KYPHON INFLATION SYRINGE
K103231 · Medtronic Spine, LLC · HRX · Jan 24, 2011 · Orthopedic
Device Facts
| Record ID | K103231 |
| Device Name | KYPHON INFLATION SYRINGE |
| Applicant | Medtronic Spine, LLC |
| Product Code | HRX · Orthopedic |
| Decision Date | Jan 24, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.1100 |
| Device Class | Class 2 |
Intended Use
The Kyphon® Inflation Syringe is intended to be used to inflate and deflate inflatable devices (including Inflatable Bone Tamps) and to measure the pressure within the inflatable device during the procedure.
Device Story
Kyphon® Inflation Syringe is a manual surgical instrument used to inflate/deflate inflatable devices, such as Inflatable Bone Tamps, and monitor internal pressure. Device components include a syringe barrel, threaded plunger assembly with depress-release handle, digital pressure display, outer shell, flexible high-pressure tubing, and weighted swivel-luer connector. It generates and monitors pressures up to 700 psi, with user-selectable maximums of 400 psi or 700 psi. A 30ml polycarbonate locking syringe is included for device preparation; a V-Klip™ accessory allows concurrent dual-syringe operation. Used in clinical settings by healthcare providers to facilitate orthopedic procedures. Digital pressure readout allows real-time monitoring, enabling precise control of inflation to ensure safe and effective device deployment, potentially improving patient outcomes during bone tamp procedures.
Clinical Evidence
Bench testing only. Testing included functional and mechanical performance verification, biocompatibility testing per ISO-10993, and sterilization validation. No clinical data was required or provided.
Technological Characteristics
Manual inflation device with digital pressure readout. Components: syringe barrel, threaded plunger, flexible high-pressure tubing, weighted swivel-luer. Materials: polycarbonate (locking syringe), PETG (tray). Sterilization: provided sterile, single-use. Pressure range: up to 700 psi. Connectivity: standalone digital display.
Indications for Use
Indicated for inflation and deflation of inflatable devices, including Inflatable Bone Tamps, and pressure monitoring during procedures requiring such devices.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
- Atrion Medical QL® Inflation Device (K032840)
Related Devices
- K160107 — DiamondTOUCH Inflation Device and Fluid Dispensing Syringe · Merit Medical Systems, Inc. · Jul 29, 2016
- K981251 — KYPHON INFLATABLE BONE TAMP · Kyphon, Inc. · Jul 2, 1998
- K130566 — BASIXTOUCH · Merit Medical Systems, Inc. · Jun 19, 2013
- K122321 — BASIXCONPAK ANALOG INFLATION SYRINGE · Merit Medical Ireland, Ltd. · Dec 19, 2012
- K040138 — VICEROY INFLATION DEVICE · Merit Medical Systems, Inc. · Feb 13, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
K(0323/
Traditional 510(k) Submission
Medtronic Spine LLC Page 1 of 2, Tab 5
## Tab 5 PREMARKET NOTIFICATION [510(k)] SUMMARY
| Summary Date | January 10, 2011 | JAN 24 2011 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Trade Name: | Kyphon® Inflation Syringe | |
| Common Name: | Arthroscope<br>Tamp | |
| Classification Name: | Class II, Arthroscope, 21 CFR 888.1110<br>Class I, Orthopedic Manual Surgical Instrument, 21 CFR 888.4540 | |
| Device Code: | HRX<br>HXG | |
| Manufacturer's Name: | Medtronic Spine LLC | |
| Address: | 1221 Crossman Avenue<br>Sunnyvale, CA 94089 | |
| Contact Person: | Hetal Jawahar Thakker | |
| Title: | Sr Regulatory Affairs Specialist | |
| Address: | 1221 Crossman Avenue<br>Sunnyvale, CA 94089 | |
| Phone: | (408) 548-5334 | |
| Predicate Device(s): | Atrion Medical QL® Inflation Device: K032840 (cleared on March 03, 2004) | |
| Intended Use | The Kyphon® Inflation Syringe is intended to be used to inflate and deflate<br>inflatable devices (including Inflatable Bone Tamps) and to measure the pressure<br>within the inflatable device during the procedure. | |
| Device Description: | The Kyphon® Inflation Syringe consists of a syringe barrel; a threaded plunger<br>assembly with a depress-release handle; a digital display that provides pressure<br>readout; an outer shell assembly that retains the internal components; a flexible<br>tubing designed to withstand high inflation pressures; and a weighted swivel-luer<br>for connection to the inflatable device (including Inflatable Bone Tamps).<br>The device is designed to generate and monitor pressures up to 700psi (providing<br>the user the option to be able to select the Maximum Inflation Pressure of 400psi or<br>700psi depending on the inflatable device being used). | |
{1}------------------------------------------------
| Medtronic Spine LLC<br>Page 2 of 2, Tab 5 | | Traditional 510(k) Submission<br>Kyphon® Inflation Syringe | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|--|
| The Locking Syringe (to be packaged together with the Inflation Syringe) is a 30ml<br>polycarbonate syringe and is included for preparation of the inflatable device, as<br>necessary. | | | |
| A V-Klip™ (available as part of the 2 Pack configuration) holds two Kyphon<br>Inflation Syringes for concurrent dual-syringe operation. | | | |
| Testing | Testing of the Kyphon® Inflation Syringe was completed to demonstrate that the<br>device meets the specifications and performance characteristics, and is<br>substantially equivalent to the predicate devices. The testing included functional<br>and mechanical testing. In addition, biocompatibility testing and sterilization<br>validation were completed. | | |
| Biocompatibility | Biocompatibility testing of the Kyphon® Inflation Syringe confirmed that the<br>devices meet applicable requirements of the FDA Blue Book Memorandum #G95-<br>1 entitled "Use of International Standard ISO-10993, "Biological Evaluation of<br>Medical Devices Part-1: Evaluation and Testing" and are biocompatible. | | |
| Sterilization | The Kyphon® Inflation Syringe will be provided sterile and is intended for single<br>use only. | | |
| Packaging and<br>Labeling | The components are placed in an inner and outer polyethylene terephthalate glycol<br>(PETG) Tray with a Tray lid. This configuration is sealed with a Tyvek® Lid, and<br>placed in a carton. | | |
| Substantial<br>Equivalence: | The information submitted in this pre-market notification supports a determination<br>that the Kyphon® Inflation Syringe is substantially equivalent to the predicate<br>devices, the Atrion Medical QL Inflation Device cleared under K032840. The<br>products have the same fundamental scientific technology and basic design, and<br>similar intended use and functional characteristics as the predicate Inflation<br>Syringe.<br>The results of testing demonstrate that the Kyphon® Inflation Syringe is safe and<br>effective for its intended use. | | |
:
.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized depiction of an eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic Spine, LLC % Hetal Jawahar Thakker Senior Regulatory Affairs Specialist 1211 Crossman Avenue Sunnyvale, California 94089
Re: K103231
Trade/Device Name: Kyphon® Inflation Syringe Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, HXG Dated: October 29, 2010 Received: November 1, 2010
Dear Hetal Jawahar Thakker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
JAN 2 4 250
{3}------------------------------------------------
Page 2 - Hetal Jawahar Thakker
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
g B-R
Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Tab 4
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name:
Kyphon® Inflation Syringe
Indications for Use:
The Kyphon® Inflation Syringe is intended to be used to inflate and deflate inflatable devices (including Inflatable Bone Tamps) and to measure the pressure within the inflatable device during the procedure.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
signature
(Division Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K103231