WONDFO METHYLENEDIOXYMETHAMPHETAMINE URINE TEST, WONDFO MORPHINE URINE TEST
Applicant
Guangzhou Wondfo Biotech Co., Ltd.
Product Code
LAF · Clinical Toxicology
Decision Date
Sep 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Indications for Use
Wondfo Methylenedioxymethamphetamine Urine Test is an immunochromatographic assay for the qualitative determination of MDMA in human urine at a cutoff concentration of 500 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Wondfo Morphine Urine Test is an immunochromatographic assay for the qualitative determination of MDMA in human urine at a cutoff concentration of 300 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Device Story
Wondfo Methylenedioxymethamphetamine and Morphine Urine Tests are lateral flow chromatographic immunoassays for qualitative drug detection in human urine. Devices are available as dip cards or integrated urine cups. Principle of operation: competitive binding immunoassay; urine sample migrates via capillary action through a nitrocellulose membrane containing a test line (drug antigen-protein conjugate) and a control line (goat anti-mouse IgG). If drug is present at or above cutoff, no test line forms; if drug is absent or below cutoff, a colored band forms. Used in home or clinical settings; operated by lay users or healthcare professionals. Results are visually interpreted at 5 minutes. Preliminary positive results require confirmation via GC/MS. Benefits include rapid, point-of-care screening for drug abuse.
Clinical Evidence
No clinical diagnostic studies were performed. Evidence consists of analytical bench testing (precision, specificity, interference, stability) and a lay-user study (n=140 per format) comparing device results to GC/MS. Lay-user study demonstrated high agreement with GC/MS (96.4%–97.8% total agreement) and confirmed ease of use.
Indicated for qualitative detection of MDMA (500 ng/mL) or Morphine (300 ng/mL) in human urine. Intended for prescription and over-the-counter use. Provides preliminary results; requires confirmatory testing (e.g., GC/MS).
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Acon Laboratories, Inc. One Step Drug Screen Test (k020771)
Acon Laboratories, Inc. MDMA One Step Ecstasy Test (k022589)
Related Devices
K131232 — WONDFO METHADONE URINE TEST (MTD 200), WONDOFO MORPHINE URINE TEST (MOP 100) · Guangzhou Wondfo Biotech Co., Ltd. · May 31, 2013
K160793 — First Sign Drug of Abuse Dip Card Test (MDMA, EDDP, Nortriptyline), First Sign Drug of Abuse Cup Test (MDMA, EDDP, Nortriptyline) · W.H.P.M., Inc. · Aug 17, 2016
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k112236
B. Purpose for Submission:
New Device
C. Measurand:
Methylenedioxymethamphetamine (MDMA), Morphine
D. Type of Test:
Qualitative lateral flow chromatographic Immunoassay
E. Applicant:
Guangzhou Wondfo Biotech Co., Ltd.
F. Proprietary and Established Names:
Wondfo Methylenedioxymethamphetamine Urine Test
Wondfo Morphine Urine Test
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LAF | Class II | 21 CFR §862.3610: Methamphetamine test system | Toxicology (91) |
| DJG | Class II | 21 CFR §862.3650: Opiate test system | Toxicology (91) |
H. Intended Use:
1. Intended use(s):
See Indications for use below.
{1}
2. Indication(s) for use:
**Wondfo Methylenedioxymethamphetamine Urine Test:**
Wondfo Methylenedioxymethamphetamine Urine Test is an immunochromatographic assay for the qualitative determination of MDMA in human urine at a cutoff concentration of 500 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
**Wondfo Morphine Urine Test:**
Wondfo Morphine Urine Test is an immunochromatographic assay for the qualitative determination of MDMA in human urine at a cutoff concentration of 300 ng/mL. The test is available in a dip card format and a cup format. It is intended for prescription use and over the counter use.
The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
3. Special conditions for use statement(s):
For prescription and over the counter use.
4. Special instrument requirements:
Not applicable, as the devices are visually-read single-use devices.
I. Device Description:
**Wondfo Methylenedioxymethamphetamine Urine Test:**
Two Different Formats:
1) Urine cups with built in test device and reaction activation key
2) Urine cups with dip card
{2}
3
# Wondfo Morphine Urine Test:
Two Different Formats:
1) Urine cups with built in test device and reaction activation key
2) Urine cups with dip card
Wondfo® Methylenedioxymethamphetamine and morphine urine DOA test kits contain 25 tests (urine cups with built in test device, reaction activation key/ urine cups with dip card) per kit. Each kit contains 25 security sealed labels and a leaflet with instructions for use. For over the counter test kits, the following are contained with each kit: 25 labeled vials for shipping a “preliminary” sample to be confirmed by a lab, 25 plastic transportation bags, 25 mailing boxes, and 25 personal identification numbers.
Material Required but not provided: Timer, external urine controls
## J. Substantial Equivalence Information:
1. Predicate Device Name:
Acon Laboratories, Inc. One Step Drug Screen Test
Acon Laboratories, Inc. MDMA One Step Ecstasy Test
2. Predicate 510(k) Number(s):
k020771, k022589
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| | Candidate Device: | Predicate Devices: |
| Intended Use/ Indications for Use | Qualitative detection of drugs-of-abuse in urine | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assay based on the principle of antigen antibody immunochemistry | Same |
| Type of Test | Immunoassay | Same |
{3}
| Device Design/ Performance | Positive Result | A rose-pink band in the control region | Same |
| --- | --- | --- | --- |
| | | | |
| | | | |
| | | | |
| | Negative result | | |
| | | Ovalbumin Conjugate | Same |
| | | | |
| | | | |
| | Detection reagent | | |
| Accuracy Assessment | Anti-mouse IgG Polyclonal antibody (control line reagent) | | |
| Results | | Qualitative | Same |
| Cut-off | | Morphine 300 ng/mL Methylenedioxymethamphetamine 500 ng/mL | Same |
| Configurations | | Cup, dip card | Card, dip card with an integrated cup for Morphine Strip, device for MDMA |
# K. Standard/Guidance Document Referenced (if applicable):
None were referenced.
# L. Test Principle:
Wondfo Methylenedioxymethamphetamine and Morphine Urine Tests are a one-step lateral flow immunoassay containing a conjugate pad with colloidal gold with antidrug antibodies, a nitrocellulose membrane, with a test line (T) and a control line (C). The T line is coated with drug antigen bound to duck egg protein and the C line is coated with goat anti-mouse IgG polyclonal antibodies. The test is a competitive binding immunoassay in which drugs and drug metabolites in a urine sample compete with immobilized drug conjugate for limited labeled antibody binding sites. When a sufficient amount of sample is applied, the sample migrates through the test device by capillary action. If the concentration of drug is below the cutoff level, the anti-drug antibodies in the colloidal gold particles will bind to the drug antigens coated in the
{4}
test zone producing a band which indicates a negative result. If the drug concentration is at the cutoff level or higher no band will form in the test zone (test line T) indicating a preliminary positive. A band should form in the control region regardless of the presence of drug or metabolite in the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision study was conducted by three operators. Three different lots were tested for morphine and methamphetamine using 12 devices/ lot at each control level of negative, -75%, -50%, -25%, cut-off, +25%, +50%, and +75% of the cut-off. Samples were prepared, concentrations of each sample confirmed by GC/MS, and then each sample was divided into 300 aliquots. The 300 sample aliquots were divided into 12 sets of 25 for each test format (cup/dipcard). The study was conducted over a 25 day period and ran 3 batches of each test format (cup/dipcard) in 2 runs/day. Each device was tested and interpreted by the same operator. Summaries are presented in the following tables:
Methylenedioxymethamphetamine (MDMA):
A. Cup Format
| Result
MDMA | -100%
cut off | -75%
cut off | -50%
cut off | -25%
Cut off | Cut off | +25%
cut off | +50%
cut off | +75%
cut off | +100%
cut off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| LOT W1260901CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 44+/6- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1260902CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 45+/5- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1260903CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 44+/6- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
B. Dip card Format
| Result
MDMA | -100%
cut off | -75%
Cut off | -50%
cut off | -25%
cut off | cut off | +25%
cut off | +50%
cut off | +75%
Cut off | +100%
cut off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| LOT W1260901P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 43+/7- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1260902P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 44+/6- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1260903P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 45+/5- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
{5}
6
# Morphine:
## A. Cup Format
| Result
MOP | -100% cut off | -75% cut off | -50% cut off | -25% Cut off | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| LOT W1460901CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 45+/5- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1460902CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 44+/6- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1460903CU | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 45+/5- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
## B. Dip card Format
| Result
MOP | -100% cut off | -75% cut off | -50% cut off | -25% cut off | Cut off | +25% Cut off | +50% cut off | +75% cut off | +100% cut off |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| LOT W1460901P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 43+/7- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1460902P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 45+/5- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| LOT W1460903P | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 44+/6- | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
## b. Linearity/assay reportable range:
Not applicable, the assay is intended for qualitative use.
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This device has internal process controls. A colored line appearing in the control region confirms sufficient sample volume and adequate membrane wicking. Users are informed that the test is invalid if a line fails to appear in the control region.
Control standards are not supplied with this device; however it is good laboratory practice to confirm the test procedure and to verify proper test performance. Users should follow all applicable guidelines for testing QC materials.
## Stability:
Accelerated stability and real time stability tests were performed on three batches of cups and dip card strips for the Wondfo Methylenedioxymethamphetamine (MDMA) and Morphine Urine tests. Accelerated stability was performed at $50^{\circ}\mathrm{C}$ and the data submitted supports an 18 month shelf life. Real time studies were performed by storing three lots
{6}
of cups and dip card strips at 4°C and 30°C. Performance tests were completed at defined intervals and the results support a 24 month shelf life at 4°C for both MDMA and morphine urine tests and a shelf life of 23 months when stored at 30°C.
d. Detection limit:
Analytical performance of the device around the cutoff is described in Section f. (Assay cut-off) below.
e. Analytical specificity:
To evaluate specificity of the urine Methylenedioxymethamphetamine (MDMA) and morphine urine devices, three batches of urine samples were tested by taking 30 negative urine samples and spiking these samples with three different analyte concentration levels. Two different groups of operators were assigned to test blinded urine samples (three operators tested the cup format and three operators tested the dip stick format). Percent cross reactivity of a compound was calculated by dividing the cutoff concentration by the minimum concentration required to obtain a positive result and multiplying by 100%. Summary of the results are as follows:
MDMA
| Methylenedioxymethamphetamine (MDMA)
(3,4- Methylenedioxymethamphetamine HCl, Cutoff =500 ng/mL) | Result
Positive at 500 ng/mL | % Cross-Reactivity
100% |
| --- | --- | --- |
| 3,4- Methylenedioxymethamphetamine HCl (MDA) | Positive at 3,000 ng/mL | 16.7% |
| 3,4- Methylenedioxymethamphetamine | Positive at 300 ng/mL | 166.7% |
MOP:
| MOP(Morphine)
(Morphine, Cutoff=300 ng/mL) | Result
Positive at 300 ng/mL | % Cross-Reactivity
100% |
| --- | --- | --- |
| Codeine | Positive at 300 ng/mL | 100% |
| Ethyl Morphine | Positive at 300 ng/mL | 100% |
| Hydrocodone | Positive at 5,000 ng/mL | 6% |
| Hydromorphone | Positive at 5,000 ng/mL | 6% |
| Morphine – 3 – β – d - glucuronide | Positive at 1,000 ng/mL | 30% |
| Thebaine | Positive at 30,000 ng/mL | 1% |
Interference Studies:
Interference studies were performed using the cup and dipcard tests using urine controls at +/-25% of each analyte cut-off concentration. Potential interferences to the Wondfo Methylenedioxymethamphetamine (MDMA) and morphine urine cup and dipcard tests were evaluated by adding various drugs, drug metabolites, and other compounds (structurally unrelated and
{7}
endogenous compounds) that are commonly found in the urine which may interfere with testing results. The following compounds were found not to cross react when tested at concentrations of $100~\mu \mathrm{g / mL}$ .
MDMA:
| 4-Acetamidophenol | (L) - Epinephrine | Perphenazine |
| --- | --- | --- |
| Acetophenetidin | Erythromycin | Phencyclidine |
| N-Acetylprocainamide | β-Estradiol | Phenelzine |
| Acetylsalicylic acid | Estrone-3-sulfate | Phenobarbital |
| Aminopyrine | Ethyl-p-aminobenzoate | Phentermine |
| Amitryptyline | Fenoprofen | Trans-2-phenylcyclopropylamine hydrochloride |
| Amobarbital | Furosemide | L-Phenylephrine |
| Amoxicillin | Gentisic acid | β-Phenylethylamine |
| Ampicillin | Hemoglobin | Phenylpropanolamine |
| L-Ascorbic acid | Hydralazine | Prednisolone |
| Apomorphine | Hydrochlorothiazide | Prednisone |
| Aspartame | Hydrocodone | Procaine |
| Atropine | (±) - Isoproterenol | Promazine |
| Benzilic acid | Hydrocortisone | Promethazine |
| Benzoic acid | O-Hydroxyhippuric acid | DL-Propranolol |
| Benzoylecgonine | 3-Hydroxytyramine | D-Propoxyphene |
| Bilirubin | Ibuprofen | D-Pseudoephedrine |
| (±) - Brompheniramine | Imipramine | Quinacrine |
| Buspiron | Iproniazid | Quinidine |
| Caffeine | Isoxsuprine | Quinine |
| Cannabidiol | Ketamine | Ranitidine |
| Cannabinol | Ketoprofen | Salicylic acid |
| Chloralhydrate | Labetalol | Secobarbital |
| Chloramphenicol | Levorphanol | Serotonin (5-Hydroxytyramine) |
| Chlordiazepoxide | Loperamide | Sulfamethazine |
| Chlorothiazide | Maprotiline | Sulindac |
| (±) - Chlorpheniramine | Meperidine | Sustiva |
| Chlorpromazine | Meprobamate | Temazepam |
| Chlorquine Methylphenidate | Methadone | Tetracycline |
| Cholesterol | Morphine-3-β-Dglucuronide | Tetrahydrocortisone, 3-acetate |
| Clomipramine | Morphine sulfate | Tetrahydrocortisone 3- |
{8}
9
| | | (β-Dglucuronide) |
| --- | --- | --- |
| Clonidine | Nalidixic acid | Tetrahydrozoline |
| Cocaethylene | Naloxone | Thebaine |
| Cocaine hydrochloride | Naltrexone | Theophynine |
| Codeine | Naproxen | Thiamine |
| Cortisone | Niacinamide | Thioridazine |
| (-) Cotinine | Nifedipine | Tolbutamide |
| Creatinine | Nimesulidate | Trans-2-phenylcyclopropylamine |
| Deoxycorticosterone | Norcodein | Trazodone |
| Dextromethorphan | Norethindrone | Triamterene |
| Diclofenac | D-Norpropoxyphene | DL-Tyrosine |
| Diazepam | Noscapine | Trifluoperazine |
| Diflunisal | D,L-Octopamine | Trimethoprim |
| Digoxin | Oxalic acid | Trimipramine |
| Dicylomine | Oxazepam | Tryptamine |
| Diphenhydramine | Oxolinic acid | D L-Tryptophan |
| 5,5 – Diphenylhydantoin | Oxycodone | Tyramine |
| Doxylamine | Oxymetazoline | Uric acid |
| Ecgonine hydrochloride | Papaverine | Verapamil |
| Ecgonine methylester | Penicillin-G | Zomepirac |
| (-) – Ψ-Ephedrine | Pentazocinehydrochloride | |
| 1R,2S Ephedrine | Pentobarbital | |
Morphine
| 4-Acetamidophenol | Ecaonine hydrochloride | Perphenazine |
| --- | --- | --- |
| Acetophenetidin | Ecgonine methylester | Phencyclidine |
| N-Acetvprocainamide | (-)-Ψ-Ephedrine | Phenelzine |
| Acetvsalicvlic acid | Fenoprofen | Phenobarbital |
| Aminopvrine | Furosemide | Phentermine |
| Amityptvline | Gentisic acid | L-Phenylephrine |
| Amorbarbital | Hemoglobin | β-Phenylethylamine |
| Amoxicillin | Hydrocortisone | Phenylpropanotamine |
| Ampicillin | O-Hydroxyhippuric acid | Prednisone |
| l-Ascorbic Acid | p-Hydroxy-methamphetamine | D.L-Propanolol |
{9}
| D.L-Amphetamine | 3-Hydroxytyramine | D-Propoxyphene |
| --- | --- | --- |
| Apormorphine | Ibuprofen | D-Pseudoephedrine |
| Aspartame | Imipramine | Quinine |
| Atropine | Iproniazid | Ranitidine |
| Benzillic acid | (±)Isoproterenol | Salicylic acid |
| Benzoic acid | Isoxsuprine | Secobarbital |
| Benzoylecaonine | Ketamine | Serotonin (5-Hydroxytyramine) |
| Benzphetamine | Ketoprofen | Sulfamethazine |
| Bilirubin | Labetalol | Sulindac |
| Caffeine | Loperamide | Temazepam |
| Cannabidiol | Maprotiline | Tetrahydrocortisone,3 Acetate |
| Chloralhvrate | Meperidine | Tetrahydrocortisone,(β-D glucuronide) |
| Chloramphenicol | Meprobamate | Tetrahydrozoline |
| Chlordiazepoxide | Methadone | Thiamine |
| Chlorothiazide | Methoxyphenamine | Thioridazine |
| (±)Chlorpheniramine | () 3,4-Methylenedioxy-amphetamine | D.L-Tyrosine |
| Chlorpromazine | ()3,4-Methylenedioxy-methamphetamine | Tolbutamide |
| Chlorquine | Nalidixic acid | Triamterene |
| Cholesterol | Nalorphine | Trifluoperazine |
| Clomipramine | Naloxone | Trimethoprim |
| Clonidine | Naltrexone | Triyptamine |
| Cocaine hydrochloride | Naproxen | D.L-Tryptophan |
| Cortisone | Niacinamide | Yyramine |
| (-)cotinine | Nifedipine | Uric acid |
| Creatinine | Norethindrone | Verapamil |
| Dextromethlorphan | D-Norpropoxyphene | Zomepirac |
| Diazepam | Noscapine | |
| Diclolrfenac | D.L-Octopamine | |
| Diflunisal | Oxalic acid | |
| Diaoxin | Oxazepam | |
| Diphenhydramine | Oxolinic acid | |
| Doxylamine | Pentobarbital | |
pH:
The pH of an aliquoted negative urine pool was adjusted to pH 4, pH 5, pH 6, pH 7, pH 8, or pH 9 was spiked with +/- 25% of the cut-off concentration of MDMA and Morphine. Each was individually tested using three batches of
{10}
strips and three batches of the cup format. The spiked, pH-adjusted urine was tested in triplicate. Varying the pH did not affect the accuracy of each test format (dipcard and cup).
Specific Gravity:
Twelve drug-free urine samples with specific gravities (1.000-1.035) and spiked with $+/- 25\%$ of the cut-off concentration of MDMA and Morphine using three batches of strips and three batches of the cup format. The spiked urine samples were tested in triplicate. The results show that varying the specific gravity does not affect the accuracy of each test format (dipcard and cup).
# f. Assay cut-off:
The assay cut off established by collecting 25 clinical urine samples containing $+/- 50\%$ of the morphine cutoff of $300~\mathrm{ng / ml}$ and 25 clinical urine samples containing $+/- 50\%$ of the Methylenedioxymethamphetamine (MDMA) cutoff of $500~\mathrm{ng / mL}$ . Additionally, 125 drug-free clinical urine samples spiked with cocaine and 125 drug-free clinical urine samples spiked with methamphetamine were both diluted from the International Drug Standard (Sigma) to concentrations: $-50\%$ , $-25\%$ , cutoff, $+25\%$ , and $+50\%$ of the morphine cutoff $300~\mathrm{ng / mL}$ and Methylenedioxymethamphetamine (MDMA) cutoff $500~\mathrm{ng / mL}$ . The clinical urine samples were collected from the Shenzhen Drug Addiction Recovery Center and drug concentrations were confirmed by GC/MS. Results were read by three laboratory assistants with relevant experience. There cutoff studies were performed by two separate groups of operators (one for the dipcard format and one for the cup format). Three operators in each group performed the readings and they were blinded to the samples. Each result was confirmed by two other assistants. The results of the study are as follows:
a. Cup Format: MDMA
| Concentration (ng/mL) | Cut-off range | n | Batch1 | | Batch2 | | Batch3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | - | + | - | + | - | + | - | + |
| 250 | -50% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 375 | -25% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 500 | Cutoff | 30 | 5 | 25 | 5 | 25 | 2 | 28 | 12 | 78 |
| 625 | +25% cutoff | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
| 750 | +50% | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
{11}
b. DipCard
MDMA
| Concentration (ng/mL) | Cut-off range | n | Batch1 | | Batch2 | | Batch3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | - | + | - | + | - | + | - | + |
| 250 | -50% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 375 | -25% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 500 | Cutoff | 30 | 4 | 26 | 4 | 26 | 3 | 27 | 11 | 79 |
| 625 | +25% cutoff | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
| 750 | +50% | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
a. Cup Format
Morphine
| Concentration (ng/mL) | Cut-off range | n | Batch1 | | Batch2 | | Batch3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | - | + | - | + | - | + | - | + |
| 150 | -50% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 225 | -25% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 300 | Cutoff | 30 | 4 | 26 | 4 | 26 | 3 | 27 | 11 | 79 |
| 375 | +25% cutoff | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
| 450 | +50% | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
b. DipCard
Morphine
| Concentration (ng/mL) | Cut-off range | n | Batch1 | | Batch2 | | Batch3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | - | + | - | + | - | + | - | + |
| 150 | -50% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 225 | -25% cutoff | 30 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 |
| 300 | Cutoff | 30 | 2 | 28 | 4 | 26 | 5 | 25 | 11 | 79 |
| 375 | +25% cutoff | 30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 |
{12}
The cut-off values were determined to be:
| Test | Calibrator | Cut-off (ng/mL) |
| --- | --- | --- |
| Methylenedioxymethamphetamine (MDMA) | 3,4 - Methylenedioxymethamphetamine HCl | 500 |
| Morphine (MOP) | Morphine | 300 |
# 2. Comparison studies:
# a. Method comparison with predicate device:
Performance of the Wondfo Methylenedioxymethamphetamine (MDMA) and Morphine Urine Tests (dipcard and cup formats) were established by comparing 80 samples (40 positive and 40 negative) of each analyte against GC/MS. Ten of the 80 samples came from drug-free urine samples. Each result was read by three laboratory assistants and a lay person. All urine samples were collected at the Shenzhen Drug Addiction Recovery Center. A summary of results comparing the results of the lay person to the experienced person are as follows:
# MDMA:
Cup Format:
| Wondfo Device Result | | Drug-Free urine | <-50% of the cut-off | Near cutoff (Between -50% and cutoff) | Cut-off to +50% of the cut-off | >+ 50% of the cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 10 | 18 | 0 | 0 |
| Viewer B | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 10 | 18 | 0 | 0 |
| Viewer C | + | 0 | 0 | 1 | 20 | 20 |
| | - | 10 | 10 | 19 | 0 | 0 |
| Lay Person | + | 0 | 0 | 3 | 20 | 20 |
| | - | 10 | 10 | 17 | 0 | 0 |
Dipcard:
| Wondfo Device Result | Drug-Free urine | <-50% of the cut-off | Near cutoff (Between | Cut-off to +50% of the cut-off | >+ 50% of the cut-off |
| --- | --- | --- | --- | --- | --- |
{13}
14
| | | | | -50% and cutoff) | | |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 10 | 18 | 0 | 0 |
| Viewer B | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 10 | 18 | 0 | 0 |
| Viewer C | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 10 | 18 | 0 | 0 |
| Lay Person | + | 0 | 0 | 3 | 20 | 20 |
| | - | 10 | 10 | 17 | 0 | 0 |
## Morphine:
Cup Format:
| Wondfo Device Result | | Drug-Free urine | <-50% of the cut-off | Near cutoff (Between -50% and cutoff) | Cut-off to +50% of the cut-off | >+ 50% of the cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 1 | 20 | 20 |
| | - | 10 | 19 | 10 | 0 | 0 |
| Viewer B | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 19 | 9 | 0 | 0 |
| Viewer C | + | 0 | 0 | 1 | 20 | 20 |
| | - | 10 | 19 | 10 | 0 | 0 |
| Lay Person | + | 0 | 0 | 3 | 20 | 20 |
| | - | 10 | 19 | 8 | 0 | 0 |
Dipcard:
| Wondfo Device Result | | Drug-Free urine | <-50% of the cut-off | Near cutoff (Between -50% and cutoff) | Cut-off to +50% of the cut-off | >+ 50% of the cut-off |
| --- | --- | --- | --- | --- | --- | --- |
| Viewer A | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 19 | 9 | 0 | 0 |
| Viewer B | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 19 | 9 | 0 | 0 |
| Viewer C | + | 0 | 0 | 1 | 20 | 20 |
| | - | 10 | 19 | 10 | 0 | 0 |
| Lay Person | + | 0 | 0 | 2 | 20 | 20 |
| | - | 10 | 19 | 9 | 0 | 0 |
{14}
A summary of discordant results are as follows:
# MDMA:
Cup Format:
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MDMAC33 | 405 | + |
| Viewer A | MDMAC63 | 497 | + |
| Viewer B | MDMAC63 | 497 | + |
| Viewer B | MDMA 218 | 402 | + |
| Viewer C | MDMAC33 | 405 | + |
| Lay person | MDMAC33 | 405 | + |
| Lay person | MDMAC63 | 497 | + |
| Lay person | MDMA 218 | 402 | + |
Dipcard:
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MDMAC33 | 405 | + |
| Viewer A | MDMA 218 | 402 | + |
| Viewer B | MDMAC33 | 405 | + |
| Viewer B | MDMA 218 | 402 | + |
| Viewer C | MDMAC33 | 405 | + |
| Viewer C | MDMAC63 | 497 | + |
| Lay person | MDMAC33 | 405 | + |
| Lay person | MDMAC61 | 490 | + |
| Lay person | MDMA 218 | 402 | + |
# Morphine:
Cup Format:
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MOPC32 | 227 | + |
| Viewer B | MOPC64 | 292 | + |
| Viewer B | MOPC65 | 285 | + |
| Viewer C | MOPC65 | 285 | + |
| Lay person | MOPC32 | 227 | + |
| Lay person | MOPC64 | 292 | + |
| Lay person | MOPC65 | 285 | + |
Dipcard:
| Viewer | Sample Number | GC/MS result | Viewer Result |
| --- | --- | --- | --- |
| Viewer A | MOPC64 | 292 | + |
| Viewer A | MOPC65 | 285 | + |
{15}
The results indicate a similar positive, negative, and overall agreement rates for both morphine and methamphetamine using the cup and dipcard formats.
The overall agreement between the Wondfo devices and GC/MS is represented in the following table:
The agreement between Wondfo devices and GC/MS method
| % Agreement | MDMA(cup) | MDMA (Dipcard) | MOP (Cup) | MOP (Dipcard) |
| --- | --- | --- | --- | --- |
| Positive | 100% | 100% | 100% | 100% |
| Negative | 95.8% | 95.0% | 96.7% | 95.8% |
| Total | 97.9% | 97.5% | 98.4% | 97.9% |
The agreement between lay person and experience viewer:
| % Agreement | MDMA (cup) | MDMA (Dipcard) | MOP (Cup) | MOP (Dipcard) |
| --- | --- | --- | --- | --- |
| Positive | 100% | 100% | 100% | 100% |
| Negative | 96.6% | 97.4% | 95.6% | 99.2% |
| Total | 98.3% | 98.7% | 97.8% | 99.6% |
# b. Lay-User Study:
A lay-user study was performed to assess the suitability of the device for home use. Six drug-free urine sample pools were spiked with $+75\%$ , $+50\%$ , $+25\%$ , $-25\%$ , $-50\%$ , and $-75\%$ of the cutoff for MDMA and morphine. Additionally, a negative urine pool with no drug was tested. The six spiked urine samples and the negative urine pool concentrations were confirmed by GC/MS and aliquoted into 40 individual containers per concentration ( $n = 280$ aliquots, 20 aliquots per concentration per test format). Testing was performed at three sites by 140 blinded consumers divided between three sites (140 lay-users for the cup format and 140 lay-users for the dipcard format). The lay users were chosen from likely intended users at the Shenzhen Drug Addiction Recovery Center, Baiyun Drug Addiction Recovery Voluntary Center, and Guangdong Provincial No. 2 People's Hospital. Each participant received the package insert, 1 blinded sample, and either a cup or dipcard test format. The lay persons test result was compared to the GC/MS result to demonstrate accuracy by lay-users. The following are the results of the lay-user study:
{16}
MDMA:
Comparison between GC/MS and lay person results for Cup format
| % of Cutoff | Number of samples | MDMA Concentration by GC/MS (ng/mL) | Lay person results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- | --- |
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 149.5 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 237.9 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 408.6 | 3 | 17 | 85.0 |
| +25% Cutoff | 20 | 608.8 | 19 | 1 | 95.0 |
| +50% Cutoff | 20 | 786.7 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 911.8 | 20 | 0 | 100 |
Comparison between GC/MS and lay person results for Dip card format
| % of Cutoff | Number of samples | MDMA Concentration by GC/MS (ng/mL) | Lay person results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- | --- |
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 149.5 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 237.9 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 408.6 | 2 | 18 | 90.0 |
| +25% Cutoff | 20 | 608.8 | 19 | 1 | 95.0 |
| +50% Cutoff | 20 | 786.7 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 911.8 | 20 | 0 | 100 |
Morphine:
Comparison between GC/MS and lay person results for Cup format
| % of Cutoff | Number of samples | Morphine Concentration by GC/MS (ng/mL) | Lay person results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- | --- |
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 80.7 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 169.5 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 236.4 | 3 | 17 | 85.0 |
| +25% Cutoff | 20 | 390.7 | 18 | 2 | 90.0 |
| +50% Cutoff | 20 | 439.8 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 548.6 | 20 | 0 | 100 |
{17}
Comparison between GC/MS and lay person results for Dip card format
| % of Cutoff | Number of samples | Morphine Concentration by GC/MS (ng/mL) | Lay person results | | The percentage of correct results (%) |
| --- | --- | --- | --- | --- | --- |
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 80.7 | 0 | 20 | 100 |
| -50% Cutoff | 20 | 169.5 | 0 | 20 | 100 |
| -25% Cutoff | 20 | 236.4 | 2 | 18 | 90.0 |
| +25% Cutoff | 20 | 390.7 | 19 | 1 | 95.0 |
| +50% Cutoff | 20 | 439.8 | 20 | 0 | 100 |
| +75% Cutoff | 20 | 548.6 | 20 | 0 | 100 |
The overall percent agreement between the Lay person and the GC/MS method:
| % Agreement | MDMA (cup) | MDMA(dipcard) | MOP (cup) | MOP (dipcard) |
| --- | --- | --- | --- | --- |
| Total | 97.1% | 97.8% | 96.4% | 97.8% |
c. Lay-User Questionnaire:
Cup Format:
Morphine:
The participant’s ages ranged from 21 to 63, and there were slightly more females than males. There was a variety of occupational and educational backgrounds and three of the lay-users had used a drug kit before. All participants understood the storage and expiration of the device and that the test could not be reused. They all understood that they could not insert and rotate the key until they were ready to test. They all read the test results at 5 minutes (none after 5 minutes) and all understood how to interpret the results.
Clarity of the package insert
| Remarks | Very clear | Clear | Ambiguous | Very ambiguous |
| --- | --- | --- | --- | --- |
| Instruction for use | 60 | 79 | 1 | 0 |
| Interpretation of results | 71 | 69 | 0 | 0 |
Clarity of test simplicity
| Remarks | Very easy | Easy | Difficult | Very difficult |
| --- | --- | --- | --- | --- |
| Number | 51 | 89 | 0 | 0 |
None of the participants responded that the test was not easy to do.
{18}
19
# MDMA:
The participant’s ages ranged from 22 to 63, and there were more males than females. There was a variety of occupational and educational backgrounds and one of the lay-users had used a drug kit before. All participants understood the storage and expiration of the device and that the test could not be reused. They all understood that they could not insert and rotate the key until they were ready to test. They all read the test results at 5 minutes (none after 5 minutes) and all understood how to interpret the results.
## Clarity of the package insert
| Remarks | Very clear | Clear | Ambiguous | Very ambiguous |
| --- | --- | --- | --- | --- |
| Instruction for use | 77 | 63 | 0 | 0 |
| Interpretation of results | 81 | 59 | 0 | 0 |
## Clarity of test simplicity
| Remarks | Very easy | Easy | Difficult | Very difficult |
| --- | --- | --- | --- | --- |
| Number | 80 | 58 | 2 | 0 |
Two of the participants responded that the test was not easy to do.
ii. Dipcard Format:
## Morphine:
The participant’s ages ranged from 24 to 60, and there were slightly more males than females. There was a variety of occupational and educational backgrounds and one of the lay-users had used a drug kit before. All participants understood the storage and expiration of the device and that the test could not be reused. They all understood that the dipcard couldn’t be immersed in urine above the marker line and understood the storage and expiration of the device and that the test could not be reused. They all read the test results at 5 minutes (none after 5 minutes) and all understood how to interpret the results.
### Clarity of the package insert
| Remarks | Very clear | Clear | Ambiguous | Very ambiguous |
| --- | --- | --- | --- | --- |
| Instruction for use | 60 | 80 | 0 | 0 |
| Interpretation of results | 79 | 61 | 0 | 0 |
### Clarity of test simplicity
| Remarks | Very easy | Easy | Difficult | Very difficult |
| --- | --- | --- | --- | --- |
| Number | 64 | 75 | 1 | 0 |
Only one of them responded that the test was not easy to do.
{19}
20
# MDMA:
The participant’s ages ranged from 23 to 61, and there were slightly more males than females. There was a variety of occupational and educational backgrounds and none of the lay-users had used a drug kit before. All participants understood the storage and expiration of the device and that the test could not be reused. They all understood that the dipcard couldn’t be immersed in urine above the marker line and understood the storage and expiration of the device and that the test could not be reused. They all read the test results at 5 minutes (none after 5 minutes) and all understood how to interpret the results.
Clarity of the package insert
| Remarks | Very clear | Clear | Ambiguous | Very ambiguous |
| --- | --- | --- | --- | --- |
| Instruction for use | 77 | 63 | 0 | 0 |
| Interpretation of results | 82 | 58 | 0 | 0 |
Clarity of test simplicity
| Remarks | Very easy | Easy | Difficult | Very difficult |
| --- | --- | --- | --- | --- |
| Number | 64 | 75 | 1 | 0 |
Only one of them responded that the test was not easy to do.
Additionally, a Flesch-Kincaid reading analysis revealed that both package inserts had a reading grade level of 7.
c. Matrix comparison:
Not applicable; these devices are for use with urine only.
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable.
b. Clinical specificity:
Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable.
4. Clinical cut-off:
Not Applicable.
{20}
5. Expected values/Reference range:
Not Applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
21
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.