KMC KYPHOPLASTY SYSTEM

K113742 · Shanghai Kinetic Medical Co., Ltd. · NDN · Sep 17, 2012 · Orthopedic

Device Facts

Record IDK113742
Device NameKMC KYPHOPLASTY SYSTEM
ApplicantShanghai Kinetic Medical Co., Ltd.
Product CodeNDN · Orthopedic
Decision DateSep 17, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3027
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The KMC Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.

Device Story

KMC Kyphoplasty System is a surgical instrument set for percutaneous vertebral augmentation; used by physicians in clinical settings. Components include balloon catheter, puncture needle, syringe pump, and multifunctional tool. System creates void in spinal cancellous bone via balloon inflation; facilitates subsequent delivery of cleared PMMA bone cement for fracture reduction and fixation. Device is single-use and ETO-sterilized. Clinical benefit derived from restoring vertebral height and stabilizing fractures. Healthcare providers use system under fluoroscopic guidance to position balloon, inflate to create cavity, and prepare site for cement injection.

Clinical Evidence

Multicenter study of 25 patients at two sites. Results demonstrated the system performed per specification when used according to instructions for use.

Technological Characteristics

System components: balloon catheter, puncture needle, syringe pump, multifunctional tool. Materials: biocompatible, previously cleared for similar use. Sterilization: ETO gas (SAL 10^-6). Packaging: PETG/PET tray with Tyvek lid. Mechanical operation: balloon inflation for bone void creation. No software or electronic components.

Indications for Use

Indicated for patients requiring reduction and fixation of spinal fractures or creation of a void in cancellous bone during percutaneous vertebral augmentation. Used with cleared spinal PMMA bone cements.

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113742: PAGE 1 OF 3 SEP 1 7 2012 # 510(k) Summary General Information Classification Regulation Number 21CFR 888.1100 21CFR 888.3027 Arthroscope Class II Regulation Name Product Code HRX NDN Trade Name Submitter Submitted by KMC Kyphoplasty System Cement/Bone Vertebroplasty Shanghai Kinetic Medical Co., Ltd Regulatory Strategies, Inc. 3924 Cascade Beach Road Lutsen, MN 55612 Tel: (218) 387-1559 Gregory Mathison Contact Date Prepared April 20, 2011 President #### Indications for Use The KMC Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty. ## Predicate Devices | K061222 | Kyphx Inflatable Bone Tamp | Kyphon Inc. | |---------|----------------------------|-------------| | K103231 | Kyphon Inflation Syringe | Kyphon Inc | {1}------------------------------------------------ # Device Description KMC Kyphoplasty System consists of five components: - Balloon Catheter . - Tool Kit - . Puncture Needle - Balloon Catheter Syringe Pump . - Kyphoplasty Multifunctional Tool . The Kyphoplasty Multifunctional Tool can be substituted for the Puncture Needle and Kyphoplasty Tool Kit. ## Sterilization The system is provided sterile and is for single use only. The ETO gas sterilization process is validated with a resulting sterility assurance level (SAL) of 10°. K13742 AGE 2 OF 3 ETO residual testing was also performed and the products met specification. ## Packaging The components are placed in a thermoformed PETG/PET tray with a heat sealed Tyvek lid. The sealed tray is placed in a Tyvek flexible peel pouch and heat sealed. The heat sealed pouches are placed in white shelf cartons and then packaged in a corrugated shipper box. Expiration dating testing was conducted using the industry standard ASTM method for accelerated aging and real-time aging was also performed. #### Materials All materials used in the manufacture of the KMC Kyphoplasty System are suitable for this use and have been used in s previously cleared products. ## Testing Product testing was performed on final sterilized devices. Testing was completed, including: dimensional, inflation/deflation, balloon size and system compatibility. All testing met the acceptance criteria. Biocompatibility testing was performed per ISO10993 as required for a temporary bone/tissue contacting device. All materials were found to be biocompatible and suitable for this use. #### Clinical Evaluation The KMC Kyphoplasty system was evaluated in a multicenter study. A total of 25 patients were treated at two sites. The conclusions from the study show the system performed per specification when used according to the instructions for use. {2}------------------------------------------------ K113742 PAGE 3 OF 3 . · # Summary of Substantial Equivalence The KMC Kyphoplasty System is equivalent to the features of the predicate products. The indications for use, basic overall function, methods of manufacturing, and materials used are substantially equivalent. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 17 2012 Shanghai Kinetic Medical Company, Limited % Regulatory Strategies, Incorporated Mr. Gregory Mathison President 3924 Cascade Beach Road Lutsen, Minnesota 55612 Re: K113742 Trade/Device Name: KMC Kyphoplasty System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: August 20, 2012 Received: August 29, 2012 Dear Mr. Mathison: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {4}------------------------------------------------ # Page 2 – Mr. Gregory Mathison CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, For Deb Lin Din Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K113742 Device Name: KMC Kyphoplasty System Indications for Use: The KMC Kyphoplasty System is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty. Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) thumor M. Taylor (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113742 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...