K Number
K192449

Validate with FDA (Live)

Date Cleared
2020-05-27

(264 days)

Product Code
Regulation Number
888.3027
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Joline Kyphoplasty System Allevo is intended to be used for the reduction of fixation of fractures and or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacy late (PMMA) bone cements).

Device Description

The Joline® Kyphoplasty System Allevo is a sterile, single-use system of instruments that facilitates balloon kyphoplasty. The subject device includes both single and double balloon catheter options that allows the introduction of one, or two balloons, into each pedicle at a time, for a total of two or four balloons per vertebrae. The balloons can be expanded or deflated separately using contrast medium, a balloon adapter, and an inflation device. The Joline® Kyphoplasty System Allevo includes instruments and accessories that facilitate the kyphoplasty procedure, including devices and curettes for creation of access channels within the bone. Subsequent to creating the cavities within the vertebral bone, the vertebral body is stabilized with an FDA cleared PMMA bone cement that cures within a few minutes. The Joline Kyphoplasty System Allevo is offered in kits that are comprised of various combinations of instruments that address surgeon preference and specific clinical condition of the patient.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Test Performed)RequirementReported Device Performance (Test Result and Conclusion)
Balloon Burst Pressure$\ge$ 27 bar ($\approx$ 400 psi)Passed all conditions; Equivalent to predicate
Balloon Burst Volume$\ge$ 3 mlPassed all conditions; Equivalent to predicate
Balloon Inflation Behavior (Average Diameter at 6ml inflation)Average diameter = 16 mm (at 6ml inflation)Passed all conditions; Equivalent to predicate
Balloon Inflation Behavior (Diameter constraint)Balloon Diameter $\le$ 19 mmPassed all conditions; Equivalent to predicate
Balloon Inflation Behavior (Length Change)Least possible length changePassed all conditions; Equivalent to predicate
Tensile Force Balloon-Shaft$\ge$ 15 NPassed all conditions; Equivalent to predicate
Tensile Force Hub-Shaft$\ge$ 15 NPassed all conditions; Equivalent to predicate
Tensile Force Luer Lock$\ge$ 15 NPassed all conditions; Equivalent to predicate
Balloon Repeated InflationMust not burst within 3 inflation/deflation cyclesNot explicitly reported in the "Test Result and Conclusion" column in the smaller table, but the general statement "The results of the non-clinical tests show that the inflatable bone tamp of the Joline Kyphoplasty System Allevo meets or exceeded all performance requirements" implies it passed.
Balloon Deflation Time$\le$ 3 sNot explicitly reported in the "Test Result and Conclusion" column in the smaller table, but the general statement "The results of the non-clinical tests show that the inflatable bone tamp of the Joline Kyphoplasty System Allevo meets or exceeded all performance requirements" implies it passed.

2. Sample size used for the test set and the data provenance

The document does not specify the sample size used for the tests. It refers to "non-clinical tests," which are typically laboratory bench tests rather than tests performed on human subjects. Therefore, provenance regarding country of origin or retrospective/prospective data is not applicable in the human subject sense. These are engineering design verification tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable. The tests performed are non-clinical, evaluating the physical and mechanical properties of the device against predefined engineering specifications. Ground truth in the context of expert consensus or medical diagnosis is not relevant here.

4. Adjudication method for the test set

This information is not applicable as the tests are non-clinical engineering tests, not involving human interpretation or adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC study or any AI assistance. This document describes a medical device (a kyphoplasty system) that is a physical instrument, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical kyphoplasty system, not an algorithm.

7. The type of ground truth used

The "ground truth" for these tests is the established engineering specifications and performance requirements for a kyphoplasty balloon system, derived from industry standards, predicate device characteristics, and safety considerations. These are objective measurements of physical properties.

8. The sample size for the training set

This is not applicable. There is no concept of a "training set" as this is not a machine learning or AI-based device.

9. How the ground truth for the training set was established

This is not applicable for the same reasons as point 8.

{0}------------------------------------------------

May 27, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION".

Joline GmbH & Co. KG % Robert Poggie President BioVera Inc. 65 Promenade Saint Louis Notre-Dame-del-L'Ile-Perrot, CA J7V 7P2 QC

Re: K192449

Trade/Device Name: Joline® Kyphoplasty System Allevo Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope; Cement, Bone Vertebroplasty Regulatory Class: Class II Product Code: HRX, NDN Dated: April 24, 2020 Received: April 27, 2020

Dear Dr. Poggie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K192449

Device Name

Joline® Kyphoplasty System Allevo

Indications for Use (Describe)

The Joline Kyphoplasty System Allevo is intended to be used for the reduction of fixation of fractures and or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacy late (PMMA) bone cements).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for the Joline® Kyphoplasty System Allevo

In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of the Joline® Kyphoplasty System Allevo.

A. SUBMITTERS INFORMATION

Submitter Name:BioVera, Inc.
Submitter Address:65 Promenade Saint-Louis, Notre-Dame-De-L'lle-Perrot,Quebec, J7V 7P2, CANADA
Contact Person:Robert A Poggie, PhD
Phone Number:(514) 901-0796
Fax Number:(514) 901-0796
Date of Submission:April 24, 2020

B. DEVICE IDENTIFICATION & MANUFACTURER

Manufacturer Name:Joline GmbH & Co. KG
Manufacturer Address:Neue Rottenburger Strasse 50, 72379 Hechingen, Germany
Registration Number:3004734264
Contact Name:Peter Kohlbecher
Title:Director Sales & Marketing
Device Trade Name:Joline® Kyphoplasty System Allevo
Device Common Name:Vertebroplasty System; Inflatable Bone Tamp
Classification Names:Arthroscope; Cement, Bone, Vertebroplasty
Classification Codes:NDN and HRX
Classification Panel:Orthopedic
Regulation Number:21 CFR sections 888.1100 and 888.3027

C1. PRIMARY PREDICATE DEVICE

K150607 Osseoflex SB

C2. REFERENCE DEVICE K140937

Osseoflex SB

{4}------------------------------------------------

D. DEVICE DESCRIPTION

The Joline® Kyphoplasty System Allevo is a sterile, single-use system of instruments that facilitates balloon kyphoplasty. The subject device includes both single and double balloon catheter options that allows the introduction of one, or two balloons, into each pedicle at a time, for a total of two or four balloons per vertebrae. The balloons can be expanded or deflated separately using contrast medium, a balloon adapter, and an inflation device. The Joline® Kyphoplasty System Allevo includes instruments and accessories that facilitate the kyphoplasty procedure, including devices and curettes for creation of access channels within the bone. Subsequent to creating the cavities within the vertebral bone, the vertebral body is stabilized with an FDA cleared PMMA bone cement that cures within a few minutes. The Joline Kyphoplasty System Allevo is offered in kits that are comprised of various combinations of instruments that address surgeon preference and specific clinical condition of the patient.

E. INDICATIONS FOR USE

The Joline® Kyphoplasty System Allevo is intended to be used for the reduction and fixation of fractures and/or creation of a void in cancellous bone in the spine during balloon kyphoplasty (for use with cleared spinal polymethylmethacrylate (PMMA) bone cements).

F. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The Joline® Kyphoplasty System Allevo has similar device characteristics and clinical indications for use as the predicate device. The table below compares the technical and performance characteristics of the subject, reference, and predicate devices. The subject and predicate devices are comprised of instruments that facilitate balloon kyphoplasty in the spine, including balloons for creating cavithin bone, cannula to transfer saline, means to deliver FDA cleared bone cement, means to create a channel of access within the bone, and are single use and provided sterile.

PerformanceCharacteristicRequirement(Joline, Single Balloon 22/16)Requirement(Predicate Osseoflex)Test Result andConclusion
Balloon BurstPressure and Volume(Constrained)Burst Pressure ≥ 27 bar(≈ 400 psi)Burst Volume ≥ 6 mlBurst Pressure ≥ 27 atm(≈ 400 psi)Burst Volume ≥ 4 mlPassed all conditionsEquivalent to predicate
Tensile ForceBalloon - ShaftDetachment Force≥ 15 NBond tensile strength≥ 15 NPassed all conditionsEquivalent to predicate
Tensile ForceHub – ShaftDetachment Force≥ 15 NBond tensile strength≥ 15 NPassed all conditionsEquivalent to predicate
Tensile ForceLuer LockDetachment Force≥ 15 NBond tensile strength≥ 15 NPassed all conditionsEquivalent to predicate
Balloon InflationBehavior(UnconstrainedBalloon Compliance)Average Diameter16 mmLength Change: least possibleBalloon Diameter ≤ 19 mmBalloon Diameter≤ 19 mmBalloon working length≤ 24 mmPassed all conditionsEquivalent to predicate

{5}------------------------------------------------

G. PERFORMANCE DATA

The following tests were performed in support of the substantial equivalence determination.

Test PerformedAcceptance Criteria
Balloon Burst PressureBurst pressure (pburst) ≥ 27 bar (~ 400 psi)
Balloon Burst VolumeBurst volume (Vburst) ≥ 3 ml
Balloon Inflation BehaviorInflated with 6 ml, the average diameter of the balloon should be 16 mm
Tensile Force Balloon-ShaftMinimum tensile strength (Fballoon-shaft) ≥ 15 N
Tensile Force Hub-ShaftMinimum tensile strength (Fhub-shaft) ≥ 15 N
Balloon Repeated InflationThe balloon must not burst within 3 inflation/deflation cycles
Balloon Deflation TimeDeflation time (tdef) ≤ 3 s

The results of the non-clinical tests show that the inflatable bone tamp of the Joline Kyphoplasty System Allevo meets or exceeded all performance requirements.

H. CONCLUSION

The Joline® Kyphoplasty System Allevo is substantially equivalent to the identified predicate device based on the indications for use, materials, design, and performance data presented in this 510(k) notification.

§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.

(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”