K Number
K140089

Validate with FDA (Live)

Date Cleared
2014-04-08

(84 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CR3 Keyless Split Sample Cup Morphine-Oxazepam is a rapid test for the qualitative detection of Morphine (a drug in the opiate class) and Oxazepam(a drug in the benzodiazepine class) in human urine at a cutoff concentration of 2000ng/mL and 300ng/mL, respectively.

The tests may yield preliminary positive results even when prescription drugs including Morphine and Oxazepam are ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of these drugs. There are no uniformly recognized cutoff concentration levels for morphine and oxazepam in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Device Description

Immunochromatograph assays for Opiate and Benzodiazepines Urine Tests use a lateral flow, one step system for the qualitative detection of Morphine and Oxazepam in human urine. Each assay uses a monoclonal antibody-dye conjugate against drugs with gold chloride and fixed drug-protein conjugates and anti-mouse IgG polyclonal antibody in membranes.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

CriteriaAcceptance Range (Implicit)Reported Device Performance (Morphine)Reported Device Performance (Oxazepam)
PrecisionShould consistently detect positive samples above cutoff and negative samples below cutoff. A small number of discordant results near the cutoff may be acceptable within a certain range.For samples at -25% cut-off, all 50 tests were negative (0+). At +25% cut-off, all 50 tests were positive (50+/0-). For samples at the exact cutoff, Lot W11510201CU5 showed 43 positive and 7 negative results (86% positive). Lot W11510202CU5 showed 44 positive and 6 negative results (88% positive). Lot W11510203CU5 showed 43 positive and 7 negative results (86% positive).For samples at -25% cut-off, all 50 tests were negative (0+). At +25% cut-off, all 50 tests were positive (50+/0-). For samples at the exact cutoff, Lot W11510201CU5 showed 43 positive and 7 negative results (86% positive). Lot W11510202CU5 showed 43 positive and 7 negative results (86% positive). Lot W11510203CU5 showed 42 positive and 8 negative results (84% positive).
StabilityStable for a specified duration at certain temperatures.Stable at 4-30°C for 18 months.Stable at 4-30°C for 18 months.
Cut-off ValuesDevice must employ specific cutoff concentrations for detection.Morphine: 2000 ng/mLOxazepam: 300 ng/mL
InterferenceShould show no interference from common compounds at specified concentrations.Numerous compounds listed show no interference at 100 µg/mL.Numerous compounds listed show no interference at 100 µg/mL.
Specificity/Cross-ReactivityShould accurately detect the target drug and demonstrate anticipated cross-reactivity with related compounds.- Morphine: 100% at 2000 ng/mL - Codeine: 100% at 2000 ng/mL - Ethylmorphine: 40% at 5000 ng/mL (and other cross-reactive compounds listed)- Oxazepam: 100% at 300 ng/mL - Alprazolam: 150% at 200 ng/mL - Clobazam: 300% at 100 ng/mL (and other cross-reactive compounds listed)
Effect of Urine DensityTest results should not be affected by varying urine densities.No effect on test results for samples with density ranges (1.000-1.035).No effect on test results for samples with density ranges (1.000-1.035).
Effect of Urine pHTest results should not be affected by varying urine pH levels.No interference for samples with pH range of 4 to 9.No interference for samples with pH range of 4 to 9.
Comparison Study (Professional Users)Device results should show good agreement with GC/MS results, especially for samples significantly above or below the cutoff. Minor discrepancies expected near cutoff.For Morphine: - Negative by GC/MS (all categories): 100% negative results by all professional users. - High Positive by GC/MS: 100% positive results by all professional users. - Near Cutoff Negative: 2-3 positive results observed among 10-test sets for each viewer. - Near Cutoff Positive: 1-2 negative results observed among 20-test sets for each viewer.For Oxazepam: - Negative by GC/MS (all categories): 100% negative results by all professional users. - High Positive by GC/MS: 100% positive results by all professional users. - Near Cutoff Negative: 2-3 positive results observed among 30-test sets for each viewer. - Near Cutoff Positive: 1-4 negative results observed among 20-test sets for each viewer.
Lay-User StudyHigh percentage agreement with GC/MS results, especially for samples significantly above or below the cutoff. Some discrepancy near cutoff is expected.For Morphine: - -100%, -75%, -50% cut-off: 100% agreement. - -25% cut-off: 90% agreement. - +25% cut-off: 85% agreement. - +50%, +75% cut-off: 100% agreement.For Oxazepam: - -75%, -50% cut-off: 100% agreement. - -25% cut-off: 85% agreement. - +25% cut-off: 85% agreement. - +50%, +75% cut-off: 100% agreement.

Study Details:

  1. Test Set Sample Size and Data Provenance:

    • Precision Study: For each concentration level (-100%cut off, -75%cut off, -50%cut off, -25%cut off, +25%cut off, +50%cut off , +75%cut off and +100%cut off), tests were performed two runs per day for 25 days with three different lots of devices. This means 50 tests per concentration per lot for a total of 450 tests per lot for 9 concentrations. With 3 lots, this amounts to 1350 tests for Morphine and 1350 tests for Oxazepam in the precision study.
    • Comparison Studies (Professional Users): 80 unaltered clinical samples in total (40 negative and 40 positive) were used for each drug (Morphine and Oxazepam). The provenance is "in-house," implying these were likely collected for the purpose of the study. The text does not specify the country of origin, but the company is based in Guangzhou, P.R. China. The samples were "clinical samples," and the study compared device results to GC/MS, making it a prospective assessment within the study's scope, though the collection method of the initial clinical samples isn't fully detailed.
    • Lay-User Study: 260 lay persons participated. For each drug, urine samples were prepared at 7 different concentrations, with 20 samples per concentration. This means 140 samples were tested for Morphine and 140 samples were tested for Oxazepam within this study using lay users. The samples were "prepared" by spiking drug-free pooled urine, indicating controlled, laboratory-prepared samples. Data provenance is implied as "in-house."
  2. Number of Experts and Qualifications:

    • Comparison Studies (Professional Users): "three laboratory assistants with relevant experience" were used. Specific qualifications (e.g., number of years of experience, specific certifications) are not provided beyond "relevant experience."
    • Lay-User Study: One "lay person" was also included in the comparison study for context, but not to establish ground truth.
  3. Adjudication Method:

    • For the precision study and lay-user study, the comparison for correctness was directly against the known prepared concentrations.
    • For the comparison studies with professional users, the device results were compared directly against independently confirmed GC/MS results, meaning GC/MS served as the definitive ground truth, and no explicit expert adjudication method (like 2+1 or 3+1) was used between the viewers. The viewers independently interpreted the device results, and these were then compared to the GC/MS.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No MRMC comparative effectiveness study was explicitly performed to measure the effect size of how much human readers improve with AI vs. without AI assistance. This device is a rapid test kit, not an AI-assisted diagnostic tool.
  5. Standalone (Algorithm Only) Performance:

    • This is not applicable as the device is a lateral flow immunoassay, a chemical test, and does not involve an algorithm or AI. The results are visually interpreted.
  6. Type of Ground Truth Used:

    • Precision and Lay-User Studies: Ground truth was established by known concentrations of the drugs (spiked samples), confirmed by GC/MS.
    • Comparison Studies (Professional Users): Ground truth was established by GC/MS (Gas Chromatography/Mass Spectrometry) results for unaltered clinical samples. GC/MS is considered the "preferred confirmatory method" and is an objective chemical analysis.
  7. Training Set Sample Size:

    • The document does not explicitly mention a "training set" in the context of machine learning or algorithm development, as this is a chemical diagnostic device. The performance characteristics describe validation studies, not training data for an algorithm.
  8. How Ground Truth for Training Set was Established:

    • Not applicable, as there is no "training set" for an algorithm. The development of the immunochromatographic assay itself would involve laboratory testing and optimization, but this wouldn't be referred to as establishing ground truth for a training set in the AI sense.

{0}------------------------------------------------

APR 0 8 2014

510(k) SUMMARY

1.Date:February 28, 2014
2.Submitter:Guangzhou Wondfo Biotech Co., Ltd.South China University of TechnologyGuangzhou, P.R. China 510641
3.Contact person:Joe ShiaLSI International Inc.504 East Diamond Ave., Suite IGaithersburg, MD 20877Telephone: 240-505-7880Fax: 301-916-6213Email:shiajl@yahoo.com
    1. Device Name:
      CR3 Keyless Split Sample Cup Morphine-Oxazepam

Classification:

Product CodeCFR #Panel
DJG21 CFR, 862.3650 Opiate Test SystemToxicology
JXM21 CFR, 862.3170 Benzodiazepine Test SystemToxicology
    1. Predicate Devices:
      Guangzhou Wondfo Biotech Co., Ltd. Wondfo Multi-Drug Urine Test Cup (Panel) (K130665)
    1. Intended Use
      CR3 Keyless Split Sample Cup Morphine-Oxazepam is a rapid test for the qualitative detection of Morphine (a drug in the opiate class) and Oxazepam(a drug in the benzodiazepine class) in human urine at a cutoff concentration of 2000ng/mL and 300ng/mL, respectively.

The tests may yield preliminary positive results even when prescription drugs including Morphine and Oxazepam are ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of these drugs. There are no uniformly recognized cutoff concentration levels for morphine and oxazepam in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

    1. Device Description
      Immunochromatograph assays for Opiate and Benzodiazepines Urine Tests use a lateral flow, one step system for the qualitative detection of Morphine and Oxazepam in human urine. Each assay uses a monoclonal antibody-dye conjugate against drugs with gold chloride and fixed drug-protein conjugates and anti-mouse IgG polyclonal antibody in membranes.
  • Substantial Equivalence Information 8.

{1}------------------------------------------------

ItemDevicePredicate
Indication(s) for useFor the qualitative determination ofMorphine, Oxazepam individual inhuman urine.Same (but the number ofdrugs detected different)
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Type Of TestImmunoassay principles that rely onantigen-antibody interactions to indicatepositive or negative resultSame
ResultsQualitativeSame
Specimen TypeHuman urineSame
Cut Off ValuesMorphine: 2000ng/mlOxazepam: 300ng/mlSame (but the numberof drugs detecteddifferent)
ConfigurationsSplit Keyless CupCup, Dip Card
Intended UseOTC Use & Prescription UseSame

9. Test Principle

It is a rapid test for the qualitative detection of Morphine and Oxazepam in urine samples. It is a lateral flow chromatographic immunoassay. When the absorbent end is immersed into a urine sample, the urine is absorbed into the device by capillary action and mixes with the antibody-dye conjugate, flowing across the pre-coated membrane. At analyte concentration below the target cut off, antibody-dye conjugates bind to the drug-protein conjugate immobilized in the Test Region (T) of the device. This produces a colored test line that indicates a negative result. When analyte concentration is above the cutoff, analyte molecules bind to the antibody-dye conjugate, preventing the antibody-dye conjugate from binding to the drug-protein conjugate immobilized in the Test Region (T) of the device. No colored band shows in the test region, indicating a preliminary positive result.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100%cut off, -75%cut off, -50%cut off, -25%cut off, +25%cut off, +50%cut off , +75%cut off and +100%cut off. For each concentration, tests were performed two runs per day for 25 days with three different lots of devices. The results obtained are summarized in the following table.

A. For Morphine testing

{2}------------------------------------------------

ResultOPI-100%cut off-75%cut off-50%cut off-25%Cut offcut off+25%cut off+50%cut off+75%cut off+100%cut off
W11510201CU550-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-
W11510202CU550-/0+50-/0+50-/0+50-/0+44+/6-50+/0-50+/0-50+/0-50+/0-
W11510203CU550-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-

B. For Oxazepam testing

ResultBZO-100%cut off-75%cut off-50%cut off-25%Cut offcut off+25%cut off+50%cut off+75%cut off+100%cut off
W11510201CU550-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-
W11510202CU550-/0+50-/0+50-/0+50-/0+43+/7-50+/0-50+/0-50+/0-50+/0-
W11510203CU550-/0+50-/0+50-/0+50-/0+42+/8-50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable ·

c. Stability

It is stable at 4-30°C for 18 months.

d. Cut-off

TestCalibratorCut-off (ng/ml)
MorphineMorphine2000
OxazepamOxazepam300

e. Interference

Compounds that show no interference at a concentration of 100 µg/mL are summarized in the following tables.

Morphine

4-AcetamidophenolEcgonine methylesterOxolinic acid
Acetophenetidin(-) -Y -EphedrineOxymetazoline
N-AcetylprocainamideErythromycinPapaverine
Acetylsalicylic acidβ-EstradiolPenicillin-G
AminopyrineEstrone-3-sulfatePentazocine
AmitryptylineEthyl-p-aminobenzoatePentobarbital
AmobarbitalFenoprofenPerphenazine
AmoxicillinFurosemidePhencyclidine

{3}------------------------------------------------

Ampicillin Ascorbic acid D,L-Amphetamine Apomorphine Aspartame Atropine Benzilic acid Benzoic acid Benzovlecgonine Benzphetamine Bilirubin (±) Brompheniramine Caffeine Cannabidiol Chloralhydrate Chloramphenicol Chlordiazepoxide Chlorothiazide (±) Chlorpheniramine Chlorpromazine Chlorquine Cholesterol Clomipramine Clonidine Cocaine hydrochloride Cortisone (-) Cotinine Creatinine Deoxycorticosterone Dextromethorphan Diazepam Diclofenac Diflunisal Digoxin Diphenhydramine Doxylamine Ecgonine hydrochloride

Oxazepam

4-Acetamidophenol Acetophenetidin N-Acetvprocainamide Acetvsalicvlic acid Aminopvrine

Gentisic acid Hemoglobin Hydralazine Hydrochlorothiazide Hydrocortisone O-Hvdroxyhippuric acid p-Hydroxy methamphetamine 3-Hydroxytyramine Ibuprofen Imipramine Iproniazid Isoprotereno] Isoxsuprine Ketamine Ketoprofen Labetalol Loperamide Maprotiline Meperidine Meprobamate Methadone Methoxyphenamine (+) 3,4-Methylenedioxyamphetamine (+)3,4-Methylenedioxymethamphetamine Nalidixic acid Nalorphine Naloxone Naltrexone Naproxen Niacinamide Nifedipine Norethindrone D-Norpropoxyphene Noscapine D,L-Octopamine Oxalic acid Oxazepam

Doxylamine Ecaonine dydrochloride Ecqonine methylester (-)-Y-Ephedrine Fenoprofen

Phenelzine Phenobarbital Phentermine L-Phenylephrine ß-Phenylethylamine Phenylpropanolamine Prednisone D.L-Propanolol D-Propoxyphene D-Pseudoephedrine Quinidine Quinine Ranitidine Salicylic acid Secobarbital Serotonin (5-Hydroxytyramine) Sulfamethazine Sulindac Temazepam Tetracycline Tetrahydrocortisone, 3 Acetate Tetrahydrocortisone3 ( B -D glucuronide) Tetrahydrozoline Thiamine Thioridazine D. L-Tyrosine Tolbutamide Triamterene Trifluoperazine Trimethoprim Trimipramine Tryptamine

D. L-Tryptophan Tyramine Uric acid Verapamil Zomepirac

Oxolinic acid Pentobarbital Perphenazine Phencyclidine Phenelzine

{4}------------------------------------------------

Amityptvline Furosemide Amorbarbital Gentisic acid Amoxicillin Hemoglobin Ampicillin Hydrocortisone l-Ascorbic Acid O-Hydroxyhippuric acid D.L-Amphetamine p-Hydroxy- methamphetamine Apormorphine 3-Hydroxytyramine Aspartame Ibuprofen Atropine Imipramine Benzillic acid Iproniazid Benzoic acid (±)Isoproterenol Benzoylecaonine Isoxsuprine Benzphetamine Ketamine Bilirubin Ketoprofen (±) Chlorpheniramine Labetalol Caffeine Loperamide Cannabidiol Maprotiline Chloralhvdrate Meperidine Chloramphenicol Meprobamate Chlorothiazide Methadone (±)Chlorpheniramine Methoxyphenamine Chlorpromazine (+) 3,4-Methylenedioxyamphetamine (+)3,4-Methylenedioxy-Chlorquine methamphetamine Cholesterol Nalidixic acid Clomipramine Nalorphine Clonidine Naloxone Cocaine hydrochloride Naltrexone Cortisone Naproxen (-)cotinine Niacinamide Creatinine Nifedipine Dextromethlorphan Norethindrone DicloIrfenac D-Norpropoxyphene Diflunisal Noscapine Diaoxin D.L-Octopamine Diphenhydramine Oxalic acid

Phenobarbital Phentermine L-Phenylephrine ß-Phenylethylamine Phenylpropanotamine Prednisone D.L-Propanolol D-Propoxyphene D-Pseudoephedrine Quinine Ranitidine Salicylic acid Secobarbital Serotonin (5-Hydroxytyramine) Sertraline Sulfamethazine Sulindac Tetrahydrocortisone,3 Acetate Tetrahydrocortisone.(B-D glucuronide) Tetrahydrozoline Thiamine Thioridazine

D.L-Tyrosine

Tolbutamide Triamterene Trifluoperazine Trimethoprim Triyptamine D.L-Tryptophan Tyramine Uric acid Verapamil Zomepirac

f. Specificity/Cross-Reactivity

To test the specificity/cross-reactivity, drug metabolites and other components that are likely to be present in urine samples were tested. Compounds that produced positive results are listed below.

7

{5}------------------------------------------------

Morphine, Cutoff=2000 ng/mLResult%Cross-Reactivity
Positive at 2,000 ng/mL100%
CodeinePositive at 2,000 ng/mL100%
EthylmorphinePositive at 5,000 ng/mL40%
HydrocodonePositive at 12,500 ng/mL16%
HydromorphinePositive at 5,000 ng/mL40%
LevorphanolPositive at 75,000 ng/mL2.7%
σ-MonoacetylmorphinePositive at 5,000 ng/mL40%
Morphine 3-b-D-glucuronidePositive at 2,000 ng/mL100%
NorcodeinePositive at 12,500 ng/mL16%
NormorphonePositive at 50,000 ng/mL4%
OxycodonePositive at 25,000 ng/mL8%
OxymorphonePositive at 25,000 ng/mL8%
ProcainePositive at 150,000 ng/mL1.3%
ThebainePositive at 100,000 ng/mL2%
Oxazepam, Cutoff=300 ng/mLResult%Cross-Reactivity
Positive at 300 ng/mL100%
AlprazolamPositive at 200 ng/mL150%
a-HydroxyalprazolamPositive at 1,500 ng/mL20%
BromazepamPositive at 1,500 ng/mL20%
ChlordiazepoxidePositive at 1,500 ng/mL20%
Clonazepam HClPositive at 800 ng/mL37.5%
ClobazamPositive at 100 ng/mL300%
ClonazepamPositive at 800 ng/mL37.5%
Clorazepate dipotassiumPositive at 200 ng/mL150%
DelorazepamPositive at 1,500 ng/mL20%
DesalkylflurazepamPositive at 400 ng/mL75%
DiazepamPositive at 200 ng/mL150%
EstazolamPositive at 2,500 ng/mL12%
FlunitrazepamPositive at 400 ng/mL75%
D,L-LorazepamPositive at 1,500 ng/mL20%
MidazolamPositive at 12,500 ng/mL2.4%
NitrazepamPositive at 100 ng/mL300%
NorchlordiazepoxidePositive at 200 ng/mL150%
NordiazepamPositive at 400 ng/mL75%
TemazepamPositive at 100 ng/mL300%
TrazolamPositive at 2,500 ng/mL12%

g. Effects of Urine Density and pH

Density (Specific Gravity)

{6}------------------------------------------------

12 urine samples with density ranges (1.000-1.035) are collected and spiked with each drug at 25% below and 25% above cutoff levels. Each sample was tested by three batches of CR3 Keyless Split Sample Cup Morphine-Oxazepam. It shows that urine density does not affect test results.

Effect of Urine pH

The pH of an aliquot negative urine pool is adjusted to a pH range of 4 to 9 in 1 pH unit increments and spiked with each drug at 25% below and 25% above cutoff levels. Each sample was tested by three batches of CR3 Keyless Split Sample Cup Morphine-Oxazepam. It shows that urine pH does not interfere with the performance of the test.

2. Comparison Studies

The method comparison for the CR3 Keyless Split Sample Cup Morphine-Oxazepam was performed in-house with three laboratory assistants with relevant experience and a lay person with no experience other than reading the instructions for use. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples were blind labeled and compared to GC/MS results. The results are presented in the table below:

Morphine

GroupNegativeLow Negativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
OperatorsPositive0021920
Negative1020810
Viewer BPositive0031820
Negative1020720
Viewer CPositive0021820
Negative1020820
LayPersonPositive0031820
Negative1020720

Discordant table:

ViewerSample numberGC/MS resultViewer result
Viewer AOPIC10611997positive
Viewer AOPIC10641994positive
Viewer AOPIC10652025negative
Viewer BOPIC10611997positive
Viewer BOPIC10621943positive
Viewer BOPIC10632043negative
Viewer BOPIC10641994positive
Viewer BOPIC10652025negative
Viewer COPIC10611997positive

{7}------------------------------------------------

Viewer COPIC10632043negative
Viewer COPIC10641994positive
Viewer COPIC10652025negative
Lay personOPIC10611997positive
Lay personOPIC10621943positive
Lay personOPIC10632043negative
Lay personOPIC10641994positive
Lay personOPIC10652025negative

Oxazepam

GroupOperatorsNegativeLow Negativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcutoff)Near CutoffPositive byGC/MS(Between thecutoff and+50%)High Positiveby GC/MS(greater than+50%)
Viewer APositive0021721
Negative10101820
Viewer BPositive0031821
Negative10101710
Viewer CPositive0021621
Negative10101830
LayPersonPositive0031521
Negative10101740

Discordant table:

ViewerSample numberGC/MS resultviewer results
Viewer ABZOC1062291positive
Viewer ABZOC1064296positive
Viewer ABZOC1093358negative
Viewer ABZOC1095344negative
Viewer BBZOC1062291positive
Viewer BBZOC1063312negative
Viewer BBZOC1064296positive
Viewer BBZOC1065288positive
Viewer CBZOC1063312negative
Viewer CBZOC1064296positive
Viewer CBZOC1065288positive
Viewer CBZOC1092357negative
Viewer CBZOC1093358negative
Viewer CBZOC1095344negative
Lay PersonBZOC1062291positive
Lay personBZOC1064296positive
Lay personBZOC1065288positive

は、

:

{8}------------------------------------------------

Lay PersonBZOC1063312negative
Lay PersonBZOC1092357negative
Lay PersonBZOC1093358negative
Lay PersonBZOC1095344negative

Lay-user study

A lay user study was performed at three intended user sites with 260 lay persons. Participants in the study were 79 males and 47 females tested the Morphine samples, 74 males and 60 females tested the Oxazepam samples. They had diverse educational and professional backgrounds and ranged in age from 21 to >50. Urine samples' were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, I blind labeled samples and a device. The results are summarized below.

Cup formatNumber ofOTC user%Agreement
DrugConcentrationsamplesNegativePositiveWith GC/MS
Drug -free-100%20200100%
-75%20200100%
-50%20200100%
-25%2018290%
Morphine+25%2031785%
+50%20020100%
+75%20020100%
-75%20200100%
-50%20200100%
-25%2017385%
Oxazepam+25%2031785%
+50%20020100%
+75%20020100%
    1. Clinical Studies
      Not applicable
  • ll. Conclusion
    Based on the test principle and performance characteristics of the device, it's concluded that CR3 Keyless Split Sample Cup Morphine-Oxazepam is substantially equivalent to the predicate.

11

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized human figure with outstretched arms, representing care and protection, above three horizontal lines that symbolize the department's various services and programs.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

April 8, 2014

GUANGZHOU WONDFO BIOTECH CO., LTD. C/O JOE SHIA 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20878

Re: K140089

Trade/Device Name: CR3 Keyless Split Sample Cup Morphine-Oxazepam Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, JXM Dated: January 06, 2014 Received: January 14, 2014

Dear Mr. Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{10}------------------------------------------------

Page 2-Mr. Shia

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K140089

Device Name

CR3 Keyless Split Sample Cup Morphine-Oxazepan

Indications for Use (Describe)

CR3 Keyless Split Sample Cup Morphine-Oxazepan is a rapid test for the qualitative detection of Morphine (a drug in the opiate class) and Oxazepam(a drug in the benzodinzepine class) in human urine at a cutoff concentration of 2000ng/mL and 300ng/mL. respectively.

The tests may vield preliminary positive results even when prescription drugs including Morphine and Oxazepam are ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of there are no uniformly recognized cutoff concentration levels for morphine and oxazepan in urinc. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confimed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Avis T. Danishefsky -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response. Including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@Ida.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).