-Run), ~4.3-5.9% CV (Total)
    Mid: ~1.7-2.6% CV (Within-Run), ~1.8-3.0% CV (Total)
    High: ~1.0-1.4% CV (Within-Run), ~1.6-1.7% CV (Total)
    Plasma:
    Low: ~2.4-4.7% CV (Within-Run), ~4.1-6.8% CV (Total)
    Mid: ~2.4-2.8% CV (Within-Run), ~2.6-3.7% CV (Total)
    High: ~0.9-1.6% CV (Within-Run), ~1.8-1.9% CV (Total)
    Precision (POL sites)(Implied to be comparable to in-house and acceptable for clinical use)ACE:
    Low (Sample 1): ~3.3-4.6% CV (Within-Run), ~5.0-6.3% CV (Total)
    Mid (Sample 2): ~1.3-2.5% CV (Within-Run), ~2.8-3.5% CV (Total)
    High (Sample 3): ~1.1-1.8% CV (Within-Run), ~1.4-3.1% CV (Total)
    ACE Alera:
    Low (Sample 1): ~3.0-6.0% CV (Within-Run), ~4.5-8.4% CV (Total)
    Mid (Sample 2): ~2.0-2.9% CV (Within-Run), ~2.5-5.2% CV (Total)
    High (Sample 3): ~0.9-1.9% CV (Within-Run), ~1.6-4.8% CV (Total)
    Method Comparison (Serum vs. Plasma)(Slope near 1, intercept near 0, high correlation)ACE: Slope: 0.957, Intercept: 0.04, Correlation: 0.9765
    ACE Alera: Slope: 0.986, Intercept: 0.05, Correlation: 0.9817
    ACE Axcel: Slope: 0.986, Intercept: 0.025, Correlation: 0.9892
    Method Comparison (POL vs. In-House)(Slope near 1, intercept near 0, high correlation)ACE (POL 1-3 vs. In-House ACE): Slopes: 0.970-1.026, Intercepts: -0.04-0.16, Correlations: 0.9902-0.9927
    ACE Alera (POL 1-3 vs. In-House ACE): Slopes: 0.990-1.010, Intercepts: -0.11-0.00, Correlations: 0.9870-0.9930
    Detection Limits (ACE Alera)(Appropriate for clinical use)LoB: 0.26 mg/dL, LoD: 0.37 mg/dL, LoQ: 0.37 mg/dL
    Linearity (ACE Alera)(Linearity up to/beyond desired measuring range)Linear to: 6.1 mg/dL (Equation: y=0.959x + 0.27)
    Interferences (ACE Alera)(No significant interference from common interferents)No significant interference at or below Icterus 50 mg/dL, Hemolysis 500 mg/dL, Lipemia 620 mg/dL, Ascorbic Acid 6 mg/dL

    2. Sample Size Used for the Test Set and the Data Provenance

    The document describes several types of studies:

    • In-House Precision:

      • CO2-LC: Low, Mid, High serum and plasma samples were tested (number of replicates per sample and runs is implicitly part of SD/CV calculation, but not explicitly stated).
      • Direct Bilirubin: Low, Mid, High serum and plasma samples.
      • Total Bilirubin: Low, Mid, High serum and plasma samples.
      • Magnesium: Low, Mid, High serum and plasma samples.
      • Data Provenance: In-house (Alfa Wassermann Diagnostic Technologies, LLC, West Caldwell, NJ), prospective testing.
    • POL (Physician Office Laboratory) Precision: Studies conducted at 3 POL sites.

      • CO2-LC: 3 samples at each of 3 POL sites and in-house.
      • Direct Bilirubin: 3 samples at each of 3 POL sites and in-house.
      • Total Bilirubin: 3 samples at each of 3 POL sites and in-house.
      • Magnesium: 3 samples at each of 3 POL sites and in-house.
      • Data Provenance: Not explicitly stated but inferred to be from POLs in the USA (prospective testing under typical POL conditions).
    • In-House Matrix Comparison (Serum vs. Plasma):

      • CO2-LC: 53-54 pairs (serum/plasma) on ACE and ACE Alera; 51 pairs on ACE Axcel.
      • Direct Bilirubin: 102 pairs on ACE; 101 pairs on ACE Alera; 56 pairs on ACE Axcel.
      • Total Bilirubin: 102 pairs on ACE and ACE Alera; 56 pairs on ACE Axcel.
      • Magnesium: 101 pairs on ACE and ACE Alera; 55 pairs on ACE Axcel.
      • Data Provenance: In-house, retrospective (presumably collected for a range of values).
    • POL Method Comparison (In-House ACE vs. POL ACE/Alera):

      • CO2-LC: 45-46 samples per POL site comparison.
      • Direct Bilirubin: 49-51 samples per POL site comparison.
      • Total Bilirubin: 48-50 samples per POL site comparison.
      • Magnesium: 50-52 samples per POL site comparison.
      • Data Provenance: Not explicitly stated but inferred to be from POLs in the USA (prospective testing under typical POL conditions) compared against in-house data.
    • Detection Limits (LoB, LoD, LoQ), Linearity, Interferences (ACE Alera):

      • Sample sizes for detection limits and linearity: Not explicitly stated, typically involves multiple replicates at various concentrations.
      • Sample sizes for interferences: Not explicitly stated, typically involves samples spiked with various concentrations of interferents.
      • Data Provenance: In-house.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. For in vitro diagnostic assays, the "ground truth" is typically the reference method or established clinical laboratory results obtained from a highly accurate and calibrated instrument or laboratory using validated methods, rather than human expert consensus for image or clinical interpretation. The document compares performance against other (presumably established) methods and predicate devices.

    4. Adjudication Method for the Test Set

    This concept (e.g., 2+1, 3+1 for resolving discrepancies) is not applicable to these types of in vitro diagnostic device studies. Performance is measured numerically and objectively.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This is an in vitro diagnostic assay, not an AI-powered diagnostic imaging device.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. The performance data presented are for the reagent and instrument system.

    7. The Type of Ground Truth Used

    For precision studies, the "ground truth" is the true concentration of the analyte in the control material or patient sample, which is established by reference methods or manufacturing specifications of the control materials. For method comparison studies, the predicate device's results or an established in-house method are used as the comparative reference. The document states the intended use is for "quantitative determination" of analytes, implying comparison to a quantitative gold standard.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning device and therefore does not have a "training set" in that context. The development of reagents and the establishment of their performance characteristics do not involve machine learning training sets.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" for these reagents in the context of AI/ML.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Alera Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative measurement of general chemistry assays, such as glucose, sodium, potassium, and chloride, for clinical use in physician office laboratories or clinical laboratories. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    ACE Glucose Reagent is intended for the quantitative determination of glucose in serum and lithium heparin plasma using the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    The ACE Ion Selective Electrode (ISE) module on the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems is used to measure concentrations of sodium, potassium, and chloride in undiluted serum and lithium heparin plasma. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. This test is intended for use in clinical laboratories and physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Alera Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative determination of general chemistry assays for clinical use in physician office laboratories or clinical laboratories. The ACE Alera Clinical Chemistry System consists of a bench-top analyzer and an internal computer. The bench-top analyzer includes a single pipettor (syringe module/fluid arm/probe), a temperature-controlled reagent compartment, a reaction wheel and a holographic diffraction grating spectrophotometer.

    In the ACE Glucose Reagent assay, glucose in serum or heparin plasma reacts with adenosine triphosphate in the presence of hexokinase and magnesium with the formation of glucose-6-phosphate and adenosine diphosphate. Glucose-6-phosphate dehydrogenase catalyzes the oxidation of glucose-6-phosphate with NAD+ to form 6-phosphogluconate and NADH. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The total amount of NADH formed is proportional to the concentration of glucose in the sample. The increase in absorbance is measured bichromatically at 340 nm/378 nm.

    The ACE Ion Selective Electrode (ISE) Module, as part of the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems, uses a potentiometric method via ion-specific electrodes to simultaneously measure sodium, potassium and chloride in undiluted serum. Ion-specific membranes measure the difference in ionic concentrations between an inner electrolyte solution and the sample. The connection of the amplifier and ground (reference electrode) to the ion selective electrode forms the measuring system. A two-point calibration utilizes ACE CAL A and CAL B undiluted ISE Calibration Solutions with precisely known ion concentrations. The measured voltage difference of the sample and the CAL A and CAL B solutions determines the ion concentration in the sample on the ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems.

    AI/ML Overview

    The device is the ACE Alera Clinical Chemistry System, ACE Glucose Reagent, and ACE Ion Selective Electrode (ISE) Module. The study assesses the performance of these components, focusing on the quantitative measurement of glucose, sodium, potassium, and chloride.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets that the device must meet in a formal, quantifiable way (e.g., "Accuracy must be > 95%"). Instead, the study aims to demonstrate substantial equivalence to predicate devices, showing that the performance of the ACE Alera system is comparable to established systems. The performance data presented focuses on precision (reproducibility) and method comparison with existing devices.

    Since specific numerical acceptance criteria were not listed, one reasonable interpretation for implied acceptance criteria for laboratory diagnostic devices typically includes:

    • Acceptable Precision: Coefficients of Variation (CV) or Standard Deviations (SD) for within-run and total precision across different concentration levels should be within generally accepted laboratory limits for each analyte. For clinical chemistry, these are often defined considering medical usefulness.
    • Acceptable Method Agreement: Linear regression analysis (slope, intercept, correlation coefficient) and standard error between the new device and a reference method (or predicate device) should indicate good agreement. Slopes close to 1, intercepts close to 0, and high correlation coefficients (e.g., >0.975) are generally desired.
    • No Significant Interference: The device should not be significantly affected by common interfering substances (icterus, hemolysis, lipemia, ascorbic acid) at clinically relevant levels.

    Here's the performance data as reported, which serves as the evidence that these implicit acceptance criteria are met:

    AnalytePerformance MetricAcceptance Criteria (Implied)Reported Device Performance (ACE Alera)
    GlucosePrecision (SD, %CV)Low CVs and SDs across different concentrations.Serum Low (62 mg/dL): Within-Run SD 0.6, CV 0.9%; Total SD 0.8, CV 1.3%
    Serum Mid (121 mg/dL): Within-Run SD 1.2, CV 1.0%; Total SD 1.5, CV 1.3%
    Serum High (366 mg/dL): Within-Run SD 6.4, CV 1.8%; Total SD 6.9, CV 1.9%
    POL PrecisionSimilar precision across different lab settings.In-House Sample 1 (63.5 mg/dL): Within-Run SD 1.2, CV 1.9%; Total SD 1.3, CV 2.1%
    POL 1 Sample 1 (64.3 mg/dL): Within-Run SD 1.1, CV 1.7%; Total SD 1.5, CV 2.3%
    (Similar data for other POLs and samples)
    Method ComparisonSlope close to 1, Intercept close to 0, R > 0.975.POL 1: Slope 1.015, Intercept 0.1, R 0.9993
    POL 2: Slope 1.005, Intercept 3.1, R 0.9995
    POL 3: Slope 0.988, Intercept 3.2, R 0.9993
    SodiumPrecision (SD, %CV)Low CVs and SDs across different concentrations.Serum Low (111.2 mmol/L): Within-Run SD 0.59, CV 0.5%; Total SD 0.93, CV 0.8%
    Serum Mid (139.0 mmol/L): Within-Run SD 0.80, CV 0.6%; Total SD 0.87, CV 0.6%
    Serum High (159.9 mmol/L): Within-Run SD 0.38, CV 0.2%; Total SD 0.90, CV 0.6%
    POL PrecisionSimilar precision across different lab settings.In-House Sample 1 (107.5 mmol/L): Within-Run SD 0.80 CV 0.7%; Total SD 1.50 CV 1.4%
    POL 1 Sample 1 (108.4 mmol/L): Within-Run SD 0.93 CV 0.9%; Total SD 1.44 CV 1.3%
    (Similar data for other POLs and samples)
    Method ComparisonSlope close to 1, Intercept close to 0, R > 0.975.POL 1: Slope 1.025, Intercept -1.74, R 0.9974
    POL 2: Slope 1.021, Intercept -2.92, R 0.9958
    POL 3: Slope 1.044, Intercept -6.27, R 0.9979
    PotassiumPrecision (SD, %CV)Low CVs and SDs across different concentrations.Serum Low (2.2 mmol/L): Within-Run SD 0.04, CV 1.6%; Total SD 0.05, CV 2.4%
    Serum Mid (4.0 mmol/L): Within-Run SD 0.07, CV 1.8%; Total SD 0.07, CV 1.8%
    Serum High (7.9 mmol/L): Within-Run SD 0.07, CV 0.9%; Total SD 0.11, CV 1.4%
    POL PrecisionSimilar precision across different lab settings.In-House Sample 1 (3.70 mmol/L): Within-Run SD 0.06 CV 1.6%; Total SD 0.06 CV 1.7%
    POL 1 Sample 1 (3.73 mmol/L): Within-Run SD 0.07 CV 1.8%; Total SD 0.08 CV 2.2%
    (Similar data for other POLs and samples)
    Method ComparisonSlope close to 1, Intercept close to 0, R > 0.975.POL 1: Slope 1.032, Intercept -0.108, R 0.9983
    POL 2: Slope 1.008, Intercept -0.054, R 0.9971
    POL 3: Slope 0.984, Intercept 0.150, R 0.9942
    ChloridePrecision (SD, %CV)Low CVs and SDs across different concentrations.Serum Low (75.0 mmol/L): Within-Run SD 0.80, CV 1.1%; Total SD 1.50, CV 2.0%
    Serum Mid (99.2 mmol/L): Within-Run SD 0.80, CV 0.8%; Total SD 0.90, CV 0.9%
    Serum High (119.3 mmol/L): Within-Run SD 0.50, CV 0.4%; Total SD 1.10, CV 0.9%
    POL PrecisionSimilar precision across different lab settings.In-House Sample 1 (77.3 mmol/L): Within-Run SD 0.50 CV 0.6%; Total SD 1.20 CV 1.6%
    POL 1 Sample 1 (78.1 mmol/L): Within-Run SD 0.76 CV 1.0%; Total SD 1.30 CV 1.7%
    (Similar data for other POLs and samples)
    Method ComparisonSlope close to 1, Intercept close to 0, R > 0.975.POL 1: Slope 1.004, Intercept 0.96, R 0.9972
    POL 2: Slope 1.000, Intercept 0.29, R 0.9956
    POL 3: Slope 1.006, Intercept 0.16, R 0.9946
    InterferenceNo significant interferenceThresholds for common interferents.GLU: No significant interference at or below 26 mg/dL Icterus, 1000 mg/dL Hemolysis, 104 mg/dL Lipemia (Intralipid), 525 mg/dL Lipemia (Triglycerides), 6 mg/dL Ascorbic Acid.
    (Similar thresholds for Na, K, Cl)

    The study essentially acts as a validation against these implied criteria, demonstrating that the ACE Alera system's performance is acceptable for its intended use, comparable to the predicate devices.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Precision Studies: The document does not explicitly state the number of individual sample replicates for the core (non-POL) precision studies. However, for the POL Precision studies, for each analyte (Glucose, Sodium, Potassium, Chloride), there were 3 samples tested in each of 4 labs (In-House and 3 POLs). The tables show means, within-run, and total standard deviations/CVs, which typically imply multiple replicates per sample (e.g., 20 or more replicates are common in such studies).
    • Method Comparison Studies:
      • Glucose: n = 46 samples for each of the three POL comparisons.
      • Sodium: n = 42 samples for each of the three POL comparisons.
      • Potassium: n = 43 samples for each of the three POL comparisons.
      • Chloride: n = 41 samples for each of the three POL comparisons.
    • Data Provenance: The method comparison data is identified as "(2012 Data)" and collected from an "In-House" lab comparing against "ACE Alera system POL" data from three different Physician Office Laboratories (POLs 1, 2, 3), suggesting multi-center evaluation within the United States. The data is retrospective in the sense that it's reported for a 510(k) submission, but the studies themselves would have been conducted prospectively as a part of the device validation. The term "POL" indicates that these are real-world, clinical laboratory settings.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    There were no human experts establishing ground truth in the context of interpretation for these types of in vitro diagnostic devices. The "ground truth" or reference values for chemical assays like glucose, sodium, potassium, and chloride are established by:

    • Reference Methods: Often, a more established or gold-standard laboratory analyzer (in this case, the predicate ACE system in the In-House lab) is used to generate the "reference" values for comparison.
    • Certified Reference Materials: Calibrators and controls with precisely known concentrations are used to calibrate and verify the accuracy of the instruments.

    The qualifications of personnel operating these instruments are typically trained medical technologists or clinical laboratory scientists, but they do not establish "ground truth" in the way an expert radiologist might interpret an image.

    4. Adjudication Method for the Test Set

    Not applicable for this type of in vitro diagnostic device study. Adjudication methods (like 2+1, 3+1 consensus) are used for subjective interpretations, such as medical image analysis, where human experts might disagree. For quantitative chemical measurements, the comparison is directly between numerical results from different instruments.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is an in vitro diagnostic device for quantitative chemical analysis, not an AI-assisted diagnostic tool that involves human interpretation of "cases" or "reads" in the way an MRMC study would evaluate.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the studies presented are essentially standalone performance evaluations of the ACE Alera Clinical Chemistry System, the ACE Glucose Reagent, and the ACE Ion Selective Electrode (ISE) Module. The tables show the performance characteristics (precision, method comparison, interference) of the device itself in generating quantitative results. Human involvement is limited to operating the instrument, performing quality control, and routine maintenance, not subjective interpretation of results. The output (e.g., glucose concentration) is a direct numerical value from the instrument.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the test set (the samples used in the method comparison studies) was established by comparison against a legally marketed predicate device, the Alfa Wassermann ACE system (specifically the ACE plus ISE/Clinical Chemistry System, K930140, K933862), effectively treating the predicate device's measurements as the reference standard. This is a common approach for demonstrating substantial equivalence for new IVD devices.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" in the context of a machine learning algorithm. For clinical chemistry analyzers, the "training" analogous to machine learning would be:

    • Instrument Calibration: The device is calibrated using commercially available calibrator solutions with known concentrations. The specific number of calibration points is not detailed but is typically specified by the manufacturer.
    • Reagent Development and Optimization: The reagents themselves (like ACE Glucose Reagent) undergo extensive development and optimization, which involves testing on numerous samples to establish their performance characteristics (e.g., linearity, stability, interference). The exact sample sizes used during this development are not provided in this regulatory summary.

    9. How the Ground Truth for the Training Set Was Established

    As above, for an IVD analyzer, the "ground truth" for calibration or reagent development typically relies on:

    • Certified Reference Materials: These are materials with highly accurate and traceable analyte concentrations, used to set the instrument's measurement scale.
    • Validated Reference Methods: Established laboratory methods, often more complex or time-consuming, that are known to be highly accurate and precise for measuring the analyte.

    The document implies that the ground truth for comparison samples was the predicate ACE system, and it is reasonable to assume that the calibration and internal controls for the ACE Alera system would rely on industry-standard reference materials and methods to establish accurate known values.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122757
    Date Cleared
    2012-10-05

    (28 days)

    Product Code
    Regulation Number
    862.1175
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Cholesterol Reagent is intended for the quantitative determination of cholesterol concentration in serum and lithium heparin plasma using the ACE Axcel Clinical Chemistry System. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE HDL-C Reagent is intended for the quantitative determination of high density lipoprotein cholesterol (HDL-C) concentration in serum and lithium heparin plasma using the ACE Axcel Clinical Chemistry System. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE LDL-C Reagent is intended for the quantitative determination of low density lipoprotein cholesterol (LDL-C) concentration in serum and lithium heparin plasma using the ACE Axcel Clinical Chemistry System. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Triglycerides Reagent is intended for the quantitative determination of triglyceride concentration in serum and lithium heparin plasma using the ACE Axcel Clinical Chemistry System. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Cholesterol Reagent is composed of a single reagent bottle. The reagent contains 4-aminoantipyrine, p-hydroxybenzoic acid, cholesterol oxidase, cholesterol esterase and peroxidase.

    The HDL-C Reagent assay utilizes two reagent bottles, the second containing a unique detergent. This detergent solubilizes only the HDL lipoprotein particles, thus releasing HDL cholesterol to react with the cholesterol esterase and cholesterol oxidase, in the presence of a chromogen to produce color. The detergent also inhibits the reaction of the cholesterol enzymes with LDL, VLDL and chylomicron lipoproteins by adsorbing to their surfaces. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 592/692 nm, is directly proportional to the HDL cholesterol concentration in the sample.

    In the ACE LDL-C Reagent assay, detergent 1 solubilizes non-LDL lipoprotein particles (HDL, VLDL and chylomicrons) and releases cholesterol. The cholesterol is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. In a second reaction, detergent 2 solublizes the remaining LDL particles and forms peroxide, via the enzymes cholesterol esterase and cholesterol oxidase. The peroxide, in the presence of peroxidase and two peroxidase substrates, 4-aminoantipyrine and DSBmT, results in a purple-red color. The amount of color formed, determined by measuring the increase in absorbance bichromatically at 544/692 nm, is directly proportional to the LDL cholesterol concentration in the sample.

    In the ACE Triglycerides Reagent assay, triglycerides in serum are hydrolyzed by lipase to form glycerol and free fatty acids. In the presence of adenosine triphosphate (ATP) and glycerol kinase, the glycerol is converted to glycerol-1-phosphate and the ATP to adenosine diphosphate. Glycerol-1-phosphate is oxidized by glycerol phosphate oxidase to yield hydrogen peroxide. The hydrogen peroxide then acts to oxidatively couple p-chlorophenol and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 505 nm/692 nm, is directly proportional to the triglycerides concentration in the sample.

    AI/ML Overview

    The provided text describes a 510(k) submission for the ACE Axcel Clinical Chemistry System and its associated reagents for Cholesterol, HDL-C, LDL-C, and Triglycerides. The submission focuses on demonstrating substantial equivalence to a predicate device (K113262) by showing that the new device has "Same" intended use, instrument platform, basic principle, and reagent composition, with the only difference being the expanded sample type (serum and lithium heparin plasma for the candidate device vs. serum only for the predicate device).

    The acceptance criteria are implicitly defined by the performance characteristics demonstrated in the study, which aim to show that the expanded sample type (lithium heparin plasma) does not negatively impact the accuracy and precision of the measurements compared to serum. The study largely relies on analytical performance data rather than clinical outcomes or expert consensus on interpretations.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as numerical targets in the document; instead, the study intends to demonstrate comparable performance to the predicate device and acceptable analytical characteristics. The reported device performance for precision and matrix comparison is provided below, which implicitly became the "accepted" performance for the expanded sample type.

    AnalyteMetric / Acceptance Criteria (Implied: Acceptable analytical performance and comparability)Reported Device Performance (Precision)Reported Device Performance (Matrix Comparison: Serum vs. Plasma)
    CholesterolPrecision (SD, %CV) at various concentrations for serum and plasmaSerum: Low: 2.4, 1.6%; Mid: 3.6, 1.4%; High: 6.8, 1.3%
    Plasma: Low: 2.7, 2.1%; Mid: 4.1, 1.2%; High: 7.9, 1.4%Slope: 0.987, Intercept: -1.9, Correlation: 0.9987 (54 pairs)
    HDL-CPrecision (SD, %CV) at various concentrations for serum and plasmaSerum: Low: 2.0, 4.3%; Mid: 2.0, 2.6%; High: 2.4, 2.2%
    Plasma: Low: 1.3, 3.1%; Mid: 1.2, 1.7%; High: 2.7, 2.6%Slope: 1.011, Intercept: -1.1, Correlation: 0.9981 (53 pairs)
    LDL-CPrecision (SD, %CV) at various concentrations for serum and plasmaSerum: Low: 2.4, 2.6%; Mid: 3.7, 2.3%; High: 7.1, 2.1%
    Plasma: Low: 1.8, 2.3%; Mid: 5.6, 2.6%; High: 9.6, 2.6%Slope: 1.006, Intercept: -1.6, Correlation: 0.9981 (54 pairs)
    TriglyceridesPrecision (SD, %CV) at various concentrations for serum and plasmaSerum: Low: 1.4, 2.1%; Mid: 3.4, 1.0%; High: 4.3, 0.7%
    Plasma: Low: 2.2, 3.2%; Mid: 3.5, 1.0%; High: 13.5, 2.3%Slope: 0.992, Intercept: -3.6, Correlation: 0.9993 (55 pairs)

    2. Sample size used for the test set and the data provenance

    • Precision/Reproducibility Study (Test Set):
      • For each analyte (Cholesterol, HDL-C, LDL-C, Triglycerides), for both serum and plasma, 3 levels of samples were used.
      • Each level was tested with 2 replicates, twice a day, on 5 separate days, yielding a total of 20 replicates per level (3 levels * 2 sample types * 20 replicates/level = 120 total measurements per analyte category, e.g., Cholesterol on Serum).
      • Data Provenance: Not explicitly stated, but typically these studies are conducted in a laboratory setting, likely in the US (given the FDA submission). It is a prospective analytical study designed to evaluate device performance under controlled conditions.
    • Matrix Comparison Study (Test Set):
      • Cholesterol: 54 paired serum and lithium heparin plasma specimens.
      • HDL-C: 53 paired serum and lithium heparin plasma specimens.
      • LDL-C: 54 paired serum and lithium heparin plasma specimens.
      • Triglycerides: 55 paired serum and lithium heparin plasma specimens.
      • These specimens covered the assay's dynamic range.
      • Data Provenance: Not explicitly stated, but likely from a clinical laboratory setting, potentially within the US. The samples are retrospective specimens collected for analytical comparison.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of submission for in vitro diagnostic reagents does not typically involve human experts establishing "ground truth" through interpretation. The "ground truth" for the test set is established by comparative measurements against a reference method or the predicate device, and by the inherent chemical/physical properties of the samples used in reproducibility studies. No information about experts or their qualifications is provided or relevant in this context.

    4. Adjudication method for the test set

    Not applicable. This is an analytical performance study for laboratory reagents, not a clinical study involving human interpretation that would require an adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/imaging device, nor does it involve human readers or case interpretations. It is an in vitro diagnostic reagent.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is an analytical device, and its performance is inherently standalone in terms of generating a quantitative result. The results are then interpreted by clinicians in the overall diagnostic process. The study evaluates the standalone performance of the reagents on the ACE Axcel Clinical Chemistry System.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this type of analytical validation is established by:

    • Reference methods and/or the predicate device: For the matrix comparison, the serum measurements on the candidate device (which is substantially equivalent to the predicate) serve as the reference against plasma measurements. The predicate device's performance also implicitly serves as a benchmark for comparison.
    • Known concentrations: For precision studies, samples are "clinically relevant decision levels" meaning they have known or well-characterized concentrations of the analytes. These concentrations are typically determined by highly accurate laboratory methods.

    8. The sample size for the training set

    Not applicable. This is not an AI or machine learning device that requires a training set. The reagents are chemical formulations, and the system is an automated analyzer.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113389
    Date Cleared
    2012-08-10

    (268 days)

    Product Code
    Regulation Number
    862.1770
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum using the ACE Axcel Clinical Chemistry System. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum using the ACE Axcel Clinical Chemistry System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum using the ACE Axcel Clinical Chemistry System. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum using the ACE Axcel Clinical Chemistry System. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE BUN/Urea Reagent assay, urea in serum is hydrolyzed to yield ammonia and carbon dioxide in the presence of urease. The ammonia formed then reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase to yield glutamate and NAD. Two moles of NADH are oxidized for each mole of urea present. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The initial rate of decrease in absorbance, monitored bichromatically at 340 nm/647 nm, is proportional to the urea concentration in the sample.

    In the ACE Creatinine Reagent assay, creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation, determined by measuring the increase in absorbance bichromatically at 505 nm/573 nm during a fixed time interval, is directly proportional to the creatinine concentration in the sample.

    In the ACE Uric Acid Reagent assay, uric acid in serum is oxidized by uricase to allantoin and hydrogen peroxide. The hydrogen peroxide then acts to oxdatively couple dichlorohydroxybenzene sulfonic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex, which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 505 nm/610 nm, is directly proportional to the uric acid concentration in the sample.

    In the ACE CK Reagent assay, serum creatine kinase initiates the conversion of creatine phosphate to creatine with the transfer of a phosphate group to adenosine diphosphate (ADP), forming ATP. The ATP is then used in the phosphorylation of D-glucose to form D-glucose-6-phosphate and ADP. This reaction is catalyzed by hexokinase. The enzyme glucose-6-phosphate dehydrogenase catalyzes the reduction of D-glucose-6-phosphate and nicotinamide adenine dinucleotide phosphate (NADP+). The series of reactions triggered by serum creatine kinase and ending in the formation of NADPH. NADPH strongly absorbs at 340 nm, whereas NADP+ does not. Therefore, the rate of conversion of NADP+ to NADPH can be determined by monitoring the increase in absorbance bichromatically at 340 nm/378 nm. This rate of conversion from NADP+ to NADPH is a function of the activity of CK in the sample.

    The ACE BUN/Urea Reagent consists of a single reagent bottle. The reagent contains alpha-ketoglutarate, urease, glutamate dehydrogenase, adenosine diphosphate (ADP), nicotinamide adenine dinucleotide and reduced (NADH).

    The ACE Creatinine Reagent consists of two reagent bottles (Sodium Hydroxide Reagent and Picric Acid Reagent). The Sodium Hydroxide Reagent (R1) contains sodium hydroxide. The Picric Acid Reagent (R2) contains picric Acid.

    The ACE Uric Acid Reagent consists of a single reagent bottle. The reagent contains 4-aminoantipyrine, dichlorohydroxybenzene sulfonic acid, peroxidase and uricase.

    The ACE CK Reagent consists of two reagent bottles (Buffer and Substrate). The Buffer Reagent (R1) contains: imidazole buffer, glucose, N-acetyl-cysteine, magnesium acetate, EDTA, NADP and hexokinase. The Substrate Reagent (R2) contains: creatine phosphate, ADP, AMP, diadenosine pentaphosphate, EDTA and glucose-6-phosphate dehydrogenase.

    AI/ML Overview

    This 510(k) summary describes the analytical performance of the Alfa Wassermann ACE BUN, Creatinine, Uric Acid, and CK Reagents when used with the ACE Axcel Clinical Chemistry System. The study aims to demonstrate substantial equivalence to a predicate device by evaluating precision, accuracy, and detection limits.

    1. Table of Acceptance Criteria (Implied) and Reported Device Performance

    The acceptance criteria for this type of device are generally understood to be that the performance of the new device (ACE Axcel System with new reagents) should be comparable to or better than a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System). While explicit numerical acceptance criteria are not strictly stated as "acceptance criteria" but rather as "reported performance," the goal is to show the device performs within acceptable analytical limits for clinical chemistry assays and is strongly correlated with the predicate.

    Reagent (Analyte)Performance MetricImplied Acceptance Criteria (Comparison to Predicate)Reported Device Performance (ACE Axcel vs. ACE Clinical Chemistry System)
    ACE BUN/UreaPrecision (Within-run CV)0.975 (strong correlation)0.9963 (lab), 0.9982 to 0.9988 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)0.995 to 1.028 (lab), 0.983 to 1.039 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -5 to 5)-0.3 to 0.6 (lab), -0.7 to 1.6 (POL)
    Detection LimitClinically relevant low level1.1 mg/dL
    ACE CreatininePrecision (Within-run CV)0.975 (strong correlation)0.9998 (lab), 0.9994 to 0.9998 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)0.975 to 0.983 (lab), 0.961 to 1.027 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -0.1 to 0.1)-0.022 to 0.010 (lab), -0.136 to 0.001 (POL)
    Detection LimitClinically relevant low level0.19 mg/dL
    ACE Uric AcidPrecision (Within-run CV)0.975 (strong correlation)0.9958 (lab), 0.9858 to 0.9961 (POL)
    Accuracy (Slope CI)Close to 1 (e.g., 0.95-1.05)1.023 to 1.060 (lab), 0.972 to 1.054 (POL)
    Accuracy (Intercept CI)Close to 0 (e.g., -0.5 to 0.5)-0.18 to 0.07 (lab), -0.31 to 0.28 (POL)
    Detection LimitClinically relevant low level1.13 mg/dL

    Note: Acceptance criteria are implied based on typical expectations for clinical chemistry assays and the intent to demonstrate substantial equivalence to a predicate device. Specific numerical targets for acceptance were not explicitly stated in the provided text, but the strong correlation and low CVs indicate meeting such criteria.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • ACE BUN/Urea Reagent:

      • Accuracy (Correlation Study): 113 samples (clinical laboratory), and patient correlation studies at three Physician Office Laboratory (POL) sites (number of samples not explicitly stated for POL, but implied to be sufficient for regression analysis).
      • Precision: Four BUN levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and Physician Office Laboratory (POL) settings are mentioned, suggesting human serum samples. Whether these were retrospective or prospective is not specified, but typically, method comparison studies use prospective or collected retrospective clinical samples.
    • ACE Creatinine Reagent:

      • Accuracy (Correlation Study): 136 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
      • Precision: Four creatinine levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.
    • ACE Uric Acid Reagent:

      • Accuracy (Correlation Study): 106 samples (clinical laboratory), and patient correlation studies at three POL sites (number of samples not explicitly stated for POL).
      • Precision: Four uric acid levels over 22 days (laboratory study), and three POL sites over 5 days (levels not specified for POL).
      • Data Provenance: Not explicitly stated, but clinical laboratory and POL settings are mentioned, suggesting human serum samples.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is an in vitro diagnostic (IVD) device for quantitative measurement of chemical analytes (BUN, Creatinine, Uric Acid, CK) in serum. The 'ground truth' for such devices is established by a reference method or a legally marketed predicate device, not by expert interpretation of images or clinical findings.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is an IVD device for quantitative chemical analysis. Adjudication methods are typically used for qualitative or interpretive diagnostic devices where human expert disagreement might occur (e.g., radiology, pathology). Here, the comparison is directly numerical between the candidate device and the predicate device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is an IVD device for laboratory chemical analysis, not an imaging or interpretive diagnostic device that involves human readers or AI assistance in interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The described studies evaluate the performance of the algorithm/system only (the ACE Axcel Clinical Chemistry System with the new reagents) in quantifying the analytes in serum. The performance data (precision, accuracy, detection limit) are intrinsic to the device's analytical capability, independent of human interpretation of the results for the purpose of generating the values themselves. While trained personnel operate the system, the analytical performance is measured as a standalone function of the device.

    7. The Type of Ground Truth Used

    The "ground truth" for the accuracy studies was established by comparing the results from the Alfa Wassermann ACE Axcel Clinical Chemistry System (the new device, 'y') to a legally marketed predicate device, the Alfa Wassermann ACE Clinical Chemistry System ('x'). This is a common method for IVD substantial equivalence, where the predicate is considered the accepted reference for performance. For detection limits, it would typically involve analyzing samples with known, very low concentrations of the analytes or diluting higher concentration samples to determine the lowest measurable level.

    8. The Sample Size for the Training Set

    The provided text describes performance validation studies, not the development or training of an algorithm in the machine learning sense. Therefore, there is no "training set" for an algorithm to learn from in this context. The study focuses on verifying the performance of the already-developed reagent and instrument system.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the machine learning sense for this type of IVD device submission.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative determination of constituents in blood and other fluids.

    ACE Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of hemoglobin A1c (µmol/L) and total hemoglobin (g/dL) in human EDTA whole blood for the calculation of percent hemoglobin A1c using the ACE Axcel Clinical Chemistry System. The test is intended for use in clinical laboratories or physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only.

    The ACE CEDIA T Uptake homogenous enzyme immunoassay is intended for the quantitative determination of unoccupied binding sites of thyroxine-binding proteins in serum using the ACE Axcel Clinical Chemistry System. Measurements of triiodothyronine uptake are used in the diagnosis and treatment of thyroid disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE T4 Reagent is intended for the quantitative determination of total thyroxine (T4) concentration in serum using the ACE Axcel Clinical Chemistry System. Total thyroxine measurements are used in the diagnosis and treatment of thyroid diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Ferritin Reagent is intended for the quantitative determination of ferritin concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency anemia. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    Prior to the ACE Hemoglobin A1c (HbA1c) Reagent assay, whole blood samples require a pretreatment step, which is done on-board the analyzer. The red blood cells in the sample are lysed by the Hemoglobin Denaturant and the hemoglobin chains are hydrolyzed. For determination of HbA1c, a latex agglutination inhibition assay is used. In the absence of HbA1c in the sample, the agglutinator (synthetic polymer containing the immunoreactive portion of HbA1c) in the HbA1c Agglutinator Reagent and the antibody-coated microparticles in the HbA1c Antibody Reagent will agglutinate. The presence of HbA1c in the sample competes for the antibody binding sites and inhibits agglutination. The increase in absorbance, monitored monochromatically at 592 nm, is inversely proportional to the HbA1c present in the sample. For the determination of total hemoglobin, all hemoglobin derivatives in the sample are converted to alkaline hematin. The reaction produces a green colored solution, which is measured bichromatically at 573 nm/692 nm. The intensity of color produced is directly proportional to the total hemoglobin concentration in the sample. The concentrations of both HbA1c and total hemoglobin are measured, the ratio is calculated and the result reported as percent HbA1c.

    The CEDIA T Uptake assay uses recombinant DNA technology to produce a unique homogeneous enzyme immunoassay system. The assay is based the bacterial enzyme β-galactosidase, which has been genetically engineered into two inactive fragments. These fragments spontaneously re-associate to form fully active enzyme which, in the assay format, cleaves a substrate, generating a color change that can be measured spectrophotometrically. In the assay, enzyme donor thyroxine conjugate binds directly to the unoccupied thyroxine-binding sites in the sample, preventing the spontaneous re-association of the enzyme fragments to form the active enzyme. Thus, thyroxine-binding proteins regulate the amount of β-galactosidase formed from the reassembly of the remaining donor and enzyme acceptor as monitored by the hydrolysis of the substrate o-nitrophenyl-β-galactopyranoside.

    The ACE T4 Assay is a homogeneous enzyme immunoassay using ready-to-use liquid ACE T4 Reagent. The assay uses 8-anilino-1-naphthalene sulfonic acid (ANS) to dissociate thyroxine from the plasma binding proteins. Using specific antibodies to thyroxine, this assay is based on the competition of glucose-6-phosphate dehydrogenase (G6PD) labeled thyroxine and the dissociated thyroxine in the sample for a fixed amount of specific antibody binding sites. In the absence of thyroxine from the sample, the thyroxine labeled G6PD in the second reagent is bound by the specific antibody in the first reagent, inhibiting the enzyme's activity. The enzyme G6PD catalyzes the oxidation of glucose-6-phosphate (G6P) with nicotinamide adenine dinucleotide (NAD+) to form 6-phosphogluconate and reduced nicotinamide adenine dinucleotide (NADH). NADH strongly absorbs at 340 nm whereas NAD+ does not. The rate of conversion, determined by measuring the increase in absorbance bichromatically at 340 nm/505 nm during a fixed time interval, is directly proportional to the amount of thyroxine in the sample. The concentration of thyroxine is determined automatically by the ACE Clinical Chemistry System using a logarithmic calibration curve established with calibrators, which are provided separately.

    In the Ferritin Assay, serum ferritin, in the presence of anti-ferritin conjugated latex micorparticles, and a buffer promoting aggregation, initiates an antigen-antibody reaction, resulting in the agglutination of the latex microparticles. The agglutination is detected turbidometrically by an absorbance change measured at a wavelength of 592 nm. The magnitude of the absorbance change is proportional to the ferritin concentration in the sample.

    AI/ML Overview

    The provided text is a 510(k) summary for the Alfa Wassermann Diagnostic Technologies ACE Axcel Clinical Chemistry System and several associated reagents. It describes the devices, their intended uses, and technological characteristics. However, the document does not contain any information about acceptance criteria or a study that proves the device meets specific acceptance criteria.

    The content of the document focuses on:

    • Identification of the device and reagents: Trade names, classifications, common names, and product codes.
    • Predicate devices: Listing the previously approved systems and reagents used for comparison in the 510(k) submission.
    • Device descriptions: Detailed explanations of the ACE Axcel Clinical Chemistry System's functionality and the biochemical principles of each reagent (HbA1c, CEDIA T Uptake, T4, Ferritin).
    • Intended Use/Indications for Use: What each device/reagent is designed to measure and for what clinical purpose.
    • Technological Characteristics: Specifications of the analyzer (throughput, reagent capacity, cooling, sample handling, optical system).
    • Regulatory approval notice: A letter from the FDA indicating substantial equivalence.

    Therefore, I cannot provide a table of acceptance criteria or details of a study proving the device meets those criteria from the provided text. The requested information about sample sizes, data provenance, expert qualifications, ground truth, MRMC studies, or standalone performance studies is not present in this document.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum using the ACE Axcel Clinical Chemistry System. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to form a green colored complex, which is measured bichromatically at 629 nm/692 nm. The intensity of color produced is directly proportional to the albumin concentration in the sample.

    In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of proteins under alkaline conditions to form a violet colored complex which is measured bichromatically at 544 nm/692 nm. The intensity of color produced is directly proportional to the total protein concentration in the sample.

    In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an acidic solution to form a blue-purple colored complex, which is measured bichromatically at 647 nm/692 nm. The intensity of color produced is directly proportional to the calcium concentration in the sample.

    In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions, inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the amount of phosphorus in the sample.

    The ACE Albumin Reagent consists of a single reagent bottle. The reagent contains Bromcresol green and acetate buffer.

    The ACE Total Protein Reagent consists of a single reagent bottle. The reagent contains copper sulfate, sodium potassium tartrate, potassium iodide and sodium hydroxide.

    The ACE Calcium-Arsenazo Reagent consists of a single reagent bottle. The Reagent contains Arsenazo III.

    The ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent bottle. The reagent contains ammonium molybdate and sulfuric acid.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for ACE Reagents

    The provided 510(k) summary (K113374) describes the performance of four reagents: ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, and ACE Inorganic Phosphorus U.V. Reagent, when used with the Alfa Wassermann ACE Axcel Clinical Chemistry System. The study establishes the substantial equivalence of these devices to their predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated in numerical terms (e.g., "CV must be

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative determination of constituents in blood and other fluids.

    The ACE Cholesterol Reagent is intended for the quantitative determination of cholesterol concentration in serum using the ACE Axcel Clinical Chemistry System. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE HDL-C Reagent is intended for the quantitative determination of high density lipoprotein cholesterol (HDL-C) concentration in serum using the ACE Axcel Clinical Chemistry System. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE LDL-C Reagent is intended for the quantitative determination of low density lipoprotein cholesterol (LDL-C) concentration in serum using the ACE Axcel Clinical Chemistry System. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Triglycerides Reagent is intended for the quantitative determination of triglyceride concentration in serum using the ACE Axcel Clinical Chemistry System. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism or various endocrine disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE Cholesterol Reagent assay, cholesterol esters in serum are completely hydrolyzed by cholesterol esterase to free cholesterol and free fatty acids. The cholesterol liberated by the esterase, plus any endogenous free cholesterol, are both oxidized by cholesterol oxidase to yield hydrogen peroxide. The hydrogen peroxide then acts to oxidatively couple p-hydroxybenzoic acid and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance, bichromatically at 505 nm/647 nm, is directly proportional to the cholesterol concentration in the sample.

    The HDL-C Assay utilizes two reagents, the second containing a unique detergent. This detergent solubilizes only the HDL lipoprotein particles, thus releasing HDL cholesterol to react with the cholesterol esterase and cholesterol oxidase, in the presence of a chromogen to enzymes with LDL, VLDL and chylomicron lipoproteins by adsorbing to their surfaces. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 592/692 nm, is directly proportional to the HDL cholesterol concentration in the sample.

    In the ACE LDL-C Reagent assay, detergent 1 solubilizes non-LDL lipoprotein particles (HDL, VLDL and chylomicrons) and releases cholesterol. The cholesterol is consumed by cholesterol esterase and cholesterol oxidase in a non-color forming reaction. In a second reaction, detergent 2 solublizes the remaining LDL particles and forms peroxide, via the enzymes cholesterol esterase and cholesterol oxidase. The peroxide, in the presence of peroxidase and two peroxidase sub- strates, 4-aminoantipyrine and DSBmT, results in a purple-red color. The amount of color formed, determined by measuring the increase in absorbance bichromatically at 544/692 nm, is directly proportional to the LDL cholesterol concentration in the sample.

    In the ACE Triglycerides Reagent Assay, triglycerides in serum are hydrolyzed by lipase to form glycerol and free fatty acids. In the presence of adenosine triphosphate (ATP) and glycerol kinase, the glycerol is converted to glycerol-1-phosphate and the ATP to adenosine diphosphate. Glycerol-1-phosphate is oxidized by glycerol phosphate oxidase to yield hydrogen peroxide. The hydrogen peroxide then acts to oxidatively couple p-chlorophenol and 4-aminoantipyrine in a reaction catalyzed by peroxidase, producing a red colored quinoneimine complex which absorbs strongly at 505 nm. The amount of chromogen formed, determined by measuring the increase in absorbance bichromatically at 505 nm/692 nm, is directly proportional to the triglycerides concentration in the sample.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Alfa Wassermann ACE Axcel Clinical Chemistry System and its associated reagents, based on the provided 510(k) summary:

    Overview of Device and Study Purpose:

    This submission describes the ACE Axcel Clinical Chemistry System and four associated reagents (ACE Cholesterol Reagent, ACE HDL-C Reagent, ACE LDL-C Reagent, ACE Triglycerides Reagent). The study's primary purpose is to demonstrate the substantial equivalence of the new ACE Axcel System and its reagents to predicate devices (Alfa Wassermann ACE plus ISE/Clinical Chemistry System and its reagents) by showing comparable performance in terms of precision, accuracy, and detection limits.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the comparison to the predicate device and the typical expectations for clinical chemistry analyzers. The document focuses on demonstrating that the new system's performance is equivalent or better than the predicate.

    Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (ACE Axcel Clinical Chemistry System with specific reagents)
    ACE Cholesterol Reagent
    Precision (Within-run CV)Acceptably low CV, comparable to predicate, demonstrating consistency.Ranged from 1.3 to 2.0% (lab), 0.7 to 1.5% (POL sites).
    Precision (Total CV)Acceptably low CV, comparable to predicate, demonstrating consistency over time.Ranged from 1.6 to 2.2% (lab), 1.0 to 1.7% (POL sites).
    Accuracy (Correlation Coeff.)High correlation (e.g., >0.98) with predicate device, indicating agreement.0.9977 (lab), 0.9945 to 0.9995 (POL sites).
    Accuracy (Slope CI)Confidence interval for slope close to 1.0, indicating proportional agreement.0.999 to 1.026 (lab), 0.964 to 1.034 (POL sites).
    Accuracy (Intercept CI)Confidence interval for intercept close to 0, indicating minimal constant bias.-6.2 to -0.5 (lab), -7.3 to 6.7 (POL sites).
    Detection LimitClinically acceptable lower limit of detection.3.6 mg/dL.
    ACE HDL-C Reagent
    Precision (Within-run CV)Acceptably low CV.Ranged from 1.4 to 2.7% (lab), 0.7 to 2.6% (POL sites).
    Precision (Total CV)Acceptably low CV.Ranged from 3.2 to 4.8% (lab), 1.1 to 3.5% (POL sites).
    Accuracy (Correlation Coeff.)High correlation with predicate.0.9959 (lab), 0.9898 to 0.9970 (POL sites).
    Accuracy (Slope CI)Confidence interval for slope close to 1.0.0.956 to 0.990 (lab), 0.936 to 1.061 (POL sites).
    Accuracy (Intercept CI)Confidence interval for intercept close to 0.-0.5 to 1.4 (lab), -3.8 to 2.0 (POL sites).
    Detection LimitClinically acceptable lower limit of detection.1.5 mg/dL.
    ACE LDL-C Reagent
    Precision (Within-run CV)Acceptably low CV.Ranged from 2.5 to 4.6% (lab), 1.7 to 4.4% (POL sites).
    Precision (Total CV)Acceptably low CV.Ranged from 3.2 to 4.9% (lab), 2.4 to 5.9% (POL sites).
    Accuracy (Correlation Coeff.)High correlation with predicate.0.9973 (lab), 0.9940 to 0.9974 (POL sites).
    Accuracy (Slope CI)Confidence interval for slope close to 1.0.0.968 to 0.996 (lab), 0.991 to 1.071 (POL sites).
    Accuracy (Intercept CI)Confidence interval for intercept close to 0.-3.1 to 1.0 (lab), -8.4 to 4.0 (POL sites).
    Detection LimitClinically acceptable lower limit of detection.4.0 mg/dL.
    ACE Triglycerides Reagent
    Precision (Within-run CV)Acceptably low CV.Ranged from 1.2 to 2.9% (lab), 0.5 to 2.3% (POL sites).
    Precision (Total CV)Acceptably low CV.Ranged from 1.8 to 3.2% (lab), 0.6 to 4.1% (POL sites).
    Accuracy (Correlation Coeff.)High correlation with predicate.0.9995 (lab), 0.9992 to 0.9996 (POL sites).
    Accuracy (Slope CI)Confidence interval for slope close to 1.0.1.025 to 1.037 (lab), 0.989 to 1.024 (POL sites).
    Accuracy (Intercept CI)Confidence interval for intercept close to 0.-2.7 to 1.1 (lab), -6.3 to 0.7 (POL sites).
    Detection LimitClinically acceptable lower limit of detection.11.6 mg/dL.

    2. Sample Size Used for the Test Set and Data Provenance

    • ACE Cholesterol Reagent (Accuracy): 110 samples.
    • ACE HDL-C Reagent (Accuracy): 109 samples.
    • ACE LDL-C Reagent (Accuracy): 108 samples.
    • ACE Triglycerides Reagent (Accuracy): 111 samples.

    The samples for the accuracy ("correlation") studies were assayed on both the new ACE Axcel System (y) and the predicate Alfa Wassermann ACE Clinical Chemistry System (x).

    Data Provenance: The document does not explicitly state the country of origin. The studies were conducted at a main lab and three separate Physician Office Laboratory (POL) sites, suggesting real-world clinical samples. There is no indication of whether the data was retrospective or prospective, but given the nature of a comparability study for a new device, it is typically prospective, involving fresh samples run on both systems simultaneously.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This type of device (clinical chemistry analyzer) does not typically involve human experts establishing ground truth in the way radiological or pathological devices do. The "ground truth" for the test set is established by the predicate device (Alfa Wassermann ACE Clinical Chemistry System) itself, which is assumed to be accurate and clinically acceptable. The study's goal is to show agreement with this established method. Therefore, no external human experts are used for ground truth establishment for individual samples.


    4. Adjudication Method for the Test Set

    Not applicable. As explained above, this is a quantitative comparison study against a predicate device, not an interpretation-based study requiring expert adjudication of results. The predicate device's readings serve as the comparator.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a study for an automated clinical chemistry analyzer which provides quantitative measurements, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.


    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the studies presented are generally "standalone" in the sense that they evaluate the performance of the automated ACE Axcel Clinical Chemistry System (device and reagents) directly. The system provides quantitative results, and the performance metrics (precision, accuracy, detection limit) are intrinsic to the device's operation. There isn't a human-in-the-loop component being evaluated for diagnostic decision-making, rather the system is automated to provide a numerical result.


    7. The Type of Ground Truth Used

    The ground truth used for accuracy (correlation) studies is the measurement obtained from the predicate device, the Alfa Wassermann ACE Clinical Chemistry System. This is a common approach for demonstrating substantial equivalence for clinical chemistry analyzers, where the new device's performance is compared against another legally marketed device's performance.


    8. The Sample Size for the Training Set

    Not applicable. This document describes the validation of a clinical chemistry system and its reagents, not an AI/ML algorithm that typically requires a separate training set. The "training" for such systems would involve chemical formulation and instrument calibration, not data-driven machine learning.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The calibration of the instrument and formulation of reagents are based on established chemical and engineering principles rather than data-driven ground truth for machine learning.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative determination of constituents in blood and other fluids.

    The ACE TIBC Reagent is intended for the quantitative determination of total iron-binding capacity in serum using the ACE Axcel Clinical Chemistry System. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Serum Iron Reagent is intended for the quantitative determination of iron concentration in serum using the ACE Axcel Clinical Chemistry System. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    The ACE Lipase Reagent is intended for the quantitative determination of lipase activity in serum using the ACE Axcel Clinical Chemistry System. Lipase measurements are used in diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE Direct Total Iron-Binding Capacity (TIBC) Reagent assay, Direct TIBC Color Reagent, an acidic buffer containing an iron-binding dye and ferric chloride, is added to the serum sample. The low pH of Direct TIBC Color Reagent releases iron from transferrin. The iron then forms a colored complex with the dye. The colored complex at the end of the first step represents both the serum iron and excess iron already present in Direct TIBC Color Reagent. Direct TIBC Buffer, a neutral buffer, is then added, shifting the pH and resulting in a large increase in the affinity of transferrin for iron. The serum transferrin rapidly binds the iron by abstracting it from the dye-iron complex. The observed decrease in absorbance of the colored dye-iron complex is directly proportional to the total iron-binding capacity of the serum sample. The absorbance is measured at 647 nm.

    In the ACE Serum Iron Reagent assay, transferrin-bound iron in serum is released at an acidic pH and reduced from ferric to ferrous ions. These ions react with ferrozine to form a violet colored complex, which is measured bichromatically at 554 nm/692 nm. The intensity of color produced is directly proportional to the serum iron concentration.

    In the ACE Lipase Reagent Assay, serum lipase acts on a natural substrate, 1,2-diglyceride, to liberate 2-monoglyceride. This is hydrolyzed by monoglyceride lipase (a highly specific enzyme for monoglyceride) into glycerol and free fatty acid. Glycerol kinase acts on glycerol to form glycerol-3-phosphate, which is in turn acted on by glycerol-3-phosphate oxidase to generate hydrogen peroxide. Peroxidase converts the hydrogen peroxide, 4-Aminoantipyrine and TOOS (N-ethyl-N-(2-hydroxy-3-sulfopropyl)-m-toluidine) into a quinine dye. The rate of formation of the dye, determined bichromatically at an absorbance of 573 nm/692 nm, is proportional to the lipase activity in the sample.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ACE Direct Total Iron-Binding Capacity (TIBC) Reagent, ACE Serum Iron Reagent, and ACE Lipase Reagent, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Device/ParameterAcceptance Criteria (Implied)Reported Device Performance and Confidence Intervals
    ACE Direct TIBC Reagent
    PrecisionLow within-run and total CV for various TIBC levels.Lab Testing:
    • Within-run CV: 0.9% to 2.2%
    • Total CV: 2.0% to 3.3%
      POL Sites:
    • Within-run CV: 0.9% to 3.4%
    • Total CV: 0.9% to 4.1% |
      | Accuracy (Correlation to Predicate) | High correlation coefficient, low standard error, slope near 1, intercept near 0 when compared to predicate device. | Lab (109 samples):
    • Correlation Coefficient: 0.9950
    • Standard Error Estimate: 9.1
    • Confidence Interval Slope: 0.961 to 0.998
    • Confidence Interval Intercept: -9.2 to 4.3
      POL Sites:
    • Correlation Coefficients: 0.9902 to 0.9987
    • Standard Error Estimates: 6.1 to 11.2
    • Confidence Interval Slopes: 0.923 to 1.006
    • Confidence Interval Intercepts: -8.2 to 19.4 |
      | Detection Limit | Low enough to be clinically useful. | 42.21 µg/dL |
      | ACE Serum Iron Reagent |
      | Precision | Low within-run and total CV for various Serum Iron levels. | Lab Testing:
    • Within-run CV: 1.2% to 5.2%
    • Total CV: 1.3% to 5.4%
      POL Sites:
    • Within-run CV: 1.2% to 4.1%
    • Total CV: 1.2% to 4.2% |
      | Accuracy (Correlation to Predicate) | High correlation coefficient, low standard error, slope near 1, intercept near 0 when compared to predicate device. | Lab (130 samples):
    • Correlation Coefficient: 0.9995
    • Standard Error Estimate: 3.3
    • Confidence Interval Slope: 1.000 to 1.012
    • Confidence Interval Intercept: -2.7 to -1.0
      POL Sites:
    • Correlation Coefficients: 0.9992 to 0.9998
    • Standard Error Estimates: 6.1 to 11.2
    • Confidence Interval Slopes: 0.997 to 1.041
    • Confidence Interval Intercepts: -2.7 to 9.2 |
      | Detection Limit | Low enough to be clinically useful. | 5.08 µg/dL |
      | ACE Lipase Reagent |
      | Precision | Low within-run and total CV for various lipase levels. | Lab Testing:
    • Within-run CV: 1.1% to 6.5%
    • Total CV: 6.0% to 10.7%
      POL Sites:
    • Within-run CV: "to 7.3%" (lower bound not specified)
    • Total CV: 1.9% to 7.3% |
      | Accuracy (Correlation to Predicate) | High correlation coefficient, low standard error, slope near 1, intercept near 0 when compared to predicate device. | Lab (107 samples):
    • Correlation Coefficient: 0.9980
    • Standard Error Estimate: 9.06
    • Confidence Interval Slope: 0.970 to 0.994
    • Confidence Interval Intercept: 1.97 to 5.97
      POL Sites:
    • Correlation Coefficients: 0.9993 to 0.9997
    • Standard Error Estimates: 4.44 to 7.89
    • Confidence Interval Slopes: 1.002 to 1.047
    • Confidence Interval Intercepts: -4.74 to 3.41 |
      | Detection Limit | Low enough to be clinically useful. | 10.63 U/L |

    Note: The acceptance criteria are "implied" because the document primarily presents the results of the performance data without explicitly stating the pre-defined target values or ranges that were aimed for. However, the context of a 510(k) submission requires demonstrating substantial equivalence, meaning the performance should be comparable to the predicate device. Therefore, the reported data, particularly the high correlation coefficients, slopes near 1, and intercepts near 0 for accuracy, indicate that these outcomes met whatever internal acceptance criteria were set for demonstrating equivalency. For precision, low CVs are generally accepted as good performance.


    2. Sample Size Used for the Test Set and Data Provenance

    • ACE Direct Total Iron-Binding Capacity (TIBC) Reagent:

      • Sample Size:
        • Accuracy (correlation study): 109 samples
        • Precision (lab): 4 TIBC levels tested for 22 days.
        • Precision (POL sites): 3 separate POL sites, testing over 5 days (number of samples not specified, but likely multiple runs per site per day).
      • Data Provenance: Not explicitly stated, but the testing occurred at "Physician Office Laboratory (POL) sites" and an unnamed central lab. It is not specified if the data is retrospective or prospective, nor the country of origin.
    • ACE Serum Iron Reagent:

      • Sample Size:
        • Accuracy (correlation study): 130 samples
        • Precision (lab): 4 Serum Iron levels tested for 22 days.
        • Precision (POL sites): 3 separate POL sites, testing over 5 days (number of samples not specified).
      • Data Provenance: Not explicitly stated, but testing occurred at "Physician Office Laboratory (POL) sites" and an unnamed central lab. Retrospective or prospective nature and country of origin are not specified.
    • ACE Lipase Reagent:

      • Sample Size:
        • Accuracy (correlation study): 107 samples
        • Precision (lab): 3 lipase levels tested for 22 days.
        • Precision (POL sites): 3 separate POL sites, testing over 5 days (number of samples not specified).
      • Data Provenance: Not explicitly stated, but testing occurred at "Physician Office Laboratory (POL) sites" and an unnamed central lab. Retrospective or prospective nature and country of origin are not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given text. The ground truth for these types of in vitro diagnostic tests is typically established by measurements from a reference method or a predicate device. The text indicates that the "Alfa Wassermann ACE Clinical Chemistry System" was used as the comparator (predicate device) (referred to as 'x' in the regression analyses).


    4. Adjudication Method for the Test Set

    This information is not applicable and therefore, not provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving subjective interpretation, such as by human readers of medical images, to resolve discrepancies in diagnoses. These clinical chemistry devices produce quantitative numerical results, which are then compared statistically to a reference method or predicate device, rather than adjudicated.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable as the described devices are in vitro diagnostic clinical chemistry reagents and an automated system (ACE Axcel Clinical Chemistry System), not AI-assisted imaging or diagnostic tools designed for human readers to interpret. Therefore, an MRMC study and effects on human reader performance are not relevant to this submission.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are standalone performance studies of the device and reagents. The ACE Axcel Clinical Chemistry System is an "automated, discrete, bench-top, random access analyzer." The performance data presented (precision, accuracy, detection limit) are measurements of the system's ability to quantitatively determine analytes directly, without a human interpretation step that would introduce a "human-in-the-loop" component in the result generation itself. The results quantify the device's inherent measurement capabilities.


    7. The Type of Ground Truth Used

    The ground truth for these studies was established by comparison to a legally marketed predicate device, the "Alfa Wassermann ACE Clinical Chemistry System" (specifically, the ACE Reagents K000781, K944911 run on the K931786 system). This is a common method for demonstrating substantial equivalence for in vitro diagnostic devices in 510(k) submissions. The new device's measurements (y) were correlated against the predicate device's measurements (x).


    8. The Sample Size for the Training Set

    This information is not provided and is generally not applicable in the context of these types of in vitro diagnostic submissions for clinical chemistry reagents and analyzers. The device described does not employ a machine learning algorithm that requires a "training set" in the conventional sense. The "training" of such a device primarily involves rigorous internal calibration procedures and validation during its development and manufacturing, which are distinct from the concept of a "training set" for AI/ML models.


    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set" requiring ground truth establishment in this manner is not applicable to this type of device and submission. The device's operational parameters are set through design, engineering, and calibration processes, not machine learning model training.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    ALFA WASSERMANN DIAGNOSTIC TECHNOLOGIES, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative measurement of general chemistry assays for clinical use in physician office laboratories or clinical laboratories.

    The ACE Axcel Clinical System includes an Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    • Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance
    • Potassium measurements are used to monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
    • Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

    The ACE Glucose Reagent is intended for the quantitative determination of glucose concentration in serum using the ACE Axcel Clinical Chemistry System. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

    Device Description

    The ACE Axcel Clinical Chemistry System consists of two major components, the chemistry instrument and an integrated Panel PC. The instrument accepts the physical patient samples, performs the appropriate optical or potentiometric measurements on those samples and communicates that data to an integral Panel PC. The Panel PC uses keyboard or touch screen input to manually enter a variety of data, control and accept data from the instrument, manage and maintain system information and generate reports relative to patient status and instrument performance. The Panel PC also allows remote download of patient requisitions and upload of patient results via a standard interface.

    In the ACE Glucose Reagent assay, glucose in serum reacts with adenosine triphosphate in the presence of hexokinase and magnesium with the formation of glucose-6-phosphate and adenosine diphosphate. Glucose-6-phosphate dehydrogenase catalyzes the oxidation of glucose-6-phosphate with NAD+ to form 6-phosphogluconate and NADH. NADH absorbs strongly at 340 nm, whereas NAD+ does not. The total amount of NADH formed is proportional to the concentration of glucose in the sample. The increase in absorbance is measured bichromatically at 340 nm/378 nm.

    The ACE Ion Selective Electrode (ISE) Module is used with ACE CAL A and CAL B Calibration Solutions in the performance of a two-point calibration in order to measure concentrations of sodium, potassium and chloride in undiluted serum. The ISE module uses a potentiometric method to simultaneously measure sodium, potassium and chloride in undiluted serum. Each electrode uses an ion-specific membrane to measure the difference in ionic concentration between an inner electrolyte solution and the sample. This difference causes an electro-chemical potential to form on the membrane of the active electrode. The connection of the amplifier and ground (reference electrode) to the ion selective electrode forms the measuring system. The two-point calibration with CAL A and CAL B with precisely known ion concentrations (two-point calibration) and the measured voltage difference of the sample and CAL A are used to determine the ion concentration in the sample.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Alfa Wassermann ACE Axcel Clinical Chemistry System, ACE Ion Selective Electrode (ISE) Module, and ACE Glucose Reagent, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Accuracy" data, where the device (y) is compared to a predicate device (x). The close correlation, with slopes near 1 and intercepts near 0, along with high correlation coefficients, indicates acceptable accuracy. Precision is evaluated by the Coefficient of Variation (CV%).

    Measurement (Reagent/Module)Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (ACE Axcel Clinical Chemistry System)
    ACE Glucose ReagentPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)1.0 to 1.4% (lab setting, 4 levels, 22 days); 0.3 to 2.2% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.0 to 1.9% (lab setting); 0.5 to 2.2% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9998 (lab setting, 122 samples); 0.9992 to 0.9998 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.01.001 to 1.009 (lab setting); 0.972 to 1.021 (POL sites)
    Accuracy (Intercept)Close to 0-1.5 to 0.1 (lab setting); -3.0 to 4.1 (POL sites)
    Detection Limit(Not explicitly stated, but low is desirable)3.1 mg/dL
    ACE Axcel Sodium ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.4 to 1.0% (lab setting, 4 levels, 21 days); 0.6 to 1.0% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)0.8 to 1.4% (lab setting); 0.8 to 1.4% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9963 (lab setting, 113 samples); 0.9917 to 0.9995 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.992 to 1.024 (lab setting); 0.989 to 1.067 (POL sites)
    Accuracy (Intercept)Close to 0-3.60 to 0.92 (lab setting); -8.85 to 2.30 (POL sites)
    ACE Axcel Potassium ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.6 to 3.5% (lab setting, 4 levels, 21 days); 1.0 to 1.6% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.3 to 3.5% (lab setting); 1.1 to 1.6% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9974 (lab setting, 115 samples); 0.9973 to 0.9996 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.989 to 1.015 (lab setting); 0.960 to 1.035 (POL sites)
    Accuracy (Intercept)Close to 0-0.050 to 0.095 (lab setting); -0.194 to 0.216 (POL sites)
    ACE Axcel Chloride ISEPrecision (Within-run CV)(Not explicitly stated, but low CV is desirable)0.5 to 1.0% (lab setting, 4 levels, 21 days); 0.9 to 1.5% (POL sites, 3 locations, 5 days)
    Precision (Total CV)(Not explicitly stated, but low CV is desirable)1.1 to 1.5% (lab setting); 1.1 to 2.6% (POL sites)
    Accuracy (Correlation Coefficient)Close to 1.00.9855 (lab setting, 111 samples); 0.9885 to 0.9996 (POL sites, 3 locations)
    Accuracy (Slope)Close to 1.00.939 to 1.002 (lab setting); 0.976 to 1.088 (POL sites)
    Accuracy (Intercept)Close to 0-1.07 to 5.63 (lab setting); -8.16 to 2.22 (POL sites)

    2. Sample Size Used for the Test Set and Data Provenance

    • ACE Glucose Reagent:

      • Accuracy (Correlation Study): 122 samples (ranging from 6 to 729 mg/dL).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites. Specific sample numbers per POL site are not provided, but the combined sites yielded multiple correlation coefficients, standard error estimates, and confidence intervals for slope and intercept.
      • Provenance: Not explicitly stated, but likely from a laboratory setting and Physician Office Laboratories, presumably within the US given the submission to the FDA. The nature of the samples (e.g., patient samples, control materials) is not specified as prospective or retrospective.
    • ACE Axcel Sodium ISE:

      • Accuracy (Correlation Study): 113 samples (ranging from 45.1 to 194.0 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
    • ACE Axcel Potassium ISE:

      • Accuracy (Correlation Study): 115 samples (ranging from 1.57 to 14.20 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.
    • ACE Axcel Chloride ISE:

      • Accuracy (Correlation Study): 111 samples (ranging from 63.4 to 176.0 mmol/L).
      • Accuracy (Patient Correlation Studies): Data from three separate Physician Office Laboratory (POL) sites.
      • Provenance: Same as Glucose – likely US lab/POL, nature of samples not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. For this type of in vitro diagnostic device, "ground truth" is established by comparing the performance of the new device to a legally marketed predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents) using quantitative measurements, not by expert interpretation. The predicate device itself acts as the reference method in these correlation studies.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this is a quantitative comparison against a predicate device, not an interpretation-based ground truth requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a clinical chemistry analyzer, an automated system for measuring analytes in samples. It does not involve human readers interpreting images or data where AI assistance would be relevant in the context of MRMC studies.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described are inherently "standalone" in the sense that they evaluate the performance of the ACE Axcel Clinical Chemistry System (including its reagents and ISE module) in producing quantitative results independently. The accuracy studies compare its output directly against a predicate device's output. There isn't a "human-in-the-loop" aspect to the core measurement performance itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the accuracy studies is the quantitative result obtained from the predicate device (Alfa Wassermann ACE Clinical Chemistry System and ACE Reagents). The studies are correlation studies comparing the new device's measurements (y) to the predicate device's measurements (x).

    8. The Sample Size for the Training Set

    Not applicable. This document describes a traditional medical device (clinical chemistry analyzer), not a machine learning or AI-based device that typically has a "training set." The device is intended to perform measurements based on established chemical and electrochemical principles.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" in the context of this device's development as described in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3