Lux HD 35 Detector (Lux HD 35); Lux HD 43 Detector (Lux HD 43)
Applicant
Iray Imaging Technology (Haining) Limited
Product Code
MQB · Radiology
Decision Date
Mar 18, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.
Device Story
Lux HD 35 and Lux HD 43 are digital flat panel detectors for general-purpose radiography. They utilize a direct-deposited CsI scintillator to convert X-rays into visible photons, which are then transformed into electron signals by a TFT/diode capacitor array. The device automatically detects X-ray exposure via an internal AED trigger module. Data is transmitted to a PC via wired (Gigabit Ethernet) or wireless (IEEE 802.11a/b/g/n/ac/ax) interfaces. The system is operated by healthcare professionals in clinical settings. The acquisition software platform manages equipment, image acquisition, and processing, outputting digital images for diagnostic review. By replacing film/screen systems, the device enables high-resolution digital imaging, facilitating faster diagnosis of disease or injury.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety, EMC, wireless coexistence, biological evaluation of patient-contact materials, and cybersecurity testing. Performance characteristics (spatial resolution, DQE) were compared to the predicate.
Technological Characteristics
Flat panel detector with direct-deposited CsI scintillator on TFT array. Dimensions: 384x460x15mm (Lux HD 35) and 460x460x15mm (Lux HD 43). Connectivity: Gigabit Ethernet and IEEE 802.11a/b/g/n/ac/ax. IP67 ingress protection. Carbon fiber frame. 16-bit ADC. Power: Max 30W. Air cooling. Software: Digital Radiographic Imaging Acquisition Software Platform.
Indications for Use
Indicated for digital radiographic diagnosis of human anatomy in adult and pediatric patients. Not intended for mammography or dental applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K252911 — Lux HD 2530 detector (Lux HD 2530) · Iray Imaging Technology (Haining) Limited · Oct 9, 2025
K201043 — Wireless Digital Flat Panel Detector (Model: Mars1717V-VSI, Mano4343W) · Iray Technology Taicang , Ltd. · May 12, 2020
K221345 — Wireless Digital Flat Panel Detector · Iray Technology Taicang , Ltd. · Jun 16, 2022
K182551 — Wireless Digital Flat Panel Detector · Iray Technology Taicang , Ltd. · Oct 17, 2018
K201932 — X-ray Flat Panel Detectors (CareView 1800Cwe /CareView 1500Cwe) · Careray Digital Medical System Co., Ltd. · Aug 7, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
March 18, 2025
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the name "U.S. Food & Drug Administration" written out next to it.
iRay Imaging Technology (Haining) Limited % Junjie Oian Registration & Regulatory Affairs Engineer No. 2, Caohejing Rd. Haining 314499 JIAXING, ZHEJIANG CHINA
Re: K243556
Trade/Device Name: Lux HD 35 Detector (Lux HD 35); Lux HD 43 Detector (Lux HD 43) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: August 20, 2024 Received: December 31, 2024
Dear Junjie Qian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by for Gabriela M. Rodal -S Gabriela M. Rodal -5/7
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K243556
Device Name
Lux HD 35 Detector (Lux HD 35); Lux HD 43 Detector (Lux HD 43)
Indications for Use (Describe)
Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# iRay Imaging Technology ( Haining ) Limited [510(k)] Application
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
- 1. Date Prepared [21 CFR 807.92(a)(1)]
August 13, 2024
# 2. Submitter's Information [21 CFR 807.92(a)(1)]
| Company Name: | iRay Imaging Technology (Haining) Limited |
|------------------|----------------------------------------------------------------|
| Company Address: | No. 2, Caohejing RD., Haining 314499, Jiaxing, Zhejiang, China |
| Contact Person: | Junjie Qian |
| Phone: | 0573-87399739 |
| Email: | Junjie.qian@iraygroup.com |
# 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name: | Lux HD 35 Detector (Lux HD 35 |
|----------------------|-------------------------------|
| | Lux HD 43 Detector (Lux HD 43 |
| Common Name: | Solid State X-Ray Imager |
| Model Name: | Lux HD 35 |
| | Lux HD 43 |
| Classification Name: | Stationary X-Ray System |
| Product Code: | MQB |
| Regulation Number: | 21 CFR 892.1680 |
| Device Class: | Class II |
{5}------------------------------------------------
## 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)[
The identification predicates within this submission are as follows:
| Manufacturer: | Carestream Health, Inc |
|----------------------|-------------------------|
| Trade Name: | Focus HD 43 Detector |
| Model Name: | Focus HD 43 |
| Product Code: | MQB |
| Classification Name: | Stationary X-Ray System |
| Regulation Number: | 21 CFR 892.1680 |
| Device Class: | Class II |
| FDA 510 (k) #: | K213529 |
### 5. Description of the Device [21 CFR 807.92(a)(4)]
Lux HD 35 Detector and Lux HD 43 Detector are digital flat panel detector. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 35cm×43cm (Lux HD 35 Detector)/42.67cm × 42.67cm (Lux HD 43 Detector) .The differences between two models are overall change in the dimensions of the image receptor
The sensor plate of Lux HD 35 Detector and Lux HD 43 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.
The major function of the Lux HD 35 Detector and Lux HD 43 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. Both kinds of detectors are the key component of DR system.
# 2 of 12
{6}------------------------------------------------
The Digital Radiographic Imaging Acquisition Software Platform - DR is part of the system, it is used to acquire, enhance, view image from Lux HD 35 Detector and Lux HD 43 Detector. Based on the risks and intended use, documentation level of the software is basic
# 6. Intended Use [21 CFR 807.92(a)(5)]
## 6.1. Indications for use
Lux HD 35 Detector and Lux HD 43 Detector are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 35 Detector and Lux HD 43 Detector are not intended for mammography or dental applications.
## 6.2. Suitable patient
Lux HD 35 Detector and Lux HD 43 Detector are suitable for providing digital Xray imaging for DR system to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients, but not intended for mammography or dental applications. The remaining notes depend on the DR system.
## 6.3. Processing of input and output
When Lux HD 35 Detector and Lux HD 43 Detector work continuously, it can automatically distinguish X-ray and output an imaging for diagnosis of disease, injury, or of any applicable health problem.
{7}------------------------------------------------
| Item | Predicate Device: | Proposed Device: | Reference device |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Focus HD 43 Detector | Lux HD 35 Detector | Mars1417V-PSI |
| | | Lux HD 43 Detector | |
| 510(K)<br>Number | K213529 | K243556 | K161730 |
| Intended Use | The Focus HD 43 Detector is<br>indicated for digital imaging<br>solution designed for<br>providing general radiographic<br>system in all general-purpose<br>diagnostic procedures. | Same | Same |
| Indications<br>for Use | Focus HD 43 Detector is<br>indicated for digital imaging<br>solutions designed to provide<br>general radiographic diagnosis<br>for human anatomy including<br>both adult and pediatric<br>patients. It is intended to<br>replace film/screen systems in<br>all general-purpose diagnostic<br>procedures. The device is not<br>intended for mammography or<br>dental applications. | Lux HD 35 Detector and<br>Lux HD 43 Detector are<br>indicated for digital imaging<br>solutions designed to<br>provide general radiographic<br>diagnosis for human<br>anatomy including both<br>adult and pediatric patients.<br>They are intended to replace<br>film/screen systems in all<br>general-purpose diagnostic<br>procedures.<br>The devices are not intended<br>for mammography or dental<br>applications. | Mars1417V-PSI<br>Wireless Digital Flat<br>Panel Detector is<br>indicated for digital<br>imaging solution<br>designed for providing<br>general radiographic<br>diagnosis of human<br>anatomy.<br>It is intended to<br>replace radiographic<br>film/screen systems in<br>all general-purpose<br>diagnostic procedures.<br>This device is not<br>intended for<br>mammography or<br>dental applications. |
# 7. Technological Characteristic [21 CFR 807.92(a)(6)]
{8}------------------------------------------------
#### Predicate Device: Proposed Device: Reference device Item Lux HD 35 Detector Focus HD 43 Detector Mars1417V-PSI Lux HD 43 Detector Classification Stationary X-ray system Same Same Name Product Code MQB Same Same Regulation 21 CFR 892.1680 Same Same Number Panel Radiology Same Same Classification II Same Same X-Ray Gd2O2S CsI CSI Absorber (Scintillator) Installation Wireless, Portable Wireless or Wired Wireless, Portable Type Readout Thin Film Transistor Same Same Mechanism Lux HD 35 Detector: 2304 × 2800 pixels Image Matrix 4267 × 4267 pixels 3500 × 4300pixels Size Lux HD 43 Detector: 4267 × 4267 pixels Pixel Size 100um 100um 150um ADC 16 bit 16 bit 14 bit Digitization Effective Lux HD 35 Detector: 426.7 mm × 426.7mm 355 mm × 434mm Imaging Area 350mm×430mm Lux HD 43 Detector: 426.7 mm × 426.7mm
#### [510(k)] Application iRay Imaging Technology (Haining)Limited
{9}------------------------------------------------
| Item | Predicate Device:<br>Focus HD 43 Detector | Proposed Device:<br>Lux HD 35 Detector<br>Lux HD 43 Detector | Reference device<br>Mars1417V-PSI |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Spatial<br>Resolution | 5.0 lp/mm | 5.0 lp/mm | 3.4 lp/mm |
| Detective<br>Quantum<br>Efficiency<br>(DQE) | 0.54 at 1 lp/mm (RQA5,<br>2.5µGy) | 0.54 at 1 lp/mm (RQA5,<br>2.5µGy) | 0.27 at 0.5 lp/mm<br>(RQA5, 3.2µGy) |
| Power<br>Consumption | Max. 42W | Lux HD 35 Detector:<br>Max. 30W<br>Lux HD 43 Detector:<br>Max. 30W | Max. 13W |
| Communicati<br>ons | a) Wired (only for service) : Gigabit<br>Ethernet (1000BASE-T)<br>b) Wireless: IEEE<br>802.11a/b/g/n/ac (2.4 GHz / 5 GHz) | a) Wired: Gigabit Ethernet (1000BASE-T)<br>b) Wireless: IEEE<br>802.11a/b/g/n/ac/ax (2.4 GHz / 5 GHz) | a) Wired: Gigabit<br>Ethernet<br>(1000BASE-T)<br>b) Wireless: IEEE<br>802.11a/b/g/n (2.4<br>GHz / 5 GHz) |
| Imaging<br>protect Plate | Carbon Fiber Plate | Same | Same |
| Cooling | Air cooling | Same | Same |
| Dimensions | 460 mm × 460 mm × 15mm | Lux HD 35 Detector<br>384mm× 460 mm × 15mm<br>Lux HD 43 Detector<br>460mm× 460 mm × 15mm | 384 mm × 460 mm ×<br>15mm |
| Detector IP<br>grade | IP56 | IP67 | / |
| Item | Predicate Device:<br>Focus HD 43 Detector | Proposed Device:<br>Lux HD 35 Detector<br>Lux HD 43 Detector | Reference device<br>Mars1417V-PSI |
| Power input<br>port | 10 pin port | 10 pin port | 4 Pin port |
| Frame<br>material | Aluminum alloy | Carbon fiber | Aluminum alloy |
| Surface<br>pressure | Uniform load: 300 kg over<br>the whole area of the surface;<br>Local load: 100 kg on an area<br>4 cm diameter of center | Uniform load: 300 kg over<br>the whole area of the<br>surface;<br>Local load: 150 kg on an<br>area 4 cm diameter of center | / |
| Operation | Temperature: +5 ~ +35°C<br>Humidity: 5 ~ 90%<br>(Non-Condensing)<br>Atmospheric pressure: 70 ~<br>106 kPa<br>Altitude: Max. 3000 meters | Temperature: +5 ~ +35°C<br>Humidity: 5 ~ 90%<br>(Non-Condensing)<br>Atmospheric pressure: 70 ~<br>106 kPa<br>Altitude: Max. 3000 meters | Temperature: +5 ~<br>+35°C<br>Humidity: 30 ~ 75%<br>(Non-Condensing)<br>Atmospheric pressure:<br>70 ~ 106 kPa<br>Altitude: Max. 3000<br>meters |
| Storage and<br>Transportation:<br>( detector ) | Temperature: -20 ~ +55°C<br>Humidity: 5 ~ 95%<br>(Non-Condensing)<br>Atmospheric pressure: 70 ~<br>106 kPa<br>Altitude: Max. 3000 meters | Temperature: -10 to + 60°C.<br>(Excluding battery)<br>Relative Humidity: 5 to<br>95%, non-condensing.<br>Atmosphere: 700mBar<br>~1060mBar<br>Altitude: Max. 3000 meter | Temperature: -20 ~<br>+55°C<br>Humidity: 10 ~ 90%<br>(Non-Condensing)<br>Atmospheric pressure:<br>70 ~ 106 kPa<br>Altitude: Max. 3000<br>meters |
| Item | Predicate Device:<br>Focus HD 43 Detector | Proposed Device:<br>Lux HD 35 Detector<br>Lux HD 43 Detector | Reference device<br>Mars1417V-PSI |
| Software | SDK(include iDetector)<br>The software is intend to<br>supply API interface for DR<br>system manufacturers. DR<br>system manufacturer control<br>the detector by SDK interface.<br>SDK is not intend to use<br>directly by other users beside<br>DR system manufacturers. | Digital Radiographic<br>Imaging Acquisition<br>Software Platform – DR<br>The software used for<br>getting Digital X-ray<br>radiography images from<br>the flat panel detectors.<br>The software is used to<br>handle the DICOM protocol<br>(DICOM 3.0).<br>The software is responsible<br>for the DR equipment<br>management, acquisition<br>and processing functions, to<br>provide patient registration,<br>scanning, image processing<br>and other functions. | iRayDR<br>The software used for<br>getting Digital X-ray<br>radiography images<br>from the flat panel<br>detectors.<br>The software is used<br>to handle the DICOM<br>protocol (DICOM<br>3.0).<br>The software is<br>responsible for the DR<br>equipment<br>management,<br>acquisition and<br>processing functions,<br>to provide patient |
| | | | registration, scanning,<br>image processing,<br>image forwarding,<br>image printing and<br>other functions. |
| Item | Predicate Device: | Proposed Device: | Reference device |
| | Focus HD 43 Detector | Lux HD 35 Detector<br>Lux HD 43 Detector | Mars1417V-PSI |
| Utilized FDA<br>guidance<br>documents | 1. Guidance for the<br>Submission of 510(k)'s for<br>Solid State X-ray Imaging<br>Devices; | Same | / |
| | 2. The 510(k) Program:<br>Evaluating Substantial<br>Equivalence in Premarket<br>Notifications[510(k)]; | Additionally, the following<br>FDA guidance documents<br>are applied<br><br>1. Content of Premarket<br>Submissions for Device<br>Software Functions<br>Guidance for Industry<br>and Food and Drug<br>Administration Staff<br>June 2023<br><br>2. Pediatric Information<br>for X-ray Imaging<br>Device Premarket<br>Notifications Guidance<br>for Industry and Food<br>and Drug<br>Administration Staff<br>November 2017<br><br>3. Cybersecurity in<br>Medical Devices:<br>Quality System<br>Considerations and<br>Content of Premarket | |
| | 3. Content of Premarket<br>Submissions for<br>Management of<br>Cybersecurity in Medical<br>Devices; | | |
| | 4. Radio Frequency Wireless<br>Technology in Medical<br>Devices. | | |
| | 5. Guidance for “Premarket<br>Assessment of Pediatric<br>Medical Devices”; | | |
| | 6. Guidance for “Pediatric<br>Information for X-ray<br>Imaging Device Premarket<br>Notifications”. | | |
| | 7. Design Control Guidance<br>For Medical Device<br>Manufacturers | | |
| Item | Predicate Device: | Proposed Device: | Reference device |
| | Focus HD 43 Detector | Lux HD 35 Detector | Mars1417V-PSI |
| | | Lux HD 43 Detector | |
| 8. | Appropriate Use of<br>Voluntary Consensus<br>Standards in Premarket<br>Submissions for Medical<br>Devices. | Submissions Guidance<br>for Industry and Food<br>and Drug<br>Administration Staff<br>September 2023 | |
| 9. | The Special 510(k)<br>Program | | |
#### [510(k)] Application iRay Imaging Technology ( Haining ) Limited
{10}------------------------------------------------
# iRay Imaging Technology ( Haining ) Limited [510(k)] Application
{11}------------------------------------------------
#### [510(k)] Application iRay Imaging Technology ( Haining ) Limited
{12}------------------------------------------------
#### iRay Imaging Technology ( Haining ) Limited [510(k)] Application
{13}------------------------------------------------
# iRay Imaging Technology ( Haining ) Limited [510(k)] Application
## 8. System requirements to operate with other radiographic system components
- 1) Recommended Generator Specification:
Energy range: 40~150kVp
mA range: 10~1000mA (depending on the generator power)
ms range: 10~6300ms to produce 0.1~1000mAs (depending on the generator power) Note: To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have any questions regarding the compatibility issue for other generators, please contact the distributor or manufacturer's service office.
- 2) Minimum configuration: Lux HD 43 Detector and Lux HD 35 Detector connected via wireless and wired communication.
| Operating System: | Windows 11 |
|-------------------|-----------------------------------|
| CPU: | Intel Core i3- 8100 3.6GHz 4C 65W |
| Memory: | 16GB (2x8GB) DDR4 2666 DIMM |
| Hard Disk: | 1TB |
# 10 of 12
{14}------------------------------------------------
- 3) X-ray exposure mode
The AED trigger module is a unit can connect X-ray signal in the Lux HD 35 Detector and Lux HD 43 Detector. Once there is X-ray generator exposure exist, the AED trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image.
### 9. Non-clinical study
- Electrical Safety and EMC testing: 1)
Electrical, mechanical, environmental safety according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC 60601-1-2. All test results are meet the standard requirements.
- 2) Biological Evaluation:
The materials of the detector which contact operators' or patients' skin have been evaluated with the FDA guidance "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". And the evaluation result assured the biological safety.
- 3) Non-clinical Considerations:
The non-clinical studies have been performed and the results have shown that sections of the non-clinical consideration mentioned in the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' are substantially equivalent to the non-clinical consideration of predicate device on the Market (Focus HD 43 detector, K213529).
- 4) Clinical Consideration:
Intended use, fundamental scientific technology, regulatory requirement, non-clinical performance, labeling, quality-assurance program keep the same with those of predicate device. Software keep the same with that of reference device except "image forwarding, image printing" function and literal name. There is no any negative change about clinical performance from predicate device.
# 11 of 12
{15}------------------------------------------------
- 5) Wireless testing
Wireless functionality and wireless coexistence testing in accordance with ANSI IEEE C63.27-2017 was performed. All test results are meet the standard requirements.
- 6) Cybersecurity testing
Cybersecurity threat modeling, risk assessment, and controls and testing were performed to comply with requirements specified in section 524B(b)(2) of the Federal Food, Drug, and Cosmetics Act to provide a reasonable assurance that the subject device with its wireless capabilities are cybersecure.
# 10. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, applicant concludes that Lux HD 35 Detector and Lux HD 43 Detector are substantially equivalent to predicate device with regards to safety and effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.